Abstract
The use of replacement heart valves continues to grow due to the increased prevalence of valvular heart disease resulting from an ageing population. Since bioprosthetic heart valves (BHVs) continue to be the preferred replacement valve, there continues to be a strong need to develop better and more reliable BHVs through and improved the general understanding of BHV failure mechanisms. The major technological hurdle for the lifespan of the BHV implant continues to be the durability of the constituent leaflet biomaterials, which if improved can lead to substantial clinical impact. In order to develop improved solutions for BHV biomaterials, it is critical to have a better understanding of the inherent biomechanical behaviors of the leaflet biomaterials, including chemical treatment technologies, the impact of repetitive mechanical loading, and the inherent failure modes. This review seeks to provide a comprehensive overview of these issues, with a focus on developing insight on the mechanisms of BHV function and failure. Additionally, this review provides a detailed summary of the computational biomechanical simulations that have been used to inform and develop a higher level of understanding of BHV tissues and their failure modes. Collectively, this information should serve as a tool not only to infer reliable and dependable prosthesis function, but also to instigate and facilitate the design of future bioprosthetic valves and clinically impact cardiology.
Similar content being viewed by others
Abbreviations
- AHA:
-
American Heart Association
- AV:
-
Aortic valve
- BHV:
-
Bioprosthetic heart valve
- BP:
-
Bovine pericardium
- ECM:
-
Extracellular matrix
- FE:
-
Finite element
- GAG:
-
Glycosaminoglycan
- GLBP:
-
Glutaraldehyde bovine pericardium
- GLUT:
-
Gluraraldehyde treatment
- LEHI:
-
Linear elastic homogeneous incompressible
- microCT:
-
Micro X-ray computed tomography
- MRI:
-
Magnetic resonance images
- MV:
-
Mitral valve
- PAV:
-
Porcine aortic valve
- PBS:
-
Phosphate buffered saline
- PD:
-
Preferred direction
- RVE:
-
Representative volume element
- SALS:
-
Small angle light scattering
- TEHV:
-
Tissue-engineered heart valve
- UTS:
-
Ultimate tensile stregth
- VEC:
-
Valvular endothelial cell
- VIC:
-
Valvular interstitial cell
- XD:
-
Cross-preferred direction
- ΔM:
-
Infinitesimal mass
- ΔV:
-
Infinitesimal volume
- ρ:
-
Mass density
- \( \varepsilon \) :
-
Fiber uniaxial strain (in structural constitutive model)
- \( \theta \) :
-
Fiber orientation angle (in structural constitutive model)
- cf :
-
Volume fraction of fibers (in structural constitutive model)
- C :
-
Left Cauchy-Green stretch tensor
- \( D(\varepsilon ) \) :
-
Fiber recruitment statistical distribution (in structural constitutive model)
- E :
-
Green–Lagrange strain tensor
- E ij :
-
Components of the Green-Lagrange strain tensor
- \( I(\theta ) \) :
-
Angular distribution of scattered light (in SALS analysis)
- \( R(\theta ) \) :
-
Fiber angular distribution (in structural constitutive model)
- S :
-
Second Piola–Kirchhoff stress tensor
- Sij :
-
Components of the second Piola–Kirchhoff stress tensor
- \( {\mathbf{S}}^{f} \) :
-
Second Piola–Kirchhoff stress tensor in the fiber (in structural constitutive model)
- \( S_{nn}^{f} \) :
-
Component of the Piola–Kirchhoff stress tensor in the fiber along fiber direction (in structural constitutive model)
- W:
-
Stored energy function
- \( w(\varepsilon ) \) :
-
Stored energy function of fiber (in structural constitutive model)
- Wf :
-
Stored energy function of the fiber ensemble (in structural constitutive model)
- Wm :
-
Stored energy function of the matrix (in structural constitutive model)
- I1, I2, I3 :
-
Principal invariants of the left Cauchy-Green stretch tensor
References
Aggarwal, A., G. Ferrari, E. Joyce, M. J. Daniels, R. Sainger, J. H. Gorman, 3rd, et al. Architectural trends in the human normal and bicuspid aortic valve leaflet and its relevance to valve disease. Ann. Biomed. Eng. 42(5):986–998, 2014. doi:10.1007/s10439-014-0973-0.
Aggarwal, A., and M. Sacks. A framework for determination of heart valves’ mechanical properties using inverse-modeling approach. In: Functional Imaging and Modeling of the Heart, edited by H. van Assen, P. Bovendeerd, and T. Delhaas. Lecture Notes in Computer Science: Springer International Publishing, 2015, pp. 285–294.
Ali, A., J. C. Halstead, F. Cafferty, L. Sharples, F. Rose, R. Coulden, et al. Are stentless valves superior to modern stented valves? A prospective randomized trial. Circulation. 114(1 Suppl):I535–I540, 2006. doi:10.1161/CIRCULATIONAHA.105.000950.
Amini, R., C. E. Eckert, K. Koomalsingh, J. McGarvey, M. Minakawa, J. H. Gorman, et al. On the in vivo deformation of the mitral valve anterior leaflet: effects of annular geometry and referential configuration. Ann. Biomed. Eng. 40(7):1455–1467, 2012. doi:10.1007/s10439-012-0524-5.
Astorino, M., J.-F. Gerbeau, O. Pantz, and K.-F. Traoré. Fluid-structure interaction and multi-body contact: application to aortic valves. Comput. Method Appl. Mech. Eng. 198:3603–3612, 2009.
Baaijens, F. P. T. A fictitious domain/mortar element method for fluid-structure interaction. Int. J. Numer. Methods Fluids 35(7):743–761, 2001.
Billiar, K. L., and M. S. Sacks. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp–Part I: Experimental results. J. Biomech. Eng. 122(1):23–30, 2000.
Billiar, K. L., and M. S. Sacks. Biaxial mechanical properties of the native and glutaraldehyde-treated aortic valve cusp: Part II–A structural constitutive model. J. Biomech. Eng. 122(4):327–335, 2000.
Black, M. M., I. C. Howard, X. Huang, and E. A. Patterson. A three-dimensional analysis of a bioprosthetic heart valve. J. Biomech. 24(9):793–801, 1991.
Blackstone, E. H. Could it happen again? The Bjork-Shiley convexo-concave heart valve story. Circulation 111(21):2717–2719, 2005.
Bogaers, A. E. J., S. Kok, B. D. Reddy, and T. Franz. Quasi-Newton methods for implicit black-box FSI coupling. Comput. Method Appl. Mech. Eng. 279:113–132, 2014.
Borazjani, I. Fluid-structure interaction, immersed boundary-finite element method simulations of bio-prosthetic heart valves. Comput. Method Appl. Mech. Eng. 257:103–116, 2013.
Borazjani, I., L. Ge, and F. Sotiropoulos. Curvilinear immersed boundary method for simulating fluid structure interaction with complex 3D rigid bodies. J. Comput. Phys. 227(16):7587–7620, 2008.
Braunwald, N. S., T. Cooper, and A. G. Morrow. Complete replacement of the mitral valve. Successful clinical application of a flexible polyurethane prosthesis. J. Thorac. Cardiovasc. Surg. 40:1–11, 1960.
Buchanan, R. M., and M. S. Sacks. Interlayer micromechanics of the aortic heart valve leaflet. Biomech. Model. Mechanobiol. 2013. doi:10.1007/s10237-013-0536-6.
