Abstract
Background
A new fecal diverting device (FDD) was fabricated for fecal diversion from the proximal colon above the anastomosis to outside the anus for protecting the rectal anastomosis. The aim of the study is to evaluate the safety and effectiveness of the FDD.
Methods
After a pilot study, a prospective observational trial was performed in 34 mongrel dogs. The experiment comprised of segmental resection and anastomosis of the colon, fixation of the FDD, and observation for 3 weeks (n = 15) and more than 3 weeks (n = 19) without initiation of parenteral nutrition.
Results
Four cases of perioperative death unrelated to the FDD were excluded. Twenty-six (87 %) of the 30 dogs survived. Sixteen (53 %) dogs were able to retain the FDD for more than 3 weeks until 82 days. The autopsy findings revealed that four (15 %) dogs showed colonic wall erosions and mucosal scarring respectively at the band fixation area without evidence of serious septic complications. The surviving dogs retained the FDD for more than 6 days. Mortality occurred in four of the five dogs that expelled the FDD within three postoperative days. A closed abscess cavity as the evidence of anastomotic leakage was noted in seven (23 %) of the surviving dogs.
Conclusions
The newly designed fecal diverting device can be retained for more than 3 weeks until 82 days without any serious complications. The FDD may prevent sepsis in case of anastomotic leakage if it is retained for more than 6 days.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Background
Anastomotic leakage is the most serious complication of bowel surgery, especially low rectal anastomosis. Various factors contribute to anastomotic leakage. Irving and Goliger thought that residual fecal matter might be a critical cause of anastomotic leakage and that preoperative bowel preparation can reduce the incidence of anastomotic leakage [1]. Recently, there have been doubts whether residual fecal matter can be a major cause of anastomotic leakage [2]. However, most surgeons perform traditional preoperative bowel preparation in patients undergoing colorectal surgery [3].
Defunctioning stoma, a conventional fecal diversion method, is commonly used when anastomotic leakage is a concern. Theoretically, defunctioning stoma prevents catastrophic septic complications by protecting the anastomotic area against further fecal contamination, even if there is anastomotic leakage [4].
Ravo and Ger introduced a technique that could prevent anastomotic complications by diverting the fecal stream from the anastomotic area using a tube device, the Coloshield™ (Deknatel, Inc., Fall River, MA) [5, 6]. Other trials using a condom or a Valtrac™ (biofragmentable anastomosis ring (BAR; United States Surgical, Princeton, NJ))-secured intracolonic bypass (VIB) have reported good results [7, 8]. However, despite successful reports of various devices and methods, there is currently no single device that can be used clinically. The reasons for this could be technical difficulties, lack of control over the duration of device retention inside the bowels, and serious complications, such as bowel perforation. We have designed a new device that diverts the fecal stream similar to that with the Coloshield™, but a different fixation method is used for the application of this new device. Thus, not only can the device be retained inside the bowels as long as required, but it can also be removed anytime. In this animal study, we attempted to evaluate the safety and effectiveness of this experimental fecal diverting device.
Methods
Design
This study was designed as a prospective observational trial. A pilot study was performed prior to this study for determination of a stable maintenance schedule for diet and bowel movements while using the experimental device in eight mongrel dogs.
A newly designed device
A newly designed device called the fecal diverting device (FDD) is a tubular device made of silicone. There are two tire-like dumbbell-shaped outer balloons on the head portion of the device which help to fix the device on the colon proximal to the anastomotic area without sutures. A nonabsorbable polyester mesh (Parietex®; Sofradim Co., Trevoux, France) band was used to fix the head portion of the device externally on the colon (Fig. 1). Inside the head portion of the device, there is an inner balloon, which permits or blocks the flow of the bowel contents. The tail, i.e., the thin tube below the head portion, is long enough to hang out from the anus. Fecal matter enters from the proximal colon into the head portion of the device and exits from the anus through the tail portion of the device, thereby avoiding fecal contamination of the anastomosis.
