Consumption of fruits leads to increase in glucose level in blood for diabetic patients, which in turn leads to peripheral, vascular, ocular complications and cardiac diseases. In this context, a non-enzymatic hybrid glucose biosensor was fabricated for the first time to detect glucose by immobilizing titanium oxide–manganese oxide (TiO2–Mn3O4) nanocomposite and chitosan membrane on to the surface of Pt working electrode (Pt/TiO2–Mn3O4/chitosan). TiO2–Mn3O4 nanocomposite catalyzed the oxidation of glucose to gluconolactone in the absence of glucose oxidase enzyme with high electron transfer rate, good biocompatibility and large surface coverage. Electrochemical measurements revealed the excellent sensing response of the developed biosensor towards glucose with a high sensitivity of 7.073 µA mM−1, linearity of 0.01–0.1 mM, low detection limit of 0.01 µM, reproducibility of 1.5% and stability of 98.8%. The electrochemical parameters estimated from the anodic process were subjected to linear regression models for the detection of unknown concentration of glucose in different fruit samples.
Similar content being viewed by others
Introduction
Glucose, a monomer with the molecular weight of 180.16 g, is the most significant source of energy (Ameen et al. 2016) in living organisms. The blood glucose level in normal person ranges between 70 and 80 mg dL−1 (Gao et al. 2015; Heli and Amirizadeh 2016). The increase in the amount of glucose in blood results in type 1 and type 2 diabetes (Kim et al. 2015; Hsu et al. 2016; Li et al. 2016). The consumption of fruits raises blood glucose levels in diabetic patients, which in-turn leads to various complications including peripheral, vascular, ocular and cardiac diseases (Ameen et al. 2016; Chou et al. 2016). In this context, development of a biosensor for the quantification and detection of glucose in fruit samples will add a significant value to the field of biosensors.
Most of the electrochemical glucose biosensors are based on the glucose oxidase enzyme, which electrochemically catalyzes glucose into gluconic acid and hydrogen peroxide (H2O2) (Lu et al. 2016; Samdani et al. 2016; Vargas et al. 2016). The concentration of glucose in blood serum of patients with type I and type II diabetes can be calculated from the current density of enzymatically released H2O2 (Xia et al. 2014; Kim et al. 2015; Zhang et al. 2016). Compared to electrochemical glucose oxidase enzyme-based biosensors, non-enzymatic glucose biosensors based on metal oxide nanoparticles offer good stability against denaturation, easy to fabricate and facilitate effective glucose oxidase enzyme like catalysis over a wide temperature and pH ranges (Eguílaz et al. 2015; Heli and Amirizadeh 2016; Li et al. 2016). For these reasons, electrode materials showing enhanced catalytic activity are very important to attain extremely sensitive detection of glucose. Notably, NiCo2O4–polyaniline, Ag–CNT, Ni–CNT, graphene–CNT, graphene–polypyrrole, MnO2–rGO, NiO–CNT, TiO2–MWCNT, CdTe–CdS, graphene–CdS, PtM–CNT, Ni–carbon nanofiber and Pt-polypyrrole have been used as nanohybrid materials in the determination of glucose (Govindhan et al. 2015; Baghayeri et al. 2016; Yu et al. 2016). Recently, nanostructured TiO2 and Mn3O4 have attracted enormous research interest in the design of electrochemical biosensors owing to their large electroactive surface area, electrocatalytic activity and enhanced electron transfer rate (Hu et al. 2011; Reza et al. 2014). Reza et al. (2014) developed xanthine biosensor by immobilizing Mn3O4–chitosan nanocomposites on indium tin oxide substrate in which Mn3O4 nanoparticles exhibited good biocompatibility, high carrier mobility and efficient catalytic activity. Similarly, Wang et al. (2015) developed hydrogen peroxide (H2O2) biosensor by immobilizing Ti3C2 MXene nanocomposite on glassy carbon electrode in which TiO2 enhanced the electron transfer rate of hemoglobin and facilitated the electrocatalysis of H2O2. To enhance the electrocatalysis of glucose in the absence of glucose oxidase enzyme, a hybrid interface of TiO2 and Mn3O4 was preferred as an electrocatalyst. To the best of our knowledge, electrochemical glucose biosensor based on TiO2 and Mn3O4 hybrid interface has been reported for the first time, that too for glucose level detection in fruits. Hence, in the present work, a non-enzymatic hybrid interfaced electrochemical biosensor was developed for glucose detection in different fruits by modifying platinum (Pt) electrode with titanium oxide (TiO2)–manganese oxide (Mn3O4) nano interface and chitosan membrane.
