Skip to main content
Log in

Optimal Depth of NIM EMG Endotracheal Tube for Intraoperative Neuromonitoring of the Recurrent Laryngeal Nerve During Thyroidectomy

  • Published:
World Journal of Surgery Aims and scope Submit manuscript

Abstract

Background

Malpositioning of the endotracheal surface electrodes can result in dysfunction of intraoperative neuromonitoring (IONM) and increase the risk of recurrent laryngeal nerve injury. The purpose of this study was to investigate the optimal depth of the nerve integrity monitor (NIM) EMG endotracheal tube.

Methods

We enrolled 105 adult patients undergoing elective thyroidectomy. Each Medtronic Xomed NIM EMG endotracheal tube was placed with the middle of the exposed electrodes well in contact with the true vocal cords under direct laryngoscopy. Function of IONM was documented and the insertion depth was measured and analyzed.

Results

Ninety-nine (94.3%) patients had successful IONM with the initial endotracheal tube position. Six (5.7%) patients needed further tube depth adjustment under fiberoptic bronchoscopy. All patients were finally had successful IONM. The optimal mean depth was 20.6 ± 0.97 cm in men and 19.6 ± 1.0 cm in women (p < 0.01). There was the trend that taller subjects had a deeper tube depth (p < 0.05).

Conclusion

We concluded that the mean depth of the NIM EMG tube would be a useful reference value for detecting the malposition of electrodes and adjusting the depth of tube during the operation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Shindo M, Chheda NN (2007) Incidence of vocal cord paralysis with and without recurrent laryngeal nerve monitoring during thyroidectomy. Arch Otolaryngol Head Neck Surg 133:481–485

    Article  PubMed  Google Scholar 

  2. Tomoda C, Hirokawa Y, Uruno T et al (2006) Sensitivity and specificity of intraoperative recurrent laryngeal nerve stimulation test for predicting vocal cord palsy after thyroid surgery. World J Surg 30:1230–1233

    Article  PubMed  Google Scholar 

  3. Chan WF, Lo CY (2006) Pitfalls of intraoperative neuromonitoring for predicting postoperative recurrent laryngeal nerve function during thyroidectomy. World J Surg 30:806–812

    Article  PubMed  Google Scholar 

  4. Witt RL (2005) Recurrent laryngeal nerve electrophysiologic monitoring in thyroid surgery: the standard of care? J Voice 19:497–500

    Article  PubMed  Google Scholar 

  5. Snyder SK, Hendricks JC (2005) Intraoperative neurophysiology testing of the recurrent laryngeal nerve: plaudits and pitfalls. Surgery 138:1183–1191

    Article  PubMed  Google Scholar 

  6. Pearlman RC, Isley MR, Ruben GD et al (2005) Intraoperative monitoring of the recurrent laryngeal nerve using acoustic, free-run, and evoked electromyography. J Clin Neurophysiol 22:148–152

    Article  PubMed  Google Scholar 

  7. Yarbrough DE, Thompson GB, Kasperbauer JL et al (2004) Intraoperative electromyographic monitoring of the recurrent laryngeal nerve in reoperative thyroid and parathyroid surgery. Surgery 136:1107–1115

    Article  PubMed  Google Scholar 

  8. Randolph G.W, Kobler JB, Wilkins J (2004) Recurrent laryngeal nerve identification and assessment during thyroid surgery: laryngeal palpation. World J Surg 28:755–760

    Article  PubMed  Google Scholar 

  9. Dralle H, Sekulla C, Haerting J et al (2004) Risk factors of paralysis and functional outcome after recurrent laryngeal nerve monitoring in thyroid surgery. Surgery 136:1310–1322

    Article  PubMed  Google Scholar 

  10. Marcus B, Edwards B, Yoo S et al (2003) Recurrent laryngeal nerve monitoring in thyroid and parathyroid surgery: the University of Michigan experience. Laryngoscope 113:356–361