Animal experimental procedures
Thirty-four healthy mongrel dogs (males 12, 15–27 kg, median weight 19 kg) were used in the study. Dogs were kept in the cages where they would be placed after surgery for at least 5 days for acclimatization to the environment prior to surgery. There was no preoperative fasting or bowel preparation. However, second-generation cephalosporin was injected 30 min prior to surgery.
The operation was performed using a low midline incision under intravenous ketamine anesthesia. After laparotomy, intraoperative colonic lavage was performed through the anus using the New Intraoperative Colonic Irrigator (NICI®, MITech Co., Seoul, Korea) [9] in order to achieve easy manipulation of the colon. For irrigation, approximately 2,000 cm3 of warm tepid water was used; the washout procedure took 5 min. We designed an ischemic anastomotic model for easier anastomotic disruption in order to be able to assess the efficacy of the FDD more accurately.
Colonic resection and anastomosis procedure was performed at approximately 10 cm from the anus, around the inferior mesenteric artery. An avascular colonic segment measuring approximately 5 cm was created by devascularization of the pericolic mesenteric vessels and inferior mesenteric artery for inducing ischemia at the anastomotic area. A circular stapler (Ethicon Endo-Surgery, Johnson & Johnson, USA) was introduced through the anus and advanced up to the devascularized segment. On this ischemic colonic segment, a 5–0 silk tie was placed between the anvil and the stapler body for resection and anastomosis. Approximately 2 cm of the colonic segment was resected using this stapling procedure.
After stapling, the head portion of the FDD was inserted into the anus with the inner balloon inflated and the outer balloons deflated. The FDD was passed through the anastomotic ring and was placed in the colon at approximately 5–10 cm proximal to the anastomotic ring. In order to fix the device, first, a mesh band was placed externally between the two outer balloons. Afterwards, the two outer balloons were inflated into a dumbbell shape so that the FDD would not move away from the site. Several silk 3–0 stitches were applied for fixing the mesh band on the colon in order to prevent the movement of the mesh band (Fig. 2). The tail portion of the device that was hanging outside the anus was severed at the anal level in order to protect the device from damage due to external factors or by the dogs themselves. After the surgical procedure, dogs were kept in cages for a duration of 3 weeks (n = 15) and more than 3 weeks (n = 19). Water and dog food were kept in the cages postoperatively. In order to ease the flow of the fecal matter through the device after surgery, lactulose (2 cm3/kg/day) and polyethylene glycol solution (50 cm3/kg/day) were thoroughly mixed in food and water, either individually or in combination. Initially, 15 dogs were kept under observation for a period of 3 weeks. Then, 19 dogs were kept under observation for as long as possible, without any set timeframe. The dogs were monitored for unexpected events throughout the experimental period over closed-circuit television (CCTV). Dogs were killed by administration of excessive doses of ketamine for evaluating the intraabdominal findings.
If the device was expelled before 3 weeks, the observation period was 5–7 days longer in order to examine the follow-up effect of early expulsion of the FDD. We removed the FDD in 15 dogs if it was retained until 3 weeks. We experienced successful removal of the FDD without anesthesia in the pilot study by pulling the tail portion (composed of a thin silicone tube and catheter) of the FDD after deflating both the outer balloons.
Objectives
The primary end points of the study were identification of possible bowel wall injury at the FDD fixation area and duration of FDD retention inside the bowels. The secondary end point was anastomotic complications. In addition, intraabdominal findings and wound complications were also evaluated.
Gross findings on examination were intraperitoneal adhesions, bowel injury, inflammation around the mesh area or anastomotic area, and any other unexpected findings. The segment including the mesh band and anastomosis was resected and evaluated macroscopically and microscopically for identification of the various mucosal and bowel wall changes. After macroscopic examination, representative sections were taken from the colon for light microscopic examination. Tissues were fixed in 10 % formalin and embedded in paraffin. Four-micrometer sections were stained with hematoxylin and eosin (H&E).
