Introduction

A recently discovered human-transferrable virus called coronavirus 2 (SARS-CoV-2) causes severe respirational syndrome. Globally, hundreds of millions of SARS-CoV-2 positive cases have been found, killing millions of people. The human-transferrable SARS-CoV-2 virus outbreak was announced as a worldwide pandemic by the WHO. The SARS-CoV-2 virus typically consists of a single-stranded RNA and four important proteins: (1) nucleocapsid (N) protein, (2) membrane (M), (3) envelope (E) and (4) spike (S) glycoprotein [1]. The S-glycoprotein can be divided into two types of subunits (S1 and S2), where S1 binds with the host cell receptor human angiotensin-converting enzyme 2 (ACEZ2) and S2 is liable for membrane fusion [2]. The human heart, kidneys, lungs, and other organs all contain ACEZ2, which enables the viral spike protein to enter cells [1]. The brain, lungs, heart, and kidneys experience significant damage as a result of the viral spike protein's activation [3]. As a result, clinical research on the development of vaccines and detection methods to stop the spread of the infectious disease must focus on neutralizing the SARS-CoV-2 virus spike protein. In clinical studies, dozens of vaccines have been developed to neutralize the SARS-CoV-2 viral spike's receptor-binding domain [4]. More than 90% of SARS-CoV-2 vaccinations are effective against the virus [4, 5]. COVAX has been trying to accelerate the creation of SARS-CoV-2 vaccines, their commercial production, and their fair distribution [4]. However, SARS-CoV-2 viruses are expanding very rapidly. Novel variants of coronavirus have been discovered with altered mutations [6]. Early identification is essential for controlling the pandemic since the vaccine is not commercially available everywhere due to economic situations. To effectively monitor infected individuals for successful quarantine and prompt treatment, it is crucial to have extremely sensitive, quick-test findings, and affordable analytical techniques. The SARS-CoV-2 virus is currently detected using the real-time reverse-transcriptase-polymerase chain reaction (RTPCR) technology as a standard reference. Results from the RTPCR method are obtained in 1–3 days [7]. The lengthy extraction of virus ribonucleic acid (RNA) has an impact on the accuracy of detection. Zhao et al. developed a pcMNPs methodology for virus RNA extraction [8] to overcome the lengthy RNA extraction process. The sensitivity of the RTPCR technique varies between 45 and 60% for RNA extraction whereas employing pcMNPs increases the sensitivity to about 92% [7]. The COVID-19 IgG/IgM fast test kit is utilized in clinical settings to identify antibodies (IgG/IgM), with IgG becoming detectable in an affected patient blood after 3-6 days and IgM becoming detectable after 8 days [9, 10]. Although less sensitive than the RTPCR method, the SARS-CoV-2 antigen fast test kit is still applied to identify the COVID-19 virus [11, 12]. Moreover, the chest computed tomography (CT) scan is checked to corroborate the test results. If the fast test kit is not handled properly, it produces false findings [13].

Surface plasmon sensors (SPREs)-based biosensors have been proposed for future COVID-19 viral detections [14,15,16,17,18]. Generally, SPRE-based sensors are characterized by rapid detection, accuracy and high sensitivity [19,20,21,22].