Experimental
Materials
Mannitol, xylose, sucrose, starch, maltose, galactose and fructose were purchased from HiMedia Laboratories, India. TiO2 and Mn3O4 nanomaterials were procured from Sigma-Aldrich, USA. Chemicals, namely glucose, lactic acid, potassium hydroxide, sodium hydroxide 0.5 wt% chitosan in 1% acetic acid (degree of deacetylation of 82.5%, molecular weight 140,000 g mol−1), dibasic sodium phosphate dihydrate, ascorbic acid, and monobasic sodium phosphate monohydrate were purchased from Merck India Ltd., India. Cadmium acetate dihydrate was procured from Loba Chemie Pvt. Ltd., India. All other chemicals, namely cupric acetate, nickel chloride, urea were procured from Thermo Fisher Scientific Pvt. Ltd., India. Pt wire which was used as counter electrode (CHI115, 0.5 mm diameter), KCl saturated Ag/AgCl which was used as reference electrode (CHI111, 0.5 mm diameter) and Pt used as working electrode (CHI102, 2 mm diameter) were procured from CH Instruments, Inc. Deionized water was used to prepare all solutions and reagents (AQUA Purifications Systems, India).
Fabrication of Pt/TiO2–Mn3O4/chitosan modified working electrode
TiO2 and Mn3O4 nanoparticles were taken in the ratio of 1:1 and ground into fine powder in a mortar and pestle. Later, a novel composite of TiO2 and Mn3O4 nanoparticles was prepared by ultrasonically mixing 1:1 ratio of TiO2 and Mn3O4 nanoparticles in 100 mL PBS containing 10 µL of chitosan for 1 h. The Pt/TiO2–Mn3O4/chitosan modified bioelectrode was fabricated by drop casting 3 µL of this suspension on to the exposed surface of Pt working electrode.
Characterization techniques
The morphologies of TiO2, Mn3O4 and TiO2–Mn3O4 nanocomposites were observed using field emission scanning electron microscope (FE-SEM, Model JSM 6701F, JEOL, Japan). Image J 1.48q software was used to study the size distribution of TiO2, Mn3O4 and TiO2–Mn3O4 nanocomposites. All linear sweep voltammetric experiments were carried out at ambient temperature with a conventional three-electrode system comprising a Pt/TiO2–Mn3O4/chitosan, a Pt wire and Ag/AgCl saturated in 0.1 M KCl was taken as working, counter and reference electrodes, respectively.
Preparation of real samples
For the estimation of glucose in banana (Musa acuminate), strawberry (Fragaria ananassa), apple (Malus domestica), grape (Vitis vinifera) and pomegranate (Punica granatum), samples are as follows: (1) initially, the flesh of fruit samples was ground using mortar and pestle. (2) 20 mg flesh of each sample was dispersed in 100 mL PBS solution and kept for centrifugation at 6000 rpm for 1 h. Later, 100 mL of deionized water was added to 0.4 mL of the supernatant for dilution. To investigate the performance of the Pt/TiO2–Mn3O4/chitosan bioelectrode in fruit samples, Pt/TiO2–Mn3O4/chitosan bioelectrode was exposed to recovery experiments by spiking various concentrations of glucose in 0.1 M PBS. Later, a calibration curve was plotted using the known concentration of glucose vs current density. Finally, the unknown glucose concentration in the flesh of fruit samples can be estimated by comparing the measured current density with the calibration curve.
Results and discussion
Morphological studies
The morphologies of TiO2, Mn3O4 and TiO2–Mn3O4 nanocomposites were characterized using FE-SEM (Fig. 1a–c). The spherical-like morphologies of TiO2 and Mn3O4 nanoparticles were observed from FE-SEM studies. From the FE-SEM image of TiO2–Mn3O4 nanocomposite, it was confirmed that TiO2 nanospheres were closely anchored and uniformly distributed on the surface of Mn3O4 nanospheres. The average diameters of TiO2, Mn3O4 and TiO2–Mn3O4 were estimated to be 275 ± 4.7, 325 ± 2.4 and 342 ± 3.7 nm, respectively.