Expelled FDD, CCTV findings, and mesh band status were analyzed for evaluation of the causes of premature FDD expulsion. In this experiment, anastomotic leakage was defined as visible disruption of anastomosis, secondary peritonitis, and presence of intraperitoneal gas or fecal soilage. Evidence of anastomotic leakage was defined as the presence of a closed abscess cavity, mucosal or muscular discontinuation in the anastomotic area.
In case of premature death before 3 weeks, an autopsy was performed within 24 h after death. The dogs that died during surgery or those that died due to any cause other than problems caused by the FDD and mesh band were excluded from the study.
Statistical analysis
We measured the mortality incidence rate, bowel injury rate, duration of FDD retention inside the bowels for more than 3 weeks, anastomotic leakage rate, and the 95 % confidence interval (CI) for each rate. These data were obtained using statistical package SPSS version 19.0.
Results
A total of 34 dogs were enrolled in this study. The reasons for exclusion from the study, the duration of FDD retention, and FDD failure are presented in Fig. 3.
Exclusion
Four of the 34 dogs were excluded from this study due to the occurrence of major wound evisceration within 5–8 postoperative days. Two dogs were found dead, and the possible cause of death was hypovolemic shock due to massive hemorrhage from multiple injured visceral vessels. Another two dogs were found alive, however, with herniated and injured multiple visceral organs. They were euthanized with a high dose of ketamine. We think that the cause of wound evisceration was excessive straining in order to push the foreign body out from the anorectum, and this was frequently noted in the CCTV records. In autopsy findings, the FDD, FDD fixation area, and anastomotic area were intact. We included the remaining 30 dogs in this study.
Survival and duration of the FDD retention inside the bowels
Twenty-six (87 %; 95 % CI, 62.3–96.2 %) of the 30 dogs survived. In 16 of the 30 dogs (53 %; 95 % CI, 34.3–71.7 %), the device was retained inside the bowels for more than 3 weeks without mortality. The longest time period for which the device was retained inside the bowels was 82 days.
Among the 14 dogs (47 %; 95 % CI, 28.3–65.6 %) in whom the device was expelled within less than 3 weeks, 4 dogs (13 %, 95 % CI, 3.7–30.7 %) died within 6 postoperative days after expulsion of the device within 3 postoperative days. The causes of expulsion of the device were considered to be FDD prototype manufacturing errors and low mesh band tension (Table 1). Autopsy findings showed the accumulation of purulent ascitic fluid and fecal matter with a foul odor in the abdominal cavity. Partial anastomotic disruption was clearly noted, and it was considered to be the cause of generalized peritonitis. In the remaining ten dogs that survived, the device was expelled after 6–19 postoperative days in nine dogs and after 2 postoperative days in the remaining one dog. None of the dogs in whom the device was retained inside the bowels for more than 6 days died.
Bowel wall changes in the FDD fixation area of the resected bowel segment
Bowel wall erosion was noted in 4 of the 26 dogs that survived (15 %; 95 % CI, 4.3–34.8 %). Linear mucosal scars were noted in four dogs (15 %; 95 % CI, 4.3–34.8 %; Table 2). Bowel wall changes appeared to be aggravated with an increase in the duration of FDD fixation. However, despite the development of bowel wall changes, there was no gross evidence of local sepsis around the erosion or the mucosal defect site. Microscopic findings revealed that the erosion caused only a minimal to mild inflammatory response, rather than severe inflammation. Loss of mucosa and fibrotic tissue changes in the muscle layer were noted at the site of erosion and mucosal scarring of the bowel wall (Fig. 4a, a’, b, b’).
Findings in the anastomotic area showed evidence of anastomotic leakage
Continuous bowel wall defects with a closed abscess cavity were observed in seven dogs (23 %; 95 % CI, 9.9–42.3 %), and mucosal discontinuation was observed in three dogs (10 %; 95 % CI, 2.7–28.1 %). Closed abscess cavities were found in the dogs in which the FDD was retained inside the bowels for 7–21 postoperative days (Table 3). Most of the dogs who had a closed abscess cavity or mucosal disruption showed symptoms of postoperative ileus such as poor oral intake and decreased general activity for several days. Microscopic examination revealed evidence of a severe inflammatory process, such as cluster of inflammatory cells, foreign body granuloma, and severe extraluminal fibrosis of the closed abscess cavity.
