Abstract
To evaluate cardiac output (CO), both invasive and semi-invasive monitors are used in critical care medicine. The pulmonary artery catheter is an invasive tool to assess CO with the major criticism that the level of its invasiveness is not supported by an improvement in patients’ outcomes. The interest in a lesser invasive techniques is high. Therefore, alternative techniques have been developed recently, and are used frequently in critical care medicine. Cardiac output can be monitored continuously by different devices that analyze the stroke volume and CO. The purpose of this review is to understand these new technologies and their applications and limitations.
Similar content being viewed by others
References
Fegler G (1954) Measurement of cardiac output in anaesthetized animals by a thermodilution method. Q J Exp Physiol Cogn Med Sci 39:153–164
Hamilton W, Moore J, Kinsman J, Spurling R (1928) Simultaneous determination of the greater and lesser circulation time, of the mean velocity of blood flow through the heart and lungs, of the cardiac output and an approximation of the amount of blood actively circulating in the heart and lungs. Am J Physiol 85:377–378
Bowdle TA (2002) Complications of invasive monitoring. Anesthesiol Clin North Am 20:571–588
Pinsky MR, Vincent JL (2005) Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med 33:1119–1122
Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897
Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F, Boyer A, Brochard L, Teboul JL (2003) Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 290:2713–2720
Calvin JE, Driedger AA, Sibbald WJ (1981) Does the pulmonary capillary wedge pressure predict left ventricular preload in critically ill patients? Crit Care Med 9:437–443
Ivanov R, Allen J, Calvin JE (2000) The incidence of major morbidity in critically ill patients managed with pulmonary artery catheters: a meta-analysis. Crit Care Med 28:615–619
Ehlers KC, Mylrea KC, Waterson CK, Calkins JM (1986) Cardiac output measurements. A review of current techniques and research. Ann Biomed Eng 14:219–239
Gama de Abreu M, Quintel M, Ragaller M et al (1997) Partial carbon dioxide rebreathing: a reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow. Crit Care Med 25:675–683
Kuck K, Ing D, Haryadi DG et al (1998) Evaluation of partial re-breathing cardiac output measurement during surgery. Anesthesiology 89:A542
Side CD, Gosling RJ (1971) Nonsurgical assessment of cardiac function. Nature 232:335–336
Mythen MG, Webb AR (1995) Perioperative plasma volume expansion reduces the incidence of gut mucosal hypoperfusion during cardiac surgery. Arch Surg 130:423–429
Grasso S, Mascia L, Del Turco M et al (2002) Effects of recruiting maneuvers in patients with acute respiratory distress syndrome ventilated with protective ventilatory strategy. Anesthesiology 96:795–802
Gan TJ, Soppitt A, Maroof M et al (2002) Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery. Anesthesiology 97:820–826
Mackay M, Saberi D, Caudwell E et al (2004) Randomised controlled trial assessing the impact of a nurse delivered, flow monitored protocol for optimisation of circulatory status after cardiac surgery. BMJ 329:258
Dark PM, Singer M (2004) The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults. Intensive Care Med 30:2060–2066
Royse CF (2009) Ultrasound-guided haemodynamic state assessment. Best Pract Res Clin Anaesthesiol 23:273–283
Perrino AC Jr, Harris SN, Luther MA (1998) Intraoperative determination of cardiac output using multiplane transesophageal echocardiography: a comparison to thermodilution. Anesthesiology 89:350–357
Darmon PL, Hillel Z, Mogtader A, Mindich B, Thys D (1994) Cardiac output by transesophageal echocardiography using continuous-wave Doppler across the aortic valve. Anesthesiology 80:796–805 (discussion 725A)
Desciak MC, Martin DE (2011) Perioperative pulmonary embolism: diagnosis and anesthetic management. J Clin Anesth 23(2):153–165
Wesseling KH, de Wit B, Weber JAP, Smith NT (1983) A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys 5:16–52
Eisenberg PR, Hansbrough JR, Anderson D, Schuster DP (1987) A prospective study of lung water measurements during patient management in an intensive care unit. Am Rev Respir Dis 136:662–668
Bindels AJ et al (2000) Relationships between volume and pressure measurements and stroke volume in critically ill patients. Crit Care 4(3):193–199