Electrochemical characterization of various modified Pt electrodes
Linear sweep voltammetric response of bare Pt, Pt/TiO2/chitosan, Pt/Mn3O4/chitosan and Pt/TiO2–Mn3O4/chitosan electrodes in pH 7.4 PBS (0.1 M) having 0.01 mM glucose was executed at a scan rate of 0.1 V s−1 are shown in Fig. 2a. Comparison of electrochemical parameters for various modified Pt electrodes is shown in Table 1. The bare Pt electrode showed no anodic peak current. On adding Mn3O4 and TiO2 nanospheres individually to the surface of Pt electrode, an anodic peak current was observed indicating oxidation of glucose to gluconolactone. The Pt/TiO2–Mn3O4/chitosan electrode exhibited increased peak current in the anodic process when compared with Pt/Mn3O4/chitosan and Pt/TiO2/chitosan electrodes. This can be ascribed to the synergic effect of TiO2 and Mn3O4 nanospheres in TiO2–Mn3O4 nanocomposite. It also depicted that the TiO2–Mn3O4 nanocomposite was well immobilized on to the surface of Platinum working electrode and provided necessary electron conduction pathways between Platinum working electrode and surface confined glucose biomolecules.
The anodic peak potential (E pa) of Pt/TiO2–Mn3O4/chitosan electrode (117 mV) was lesser than the E pa values of Pt/TiO2/chitosan (130 mV) and Pt/Mn3O4/chitosan (153 mV) electrodes. The lower E pa value signified that the TiO2–Mn3O4 nanocomposites decreased the working potential needed for the oxidation of glucose to gluconolactone.
The electron transfer rate constant of different modified Pt electrodes was calculated using Eq. (1):
where \(I_{\text{pa}}\) is the anodic peak current and \(Q\) is the amount of charge consumed. The electron transfer rate constant of TiO2–Mn3O4 nanocomposites immobilized on the surface of Platinum working electrode in the anodic process was calculated as 3.35 s−1, which was higher than the value reported for TiO2 (K s = 2.42 s−1) and Mn3O4 nanospheres (K s = 1.44 s−1) immobilized on the surface of Platinum working electrode. The higher K s value depicted the enhanced electron transfer rate by TiO2–Mn3O4 nanocomposites during the anodic process.
The surface coverage of adsorbed electroactive glucose biomolecule on various modified Pt electrodes was calculated using Eq. (2):
where \(n_{\text{a}}\) is the number of electrons involved in the charge transfer step,\(T\) is the room temperature, \(R\) is the gas constant, \(A\) is the area of Pt working electrode, \(\varGamma\) is the surface density of reactant species, \(I_{\text{p}} \left( {{\text{irrev}}.} \right)\) is the irreversible peak current, \(F\) is the Faraday’s constant, \(n\) is the number of electrons in the redox reaction, \(\alpha\) is the transfer coefficient and \(\nu\) is the scan rate. The estimated surface coverage of electroactive glucose biomolecule on Pt/TiO2–Mn3O4/chitosan bioelectrode surface (1.54 × 10−1 mol cm−2) was higher than those observed for other modified electrodes (Table 1). Owing to the enhanced electron transfer rate, high surface coverage of glucose and enhanced oxidation current, Pt/TiO2–Mn3O4/chitosan electrode was chosen for further electrochemical analysis.
Effect of scan rate
The electrochemical response of Pt/TiO2–Mn3O4/chitosan electrode in PBS (pH 7.4, 0.1 M) containing 0.01 mM glucose with the effect of scan rate is shown in Fig. 2b. From Fig. 2b, one can observe the applied scan rate-dependent anodic peak current. The anodic peak was shifted towards negative potential with the increased scan rate, which might be due to the electron transfer limitation of glucose at the surface of Pt/TiO2–Mn3O4/chitosan electrode. The anodic current (I pa (µA) = 0.131 [ν] (µA/V s−1) + 0.116) increased linearly with the scan rate from 0.01 to 0.1 V s−1 (Fig. 2c). The linear dependency between scan rate and oxidation current indicated that electroactive glucose biomolecule adsorbed onto the surface of Pt/TiO2–Mn3O4/chitosan electrode has experienced a surface confined electron transfer.
