Background

Coronary computed tomography angiography (CCTA) is a robust non-invasive tool to exclude coronary artery disease (CAD) [1, 2]. Conversely, its performance is moderate in assessing the functional relevance of a coronary stenosis [3]. Documenting evidence of functional relevance of a coronary stenosis is recommended prior to any revascularization procedure to improve outcome [4,5,6]. Hence, functional information should complement pure anatomic characterization of a coronary stenosis for appropriate clinical decision-making [7]. Cardiac hybrid imaging is a method to assess coronary anatomy and function at the same time by combining CCTA and single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI). The added clinical and prognostic value of cardiac hybrid imaging has been demonstrated [8,9,10,11]. Despite these developments, there is growing interest in a single imaging modality that allows comprehensive morphological and functional assessment. Besides static and dynamic CT perfusion [12], recent studies have suggested that functional relevance of a coronary lesion can be estimated by fractional flow reserve (FFR) derived from CCTA (FFRCT) with high diagnostic accuracy [13] and positive impact on downstream resource utilization [14]. However, since the computation of FFRCT is a cumbersome process with limited availability, faster and less complex alternative CT-derived parameters have evolved, such as the decrease in corrected contrast opacification (CCO) across a stenosis and the transluminal attenuation gradient (TAG) along a coronary vessel [15,16,17,18]. Despite the widespread adoption of SPECT MPI into clinical routine and its utility in appropriate clinical decision-making [19, 20], none of the previous studies investigating CCO decrease and TAG has used SPECT MPI as the standard of reference. Therefore, the present study aims at evaluating the diagnostic accuracy of CCO decrease and TAG in comparison to SPECT MPI.

Methods

Study population

The present retrospective study consists of 72 consecutive patients who underwent hybrid CCTA/SPECT MPI and had at least one coronary stenosis (i.e., luminal diameter narrowing ≥ 50% as assessed in CCTA). Exclusion criteria were history of coronary artery bypass graft (CABG) or stenting.

CCTA acquisition and interpretation

On a stand-alone 64-slice CT scanner (LightSpeed VCT, GE Healthcare), patients underwent contrast-enhanced CCTA with prospective electrocardiography (ECG) triggering [21]. The following scanning parameters were applied: slice acquisition 64 × 0.625 mm, smallest X-ray window (only 75% of the RR-cycle), z-coverage value of 40 mm with an increment of 35 mm, gantry rotation time 350 ms, body mass index (BMI) adapted tube voltage (100 kV, BMI < 25 kg/m2; 120 kV, BMI ≥ 25 kg/m2), and effective tube-current (450 mA, BMI < 22.5 kg/m2; 500 mA, BMI 22.5–25 kg/m2; 550 mA, BMI 25–27.5 kg/m2; 600 mA, BMI 27.5–30 kg/m2; 650 mA, BMI > 30 kg/m2). Bolus tracking was performed with a region of interest (ROI) placed into the ascending aorta, and image acquisition was started 4 s after the signal density reached a predefined threshold of 120 Hounsfield units (HU). In order to achieve a target heart rate < 65 bpm, intravenous metoprolol (5–20 mg) was administered prior to the CCTA examination if necessary. Furthermore, all patients received 2.5 mg sublingual isosorbiddinitrate 2 min prior to the scan.

CCTA images were analyzed by consensus of two experienced readers with regard to morphologically significant lesions (≥ 50%). CCO decrease was measured for each stenosis as previously described [17]. In brief, a region of interest (ROI) with a diameter of 1 mm was placed in the center of the coronary lumen and a ROI with a diameter of 10 mm was placed in the descending aorta (on the same axial slice). CCO was calculated as the ratio of mean attenuation in the coronary ROI over the aorta ROI. CCO was measured twice within 2 cm proximal and distal of a lesion. The difference between the proximal and the distal CCO (using the lower of the two CCO values) was defined as CCO decrease (Fig. 1). If multiple lesions were present within one vessel, only the highest CCO decrease was used for further calculation of diagnostic accuracy.