Buehler, M. J. Atomistic and continuum modeling of mechanical properties of collagen: Elasticity, fracture, and self-assembly. J. Mater. Res. 21(08):1947–1961, 2006. doi:10.1557/jmr.2006.0236.
Burriesci, G., I. C. Howard, and E. A. Patterson. Influence of anisotropy on the mechanical behaviour of bioprosthetic heart valves. J. Med. Eng. Technol. 23(6):203–215, 1999.
Cacciola, G., G. W. Peters, and F. P. Baaijens. A synthetic fiber-reinforced stentless heart valve. J. Biomech. 33(6):653–658, 2000.
Cacciola, G., G. W. Peters, and P. J. Schreurs. A three-dimensional mechanical analysis of a stentless fibre-reinforced aortic valve prosthesis. J. Biomech. 33(5):521–530, 2000.
Carew, E. O., A. Garg, J. E. Barber, and I. Vesely. Stress relaxation preconditioning of porcine aortic valves. Ann. Biomed. Eng. 32(4):563–572, 2004.
Carmody, C. J., G. Burriesci, I. C. Howard, and E. A. Patterson. An approach to the simulation of fluid-structure interaction in the aortic valve. J. Biomech. 39:158–169, 2006.
Cataloglu, A., R. E. Clark, and P. L. Gould. Stress analysis of aortic-valve leaflets with smoothed geometrical data. J. Biomech. 10(3):153–158, 1977. doi:10.1016/0021-9290(77)90053-7.
Chandran, K. B. Role of computational simulations in heart valve dynamics and design of valvular prostheses. Cardiovascular engineering and technology. 1(1):18–38, 2010. doi:10.1007/s13239-010-0002-x.
Chew, G. G., I. C. Howard, and E. A. Patterson. Simulation of damage in a porcine prosthetic heart valve. J. Med. Eng. Technol. 23(5):178–189, 1999.
Chew, P. H., F. C. Yin, and S. L. Zeger. Biaxial stress-strain properties of canine pericardium. J. Mol. Cell. Cardiol. 18(6):567–578, 1986.
Choi, H. S., and R. P. Vito. Two-dimensional stress-strain relationship for canine pericardium. J. Biomech. Eng. 112(2):153–159, 1990.
Christie, G. W. Anatomy of aortic heart valve leaflets: the influence of glutaraldehyde fixation on function. Eur. J. Cardiothorac. Surg. 6:S25–S33, 1992.
Christie, G., and B. Barratt-Boyes, editors Time-dependent changes to the leaflet elasticity of the Medtronic Intact valve in vivo. World Symposium on Heart Valve Disease, London; 1999.
Chuong, C. J., and Y. C. Fung. Three-dimensional stress distribution in arteries. J. Biomech. Eng. 105(3):268–274, 1983.
Chuong, C. J., and Y. C. Fung. On residual stresses in arteries. J. Biomech. Eng. 108(2):189–192, 1986.
Cohn, L. H. Cutler, Elliot, Carr mitral-valve surgery at Peter-Bent-Brigham-Hospital 1923. J. Card. Surg. 9:137–138, 1994.
Cohn, L. H., J. J. Collins, Jr, V. J. Disesa, G. S. Couper, P. S. Peigh, W. Kowalker, et al. Fifteen-year experience with 1678 Hancock Porcine bioprosthetic heart vavle replacements. Ann. Surg. 210(4):435–442, 1989.
Conti, C. A., E. Votta, A. Della Corte, L. Del Viscovo, C. Bancone, M. Cotrufo, et al. Dynamic finite element analysis of the aortic root from MRI-derived parameters. Med. Eng. Phys. 32(2):212–221, 2010. doi:10.1016/j.medengphy.2009.12.003.
Courant, R., K. Friedrichs, and H. Lewy. Über die partiellen Differenzengleichungen der mathematischen Physik. Math. Ann. 100(1):32–74, 1928.
Courant, R., K. Friedrichs, and H. Lewy. On the partial difference equations of mathematical physics. IBM J. Res. Develop. 11:215–234, 1967.
Crofts, C. E., and E. A. Trowbridge. The tensile strength of natural and chemically modified bovine pericardium. J. Biomed. Mater. Res. 22:89–98, 1988.
de Hart, J. Fluid-Structure Interaction in the Aortic Heart Valve: a three-dimensional computational analysis (Ph.D. Thesis). Eindhoven: Technische Universiteit Eindhoven, 2002.
De Hart, J., F. P. Baaijens, G. W. Peters, and P. J. Schreurs. A computational fluid-structure interaction analysis of a fiber-reinforced stentless aortic valve. J. Biomech. 36(5):699–712, 2003.
De Hart, J., G. W. M. Peters, P. J. G. Schreurs, and F. P. T. Baaijens. A three-dimensional computational analysis of fluid-structure interaction in the aortic valve. J. Biomech. 36:103–112, 2003.
Donea, J., S. Giuliani, and J. P. Halleux. An arbitrary Lagrangian-Eulerian finite element method for transient dynamic fluid-structure interactions. Comput. Methods Appl. Mech. Eng. 33:689–723, 1982.
Donea, J., A. Huerta, J.-P. Ponthot, and A. Rodriguez-Ferran. Arbitrary Lagrangian-Eulerian Methods. Encyclopedia of Computational Mechanics. New York: Wiley, 2004.
Driessen, N. J., C. V. Bouten, and F. P. Baaijens. A structural constitutive model for collagenous cardiovascular tissues incorporating the angular fiber distribution. J. Biomech. Eng. 127(3):494–503, 2005.
Eckert, C. E., R. Fan, B. Mikulis, M. Barron, C. A. Carruthers, V. M. Friebe, et al. On the biomechanical role of glycosaminoglycans in the aortic heart valve leaflet. Acta Biomater. 9(1):4653–4660, 2013. doi:10.1016/j.actbio.2012.09.031.
Engelmayr, Jr, G. C., D. K. Hildebrand, F. W. Sutherland, J. E. Mayer, Jr, and M. S. Sacks. A novel bioreactor for the dynamic flexural stimulation of tissue engineered heart valve biomaterials. Biomaterials 24(14):2523–2532, 2003.
Engelmayr, G. C., and M. S. Sacks. A structural model for the flexural mechanics of nonwoven tissue engineering scaffolds. J. Biomech. Eng. 128:610–622, 2006.
Ennker, J., A. Albert, and I. C. Ennker. Stentless aortic valves. Current aspects. HSR Proc. Intens. Care Cardiovasc. Anesthesia. 4(2):77–82, 2012.
Fan, R., and M. S. Sacks. Simulation of planar soft tissues using a structural constitutive model: finite element implementation and validation. J. Biomech. 47:2043–2054, 2014.
Fata, B., C. A. Carruthers, G. Gibson, S. Watkins, D. Gottlieb, J. E. Mayer, Jr, et al. Regional structural and biomechanical alterations of the ovine main pulmonary artery during postnatal growth. J. Biomech. Eng. 135(2):0210221–02102211, 2013.
Fata, B., W. Zhang, R. Amini, and M. Sacks. Insights into regional adaptations in the growing pulmonary artery using a meso-scale structural model: effects of ascending aorta impingement. J. Biomech. Eng. 2014. doi:10.1115/1.4026457.
Freed, A. D., and T. C. Doehring. Elastic model for crimped collagen fibrils. J. Biomech. Eng. 127(4):587–593, 2005.
