Introduction

Selenium is known to be an essential trace element in terms of antioxidant activity, but it exhibits toxicity at high concentrations (Lobinski et al. 2000). Selenium is widely used not only in photography, pharmaceutical production, and rubber vulcanization but also in the fields of ceramics, glass, photovoltaics, pigments, rectifier semiconductors, and steel (Foster and Sumar 1997). However, there is a very narrow range of difference between beneficial and harmful levels of health. Daily selenium intake less than 0.1 mg per 1 kg of body causes selenium deficiency. However, the daily ingestion of selenium exceeding 1 mg per 1 kg of body can be toxic to an average person (Suzuki et al. 1995). The diseases caused by selenium deficiency were cardiovascular disease, impaired immune function, aging promotion, etc. (Alissa et al. 2003). Poisoning symptoms are a typical metallic taste in the mouth, severe irritation in the respiratory system, pulmonary edema, and garlic-specific odors from respiration and sweat (Navarro-Alarcón and López-Martínez 2000). The recommended intake of selenium by the World Health Organization was 30 to 40 mg per day for adult men and women (Trace elements in human nutrition and health. World Health Organ (WHO) 1996).

In general, the main route through which people are exposed to selenium is food, followed by water and air (Selinus 2013). The selenium content of food produced from the soil is highly dependent on the amount of selenium present in the soil. Selenium concentrations in soils around the world are generally varied between 0.01 and 1000 mg/kg (Thomson 2004). The amount of selenium present in the soil varied depending on the country and region, and the amount of selenium in a plant varied depending on the ability of plants to absorb and accumulate selenium (Dumont et al. 2006). We intake most selenium from bread, cereal, meat, and poultry. Selenium concentrations are lower in animal foods than in vegetable foods (Mahima et al. 2012). Plants preferentially absorb selenate and convert it to selenium organic compounds (Finley 2005). In the case of aquatic organisms, selenium can be accumulated and expanded through the food chain (Sandholm et al. 1973). Selenium is abundant in fish viscera and meat, but the utilization rate of selenium in the body is the largest by selenium rich cereals and grains. Dairy products, fruits, and vegetables have the lowest selenium content (Brown and Arthur 2001).

The four natural oxidation states of selenium are elemental selenium, selenide (Se2−), selenite (Se4+), and selenate (Se6+) (Fio and Fujii 1990). Organic selenium compounds are the major selenium species in grains and vegetables, while inorganic selenium is abundant in water (Xiong et al. 2018). Inorganic selenium species such as Se (VI) and Se (IV) are more toxic than organic selenium such as dimethyl selenide, dimethyl diselenide, trimethyl selenonium ion, Se-cystine (Se-Cys), selenourea, selenobetaine, and Se-methionine (Se-Met) (Rohn et al. 2018). Many studies showed that selenium existed in various forms of inorganic and organic compounds in living organisms and foods, and its toxicity varied depending on the chemical species (Garousi 2015). Thus, it is difficult to evaluate the exact toxicity only by total amount of selenium in a sample. Accordingly, it is very important and meaningful to separate and quantify the different selenium species rather than analyzing the total selenium in the various samples.

In the last decade, the most work for selenium speciation has been done by the chromatographic method for separation of the selenium species using IC-ICP-MS, HPLC-ICP-MS, and ESI-MS for the detection of the separated selenium species (Michalke 2003; Roberge et al. 2003; Li et al. 2010). The method has been universally used because it was very sensitive and accurate. LIBS has been used for chromium and arsenic speciation after the elemental species were separated by ion exchange membrane (Dockery et al. 2005; Nam et al. 2018). LIBS has several advantage in terms of the simple sample preparation, fast analysis time, and cost advantages.

In this study, the LIBS coupled with ion exchange membrane was used for the determination of inorganic selenium. The solid phase extraction (SPE) membrane was used to separate inorganic selenium, and those were detected by laser-induced breakdown spectroscopy (LIBS). A syringe-type SPE membrane filter has been widely used to remove the impurities in the sample solution for the sample preparation. However, in our study, the inorganic selenium retained on the disk was directly analyzed by LIBS. The pH value of the sample was adjusted to 4 using ammonium hydroxide and phosphoric acid for the efficient separation of inorganic selenium species. The recovery efficiency of Se (VI) by LIBS was 100.2%. From the results, it was confirmed that inorganic selenium species could be separated and detected using SPE membrane and LIBS.

