Abstract
Transpulmonary pressure (PL) calculation requires esophageal pressure (PES) as a surrogate of pleural pressure (Ppl), but its calibration is a cumbersome technique. Central venous pressure (CVP) swings may reflect tidal variations in Ppl and could be used instead of PES, but the interpretation of CVP waveforms could be difficult due to superposition of heartbeat-induced pressure changes. Thus, we developed a digital filter able to remove the cardiac noise to obtain a filtered CVP (f-CVP). The aim of the study was to evaluate the accuracy of CVP and filtered CVP swings (ΔCVP and Δf-CVP, respectively) in estimating esophageal respiratory swings (ΔPES) and compare PL calculated with CVP, f-CVP and PES; then we tested the diagnostic accuracy of the f-CVP method to identify unsafe high PL levels, defined as PL>10 cmH2O. Twenty patients with acute respiratory failure (defined as PaO2/FiO2 ratio below 200 mmHg) treated with invasive mechanical ventilation and monitored with an esophageal balloon and central venous catheter were enrolled prospectively. For each patient a recording session at baseline was performed, repeated if a modification in ventilatory settings occurred. PES, CVP and airway pressure during an end-inspiratory and -expiratory pause were simultaneously recorded; CVP, f-CVP and PES waveforms were analyzed off-line and used to calculate transpulmonary pressure (PLCVP, PLf-CVP, PLPES, respectively). Δf-CVP correlated better than ΔCVP with ΔPES (r = 0.8, p = 0.001 vs. r = 0.08, p = 0.73), with a lower bias in Bland Altman analysis in favor of PLf-CVP (mean bias − 0.16, Limits of Agreement (LoA) -1.31, 0.98 cmH2O vs. mean bias − 0.79, LoA − 3.14, 1.55 cmH2O). Both PLf-CVP and PLCVP correlated well with PLPES (r = 0.98, p < 0.001 vs. r = 0.94, p < 0.001), again with a lower bias in Bland Altman analysis in favor of PLf-CVP (0.15, LoA − 0.95, 1.26 cmH2O vs. 0.80, LoA − 1.51, 3.12, cmH2O). PLf-CVP discriminated high PL value with an area under the receiver operating characteristic curve 0.99 (standard deviation, SD, 0.02) (AUC difference = 0.01 [-0.024; 0.05], p = 0.48). In mechanically ventilated patients with acute respiratory failure, the digital filtered CVP estimated ΔPES and PL obtained from digital filtered CVP represented a reliable value of standard PL measured with the esophageal method and could identify patients with non-protective ventilation settings.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Background
Mechanical ventilation (MV) is a life-saving treatment in patients with acute respiratory failure; however, MV can also contribute to additional lung damage, which has been named “ventilator-induced lung injury” (VILI) [1]. To minimize VILI, current guidelines recommend a “protective” ventilation strategy, using low tidal volume and a plateau pressure < 30 cmH2O, especially in patients suffering from acute respiratory distress syndrome (ARDS) [2]. Nevertheless, these ventilatory settings might not be appropriate in all patients on MV [3, 4]; indeed, setting of MV parameters cannot rely solely on airway pressures evaluation, because this is an imprecise index of the stress applied to the lung, which is best estimated by the transpulmonary pressure (PL) [5]. The correct assessment of PL would request the measurement of pleural pressure (Ppl), which is not easy to obtain in the clinical setting [6]. Thus, esophageal pressure (PES), measured through an esophageal balloon, has been used as a surrogate of Ppl [7] and PL calculated as the difference between airway pressure (Paw) and PES [6].
Despite the solid physiological background of PL, a large observational study showed that PES is rarely monitored in clinical practice [1]; this is due to difficulties in placing the esophageal balloon, validating the signal and interpreting the data [8]. For such reasons, an estimation of PL available without the need of placing an esophageal balloon would be attractive for clinicians. Due to its low elastance and close anatomical location, superior vena cava could be an alternative site of measure of pleural pressure: previous studies have reported that tidal excursions in central venous pressure (CVP) may reflect pleural swings (ΔPpl) [9], although the comparison between ΔCVP and the reference method, i.e. ΔPES showed a poor agreement [10, 11]. Due to this uncertainty, the idea of using respiratory swings in CVP as a surrogate of ΔPpl has not been further implemented, and, to the best of our knowledge, only one recent study has attempted to estimate transpulmonary pressure from the respiratory fluctuations of CVP [12].
Some of the pitfalls in the CVP wave analysis are related to the assessment of CVP tracings. Indeed, CVP signal, as well as PES, is influenced by the changes in intrathoracic pressure [9] and shows ventilation-dependent oscillations, mainly in the low-frequency domain. Although, this waveform also presents distortions related to the high-frequency cardiac oscillations and altered shape from peaks and troughs. As such, we developed a digital signal processing technique able to remove from the CVP signal the cardiac components and to extract therefore the respiratory ones; this method could be a simple and effective alternative to estimate PL from the CVP analysis (f-CVP).
The aim of the present study was to assess the validity of CVP and filtered CVP respiratory swings (ΔCVP and Δf-CVP) as an estimate of esophageal respiratory swings (ΔPES) in a cohort of mechanically ventilated patients with acute respiratory failure. Then, we obtained PL using ΔCVP and Δf-CVP and compared them to PL calculated with the reference PES method. For this purpose, we used the transpulmonary pressure formula that represents the tidal lung stress, the effective distending pressure of the lungs when they are inflated with tidal volume [13, 14]. Secondary outcome was to test the diagnostic accuracy of the f-CVP method to identify patients at risk of VILI with high PL levels.