Autopsy findings in the dogs that survived
On gross examination, mostly minimal to moderate adhesions were observed around the band fixation area and anastomotic area. Most of the times, no difficulties were encountered during the dissection of these adhesions. However, in ten dogs with evidence of anastomotic leakage, moderate to severe adhesions were observed around the anastomotic area.
Removal of the FDD
In the 15 dogs kept under observation for 3 weeks, the FDD was expelled spontaneously in 10 dogs. In the remaining five dogs, the FDD was removed manually after it was retained until 3 weeks. Even though there was minimal to mild resistance while pulling the FDD out, we successfully explantated the FDD in all the dogs. And there were no evidence of intraabdominal injuries around the FDD fixation area or anastomotic area in the autopsy findings after manual removal of the FDD. In the remaining 19 dogs, spontaneous expulsion of the device occurred before or after 3 weeks.
Other findings
The causes of early expulsion of the device were evaluated. Device failure occurred in 7 (50 %) of 14 dogs. One or both the outer balloons collapsed due to tearing or incomplete sealing of this prototype balloon device. Early device expulsion occurred in five (36 %) of seven remaining dogs as a result of the body portion of the FDD or catheters for outer balloons being bitten off by the dogs. In the remaining two dogs, the device was expelled easily as a result of a loose mesh band (Table 4.). Excessive straining in order to expel the device was noted in almost all dogs that were identified in CCTV recordings. Wound infection and partial wound disruption occurred in seven dogs (23 %) and two dogs (6 %), respectively.
Discussion
As reported in the literature, the first attempt towards prevention of anastomotic leakage by fecal diversion was performed by Lanfrank in 1565 [10]. In 1984, Rabo and Ger introduced a unique method for preventing fecal contamination of the anastomotic area by diverting the fecal matter through a silicone tube. This device was named Coloshield™, and it showed successful clinical results [11–18]. The same method with the use of a condom rather than the Coloshield™ also reported successful results [7, 19]. The VIB was applied in the animal experiments and clinical study with successful results [8, 20]. However, despite these successes, currently, there is no single device which is routinely used clinically.
The shortcomings of these various methods in which unique devices are used are that they are technically demanding, are associated with a risk for bowel perforation, and there is lack of control over the duration of device retention inside the bowels. In the case of Coloshield™ and condom, continuous suturing of only the mucosa and muscularis mucosa layer is recommended. Egozi et al. [21] reported a case of peritonitis caused by the use of Coloshield™ which led to erosion and perforation of the colonic wall because of a comparably solid tube material. This phenomenon has already been described in the Ross’s report of an animal experiment [22]. Lack of control over the duration of the device retention inside the bowels is a major problem associated with the use of these alternatives. Coloshield™ is expelled naturally when necrosis of the sutured mucosal and submucosal tissue occurs. The duration of the Valtrac™ ring inside the colon is also limited.
The density of collagen, which is the most crucial for healing of anastomosis, is the highest at 1 week after anastomosis [23]. The time duration of retention of Coloshield™ is 1–3 weeks and that for Valtrac-secured intracolonic bypass is usually 2 weeks. Therefore, in most cases, theoretically it allows safe anastomosis. However, if partial dehiscence occurs due to ischemic tissue necrosis, the anastomotic healing period can be prolonged. In various situations, the time duration required for a safe anastomosis might be prolonged compared to that in a normal healthy intestine. In the case of preoperative radiation therapy or chronic use of steroids, a period of 1–2 weeks cannot be considered sufficient for wound healing [24].