Methods

Ion exchange membrane

Inorganic selenium (Se (VI)) can be separated using ion exchange membranes as anionic form at pH 7. Anion exchange was performed on 47-mm round segments of a solid phase extraction disk (EmporeTM, 3 M Corp, St. Paul, MN). The anion exchange polymer has a polymer support functionalized by trimethyl ammonium propyl group. Filtration was performed using a locally glass funnel with a diameter of 47 mm. The membrane was conditioned according to the manufacturer’s instruction that was described in procedure in detail. The standards were directly filtered without any pretreatment. Since the samples were prepared from Se (IV) and Se (VI), Se (IV) was oxidized to Se (VI) and passed through the membrane. Then, the membranes were dried in the oven, of which temperature was 40°C, for about 24 h and analyzed by LIBS.

Apparatus

The commercial LIBS instrument (RT100-EC, Applied Spectra, Inc.) was used with a Q-switched Nd:YAG laser (1064 nm, 7 ns, maximum 50 mJ/pulse, Polaris, New Wave Research) and a gateable 5-channel charge-coupled device (CCD) spectrometer (190–800 nm wavelength coverage, ~ 0.1 nm resolution) for recording the LIBS spectra. The laser beam was focused on the substrate surface through an objective lens (× 5 magnification, 35 mm working distance). The calibration curve was obtained by focusing 10-mJ laser pulse energy on the substrate surface. The plasma emission was collected through two lenses and sent to the spectrometer via an optical fiber bundle. The CCD detection gate with a 1.05-ms width was delayed from the Q-switching event of the laser by 500 ns to optimize the signal-to-background ratio. The laser was operated at the repetition rate of 10 Hz. The laser beam spot size was set to 150 μm in diameter on the sample surface. To record a single LIBS spectrum, signals from 1000 laser shot were accumulated. During the measurement, the sample stage was translated at the rate of 1 mm/s.

Chemicals and reagents

Stock solutions of Se (VI) were prepared by sodium selenate (≥ 95%, Sigma-Aldrich Co.), and Se (IV) were prepared by sodium selenite (98%, Sigma-Aldrich Co.) by using distilled deionized water as a solvent. The concentration of Se in each stock solution was set to 1000 mg/kg. The stock solutions were kept in the refrigerator of which temperature was set to 2 °C. Calibration standards and samples were prepared by using distilled deionized water from the stock and ammonium hydroxide (≥ 28%, Sigma-Aldrich), and phosphoric acid (85–90%, Sigma-Aldrich) was used to adjust pH (Martínez-Bravo et al. 2001). Hydrogen peroxide (35%, 1 mL) was used to oxidize Se (IV) to Se (VI).

Procedure

The membranes need to be conditioned before they are used. Acetone (10 mL), methanol (10 mL), distilled deionized water (10 mL), aqueous solution of sodium hydroxide (1 M, 10 mL), and distilled deionized water (10 mL) were flowed subsequently through them for their conditioning. Inorganic selenium solutions were filtered through the conditioned extraction membrane filter disk (~ 2 mL/min) by suction using a hand vacuum pressure pump. The volume of each solution was set to 10 mL. After the filtration, Se (VI) was retained on the extraction membrane filter disk as it was shown in Fig. 1.

