A comparative test of differential scanning calorimeters

  • Instrumentation
  • Published:
Journal of thermal analysis Aims and scope Submit manuscript

Abstract

The Dutch Society for Thermal Analysis has developed tests to measure the resolution and the sensitivity of Differential Scanning Calorimeters. For this test the substance 4,4′-azoxyanisole is used. This substance shows two transitions: a solid to liquid crystal transition at about 117°C (ΔH≈120 J g−1) and a liquid crystal to isotropic liquid transition at about 134°C (ΔH≈2 J g−1). The resolution test is performed using an amount of 5 mg substance and a high heating rate of 20°C min−1. The resolution is evaluated by measuring how well the two peaks are separated. An amount of 0.25 mg substance and a low heating rate of 0.1 °C min−1 is used for the sensitivity test. The sensitivity is evaluated as the ratio of the peak height of the LC-transition and the top-top noise level.

Members of the TAWN were asked to participate in the test. Each participant was provided with an amount of sample and a test procedure. 47 Contributions were received and these results are presented.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

All members of the TAWN who participated in the tests are thanked for their contributions.

Rights and permissions

Reprints and permissions

About this article

Cite this article

van Ekeren, P.J., Holl, C.M. & Witteveen, A.J. A comparative test of differential scanning calorimeters. Journal of Thermal Analysis 49, 1105–1114 (1997). https://doi.org/10.1007/BF01996799

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01996799

Keywords

Navigation