2 Materials and methods
2.1 Study population
This was a retrospective analysis of prospectively collected data from patients admitted to a 7-bed Intensive Care Unit (ICU) at the University Hospital of Siena (Italy). The study was approved by the local Ethics Committee (Comitato Etico Regionale per la Sperimentazione Clinica della Regione Toscana; PPL2-14925) and registered on Clinicaltrials.gov (NCT04443491). Written informed consent was obtained from the patient’s next of kin before starting the study protocol.
Between May 2016 and March 2019, all patients with a diagnosis of acute hypoxemic respiratory failure (defined as PaO2/FiO2 ratio below 200) requiring invasive mechanical ventilation and equipped with both a central venous catheter (CVC) and an esophageal balloon were enrolled within 48 h from the starting of MV. The decision to insert the esophageal balloon was made by the attending physician and PES was used to set MV, by reducing tidal volume (VT) (if PL was equal or above 10 cmH2O) [15], by increasing tidal volume in case of respiratory acidosis (if PL was within safe values), by adjusting PEEP level during a decremental PEEP trial test or to maintain an end-expiratory PL equal or above 0 cmH2O or a combination of the above.
The exclusion criteria were: patients < 18 years; arrhythmias; presence of pulmonary air leakage (i.e. chest trauma and/or chest tube with active aspiration); pregnancy. Patients with arrhythmias were excluded due to difficult readings of the CVP trace. Chest tube with negative aspiration could impair homogeneous transmission of pleural pressure across the intrathoracic space, so those cases were excluded. Pregnant patients were excluded in according to the local Ethics Committee policy.
2.2 Setting
During the measurements, patients were in the semi recumbent position. PES was measured using a double-balloon, graduated feeding catheter (NutriVent®, Mirandola, Modena, Italy) [16] connected to a pressure transducer filled with air; the esophageal balloon was initially positioned in the stomach to check for the presence of positive deflection; then, it was withdrawn until it reached the lower third of the esophagus when cardiac artifacts appeared on the esophageal pressure trace. To check the correct position of the esophageal balloon, external manual compressions on the rib cage were applied during an expiratory pause and simultaneous positive deflections of airway and esophageal pressures compared (i.e. “positive pressure occlusion test”) [6]. Briefly, an occlusion maneuver was performed at end-expiration; changes in PES (ΔPES) and Paw (ΔPaw) during gentle chest compressions were simultaneously recorded, and the test suggested correct positioning if ΔPES/ΔPaw ratio ranged from 0.9 to 1.1. Also, correct positioning was confirmed on chest X-ray, with the direct visualization of two radiopaque markers, one above and one below the diaphragm. The balloon was inflated at a volume of 4 ml according to the recommendations of the manufacturer [16] and the balloon inflation was checked before every measurement to ensure it contained the adequate amount of air.
A triple-lumen central venous catheter (CVC; Certofix Trio V720 Braun, Melsungen Germany) was placed, as routinely protocol in our department, using ultrasound and electrocardiogram (ECG)-guided technique, through the right jugular vein with the tip located in the superior vena cava upstream to the right atrium [17]. Correct positioning of the CVC was also confirmed on chest X-ray. Measure of CVP was obtained from the distal port of a fluid-filled CVC connected to an electronic pressure transducer connected to the bedside multiparametric monitor. The transducer was zeroed at the phlebostatic point at 5 cm below the sternal angle (thus at this level patients do not have to be supine for the measurements) [18]. CVP values were taken at the base of the “c” wave at the end-inspiratory and end-expiratory pauses [19].
Continuous ECG tracing, PES and CVP curves were displayed on the bedside multi-parametric monitor (Intellivue MP60, Philips), while simultaneously PES and CVP were acquired as “rough” signals by a dedicated software developed by the authors (LabVIEW, National Instruments Corporation, Assago-Milano, Italy) before entering in the ICU monitor [20]. All traces were acquired according to the default bedside monitor sampling frequencies (500 Hz for the ECG traces, 125 Hz for the others).
2.3 Study protocol
All patients were deeply sedated (i.e. Richmond Agitation and Sedation Scale, RASS of -4 or -5 [21]), paralyzed and ventilated with a square flow waveform with tidal volume, respiratory rate and PEEP set by the attending physician, while the oxygen fraction was selected to keep a SpO2 between 94 and 96%. At baseline (T1), PES and CVP curves were simultaneously recorded for 5 min; if ventilator settings were changed, all measurements were repeated within one hour after such modification (T2). Each measure was taken after reaching an equilibrium (i.e. at least 15 min after the last change in ventilator setting) [22]. In all patients, an expiratory and inspiratory pause of 3–5 s were performed to achieve static respiratory mechanics. We recorded plateau pressure (Pplat), total PEEP (PEEPtot) and intrinsic PEEP values. The driving pressure was calculated as Pplat minus PEEPtot [23]. All these data were recorded as routine practice and therefore analyzed off-line.