The reason for setting the time duration for device retention inside the bowels in this study was based on the fact that the majority of anastomotic leaks occur within 1–2 weeks. We thought that the 3-week period of fecal diversion was sufficient for anastomotic wound healing provided there were no other factors which could inhibit wound healing. As a result, in 53 % of dogs, the device was successfully retained inside the bowels for more than 3 weeks, and the longest period of device retention was 82 days. This result proves that the duration of fecal diversion can be controlled for extended periods using the FDD if the device is designed using high standards. Thus, it is thought that the FDD might function like an abdominal stoma in the patients, even in those who require an extended period of fecal diversion.
We did not evaluate the adverse effects of the FDD on the anal sphincter complex in this study; however, if we use the FDD in a clinical setting then this might be an issue that needs to be addressed. Similar examples found in the literature are the application of continent anal plug and Zassi Bowel Management System (Zassi Medical Evolutions, Fernandina Beach, Florida) [25, 26]. It has been reported that there was no sphincter injury at a maximum duration of retention of 37 and 59 days.
In a clinical setting, if there is postoperative anastomotic leakage, preceding stoma creation or Hartmann’s procedure is performed conventionally. Even though these are well-known techniques, another operation is needed in the near future for reducing the stoma. Furthermore, patients could experience various stoma-related complications before and after reduction of the stoma. The situation in our study is not similar to these situations of postoperative anastomotic leakage without preceding stoma creation; however, we can consider application of the FDD as an alternative procedure to defunctioning stoma or Hartmann’s operation. The benefits of application of the FDD could be avoiding stoma-related complications and another operation in the near future. However, we cannot be sure of the safety of the FDD because of the possible occurrence of unknown complications due to the use of the FDD. The most critical complication is the development of anastomotic stricture after leakage. Subsequent local sepsis would be more serious in cases without a stoma than in those with a stoma. In the current study, there was no case of failure of explantation of the device even though there was development of strictures after leakage. However, we need to perform further studies concerning this issue to confirm the safety of the FDD.
The most critical issue in this study is the safety of the fixation method of the FDD. It is well known that the bowel is very sensitive to compression. Possible injury to the band fixation area is the most important concern if the duration of device retention inside the bowels is prolonged. Erosion occurs following adjustable gastric banding procedures performed for morbid obesity or artificial anal sphincter procedures performed for fecal incontinence. As has been reported, the frequency of bowel wall erosion, as a long-term complication, also increases with time; in particular, the frequency of bowel wall erosion increases with the use of an artificial anal sphincter that compresses the bowel wall more than the adjustable gastric banding [27, 28]. In our study, the erosion supposedly resulted from continuous pressure between the band and the solid head portion of the device. If the band is kept too loose so as to decrease the pressure, the device would be expelled easily. On the other hand, if the band is kept too tight so as to hold the device securely, the pressure rises inevitably. We applied a slightly longer mesh band than the circumference of the bowel retaining the FDD. It is very encouraging to note that the fixation of the device with the application of tension resulted only in several erosions and mild mucosal scars, and did not lead to any critical complications or mortality. We used a nonabsorbable mesh band in this study because we could not find an ideal absorbable mesh material. It is also very encouraging to note that 85 % of the dogs that survived were erosion-free. We think that the problem of erosion can be solved if we use an ideal absorbable mesh material with an appropriate shape and tension in the future. We think that the use of absorbable mesh materials could reduce the long-term problem of erosion even if erosion develops.
Successful removal of the FDD was observed in the pilot study even though there were some difficulties because of poor cooperation of the dogs. The difficulties faced during the FDD removal in this study were not only the animal’s poor cooperation but also minimal to mild resistance while pulling the FDD out. We think that the FDD removal could be painful for the patient if it would be applied clinically although the explantation procedure might be successful. This study has a limitation regarding the safety with respect to the manual removal of the FDD because of the small sample size. However, the fact that there was no evidence of intraabdominal injuries around the FDD fixation area or anastomotic area in autopsy findings after manual removal of the FDD indicates that serious complications may not occur.