Fig. 1
figure 1

Diagram of procedure

Results and discussion

Optimum pH and flow rate

The pKa values of Se (VI) and Se (IV) are 2.70 and − 2.01, respectively (Peng et al. 2015; Wu and Sun 2016). Ionic characteristics of Se (IV) and Se (VI) highly depend on the pH of the solution. In order to separate Se (IV) and Se (VI) using ion exchange membrane, the two inorganic selenium species must exist in different forms such as neutral and ionic forms. But, in order to separate two kinds of inorganic selenium species using this principle, the pH value of the sample should be 2 or less. It is not suitable for this method because of its strong acidity. The recommended pH value ranges from 2 to 12 for the ion exchange membrane. Thus, Se (IV) was oxidized to Se (VI) and quantified in the form of total inorganic selenium. The ionic characteristics of Se (VI) were highly depend on the pH of the solution. So, the recoveries of Se (VI) collected on extraction membrane filter disk were investigated with the different pH of solution. The pH of the calibration standard and sample was adjusted by the 28% ammonium hydroxide. The small amounts of the concentrated ammonium hydroxide were added to the sample solutions in which the concentrations of Se (VI) had been set to sample concentration. Four sample solutions were prepared with the same Se (VI) concentrations at different pH values, 3.0, 5.0, 7.0, and 9.0. Figure 2a shows the variation of LIBS signal intensities of Se (VI) collected on the extraction membrane filter disk at the different pH. The error bars represent standard deviations for the corresponding five repetitive measurements. The signal intensity was the highest at pH 7. In the case of pH 3, 5, and 9, the signal intensity was lower than that measured at pH 7; so, Se (VI) could not be recovered in its entirety.

Fig. 2
figure 2

The signal intensity of Se (VI) from LIBS spectra. a Signal intensities of Se (VI) at different pH values (3.0, 5.0, 7.0, and 9.0) and b those at different flow rates (1.0, 2.0, 4.0, and 8.0 mL/min). A 10-mL solution was used for each analysis

In addition, the flow rate of sample influenced the recovery. Thus, the results were analyzed by adjusting the four different flow rates: 1, 2, 4, and 8 mL/min. Figure 2b shows the signal intensities at different flow rates. The error bars represent standard deviations for the corresponding five repetitive measurements. The flow rates were changed from 1 to 8 mL/min for optimization, and the signal intensity was the highest at 2 mL/min. Since the signal intensity at different flow rates (1, 4, and 8 mL/min) is lower than the signal intensity at the flow rate 2 mL/min, it can also be assumed that not all of Se (VI) has been recovered.

Determination of Se (VI)

Standard solutions for the determination of Se (VI) were prepared. A Se (IV) and Se (VI) standard was oxidized to Se (VI) before passed through the extraction membrane filter disk, and the anionic Se (VI) remained on the disk was measured by LIBS. Table 1 shows the recovery of Se (VI). The recoveries of 50 mg/kg and 100 mg/kg oxidized Se (VI) standard recovery were 98.5 % and 101.8 %, respectively.

Table 1 Determination of Se (VI) in pH 7

Calibration curves and detection limits

Five standard solutions were prepared with the pH, 7.0. Each solution was flowed through a membrane filter. The dried membrane filter was analyzed by LIBS. Membrane filters containing the standard Se (VI) with the concentration range from 10 to 200 mg/kg were prepared for LIBS calibration. Emission was collected for 200 laser shots per spectrum. The spectra in the wavelength regions, where the Se (I) and C (I) emission lines were observed, are shown in Fig. 3. Two Se (I) emission lines at 203.99 and 206.28 nm, indicated by “*” in Fig. 3a, were assigned based on the NIST Atomic Spectra Database. The intensities of these Se (I) lines show clear positive correlations with the concentration of Se in the calibration standards. Among them, one of the strongest lines at 203.99 nm was chosen for developing the calibration curve. For the five calibration standards, the C (I) line at 247.86 nm was observed with very similar intensities. This C (I) emission originates from the ablation of the filter material. This indicates that our measurement condition is found to be quite consistent from sample to sample. Thus, any additional data processing of intensity normalization was not applied.

Fig. 3
figure 3

LIBS spectra of the LIBS calibration standards with five different Se (VI) concentrations (10, 20, 50, 150, and 200 mg/kg) in the wavelength regions where the a Se (I) and b C (I) emission lines were observed. In (a), the two peaks denoted by “*” are the Se (I) emission lines at 203.99 and 206.28 nm. The C (I) lines at 247.86 nm, shown in (b), originated from the membrane filter material

The analytical calibration curve for Se (VI) was shown in Fig. 4. The calibration curves of Se (VI) showed good linearity in the range of 10–200 mg/kg. The measured Se (I) line intensities are fitted well with a linear function (R2 = 0.9983). The calibration curve shows linear dynamic ranges of 1–2 orders of magnitude for Se (VI). The correlation coefficients ranged from 0.99 to 1.00. The error bars represent two times of standard deviations. The limit of detection for Se (VI) was 11 mg/kg. The detection limit was defined as the concentration giving a signal equivalent to three times the noise, calculated from the standard deviation of 3 repetitive measurement of the background intensities.

