The John Charnley Award: Risk Factors for Cup Malpositioning: Quality Improvement Through a Joint Registry at a Tertiary Hospital

  • Symposium: Papers Presented at the Hip Society Meetings 2010
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Few studies have examined factors that affect acetabular cup positioning. Since cup positioning has been linked to dislocation and increased bearing surface wear, these factors affecting cup position are important considerations.

Question/purposes

We determined the percent of optimally positioned acetabular cups and whether patient and surgical factors affected acetabular component position.

Methods

We obtained postoperative AP pelvis and cross-table lateral radiographs on 2061 consecutive patients who received a THA or hip resurfacing from 2004 to 2008. One thousand nine hundred and fifty-two hips had AP pelvic radiographs with correct position of the hip center, and 1823 had both version and abduction angles measured. The AP radiograph was measured using Hip Analysis Suite™ to calculate the cup inclination and version angles, using the lateral film to determine version direction. Acceptable ranges were defined for abduction (30°–45°) and version (5°–25°).

Results

From the 1823 hips, 1144 (63%) acetabular cups were within the abduction range, 1441 (79%) were within the version range, and 917 (50%) were within the range for both. Surgical approach, surgeon volume, and obesity (body mass index > 30) independently predicted malpositioned cups. Comparison of low versus high volume surgeons, minimally invasive surgical versus posterolateral approach, and obesity versus all other body mass index groups showed a twofold (1.5–2.8), sixfold (3.5–10.7), and 1.3-fold (1.1–1.7) increased risk for malpositioned cups, respectively.

Conclusions

Factors correlated to malpositioned cups included surgical approach, surgeon volume, and body mass index with increased risk of malpositioning for minimally invasive surgical approach, low volume surgeons, and obese patients. Further analyses on patient and surgical factors’ influence on cup position at a lower volume medical center would provide a valuable comparison.

Level of Evidence

Level II, prognostic study. See Guidelines for Authors for a complete description of levels of evidence.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Ali Khan MA, Brakenbury PH, IS R. Dislocation following total hip replacement. J Bone Joint Surg Br. 1981;63:214–218.

    PubMed  Google Scholar 

  2. Arthursson AJ, Furnes O, Espehaug B, Havelin LI, Soreide JA. Prosthesis survival after total hip arthroplasty—does surgical approach matter? Analysis of 19,304 Charnley and 6,002 Exeter primary total hip arthroplasties reported to the Norwegian Arthroplasty Register. Acta Orthop. 2007;78:719–729.

    Article  PubMed  Google Scholar 

  3. Bartz RL, Noble PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.

    CAS  PubMed  Google Scholar 

  4. Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87:762–769.

    Article  CAS  PubMed  Google Scholar 

  5. Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ. Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg. 2007;127:375–379.

    Article  CAS  PubMed  Google Scholar 

  6. Bradford L, Kurland R, Sankaran M, Kim H, Pruitt LA, Ries MD. Early failure due to osteolysis associated with contemporary highly cross-linked ultra-high molecular weight polyethylene. A case report. J Bone Joint Surg Am. 2004;86:1051–1056.

    Google Scholar 

  7. Bragdon CR, Martell JM, Greene ME, Estok DM II, Thanner J, Karrholm J, Harris WH, Malchau H. Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res. 2006;448:52–57.

    Article  PubMed  Google Scholar 

  8. Broers H, Jansing N. How precise is navigation for minimally invasive surgery? Int Orthop. 2007;31 Suppl 1:S39–S42.

    Article  PubMed  Google Scholar 

  9. Colwell CWJ. Instability after total hip arthroplasty. Current Orthop Prac. 2009;20:8–14.

    Article  Google Scholar 

  10. Conroy JL, Whitehouse SL, Graves SE, Pratt NL, Ryan P, Crawford RW. Risk factors for revision for early dislocation in total hip arthroplasty. J Arthroplasty. 2008;23:867–872.

    Article  PubMed  Google Scholar 

  11. De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.

    Article  PubMed  Google Scholar 

  12. Dorr LD. Acetabular cup position: the imperative of getting it right. Orthopedics. 2008;31:898–899.

    Article  PubMed  Google Scholar 

  13. Enocson ATJ, Tornkvist H, Lapidus L. Dislocation of hemiarthroplasty after femoral neck fracture: better outcome after the anterolateral approach in a prospective cohort study on 739 consecutive hips. Acta Orthop. 2008;79:211–217.