Funder JA. Current status on stentless aortic bioprosthesis: a clinical and experimental perspective. Eur. J. Cardiothorac. Surg. 41(4):790–799, 2012. doi:10.1093/ejcts/ezr141.
Gabbay, S., U. Bortolotti, F. Wasserman, N. Tindel, S. M. Factor, and R. W. Frater. Long-term follow-up of the Ionescu-Shiley mitral pericardial xenograft. J. Thorac. Cardiovasc. Surg. 88(5 Pt 1):758–763, 1984.
Gao, B. Z., S. Pandya, C. Arana, and N. H. C. Hwang. Bioprosthetic heart valve leaflet deformation monitored by double-pulse stereo photogrammetry. Ann. Biomed. Eng. 30(1):11–18, 2002.
Gao, Z. B., S. Pandya, N. Hosein, M. S. Sacks, and N. H. C. Hwang. Bioprosthetic heart valve leaflet motion monitored by dual camera stereo photogrammetry. J. Biomech. 33(2):199–207, 2000. doi:10.1016/S0021-9290(99)00165-7.
Garikipati, K., S. Göktepe, and C. Miehe. Elastica-based strain energy functions for soft biological tissue. J. Mech. Phys. Solids 56(4):1693–1713, 2008. doi:10.1016/j.jmps.2007.07.005.
Ge, L., and F. Sotiropoulos. A numerical method for solving the 3D unsteady incompressible Navier-Stokes equations in curvilinear domains with complex immersed boundaries. J. Comput. Phys. 225(2):1782–1809, 2007.
Gilmanov, A., T. B. Le, and F. Sotiropoulos. A numerical approach for simulating fluid structure interaction of flexible thin shells undergoing arbitrarily large deformations in complex domains. J. Comput. Phys. 300:814–843, 2015. doi:10.1016/j.jcp.2015.08.008.
Gilmanov, A., and F. Sotiropoulos. Comparative hemodynamics in an aorta with bicuspid and trileaflet valves. Theoret. Comput. Fluid Dyn. 30(1):67–85, 2015. doi:10.1007/s00162-015-0364-7.
Gingold, R. A., and J. J. Monaghan. Smoothed particle hydrodynamics: theory and application to non-spherical stars. Mon. Not. R. Astron. Soc. 181(3):375–389, 1977. doi:10.1093/mnras/181.3.375.
Girardot, M. N., M. Torrianni, and J. M. Girardot. Effect of AOA on glutaraldehyde-fixed bioprosthetic heart vavlve cusps and walls: binding and calcification studies. Int. J. Artif. Organs 17:76–82, 1994.
Gloeckner, D. C., K. L. Billiar, and M. S. Sacks. Effects of mechanical fatigue on the bending properties of the porcine bioprosthetic heart valve. ASAIO J. 45(1):59–63, 1999. doi:10.1097/00002480-199901000-00014.
Gnyaneshwar, R., R. K. Kumar, and K. R. Balakrishnan. Dynamic analysis of the aortic valve using a finite element model. Ann. Thorac. Surg. 73(4):1122–1129, 2002.
Go, A. S., D. Mozaffarian, V. L. Roger, E. J. Benjamin, J. D. Berry, M. J. Blaha, et al. Heart disease and stroke statistics—2014 update: a report from the American Heart Association. Circulation 129(3):e28–e292, 2014. doi:10.1161/01.cir.0000441139.02102.80.
Go, A., D. Mozaffarian, V. Roger, E. Benjamin, J. Berry, W. Borden, et al. Heart disease and stroke statistics—2013 update: a report from the American Heart Association. Circulation 127(1):e6–e245, 2013.
Gorman, 3rd, J. H., R. C. Gorman, B. M. Jackson, Y. Enomoto, M. G. St John-Sutton, and L. H. Edmunds, Jr. Annuloplasty ring selection for chronic ischemic mitral regurgitation: lessons from the ovine model. Ann. Thorac. Surg. 76(5):1556–1563, 2003.
Gorman, 3rd, J. H., K. B. Gupta, J. T. Streicher, R. C. Gorman, B. M. Jackson, M. B. Ratcliffe, et al. Dynamic three-dimensional imaging of the mitral valve and left ventricle by rapid sonomicrometry array localization. J. Thorac. Cardiovasc. Surg. 112(3):712–726, 1996.
Gould, P. L., A. Cataloglu, G. Dhatt, A. Chattopadhyay, and R. E. Clark. Stress analysis of the human aortic valve. Comput. Struct. 3(2):377–384, 1973.
Grande, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Stress variations in the human aortic root and valve: The role of anatomic asymmetry. Ann. Biomed. Eng. 26(4):534–545, 1998. doi:10.1114/1.122.
Grande, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Mechanisms of aortic valve incompetence in aging: a finite element model. J. Heart Valve Dis. 8(2):149–156, 1999.
Grande, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Mechanisms of aortic valve incompetence: finite element modeling of aortic root dilatation. Ann. Thorac. Surg. 69(6):1851–1857, 2000.
Grande-Allen, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Re-creation of sinuses is important for sparing the aortic valve: a finite element study. J. Thorac. Cardiovasc. Surg. 119(4 Pt 1):753–763, 2000.
Grande-Allen, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Mechanisms of aortic valve incompetence: finite-element modeling of Marfan syndrome. J. Thorac. Cardiovasc. Surg. 122(5):946–954, 2001. doi:10.1067/mtc.2001.116314.
Grande-Allen, K. J., R. P. Cochran, P. G. Reinhall, and K. S. Kunzelman. Finite-element analysis of aortic valve-sparing: influence of graft shape and stiffness. IEEE Trans Biomed. Eng. 48(6):647–659, 2001. doi:10.1109/10.923783.
Grant, C. W., and B. G. Barrat-Boyes. Mechanical properties of porcine pulmonary valve leaflets: how do they differ from aortic leaflets. Ann. Thorac. Surg. 60(Supplement 2):S195–S199, 1995.
Grashow, J. S., M. S. Sacks, J. Liao, and A. P. Yoganathan. Planar biaxial creep and stress relaxation of the mitral valve anterior leaflet. Ann. Biomed. Eng. 34(10):1509–1518, 2006.
Grashow, J. S., A. P. Yoganathan, and M. S. Sacks. Biaixal stress-stretch behavior of the mitral valve anterior leaflet at physiologic strain rates. Ann. Biomed. Eng. 34(2):315–325, 2006. doi:10.1007/s10439-005-9027-y.
Griffith, B. E. Immersed boundary model of aortic heart valve dynamics with physiological driving and loading conditions. Int. J. Numer. Methods Biomed. Eng. 28(3):317–345, 2012.
Hamid, M. S., H. N. Sabbah, and P. D. Stein. Comparison of finite-element stress-analysis of aortic-valve leaflet using either membrane elements or solid elements. Comput. Struct. 20(6):955–961, 1985. doi:10.1016/0045-7949(85)90015-X.
Hamid, M. S., H. N. Sabbah, and P. D. Stein. Influence of stent height upon stresses on the cusps of closed bioprosthetic valves. J. Biomech. 19:759–769, 1986.
Hamid, M. S., H. N. Sabbah, and P. D. Stein. Influence of stent height upon stresses on the cusps of closed bioprosthetic valves. J. Biomech. 19(9):759–769, 1986. doi:10.1016/0021-9290(86)90199-5.
