Abstract
Percutaneous aortic valve (PAV) replacement is currently being investigated as an endovascular alternative to conventional open-chest valve surgery for patients with severe aortic stenosis. The results of multi-center clinical trials of PAV devices have been encouraging. However, there are serious adverse events associated with this procedure. Furthermore, long-term durability and safety of PAV need to be studied carefully. In this study, we developed a thin pericardial bioprosthetic valve model, which has similar design features of PAV. We utilized this model to investigate PAV deformation under static, pressure-only loading conditions using Finite Element method. Mechanical properties of PAV leaflet were obtained from planar biaxial testing of glutaraldehyde treated thin bovine pericardium (BP) and porcine pericardium (PP), and characterized by the Fung-elastic model. Simulations were performed to examine the effects of tissue thickness and anisotropy on the valve deformation and stress distribution. The results indicated peak stress and strain occurred in the vicinity of commissures. The peak maximum principal stresses (MPS) were reduced with the increase of leaflet tissue thickness, by 36% and 59% from the mean thickness to 0.35 mm for BP and PP, respectively. The PAV with BP leaflet had a lower peak MPS than that with PP leaflet. Moreover, leaflet material orientation had a significant influence on the peak MPS of PAV.
Similar content being viewed by others
References
AHA, American Heart Association: Heart Disease and Stroke Statistics—2010 Update, 2010.
Balentine, J., and A. Eisenhart, Aortic Stenosis, Emergency Medicine, 2007.
Berland, G., P. Block, T. DeLoughery, and G. Grunkemeier. Clinical one-year outcomes after stenting in acute myocardial infarction. Cathet. Cardiovasc. Diagn. 40(4):337–341, 1997.
Berry, C., A. Asgar, Y. Lamarche, B. Marcheix, P. Couture, A. Basmadjian, A. Ducharme, J. Laborde, R. Cartier, and R. Bonan, Novel therapeutic aspects of percutaneous aortic valve replacement with the 21F CoreValve Revalving System. Catheter. Cardiovasc. Interv., 4(70):610–616, 2007.
Billiar, K. L., and M. S. Sacks. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp—part I: experimental results. J. Biomech. Eng. 122(1):23–30, 2000.
Cacciola, G., G. W. Peters, and F. P. Baaijens. A synthetic fiber-reinforced stentless heart valve. J. Biomech. 33(6):653–658, 2000.
Cacciola, G., G. W. Peters, and P. J. Schreurs. A three-dimensional mechanical analysis of a stentless fibre-reinforced aortic valve prosthesis. J. Biomech. 33(5):521–530, 2000.
Christie, G. W. Computer modelling of bioprosthetic heart valves. Eur. J. Cardio-Thorac. Surg. 6:S95–S101, 1992.
Christie, G. W., and B. G. Barratt-Boyes. On stress reduction in bioprosthetic heart valve leaflets by the use of a flexible stent. J. Card. Surg. 6(4):476–481, 1991.
Fung, Y. C. Biomechanics: Mechanical Properties of Living Tissues. New York: Springer Verlag, 1993.
Gnyaneshwar, R., R. K. Kumar, and K. R. Balakrishnan. Dynamic analysis of the aortic valve using a finite element model. Ann. Thorac. Surg. 73(4):1122–1129, 2002.
Grube, E., J. C. Laborde, U. Gerckens, T. Felderhoff, B. Sauren, L. Buellesfeld, R. Mueller, M. Menichelli, T. Schmidt, B. Zickmann, S. Iversen, and G. W. Stone. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study. Circulation. 114(15):1616–1624, 2006; Epub 2006 Oct.
Grube, E., J. C. Laborde, B. Zickmann, U. Gerckens, T. Felderhoff, B. Sauren, A. Bootsveld, L. Buellesfeld, and S. Iversen. First report on a human percutaneous transluminal implantation of a self-expanding valve prothesis for interventional treatment of aortic valve stenosis. Catheter. Cardiovasc. Interv. 66(4):465–469, 2005.
Grube, E., G. Schuler, L. Buellesfeld, U. Gerckens, A. Linke, P. Wenaweser, B. Sauren, F. Mohr, T. Walther, B. Zickmann, S. Iversen, T. Felderhoff, R. Cartier, and R. Bonan. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome. J. Am. Coll. Cardiol. 1(50):69–76, 2007.
Iung, B., G. Baron, E. G. Butchart, F. Delahaye, C. Gohlke-Barwolf, O. W. Levang, P. Tornos, J. L. Vanoverschelde, F. Vermeer, E. Boersma, P. Ravaud, and A. Vahanian. A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart Disease. Eur. Heart J. 24(13):1231–1243, 2003.
Kim, H., J. Lu, M. S. Sacks, and K. B. Chandran. Dynamic Simulation of bioprosthetic heart valves using a stress resultant shell model. Ann. Biomed. Eng. 36(2):262–275, 2008.
Krucinski, S., I. Vesely, M. A. Dokainish, and G. Campbell. Numerical simulation of leaflet flexure in bioprosthetic valves mounted on rigid and expansile stents. J. Biomech. 26(8):929–943, 1993.