Effect of glucose
The electroanalytical performance of Pt/TiO2–Mn3O4/chitosan electrode was investigated towards glucose by successive addition of glucose into electrochemical cell containing 0.1 M PBS (pH 7.4). The intensity of anodic current response increased with an increase of glucose concentration from 0.01 to 0.1 mM (Fig. 2d). The current response of the developed Pt/TiO2–Mn3O4/chitosan electrode was linear (I pa (µA) = 7.073 [glucose] (µA mM−1) + 17.468) towards glucose in the concentration range between 0.01 and 0.1 mM with the correlation coefficient (r) of 0.99. Sensitivity and response time of the developed bioelectrode were estimated to be 7.073 µA mM−1 and less than 5 s, respectively. The enhanced sensitivity can be attributed to the large specific surface area of the TiO2–Mn3O4 nanocomposites, which allows the oxidation of substantial amount of glucose. The developed glucose biosensor displayed detection and quantification limits of 0.01 and 0.03 µM, respectively. The Michaelis–Menten constant estimated using Lineweaver–Burk plot was 32 µM, which indicated the good affinity between TiO2–Mn3O4 nanocomposites and surface confined glucose biomolecules. From the Hill plot, the degree of cooperativity was calculated as 2.1. The estimated degree of cooperativity was greater than 1, which indicated that glucose bound on the surface of TiO2–Mn3O4 nanocomposites drives the other incoming glucose molecules to bind to the surface of the developed electrode.
Model fitting and validation
Table 2 shows the electrochemical parameters estimated from the anodic process for various glucose concentrations. Added glucose was considered as dependent variable and other electrochemical parameters as independent variables (K s, I pa, Γ, αn and Q). K s vs [glucose], I pa vs [glucose], Γ vs [glucose], αn vs [glucose] and Q vs [glucose] models exhibited linear behavior. Therefore, in this work, glucose quantification and detection in different fruit samples were performed using linear regression models.
Table 3 shows estimated parameters for various linear regression models. ANOVA results indicated the significance of the proposed linear regression models at 95% confidence limit. R 2 value obtained for the five linear regression models was greater than 99%, which proved the predictive ability of these five models. Root mean square error for cross validation (RMSECV), relative prediction error (RPE), % recovery and correlation (r) were compared for the predicted and added concentrations of glucose, to assess the accuracy of the five proposed linear regression models. Only the Q vs [glucose] model showed (RMSECV = 6.917 × 10−4, r = 0.99 and % recovery = 99.283) best results in validation, whereas K s vs [glucose], I pa vs [glucose], Γ vs [glucose] and αn vs [glucose] models showed poor recovery, RMSECV and RPE. Based on the error analysis, Q vs [glucose] model was used for the estimation of glucose in fruit samples (Fig. 3a).
Precision and accuracy studies
The repeatability and reproducibility of Pt/TiO2–Mn3O4/chitosan electrode were studied by measuring anodic current in 0.1 mM of glucose using eight different Pt/TiO2–Mn3O4/chitosan bioelectrodes. The relative standard deviation (RSD) of linear sweep voltammetric responses for intra-assay and inter-assay analysis was 1.2 and 1.5%, respectively. The observed results showed that the embedding between chitosan and TiO2–Mn3O4 nanocomposite was good. The glucose concentration ± relative error values for the assay of 0.01, 0.05 and 0.1 mM glucose added into 0.1 M PBS (pH 7.4) were 0.01 ± 1.2 × 10−4, 0.05 ± 1.1 × 10−4 and 0.01 ± 1.2 × 10−4 mM for five measurements, suggesting good accuracy of the glucose biosensor.