Harada, Y., M. Kawada, K. Ishihara, M. Higashidate, H. Kurosawa, and Y. Imai. A new valved conduit with commissures using a glutaraldehyde preserved equine pericardium. Kyobu Geka. 42(6):457–459, 1989.
Harken, D. E., H. S. Soroff, W. J. Taylor, A. A. Lefemine, S. K. Gupta, and S. Lunzer. Partial and complete prostheses in aortic insufficiency. J. Thorac. Cardiovasc. Surg. 40:744–762, 1960.
Hiester, E. D., and M. S. Sacks. Optimal bovine pericardial tissue selection sites. I. Fiber architecture and tissue thickness measurements. J. Biomed. Mater. Res. 39(2):207–214, 1998.
Hiester, E. D., and M. S. Sacks. Optimal bovine pericardial tissue selection sites. II. Cartographic analysis. J. Biomed. Mater. Res. 39(2):215–221, 1998.
Hilbert, S. L., V. J. Ferrans, and W. M. Swanson. Optical methods for the nondestructive evaluation of collagen morphology in bioprosthetic heart valves. J. Biomed. Mater. Res. 20(9):1411–1421, 1986. doi:10.1002/jbm.820200914.
Holzapfel, G. A., T. C. Gasser, and R. W. Ogden. A new constitutive framework for arterial wall mechanics and a comparative study of material models. J. Elasticity. 61(1–3):1–48, 2000. doi:10.1023/A:1010835316564.
Howard, I. C., E. A. Patterson, and A. Yoxall. On the opening mechanism of the aortic valve: some observations from simulations. J. Med. Eng. Technol. 27(6):259–266, 2003. doi:10.1080/0309190031000096621.
Hsu, M.-C., D. Kamensky, Y. Bazilevs, M. S. Sacks, and T. J. R. Hughes. Fluid-structure interaction analysis of bioprosthetic heart valves: significance of arterial wall deformation. Comput. Mech. 54:1055–1071, 2014.
Hsu, M.-C., D. Kamensky, F. Xu, J. Kiendl, C. Wang, M. C. H. Wu, et al. Dynamic and fluid-structure interaction simulations of bioprosthetic heart valves using parametric design with T-splines and Fung-type material models. Comput. Mech. 55:1211–1225, 2015.
Huang, X., M. M. Black, I. C. Howard, and E. A. Patterson. A two-dimensional finite element analysis of a bioprosthetic heart valve. J. Biomech. 23(8):753–762, 1990.
Huang, H. Y., J. Liao, and M. S. Sacks. In-situ deformation of the aortic valve interstitial cell nucleus under diastolic loading. J. Biomech. Eng. 129(6):880–889, 2007. doi:10.1115/1.2801670.
Hughes, T. J. R., J. A. Cottrell, and Y. Bazilevs. Isogeometric analysis: CAD, finite elements, NURBS, exact geometry, and mesh refinement. Comput. Methods Appl. Mech. Eng. 194:4135–4195, 2005.
Hughes, T. J. R., W. K. Liu, and T. K. Zimmermann. Lagrangian-Eulerian finite element formulation for incompressible viscous flows. Comput. Method Appl. Mech. Eng. 29:329–349, 1981.
Humphrey, J. D. Cardiovascular solid mechanics: cells, tissues, and organs. Lightning Source: UKltd, 2002.
Humphrey, J. D., R. K. Strumpf, and F. C. Yin. Determination of a constitutive relation for passive myocardium: I. A new functional form. J. Biomech. Eng. 112(3):333–339, 1990.
Humphrey, J. D., R. K. Strumpf, and F. C. Yin. Determination of a constitutive relation for passive myocardium: II. Parameter estimation. J Biomech Eng. 112(3):340–346, 1990.
Humphrey, J. D., R. K. Strumpf, and F. C. Yin. A constitutive theory for biomembranes: application to epicardial mechanics. J. Biomech. Eng. 114(4):461–466, 1992.
Humphrey, J. D., D. L. Vawter, and R. P. Vito. Mechanical behavior of excised canine visceral pleura. Ann. Biomed. Eng. 14(5):451–466, 1986.
Iyengar, A. K. S., H. Sugimoto, D. B. Smith, and M. S. Sacks. Dynamic in vitro quantification of bioprosthetic heart valve leaflet motion using structured light projection. Ann. Biomed. Eng. 29(11):963–973, 2001.
Jamieson, W. R. L. H., L. H. Burr, G. J. Fradet, R. T. Miyagishima, M. T. Janusz, and S. V. Lichtenstein. Carpentier-Edwards supraannular porcine bioprosthesis evaluation over 15 years. Ann. Thorac. Surg. 66(6 Suppl):S49–S52, 1998.
Jennings, L. M., M. Butterfield, C. Booth, K. G. Watterson, and J. Fisher. The pulmonary bioprosthetic heart valve: its unsuitability for use as an aortic valve replacement. J. Heart Valve Dis. 11(5):668–678, 2002.
Johnson, A. A., and T. E. Tezduyar. Parallel Computation of Incompressible Flows with Complex Geometries. Int J Numer Meth Fl. 24:1321–1340, 1997.
Johnson, A. A., and T. E. Tezduyar. 3d simulation of fluid-particle interactions with the number of particles reaching 100. Comput. Method Appl. Mech. Eng. 145:301–321, 1997.
Johnson, A. A., and T. E. Tezduyar. Advanced mesh generation and update methods for 3D flow simulations. Comput. Mech. 23:130–143, 1999.
Jones, M., E. E. Eidbo, S. L. Hilbert, V. J. Ferrans, and R. E. Clark. Anticalcification treatments of bioprosthetic heart valves: in vivo studies in sheep. J. Card. Surg. 4(1):69–73, 1989.
Jouan, J. Mitral valve repair over five decades. Ann. Cardiothorac. Surg. 4(4):322–334, 2015. doi:10.3978/j.issn.2225-319X.2015.01.07.
Kamensky, D., M.-C. Hsu, D. Schillinger, J. A. Evans, A. Aggarwal, Y. Bazilevs, et al. An immersogeometric variational framework for fluid-structure interaction: Application to bioprosthetic heart valves. Comput. Method Appl. Mech. Eng. 284:1005–1053, 2015.
Kiendl, J. Isogeometric Analysis and Shape Optimal Design of Shell Structures. Lehrstuhl für Statik: Technische Universität München, München, 2011.
Kiendl, J., K.-U. Bletzinger, J. Linhard, and R. Wüchner. Isogeometric shell analysis with Kirchhoff-Love elements. Comput. Method Appl. Mech. Eng. 198:3902–3914, 2009.
Kiendl, J., M.-C. Hsu, M. C. H. Wu, and A. Reali. Isogeometric Kirchhoff-Love shell formulations for general hyperelastic materials. Comput. Method Appl. Mech. Eng. 291:280–303, 2015.
Kim, H., K. B. Chandran, M. S. Sacks, and J. Lu. An experimentally derived stress resultant shell model for heart valve dynamic simulations. Ann. Biomed. Eng. 35(1):30–44, 2007. doi:10.1007/s10439-006-9203-8.
Kim, H., J. Lu, M. S. Sacks, and K. B. Chandran. Dynamic simulation of bioprosthetic heart valves using a stress resultant shell model. Ann. Biomed. Eng. 36(2):262–275, 2008. doi:10.1007/s10439-007-9409-4.