Boussat S et al (2002) Intravascular volume monitoring and extravascular lung water in septic patients with pulmonary edema. Intensive Care Med 28(6):712–718
Eisenberg PR et al (1987) A prospective study of lung water measurements during patient management in an intensive care unit. Am Rev Respir Dis 136(3):662–668
Sakka SG et al (2002) Prognostic value of extravascular lung water in critically ill patients. Chest 122(6):2080–2086
Della Rocca G, Costa MG, Coccia C et al (2003) Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 50:707–711
Halvorsen PS, Espinoza A, Lundblad R et al (2006) Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand 50:1050–1057
Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK (2007) Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 99:329–336
Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P (2002) Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 88:350–356
Goepfert MS, Reuter DA, Akyol D, Lamm P, Kilger E, Goetz AE (2007) Goal-directed fluid management reduces vasopressor and catecholamine use in cardiac surgery patients. Intensive Care Med 33:96–103
Mayer J, Boldt J, Wolf MW et al (2008) Cardiac output derived from arterial pressure waveform analysis in patients undergoing cardiac surgery: validity of a second generation device. Anesth Analg 106:867–872
Scheeren TW, Wiesenack C, Compton FD et al (2008) Performance of a minimally invasive cardiac output monitoring system (Flotrac/Vigileo). Br J Anaesth 101:279–280
Mehta Y, Chand RK, Sawhney R et al (2008) Cardiac output monitoring: comparison of a new arterial pressure waveform analysis to the bolus thermodilution technique in patients undergoing off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 22:394–399
Cannesson M, Attof Y, Rosamel P et al (2007) Comparison of FloTrac™ cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements. Eur J Anaesthesiol 24:832–839
Breukers RM, Sepehrkhouy S, Spiegelenberg SR, Groeneveld AB (2007) Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth 21:632–635
De Waal EE, Kalkman CJ, Rex S, Buhre WF (2007) Validation of a new arterial pulse contour-based cardiac output device. Crit Care Med 35:1904–1909
Opdam HI, Wan L, Bellomo R (2007) A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med 33:344–349
Prasser C, Bele S, Keyl C et al (2007) Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration. BMC Anesthesiol 7:9
Yeo LL et al (2010) Comparison of cardiac output measurement by arterial waveform analysis and pulmonary artery catheter in mitral stenosis. Ann Acad Med Singap 39(8):655–657
Linton RA, Band DM, Haire KM (1993) A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 71:262–266
Costa MG, Della Rocca G, Chiarandini P et al (2008) Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 34:257–263
Cecconi M, Dawson D, Grounds RM, Rhodes A (2009) Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med 35:498–504
Beaulieu KE, Kerr CL, McDonell WN (2005) Evaluation of a lithium dilution cardiac output technique as a method for measurement of cardiac output in anesthetized cats. Am J Vet Res 66:1639–1645
Raaijmakers E, Faes TJ, Scholten RJ et al (1999) A meta-analysis of three decades of validating thoracic cardiography. Crit Care Med 27:1203–1213
Gujjar AR, Muralidhar K, Banakal S et al (2008) Non-invasive cardiac output by transthoracic electrical bioimpedance in post-cardiac surgery patients: comparison with thermodilution method. J Clin Monit Comput 22:175–180
Van De Water JM, Miller TW, Vogel RL et al (2003) Impedance cardiography: the next vital sign technology? Chest 123:2028–2033
Corniea J, Crowley R, Duvvuri R, Ho J, Mahajan A, Salehi A, Sanchez E (2008) Evaluation of an endotracheal continuous cardiac output monitor in patients undergoing cardiac surgery. Anesth Analg 106(SCA Suppl):1–128 SCA60
Kossari N, Hufnagel G, Squara P (2009) Bioreactance: a new tool for cardiac output and thoracic fluid content monitoring during hemodialysis. Hemodial Int 13:512–517
Raval NY, Squara P, Cleman M et al (2008) Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 22:113–119
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Porhomayon, J., El-Solh, A., Papadakos, P. et al. Cardiac output monitoring devices: an analytic review. Intern Emerg Med 7, 163–171 (2012). https://doi.org/10.1007/s11739-011-0738-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11739-011-0738-9