Interferent and stability studies
To investigate the interference of biomolecules with Pt/TiO2–Mn3O4/chitosan electrode, interferents, namely mannitol, xylose, sucrose, starch, maltose, galactose, fructose, lactic acid, ascorbic acid, urea, Ca2+, Zn2+, Cd2+ and Ni2+ at a relatively high concentration of 0.1 mM were added to electrochemical cell containing 0.1 mM glucose. No significant change in current response was observed (I(%) ≤ 2.15) in the presence of interferents like mannitol (I(%) = 0.25), xylose (I(%) = 1.32), sucrose (I(%) = 2.15), starch (I(%) = 1.58), maltose (I(%) = 1.76), galactose (I(%) = 0.74), fructose (I(%) = 2.12), lactic acid (I(%) = 1.14), ascorbic acid (I(%) = 1.48), urea (I(%) = 0.58), Ca2+ (I(%) = 2.31), Zn2+ (I(%) = 2.67), Cd2+ (I(%) = 1.41) and Ni2+ (I(%) = 1.54) indicating the ability of developed electrode to overcome tested interferents. The Pt/TiO2–Mn3O4/chitosan electrode was preserved in PBS (0.1 M, pH 7.4) at room temperature after each linear sweep voltammetric measurement. It was observed that the oxidation current response decreased by about 1.2% after 10 days. These results indicated the good stability and reproducibility of the Pt/TiO2–Mn3O4/chitosan electrode. It also indicated that the fabrication technique and embedding between TiO2–Mn3O4 nanocomposite and chitosan were good.
Glucose detection in different fruit samples
The concentration of glucose in the flesh of fruit samples was estimated using the calibrated Q vs [glucose] model (Fig. 3b) and Eq. (3),
To examine the potential application of the developed Pt/TiO2–Mn3O4/chitosan electrode in the flesh of fruit samples, the extract of flesh of different fruits was diluted in the ratio of 1:10, 1:100 and 1:1000. From the calibration curve, the amount of glucose present in the banana, strawberry, apple, grape and pomegranate samples was found to be 0.02, 0.03, 0.02, 0.04 and 0.03 mM. The observed results demonstrated that the TiO2–Mn3O4 nanocomposites can be used as a novel sensing material for quantification of glucose in different fruit samples.
Conclusion
In the present work, an enzyme-free glucose biosensor was fabricated by immobilizing TiO2–Mn3O4 nanocomposite on the surface of Pt working electrode. The TiO2–Mn3O4 nanocomposites catalyzed the oxidation of glucose to gluconolactone in the absence of glucose oxidase enzyme. The TiO2–Mn3O4 enhanced the electron transfer rate between adsorbed glucose and Pt working electrode and, thereby, minimized the working potential needed for the oxidation of glucose to gluconolactone. The observed results are encouraging; hence, in future, this sensor can also be implemented for the detection and quantification of wide range of glucose in different food samples, packaged food, beverages and also in blood serum.
References
Ameen S, Akhtar MS, Shin HS (2016) Nanocages-augmented aligned polyaniline nanowires as unique platform for electrochemical non-enzymatic glucose biosensor. Appl Catal A Gen 517:21–29
Baghayeri M, Amiri A, Farhadi S (2016) Development of non-enzymatic glucose sensor based on efficient loading Ag nanoparticles on functionalized carbon nanotubes. Sens Actuators B Chem 225:354–362. doi:10.1016/j.snb.2015.11.003
Chou J-C, Chen J-S, Liao Y-H et al (2016) Effect of different contents of magnetic beads on enzymatic IGZO glucose biosensor. Mater Lett 175:241–243. doi:10.1016/j.matlet.2016.04.050
Eguílaz M, Villalonga R, Pingarrón JM et al (2015) Functionalization of Bamboo-like carbon nanotubes with 3-Mercaptophenylboronic acid-modified gold nanoparticles for the development of a hybrid glucose enzyme electrochemical biosensor. Sens Actuators B Chem 216:629–637. doi:10.1016/j.snb.2015.03.112
Gao ZF, Chen DM, Lei JL et al (2015) A regenerated electrochemical biosensor for label-free detection of glucose and urea based on conformational switch of i-motif oligonucleotide probe. Anal Chim Acta 897:10–16. doi:10.1016/j.aca.2015.09.045