Kronick, P. L., and M. S. Sacks. Matrix macromolecules that affect the viscoelasticity of calfskin. J Biomech Eng-T ASME. 116(2):140–145, 1994. doi:10.1115/1.2895712.
Krucinski, S., I. Vesely, M. A. Dokainish, and G. Campbell. Numerical simulation of leaflet flexure in bioprosthetic valves mounted on rigid and expansile stents. J. Biomech. 26(8):929–943, 1993.
Kunzelman, K. S., D. R. Einstein, and R. P. Cochran. Fluid-structure interaction models of the mitral valve: function in normal and pathological states. Philos. Trans. R. Soc. Lond. B Biol. Sci. 362(1484):1393–1406, 2007. doi:10.1098/rstb.2007.2123.
Labrosse, M. R., C. J. Beller, M. Boodhwani, C. Hudson, and B. Sohmer. Subject-specific finite-element modeling of normal aortic valve biomechanics from 3D + t TEE images. Med. Image Anal. 20(1):162–172, 2015. doi:10.1016/j.media.2014.11.003.
Lanir, Y. A structural theory for the homogeneous biaxial stress-strain relationships in flat collagenous tissues. J. Biomech. 12(6):423–436, 1979.
Lanir, Y. Plausibility of structural constitutive equations for swelling tissues–implications of the C-N and S-E conditions. J. Biomech. Eng. 118(1):10–16, 1996.
Leat, M. E., and J. Fisher. Comparative-study of the function of the abiomed polyurethane heart-valve for use in left-ventricular assist devices. J. Biomed. Eng. 15(6):516–520, 1993. doi:10.1016/0141-5425(93)90068-A.
Lee, J. M., and D. R. Boughner. Tissue mechanics of canine pericardium in different test environments. Circ. Res. 49:533–544, 1981.
Lee, J. M., D. R. Boughner, and D. W. Courtman. The glutaraldehyde-stabilized porcine aortic valve xenograft. II. Effect of fixation with or without pressure on the tensile viscoelastic properties of the leaflet material. J. Biomed. Mater. Res. 18:79–98, 1984.
Lee, J. M., S. A. Haberer, and D. R. Boughner. The bovine pericardial xenograft: I. Effect of fixation in aldehydes without constraint on the tensile properties of bovine pericardium. J. Biomed. Mater. Res. 23:457–475, 1989.
Lee, C.-H., W. Zhang, J. Liao, C. A. Carruthers, J. I. Sacks, and M. S. Sacks. On the presence of affine fibril and fiber kinematics in the mitral valve anterior leaflet. Biophys. J. 108(8):2074–2087, 2015. doi:10.1016/j.bpj.2015.03.019.
Leeson-Dietrich, J., D. Boughner, and I. Vesely. Porcine pulmonary and aortic valves: a comparison of their tensile viscoelastic properties at physiological strain rates. J. Heart Valve Dis. 4:88–94, 1995.
Leon, M. B., C. R. Smith, M. J. Mack, R. R. Makkar, L. G. Svensson, S. K. Kodali, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. New Engl. J. Med. 2016. doi:10.1056/NEJMoa1514616.
Li, J., X. Y. Luo, and Z. B. Kuang. A nonlinear anisotropic model for porcine aortic heart valves. J. Biomech. 34(10):1279–1289, 2001. doi:10.1016/S0021-9290(01)00092-6.
Li, K., and W. Sun. Simulated thin pericardial bioprosthetic valve leaflet deformation under static pressure-only loading conditions: implications for percutaneous valves. Ann. Biomed. Eng. 38(8):2690–2701, 2010. doi:10.1007/s10439-010-0009-3.
Liao, J., E. M. Joyce, and M. S. Sacks. Effects of decellularization on the mechanical and structural properties of the porcine aortic valve leaflet. Biomaterials 29(8):1065–1074, 2008. doi:10.1016/j.biomaterials.2007.11.007.
Liao, J., L. Yang, J. Grashow, and M. S. Sacks, editors. Collagen fibril kinematics in mitral valve leaflet under biaxial elongation, creep, and stress relaxation. Society for Heart Valve Disease Third Biennial Meeting, 2005; Vancouver: SHVD.
Liao, J., L. Yang, J. Grashow, and M. S. Sacks. The relation between collagen fibril kinematics and mechanical properties in the mitral valve anterior leaflet. J. Biomech. Eng. 129(1):78–87, 2007. doi:10.1115/1.2401186.
Lo, D., and I. Vesely. Biaxial strain analysis of the porcine aortic valve. Ann. Thorac. Surg. 60(2 Suppl):S374–S378, 1995.
Lovekamp, J. J., D. T. Simionescu, J. J. Mercuri, B. Zubiate, M. S. Sacks, and N. R. Vyavahare. Stability and function of glycosaminoglycans in porcine bioprosthetic heart valves. Biomaterials 27(8):1507–1518, 2006. doi:10.1016/j.biomaterials.2005.08.003.
Maestro, M. M., J. Turnay, N. Olmo, P. Fernández, D. Suárez, J. M. García Páez, et al. Biochemical and mechanical behavior of ostrich pericardium as a new biomaterial. Acta Biomater. 2(2):213–219, 2006.
Makhijani, V. B., H. Q. Yang, P. J. Dionne, and M. J. Thubrikar. Three-dimensional coupled fluid-structure simulation of pericardial bioprosthetic aortic valve function. ASAIO J. 43(5):M387–M392, 1997.
Mako, W. J., and I. Vesely. In vivo and in vitro models of calcification in porcine aortic valve cusps. J. Heart Valve Dis. 6(3):316–323, 1997.
Martin, C., and W. Sun. Biomechanical characterization of aortic valve tissue in humans and common animal models. J. Biomed. Mater. Res. A 100(6):1591–1599, 2012.
Mercuri, J. J., J. J. Lovekamp, D. T. Simionescu, and N. R. Vyavahare. Glycosaminoglycan-targeted fixation for improved bioprosthetic heart valve stabilization. Biomaterials 28(3):496–503, 2007. doi:10.1016/j.biomaterials.2006.09.005.
Merryman, W. D., H. Y. Huang, F. J. Schoen, and M. S. Sacks. The effects of cellular contraction on aortic valve leaflet flexural stiffness. J. Biomech. 39(1):88–96, 2006.
Michler, C., H. van Brummelen, and R. de Borst. An investigation of Interface-GMRES(R) for fluid-structure interaction problems with flutter and divergence. Comput. Mech. 47(1):17–29, 2011.
Mirnajafi, A., J. M. Raymer, L. R. McClure, and M. S. Sacks. The flexural rigidity of the aortic valve leaflet in the commissural region. J. Biomech. 39(16):2966–2973, 2006. doi:10.1016/j.jbiomech.2005.10.026.
Mirnajafi, A., J. Raymer, M. J. Scott, and M. S. Sacks. The effects of collagen fiber orientation on the flexural properties of pericardial heterograft biomaterials. Biomaterials 26(7):795–804, 2005. doi:10.1016/j.biomaterials.2004.03.004.
Mirnajafi, A., B. Zubiate, and M. S. Sacks. Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials. J. Biomed. Mater. Res. A. 94(1):205–213, 2010. doi:10.1002/jbm.a.32659.
Mittal, R., and G. Iaccarino. Immersed boundary methods. Annu. Rev. Fluid Mech. 37:239–261, 2005.
Moore, M. A., R. E. Phillips, Jr, B. K. McIlroy, V. M. Walley, and P. J. Hendry. Evaluation of porcine valves prepared by dye-mediated photooxidation. Ann. Thorac. Surg. 66(6 Suppl):S245–S248, 1998.
