A Brief History of Nuclear Medicine Physics, Instrumentation, and Data Sciences in Korea

  • Review
  • Published:
Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

We review the history of nuclear medicine physics, instrumentation, and data sciences in Korea to commemorate the 60th anniversary of the Korean Society of Nuclear Medicine. In the 1970s and 1980s, the development of SPECT, nuclear stethoscope, and bone densitometry systems, as well as kidney and cardiac image analysis technology, marked the beginning of nuclear medicine physics and engineering in Korea. With the introduction of PET and cyclotron in Korea in 1994, nuclear medicine imaging research was further activated. With the support of large-scale government projects, the development of gamma camera, SPECT, and PET systems was carried out. Exploiting the use of PET scanners in conjunction with cyclotrons, extensive studies on myocardial blood flow quantification and brain image analysis were also actively pursued. In 2005, Korea’s first domestic cyclotron succeeded in producing radioactive isotopes, and the cyclotron was provided to six universities and university hospitals, thereby facilitating the nationwide supply of PET radiopharmaceuticals. Since the late 2000s, research on PET/MRI has been actively conducted, and the advanced research results of Korean scientists in the fields of silicon photomultiplier PET and simultaneous PET/MRI have attracted significant attention from the academic community. Currently, Korean researchers are actively involved in endeavors to solve a variety of complex problems in nuclear medicine using artificial intelligence and deep learning technologies.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12
Fig. 13
Fig. 14
Fig. 15
Fig. 16
Fig. 17
Fig. 18
Fig. 19
Fig. 20
Fig. 21
Fig. 22
Fig. 23

Similar content being viewed by others

References

  1. Lee MC, Oh SW, Chung J-K, Lee DS. The current status and future perspectives of nuclear medicine in Korea. Nucl Med Mol Imaging. 2010;44:95–101.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Lee MH, Hahn SS, Lee SH, Lee JK, Kang SS, Koh CS, et al. Investigation and treatment of disease of the thyroid gland with radioactive iodine: the first report. Korean J Med. 1961;4:29–49.

    Google Scholar 

  3. Toh SH. Thyroid and radioactive I-131. J Korean Med Assoc. 1961;4:72–83.

    CAS  Google Scholar 

  4. Lee M-H. The present status of nuclear medicine in Korea. Korean J Nucl Med. 1968;2:9–14.

    Google Scholar 

  5. Cho ZH, Chan JK, Eriksson L. Circular ring transverse axial positron camera for 3-dimensional reconstruction of radionuclides distribution. IEEE Trans Nucl Sci. 1976;23:613–22.

    Article  Google Scholar 

  6. Cho Z, Farukhi M. Bismuth germanate as a potential scintillation detector in positron cameras. J Nucl Med. 1977;18:840–4.

    CAS  PubMed  Google Scholar 

  7. Cho Z, Hong K, Hilal S. Spherical positron emission tomograph (S-PET) I-performance analysis. Nucl Instrum Methods Phys Res. 1984;225:422–38.

    Article  CAS  Google Scholar 

  8. Cho Z, Hong K, Ra JB, Lee S, Hilal S, Correll J. A new sampling scheme for the ring positron camera: Dichotomic ring sampling. IEEE Trans Nucl Sci. 1981;28:94–8.

    Article  Google Scholar 

  9. Cho ZH, Yi W, Jung KJ, Lee BU, Min HB, Song HB. Performance of single photon tomographic system - GAMMATOM-1. IEEE Trans Nucl Sci. 1982;29:484–7.

    Article  Google Scholar 

  10. Ra JB, Lim C, Cho Z, Hilal S, Correll J. A true three-dimensional reconstruction algorithm for the spherical positron emission tomograph. Phys Med Biol. 1982;27:37–50.

    Article  Google Scholar 

  11. Cho ZH, Son YD, Kim HK, Kim KN, Oh SH, Han JY, et al. A fusion PET–MRI system with a high-resolution research tomograph-PET and ultra-high field 7.0 T-MRI for the molecular-genetic imaging of the brain. Proteomics. 2008;8:1302–23.

    Article  CAS  PubMed  Google Scholar 

  12. Cho ZH, Son YD, Kim HK, Kim NB, Choi EJ, Lee SY, et al. Observation of glucose metabolism in the thalamic nuclei by fusion PET/MRI. J Nucl Med. 2011;52:401–4.

    Article  CAS  PubMed  Google Scholar 

  13. Hong IK, Chung ST, Kim HK, Kim YB, Son YD, Cho ZH. Ultra fast symmetry and SIMD-based projection-backprojection (SSP) algorithm for 3-D PET image reconstruction. IEEE Trans Med Imaging. 2007;26:789–803.

    Article  CAS  PubMed  Google Scholar 

  14. Lim C, Gottschalk S, Walker R, Schreiner R, Valentino F, Pinkstaff C, et al. Triangular SPECT system for 3-D total organ volume imaging: design concept and preliminary imaging results. IEEE Trans Nucl Sci. 1985;32:741–7.

    Article  Google Scholar 

  15. Lim C, Walker R, Pinkstaff C, Kim K, Anderson J, King S, et al. Triangular SPECT system for 3-D total organ volume imaging: performance results and dynamic imaging capability. IEEE Trans Nucl Sci. 1986;33:501–4.

    Article  Google Scholar 

  16. Yoo H, Min BG, Lee MC, Kim SK, Lee JS, Koh C-S. A study on the quantitative evaluation of renal function using compartment model. Seoul J Med. 1983;24:437–41.

    Google Scholar 

  17. Min B-G, Kim Y-H, Koh C-S. A study on left ventricular volume change measurement using nuclear stethoscope. J Biomed Eng Res. 1984;5:167–72.

    Google Scholar 

  18. Bong JG, Kim HJ, So SG, Kim HM, Lee JD, Gwon SI. Development of imaging gamma probe using the position sensitive PMTube. J Biomed Eng Res. 1999;20:107–13.

    Google Scholar 

  19. Yang MG, Kwark C, Shim YG, Kim H-J, Choi Y, Chung J-K, et al. Design and performance evaluation of small size counting and imaging gamma probe system. J Biomed Eng Res. 1997;18:291–9.