2.4 Off-line analysis
From a spectral analysis perspective, CVP signal is characterized by low frequency (respiratory-dependent) as well as high frequency (cardiac-dependent) components [24]. Therefore, we used a dedicated low pass digital filter to remove the high frequency components to obtain a f-CVP. Technical details regarding the low-pass filter design have been described in recent works [19]. Using a linear filtering approach, we assumed that CVP is the sum of three pressure components related to the heart activity (PH), the breath activity (PB) and other noisy components (N):
where P0 is a biasing pressure offset (the CVP mean value), whereas the remaining signals have zero-mean.
In the linear filtering approach, it is hypothesized that the pressures PH and PB are signals with separable high-frequency and low-frequency spectral components, i.e., it was assumed that for a certain frequency f0.
where PB and PH are the Fourier transforms of PB and PH, respectively. Following the above assumption, a theoretical linear low-pass filter with unity gain HLP exists such that, neglecting the noisy term N, a reliable estimation P´B of the spectrum PB can be obtained as.
The values of f-CVP obtained during end-inspiratory and end-expiratory pauses were reported in the database and used to calculate transpulmonary pressure. Figure 1A and B shows examples of curves analysis with the proposed method: it could be noted that a temporal delay exists in the filtered curve due to the processing of the signal, however, for the purpose of the calculation of PL, this delay is negligible.
During the same respiratory pauses, we recorded end-inspiratory and end-expiratory values of PES. These waveforms have cardiogenic oscillations: we measured the values of pressure at the peak of these oscillations for calculation. Filtered CVP, instead, appeared as a “clean” signal, without the typical waves, so we just selected the numeric values displayed during the respiratory pauses. Both PES and f-CVP waveforms were analyzed offline with LabVIEW software, as shown in Fig. 1.
The transpulmonary driving pressure was calculated as [13, 14]:
The same formula was applied for CVP and filtered CVP obtaining PLCVP and PLf-CVP, respectively.
2.5 Statistical analysis
Normal distribution of data was tested by the Shapiro–Wilk Normality Test. Data are reported as mean ± standard deviation or median [interquartile range] as appropriate. The paired t-test was used to compare the variables.
The association between ΔCVP and Δf-CVP with ΔPES and then between transpulmonary pressure values (calculated with PES, CVP and f-CVP) were expressed by correlation coefficient (r). The agreement between ΔCVP and Δf-CVP with ΔPES and between PL obtained with the reference PES method and with CVP and f-CVP was assessed using the Bland-Altman analysis corrected for repeated measures when appropriate. Limits of agreement (LoA) (as 2 times standard deviation, SD, of the bias) were computed.
The ability of f-CVP derived PL to follow variations or trends in PES derived PL after an adjustment in ventilator setting was assessed analyzing the correlation between the changes (∆) in PL calculated by subtracting the first baseline from the second measurement (T2–T1), obtaining ∆PLf-CVP and ∆PLPES, respectively [25, 26]. We analyzed the direction of change between the pre-post differences to assess the percentage of concordance between the two methods. Two-tailed statistical hypothesis testing was performed with p-values ≤ 0.05 considered statistically significant.
The diagnostic accuracy of the f-CVP method to identify patients at risk of VILI was displayed as the area under the Receiver Operating Characteristic (ROC) curve (AUC). For this purpose, a threshold of 10 cmH2O of transpulmonary driving pressure was considered [15]. DeLong test was used to compare the AUCs of ROC [27].
As previously reported [11], compliance of the superior vena cava could influence the transmission of pleural pressure swings to the point that the ΔCVP relationship to ΔPpl decreased at higher values of CVP [10] and ΔCVP may underestimate ΔPES. Thus, we tested the effect of CVP value on the correlation and agreement between the respiratory swings and the transpulmonary pressures by dichotomizing patients into two groups based on the median end-expiratory CVP value (low vs. high CVP level).
Statistical analysis was performed using MedCalc software (MedCalc Software bv, Ostend, Belgium; https://www.medcalc.org; 2016) and GraphPad PRISM version 6.0 (San Diego, CA, USA).
3 Results
3.1 Study population
During the study period, 163 patients had a diagnosis of acute hypoxemic respiratory failure requiring invasive mechanical ventilation and an esophageal balloon was placed in 46 (28%) of them. Due to technical issues on the data acquisition of the system software, the recordings of 22 patients, performed between October 2016 to December 2017, were lost. Thus, 24 patients were eligible for this study, but four were eventually excluded because of cardiac arrhythmia during the recording (n = 1) or poor quality of the waveform signals (n = 3). The remaining 20 patients represented the study cohort; demographic and clinical characteristics of these are shown in Tables 1 and 2; ARDS was diagnosed in seven patients, two with moderate and five with severe forms [28]. Rescue therapies were necessary in two patients (one required prone position and inhaled nitric oxide and another needed venous-venous extracorporeal membrane oxygenation); ICU mortality was 25% (n = 5). Ventilatory setting, as well as lung mechanics, determined before and after variation of the ventilator strategy, are shown in Table 3.
3.2 Baseline measurements
Median values of PES, CVP, f-CVP and relative respiratory tidal swings and transpulmonary driving pressures obtained are reported in Table 3. Δf-CVP (Fig. 2) correlated better than ΔCVP with ΔPES (r = 0.75, p = 0.001 vs. r = 0.08, p = 0.73 respectively). Also, the agreement with ΔPES was better for Δf-CVP than ΔCVP as revealed by the Bland-Altman analysis (mean bias − 0.16, LoA − 1.31, 0.98 cmH2O vs. mean bias − 0.79, LoA − 3.14, 1.55 cmH2O), as shown in Fig. 3A and B.