    Article  PubMed  Google Scholar 

  14. Gallo J, Havranek V, Zapletalova J. Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty. Int Orthop. 2010;34:19–26.

    Article  PubMed  Google Scholar 

  15. Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467:2362–2370.

    Article  PubMed  Google Scholar 

  16. Holt G, Murnaghan C, Reilly J, Meek RM. The biology of aseptic osteolysis. Clin Orthop Relat Res. 2007;460:240–252.

    CAS  PubMed  Google Scholar 

  17. Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.

    PubMed  Google Scholar 

  18. Jasty M, Goetz DD, Bragdon CR, Lee KR, Hanson AE, Elder JR, Harris WH. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J Bone Joint Surg Am. 1997;79:349–358.

    Article  CAS  PubMed  Google Scholar 

  19. Jolles BM, Genoud P, Hoffmeyer P. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res. 2004;426:174–179.

    Article  PubMed  Google Scholar 

  20. Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163–170.

    Article  PubMed  Google Scholar 

  21. Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–534.

    Article  CAS  PubMed  Google Scholar 

  22. Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29:465–471.

    Article  PubMed  Google Scholar 

  23. Kummer FJ, Shah S, Lyer S, DiCesare PE. The effect of acetabular cup orientations on limiting hip rotation. J Arthroplasty. 1999;14:509–513.

    Article  CAS  PubMed  Google Scholar 

  24. Leichtle U, Gosselke N, Wirth CJ, Rudert M. Radiologic evaluation of cup placement variation in conventional total hip arthroplasty [in German]. Rofo. 2007;179:46–52.

    CAS  PubMed  Google Scholar 

  25. Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res. 2009;467:2259–2265.

    Article  PubMed  Google Scholar 

  26. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.

    CAS  PubMed  Google Scholar 

  27. Lubbeke A, Moons KG, Garavaglia G, Hoffmeyer P. Outcomes of obese and nonobese patients undergoing revision total hip arthroplasty. Arthritis Rheum. 2008;59:738–745.

    Article  PubMed  Google Scholar 

  28. Mallory TH, Lombardi AV Jr, Fada RA, Herrington SM, Eberle RW. Dislocation after total hip arthroplasty using the anterolateral abductor split approach. Clin Orthop Relat Res. 1999;358:166–172.

    Article  PubMed  Google Scholar 

  29. Malviya A, Holland JP. Pseudotumours associated with metal-on-metal hip resurfacing: 10-year Newcastle experience. Acta Orthop Belgica. 2009;75:477–483.

    Google Scholar 

  30. Manley M, Ong K, Lau E, Kurtz SM. Effect of volume on total hip arthroplasty revision rates in the United States Medicare population. J Bone Joint Surg. 2008;90:2446–2451.

    Article  PubMed  Google Scholar 

  31. Marx A, von Knoch M, Pfortner J, Wiese M, Saxler G. Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg. 2006;126:487–492.

    Article  PubMed  Google Scholar 

  32. Morlock MM, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90 Suppl 3:89–95.

    Article  PubMed  Google Scholar 

  33. Morrey BF. Difficult complications after hip joint replacement: dislocation. Clin Orthop Relat Res. 1997;344:179–187.

    Article  PubMed  Google Scholar 

  34. Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.

    CAS  PubMed  Google Scholar 

  35. Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM II, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;16(8 Suppl 1):24–30.

    Google Scholar 

  36. Myers GJ, Morgan D, McBryde CW, O’Dwyer K. Does surgical approach influence component positioning with Birmingham Hip Resurfacing? Int Orthop. 2009;33:59–63.

    Article  CAS  PubMed  Google Scholar 

  37. Najarian BC, Kilgore JE, Markel DC. Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplasty. 2009;24:15–21.

    Article  PubMed  Google Scholar 

  38. Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci. 2001;12:141–144.

    CAS  Google Scholar 

  39. Newington DP, Bannister GC, Fordyce M. Primary total hip replacement in patients over 80 years of age. J Bone Joint Surg Br. 1990;72:450–452.

