Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras

  • Original Article
  • Published:
Journal of Nuclear Cardiology Aims and scope

Abstract

Purpose

Estimation of left ventricular ejection fraction (LVEF) with equilibrium 99MTc-HSA equilibrium radionuclide angiography (MUGA) is frequently used for assessing cardiac function. The purpose of this study was to compare intra- and interobserver variation between three different gamma cameras.

Materials and Methods

Eighty-two patients, scanned in the same sequential order on the three cameras. Each acquisition was analyzed twice by two technologists. Inter- and intraobserver variations were calculated as the coefficient of variation and the 95% confidence interval for limits of agreement between each sequence of analyses for each of the three cameras.

Results

The lowest intraobserver variations in LVEF for the two NaI-detector cameras were 3.1% (−4.0% to 3.5%) for the planar and 3.4% (−4.2% to 4.5%) for SPECT (P ≤ 0.001-0.019), the highest result for the CZT SPECT camera was 2.6% (−2.9% to 3.1%). Similarly, interobserver variation was 4.8% (−4.8% to 6.4%) and 4.9% (−5.4% to 7.5%), respectively, for each of the NaI-detector cameras and 3.3% (−3.4% to 4.3%) for the CZT SPECT camera (P ≤ 0.001-0.008).

Discussion

The CZT detector camera was superior to both NaI detector cameras regarding intra- and interobserver variation. The CZT SPECT camera may identify changes in LVEF with greater certainty than its NaI detector-equipped counterparts.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3
Figure 4
Figure 5

Similar content being viewed by others

References

  1. Wackers FJ, Berger HJ, Johnstone DE, Goldman L, Reduto LA, Langou RA, et al. Multiple gated cardiac blood pool imaging for left ventricular ejection fraction: validation of the technique and assessment of variability. Am J Cardiol 1979;43:1159-66.

    Article  PubMed  CAS  Google Scholar 

  2. Harel F, Finnerty V, Gregoire J, Thibault B, Marcotte F, Ugolini P, et al. Gated blood-pool SPECT versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction. J Nuclear Cardiol 2010;17:427-34.

    Article  Google Scholar 

  3. Hiscock SC, Evans MJ, Morton RJ, Hall DO. Investigation of normal ranges for left ventricular ejection fraction in cardiac gated blood pool imaging studies using different processing workstations. Nuclear Med Commun 2008;29:103-9.

    Article  Google Scholar 

  4. Daou D, Coaguila C, Benada A. Comparison of interstudy reproducibility of equilibrium electrocardiography-gated SPECT radionuclide angiography versus planar radionuclide angiography for the quantification of global left ventricular function. J Nuclear Cardiol 2006;13:233-43.

    Article  Google Scholar 

  5. Harel F, Finnerty V, Ngo Q, Gregoire J, Khairy P, Thibault B. SPECT versus planar gated blood pool imaging for left ventricular evaluation. J Nuclear Cardiol 2007;14:544-9.

    Article  Google Scholar 

  6. Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential. Eur J Nuclear Med Mol Imaging 2010;37:1887-902.

    Article  Google Scholar 

  7. Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images. J Nuclear Med 2012;53:1897-903.

    Article  Google Scholar 

  8. Daou D, Coaguila C, Vilain D. Electrocardiograph-gated single photon emission computed tomography radionuclide angiography presents good interstudy reproducibility for the quantification of global systolic right ventricular function. Nuclear Med Commun 2007;28:391-9.

    Article  Google Scholar 

  9. Acampa W, Caprio MG, Nicolai E, Liuzzi R, De Luca S, Capasso E, et al. Assessment of poststress left ventricular ejection fraction by gated SPECT: comparison with equilibrium radionuclide angiocardiography. Eur J Nuclear Med Mol Imaging 2010;37:349-56.

    Article  Google Scholar 

  10. Godkar D, Bachu K, Dave B, Megna R, Niranjan S, Khanna A. Comparison and co-relation of invasive and noninvasive methods of ejection fraction measurement. J Nat Med Assoc 2007;99:1227-8.

