Abstract
Background
The success of THA largely depends on correct placement of the individual components. Traditionally, these have been placed freehand using anatomic landmarks, but studies have shown poor accuracy with this method.
Questions/purposes
Specifically, we asked (1) does using fluoroscopy lead to more accurate and greater likelihood of cup placement with the Lewinnek safe zone than does freehand cup placement; (2) is there a learning curve associated with the use of fluoroscopy for cup placement; (3) does the use of fluoroscopy increase operative time; and (4) is there a difference in leg length discrepancy between freehand and fluoroscopic techniques?
Methods
This series consisted of 109 consecutive patients undergoing primary THA, conversion of a previous hip surgery to THA, and revision THA during a 24-month period. No patients were excluded from analysis during this time. The first 52 patients had cups placed freehand, and then the next 57 patients had acetabular components placed using fluoroscopy; the analysis began with the first patient treated using fluoroscopy, to include our initial experience with the technique. The abduction, version, and limb length discrepancy were measured on 6-week postoperative pelvic radiographs obtained with the patient in the supine position. Operative time, sex, age, BMI, diagnosis, operative side, and femoral head size were recorded as possible confounders.
Results
Cups inserted freehand were placed in the ideal range of abduction (30°–45°) and anteversion (5°–25°) 44% of the time. With fluoroscopy, placement in the Lewinnek safe zone for both measures significantly increased to 65%. The odds of placing the cup in the Lewinnek safe zone for abduction and version were 2.3 times greater with the use of fluoroscopy (95% CI, 1.2–5.0; p = 0.03). Patients undergoing primary THAs (32 freehand, 35 C-arm) had cup placement in the safe zone for abduction and version 44% of the time freehand and 57% of the time with fluoroscopy, which failed to reach statistical significance. There was no difference in operative time, patient age, sex, operative side, diagnosis, limb length discrepancy, or femoral head size between the two groups.
Conclusions
The use of fluoroscopy to directly observe pelvic position and acetabular component placement increased the success of placement in the Lewinnek safe zone in this cohort of patients having complex and primary THAs. This is a simple, low-cost, and quick method for increasing successful acetabular component alignment. The study population included a large proportion of patients having complex THAs, and further validation of this technique in patients undergoing straightforward, primary THAs needs to be done to understand if similar gains in accuracy for component placement can be expected in that group.
Level of Evidence
Level III, therapeutic study. See the Instructions for Authors for a complete description of levels of evidence.
Similar content being viewed by others
References
Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.
Bragdon CR, Martell JM, Greene ME, Estok DM 2nd, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.
Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H. The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011;469:319–329.
Digioia AM 3rd, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, Levison TJ, Picard F. Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty. 2002;17:359–364.
Hosmer DW, Lemeshow S. Applied Logistic Regression. New York, NY: Wiley; 2000.
Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.
Katz JN, Losina E, Barrett J, Phillips CB, Mahomed NN, Lew RA, Guadagnoli E, Harris WH, Poss R, Baron JA. Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States medicare population. J Bone Joint Surg Am. 2001;83:1622–1629.
Katz MH. Multivariable Analysis : A Practical Guide for Clinicians and Public Health Researchers. New York, NY: Cambridge University Press; 2011.
Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.
Kjellberg M, Al-Amiry B, Englund E, Sjoden GO, Sayed-Noor AS. Measurement of leg length discrepancy after total hip arthroplasty: the reliability of a plain radiographic method compared to CT-scanogram. Skeletal Radiol. 2012;41:187–191.
Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.
Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.
Martell JM, Berkson E, Berger R, Jacobs J. Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:1111–1117.
Meermans G, Malik A, Witt J, Haddad F. Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty. Clin Orthop Relat Res. 2011;469:1677–1682.
Moskal JT, Capps SG. Acetabular component positioning in total hip arthroplasty: an evidence-based analysis. J Arthroplasty. 2011;26:1432–1437.
Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.
Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H. Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res. 2003;411:140–151.
Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD. Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am. 2003;85(suppl 4):56–63.
Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt on acetabular retroversion: a study of pelves from cadavers. Clin Orthop Relat Res. 2003;407:241–248.
Tsalafoutas IA, Tsapaki V, Kaliakmanis A, Pneumaticos S, Tsoronis F, Koulentianos ED, Papachristou G. Estimation of radiation doses to patients and surgeons from various fluoroscopically guided orthopaedic surgeries. Radiat Prot Dosimetry. 2008;128:112–119.
Wera GD, Ting NT, Moric M, Paprosky WG, Sporer SM, Della Valle CJ. Classification and management of the unstable total hip arthroplasty. J Arthroplasty. 2012;27:710–715.
Zhu J, Wan Z, Dorr LD. Quantification of pelvic tilt in total hip arthroplasty. Clin Orthop Relat Res. 2010;468:571–575.
Acknowledgments
We thank David Zurakowski PhD (Departments of Anesthesiology and Surgery, Boston Children’s Hospital, Boston, MA) for assistance with the statistical analyses in the current study.
Author information
Authors and Affiliations
Corresponding author
Additional information
Each author certifies that he or she, or a member of his or her immediate family, has no funding or commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.
All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research editors and board members are on file with the publication and can be viewed on request.
Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.
About this article
Cite this article
Beamer, B.S., Morgan, J.H., Barr, C. et al. Does Fluoroscopy Improve Acetabular Component Placement in Total Hip Arthroplasty?. Clin Orthop Relat Res 472, 3953–3962 (2014). https://doi.org/10.1007/s11999-014-3944-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11999-014-3944-8