    Google Scholar 

  20. Kim J-H, Choi Y, Kim J-Y, Im K-C, Kim S-E, Choi Y-S, et al. Monte Carlo simulations of detection efficiency and position resolution of NaI(TI)-PMT detector used in small gamma camera. Prog Med Phys. 1997;8:67–76.

    Google Scholar 

  21. Kim JH, Choi Y, Joo KS, Sihn BS, Chong JW, Kim SE, et al. Development of a miniature scintillation camera using an NaI(Tl) scintillator and PSPMT for scintimammography. Phys Med Biol. 2000;45:3481–8.

    Article  CAS  PubMed  Google Scholar 

  22. Choi Y, Kim J-H, Kim J-Y, Im K-C, Kim S-E, Choe Y-S, et al. Investigation of the signal characteristics of a small gamma camera system using NaI(Tl)-position sensitive photomultiplier tube. Korean J Nucl Med. 2000;34:82–93.

    Article  Google Scholar 

  23. Kim H-J, So S-G, Bong J-G, Kim H-M, Kim J-H, Ju G-S, et al. A study of digitalizing analog gamma camera using Gamma-PF board. J Biomed Eng Res. 1998;19:351–60.

    Google Scholar 

  24. Bong J-K, Kim H-J, Im J-H, Yang S-O, Moon D-H, Ryu J-S, et al. A study of dopamine transporter imaging and comparison of noninvasive simplified quantitative methods in normal controls and Parkinson’s patients. Korean J Nucl Med. 1996;30:315–24.

    Google Scholar 

  25. Kim H-J, Son H-K, Bong J-K, Nam K-P, Lee H-K. Collimator selection in nuclear medicine imaging using I-123 generated by Te-124 reaction. Korean J Nucl Med. 1996;30:372–8.

    Google Scholar 

  26. Son H-K, Nha S-K, Lee H-K, Kim H-J. The measurement of sensitivity and comparative analysis of simplified quantitation methods to measure dopamine transporters using [I-123] IPT pharmacokinetic computer simulations. Korean J Nucl Med. 1997;31:19–29.

    Google Scholar 

  27. Kim HJ, Im JH, Yang SO, Moon DH, Ryu JS, Bong JK, et al. Imaging and quantitation of dopamine transporters with iodine-123-IPT in normal and Parkinson’s disease subjects. J Nucl Med. 1997;38:1703–11.

    CAS  PubMed  Google Scholar 

  28. Wagner HN Jr. SNM annual meeting: medical problem solving. J Nucl Med. 1996;37:11N-4N, 7N, 26N

  29. Lee D-S, Lee T-H, Kim K-M, Chung J-K, Lee M-C, Koh C-S. Optimization of subtraction brain perfusion SPECT with basal/acetazolamide consecutive acquisition. Korean J Nucl Med. 1997;31:330–8.

    Google Scholar 

  30. Kim K-M, Lee D-S, Kim S-K, Lee J-S, Kang K-W, Yeo J-S, et al. Quantification of cerebrovascular reserve using Tc-99m HMPAO brain SPECT and lassen’s algorithm. Korean J Nucl Med. 2000;34:322–35.

    Article  Google Scholar 

  31. Lee DS, Kim KM, Kim SK, Cheon GJ, Kim YK, Chung JK, et al. Development of a method for measuring myocardial contractility with gated myocardial SPECT and arterial tonometry. J Nucl Cardiol. 1999;6:657–63.

    Article  CAS  PubMed  Google Scholar 

  32. Ahn JY, Lee DS, Lee JS, Kim SK, Cheon GJ, Yeo JS, et al. Quantification of regional myocardial blood flow using dynamic H2(15)O PET and factor analysis. J Nucl Med. 2001;42:782–7.

    CAS  PubMed  Google Scholar 

  33. Lee BI, Lee JS, Lee DS, Choi S. Myocardial blood flow quantification in dynamic PET: An ensemble ICA approach. Lect Notes Comput Sci. 2005;3697:709–14.

    Article  Google Scholar 

  34. Lee JS, Lee DD, Choi S, Park KS, Lee DS. Non-negative matrix factorization of dynamic images in nuclear medicine. IEEE Nucl Sci Symp Conf Rec. 2001;4:2027–30.

    Google Scholar 

  35. Lee JS, Lee DS, Ahn JY, Cheon GJ, Kim SK, Yeo JS, et al. Parametric image of myocardial blood flow generated from dynamic H2(15)O PET using factor analysis and cluster analysis. Med Biol Eng Comput. 2005;43:678–85.

    Article  CAS  PubMed  Google Scholar 

  36. Lee JS, Lee DS, Ahn JY, Cheon GJ, Kim SK, Yeo JS, et al. Blind separation of cardiac components and extraction of input function from H(2)(15)O dynamic myocardial PET using independent component analysis. J Nucl Med. 2001;42:938–43.

    CAS  PubMed  Google Scholar 

  37. Lee JS, Lee DS, Ahn JY, Yeo JS, Cheon GJ, Kim SK, et al. Generation of parametric image of regional myocardial blood flow using H(2)(15)O dynamic PET and a linear least-squares method. J Nucl Med. 2005;46:1687–95.

    PubMed  Google Scholar 

  38. Choi Y, Huang S-C, Hawkins RA, Kim JY, Kim B-T, Hoh CK, et al. Quantification of myocardial blood flow using 13N-ammonia and PET: comparison of tracer models. J Nucl Med. 1999;40:1045–55.

    CAS  PubMed  Google Scholar 

  39. Choi Y, Kim J-Y, Im K-C, Kim J-H, Woo S-K, Lee K-H, et al. Quantification of myocardial blood flow using dynamic N-13 ammonia PET and factor analysis. Korean J Nucl Med. 1999;33:316–26.

    Google Scholar 

  40. Kim J-Y, Lee K-H, Kim S-E, Choe Y-S, Ju H-K, Kim Y-J, et al. A refined method for quantification of myocardial blood flow using N-13 ammonia and dynamic PET. Korean J Nucl Med. 1997;31:73–82.

    Google Scholar 

  41. Kim SJ, Lee JS, Kim YK, Frost J, Wand G, McCaul ME, et al. Multiple linear analysis methods for the quantification of irreversibly binding radiotracers. J Cereb Blood Flow Metab. 2008;28:1965–77.

    Article  CAS  PubMed  Google Scholar 

  42. Seo S, Kim SJ, Yoo HB, Lee JY, Kim YK, Lee DS, et al. Noninvasive bi-graphical analysis for the quantification of slowly reversible radioligand binding. Phys Med Biol. 2016;61:6770–90.

    Article  CAS  PubMed  Google Scholar 

  43. Kim JW, Lee JS, Kim SJ, Hoigebazar L, Shin K-H, Yu K-S, et al. Compartmental modeling and simplified quantification of [11C] sertraline distribution in human brain. Arch Pharmacal Res. 2012;35:1591–7.

    Article  CAS  Google Scholar 

  44. Kim JH, Kim Y-H, Kim YJ, Yang BY, Jeong JM, Youn H, et al. Quantitative positron emission tomography imaging of angiogenesis in rats with forelimb ischemia using 68 Ga-NOTA-c(RGDyK). Angiogenesis. 2013;16:837–46.

    Article  CAS  PubMed  Google Scholar 

  45. Seo S, Kim SJ, Kim YK, Lee J-Y, Jeong JM, Lee DS, et al. Comparative assessment of parametric neuroreceptor mapping approaches based on the simplified reference tissue model using [11C] ABP688 PET. J Cereb Blood Flow Metab. 2015;35:2098–108.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  46. Seo S, Kim SJ, Lee DS, Lee JS. Recent advances in parametric neuroreceptor mapping with dynamic PET: basic concepts and graphical analyses. Neurosci Bull. 2014;30:733–54.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Kim SJ, Lee JS, Im KC, Kim S-Y, Park S-A, Lee SJ, et al. Kinetic modeling of 3′-deoxy-3′-18F-fluorothymidine for quantitative cell proliferation imaging in subcutaneous tumor models in mice. J Nucl Med. 2008;49:2057–66.

    Article  PubMed  Google Scholar 

  48. Lim KS, Kwon JS, Jang IJ, Jeong JM, Lee JS, Kim HW, et al. Modeling of brain D2 receptor occupancy-plasma concentration relationships with a novel antipsychotic, YKP1358, using serial PET scans in healthy volunteers. Clin Pharmacol Ther. 2007;81:252–8.

    Article  CAS  PubMed  Google Scholar 

  49. Lee DS, Lee JS, Kim KM, Chung J-K, Lee MC. Functional brain mapping using H2(15)O positron emission tomography (I): Statistical parametric mapping method. Korean J Nucl Med. 1998;32:225–37.

    Google Scholar 

  50. Lee JS, Lee DS, Lee SK, Nam HW, Kim S-K, Park KS, et al. Functional brain mapping using H2(15)O positron emission tomography (II): Mapping of human working memory. Korean J Nucl Med. 1998;32:238–49.

    Google Scholar 

  51. Kwon JS, Kim JJ, Lee DW, Lee JS, Lee DS, Kim MS, et al. Neural correlates of clinical symptoms and cognitive dysfunctions in obsessive-compulsive disorder. Psychiatry Res. 2003;122:37–47.

    Article  PubMed  Google Scholar 

  52. Lee DS, Lee JS, Oh SH, Kim SK, Kim JW, Chung JK, et al. Cross-modal plasticity and cochlear implants. Nature. 2001;409:149–50.

    Article  CAS  PubMed  Google Scholar 

  53. Lee JS, Kim BN, Kang E, Lee DS, Kim YK, Chung JK, et al. Regional cerebral blood flow in children with attention deficit hyperactivity disorder: comparison before and after methylphenidate treatment. Hum Brain Mapp. 2005;24:157–64.

    Article  PubMed  Google Scholar 

  54. Lee JS, Lee DS, Oh SH, Kim CS, Kim JW, Hwang CH, et al. PET evidence of neuroplasticity in adult auditory cortex of postlingual deafness. J Nucl Med. 2003;44:1435–9.

    PubMed  Google Scholar 

  55. Lee JS, Lee DS, Kim J, Kim YK, Kang E, Kang H, et al. Development of Korean standard brain templates. J Korean Med Sci. 2005;20:483–8.

    Article  PubMed  PubMed Central  Google Scholar 

  56. Lee JS, Lee DS, Kim YK, Kim J, Lee HY, Lee SK, et al. Probabilistic map of blood flow distribution in the brain from the internal carotid artery. Neuroimage. 2004;23:1422–31.

    Article  PubMed  Google Scholar 

  57. Kim SJ, Kim IJ, Kim YK, Lee TH, Lee JS, Jun S, et al. Probabilistic anatomic mapping of cerebral blood flow distribution of the middle cerebral artery. J Nucl Med. 2008;49:39–43.

    Article  PubMed  Google Scholar 

  58. Lee JS, Lee DS. Analysis of functional brain images using population-based probabilistic atlas. Curr Med Imaging Rev. 2005;1:81–7.

    Article  Google Scholar 

  59. Koo B, Lee J, Kim J, Lee J, Kim I, Kim J, et al. Developing a Korean standard brain atlas on the basis of statistical and probabilistic approach and visualization tool for functional image analysis. Korean J Nucl Med. 2003;37:162–70.

    Google Scholar 

  60. Lee J-S, Lee D-S, Kim Y-K, Kim J-S, Lee J-M, Koo B-B, et al. Quantification of brain images using Korean standard templates and structural and cytoarchitectonic probabilistic maps. Korean J Nucl Med. 2004;38:241–52.

    Google Scholar 

  61. Choi JY, Lee KH, Na DL, Byun HS, Lee SJ, Kim H, et al. Subcortical aphasia after striatocapsular infarction: quantitative analysis of brain perfusion SPECT using statistical parametric mapping and a statistical probabilistic anatomic map. J Nucl Med. 2007;48:194–200.

    PubMed  Google Scholar 

  62. Lee JS, Lee DS, Kim SK, Lee SK, Chung JK, Lee MC, et al. Localization of epileptogenic zones in F-18 FDG brain PET of patients with temporal lobe epilepsy using artificial neural network. IEEE Trans Med Imaging. 2000;19:347–55.