    Article  PubMed  Google Scholar 

  11. Tschopp KP, Gottardo C (2002) Comparison of various methods of electromyographic monitoring of the recurrent laryngeal nerve in thyroid surgery. Ann Otol Rhinol Laryngol 111:811–816

    PubMed  Google Scholar 

  12. Otto RA, Cochran CS (2002) Sensitivity and specificity of intraoperative recurrent laryngeal nerve stimulation in predicting postoperative nerve paralysis. Ann Otol Rhinol Laryngol 111:1005–1007

    PubMed  Google Scholar 

  13. Djohan RS, Rodriguez HE, Connolly MM et al (2000) Intraoperative monitoring of recurrent laryngeal nerve function. Am Surg 66:595–597

    PubMed  CAS  Google Scholar 

  14. Eisele DW (1996) Intraoperative electrophysiologic monitoring of the recurrent laryngeal nerve. Laryngoscope 106:443–449

    Article  PubMed  CAS  Google Scholar 

  15. Karlan MS, Catz B, Dunkelman D et al (1984) A safe technique for thyroidectomy with complete nerve dissection and parathyroid preservation. Head Neck Surg 6:1014–1019

    Article  PubMed  CAS  Google Scholar 

  16. Jatzko GR, Lisborg BH, Muller MG et al (1994) Recurrent nerve palsy after thyroid operations—principal nerve identification and a literature review. Surgery 115:139–144

    PubMed  CAS  Google Scholar 

  17. Chiang FY, Wang LF, Huang YF et al (2005) Recurrent laryngeal nerve palsy after thyroidectomy with routine identification of the recurrent laryngeal nerve. Surgery 137:342–347

    Article  PubMed  Google Scholar 

  18. Thomusch O, Sekulla C, Machens A et al (2004) Validity of intra-operative neuromonitoring signals in thyroid surgery. Langenbecks Arch Surg 389:499–503

    Article  PubMed  Google Scholar 

  19. Yap SJ, Morris RW, Pybus DA (1994) Alterations in endotracheal tube position during general anaesthesia. Anaesth Intensive Care 22:586–588

    PubMed  CAS  Google Scholar 

  20. Kim JH, Hong DM, Oh AY et al (2007) Tracheal shortening during laparoscopic gynecologic surgery. Acta Anaesthesiol Scand 51:235–238

    Article  PubMed  CAS  Google Scholar 

  21. Jin-Hee K, Ro YJ, Seong-Won M et al (2005) Elongation of the trachea during neck extension in children: implications of the safety of endotracheal tubes. Anesth Analg 101:974–977

    Article  PubMed  Google Scholar 

  22. Beldi G, Kinsbergen T, Schlumpf R (2004) Evaluation of intraoperative recurrent nerve monitoring in thyroid surgery. World J Surg 28:589–591

    Article  PubMed  Google Scholar 

  23. Cherng CH, Wong CS, Hsu CH et al (2002) Airway length in adults: estimation of the optimal endotracheal tube length for orotracheal intubation. J Clin Anesth 14:271–274

    Article  PubMed  Google Scholar 

  24. Cormack RS, Lehane J (1984) Difficult tracheal intubation in obstetrics. Anaesthesia 39:1105–1111

    Article  PubMed  CAS  Google Scholar 

  25. Koh LK, Kong CE, Ip-Yam PC (2002) The modified Cormack-Lehane score for the grading of direct laryngoscopy: evaluation in the Asian population. Anaesth Intensive Care 30:48–51

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Feng-Yu Chiang.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lu, IC., Chu, KS., Tsai, CJ. et al. Optimal Depth of NIM EMG Endotracheal Tube for Intraoperative Neuromonitoring of the Recurrent Laryngeal Nerve During Thyroidectomy. World J Surg 32, 1935–1939 (2008). https://doi.org/10.1007/s00268-008-9549-1

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00268-008-9549-1

Keywords

Navigation