Fig. 1
figure 1

Example of CCO decrease and TAG assessment. A soft plaque with high-grade stenosis is located in the proximal LAD (a, red arrow). Two coronary ROIs (a, small white circles) are placed proximal (b) and distal of a lesion (c). Corrected contrast opacification (CCO) is calculated as the ratio of mean attenuation in the coronary ROI over the ROI in the descending aorta (large white circles). Using the lower of the two values, the difference between the proximal (b) and distal (c) was defined as CCO decrease. The transluminal attenuation gradient (TAG) was measured by ROIs at 5-mm intervals from the ostium to the distal segment where the vessel cross-sectional area decreases below 2 mm2 (a, small black circles) and defined as the linear regression coefficient between intraluminal attenuation and distance from the ostium

TAG was measured in each major vessel with a coronary stenosis while side-branch vessels were not included in TAG assessment as previously described [22]. The luminal centerline was determined, and perpendicular cross-sectional images were reconstructed. At 5-mm intervals from the ostium to the distal level where the vessel cross-sectional area decreases below 2.0 mm2, a ROI with a diameter of 1 mm was manually positioned in the luminal center and mean HU were determined. In order to maintain linearity of the gradient, excessively calcified coronary segments were excluded as previously reported [22]. TAG was defined as the linear regression coefficient between intraluminal attenuation in HU and length from the ostium (Fig. 1).

SPECT MPI acquisition and interpretation

A 1-day ECG-gated stress/rest protocol was used with pharmacologic stress induced by infusion of adenosine at a standard rate of 140 μg/kg/min, and a BMI-adapted dose of 250 to 350 MBq 99mTc-tetrofosmin was injected 3 min into the pharmacologic stress [23]. After a delay of 60 min, stress images were acquired during 15 min. Immediately thereafter, a threefold higher dose 99mTc-tetrofosmin was administered and rest images were acquired during 15 min. In all patients, gated images were acquired with a dual-head camera (Millenium VG and Hawkeye or Ventri, both GE Healthcare) using standard acquisition parameters and X-ray-based attenuation correction. A commercially available software package (Cedars QGS/QPS, Los Angeles, CA, USA) was used for image analysis. Identification of ischemia and allocation to its corresponding coronary vessel was performed as previously reported [10]. In brief, SPECT myocardial tomograms were split into 20 segments for each patient. These segments were analyzed by consensus of two experienced readers using the following five-point scoring system: 0, normal; 1, equivocal; 2, moderate; 3, severe reduction of radioisotope uptake; and 4, absence of detectable tracer in a segment. A scan was scored as abnormal if two or more segments had stress scores ≥ 2. A reversible perfusion defect was categorized as one in which a stress defect was associated with a rest score ≤ 1 or a stress defect score of ≥ 4 with a rest score of 2.

Statistical analysis

Continuous variables are expressed as mean ± standard deviation (SD) or as median with interquartile range (IQR) if not normally distributed. Categorical variables are expressed as frequencies or percentages. Kolmogorov-Smirnov test was applied to assess normal distribution. P values for categorical variables were calculated by the chi-square test and for continuous variables by Mann-Whitney U test or Kruskal-Wallis test. We took into account the repeated structure of the measures and the hierarchical data structure (i.e., the fact that the segments and vessels were clusters of observations in the patients). Although generalized linear mixed modeling would be an excellent alternative, such analysis was restricted by the data structure (complexity due to multiple intrinsic hemodynamic and scan parameters) of the present study [24]. Receiver-operating characteristics (ROC) curve analysis was plotted to illustrate the performance of CCO decrease and TAG to diagnose ischemia. Youden’s index was calculated to define the optimal threshold. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of CCO decrease and TAG were assessed on a per-vessel basis, in case the ROC curve analysis was statistically significant. A p value < 0.05 was considered statistically significant. SPSS 20.0 (IBM Corporation, Armonk, NY) was used for analysis.

Results

Study population

The baseline characteristics of the study population are summarized in Table 1. The 72 patients had 183 coronary lesions in 127 vessels (108 major vessels and 19 side-branch vessels).

Table 1 Baseline characteristics

CCTA and SPECT findings

CCO decrease was successfully assessed in all 183 coronary lesions of the 127 vessels (100%). Distribution of CCO decrease according to stenosis severity is illustrated in Fig. 2a. Median CCO decrease was 0.094 (IQR, − 0.017 to 0.209), 0.128 (IQR, 0.005 to 0.205), and 0.530 (IQR, 0.405 to 0.678) for lesions with 50–69, 70–89, and 90–100% stenosis severity, respectively. While CCO decrease was significantly different across stenosis severities (p < 0.001), comparison between lesions of 50–69% and 70–89% stenosis severity revealed no significant difference (p = 0.470).