Langdon, S. E., R. Chernecky, C. A. Pereira, D. Abdulla, and J. M. Lee. Biaxial mechanical/structural effects of equibiaxial strain during crosslinking of bovine pericardial xenograft materials. Biomaterials 20(2):137–153, 1999.
Lee, J. M., and S. E. Langdon. Thickness measurements of soft tissue biomaterials: a comparison of five methods. J. Biomech. 29(6):829–932, 1996.
Libby, P., R. O. Bonow, D. P. Zipes, and D. L. Mann. Braunwald’s Heart Disease: A Textbook of Cardiovascular Medicine. Philadelphia, PA: Saunders, An imprint of Elsevier, 2007.
Lutter, G., D. Kuklinski, G. Berg, P. von Samson, J. Martin, M. Handke, P. Uhrmeister, and F. Beyersdorf. Percutaneous aortic valve replacement: an experimental study. I. Studies on implantation. J. Thorac. Cardiovasc. Surg. 123(4):768–776, 2002.
Marcheix, B., Y. Lamarche, C. Berry, A. Asgar, J. Laborde, A. Basmadjian, A. Ducharme, A. Denault, R. Bonan, and R. Cartier. Surgical aspects of endovascular retrograde implantation of the aortic CoreValve bioprosthesis in high-risk older patients with severe symptomatic aortic stenosis. J. Thorac. Cardiovasc. Surg. 5(134):1150–1156, 2007.
Movahed, M. Where are we going with percutaneous aortic valve replacement? Expert Rev. Cardiovasc. Ther. 6(5):997–998, 2007.
Sacks, M. S., and C. J. Chuong. Orthotropic mechanical properties of chemically treated bovine pericardium. Ann. Biomed. Eng. 26(5):892–902, 1998.
Sacks, M. S., and F. J. Schoen. Collagen fiber disruption occurs independent of calcification in clinically explanted bioprosthetic heart valves. J. Biomed. Mater. Res. 62(3):359–371, 2002.
Schultz, C. J., A. Weustink, N. Piazza, A. Otten, N. Mollet, G. Krestin, R. J. van Geuns, P. de Feyter, P. W. J. Serruys, and P. de Jaegere. Geometry and degree of apposition of the CoreValve ReValving System with multislice computed tomography after implantation in patients with aortic stenosis. J. Am. Coll. Cardiol. 54(10):911–918, 2009.
Simulia, Abaqus Analysis User’s Manual. Providence, RI, 2009.
Sun, W., A. Abad, and M. S. Sacks. Simulated bioprosthetic heart valve deformation under quasi-static loading. J. Biomech. Eng. 127(6):905–914, 2005.
Sun, W., and M. S. Sacks. Finite element implementation of a generalized Fung-elastic constitutive model for planar tissues. Biomech. Model. Mechanobiol. 4(2–3):190–199, 2005.
Sun, W., M. S. Sacks, T. L. Sellaro, W. S. Slaughter, and M. J. Scott. Biaxial mechanical response of bioprosthetic heart valve biomaterials to high in-plane shear. J. Biomech. Eng. 125:372–380, 2003.
Sung, H. W., Y. Chang, C. T. Chiu, C. N. Chen, and H. C. Liang. Crosslinking characteristics and mechanical properties of a bovine pericardium fixed with a naturally occurring crosslinking agent. J. Biomed. Mater. Res. 47(2):116–126, 1999.
US Market for Cardiac Surgery Devices, iData Research Inc., 2009.
Vesely, I. The evolution of bioprosthetic heart valve design and its impact on durability. Cardiovasc. Pathol. 12(5):277–286, 2003.
Webb, J. G., L. Altwegg, J.-B. Masson, S. A. Bugami, A. A. Ali, and R. A. Boone. A new transcatheter aortic valve and percutaneous valve delivery system. J. Am. Coll. Cardiol. 53(20):1855–1858, 2009.
Webb, J., S. Pasupati, K. Humphries, C. Thompson, L. Altwegg, R. Moss, A. Sinhal, R. Carere, B. Munt, D. Ricci, J. Ye, A. Cheung, and S. Lichtenstein. Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis. Circulation 7(116):755–763, 2007.
Zajarias, A., and A. G. Cribier. Outcomes and safety of percutaneous aortic valve replacement. J. Am. Coll. Cardiol. 53(20):1829–1836, 2009.
Acknowledgments
This research was supported in part by the State of Connecticut Department of Public Health Biomedical Research Grant DPH2010-0085 and the AHA SDG award 0930319N.
Author information
Authors and Affiliations
Corresponding author
Additional information
Associate Editor John H. Linehan oversaw the review of this article.
Rights and permissions
About this article
Cite this article
Li, K., Sun, W. Simulated Thin Pericardial Bioprosthetic Valve Leaflet Deformation Under Static Pressure-Only Loading Conditions: Implications for Percutaneous Valves. Ann Biomed Eng 38, 2690–2701 (2010). https://doi.org/10.1007/s10439-010-0009-3
Received:
Accepted:
Published:
Issue Date: