Abstract
Introduction
Intraoperative neuromonitoring has been widely adopted to facilitate the identification and preservation of recurrent laryngeal nerve (RLN) function during thyroid surgery. The present prospective study validated the ability of this technique to predict postoperative RLN outcomes in a single endocrine surgical unit.
Methods
Neuromonitoring was performed using Neurosign 100 with laryngeal surface electrodes in 171 patients with 271 nerves at risk during thyroidectomy. Vocal cord function was routinely documented perioperatively. Patients were also stratified to low risk (primary surgery for benign disease) and high risk (malignancy and recurrent disease) for subgroup analysis.
Results
Unilateral vocal cord palsy occurred in 15 patients (5.5%) postoperatively. The incidence of postoperative nerve palsy in the low risk and high risk groups was 4.4% and 7.8%, respectively. All but two patients had recovery of function within a median period of 4 months after the operation. The rates of transient and permanent RLN palsy based on nerves at risk were 4.8% (n = 13) and 0.7% (n = 2), respectively. There were 241 true-negative (positive signal and no cord palsy), 15 false-positive (negative signal but no cord palsy), 8 true-positive (negative signal and cord palsy), and 7 false-negative (positive signal but cord palsy) results, as correlated with the postoperative assessment. The sensitivity, specificity, and positive and negative predictive values were 53%, 94%, 35%, and 97%, respectively. For the high risk group, the sensitivity and positive predictive value increased to 86% and 60%, respectively.
Conclusions
There are pitfalls associated with the use of intraoperative neuromonitoring during thyroid surgery. Routine application is not recommended except for selected high risk patients.
Similar content being viewed by others
References
Ready AR, Barnes AD. Complications of thyroidectomy. Br J Surg 1994;81:1555–1556
Lahey FH. Routine dissection and demonstration of recurrent laryngeal nerves in subtotal thyroidectomy. Surg Gynecol Obstet 1938;66:775–777
Jatzko GR, Lisborg PH, Mueller MG, et al. Recurrent nerve palsy after thyroid operations: principal nerve identification and a literature review. Surgery 1994;115:139–144
Hermann M, Alk G, Roka R, et al. Laryngeal recurrent nerve injury in surgery for benign thyroid disease: effect of the nerve dissection and impact of individual surgeon in more than 27,000 nerves at risk. Ann Surg 2002;235:261–268
Shedd DP, Burget GC. Identification of the recurrent laryngeal nerve: electrical method for evaluation in the human. Arch Surg 1966;92:861–864
Flisberg K, Lindholm T. Electrical stimulation of the human laryngeal nerve during thyroid operation. Acta Otolaryngol 1970;263:63–67
Premachandra DJ, Radcliffe GJ, Stearns MP. Intraoperative identification of the recurrent laryngeal nerve and demonstration of its function. Laryngoscope 1990;100:94–96
Eisele DW. Intraoperative electrophysiologic monitoring of the recurrent laryngeal nerve. Laryngoscope 1996;106:443–449
Thomusch O, Sekulla C, Walls G, et al. Intraoperative neuromonitoring of surgery for benign goiter. Am J Surg 2002;183:673–678
Lo CY, Kwok KF, Yuen PW. A prospective evaluation of recurrent laryngeal nerve paralysis during thyroidectomy. Arch Surg 2000;135:204–207
Dralle H, Sekulla C, Haerting J, et al. Risk factors of paralysis and functional outcome after recurrent laryngeal nerve monitoring in thyroid surgery. Surgery 2004;136:1310–1322
Hermann M, Hellebart C, Freissmuth M. Neuromonitoring in thyroid surgery: prospective evaluation of intraoperative electrophysiological responses for the prediction of recurrent laryngeal nerve injury. Ann Surg 2004; 240:9–17
Beldi G, Kinsbergen T, Schlumpf R. Evaluation of intraoperative recurrent nerve monitoring in thyroid surgery. World J Surg 2004;28:589–591
Otto RA, Cochran CS. Sensitivity and specificity of intraoperative recurrent laryngeal nerves stimulation in predicting postoperative nerve paralysis. Ann Otol Rhinol Laryngol 2002;111:1005–1007
Author information
Authors and Affiliations
Corresponding author
Additional information
This work was presented as a free paper during International Surgical Week 2005, Durban, South Africa, August, 2005.
Rights and permissions
About this article
Cite this article
Chan, WF., Lo, CY. Pitfalls of Intraoperative Neuromonitoring for Predicting Postoperative Recurrent Laryngeal Nerve Function during Thyroidectomy. World J. Surg. 30, 806–812 (2006). https://doi.org/10.1007/s00268-005-0355-8
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00268-005-0355-8