    Article  CAS  PubMed  Google Scholar 

  63. Lee JS, Park KS, Lee DS, Lee CW, Chung JK, Lee MC. Development and applications of a software for Functional Image Registration (FIRE). Comput Methods Programs Biomed. 2005;78:157–64.

    Article  PubMed  Google Scholar 

  64. Park HJ, Kim JJ, Youn T, Lee DS, Lee MC, Kwon JS. Independent component model for cognitive functions of multiple subjects using [15O]H2O PET images. Hum Brain Mapp. 2003;18:284–95.

    Article  PubMed  PubMed Central  Google Scholar 

  65. Park H-J, Kubicki M, Shenton ME, Guimond A, McCarley RW, Maier SE, et al. Spatial normalization of diffusion tensor MRI using multiple channels. Neuroimage. 2003;20:1995–2009.

    Article  PubMed  Google Scholar 

  66. Oh JS, Song IC, Lee JS, Kang H, Park KS, Kang E, et al. Tractography-guided statistics (TGIS) in diffusion tensor imaging for the detection of gender difference of fiber integrity in the midsagittal and parasagittal corpora callosa. Neuroimage. 2007;36:606–16.

    Article  PubMed  Google Scholar 

  67. Park HJ, Lee JD, Chun JW, Seok JH, Yun M, Oh MK, et al. Cortical surface-based analysis of 18F-FDG PET: measured metabolic abnormalities in schizophrenia are affected by cortical structural abnormalities. Neuroimage. 2006;31:1434–44.

    Article  PubMed  Google Scholar 

  68. Oh JS, Oh M, Chung SJ, Kim JS. Cerebellum-specific 18F-FDG PET analysis for the detection of subregional glucose metabolism changes in spinocerebellar ataxia. NeuroReport. 2014;25:1198–202.

    Article  CAS  PubMed  Google Scholar 

  69. Lee JD, Park HJ, Park ES, Kim DG, Rha DW, Kim EY, et al. Assessment of regional GABA(A) receptor binding using 18F-fluoroflumazenil positron emission tomography in spastic type cerebral palsy. Neuroimage. 2007;34:19–25.

    Article  PubMed  Google Scholar 

  70. Son HJ, Oh JS, Roh JH, Seo SW, Oh M, Lee SJ, et al. Differences in gray and white matter (18)F-THK5351 uptake between behavioral-variant frontotemporal dementia and other dementias. Eur J Nucl Med Mol Imaging. 2019;46:357–66.

    Article  CAS  PubMed  Google Scholar 

  71. Kim JS, Lee JS, Park MH, Kang H, Lee JJ, Lee HJ, et al. Assessment of cerebral glucose metabolism in cat deafness model: strategies for improving the voxel-based statistical analysis for animal PET studies. Mol Imaging Biol. 2008;10:154–61.

    Article  PubMed  Google Scholar 

  72. Kim JS, Lee JS, Park MH, Kim KM, Oh SH, Cheon GJ, et al. Feasibility of template-guided attenuation correction in cat brain PET imaging. Mol Imaging Biol. 2010;12:250–8.

    Article  PubMed  Google Scholar 

  73. Woo SK, Lee TS, Kim KM, Kim JY, Jung JH, Kang JH, et al. Anesthesia condition for (18)F-FDG imaging of lung metastasis tumors using small animal PET. Nucl Med Biol. 2008;35:143–50.

    Article  CAS  PubMed  Google Scholar 

  74. Kim JS, Yu AR, Kim KM, Oh SJ, Ryu JS, Kim HJ, et al. Validation of a postinjection transmission method for actual rat brain PET. Med Phys. 2012;39:5614–20.

    Article  CAS  PubMed  Google Scholar 

  75. Chung YH, Choi Y, Cho G, Choe YS, Lee KH, Kim BT. Characterization of dual layer phoswich detector performance for small animal PET using Monte Carlo simulation. Phys Med Biol. 2004;49:2881–90.

    Article  PubMed  Google Scholar 

  76. Jeong MH, Choi Y, Chung YH, Song TY, Jung JH, Hong KJ, et al. Performance improvement of small gamma camera using NaI(Tl) plate and position sensitive photo-multiplier tubes. Phys Med Biol. 2004;49:4961–70.

    Article  PubMed  Google Scholar 

  77. Song TY, Choi Y, Jung JH, Min BJ, Hong KJ, Choe YS, et al. Performance amelioration for small animal SPECT using optimized pinhole collimator and image correction technique. IEEE Trans Nucl Sci. 2005;52:1396–400.

    Article  Google Scholar 

  78. Chung YH, Choi Y, Cho G, Choe YS, Lee K-H, Kim B-T. Optimization of dual layer phoswich detector consisting of LSO and LuYAP for small animal PET. IEEE Trans Nucl Sci. 2005;52:217–21.

    Article  CAS  Google Scholar 

  79. Kim JS, Lee JS, Im KC, Kim SJ, Kim SY, Lee DS, et al. Performance measurement of the microPET focus 120 scanner. J Nucl Med. 2007;48:1527–35.

    Article  PubMed  Google Scholar 

  80. Lee BI, Lee JS, Kim JS, Lee DS, Choi CU, Lim SM. Performance characteristics of microPET R4 scanner for small animal imaging. Korean J Nucl Med. 2005;39:49–56.

    Google Scholar 

  81. Ito M, Lee JS, Kwon SI, Lee GS, Hong B, Lee KS, et al. A four-layer DOI detector with a relative offset for use in an animal PET system. IEEE Trans Nucl Sci. 2010;57:976–81.

    Article  Google Scholar 

  82. Ito M, Lee JS, Park MJ, Sim KS, Hong SJ. Design and simulation of a novel method for determining depth-of-interaction in a PET scintillation crystal array using a single-ended readout by a multi-anode PMT. Phys Med Biol. 2010;55:3827–41.

    Article  PubMed  Google Scholar 

  83. Hong SJ, Kwon SI, Ito M, Lee GS, Sim K-S, Park KS, et al. Concept verification of three-layer DOI detectors for small animal PET. IEEE Trans Nucl Sci. 2008;55:912–7.