Govindhan M, Amiri M, Chen A (2015) Au nanoparticle/graphene nanocomposite as a platform for the sensitive detection of NADH in human urine. Biosens Bioelectron 66:474–480. doi:10.1016/j.bios.2014.12.012
Heli H, Amirizadeh O (2016) Non-enzymatic glucose biosensor based on hyperbranched pine-like gold nanostructure. Mater Sci Eng C 63:150–154. doi:10.1016/j.msec.2016.02.068
Hsu C-W, Su F-C, Peng P-Y et al (2016) Highly sensitive non-enzymatic electrochemical glucose biosensor using a photolithography fabricated micro/nano hybrid structured electrode. Sens Actuators B Chem 230:559–565. doi:10.1016/j.snb.2016.02.109
Hu L, Huo K, Chen R et al (2011) Recyclable and high-sensitivity electrochemical biosensing platform composed of carbon-doped TiO2 nanotube arrays. Anal Chem 83:8138–8144. doi:10.1021/ac201639m
Kim J-H, Lee D, Bae T-S, Lee Y-S (2015) The electrochemical enzymatic glucose biosensor based on mesoporous carbon fibers activated by potassium carbonate. J Ind Eng Chem 25:192–198. doi:10.1016/j.jiec.2014.10.034
Li W, Qian D, Wang Q et al (2016) Fully-drawn origami paper analytical device for electrochemical detection of glucose. Sens Actuators B Chem 231:230–238. doi:10.1016/j.snb.2016.03.031
Lu W, Sun Y, Dai H et al (2016) Direct growth of pod-like Cu2O nanowire arrays on copper foam: highly sensitive and efficient nonenzymatic glucose and H2O2 biosensor. Sens Actuators B Chem 231:860–866. doi:10.1016/j.snb.2016.03.058
Reza KK, Singh N, Yadav SK et al (2014) Pearl shaped highly sensitive Mn3O4 nanocomposite interface for biosensor applications. Biosens Bioelectron 62:47–51. doi:10.1016/j.bios.2014.06.013
Samdani KJ, Samdani JS, Kim NH, Lee JH (2016) FeMoO4 based, enzyme-free electrochemical biosensor for ultrasensitive detection of norepinephrine. Biosens Bioelectron 81:445–453. doi:10.1016/j.bios.2016.03.029
Vargas E, Ruiz MA, Campuzano S et al (2016) Non-invasive determination of glucose directly in raw fruits using a continuous flow system based on microdialysis sampling and amperometric detection at an integrated enzymatic biosensor. Anal Chim Acta 914:53–61. doi:10.1016/j.aca.2016.02.015
Wang F, Yang C, Duan M et al (2015) TiO2 nanoparticle modified organ-like Ti3C2 MXene nanocomposite encapsulating hemoglobin for a mediator-free biosensor with excellent performances. Biosens Bioelectron 74:1022–1028. doi:10.1016/j.bios.2015.08.004
Xia L, Song J, Xu R et al (2014) Zinc oxide inverse opal electrodes modified by glucose oxidase for electrochemical and photoelectrochemical biosensor. Biosens Bioelectron 59:350–357. doi:10.1016/j.bios.2014.03.038
Yu Z, Li H, Zhang X et al (2016) Facile synthesis of NiCo2O4@Polyaniline core–shell nanocomposite for sensitive determination of glucose. Biosens Bioelectron 75:161–165. doi:10.1016/j.bios.2015.08.024
Zhang E, Xie Y, Ci S et al (2016) Porous Co3O4 hollow nanododecahedra for nonenzymatic glucose biosensor and biofuel cell. Biosens Bioelectron 81:46–53. doi:10.1016/j.bios.2016.02.027
Acknowledgements
The authors are grateful to the Department of Science and Technology, New Delhi, for their financial support [SR/NM/PG-16/2007, DST/TSG/PT/2008/28, SR/FST/ETI-284/2011(C), SR/FST/LSI-453/2010 and DST/TM/WTI/2K14/197(a)(G)]. Sasya Madhurantakam wish to acknowledge DST for providing INSPIRE fellowship (IF120812). Jayanth Babu wish to acknowledge DST-SERB for ECR Grant (File No. ECR/2016/001805). We also acknowledge SASTRA University, Thanjavur for extending infrastructural support to carry out the study.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Jayanth Babu, K., Sasya, M., Nesakumar, N. et al. Non-enzymatic detection of glucose in fruits using TiO2–Mn3O4 hybrid nano interface. Appl Nanosci 7, 309–316 (2017). https://doi.org/10.1007/s13204-017-0571-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13204-017-0571-1