For the past three decades, SPRE has been a widely used method to identify biomolecular interactions for clinical applications [23]. Several SPRE-based sensors have been employed in biosensing applications, including localized and compact SPRE biosensors [24]. There is no essential difference in the principle of operation of SPRE-based biosensors. Depending on light coupling, the SPRE structures can be classified into four sets: grating-based, fiber optic-based, waveguide-based and prism-based. Researchers preferred prism-based SPRE structure due to its accurate realization [25]. There are two possible setups for the prism-based SPRE sensor: (1) Kretschmann and (2) Otto. Kretschmann configuration is preferred for experimental work with superior results using the angular interrogation method due to less absorption loss [26, 27]. The stimulation of the surface plasmons (SPNs) can be applied to the SPRE sensors for producing the surface plasmon waves (SPWEs) on the metallic film by applying the attenuated total reflection method [28]. The SPWEs oppose the light energy of the plasmons, and that light energy is translated as reflected light in the total internal reflection principle. At the metal-dielectric contact, the SPWEs transpire the evanescent wave (EWE) whose intensity exponentially decays in the direction perpendicular to the interface. The resonance oscillation of the SPWEs is called the SPRE phenomenon which shows the minimum reflectivity. The reflectance profile depends on the surface refractive index (RIX). The resonance dip of the reflected radiation varies depending on any variations in the RIX of the materials utilized in the construction of the SPRE biosensor. The standard SPRE sensors achieve broader SPRE curves and less resonance angle change. The ultimate goal of researchers is to establish sharp curves and higher resonance angle changes to improve the sensor performance [29]. As a result, many researchers have inserted dielectric layers over the metal layer to enhance SPRE performance and provide protection for the metal film from oxidation [30]. Due to its enormous potential, high electric field absorption, and tiny gap, BiFeO3 is a promising material that has received a lot of attention [31]. Although the SPRE performance can be enhanced by using dielectric materials, it is difficult to maintain its optical properties and complex geometry. In order to mitigate this issue, researchers have proposed hybrid structures incorporating bimolecular recognizing elements (BRETs), which can provide a high SPRE performance [32]. Though, SPRE biosensors with good performance are still lacking. Several interesting two-dimensional (2D) transition metal di-chalcogenides (TMDEs) materials are available to fulfill this challenge. Due to its unique optical and electrical properties, the graphene family has attracted a lot of attention recently [33, 34]. In recent work, graphene-MoS2 layers have been used to design a biosensor for biomolecule detection. The sensitivity was observed to enhance when an additional nanolayer of silicon is used [35,36,37,38,39,40].

Despite the fact that several SPRE biosensors have already been developed by other researchers, further research is still necessary to enhance the performance characteristics and advance the development of sensors.

The novelty of the current work lies in the fact that SPRE-based sensors are characterized by rapid detection, accuracy and high sensitivity. So far, very few articles have been published on using SPRE-based sensors for the detection of SARS-CoV-2 virus. The current paper proposes an efficient structure based on SPRE for the detection of this dangerous virus that can cause severe respirational syndrome.

This work reports a novel SPRE sensor that can efficiently detect SARS-CoV-2 virus. The hybrid structure comprises a BK7 prism, Ag, BiFeO3, graphene and sensing medium. The performance of the SPRE biosensor is evaluated in terms of the full-width at half maximum (FWHM), detection accuracy, sensitivity and figure of merit. At a wavelength of 633 nm, the sensor performance is numerically examined. The current paper is structured as follows. The theoretical and numerical modeling of the suggested SPRE biosensor is explained in section "Structure Considerations". Section "Results and Discussion" reports the results obtained and the discussion. The present study is finally concluded in section "Conclusion".

Structure Considerations

Model Design

A biosensor based on SPRE is proposed (Fig. 1). It has five layers: BK7 prism, Ag, BiFeO3, graphene, and analyte layer. The thicknesses and RIXs of the silver layer, BiFeO3, and graphene are denoted as dAg, dBF and dG and nAg, nBF and nG, respectively.

Fig. 1
figure 1

SPRE device based on BiFeO3-graphene layers.

The RIX of BK7 prism is given by [41]

$${n}_{p}\left(\lambda \right)=\sqrt{1+\frac{1.03961212 {\lambda }^{2}}{{\lambda }^{2}-0.00600069867}+\frac{1.01046945 {\lambda }^{2}}{{\lambda }^{2}-103.560653}+\frac{0.231792344 {\lambda }^{2}}{{\lambda }^{2}-0.0200179144}}$$
(1)

The Ag RIX is obtained as (Drude-Lorentz model) [42]

$${n}_{Ag}\left(\lambda \right)=1-\frac{{\lambda }_{c} {\lambda }^{2}}{{{\lambda }_{p}}^{2}\left({\lambda }_{c}+i\lambda \right)}$$
(2)

with \({\lambda }_{p}=0.1454\; \mu m\) and \({\lambda }_{c}=0.1761\; \mu m\) are the plasma and collision wavelengths of the Ag material. The RIX of BiFeO3 is 2.9680.