    Article  CAS  Google Scholar 

  84. Lee CM, Hong SJ, Yoon HS, Ito M, Kwon S, Park SK, et al. Spatial and energy resolutions of a hexagonal animal PET scanner based on LGSO crystal and flat-panel PMT. Nucl Eng Technol. 2012;44:53–60.

    Article  CAS  Google Scholar 

  85. Yoon HS, Lee CM, Ito M, Kwon SI, Kim SM, Ko GB, et al. Breast-dedicated positron emission tomography systems. J Biomed Eng Res. 2010;31:259–68.

    Google Scholar 

  86. Ito M, Hong SJ, Lee JS. Positron emission tomography (PET) detectors with depth-of-interaction (DOI) capability. Biomed Eng Lett. 2011;1:70–81.

    Article  Google Scholar 

  87. Chung YH, Baek C-H, Lee S-J, Hong KJ, Kang JH, Choi Y. Preliminary experimental results of a quasi-monolithic detector with DOI capability for a small animal PET. Nucl Instrum Meth A. 2010;621:590–4.

    Article  CAS  Google Scholar 

  88. Chung YH, Lee S-J, Baek C-H, Choi Y. New design of a quasi-monolithic detector module with DOI capability for small animal PET. Nucl Instrum Meth A. 2008;593:588–91.

    Article  CAS  Google Scholar 

  89. Ito M, Lee MS, Lee JS. Continuous depth-of-interaction measurement in a single-layer pixelated crystal array using a single-ended readout. Phys Med Biol. 2013;58:1269–82.

    Article  PubMed  Google Scholar 

  90. Lee MS, Kim KY, Ko GB, Lee JS. Prototype pre-clinical PET scanner with depth-of-interaction measurements using single-layer crystal array and single-ended readout. Phys Med Biol. 2017;62:3983–96.

    Article  CAS  PubMed  Google Scholar 

  91. Lee MS, Lee JS. Depth-of-interaction measurement in a single-layer crystal array with a single-ended readout using digital silicon photomultiplier. Phys Med Biol. 2015;60:6495–514.

    Article  PubMed  Google Scholar 

  92. Kim SM, Lee JS, Lee CS, Kim CH, Lee MC, Lee DS, et al. Fully three-dimensional OSEM-based image reconstruction for Compton imaging using optimized ordering schemes. Phys Med Biol. 2010;55:5007–27.

    Article  PubMed  Google Scholar 

  93. Kim SM, Seo H, Park JH, Kim CH, Lee CS, Lee SJ, et al. Resolution recovery reconstruction for a Compton camera. Phys Med Biol. 2013;58:2823–40.

    Article  PubMed  Google Scholar 

  94. An S, Seo H, Lee J, Lee C, Lee J, Kim CH. Effect of detector parameters on the image quality of Compton camera for 99mTc. Nucl Instrum Meth A. 2007;571:251–4.

    Article  CAS  Google Scholar 

  95. Seo H, Kim CH, Park JH, Kim JK, Lee JH, Lee CS, et al. Development of double-scattering-type Compton camera with double-sided silicon strip detectors and NaI(Tl) scintillation detector. Nucl Instrum Meth A. 2010;615:333–9.

    Article  CAS  Google Scholar 

  96. Kang J, Choi Y, Hong KJ, Hu W, Jung JH, Huh Y, et al. A small animal PET based on GAPDs and charge signal transmission approach for hybrid PET-MR imaging. J Inst. 2011;6:P08012.

    Google Scholar 

  97. Kang J, Choi Y, Hong KJ, Jung JH, Hu W, Huh YS, et al. A feasibility study of photosensor charge signal transmission to preamplifier using long cable for development of hybrid PET-MRI. Med Phys. 2010;37:5655–64.

    Article  PubMed  Google Scholar 

  98. Kwon SI, Lee JS, Yoon HS, Ito M, Ko GB, Choi JY, et al. Development of small-animal PET prototype using silicon photomultiplier (SiPM): initial results of phantom and animal imaging studies. J Nucl Med. 2011;52:572–9.

    Article  CAS  PubMed  Google Scholar 

  99. Lee JS, Hong SJ. Geiger-mode avalanche photodiodes for PET/MRI Electronics for Radiation Detection. FL: CRC Press Boca Raton; 2010. p. 179–98.

    Google Scholar 

  100. Lee JS, Ito M, Sim K, Hong B, Lee KS, Muhammad J, et al. Investigation of solid-state photomultipliers for positron emission tomography scanners. J Korean Phys Soc. 2007;50:1332–9.

    Article  Google Scholar 

  101. Hong SJ, Song IC, Ito M, Kwon SI, Lee GS, Sim K-S, et al. An investigation into the use of Geiger-mode solid-state photomultipliers for simultaneous PET and MRI acquisition. IEEE Trans Nucl Sci. 2008;55:882–8.

    Article  Google Scholar 

  102. Jung JH, Choi Y, Hong KJ, Hu W, Kang JH, Min BJ, et al. Development of a position decoder circuit for PET consisting of GAPD arrays. Nucl Instrum Meth A. 2010;621:310–5.

    Article  CAS  Google Scholar 

  103. Kwon SI, Hong SJ, Ito M, Yoon HS, Song IC, Lee JS. Development of SSPM-PET with digitally controlled bias voltage supply. J Nucl Med. 2009;50(supplement 2):355.

    Google Scholar 

  104. Yoon HS, Ko GB, Kwon SI, Lee CM, Ito M, Song IC, et al. Simultaneous PET/MR imaging with a MR-compatible SiPM PET scanner. J Nucl Med. 2011;52(supplement 1):258.

    Google Scholar 

  105. Yoon HS, Ko GB, Kwon SI, Lee CM, Ito M, Song IC, et al. Initial results of simultaneous PET/MRI experiments with an MRI-compatible silicon photomultiplier PET scanner. J Nucl Med. 2012;53:608–14.

    Article  CAS  PubMed  Google Scholar 

  106. Cho G, Kim H, Sul WS, Lee C, Kang B-S. Optimum design of quenching capacitor integrated silicon photomultipliers for TOF-PET application. Phys Procedia. 2012;37:1511–7.