Fig. 4
figure 4

Calibration curve of Se (VI) from LIBS analysis of the membrane filters. The filled squares represent the experimental values, and the red line is a linear fit of them (y = a + bx, with a = − 6.85 ± 5.05, b = 10.55 ± 0.06, and R2 = 0.9983)

Performances for spike water

The procedure presented was applied to the detection of Se (VI) in natural water samples from a tap water from Muan County in Republic of Korea. As a result, Se (VI) was not detected. Thus, Se (VI) was spiked into the tap water. The concentration of the spiked Se (VI) was determined to be 50.7 ± 0.2 mg/kg. This leads to the recovery efficiency of 101.3%. The results are listed in Table 2. The result described that the developed method can be applied to the real water samples such as rainwater and drinking water for the determination of inorganic selenium.

Table 2 Analysis of Se (VI) in sample with Se (VI) added to tap water

Further improvement of LOD

For reliable analysis of selenium in natural water or blood samples, the ppb-level LOD is required. Our current method provides the lower LOD (11 mg/kg). However, it can be improved further. It should be noted that the current LOD can be lowered much even retaining the current experimental setup. We investigated variation of the signal-to-noise ratio with the laser scan number. Each laser scan was performed accumulating 200 laser shots. Nine line scans were conducted consecutively. Then, nine spectra were produced by averaging different numbers of line scans from one to nine. Then, for each spectrum, the signal intensity was obtained by taking area under the baseline-subtracted Se emission peak at 203.93 nm. The noise was estimated by the root-mean-square-error of the baseline intensities in the wavelength region between 200.634 nm and 207.159 nm. To obtain the noise, the intensities around the Se emission peak were excluded (203.806–204.070 nm). In Fig. 5, the signal-to-noise ratio values were plotted with respect to the corresponding numbers of line scans averaged. The LOD of 11 mg/kg was estimated using three line scans. The corresponding signal-to-noise ratio is 0.30. If the number of line scans were increased from 3 to 9, the signal-to-noise ratio is increased to 1.03 by a factor of 3.4. It leads the LOD of 3.2 mg/kg. As can be seen in Fig. 5, the signal-to-noise ratio seems to increase further for the larger numbers of line scans averaged than 9. In addition, the ion exchange membrane was kept unsaturated in this experiment, so it has unused binding capacity for further pre-concentration of Se. From the calibration curve shown in Fig. 4, it can be found that the ion exchange membrane has not been saturated by 10 mL of the most concentrated Se standard solution (200 mg/kg). This implies that the current LOD performance can be improved further at least by a factor of 18 (= 200/11). Averaging more line-scan data and increasing the volume of solution, the LOD of 100–200 μg/kg might be obtained. However, the lower ppb-level analysis requires the higher-performance detection systems such as ICP-MS. The ICP-MS can be used in combination with laser-ablation sampling which was used for this LIBS analysis. Currently, the research for application of our simple selenium speciation method and the laser-ablation sampling to ICP-MS analysis is in progress.

Fig. 5
figure 5

Signal-to-noise ratio values with respect to the corresponding numbers of line scans averaged. The signal intensities were taken for the Se emission peak at 203.93 nm, and the noise values were estimated from the baseline intensities nearby the Se emission peak (see text)

Conclusion

We demonstrated a simple methodology for quantification of inorganic selenium in water. The ion exchange membrane was employed for selectively collecting Se (VI) species in watery. Se (IV) was oxidized to Se (VI) and adsorbed on the membrane in the form of inorganic selenium and analyzed by LIBS. LIBS analysis for the membrane holding Se (VI) species showed reliable analytical performances: measurement precision of a few percent relative standard deviation, limit of detection of 11 mg Se out of 1 kg of water, and linear dynamic range from 10 to 200 mg/kg.

It is possible to analyze inorganic selenium by ion exchange membrane filter and LIBS. The analytical method can be faster and simpler than conventional methods. To apply our methodology to real samples such as rainwater and drinking water, further improvement in the LOD performance is definitely necessary and currently in progress.