Fig. 2
figure 2

CCO decrease and TAG according to stenosis severity. Box-and-whisker plot showing corrected contrast opacification (CCO) decrease (a) and TAG (b) of all lesions in the categories of 50 to 69, 70 to 89, and 90 to 100% diameter stenosis. The red horizontal line corresponds to the cutoff value for an abnormal CCO decrease as used in the present study (CCO decrease > 0.168). The boxes represent the interquartile range (IQR) and the dark line in the middle is the median. The whiskers are defined as 1.5 times the IQR

TAG was successfully assessed in 106 major vessels with a coronary stenosis (98%). Distribution of TAG according to stenosis severity is illustrated in Fig. 2b. Median TAG was − 0.9 (IQR, − 1.3 to − 0.6), − 1.1 (IQR, − 1.8 to − 0.8), and − 1.7 (IQR, − 3.2 to − 1.0) for vessels with the most severe lesion of 50–69, 70–89, or 90–100% stenosis severity. While TAG was significantly different across stenosis severities (p < 0.05), comparison between lesions of 50–69 and 70–89% stenosis severity only trended towards a difference (p = 0.069).

Of 127 vessels with a coronary stenosis, 38 vessels (including four side-branch vessels) (30%) were subtended by ischemic myocardium. In these vessels, median CCO decrease was significantly higher compared to vessels subtended by non-ischemic myocardium (0.266 vs. 0.115; p < 0.001; Fig. 3a). In contrast, median TAG remained unchanged in vessels subtended by ischemic myocardium (− 1.1 HU/mm vs. − 1.1 HU/mm; p = 0.616; Fig. 3b).

Fig. 3
figure 3

CCO decrease and TAG in ischemic and non-ischemic myocardium. Box-and-whisker plot showing CCO decrease (a) and TAG (b) in ischemic and non-ischemic myocardium. The red horizontal line corresponds to the cutoff value for an abnormal CCO decrease as defined in the present study (CCO decrease > 0.168). The boxes represent the interquartile range (IQR) and the dark line in the middle is the median. The whiskers are defined as 1.5 times the IQR

Diagnostic accuracy of CCO decrease and TAG

The ROC curve analysis resulted in an area under the curve (AUC) of 0.707 (p < 0.001) for CCO decrease to predict ischemia (Fig. 4a). The optimal threshold of CCO decrease was defined at 0.168. Implementing the latter, an abnormal CCO decrease correctly detected ischemia in 28 of 38 vessels (sensitivity, 74%; 95% CI 57 to 87%) and correctly ruled out ischemia in 57 of 89 (specificity, 64%; 95% CI 53 to 74%) vessels. This resulted in a negative predictive value, positive predictive value, and accuracy of 85% (95% CI 74 to 93%), 47% (95% CI 34 to 60%), and 67%, respectively. After exclusion of lesions with 90–100% stenosis severity, the ROC curve analysis resulted in an AUC of 0.649 (p < 0.05). The optimal threshold of CCO decrease was unchanged. It correctly detected ischemia in 18 of 28 vessels (64%) and correctly ruled out ischemia in 55 of 82 vessels (67%). This resulted in a sensitivity, specificity, negative predictive value, positive predictive value, and accuracy of 64% (95% CI 44 to 81%), 67% (95% CI 56 to 77%), 85% (95% CI 74 to 92%), 40% (95% CI 26 to 56%), and 66%, respectively.

Fig. 4
figure 4

ROC curve of CCO decrease and TAG. While CCO decrease (a) significantly predicted ischemia with an AUC of 0.707 (p < 0.001) in the ROC curve analysis, TAG (b) did not (indicated by an AUC of 0.469; p = 0.616). The arrows in panel a indicate the cutoff value for an abnormal CCO. By adjusting the cutoff value from 0.168 (as defined in the present study) to 0.026, sensitivity would increase to 90% (at the cost of reducing specificity to 29%). On the contrary, a cutoff value at 0.367 would increase specificity to 90% (but decrease sensitivity to 42%)

The ROC curve revealed an AUC of 0.469 (p = 0.616; Fig. 4b), indicating that TAG is no predictor for ischemia.

Discussion

The present study demonstrates that an abnormal CCO decrease has an encouraging negative predictive value to exclude functionally relevant lesions conferring potential implication in the diagnostic work-up of patients with low pre-test probability for CAD which is the population benefiting most from CCTA. While median CCO decrease was significantly higher in vessels subtended by ischemic myocardium, TAG remained unchanged. Consequently, TAG did not predict ischemia in patients with a coronary stenosis.