We observed early expulsion of the device in 14 dogs in this study. Early expulsion of the device not only results in loss of function but also in possible damage to the anastomosis as the device descends and passes through the anastomotic area. Four of 14 dogs with early device expulsion died of generalized peritonitis due to anastomotic breakdown. Expulsion of the FDD occurred within three postoperative days in these four dogs. However, expulsion of the FDD after 6 postoperative days did not result in death. This result was similar to the results reported by Ravo or Ye which show that in most cases, anastomosis was protected by fecal diversion for more than a week. We can assume that protection of the anastomosis for more than 6 days is sufficient in this dog model.
The most concerning anastomotic problem occurs during extraperitoneal rectal anastomosis in colorectal surgery. We did not use the distal rectum in the study because anastomosis near the anus can be damaged in dogs due to straining and biting. We tried to make the anastomotic complications in the colon easier for the substitution of low rectal anastomosis. The experimental animal model developed for this study had an inadequate blood supply at the anastomotic site which can result in a high anastomotic complication rate, and the efficacy of the FDD was assessed against this background. According to the results, evidence of anastomotic leakage based on autopsy findings was noted in seven dogs with a closed abscess cavity, in three dogs with mucosal discontinuation, and in four dogs with generalized peritonitis due to anastomotic leakage. We think that in this study, 46 % of anastomotic breakdown developed due to the use of this experimental model which provided an excellent background.
Regarding early expulsion of the device, we think that the problem involving experimental animals is in itself not a big one; however, we can understand that there might be some discomfort to the experimental animals because of the procedure. The problem regarding manufacturing of safer devices and applying appropriate band tension can be solved, and it is a subject for future trials.
In conclusion, to the best of our knowledge, the technique of retaining an intraluminal solid device with an extraluminal band at the site of a hollow viscus has not been reported in the literature until now. In this study, the newly designed fecal diverting device (FDD) was retained inside the bowels for more than 3 weeks, and it effectively protected the anastomosis against fecal contamination by diverting the fecal stream. Erosion developed in 15 % of cases; however, there were no cases of septic or critical complications. The longest duration of device retention was 82 days. In this animal study, despite the occurrence of anastomotic leakage, the FDD was able to prevent a catastrophic cascade if it was retained inside the bowels for more than 6 days. Further studies with the use of an absorbable mesh band and reliable device are needed in order to verify the safety of the FDD before its clinical application.
References
Irvin TT, Goligher JC (1973) Aetiology of disruption of intestinal anastomoses. Br J Surg 60:461–464
Guenaga KF, Matos D, Castro AA, Atallah AN, Wille-Jorgensen P (2011) Mechanical bowel preparation for elective colorectal surgery. Cochrane Database Syst Rev 9:CD001544
Wong NY, Eu KW (2005) A defunctioning ileostomy does not prevent clinical anastomotic leak after a low anterior resection: a prospective, comparative study. Dis Colon Rectum 48:2076–2079
Montedori A, Cirocchi R, Farinella E, Sciannameo F, Abraha I (2010) Covering ileo- or colostomy in anterior resection for rectal carcinoma. Cochrane Database Syst Rev 5:CD006878
Ravo B, Ger R (1984) A preliminary report on the intracolonic bypass as an alternative to a temporary colostomy. Surg Gynecol Obstet 159:541–545
Ravo B, Ger R (1984) Intracolonic bypass by an intraluminal tube. An experimental study. Dis Colon Rectum 27:360–365
