Abstract
Objective: The purpose of this study was to assess and compare the humidification, heating, and resistance properties of three commercially available heat-moisture exchangers (HMEs). To mimic clinical conditions, a previously validated, new, realistic experimental setup and measurement protocol was used.
Design: Prospective, comparative experimental study.
Setting: Surgical Intensive Care Unit, University Hospital of Rotterdam.
Materials: An experimental set-up consisting of a patient model, measurement systems, and ventilator and three different HME types.
Interventions: The air flow, pressure in the ventilation circuit, pressure difference over the HME, and partial water vapour pressure and temperature at each side of the HMEs were measured. Measurements were repeated every 30 min during the first 2 h and every hour up to 24 h for each HME at six different ventilator settings. The mean inspiratory and maximum expiratory resistance, flow-weighted mean absolute humidity and temperature outputs, and humidification and heating efficiencies of HMEs were calculated.
Measurements and results: The Dar Hygroster had the highest humidity output, temperature output, humidification efficiency, and heating efficiency values throughout the study (32.8 ± 21. mg/l, 32.2 ± 0.8 °C, 86.3 ± 2.3 %, and 0.9 ± 0.01 %, respectively) in comparison to the Humid-Vent Filter (25.3 ± 3.2 mg/l, 31.9 ± 0.8 °C, 72.2 ± 5.3%, 0.9 ± 0.02 %, respectively) and the Pall Ultipor BB100 breathing circuit filter (23.4 ± 3 mg/l, 28.3 ± 0.7 °C, 68.8 ± 5.9 %, 0.8 ± 0.02 %, respectively). The inspiratory and expiratory resistance of the HMEs remained below clinically acceptable maximum values (2.60 ± 0.04 and 2.45 ± 0.05 cmH2O/l per s, respectively). Conclusion: The Dar Hygroster filter was found to have the highest humidity and temperature output of all three HMEs, the Humid-Vent filter had a satisfactory humidity output only at low tidal volume flow rate and minute volume settings, whereas the Pall Ultipore BB100 never achieved a sufficient humidity and temperature output.
Article PDF
Similar content being viewed by others
References
Dery R (1973) The evolution of heat and moisture in the respiratory tract during anaesthesia with a non-rebreathing system. Can Anaesth Soc J 20: 296–309
Ingelstedt S (1956) Studies on the conditioning of air in the respiratory tract. Acta Otolaryngol Suppl 131:1–81
Burton JDK, Lond MB (1962) Effects of dry anaesthetic gases on the respiratory mucous membrane. Lancet III: 235–238
Tsuda T, Noguchi H, Takumi Y, Aochi O (1977) Optimum humidification of air administered to a tracheostomy in dogs. Br J Anaesth 49: 965–976
Rashad K, Wilson K, Hurt HH, Graff TD, Benson DW (1967) Effect of humidification of anesthetic gases on static compliance. Anesth Analg 46: 127–133
Noguchi H, Takumi Y, Aochi O (1973) A study of humidification in tracheotomized dogs. Br J Anaesth 45: 844–847
Chalon J, Patel C, Ali M, Ramanathan S, Capan L, Tang CK, Turndorf H (1979) Humidity and anesthetized patient. Anesthesiology 50:195–198
Marfatia S, Donahoe PK, Hendren WH (1975) Effect of dry and humidified gases on the respiratory epithelium in rabbits. J Pediatr Surg 10: 583–590
Fonkalsrud EW, Sanchez M, Higashijima I, Arima E (1975) A comparative study of the effects of dry vs. humidified ventilation on canine lungs. Surgery 78: 373–380
Hirsch JA, Tokayer JL, Robinson MJ, Sackner MA (1975) Effects of dry air and subsequent humidification on tracheal mucous velocity in dogs. J Appl Physiol 39: 242–246
Evaluation Report (1987) Heat and moisture exchangers. J Med Eng Technol 11: 117–127
Craven DE, Goularte TA, Make BJ (1984) Contaminated condensate in mechanical ventilator circuits. Am J Respir Crit Care Med 129: 625–628
Boys JE, Howells TH (1972) Humidification in anaesthesia. A review of the present situation. Br J Anaesth 44: 879–886
Im SWK, Fung JPH, So SY, Yu DYC (1982) Unusual dissemination of Pseudomonas by ventilators. Anaesthesia 37:1074–1077
Klein EF, Graves SA (1974) Hot pot tracheitis. Chest 65: 225–226
Ogino M, Kopotic R, Mannino FL (1985) Moisture-conserving efficiency of condenser humidifiers. Anaesthesia 40: 990–995