    CAS  PubMed  Google Scholar 

  40. Nogler M, Mayr E, Krismer M, Thaler M. Reduced variability in cup positioning: the direct anterior surgical approach using navigation. Acta orthop. 2008;79:789–793.

    Article  PubMed  Google Scholar 

  41. Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.

    Article  CAS  PubMed  Google Scholar 

  42. Pedersen DR, Callaghan JJ, Brown TD. Activity-dependence of the “safe zone” for impingement versus dislocation avoidance. Med Eng Phys. 2005;27:323–328.

    Article  CAS  PubMed  Google Scholar 

  43. Penney GP, Edwards PJ, Hipwell JH, Slomczykowski M, Revie I, Hawkes DJ. Postoperative calculation of acetabular cup position using 2-D-3-D registration. IEEE Trans Biomed Eng. 2007;54:1342–1348.

    Article  PubMed  Google Scholar 

  44. Pirard E, De Lint JA. Anteversion of the acetabular component in obese patients. Hip Int. 2007;17:99–103.

    CAS  PubMed  Google Scholar 

  45. Reize P, Geiger EV, Suckel A, Rudert M, Wulker N. Influence of surgical experience on accuracy of acetabular cup positioning in total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2008;37:360–363.

    Google Scholar 

  46. Sadr Azodi O, Adami J, Lindstrom D, Eriksson KO, Wladis A, Bellocco R. High body mass index is associated with increased risk of implant dislocation following primary total hip replacement: 2,106 patients followed for up to 8 years. Acta Orthop. 2008;79:141–147.

    Article  PubMed  Google Scholar 

  47. Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552.

    Article  CAS  PubMed  Google Scholar 

  48. Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, Michaelis U, Kemper G, Grutzner PA, Steffen R, von Knoch M, Holland-Letz T, Bernsmann K. The accuracy of free-hand cup positioning—a CT based measurement of cup placement in 105 total hip arthroplasties. Int Orthop. 2004;28:198–201.

    Article  CAS  PubMed  Google Scholar 

  49. Shervin N, Rubash HE, Katz JN. Orthopaedic procedure volume and patient outcomes: a systematic literature review. Clin Orthop Relat Res. 2007;457:35–41.

    Article  PubMed  Google Scholar 

  50. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty. 2005;20:427–435.

    Article  PubMed  Google Scholar 

  51. Shon WY, Gupta S, Biswal S, Hur CY, Jajodia N, Hong SJ, Myung JS. Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty. Skeletal Radiol. 2008;37:1119–1127.

    Article  PubMed  Google Scholar 

  52. Solomon DH, Losina E, Baron JA, Fossel AH, Guadagnoli E, Lingard EA, Miner A, Phillips CB, Katz JN. Contribution of hospital characteristics to the volume-outcome relationship: dislocation and infection following total hip replacement surgery. Arthritis Rheum. 2002;46:2436–2444.

    Article  PubMed  Google Scholar 

  53. Todkar M. Obesity does not necessarily affect the accuracy of acetabular cup implantation in total hip replacement. Acta Orthop Belg. 2008;74:206–209.

    PubMed  Google Scholar 

  54. Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–821.

    Article  PubMed  Google Scholar 

  55. Wixson RL, MacDonald MA. Total hip arthroplasty through a minimal posterior approach using imageless computer-assisted hip navigation. J Arthroplasty. 2005;20(7 Suppl 3):51–56.

    Article  PubMed  Google Scholar 

  56. Yamaguchi M, Akisue T, Bauer TW, Hashimoto Y. The spatial location of impingement in total hip arthroplasty. J Arthroplasty. 2000;15:305–313.

    Article  CAS  PubMed  Google Scholar 

  57. Zheng G, Steppacher S, Zhang X, Tannast M. Precise estimation of postoperative cup alignment from single standard X-ray radiograph with gonadal shielding. Med Image Comput Comput Assist Interv Int Conf Med Image Comput Comput Assist Interv. 2007;10(Pt 2):951–959.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Henrik Malchau MD, PhD.

Additional information

This project was funded by the Harris Orthopaedic Laboratory.

Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research.

This work was performed at Harris Orthopaedic Laboratory.