    Article  CAS  Google Scholar 

  107. Sul W-S, Kim H, Cho G. Silicon photomultiplier modules for MRI-compatible PET. J Inst. 2015;10:C04002.

    Google Scholar 

  108. Hong KJ, Choi Y, Jung JH, Kang J, Hu W, Lim HK, et al. A prototype MR insertable brain PET using tileable GAPD arrays. Med Phys. 2013;40:042503

  109. Jung JH, Choi Y, Jung J, Kim S, Lim HK, Im KC, et al. Development of PET/MRI with insertable PET for simultaneous PET and MR imaging of human brain. Med Phys. 2015;42:2354–63.

    Article  PubMed  Google Scholar 

  110. Oh C, Han Y, Kim D, Lim H, Choi Y, Cho G, et al. An optimal RF shielding method for MR-PET fusion system with insertable PET. Int J Imaging Syst Technol. 2014;24:263–9.

    Article  Google Scholar 

  111. Ko GB, Yoon HS, Kim KY, Lee MS, Yang BY, Jeong JM, et al. Simultaneous multiparametric PET/MRI with silicon photomultiplier PET and ultra-high-field MRI for small-animal imaging. J Nucl Med. 2016;57:1309–15.

    Article  CAS  PubMed  Google Scholar 

  112. Ko GB, Kim KY, Yoon HS, Lee MS, Son JW, Im HJ, et al. Evaluation of a silicon photomultiplier PET insert for simultaneous PET and MR imaging. Med Phys. 2016;43:72.

    Article  CAS  PubMed  Google Scholar 

  113. Lee S, Choi Y, Jung JH, Jung J, Kim KB, Choe HJ, et al. Performance evaluation of a small animal PET scanner a high level of multiplexing and charge-signal transmission. Phys Med Biol. 2019;64:045015

  114. Cho ZH, Son YD, Kim HK, Kwon DH, Joo YH, Ra JB, et al. Development of positron emission tomography with wobbling and zooming for high sensitivity and high-resolution molecular imaging. IEEE Trans Med Imaging. 2019;38:2875–82.

    Article  PubMed  Google Scholar 

  115. Son JW, Kim KY, Yoon HS, Won JY, Ko GB, Lee MS, et al. Proof-of-concept prototype time-of-flight PET system based on high-quantum-efficiency multianode PMTs. Med Phys. 2017;44:5314–24.

    Article  CAS  PubMed  Google Scholar 

  116. Won JY, Ko GB, Kim KY, Park H, Lee S, Son JW, et al. Comparator-less PET data acquisition system using single-ended memory interface input receivers of FPGA. Phys Med Biol. 2020;65:155007

  117. Won JY, Park H, Lee S, Son JW, Chung Y, Ko GB, et al. Development and initial results of a brain PET insert for simultaneous 7-Tesla PET/MRI using an FPGA-only signal digitization method. IEEE Trans Med Imaging. 2021;40:1579–90.

    Article  PubMed  Google Scholar 

  118. Choe H-J, Choi Y, Kwak DJ, Lee J. Prototype time-of-flight PET utilizing capacitive multiplexing readout method. Nucl Instrum Meth A. 2019;921:43–9.

    Article  CAS  Google Scholar 

  119. Hong SJ, Kang HG, Ko GB, Song IC, Rhee JT, Lee JS. SiPM-PET with a short optical fiber bundle for simultaneous PET-MR imaging. Phys Med Biol. 2012;57:3869–83.

    Article  PubMed  Google Scholar 

  120. Han YB, Song SH, Kang HG, Lee HY, Hong SJ. SiPM-based gamma detector with a central GRIN lens for a visible/NIRF/gamma multi-modal laparoscope. Opt Express. 2021;29:2364–77.

    Article  CAS  PubMed  Google Scholar 

  121. Kang HG, Lee HY, Kim KM, Song SH, Hong GC, Hong SJ. A feasibility study of an integrated NIR/gamma/visible imaging system for endoscopic sentinel lymph node mapping. Med Phys. 2017;44:227–39.

    Article  CAS  PubMed  Google Scholar 

  122. Kang HG, Song SH, Han YB, Lee HY, Kim KM, Hong SJ. Proof-of-concept of a multimodal laparoscope for simultaneous NIR/gamma/visible imaging using wavelength division multiplexing. Opt Express. 2018;26:8325–39.

    Article  CAS  PubMed  Google Scholar 

  123. Park C, Kim C, Kim J, Lee Y, Na Y, Lee K, et al. Performance comparison between ceramic Ce:GAGG and single crystal Ce:GAGG with digital-SiPM. J Inst. 2017;12:P01002.

    Google Scholar 

  124. Ullah MN, Pratiwi E, Park JH, Yamamoto S, Kamada K, Yoshikawa A, et al. Studies on sub-millimeter LYSO: Ce, Ce: GAGG, and a new Ce: GFAG block detector for PET using digital silicon photomultiplier. Nucl Instrum Meth A. 2018;911:115–22.

    Article  CAS  Google Scholar 

  125. Ullah MN, Park C, Pratiwi E, Kim C, Choi H, Yeom J-Y. A new positron-gamma discriminating phoswich detector based on wavelength discrimination (WLD). Nucl Instrum Meth A. 2019;946:162631

  126. Ullah MN, Pratiwi E, Park JH, Lee K, Choi H, Yeom J-Y. Wavelength discrimination (WLD) TOF-PET detector with DOI information. Phys Med Biol. 2020;65:055003

  127. Bae S, Bae J, Lee H, Lee K. Development of correction methods for variable pinhole single-photon emission computed tomography. J Inst. 2016;11:C02060.

    Google Scholar 

  128. Bae S, Chun J, Cha H, Yeom JY, Lee K, Lee H. Simulation study of a novel target oriented SPECT design using a variable pinhole collimator. Med Phys. 2017;44:470–8.