    Google Scholar 

  11. Walker J, Bhullar N, Fallah-Rad N, Lytwyn M, Golian M, Fang T, et al. Role of three-dimensional echocardiography in breast cancer: comparison with two-dimensional echocardiography, multiple-gated acquisition scans, and cardiac magnetic resonance imaging. J Clin Oncol 2010;28:3429-36.

    Article  PubMed  Google Scholar 

  12. Groch MW, DePuey EG, Belzberg AC, Erwin WD, Kamran M, Barnett CA, et al. Planar imaging versus gated blood-pool SPECT for the assessment of ventricular performance: a multicenter study. J Nuclear Med 2001;42:1773-9.

    CAS  Google Scholar 

  13. Wells RG, Marvin B, Kovalski G, Ruddy TD. Planar radionuclide angiography with a dedicated cardiac SPECT camera. J Nuclear Cardiol 2013;20:358-66.

    Article  Google Scholar 

  14. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10.

    Article  PubMed  CAS  Google Scholar 

  15. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135-60.

    Article  PubMed  CAS  Google Scholar 

  16. Hacker M, Hoyer X, Kupzyk S, La Fougere C, Kois J, Stempfle HU, et al. Clinical validation of the gated blood pool SPECT QBS processing software in congestive heart failure patients: correlation with MUGA, first-pass RNV and 2D-echocardiography. Int J Cardiovasc Imaging 2006;22:407-16.

    Article  PubMed  Google Scholar 

  17. Heidendal GA, Bezemer PD, Koopman PA, den Hollander W, Teule GJ, van der Wall EE, et al. Reproducibility of ejection fraction measurements by gated equilibrium blood pool scintigraphy. Eur J Nuclear Med 1983;8:467-70.

    Article  CAS  Google Scholar 

  18. Schwartz RG, McKenzie WB, Alexander J, Sager P, D’Souza A, Manatunga A, et al. Congestive heart failure and left ventricular dysfunction complicating doxorubicin therapy. Seven-year experience using serial radionuclide angiocardiography. Am J Med 1987;82:1109-18.

    Article  PubMed  CAS  Google Scholar 

  19. Tonge CM, Fernandez RC, Harbinson MT. Commentary: current issues in nuclear cardiology [commentary]. British J Radiol 2008;81:270-4.

    Article  CAS  Google Scholar 

  20. Sipola P, Vanninen E, Jantunen E, Nousiainen T, Kiviniemi M, Hartikainen J, et al. A prospective comparison of cardiac magnetic resonance imaging and radionuclide ventriculography in the assessment of cardiac function in patients treated with anthracycline-based chemotherapy. Nuclear Med Commun 2012;33:51-9.

    Article  CAS  Google Scholar 

  21. Wright GA, Thackray S, Howey S, Cleland JG. Left ventricular ejection fraction and volumes from gated blood-pool SPECT: comparison with planar gated blood-pool imaging and assessment of repeatability in patients with heart failure. J Nuclear Med 2003;44:494-8.

    Google Scholar 

  22. De Bondt P, Claessens T, Rys B, De Winter O, Vandenberghe S, Segers P, et al. Accuracy of 4 different algorithms for the analysis of tomographic radionuclide ventriculography using a physical, dynamic 4-chamber cardiac phantom. J Nuclear Med 2005;46:165-71.

    Google Scholar 

  23. Massardo T, Jaimovich R, Lavados H, Gutierrez D, Rodriguez JC, Saavedra JM, et al. Comparison of radionuclide ventriculography using SPECT and planar techniques in different cardiac conditions. Eur J Nuclear Med Mol Imaging 2007;34:1735-46.

    Article  Google Scholar 

Download references

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Maria Maj Jensen MD.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Jensen, M.M., Schmidt, U., Huang, C. et al. Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras. J. Nucl. Cardiol. 21, 384–396 (2014). https://doi.org/10.1007/s12350-013-9844-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12350-013-9844-6

Keywords

Navigation