Correlations between PLCVP and PLf-CVP with PES were r = 0.94, p < 0.001 and r = 0.98, p < 0.001 (Fig. 4), respectively. Bland-Altman analysis again revealed a lower bias for PLf-CVP than PLCVP in comparison to PLPES (0.15, LoA − 0.95, 1.26 cmH2O vs. 0.80, LoA − 1.51, 3.12, cmH2O respectively).
3.3 Measurements after change in ventilator settings
After PES evaluation, a variation in the ventilator settings were performed in 16 patients: PEEP was adjusted in four patients, VT in ten and two patients experienced changes in both. In five cases (31%), values of PL\(\ge\)10 cmH2O were reduced either by lowering tidal volume < 6 mL/kg predicted body weight (n = 4) or by reducing PEEP (n = 1). In seven cases (44%), VT was increased to correct respiratory acidosis, keeping PL below 10 cmH2O (n = 5) or unchanged (n = 2). Correlation coefficient between ΔPES, ΔCVP and Δf-CVP were r = 0.49 (p = 0.049) and r = 0.69 (p = 0.003), respectively, and for Bland-Altman analysis mean biases were − 0.67 cmH2O (LoA − 2.61, 1.27 cmH2O) and − 0.16 cmH2O (LoA − 1.46, 1.15 cmH2O), respectively. PLPES correlated well with both PLCVP and PLf-CVP (r = 0.93, p < 0.001 and r = 0.97, p < 0.001). Mean bias between PLPES and PLCVP was 0.67 cmH2O (LoA − 1.27, 2.61 cmH2O) and with PLf-CVP was 0.16 cmH2O (LoA − 1.15, 1.46 cmH2O).
The change in PL after modification of the ventilator settings (T2-T1) was determined separately for both PES and f-CVP. As shown in Fig. 5, comparison of the changes measured by the two methods demonstrated a good correlation (r = 0.97, p < 0.0001). Four PL pairs were excluded from the analysis as at least one PL value was zero, i.e., PL remained unchanged after the adjustment of MV. The concordance of PL was 100% (12 of 12 pairs of PL agreed).
3.4 Diagnostic performance
During baseline measurements, in five patients PLPES was ≥ 10 cmH2O; the same patients were identified using the CVP and f-CVP formulas. PLf-CVP was ≥ 10 cmH2O in other three patients, while PLCVP in the other five, instead in the same cases PLPES revealed lower values within normal range. After the change in ventilatory settings that were applied in four patients, in three cases PLPES was reduced below 10 cmH2O and in one patient, although lower, remained above this cut-off. In the same patients PLf-CVP acted identically (after the changes in ventilatory settings PLf-CVP had dropped in all four cases, in three of them below the established cut-off), while PLCVP resulted below 10 cmH2O in only two of the four cases.
For the baseline measurements, PLfCVP discriminated high PL value with an area under the ROC curve of 0.97 (SD, 0.06) (AUC difference = 0.03 [0.03; 0.10], p = 0.32); sensitivity was 100%, specificity 80% with positive predictive value of 62.5% and negative predictive value of 100%.
3.5 Low vs. high CVP
Considering all the 36 measurements together, the median value of end-expiratory CVP was 20.5 cmH2O. In the low CVP (CVP lower than 20.5 cmH2O) group (18 measurements), Δf-CVP maintained a good correlation with ΔPES (r = 0.81, p < 0.01), mean bias was 0.16 (LoA − 0.98, 1.29) cmH2O, while ΔCVP showed a poor correlation (r = 0.30, p = 0.27) and a worse accuracy (mean bias 0.8, LoA − 1.33, 2.89 cmH2O). In the high (CVP equal or higher than 20.5 cmH2O) group (18 measurements) correlation between Δf-CVP and ΔPES was moderate, although significant (r = 0.55, p = 0.02), for ΔCVP, instead, became weak (r = 0.23, p = 0.36). Δf-CVP, again, performed well as accuracy (mean bias 0.16, LoA − 1.10, 1.41 cmH2O) compared to ΔCVP (mean bias 0.72, LoA − 1.49, 2.93 cmH2O).
4 Discussion
In this study, a new approach to estimate respiratory variations of PES was proposed, using the digital analysis of CVP curves in patients with acute respiratory failure under controlled invasive mechanical ventilation. Δf-CVP showed a good and significant correlation with ΔPES and could estimate it with accuracy and precision, while “rough”, non-filtered, CVP values exhibited worse results.
When compared with the reference PL calculated from PES, PL derived either by f-CVP and CVP methods showed a high correlation. Then, we demonstrated the capability of the f-CVP method to track variations in PL when ventilator settings were modified. Finally, we were able to detect the same cases of unsafe high PL level as identified with the reference PES method.