The graphene RIX is given by [43]

$${n}_{G}\left(\lambda \right)=3+\lambda \frac{C}{3}i,\mathrm{ C }= 5.446\; {\mu m}^{-1}$$
(3)

Concerning the sensing medium, phosphate-buffered solution (PBSN) may be utilized as a RIX of PBSN that is deliberated as ns = 1.334 + ∆ns, where ∆ns is the index variation as a result of analyte-ligand interaction on the sensor surface. According to earlier research [44], a PBSN buffer containing SARS-CoV-2 glycoprotein may dissolve HEPES 10 mM and 120 mM NaCl solution, and the RIX of PBSN can be experimentally measured.

When used, various quantities of HEPES solution were added to a 120 mM NaCl solution and a good linear relation between the RIX of PBSN and HEPES concentration ranging from 0 to 20 mM was found, and this relationship can be expressed as [44]

$$n=0.00004y+1.334$$
(4)

where n is the RIX index of PBSN. y is the solution of HEPES (mM). In this investigation, diverse concentrations of HEPES solution have been given in 10 mM, so the RIX of the sensing medium was assumed to be 1.3344 [45].

Mathematical Modeling

Along the z-axis, all layers are organized into a stack. The dielectric constant and thickness for each layer are εk and dk. The relationship between the tangential fields of the first boundary at z = 0 and those of the final boundary at z = zN-1 is given by the characteristic matrix (W) as

$$\left[\begin{array}{c}{F}_{1}\\ {G}_{1}\end{array}\right]=W\left[\begin{array}{c}{F}_{{\varvec{N}}-1}\\ {G}_{{\varvec{N}}-1}\end{array}\right]$$
(5)

where Fi and Gi are the electric and magnetic fields at the i boundary. The characteristic matrix (Wj) of the jth layer can be written as

$${W}_{j}=\left[\begin{array}{cc}\mathrm{cos}({\beta }_{j})& -\frac{i\; \mathrm{sin}({\beta }_{j})}{{Y}_{j}}\\ -i\,{Y}_{j}\mathrm{sin}({\beta }_{j})& \mathrm{cos}({\beta }_{j})\end{array}\right]$$
(6)

\({\beta }_{j}\) is the phase variation which can be written as

$${\beta }_{j}=\frac{2\pi }{\lambda }{d}_{j}{({\varepsilon }_{j}-{{{(n}_{1}\sin\theta}_{1})}^{2})}^{0.5}$$
(7)

where \({n}_{j}\) is the RIX, \({\varepsilon }_{j}\) is the dielectric permittivity, and dj is the thickness. \({\theta }_{1} \;and\; {n}_{1}\) are the prism angle of incidence and RIX.

For TM wave \({Y}_{j}={({\varepsilon }_{j}-{{{(n}_{1}\sin\theta }_{1})}^{2})}^{0.5}/{\varepsilon }_{j}\).

WT is the characteristic matrix of the device shown in Fig. 1 which can be expressed as

$${W}_{T}={W}_{Ag} {W}_{BF}{ W}_{G}=\left[\begin{array}{cc}{W}_{11}& {W}_{12}\\ {W}_{21}& {W}_{22}\end{array}\right]$$
(8)

where Wij are the elements of the matrix.

The reflection coefficient (r) for the TM wave

$$r=\frac{\left({W}_{11}+{W}_{12}{Y}_{N}\right){Y}_{1}-({W}_{21}+{W}_{22}{Y}_{N})}{\left({W}_{11}+{W}_{12}{Y}_{N}\right){Y}_{1}+({W}_{21}+{W}_{22}{Y}_{N})}$$
(9)

The reflection intensity can be written as

$$R={\left|r\right|}^{2}$$
(10)

SPRE sensor performance can be measured by means of sensitivity (S) and figure of merit (FoM). A modification of resonance angle, Δθres, takes place as a result of the change in the analyte RI, Δns. The sensitivity can be defined in terms of Δθres and Δns as

$$S= \frac{\mathrm{\Delta \theta res}}{\mathrm{\Delta ns}}$$
(11)

The FWHM of the dip is given by

$$FWHM= {\theta }_{2}-{\theta }_{1}$$
(12)

where θ1 and θ2 are the resonant angles at 50% intensity.