Morganti, S., F. Auricchio, D. J. Benson, F. I. Gambarin, S. Hartmann, T. J. R. Hughes, et al. Patient-specific isogeometric structural analysis of aortic valve closure. Comput. Method Appl. M. 284:508–520, 2015.
Mozaffarian, D., E. J. Benjamin, A. S. Go, D. K. Arnett, M. J. Blaha, M. Cushman, et al. Heart disease and stroke statistics—2015 update: a report from the American Heart Association. Circulation 131(4):e29–e322, 2015. doi:10.1161/CIR.0000000000000152.
Neethling, W. M., S. Cooper, J. J. Van Den Heever, J. Hough, and A. J. Hodge. Evaluation of kangaroo pericardium as an alternative substitute for reconstructive cardiac surgery. J. Cardiovasc. Surg. 43(3):301–306, 2002.
Parry, D. A. The molecular and fibrillar structure of collagen and its relationship to the mechanical properties of connective tissue. Biophys. Chem. 29(1–2):195–209, 1988.
Patterson, E. A., I. C. Howard, and M. A. Thornton. A comparative study of linear and nonlinear simulations of the leaflets in a bioprosthetic heart valve during the cardiac cycle. J. Med. Eng. Technol. 20(3):95–108, 1996. doi:10.3109/03091909609008387.
Payne, D. M., H. P. Koka, P. J. Karanicolas, M. W. Chu, A. D. Nagpal, M. Briel, et al. Hemodynamic performance of stentless versus stented valves: a systematic review and meta-analysis. J. Card. Surg. 23(5):556–564, 2008. doi:10.1111/j.1540-8191.2008.00705.x.
Pereira, C. A., J. M. Lee, and S. A. Haberer. Effect of alternative crosslinking methods on the low strain rate viscoelastic properties of bovine pericardial bioprosthetic material. J. Biomed. Mater. Res. 24:345–361, 1990.
Perez de Arenaza, D., B. Lees, M. Flather, F. Nugara, T. Husebye, M. Jasinski, et al. Randomized comparison of stentless versus stented valves for aortic stenosis: effects on left ventricular mass. Circulation 112(17):2696–2702, 2005. doi:10.1161/CIRCULATIONAHA.104.521161.
Peskin, C. S. Flow patterns around heart valves: a numerical method. J. Comput. Phys. 10(2):252–271, 1972.
Pibarot, P., and J. G. Dumesnil. Valvular heart disease: changing concepts in disease management. Circulation 119:1034–1048, 2009.
Pouch, A., S. Tian, M. Takabe, H. Wang, J. Yuan, A. Cheung, et al. Segmentation of the aortic valve apparatus in 3D echocardiographic images: deformable modeling of a branching medial structure. In: Statistical Atlases and Computational Models of the Heart—Imaging and Modelling Challenges, edited by O. Camara, T. Mansi, M. Pop, K. Rhode, M. Sermesant, and A. Young. Lecture Notes in Computer Science: Springer International Publishing, 2015, pp. 196–203.
Quinonez, L. G., R. Breitbart, W. Tworetsky, J. E. Lock, A. C. Marshall, and S. M. Emani. Stented bovine jugular vein graft (Melody valve) for surgical mitral valve replacement in infants and children. J. Thorac. Cardiovasc. Surg. 148(4):1443–1449, 2014. doi:10.1016/j.jtcvs.2013.10.059.
Rabkin, E., S. P. Hoerstrup, M. Aikawa, J. E. Mayer, Jr, and F. J. Schoen. Evolution of cell phenotype and extracellular matrix in tissue-engineered heart valves during in vitro maturation and in vivo remodeling. J. Heart Valve Dis. 11(3):308–314, 2002.
Roger, V. L., A. S. Go, D. M. Lloyd-Jones, E. J. Benjamin, J. D. Berry, W. B. Borden, et al. Heart disease and stroke statistics—2012 update: a report from the American Heart Association. Circulation 125(1):e2–e220, 2012. doi:10.1161/CIR.0b013e31823ac046.
Ross, D. N. Replacement of aortic and mitral valves with a pulmonary autograft. Lancet 2(7523):956–958, 1967.
Rousseau, E. P. M., A. A. Vansteenhoven, J. D. Janssen, and H. A. Huysmans. A mechanical analysis of the closed hancock heart-valve prosthesis. J. Biomech. 21(7):545–562, 1988. doi:10.1016/0021-9290(88)90218-7.
Sacks, M. S. A structural constitutive model for chemically treated planar tissues under biaxial loading. Comput. Mech. 26(3):243–249, 2000. doi:10.1007/S004660000175.
Sacks, M. S. Incorporation of experimentally-derived fiber orientation into a structural constitutive model for planar-collagenous tissues. J. Biomech. Eng. T ASME 125(2):280–287, 2003. doi:10.1115/1.1544508.
Sacks, M. S., and C. J. Chuong. Orthotropic mechanical properties of chemically treated bovine pericardium. Ann. Biomed. Eng. 26(5):892–902, 1998. doi:10.1114/1.135.
Sacks, M. S., Y. Enomoto, J. R. Graybill, W. D. Merryman, A. Zeeshan, A. P. Yoganathan, et al. In-vivo dynamic deformation of the mitral valve anterior leaflet. Ann. Thorac. Surg. 82(4):1369–1377, 2006. doi:10.1016/j.athoracsur.2006.03.117.
Sacks, M. S., A. Mirnajafi, W. Sun, and P. Schmidt. Bioprosthetic heart valve heterograft biomaterials: structure, mechanical behavior and computational simulation. Expert Rev. Med. Devic. 3(6):817–834, 2006. doi:10.1586/17434440.3.6.817.
Sacks, M. S., and F. J. Schoen. Collagen fiber disruption occurs independent of calcification in clinically explanted bioprosthetic heart valves. J. Biomed. Mater. Res. 62(3):359–371, 2002.
Sacks, M. S., F. J. Schoen, and J. E. Mayer. Bioengineering challenges for heart valve tissue engineering. Annu. Rev. Biomed. Eng. 11:289–313, 2009. doi:10.1146/annurev-bioeng-061008-124903.
Sacks, M. S., D. B. Smith, and E. D. Hiester. A small angle light scattering device for planar connective tissue microstructural analysis. Ann. Biomed. Eng. 25(4):678–689, 1997.
Sacks, M. S., and W. Sun. Multiaxial mechanical behavior of biological materials. Annu. Rev. Biomed. Eng. 5:251–284, 2003. doi:10.1146/Annurev.Bioeng.5.011303.120714.
Sacks, M. S., W. Zhang, and S. Wognum. A novel fibre-ensemble level constitutive model for exogenous cross-linked collagenous tissues. Interface Focus. 6(1):20150090, 2016. doi:10.1098/rsfs.2015.0090.
Salgo, I. S., J. H. Gorman, R. C. Gorman, B. M. Jackson, F. W. Bowen, T. Plappert, et al. Effect of annular shape on leaflet curvature in reducing mitral leaflet stress. Circulation 106(6):711–717, 2002. doi:10.1161/01.Cir.0000025426.39426.83.
Salmasi, M. Y., M. Acharya, N. Humayun, D. Baskaran, S. Hubbard, and H. Vohra. Is valve repair preferable to valve replacement in ischaemic mitral regurgitation? A systematic review and meta-analysis. Eur. J. Cardiothorac. Surg. 2016. doi:10.1093/ejcts/ezw053.