Assessment of PL allows to discriminate lung from chest wall mechanics and consequently titrate mechanical ventilation, although often neglected in clinical practice due to technical and methodological difficulties in using esophageal balloon [1]. Almost all mechanically ventilated patients have a central venous access [29] and we showed that the presence of a correctly placed CVC allowed, together with the presented algorithm of processed CVP, to obtain an estimation of respiratory swings of Ppl and PL without further devices. The idea of using the superior vena cava as an alternative site to the esophagus to record pleural pressure goes back to the 70’s: conflicting results emerged about the correlation between respiratory changes in CVP and PES, and CVP was found to perform better [9], worse [10, 30, 31] or as good as PES [32] in providing reliable measure of Ppl. Since different catheter-manometer systems were employed and CVP signal was not always processed, we cannot ultimately speculate about these past results. However, some differences between methods employed in the negative studies need to be cited, as a possible explanation of the different results obtained. Ostrander et al. [10] reported that in ten spontaneously breathing dogs ∆Ppl was transmitted to the vena cava not only with attenuation, but also with a significant temporal delay that could alter its accuracy. This phase lag was also seen in thirteen healthy spontaneously breathing subjects [31] and was interpreted as a consequence of the pressure-raising effects of increased inspiratory venous return. This delay was abolished when external airway resistances were added, simulating a respiratory distress, where instead prevailed a fall in intrathoracic pressure during inspiration and also the increased PES amplitudes were paralleled by similar increases in CVP amplitudes. In our investigation we didn’t observe such a significant delay, probably because we studied paralyzed patients under controlled mechanical ventilation during respiratory pauses, with no airway flow, in which the effect of an eventual phase lag would be not significant. Although physiologically sound and clinically appealing, the idea of using the CVP swings (ΔCVP) as a surrogate of Ppl fluctuations still seems experimental. Recently, this approach has been reevaluated [33]: various authors have reported good and significant correlation between ΔCVP and ΔPES, although a poor agreement, and, in patients breathing spontaneously, ΔCVP was related to the level of inspiratory effort [34, 35]. Very recent experts statement suggested that swings in CVP could be monitored as surrogate of work of breathing to detect strenuous inspiratory efforts and prevent patient-self-induced lung injury (P-SILI) [36]. In our study, thanks to the application of a filtered CVP waveform, the estimation of ΔPES by ΔCVP was improved in comparison to non-filtered CVP values. A possible explanation for the better results in terms of correlation and accuracy with f-CVP may be related to superimposed cardiac oscillation and to inferences from peaks and troughs upon the CVP waveform during tidal swings that could make accurate reading difficult: these artifacts were cleared with the low-pass filter approach. Instead, transpulmonary driving pressure values were reliable when using both CVP methods (although correlation and accuracy were slightly better with f-CVP). This finding could be easily explained and depend on the calculation of the transpulmonary formula: in our study respiratory swings in Ppl (as measured by PES, CVP and f-CVP) were relatively small and subtracted from a larger common variation in airway pressure.
Another group [12] have recently focused on the same objective proposing an alternative method by applying a corrective factor derived from an occlusion test. Their results were consistent with those of the present study: swings in CVP performed better as a surrogate of ΔPES when CVP values were “cleared” and the corrected-CVP-derived PL correlated well with PES-derived PL. However, the study population significantly differs from ours, as post-cardiac surgery pediatric patients were selected and cases of severe ARDS were not included. Moreover, the calculation of transpulmonary pressure assumed the equivalence between the ratio of ΔPaw to ΔCVP and the ratio of ΔPpl to ΔCVP during an occlusion test, far from being proved in different clinical settings and possible influenced by CVP level at high values. Indeed, due to a blunt transmission of change of Ppl to cardiac cavities and superior vena cava in case of non-compliant heart [37], it has been stated that ΔCVP should not be expected to reflect ΔPES at higher values of CVP [11]. Thus, we tested our method in cases of high end-expiratory CVP level, based on a threshold derived from the median value found in our study: in cases of CVP above 20.5 cmH2O (half of the cases included in the study) the correlation between Δf-CVP and ΔPES worsened, but remained moderate and significative, and Bland Altman analysis showed a reliable accuracy. This result becomes even more important when considering that a significant group of patients included in the analysis had a diagnosis of severe ARDS (25%), a condition frequently burdened by right heart failure [38], even if in our study we didn’t report its incidence. In contrast to the cited report of Bellamare [11], our findings were consistent with the results of Walling in which ∆CVP and ∆PES were comparable in cases of higher mean CVP (defined as > 10 cmH2O) [9]. Also, ∆CVP remained a reliable index in the identification of high inspiratory efforts without being affected by high mean CVP values [35]. Again, for Lassola et al. the association between ∆CVP and ∆PES was independent from the value of CVP (with a cut-off for high CVP of 14 cmH2O) [34]. One could argue about the different values employed to define a high CVP, but, as reported in a recent expert review, no clear cut-off can be identified [39].
Targeting protective mechanical ventilation also implies the need to stay within a safety PL limit. Transpulmonary driving pressure should be kept below 10–12 cmH2O, especially in ARDS patients to minimize the stress applied to an inhomogeneous lung parenchyma [13]. We tested the diagnostic accuracy of PLf-CVP in identifying cases of high and potentially unsafe level of PL, as defined by a threshold value of 10 cmH2O of PLPES: the proposed method worked as a sensitive and specific test to detect patients at risk of VILI with the capacity of rule out such cases.