    Article  CAS  PubMed  Google Scholar 

  129. Cha H, Cho K, Jung Y-J, Bae S, Kim KM, Kim M, et al. Development of organ-specific dual-head single-photon emission computed tomography using variable pinhole collimator. Nucl Instrum Meth A. 2021;987:164822

  130. Oh JS, Kang BC, Roh JL, Kim JS, Cho KJ, Lee SW, et al. Intratumor textural heterogeneity on pretreatment (18)F-FDG PET images predicts response and survival after chemoradiotherapy for hypopharyngeal cancer. Ann Surg Oncol. 2015;22:2746–54.

    Article  PubMed  Google Scholar 

  131. Son HJ, Oh JS, Oh M, Kim SJ, Lee JH, Roh JH, et al. The clinical feasibility of deep learning-based classification of amyloid PET images in visually equivocal cases. Eur J Nucl Med Mol Imaging. 2020;47:332–41.

    Article  CAS  PubMed  Google Scholar 

  132. Kim KY, Son JW, Kim K, Chung Y, Park JY, Lee YS, et al. Performance evaluation of SimPET-X, a PET insert for simultaneous mouse total-body PET/MR imaging. Mol Imaging Biol. 2021

  133. Son JW, Kim KY, Park JY, Kim K, Lee YS, Ko GB, et al. SimPET: a preclinical PET insert for simultaneous PET/MR imaging. Mol Imaging Biol. 2020;22:1208–17.

    Article  PubMed  Google Scholar 

  134. Choi H, Cheon GJ, Kim H-J, Choi SH, Lee JS, Kim Y-I, et al. Segmentation-based MR attenuation correction including bones also affects quantitation in brain studies: an initial result of 18F-FP-CIT PET/MR for patients with parkinsonism. J Nucl Med. 2014;55:1617–22.

    Article  CAS  PubMed  Google Scholar 

  135. Kim JH, Lee JS, Song I-C, Lee DS. Comparison of segmentation-based attenuation correction methods for PET/MRI: evaluation of bone and liver standardized uptake value with oncologic PET/CT data. J Nucl Med. 2012;53:1878–82.

    Article  PubMed  Google Scholar 

  136. Yoo HJ, Lee JS, Lee JM. Integrated whole body MR/PET: where are we? Korean J Radiol. 2015;16:32–49.

    Article  PubMed  PubMed Central  Google Scholar 

  137. An HJ, Seo S, Kang H, Choi H, Cheon GJ, Kim H-J, et al. MRI-based attenuation correction for PET/MRI using multiphase level-set method. J Nucl Med. 2016;57:587–93.

    Article  PubMed  Google Scholar 

  138. Hwang D, Kang SK, Kim KY, Seo S, Paeng JC, Lee DS, et al. Generation of PET attenuation map for whole-body time-of-flight 18F-FDG PET/MRI using a deep neural network trained with simultaneously reconstructed activity and attenuation maps. J Nucl Med. 2019;60:1183–9.

    Article  PubMed  PubMed Central  Google Scholar 

  139. Hwang D, Kim KY, Kang SK, Seo S, Paeng JC, Lee DS, et al. Improving the accuracy of simultaneously reconstructed activity and attenuation maps using deep learning. J Nucl Med. 2018;59:1624–9.

    Article  CAS  PubMed  Google Scholar 

  140. Hwang D, Kang SK, Kim KY, Choi H, Seo S, Lee JS. Data-driven respiratory phase-matched PET attenuation correction without CT. Phys Med Biol. 2021;66:115009

  141. Lee JS. A review of deep-learning-based approaches for attenuation correction in positron emission tomography. IEEE Trans Radiat Plasma Med Sci. 2021;5:160–84.

    Article  Google Scholar 

  142. Kim JH, Lee JS, Kang KW, Lee HY, Han SW, Kim TY, et al. Whole-body distribution and radiation dosimetry of (68)Ga-NOTA-RGD, a positron emission tomography agent for angiogenesis imaging. Cancer Biother Radiopharm. 2012;27:65–71.

    CAS  PubMed  Google Scholar 

  143. Choi JY, Lyoo CH, Kim JS, Kim KM, Lee M, Ryu YH. Biodistribution and radiation dosimetry of [(18)F]Mefway in humans. Mol Imaging Biol. 2016;18:803–6.

    Article  CAS  PubMed  Google Scholar 

  144. Choi JY, Lyoo CH, Lee JH, Cho H, Kim KM, Kim JS, et al. Human radiation dosimetry of [(18)F]AV-1451(T807) to detect tau pathology. Mol Imaging Biol. 2016;18:479–82.

    Article  CAS  PubMed  Google Scholar 

  145. Gupta A, Lee MS, Kim JH, Lee DS, Lee JS. Preclinical voxel-based dosimetry in theranostics: a review. Nucl Med Mol Imaging. 2020;54:86–97.

    Article  PubMed  PubMed Central  Google Scholar 

  146. Gupta A, Lee MS, Kim JH, Park S, Park HS, Kim SE, et al. Preclinical voxel-based dosimetry through GATE Monte Carlo simulation using PET/CT imaging of mice. Phys Med Biol. 2019;64:095007

  147. Gupta A, Shin JH, Lee MS, Park JY, Kim K, Kim JH, et al. Voxel-based dosimetry of iron oxide nanoparticle-conjugated (177)Lu-labeled folic acid using SPECT/CT imaging of mice. Mol Pharm. 2019;16:1498–506.

    Article  CAS  PubMed  Google Scholar 

  148. Lee MS, Kim JH, Paeng JC, Kang KW, Jeong JM, Lee DS, et al. Whole-body voxel-based personalized dosimetry: the multiple voxel S-value approach for heterogeneous media with nonuniform activity distributions. J Nucl Med. 2018;59:1133–9.

    Article  CAS  PubMed  Google Scholar 

  149. Song YS, Kim JH, Lee BC, Jung JH, Park HS, Kim SE. Biodistribution and internal radiation dosimetry of (99m)Tc-IDA-D-[c(RGDfK)](2) (BIK-505), a novel SPECT radiotracer for the imaging of integrin α(v)β(3) expression. Cancer Biother Radiopharm. 2018;33:396–402.

    CAS  PubMed  PubMed Central  Google Scholar 

  150. Lee MS, Hwang D, Kim JH, Lee JS. Deep-dose: a voxel dose estimation method using deep convolutional neural network for personalized internal dosimetry. Sci Rep. 2019;9:10308.