While previous studies have reported on the diagnostic value of CCO decrease and TAG in comparison to thrombolysis in myocardial infarction (TIMI) flow, cardiac magnetic resonance imaging, and invasive FFR, our results are the first to compare their performance in the assessment of the functional relevance of a coronary stenosis with SPECT MPI. Considering the fact that invasive FFR was originally validated against SPECT MPI [25] and that there exists substantial mismatch between invasive FFR and SPECT [26], the direct comparison between CCO decrease and SPECT MPI extends our understanding of CCO decrease as a clinical tool. The diagnostic accuracy of CCO decrease in the present study is lower than that in the previous studies despite the comparable baseline characteristics [15,16,17]. Previous studies included substantially smaller study populations [15, 16] or used other techniques as a standard of reference (TIMI flow in a previous report [17] vs. SPECT MPI in the present study). Comparing CCO decrease to PET MPI, the ultimate gold standard for non-invasive assessment of myocardial blood flow has resulted in a diagnostic accuracy of 70% and a cutoff for an abnormal CCO decrease at 0.166 [18], matching well with the present study. Conversely, the literature is inconsistent concerning the diagnostic accuracy of TAG. While several reports have demonstrated an incremental value to diagnose functionally relevant lesions over anatomic CCTA findings [27,28,29], studies without single-beat acquisition rarely revealed a diagnostic benefit for TAG [30, 31]. Overall, the finding that differences in resting CCTA contrast densities may predict functional relevance is astounding, particularly in view of the evidence that an expected reduction in resting flow only occurs in lesions with a stenosis degree of 90% and above [32, 33]. The distribution of stenosis severity across the patient population, therefore, is key to the interpretation of previous studies and could potentially contribute to the inconsistencies in the literature. In the present study, however, if the analysis was limited to lesions below 90% diameter stenosis, diagnostic accuracy did only change marginally. Thus, CCO decrease seems to add moderate but significant value to the evaluation of intermediate lesions. This finding—that functional relevance of an anatomic coronary lesion can be assessed without requiring any hyperemic stress—is in line with previous studies measuring instantaneous wave-free ratio (iFR) [34]. Interestingly, iFR estimates coronary pressure during diastole when resting resistance is lowest—a timing interval similar to the one used by prospectively ECG-triggered CCTA [35].

A large body of evidence has documented that the presence of a functionally relevant stenosis impairs outcome, regardless by which modality the latter was assessed [4, 10, 19, 36, 37]. As a consequence, physiological assessment of stenosis relevance is critical in decision-making for appropriate treatment strategy as both revascularization of non-relevant lesions [6] and deferral from revascularization of hemodynamically relevant lesions may lead to a less favorable outcome [4, 5]. Although current guidelines mandate functional assessment for evidence-based revascularization [7], both non-invasive stress testing before invasive coronary angiography and FFR before percutaneous coronary intervention remain underused in daily practice [38, 39]. By its ease of use at no additional costs, CCO decrease may increase the probability for a comprehensive anatomic and functional non-invasive assessment before the patient is referred to invasive coronary angiography. Due to its moderate diagnostic accuracy, CCO decrease may not replace MPI, but it could potentially be positioned as a gatekeeper after CCTA to allocate further non-invasive diagnostic work-up. Although any test added to CCTA should ideally feature high positive predictive value, this does not apply to CCO decrease [16]. However, its encouraging negative predictive value and, as recently demonstrated, its added prognostic value [40] indicates a role to individualize clinical workflow: CCTA identifies a stenosis but if CCO decrease is normal, patients might be deferred from further MPI [41]. In contrast, if CCO decrease is abnormal over a stenosis, further non-invasive testing should be considered. In order to minimize the number of underdiagnosed patients, the cutoff value of an abnormal CCO decrease could be adjusted from 0.168 to lower values, e.g., a cutoff value at 0.026 would increase sensitivity to 90% (Fig. 4a). Through this workflow, downstream resource utilization might be influenced in a cost-effective manner. This may hold true in particular for the population which is typically referred for CCTA, namely patients with low to intermediate pre-test probability.

We acknowledge the following limitations. First, the present study is a retrospective single-center study and has all of the inherent limitations of that study design. Second, the threshold for CCO decrease as defined in the present study should be extrapolated with caution to different patient populations since differences not only in baseline characteristics but also in scan protocol and post-processing might influence analysis of CT images [42, 43]. Third, it cannot be excluded that the lack of a diagnostic value of TAG was due to the inferior performance of TAG with 64-slice scanners [27, 30] compared to wide-volume scanners [28, 44]. With the wider distribution of 256-slice CT scanners [45], future studies may assess the diagnostic value of TAG without this limitation.

Conclusions

CCTA-derived CCO decrease but not TAG predicts ischemia in SPECT MPI. The negative predictive value of CCO decrease of 85% may confer clinical implications in the diagnostic work-up of patients with a coronary stenosis.