The detection accuracy (DA) of the SPRE sensor can be obtained by using

$$DA=\frac{1}{FWHM}$$
(13)

The FoM depends on both S and DA

$$FOM=S\times DA$$
(14)

Results and Discussion

Here, we present a Kretschmann-design SPRE biosensor, which includes a prism to give the SPNs the required momentum. BK7 prism is used in this work because of its low RI. Applying TM light (λ = 633 nm), SPN coupling is realized at the contact between the metal and dielectric. A reflectance change is observed due to the variation in the analyte RIX. The reflectance curve exhibits a sharp dip at resonance conditions. This indicates that maximum energy is transferred to the SPNs at resonance conditions. The refractive indices of the materials used are \({n}_{p}= 1.5151\), nAg = \(0.0562+4.2776 i\), nBF = 2.9680 and nG = 3 + 1.149106 \(i\). \({n}_{p}\), nAg, and nG are calculated using Eqs. (1), (2) and (3), respectively, at λ = 633 nm. The thicknesses of the layers are initially considered as dAg = 30 nm, dBF = 1 nm and dG = M \(\times\) 0.34 nm (M = 1). Five different concentrations of 0, 150, 275, 400, and 525 mM mixing in the buffer channel are considered. According to Eq. (4), the analyte RIX ranges from 1.334 to 1.355. Wolfram Mathematica 11.2 software has been used to conduct numerical simulations.

Figure 2 shows the reflectance profile for various device arrangements. The findings imply that the resonant angle shifts toward a larger value in response to the increase in the sensing medium RIX. The resonance curve for a traditional SPRE sensor using Ag-Graphene layers is shown in Figure 2a. The sensitivity Eq. (11) found for the SPRE sensor (without BiFeO3) is 114.29 deg/RIU. DA and FoM Eqs. (13) and (14) obtained from the resonance curve are 191.9 \(\times {10}^{-3}\) deg-1 and 21.93 RIU-1, respectively. After introducing a 1 nm BiFeO3 layer in the conventional SPRE sensor, the sensitivity becomes 120.95 deg/RIU as illustrated in Fig. 2b which means an improvement of 5.83%. In this case, DA and FoM achieved with the BiFeO3 layer are 181.4 \(\times {10}^{-3}\) deg-1 and 21.94 RIU-1, respectively. Thus, the device with the structure BK7 prism/Ag/ BiFeO3 /graphene/ sensing medium will be used in the next parts of the paper due to its better performance.

Fig. 2
figure 2

Reflectance profile of the proposed SPRE sensor. a without BiFeO3 and b with a BiFeO3 layer

The impacts of certain factors on the sensitivity of the SPRE sensor are examined. These factors include the number of graphene layers, BiFeO3 layer thickness, metal layer thickness, and the RIX of the sensing medium. As illustrated in Fig. 3a, various thicknesses of Ag (between 30 and 60 nm) are employed. The calculated sensitivities for silver layer thicknesses of 30, 40, 50, and 60 nm, respectively, are 120.95, 126.67, 131.4, and 134.29 deg/RIU. For Ag-layer thicknesses > 60 nm, no dip can be observed in the reflectance curve. Therefore, the optimum thickness of the Ag layer is 60 nm which will be considered in the next investigation. The BiFeO3 layer thickness is varied between 1 and 7 nm and the sensitivity is calculated for each thickness (Fig. 3b). The sensitivities are found as 134.29, 152.38, 189.52 and 282.86 deg/RIU for dBF = 1, 3, 5 and 7 nm. For dBF > 7 nm, the resonant dip disappears. Hence, the optimum BiFeO3 layer thickness is dBF = 7 nm. A graphene sheet has a thickness of 0.34 nm. The graphene layer is assumed to have a number of M sheets. The reflectance profile is investigated for different M and the sensitivity is calculated and found as 282.86, 293.81, 275.2 and 228.57 deg/RIU for M = 1, 2, 3 and 4, respectively as shown in Fig. 3c. Therefore, the optimum number of graphene sheets is M = 2. Figure 3d shows the sensitivity as a function of the chemical potential (μc) of graphene. The graphene permittivity (\({\varepsilon }_{G}\)) can be written as

$${\varepsilon }_{G}=1+\frac{i \;{\sigma }_{G}}{\omega {\varepsilon }_{0}{d}_{G}}$$
(13)