Sasaki, N., and S. Odajima. Elongation mechanism of collagen fibrils and force-strain relations of tendon at each level of structural hierarchy. J. Biomech. 29(9):1131–1136, 1996.
Sasaki, N., and S. Odajima. Stress-strain curve and Young’s modulus of a collagen molecule as determined by the X-ray diffraction technique. J. Biomech. 29:655–658, 1996.
Sauren, A. A., M. C. van Hout, A. A. van Steenhoven, F. E. Veldpaus, and J. D. Janssen. The mechanical properties of porcine aortic valve tissues. J. Biomech. 16(5):327–337, 1983.
Schoen, F. Aortic valve structure-function correlations: Role of elastic fibers no longer a stretch of the imagination. J. Heart Valve Dis. 6:1–6, 1997.
Schoen, F. J. Pathologic findings in explanted clinical bioprosthetic valves fabricated from photooxidized bovine pericardium. J. Heart Valve Dis. 7(2):174–179, 1998.
Schoen, F. J., J. Fernandez, L. Gonzalezlavin, and A. Cernaianu. Causes of failure and pathological findings in surgically removed ionescu-shiley standard bovine pericardial heart-valve bioprostheses—emphasis on progressive structural deterioration. Circulation 76(3):618–627, 1987.
Siddiqui, R. F., J. R. Abraham, and J. Butany. Bioprosthetic heart valves: modes of failure. Histopathology 55(2):135–144, 2009.
Simo, J. C., and D. D. Fox. On a stress resultant geometrically exact shell-model. 1. Formulation and optimal parametrization. Comput. Method Appl. Mech. Eng. 72(3):267–304, 1989. doi:10.1016/0045-7825(89)90002-9.
Simo, J. C., D. D. Fox, and M. S. Rifai. On a stress resultant geometrically exact shell-model. 3. computational aspects of the nonlinear-theory. Comput. Method Appl. Mech. Eng. 79(1):21–70, 1990. doi:10.1016/0045-7825(90)90094-3.
Smith, D. B., M. S. Sacks, P. M. Pattany, and R. Schroeder. High-resolution magnetic resonance imaging to characterize the geometry of fatigued porcine bioprosthetic heart valves. J. Heart Valve Dis. 6(4):424–432, 1997.
Smith, D. B., M. S. Sacks, P. M. Pattany, and R. Schroeder. Fatigue-induced changes in bioprosthetic heart valve three-dimensional geometry and the relation to tissue damage. J. Heart Valve Dis. 8(1):25–33, 1999.
Sotiropoulos, F., and I. Borazjani. A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves. Med. Biol. Eng. Compu. 47(3):245–256, 2009. doi:10.1007/s11517-009-0438-z.
Sotiropoulos, F., and X. Yang. Immersed boundary methods for simulating fluid-structure interaction. Prog. Aerosp. Sci. 65:1–21, 2014.
Sripathi, V. C., R. K. Kumar, and K. R. Balakrishnan. Further insights into normal aortic valve function: Role of a compliant aortic root on leaflet opening and valve orifice area. Ann. Thorac. Surg. 77(3):844–851, 2004. doi:10.1016/S0003-4975(03)01518-2.
Starr, A. The artificial heart valve. Nat. Med. 13:1160–1164, 2007.
Stella, J. A., J. Liao, and M. S. Sacks. Time-dependent biaxial mechanical behavior of the aortic heart valve leaflet. J. Biomech. 40(14):3169–3177, 2007. doi:10.1016/j.jbiomech.2007.04.001.
Stock, U. A., T. Sakamoto, S. Hatsuoka, D. P. Martin, M. Nagashima, A. M. Moran, et al. Patch augmentation of the pulmonary artery with bioabsorbable polymers and autologous cell seeding. J Thorac Cardiovasc Surg. 120(6):1158–1167, 2000; (discussion 68).
Stradins, P., R. Lacis, I. Ozolanta, B. Purina, V. Ose, L. Feldmane, et al. Comparison of biomechanical and structural properties between human aortic and pulmonary valve. Eur. J. Cardiothorac. Surg. 26(3):634–639, 2004.
Sturla, F., E. Votta, M. Stevanella, C. A. Conti, and A. Redaelli. Impact of modeling fluid-structure interaction in the computational analysis of aortic root biomechanics. MedEngPhys. 35:1721–1730, 2013.
Sugimoto, H., and M. S. Sacks. Effects of leaflet stiffness on in vitro dynamic bioprosthetic heart valve leaflet shape. Cardiovasc. Eng. Technol. 4(1):2–15, 2013.
Sun, W., A. Abad, and M. S. Sacks. Simulated bioprosthetic heart valve deformation under quasi-static loading. J. Biomech. Eng. 127(6):905–914, 2005.
Sun, W., and M. S. Sacks. Finite element implementation of a generalized Fung-elastic constitutive model for planar soft tissues. Biomech. Model Mechan. 4(2–3):190–199, 2005. doi:10.1007/S10237-005-0075-X.
Sun, W., M. S. Sacks, and M. J. Scott. Effects of boundary conditions on the estimation of the planar biaxial mechanical properties of soft tissues. J. Biomech. Eng. T ASME 127(4):709–715, 2005. doi:10.1115/1.1933931.
Sun, W., M. S. Sacks, T. L. Sellaro, W. S. Slaughter, and M. J. Scott. Biaxial mechanical response of bioprosthetic heart valve biomaterials to high in-plane shear. J. Biomech. Eng. 125:372–380, 2003.
Sung, H. W., Y. Chang, C. T. Chiu, C. N. Chen, and H. C. Liang. Crosslinking characteristics and mechanical properties of a bovine pericardium fixed with a naturally occurring crosslinking agent. J. Biomed. Mater. Res. 47(2):116–126, 1999.
Sung, H. W., Y. Chang, C. T. Chiu, C. N. Chen, and H. C. Liang. Mechanical properties of a porcine aortic valve fixed with a naturally occurring crosslinking agent. Biomaterials 20(19):1759–1772, 1999.
Sung, H. W., W. H. Chang, C. Y. Ma, and M. H. Lee. Crosslinking of biological tissues using genipin and/or carbodiimide. J. Biomed. Mater. Res. 63(3):427–438, 2003.
Sutherland, F. W., T. E. Perry, Y. Yu, M. C. Sherwood, E. Rabkin, Y. Masuda, et al. From stem cells to viable autologous semilunar heart valve. Circulation 111(21):2783–2791, 2005.
Takizawa, K., C. Moorman, S. Wright, J. Christopher, and T. E. Tezduyar. Wall shear stress calculations in space-time finite element computation of arterial fluid-structure interactions. Comput. Mech. 46:31–41, 2010.
Takizawa, K., T. E. Tezduyar, A. Buscher, and S. Asada. Space-time interface-tracking with topology change (ST-TC). Comput. Mech. 54(4):955–971, 2013.
Takizawa, K., T. Tezduyar, A. Buscher, and S. Asada. Space-time fluid mechanics computation of heart valve models. Comput. Mech. 54(4):973–986, 2014.
Tepole, A. B., H. Kabaria, K.-U. Bletzinger, and E. Kuhl. Isogeometric Kirchhoff-Love shell formulations for biological membranes. Comput. Method Appl. Mech. Eng. 293:328–347, 2015.
