Skip to main content
Log in

Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm

  • Cardiac
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

The purpose of this study was to test parallel imaging techniques for improvement of temporal resolution in multislice single breath-hold real-time cine steady-state free precession (SSFP) in comparison with standard segmented single-slice SSFP techniques. Eighteen subjects were examined on a 1.5-T scanner using a multislice real-time cine SSFP technique using the GRAPPA algorithm. Global left ventricular parameters (EDV, ESV, SV, EF) were evaluated and results compared with a standard segmented single-slice SSFP technique. Results for EDV (r=0.93), ESV (r=0.99), SV (r=0.83), and EF (r=0.99) of real-time multislice SSFP imaging showed a high correlation with results of segmented SSFP acquisitions. Systematic differences between both techniques were statistically non-significant. Single breath-hold multislice techniques using GRAPPA allow for improvement of temporal resolution and for accurate assessment of global left ventricular functional parameters.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1a, b.
Fig. 2a, b.
Fig. 3a–d.
Fig. 4a–d.

Similar content being viewed by others

References

  1. Barkhausen J, Goyen M, Ruhm SG, Eggebrecht H, Debatin JF, Ladd ME (2002) Assessment of ventricular function with single breath-hold real-time steady-state free precession cine MR imaging. Am J Roentgenol 178:731–735

    Google Scholar 

  2. Lee VS, Resnick D, Bundy JM, Simonetti OP, Lee P, Weinreb JC (2002) Cardiac function: MR evaluation in one breath-hold with real-time true fast imaging with steady-state precession. Radiology 222:835–842

    PubMed  Google Scholar 

  3. Wintersperger BJ, Nikolaou K, Schoenberg SO, Nittka M, Huber AM, Reiser MF (2002) MRI cardiac function analysis in a single breath-hold: real-time TrueFISP with improved temporal resolution using integrated parallel acquisition techniques (iPAT). Radiology 225:P298

    Google Scholar 

  4. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1202–1210

    Article  PubMed  Google Scholar 

  5. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310

    CAS  PubMed  Google Scholar 

  6. Atkinson DJ, Edelman RR (1991) Cineangiography of the heart in a single breath-hold with a segmented turboFLASH sequence. Radiology 178:357–360

    CAS  PubMed  Google Scholar 

  7. Sakuma H, Fujita N, Foo TK, Caputo GR, Nelson SJ, Hartiala J, Shimakawa A, Higgins CB (1993) Evaluation of left ventricular volume and mass with breath-hold cine MR imaging. Radiology 188:377–380

    CAS  PubMed  Google Scholar 

  8. Lamb HJ, Doornbos J, van der Velde EA, Kruit MC, Reiber JH, de Roos A (1996) Echo planar MRI of the heart on a standard system: validation of measurements of left ventricular function and mass. J Comput Assist Tomogr 20:942–949

    Article  CAS  PubMed  Google Scholar 

  9. Barkhausen J, Ruehm SG, Goyen M, Buck T, Laub G, Debatin JF (2001) MR evaluation of ventricular function: true fast imaging with steady-state precession versus fast low-angle shot cine MR imaging: feasibility study. Radiology 219:264–269

    CAS  PubMed  Google Scholar 

  10. Moon JC, Lorenz CH, Francis JM, Smith GC, Pennell DJ (2002) Breath-hold FLASH and FISP cardiovascular MR imaging: left ventricular volume differences and reproducibility. Radiology 223:789–797

    PubMed  Google Scholar 

  11. Weissler AM, Harris WS, Schoenfeld CD (1968) Systolic time intervals in heart failure in man. Circulation 37:149–159

    CAS  PubMed  Google Scholar 

  12. Setser RM, Fischer SE, Lorenz CH (2000) Quantification of left ventricular function with magnetic resonance images acquired in real time. J Magn Reson Imaging 12:430–438

    Article  CAS  PubMed  Google Scholar 

  13. Miller S, Simonetti OP, Carr J, Kramer U, Finn JP (2002) MR imaging of the heart with cine true fast imaging with steady-state precession: influence of spatial and temporal resolutions on left ventricular functional parameters. Radiology 223:263–269

    PubMed  Google Scholar 

  14. Schalla S, Nagel E, Lehmkuhl H, Klein C, Bornstedt A, Schnackenburg B, Schneider U, Fleck E (2001) Comparison of magnetic resonance real-time imaging of left ventricular function with conventional magnetic resonance imaging and echocardiography. Am J Cardiol 87:95–99

    Google Scholar 

  15. Peters DC, Korosec FR, Grist TM, Block WF, Holden JE, Vigen KK, Mistretta CA (2000) Undersampled projection reconstruction applied to MR angiography. Magn Reson Med 43:91–101

    Google Scholar 

  16. Schaeffter T, Weiss S, Eggers H, Rasche V (2001) Projection reconstruction balanced fast field echo for interactive real-time cardiac imaging. Magn Reson Med 46:1238–1241

    Article  CAS  PubMed  Google Scholar 

  17. Larson AC, Simonetti OP (2001) Real-time cardiac cine imaging with SPIDER: steady-state projection imaging with dynamic echo-train readout. Magn Reson Med 46:1059–1066

    Article  CAS  PubMed  Google Scholar 

  18. Plein S, Smith WH, Ridgway JP, Kassner A, Beacock DJ, Bloomer TN, Sivananthan MU (2001) Qualitative and quantitative analysis of regional left ventricular wall dynamics using real-time magnetic resonance imaging: comparison with conventional breath-hold gradient echo acquisition in volunteers and patients. J Magn Reson Imaging 14:23–30

    Article  CAS  PubMed  Google Scholar 

  19. Nikolaou K, Schoenberg SO, Nittka M, Attenberger U, Behr J, Reiser MF (2003) High resolution magnetic resonance angiography and fast perfusion imaging in the diagnosis of pulmonary arterial hypertension: benefit of parallel imaging techniques. Eur Radiol 13 (Suppl 1):S161

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bernd J. Wintersperger.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wintersperger, B.J., Nikolaou, K., Dietrich, O. et al. Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13, 1931–1936 (2003). https://doi.org/10.1007/s00330-003-1982-9

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-003-1982-9

Keywords

Navigation