where \({\sigma }_{G}\) is the graphene surface conductivity which is given by

$${\sigma }_{G}=\frac{{e}^{2}}{\pi {h}^{2}}\frac{{k}_{B}T}{{\Gamma }_{G}-i\omega }[\frac{{\mu }_{C}}{{k}_{B}T}+2\mathrm{ln}({e}^{\frac{-{\mu }_{C}}{{k}_{B}T}}+1)]$$
(14)

where \({k}_{B}\), T, \({\Gamma }_{G}\) and \({\mu }_{C}\) are Boltzmann constant, temperature, graphene phenomenological scattering rate, and chemical potential. The graphene RIX is related to \({\varepsilon }_{G}\) through the relation \({n}_{G}=\sqrt{{\varepsilon }_{G}}\) . The chemical potential of graphene is varied and the sensitivity is calculated. Figure 3d shows the sensitivity of the proposed structure versus the chemical potential of graphene. The sensitivity decreases slightly as the chemical potential of graphene increases. The highest sensitivity of 293.81 is attained when \({\mu }_{C}\) = 0.3 eV.

Fig. 3
figure 3

Sensitivity optimization. a Sensitivity versus the thickness of silver at dBT = 1 nm and M = 1. b Sensitivity versus the thickness of BiFeO3 at dAg = 60 nm and M = 1. c Sensitivity versus the number of graphene sheets at dAg = 60 nm and dBF = 7 nm. d Sensitivity versus the chemical potential of graphene at dAg = 60 nm, dBF = 7 nm and M = 2

Figure 4a displays the resonant dips of the reflectance curve for various SARS-CoV-2 solution concentrations. The parameters are set as dAg = 60 nm, dBF = 7 nm, and dG = M×0.34 nm (M = 2) as optimum values. The resonant angle is observed at 78.34° when 0.0 mM is used as the sensing medium. It is noted that as SARS-CoV-2 concentration in the sensing medium solution is increased, the resonant dip moves to higher resonant angles. Resonant angles are found at 79.82º, 81.25º, 82.79º and 84.51º for concentrations of 150, 275, 400 and 525 mM, respectively. We have used the reflectance profile to calculate the sensitivity of the sensor to the sensing medium. The sensitivities are found as 246.67, 264.55, 278.12 and 293.81 deg/RIU for the samples having concentrations of 150, 275, 400 and 525 mM, respectively, as shown in Fig. 4b. SARS-CoV-2 concentrations, analyte RI, dip position, angular shift and sensitivity are presented in Table 1. The highest sensitivity of the biosensor is 293.81 deg/RIU for a concentration of 525 mM, which is an ultra-high value as compared to the recently published papers as shown in Table 2.

Fig. 4
figure 4

a Reflectance curve of the proposed SPRE sensors for different concentrations. b Sensitivity versus the analyte RIX for the proposed SPRE structure

Table 1 Variation of the sensitivity with SARS-CoV-2 solution concentration
Table 2 Comparing the current work with the most recently published SPRE biosensors

Finally, the fabrication techniques of the proposed sensor have to be illustrated. A thin layer of 60 nm of Ag metal has to be deposited on the BK7 prism. The most common deposition methods include ion plating evaporation, laser beam evaporation, vacuum thermal evaporation, and electron beam evaporation. A 7 nm layer of BiFeO3 has to be grown on the Ag film. The deposition of BiFeO3 thin film can be achieved using pulsed laser deposition. Two layers of graphene are then deposited on the top of the BiFeO3 thin film. Chemical vapor deposition can be used for graphene deposition.

Conclusion

In this work, SPRE-based biosensor has been theoretically designed and investigated for the detection of SARS-CoV-2 virus. The proposed structure has been explored by employing the conventional technique of TMM. The structure is examined without and with the BiFeO3 layer. It is found that the presence of a 1 nm BiFeO3 layer can improve the sensitivity by 5.83%. The sensitivity is then examined with the Ag and BiFeO3 layer thicknesses. It is found that the optimum values of Ag and BiFeO3 layers that correspond to the maximum sensitivity are 60 nm and 7 nm, respectively. The optimum number of graphene sheets is 2. The variation of the sensitivity is studied with the SARS-Cov-2 concentration. The highest sensitivity is found as 293 deg/RIU corresponding to a SARS-Cov-2 concentration of 525 mM (RIX of a sensing medium of 1.355). It is believed that the proposed device can open a new window for the efficient detection of SARS-CoV-2.