Tezduyar, T. E. Computation of moving boundaries and interfaces and stabilization parameters. Int. J. Numer. Methods F. 43:555–575, 2003.
Tezduyar, T., S. Aliabadi, M. Behr, A. Johnson, and S. Mittal. Massively parallel finite element computation of 3D flows—mesh update strategies in computation of moving boundaries and interfaces. Parallel Computational Fluid Dynamics. New Trends and Advances. 1995; pp. 21–30.
Tezduyar, T. E., M. Behr, and J. Liou. A new strategy for finite element computations involving moving boundaries and interfaces—the deforming-spatial-domain/space-time procedure: I1 The concept and the preliminary numerical tests. Comput. Method Appl. Mech. Eng. 94(3):339–351, 1992.
Tezduyar, T. E., M. Behr, S. Mittal, and J. Liou. A new strategy for finite element computations involving moving boundaries and interfaces—the deforming-spatial-domain/space-time procedure: II1 Computation of free-surface flows, two-liquid flows, and flows with drifting cylinders. Comput. Method Appl. Mech. Eng. 94(3):353–371, 1992.
Thornton, M. A., L. C. Howard, and E. A. Patterson. Three-dimensional stress analysis of polypropylene leaflets for prosthetic heart valves. Med. Eng. Phys. 19(6):588–597, 1997. doi:10.1016/S1350-4533(96)00042-2.
Thubrikar, M. The Aortic Valve. Boca Raton: CRC, 1990.
Thubrikar, M. J., J. D. Deck, J. Aouad, and S. P. Nolan. Role of mechanical stress in calcification of aortic bioprosthetic valves. J. Thorac. Cardiovasc. Surg. 86(1):115–125, 1983.
Thubrikar, M. J., J. L. Heckman, and S. P. Nolan. High speed cine-radiographic study of aortic valve leaflet motion. J. Heart Valve Dis. 2(6):653–661, 1993.
Thubrikar, M. J., J. R. Skinner, R. T. Eppink, and S. P. Nolan. Stress analysis of porcine bioprosthetic heart valves in vivo. J. Biomed. Mater. Res. 16(6):811–826, 1982. doi:10.1002/jbm.820160607.
Toma, M., M. O. Jensen, D. R. Einstein, A. P. Yoganathan, R. P. Cochran, and K. S. Kunzelman. Fluid-structure interaction analysis of papillary muscle forces using a comprehensive mitral valve model with 3D chordal structure. Ann. Biomed. Eng. 2015. doi:10.1007/s10439-015-1385-5.
Tong, P., and Y. C. Fung. The stress-strain relationship for the skin. J. Biomech. 9(10):649–657, 1976.
Trowbridge, E. A., and C. E. Crofts. Pericardial heterograft valves: an assessment of leaflet stresses and their implications for heart valve design. J. Biomed. Eng. 9(4):345–355, 1987.
Trowbridge, E. A., P. V. Lawford, and C. E. Crofts. Pericardial heterografts: a comparative study of suture pull-out and tissue strength. J. Biomed. Eng. 11(4):311–314, 1989.
van Brummelen, E. H. Added mass effects of compressible and incompressible flows in fluid-structure interaction. J. Appl. Mech. 76:021206, 2009.
van Loon, R. A 3D method for modelling the fluid-structure interaction of heart valves (Ph.D. thesis). Eindhoven: Technische Universiteit Eindhoven, 2005.
van Loon, R. Towards computational modelling of aortic stenosis. Int. J. Numer. Methods Biomed. Eng. 26:405–420, 2010.
van Loon, R., P. D. Anderson, and F. N. van de Vosse. A fluid-structure interaction method with solid-rigid contact for heart valve dynamics. J. Comput. Phys. 217:806–823, 2006.
Vesely, I. The role of elastin in aortic valve mechanics. J. Biomech. 31(2):115–123, 1998.
Vesely, I., J. E. Barber, and N. B. Ratliff. Tissue damage and calcification may be independent mechanisms of bioprosthetic heart valve failure. J. Heart Valve Dis. 10(4):471–477, 2001.
Vesely, I., and D. Boughner. Analysis of the bending behaviour of porcine xenograft leaflets and of neutral aortic valve material: bending stiffness, neutral axis and shear measurements. J. Biomech. 22(6–7):655–671, 1989.
Vesely, I., and R. Noseworthy. Micromechanics of the fibrosa and the ventricularis in aortic valve leaflets. J. Biomech. 25(1):101–113, 1992.
Vyavahare, N., M. Ogle, F. J. Schoen, R. Zand, D. C. Gloeckner, M. S. Sacks, et al. Mechanisms of bioprosthetic heart valve failure: Fatigue causes collagen denaturation and glycosaminoglycan loss. J. Biomed. Mater. Res. 46:44–50, 1999.
Wells, S. M., and M. S. Sacks. Effects of fixation pressure on the biaxial mechanical behavior of porcine bioprosthetic heart valves with long-term cyclic loading. Biomaterials 23(11):2389–2399, 2002.
Wiegner, A. W., O. H. Bing, T. K. Borg, and J. B. Caulfield. Mechanical and structural correlates of canine pericardium. Circ. Res. 49(3):807–814, 1981.
Wilber, J. P., and J. R. Walton. The convexity properties of a class of constitutive models for biological soft issues. Math. Mech. Solids 7(3):217–235, 2002. doi:10.1177/108128602027726.
Wu, W., D. Pott, B. Mazza, T. Sironi, E. Dordoni, C. Chiastra, et al. Fluid-structure interaction model of a percutaneous aortic valve: comparison with an in vitro test and feasibility study in a patient-specific case. Ann. Biomed. Eng. 44(2):590–603, 2016. doi:10.1007/s10439-015-1429-x.
Yahia, L. H., and G. Drouin. Microscopical investigation of canine anterior cruciate ligament and patellar tendon: collagen fascicle morphology and architecture. J. Orthop. Res. 7(2):243–251, 1989. doi:10.1002/jor.1100070212.
Ye, Q., G. Zund, S. Jockenhoevel, S. P. Hoerstrup, A. Schoeberlein, J. Grunenfelder, et al. Tissue engineering in cardiovascular surgery: new approach to develop completely human autologous tissue. Eur. J. Cardiothorac. Surg. 17(4):449–454, 2000.
Zhang, W., S. Ayoub, J. Liao, and M. S. Sacks. On the mechanical role of collagen and elastin fibers in the layers of the mitral heart valve leaflet. J. Mech. Behav. Biomed. Mater., in press.
Acknowledgments
National Institute of Health, Award Number R01 HL119297 and R01 HL63954 to MSS. National Institute of Health, Award T32 to KRF. American Heart Association, Post Doctoral Fellowship 14POST18720037 to AA.
Conflict of interest
None of the authors have any conflicts of interest to report.
Ethical Approval
No human studies were carried out by the authors for this article. No animal studies were carried out by the authors for this article.
Author information
Authors and Affiliations
Corresponding author
Additional information
Associate Editor Ajit P. Yoganathan oversaw the review of this article.
Rights and permissions
About this article
Cite this article
Soares, J.S., Feaver, K.R., Zhang, W. et al. Biomechanical Behavior of Bioprosthetic Heart Valve Heterograft Tissues: Characterization, Simulation, and Performance. Cardiovasc Eng Tech 7, 309–351 (2016). https://doi.org/10.1007/s13239-016-0276-8