Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings

  • Cardiac
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

Cardiac magnetic resonance imaging and echocardiography are currently regarded as standard modalities for the quantification of left ventricular volumes and ejection fraction. With the recent introduction of dual-source computedtomography (DSCT), the increased temporal resolution of 83 ms should also improve the assessment of cardiac function in CT. The aim of this study was to evaluate the accuracy of DSCT in the assessment of left ventricular functional parameters with cardiac magnetic resonance imaging (MRI) as standard of reference. Fifteen patients (two female, 13 male; mean age 50.8 ± 19.2 years) underwent CT and MRI examinations on a DSCT (Somatom Definition; Siemens Medical Solutions, Forchheim, Germany) and a 3.0-Tesla MR scanner (Magnetom Trio; Siemens Medical Solutions), respectively. Multiphase axial CT images were analysed with a semiautomatic region growing algorithms (Syngo Circulation; Siemens Medical Solutions) by two independent blinded observers. In MRI, dynamic cine loops of short axis slices were evaluated with semiautomatic contour detection software (ARGUS; Siemens Medical Solutions) independently by two readers. End-systolic volume (ESV), end-diastolic volume (EDV), ejection fraction (EF) and stroke volume (SV) were determined for both modalities, and correlation coefficient, systematic error, limits of agreement and inter-observer variability were assessed. In DSCT, EDV and ESV were 135.8 ± 41.9 ml and 54.9 ± 29.6 ml, respectively, compared with 132.1 ± 40.8 ml EDV and 57.6 ± 27.3 ml ESV in MRI. Thus, EDV was overestimated by 3.7 ml (limits of agreement −46.1/+53.6), while ESV was underestimated by 2.6 ml (−36.6/+31.4). Mean EF was 61.6 ± 12.4% in DSCT and 57.9 ± 9.0% in MRI, resulting in an overestimation of EF by 3.8% with limits of agreement at −14.7 and +22.2%. Rank correlation rho values were 0.81 for EDV (P = 0.0024), 0.79 for ESV (P = 0.0031) and 0.64 for EF (P = 0.0168). The kappa value of inter-observer variability were amounted to 0.85 for EDV, ESV and EF. DSCT offers the possibility to quantify left ventricular function from coronary CT angiography datasets with sufficient diagnostic accuracy, adding to the value of the modality in a comprehensive cardiac assessment. The observed differences in the measured values may be due to different post-processing methods and physiological reactions to contrast material injection without beta-blocker medication.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Hundley WG, Hamilton CA, Thomas MS, Herrington DM, Salido TB, Kitzman DW, Little WC, Link KM (1999) Utility of fast cine magnetic resonance imaging and display for the detection of myocardial ischemia in patients not well suited for second harmonic stress echocardiography. Circulation 100:1697–1702

    PubMed  CAS  Google Scholar 

  2. Juergens KU, Fischbach R (2006) Left ventricular function studied with MDCT. Eur Radiol 16:342–357

    Article  PubMed  Google Scholar 

  3. Wintersperger B, Nikolaou K, Dietrich O, Rieber J, Nittka M, Reiser M, Schoenberg S (2003) Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13:1931–1936

    Article  PubMed  Google Scholar 

  4. Rist C, Johnson TR, Becker A, Leber AW, Huber A, Busch S, Becker CR, Reiser MF, Nikolaou K (2007) Dual-source cardiac CT imaging with improved temporal resolution: impact on image quality and analysis of left ventricular function. Radiologe 47:287–294, Apr

    Article  PubMed  Google Scholar 

  5. Tanabe K, Belohlavek M, Jakrapanichakul D et al (1998) Three dimensional echocardiography: precision and accuracy of left ventricular volume measurement using rotational geometry with variable numbers of slice resolution. Echocardiography 15:575–580

    Article  PubMed  Google Scholar 

  6. Johnson T, Nikolaou K, Wintersperger BJ, Leber A, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser M, Becker C (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415

    Article  PubMed  Google Scholar 

  7. Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666

    Article  PubMed  Google Scholar 

  8. Becker CR, Knez A, Leber A, Treede H, Ohnesorge B, Schoepf UJ, Reiser MF (2002) Detection of coronary artery stenoses with multislice helical CT angiography. J Comput Assist Tomogr 26:750–755

    Article  PubMed  Google Scholar 

  9. Juergens KU, Grude M, Maintz D, Fallenberg EM, Wichter T et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experiences. Radiology 230:403–410

    Article  PubMed  Google Scholar 

  10. Mahnken AH, Koos R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kuhl HP, Gunther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720

    Article  PubMed  Google Scholar 

  11. Sugeng L, Mor-Avi V, Weinert L, Niel J, Steringer-Mascherbauer R, Schmidt F, Galuschky C, Schummers G, Lang RM, Nesser HJ (2006) Quantitative assessment of left ventricular size and function: side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference. Circulation 114:654–661

    Article  PubMed  Google Scholar 

  12. Schlosser T, Mohrs OK, Magedanz A, Voigtlander T, Schmermund A, Barkhausen J (2007) Assessment of left ventricular function and mass in patients undergoing computed tomography (CT) coronary angiography using 64-detector-row CT: comparison to magnetic resonance imaging. Acta Radiol 48:30–35

    Article  PubMed  CAS  Google Scholar 

  13. Muhlenbruch G, Das M, Hohl C, Wildberger JE, Rinck D, Flohr TG, Koos R, Knackstedt C, Gunther RW, Mahnken AH (2006) Global left ventricular function in cardiac CT. Evaluation of an automated 3D region-growing segmentation algorithm. Eur Radiol 16:1117–1123

    Article  PubMed  Google Scholar 

  14. Mahnken AH, Muhlenbruch G, Koos R, Stanzel S, Busch PS, Niethammer M, Gunther RW, Wildberger JE (2006) Automated vs. manual assessment of left ventricular function in cardiac multidetector row computed tomography: comparison with magnetic resonance imaging. Eur Radiol 16:1416–1423

    Article  PubMed  Google Scholar 

  15. Vogel-Claussen J, Finn JP, Gomes AS, Hundley GW, Jerosch-Herold M, Pearson G, Sinha S, Lima JA, Bluemke DA (2006) Left ventricular papillary muscle mass: relationship to left ventricular mass and volumes by magnetic resonance imaging. J Comput Assist Tomogr 30:426–432

    Article  PubMed  Google Scholar 

  16. Orakzai SH, Orakzai RH, Nasir K, Budoff MJ (2006) Assessment of cardiac function using multidetector row computed tomography. J Comput Assist Tomogr 30:555–563

    Article  PubMed  Google Scholar 

  17. Lasalarie JC, Serfaty JM, Carre C, Messika-Zeitoun D, Jeannot C, Schouman-Claeys E, Laissy JP (2007) Accuracy of contrast-enhanced cine-MR sequences in the assessment of left ventricular function: comparison with precontrast cine-MR sequences. Results of a bicentric study. Eur Radiol, DOI 10.1007/s00330-007-0647-5

  18. Wintersperger BJ, Sincleair S, Runge VM, Dietrich O, Huber A, Reiser MF, Schoenberg SO (2007) Dual breath-hold magnetic resonance cine evaluation of global and regional cardiac function. Eur Radiol 17:73–80

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to S. Busch.

Additional information

S. Busch and T. Johnson contributed equally to this study.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Busch, S., Johnson, T.R.C., Wintersperger, B.J. et al. Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings. Eur Radiol 18, 570–575 (2008). https://doi.org/10.1007/s00330-007-0767-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-007-0767-y

Keywords

Navigation