Superposition of heartbeat-induced pressure changes is commonly seen in the CVP tracing. Hence, we developed a digital filtering able to remove the noise added by the cardiac component on the respiratory waveforms. Similar attempts were made even for PES, because also this waveform is influenced by pressure changes within the pericardium transmitted to the esophageal balloon, hampering the accuracy of the measurements [40,41,42,43]. A filtering approach based on fixed cut-off frequencies may suffer from a systematic over-attenuation in amplitude signal and unavoidable information loss because of potential overlap between bandwidths of cardiogenic oscillations and respiratory components. For such reason some authors proposed adaptive filtering techniques [42, 43]. In our study, the drawback related to such bandpass signal did not occur because measurements of CVP were collected during respiratory pauses on MV, in which the bandwidths of heart rate and respiratory component were very distant to each other. Indeed, as one could see in Fig. 1B, the values f-CVP during the respiratory hold maneuvers were slightly larger than the respective during tidal breaths (i.e. the inspiratory pause values of f-CVP were higher than the peak during inspiration and the expiratory pause values of f-CVP were lower than the peak during expiration) in a variable magnitude, but still below 0.5 cmH2O in our report. This sort of attenuation depends on the spectral signal overlaps between the respiratory and cardiac component. Such an error could not be neglected if one considers the low-pass filter CVP swings during tidal breaths, because it would imply a significant under-estimation of respiratory efforts in spontaneous breathing patients. So, the feasibility of the proposed technique is restricted to cases where a reliable respiratory hold measure could be obtained. We are aware that during spontaneous breathing this low-pass filter would perform weakly, and a novel adaptive time-variant filtering technique of CVP is being developed [20]. In our study similar digital processing was not attempted for PES curves because our aim was to propose a method of PL calculation that could do without an esophageal balloon, not to ameliorate the reference one. We are aware that f-CVP estimation of PL was obtained with an off-line analysis which could limit its clinical bedside application. Nonetheless, it is far less time consuming than the reference method, without any need for catheter calibration and could be easily implemented in a real-time monitoring.
Our study has some limitations. First, it is single-center and was conducted in a small population, although it is consistent with a preliminary clinical report of an experimental technique. Second, our method was applied in patients on controlled invasive MV and not tested in patients in spontaneous breathing: future research may involve such category of patients. Third, we measured CVP and PES in a semi-recumbent position, while in most studies they were measured in supine position [9, 12, 16]. Nevertheless, our analysis was non-interventional and performed in a population of intubated patients with severe respiratory disease, who, as routine standard of care, were maintained in a semi-recumbent position. Other recent studies found a positive correlation in tidal swings of CVP and PES in semi-recumbent position in a real-life scenario [34, 35]. This approach is supported in literature: Flemale demonstrated that both CVP and PES swings provide accurate measurements of the change in pleural pressure in different body positions, including supine and seated, if these methods were validated by an occlusion test [32]. In addition, recently, Repessé suggested the semirecumbent over the supine position in the evaluation of esophageal pressure because in the latter PES is more altered by cardiac artifacts and also tends to overestimate the value of PES [44].
One could wonder about the feasibility of the proposed technique: indeed, four patients out of twenty-four were excluded. Actually, one patient was erroneously enrolled because of his cardiac arrhythmia, a condition that might interfere with the digital filtering analysis resulting in a disturbed waveform. The other three patients were excluded due to esophageal pressure waveforms impossible to interpret because of superimposed artifacts, while in the same patients the f-CVP signals were readable. This proved again the difficulties of having a readable PES curve and underlined the need of an alternative method of transpulmonary pressure estimate.
5 Conclusions
Respiratory swings in CVP values obtained with a dedicated digital filter could represent a surrogate of ΔPES and PL calculated with f-CVP was strongly comparable with the reference PL−derived esophageal balloon technique. Moreover, the f-CVP method identified patients with high PL levels potentially at risk of VILI. Currently, the proposed technique is an off-line analysis, but could be easily integrated in a bedside monitor. Finally, the reliability of the f-CVP method must be assessed in spontaneous breathing patients.
Data availability
The datasets analyzed during the current study are available from the corresponding author on reasonable request.
References
Bellani G, Laffey JG, Pham T, et al. Epidemiology, patterns of care, and mortality for patients with acute respiratory distress syndrome in intensive care units in 50 countries. JAMA. 2016;315:788–800.
Fan E, Del Sorbo L, Goligher EC, Hodgson CL, Munshi L, Walkey AJ, Adhikari NKJ, Amato MBP, Branson R, Brower RG, Ferguson ND, Gajic O, Gattinoni L, Hess D, Mancebo J, Meade MO, McAuley DF, Pesenti A, Ranieri VM, Rubenfeld GD, Rubin E, Seckel M, Slutsky AS, Talmor D, Thompson BT, Wunsch H, Uleryk E, Brozek J, Brochard LJ. American Thoracic Society, European Society of Intensive Care Medicine, and Society of Critical Care Medicine. An Official American Thoracic Society/European Society of Intensive Care Medicine/Society of Critical Care Medicine Clinical Practice Guideline (2017) Mechanical Ventilation in Adult Patients with Acute Respiratory Distress Syndrome. Am J Respir Crit Care Med 195(9):1253–1263.
Terragni PP, Rosboch G, Tealdi A, Corno E, Menaldo E, Davini O, Gandini G, Herrmann P, Mascia L, Quintel M, et al. Tidal hyperinflation during low tidal volume ventilation in acute respiratory distress syndrome. Am J Respir Crit Care Med. 2007;175(2):160–6.