Yoon WH, Song IS, Chang ES (1994) Intraluminal bypass technique using a condom for protection of coloanal anastomosis. Dis Colon Rectum 37:1046–1047
Ye F, Wang D, Xu X, Liu F, Lin J (2008) Use of intracolonic bypass secured by a biodegradable anastomotic ring to protect the low rectal anastomosis. Dis Colon Rectum 51:109–115
Kim JH, Shon DH, Kang SH, Jang BI, Chung MK, Kim JH, Shim MC (2005) Complete single-stage management of left colon cancer obstruction with a new device. Surg Endosc 19:1381–1387
Fraser I (1982) An historical perspective on mechanical aids in intestinal anastomosis. Surg Gynecol Obstet 155:566–574
Carpenter D, Bello J, Sokol TP, Sackier J, Carroll B, Wood C, Lugo D, Morgenstern L (1990) The intracolonic bypass tube for left colon and rectal trauma. The avoidance of a colostomy. Am Surg 56:769–773
Keane PF, Ohri SK, Wood CB, Sackier JM (1988) Management of the obstructed left colon by the one-stage intracolonic bypass procedure. Dis Colon Rectum 31:948–951
Falcone RE, Wanamaker SR, Santanello SA, Carey LC (1992) Colorectal trauma: primary repair or anastomosis with intracolonic bypass vs. ostomy. Dis Colon Rectum 35:957–963
Ravo B (1987) The intracolonic bypass procedure. Int J Colorectal Dis 2:38–42
Ravo B, Metwally N, Castera P, Polansky PJ, Ger R (1988) The importance of intraluminal anastomotic fecal contact and peritonitis in colonic anastomotic leakages. An experimental study. Dis Colon Rectum 31:868–871
Ravo B, Mishrick A, Addei K, Castrini G, Pappalardo G, Gross E, Sackier JM, Wood CB, Ger R (1987) The treatment of perforated diverticulitis by one-stage intracolonic bypass procedure. Surgery 102:771–776
Launer DP, Sackier JM (1991) Pouch-anal anastomosis without diverting ileostomy. Dis Colon Rectum 34:993–998
Rosati C, Smith L, Deitel M, Burul CJ, Baida M, Borowy ZJ, Bryden P (1992) Primary colorectal anastomosis with the intracolonic bypass tube. Surgery 112:618–622, discussion 622–613
Ruiz PL, Facciuto EM, Facciuto ME, Rodriguez Otero JC, Pigatto J, Cominelli H (1995) New intraluminal bypass tube for management of acutely obstructed left colon. Dis Colon Rectum 38:1108–1109
Chen TC, Yang MJ, Chen SR, Chang CP, Chi CH (1997) Valtrac-secured intracolonic bypass device: an experimental study. Dis Colon Rectum 40:1063–1067
Egozi L, Sorrento JJ, Golub R, Schultz EH (1993) Complication of the intracolonic bypass. Report of a case. Dis Colon Rectum 36:191–193
Ross H (1987) The effect of an intraluminal tube used as an internal drain on the healing of the rat colon. Dis Colon Rectum 30:591–594
Braskén P, Renvall S, Lehto M (1991) Fibronectin, laminin, and collagen types I, III, IV and V in the healing rat colon anastomosis. Br J Surg 78:1048–1052
Bullard KM, Trudel JL, Baxter NN, Rothenberger DA (2005) Primary perineal wound closure after preoperative radiotherapy and abdominoperineal resection has a high incidence of wound failure. Dis Colon Rectum 48:438–443
Kim J, Shim MC, Choi BY, Ahn SH, Jang SH, Shin HJ (2001) Clinical application of continent anal plug in bedridden patients with intractable diarrhea. Dis Colon Rectum 44:1162–1167
Keshava A, Renwick A, Stewart P, Pilley A (2007) A nonsurgical means of fecal diversion: the Zassi Bowel Management System. Dis Colon Rectum 50:1017–1022
Cherian PT, Goussous G, Ashori F, Sigurdsson A (2010) Band erosion after laparoscopic gastric banding: a retrospective analysis of 865 patients over 5 years. Surg Endosc 24:2031–2038
Devesa JM, Rey A, Hervas PL, Halawa KS, Larranaga I, Svidler L, Abraira V, Muriel A (2002) Artificial anal sphincter: complications and functional results of a large personal series. Dis Colon Rectum 45:1154–1163
Open Access
This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Kim, J.H., Jung, S.H., Kim, YJ. et al. Long-term fecal diverting device for the prevention of sepsis in case of colorectal anastomotic leakage: an animal experiment. Int J Colorectal Dis 28, 477–484 (2013). https://doi.org/10.1007/s00384-012-1580-x
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00384-012-1580-x