Martin C, Perrin G, Gevaudan MJ, Saux P, Gouin F (1990) Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit. Chest 97: 144–149
Ploysongsang Y, Branson R, Rashkin MC, Hurst JM (1988) Pressure flow characteristics of commonly used heat-moisture exchangers. Am J Resp Crit Care Med 138: 675–678
Hay R, Miller WC (1982) Efficacy of a new hygroscopic condenser humidifier. Crit Care Med 10: 49–51
Chalon J, Markham JP, Ali MM, Ramanathan S, Turndorf H (1984) The Pall Ultipor breathing circuit filter: an efficient heat and moisture exchanger. Anesth Analg 63: 566–567
Eckerbom B, Lindholm CE (1990) Laboratory evaluation of heat and moisture exchangers. Assessment of the draft international standard (ISO/DIS 9360) in practice. Acta Anaesthesiol Scand 34: 291–295
Eckerbom B, Lindholm CE (1990) Performance evaluation of six heat and moisture exchangers according to the draft international standard (ISO/DIN 9360). Acta Anaesthesiol Scand 34: 404–409
Heat and Moisture Exchangers (1983–1984) Evaluation. Health Devices Sourcebook 12:155–167
Vickers MD, Mecklenburgh JS, Hampson M (1992) Heat and Moisture exchangers (HME), DAR Hygrobac/Hygroboy. Evaluation 103: 1–12
Vickers MD, Mecklenburgh JS, Hampson M (1992) Heat and moisture exchangers (HME), Gibeck Humid-Vent filter. Evaluation 102:1–10
Vickers MD, Mecklenburgh JS, Hampson M (1992) Heat and moisture exchangers (HME), DAR Hygroster. Evaluation 142:1–16
Mebius C (1992) Heat and moisture exchangers with bacterial filters: a laboratory evaluation. Acta Anaesthesiol Scand 36:1–5
Kugimiya T, Phuc TG, Numata K (1989) Laboratory evaluation of heat and moisture exchangers. J Anesth 3: 80–85
Ünal N, Pompe JC, Holland WPJ, Gültuna I, Huygen PEM, Jabaaij K, Ince C, Saygìn B, Bruining HA (1995) An experimental set-up to test heat-moisture exchangers. Intensive Care Med 21:142–148
International Organization for Standardization (1988) Heat and moisture exchangers for use in humidifying respired gases in humans. Draft International Standard (ISO/DIS 9360, 1988) 121:1–13
International Organization for Standardization (1992) Anaesthetic and respiratory equipment — heat and moisture exchangers for use in humidifying respired gases in humans. International Standard (ISO 9360,1992)
McFadden ER, Pichurko BM, Bowman HF, Ingenito E, Burns S, Dowling N, Solway J (1985) Thermal mapping of the airways in humans. J Appl Physiol 58: 564–579
Ferrus L, Guenard H, Vardon G, Varene P (1980) Respiratory water loss. Respir Physiol 39: 367–381
Liese W, Warwick WJ, Cumming G (1974) Water vapour pressure in expired air. Respiration 31: 252–261
Ward JJ, Helmholz HF (1991) Applied humidity and aerosol therapy. In: Burton GG, Hodgkin JE, Ward JJ (eds) Respiratory care. A guide for clinical practice, 3th ed. Lippincott, Philadelphia, pp355–396
Cohen IL, Weinberg PF, Fein IA, Rowinski GS (1988) Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit. Crit Care Med 16: 277–279
Turtle MJ, Ilsley AH, Rutten AJ, Runciman WB (1987) An evaluation of six disposable heat and moisture exchangers. Anaesth Intensive Care 15: 317–322
Martin C, Papazian L, Perrin G, Bantz P, Gouin F (1992) Performance evaluation of 3 vaporizing humidifiers and 2 heat and moisture exchangers in patients with minute ventilation >101/ min. Chest 102:1347–1350
Miyao H, Hirokawa T, Miyasaka K, Kawazoe T (1992) Relative humidity, not absolute humidity, is of great importance when using a humidifier with a heating wire. Crit Care Med 20: 674–679
Sottiaux T, Mignolet GRT, Damas P, Lamy M (1993) Comparative evaluation of three heat and moisture exchangers during short-term postoperative mechanical ventilation. Chest 104: 220–224
Conti G, De Blasi RA, Rocco M, Pelaia P Antonelli M, Bufi M, Mattia C, Gasparetto A (1990) Effects of the heat-moisture exchangers on dynamic hyperinflation of mechanically ventilated COPD patients. Intensive Care Med 16: 441–443
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ünal, N., Kanhai, J.K.K., Buijk, S.L.C.E. et al. A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings. Intensive Care Med 24, 138–146 (1998). https://doi.org/10.1007/s001340050535
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s001340050535