    Article  PubMed  PubMed Central  Google Scholar 

  151. Lee N, Oh I, Chae SY, Jin S, Oh SJ, Lee SJ, et al. Radiation dosimetry of [(18)F]GP1 for imaging activated glycoprotein IIb/IIIa receptors with positron emission tomography in patients with acute thromboembolism. Nucl Med Biol. 2019;72–73:45–8.

    Article  PubMed  Google Scholar 

  152. Lee I, Lim I, Byun BH, Kim BI, Choi CW, Woo SK, et al. A microdose clinical trial to evaluate [(18)F]Florastamin as a positron emission tomography imaging agent in patients with prostate cancer. Eur J Nucl Med Mol Imaging. 2021;48:95–102.

    Article  CAS  PubMed  Google Scholar 

  153. Woo SK, Jang SJ, Seo MJ, Park JH, Kim BS, Kim EJ, et al. Development of (64)Cu-NOTA-trastuzumab for HER2 targeting: a radiopharmaceutical with improved pharmacokinetics for human studies. J Nucl Med. 2019;60:26–33.

    Article  CAS  PubMed  Google Scholar 

  154. Choi H, Lee DS. Generation of structural MR images from amyloid PET: application to MR-less quantification. J Nucl Med. 2018;59:1111–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  155. Kang SK, Seo S, Shin SA, Byun MS, Lee DY, Kim YK, et al. Adaptive template generation for amyloid PET using a deep learning approach. Hum Brain Mapp. 2018;39:3769–78.

    Article  PubMed  PubMed Central  Google Scholar 

  156. Park J, Bae S, Seo S, Park S, Bang JI, Han JH, et al. Measurement of glomerular filtration rate using quantitative SPECT/CT and deep-learning-based kidney segmentation. Sci Rep. 2019;9:4223.

    Article  PubMed  PubMed Central  Google Scholar 

  157. Park J, Lee JS, Oh D, Ryoo HG, Han JH, Lee WW. Quantitative salivary gland SPECT/CT using deep convolutional neural networks. Sci Rep. 2021;11:7842.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  158. Lee S, Jung JH, Kim D, Lim HK, Park MA, Kim G, et al. PET/CT for brain amyloid: a feasibility study for scan time reduction by deep learning. Clin Nucl Med. 2021;46:e133–40.

    Article  PubMed  Google Scholar 

  159. Choi JH, Kim HA, Kim W, Lim I, Lee I, Byun BH, et al. Early prediction of neoadjuvant chemotherapy response for advanced breast cancer using PET/MRI image deep learning. Sci Rep. 2020;10:21149.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  160. Jeong YJ, Park HS, Jeong JE, Yoon HJ, Jeon K, Cho K, et al. Restoration of amyloid PET images obtained with short-time data using a generative adversarial networks framework. Sci Rep. 2021;11:4825.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  161. Yie SY, Kang SK, Hwang D, Lee JS. Self-supervised PET denoising. Nucl Med Mol Imaging. 2020;54:299–304.

    Article  PubMed  Google Scholar 

  162. Hyun SH, Ahn MS, Koh YW, Lee SJ. A machine-learning approach using PET-based radiomics to predict the histological subtypes of lung cancer. Clin Nucl Med. 2019;44:956–60.

    Article  PubMed  Google Scholar 

  163. Ryoo HG, Choi H, Lee DS. Deep learning-based interpretation of basal/acetazolamide brain perfusion SPECT leveraging unstructured reading reports. Eur J Nucl Med Mol Imaging. 2020;47:2186–96.

    Article  CAS  PubMed  Google Scholar 

  164. Kim JY, Oh D, Sung K, Choi H, Paeng JC, Cheon GJ, et al. Visual interpretation of [(18)F]Florbetaben PET supported by deep learning-based estimation of amyloid burden. Eur J Nucl Med Mol Imaging. 2021;48:1116–23.

    Article  CAS  PubMed  Google Scholar 

  165. Choi H, Kim YK, Yoon EJ, Lee JY, Lee DS. Cognitive signature of brain FDG PET based on deep learning: domain transfer from Alzheimer’s disease to Parkinson’s disease. Eur J Nucl Med Mol Imaging. 2020;47:403–12.

    Article  PubMed  Google Scholar 

  166. Kang SK, Choi H, Lee JS. Translating amyloid PET of different radiotracers by a deep generative model for interchangeability. Neuroimage. 2021;232:117890

  167. Choi H, Ha S, Kang H, Lee H, Lee DS. Deep learning only by normal brain PET identify unheralded brain anomalies. EBioMedicine. 2019;43:447–53.

    Article  PubMed  PubMed Central  Google Scholar 

  168. Choi H, Jin KH. Predicting cognitive decline with deep learning of brain metabolism and amyloid imaging. Behav Brain Res. 2018;344:103–9.

    Article  CAS  PubMed  Google Scholar 

  169. Choi H, Ha S, Im HJ, Paek SH, Lee DS. Refining diagnosis of Parkinson’s disease with deep learning-based interpretation of dopamine transporter imaging. NeuroImage Clin. 2017;16:586–94.

    Article  PubMed  PubMed Central  Google Scholar 

  170. Lee S, Kim J, Kang H, Kang D-Y, Park J. Genetic algorithm based deep learning neural network structure and hyperparameter optimization. Appl Sci. 2021;11:744.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The author would like to thank Zang-Hee Cho, Kwang Suk Park, Jong Seo Chai, Yong Hyun Chung, Jong Ho Kim, and Jinhun Joung for their testimony of history and for providing pictures.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jae Sung Lee.

Ethics declarations

Ethics Approval

For this type of study, ethical approval is not required.

Informed Consent

For this type of study, informed consent is not required.

Conflict of Interest

Jae Sung Lee, Kyeong Min Kim, Yong Choi, and Hee-Joung Kim declare that they have no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lee, J.S., Kim, K.M., Choi, Y. et al. A Brief History of Nuclear Medicine Physics, Instrumentation, and Data Sciences in Korea. Nucl Med Mol Imaging 55, 265–284 (2021). https://doi.org/10.1007/s13139-021-00721-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13139-021-00721-7

Keywords

Navigation