Bellani G, Guerra L, Musch G, Zanella A, Patroniti N, Mauri T, Messa C, Pesenti A. Lung regional metabolic activity and gas volume changes induced by tidal ventilation in patients with acute lung injury. Am J Respir Crit Care Med. 2011;183(9):1193–9.
Talmor D, Sarge T, Malhotra A, O’Donnell CR, Ritz R, Lisbon A, Novack V, Loring SH. Mechanical ventilation guided by esophageal pressure in acute lung injury. N Engl J Med. 2008;359(20):2095–104.
Akoumianaki E, Maggiore SM, Valenza F, Bellani G, Jubran A, Loring SH, Pelosi P, Talmor D, Grasso S, Chiumello D, Guérin C, Patroniti N, Ranieri VM, Gattinoni L, Nava S, Terragni PP, Pesenti A, Tobin M, Mancebo J, Brochard L, PLUG Working Group, Intensive Care Medicine). The application of esophageal pressure measurement in patients with respiratory failure. Am J Respir Crit Care Med. 2014;189(5):520–31.
Cherniack RM, Farhi LE, Armstrong BW, et al. A comparison of esophageal and intrapleural pressure in man. J Appl Physiol. 1955;8:203–11.
Brochard L. Measurement of esophageal pressure at bedside: pros and cons. Curr Opin Crit Care. 2014;20(1):39–46.
Walling PT, Savege TM. A comparison of oesophageal and central venous pressures in the measurement of transpulmonary pressure change. Br J Anaesth. 1976;48(5):475–9.
Ostrander L, Schneider A, Burma G, Kreul J. A comparison of central venous pressure and pleural pressure in supine dogs. Respir Physiol. 1977;31:63–70.
Bellemare P, Goldberg P, Magder SA. Variations in pulmonary artery occlusion pressure to estimate changes in pleural pressure. Intensive Care Med. 2007;33(11):2004–8.
Kyogoku M, Takeuchi M, Inata Y, Okuda N, Shimizu Y, Hatachi T, Moon K, Tachibana K. A novel method for transpulmonary pressure estimation using fluctuation of central venous pressure. J Clin Monit Comput. 2020;34(4):725–31.
Baedorf Kassis E, Loring SH, Talmor D. Mortality and pulmonary mechanics in relation to respiratory system and transpulmonary driving pressures in ARDS. Intensive Care Med. 2016;42:1206–13.
Chen L, et al. Partition of respiratory mechanics in patients with acute respiratory distress syndrome and association with outcome: a multicentre clinical study. Intensive Care Med. 2022;48(7):888–98.
Mauri T, Yoshida T, Bellani G, Goligher EC, Carteaux G, Rittayamai N, Mojoli F, Chiumello D, Piquilloud L, Grasso S, Jubran A, Laghi F, Magder S, Pesenti A, Loring S, Gattinoni L, Talmor D, Blanch L, Amato M, Chen L, Brochard L, Mancebo J, PLeUral pressure working Group (PLUG—Acute Respiratory Failure section of the European Society of Intensive Care Medicine). Esophageal and transpulmonary pressure in the clinical setting: meaning, usefulness and perspectives. Intensive Care Med. 2016;42(9):1360–73.
Chiumello D, Gallazzi E, Marino A, Berto V, Mietto C, Cesana B, et al. A validation study of a new nasogastric polyfunctional catheter. Intensive Care Med. 2011;37:791–5.
Practice Guidelines for Central Venous Access. 2020: An Updated Report by the American Society of Anesthesiologists Task Force on Central Venous Access. Anesthesiology 2020;132(1):8–43.
Hamzaoui O, Teboul JL. Central venous pressure (CVP). Intensive Care Med. 2022;48(10):1498–500.
Roger C, et al. Comparison of different techniques of central venous pressure measurement in mechanically ventilated critically ill patients. Br J Anaesth vol. 2017;118:2: 223–31.
Addabbo T, Cataldo G, Cevenini G, Fort A, Franchi F, Moretti R, Mugnaini M, Scolletta S, Vignoli V. A Measurement System to Estimate the Pleural pressure from the CVP for Respiratory System Monitoring. IEEE Trans Instrum Meas. 2019;68(7):2469–78.
Curtis N, Sessler MS, Gosnell MJ, Grap GM, Brophy PV, O’Neal KA, Keane EP, Tesoro, Elswick RK. The Richmond agitation–sedation scale. Am J Respir Crit Care Med. 2022;10:1338–44.
Chiumello D, Coppola S, Froio S, Mietto C, Brazzi L, Carlesso E, Gattinoni L. Time to reach a new steady state after changes of positive end expiratory pressure. Intensive Care Med. 2013;39:1377–85.
Amato MB, Meade MO, Slutsky AS, Brochard L, Costa EL, Schoenfeld DA, Stewart TE, Briel M, Talmor D, Mercat A, Richard JC, Carvalho CR, Brower RG. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med. 2015;372(8):747–55.
Meagher PF, Jensen RE, Weil MH, Shubin H. Measurement of respiration rate from central venous pressure in the critically ill patient. IEEE Trans Biomed Eng. 1966;2:54–7.
Perrino AC, Harris SN, Luther MA. Intraoperative determination of cardiac output using multiplane transesophageal echocardiography: a comparison to thermodilution. Anesthesiology. 1988;89:350–716.
Jansen JR, Schreuder JJ, Mulier JP, Smith NT, Settels JJ, Wesseling KH. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth. 2001;87:212–22.
DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing areas under two or more correlated reciever operating characteristics curves: a nonparamentric approach. Biometrics. 1988;44(3):837–45.
Ranieri VM, Rubenfeld GD, Thompson BT, et al. Acute respiratory distress syndrome: the Berlin definition. JAMA. 2012;307:2526–33.
Timsit JF, Baleine J, Bernard L, et al. Expert consensus-based clinical practice guidelines management of intravascular catheters in the intensive care unit. Ann Intensiv Care. 2020;10(1):118.
Hylkema BS, Barkmeijer-Degenhart P, van der Mark TW, Peset R, Sluiter HJ. Central venous versus esophageal pressure changes for calculation of lung compliance during mechanical ventilation. Crit Care Med. 1983;11(4):271–5.
Hedstrand U, Jakobson S, Nylund U, Sterner H. The concordance of respiratory fluctuations in Oesophageal and central venous pressures. Upsala J Med Sci. 1976;81:49–53.
Flemale A, Gillard C, Dierckx J. Comparison of central venous, oesophageal and mouth occlusion pressure with water-filled catheters for estimating pleural pressure changes in healthy adults. Eur Respir J. 1988;1:51–7.
Umbrello M, Cereghini S, Muttini S. Respiratory variations of central venous pressure as indices of Pleural pressure swings: a narrative review. Diagnostics. 2022;13:1022.
Lassola S, Miori S, Sanna A, et al. Central venous pressure swing outperforms diaphragm ultrasound as a measure of inspiratory effort during pressure support ventilation in COVID-19 patients. J Clin Monit Comput. 2022;36(2):461–71.
Colombo J, Spinelli E, Grasselli G, Pesenti AM, Protti A. Detection of strong inspiratory efforts from the analysis of central venous pressure swings: a preliminary clinical study. Minerva Anestesiol. 2020;86(12):1296–304.
Gattinoni L, Chiumello D, Caironi P, Busana M, Romitti F, Brazzi L, Camporota L. COVID-19 pneumonia: different respiratory treatments for different phenotypes? Intensive Care Med. 2020;14:1–4.
Magder SA, Georgiadis G, Cheong T. Respiratory variations in right atrial pressure predict response to fluid challenge. J Crit Care. 1992;7:76–85.
Sato R, Dugar S, Cheungpasitporn W, et al. The impact of right ventricular injury on the mortality in patients with acute respiratory distress syndrome: a systematic review and meta-analysis. Crit Care. 2021;25(1):172.
De Backer D, Vincent JL. Should we measure the central venous pressure to guide fluid management? Ten answers to 10 questions. Crit Care. 2018;22(1):43.
Xuan He MS, et al. Use of Cardiac Cycle locating to minimize the influence of Cardiac artifacts on esophageal pressure measurement during dynamic occlusion test. Respir Care. 2018;63(2):169–76.
Hartford CG, van Schalkwyk JM, Rogers GG, Turner MJ. Primate pleuroesophageal tissue barrier frequency response and esophageal pressure waveform bandwidth in health and acute lung injury. Anesthesiology. 2000;92(2):550–8.
Schuessler TF, Gottfried SB, Goldberg P, Kearney RE, Bates JH. An adaptive filter to reduce cardiogenic oscillations on esophageal pressure signals. Ann Biomed Eng. 1998;26(2):260–7.
Cheng YP, Wu HD, Jan GJ, Wang CY. Removal of cardiac beat artifact in esophageal pressure measurement via a modified adaptive noise cancellation scheme. Ann Biomed Eng. 2001;29(3):236–43.
Repessé X, Vieillard-Baron A, Geri G. Value of measuring esophageal pressure to evaluate heart-lung interactions-applications for invasive hemodynamic monitoring. Ann Transl Med. 2018;6(18):351.
Acknowledgements
Not applicable.
Funding
The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.
Open access funding provided by Università degli Studi di Siena within the CRUI-CARE Agreement.
Author information
Authors and Affiliations
Contributions
FF (the corresponding author) conceived and designed the study, collected and analysed the data, drafted the manuscript, and approved the submitted version of the manuscript. ED collected the data, analysed the recorded waveforms and drafted the manuscript, and approved the submitted version of the manuscript. AF collected and analysed the data and drafted the manuscript, and approved the submitted version of the manuscript. CB and DM participated in the data collection, and approved the submitted version of the manuscript. GC, GC and TA developed the digital filtering software, were responsible for the informatic issues, and approved the submitted version of the manuscript. SS, CAV, FST and SSc revised the manuscript, and approved the submitted version of the manuscript.
Corresponding author
Ethics declarations
Ethical approval
The study was approved by the local Ethics Committee (Comitato Etico Regionale per la Sperimentazione Clinica della Regione Toscana; PPL2-14925). Written informed consent was obtained from the patient’s next of kin before starting the study protocol.
Consent to participate
Written informed consent was obtained from the parents.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Franchi, F., Detti, E., Fogagnolo, A. et al. Estimation of the transpulmonary pressure from the central venous pressure in mechanically ventilated patients. J Clin Monit Comput (2024). https://doi.org/10.1007/s10877-024-01150-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10877-024-01150-5