Numerical method for axial motion artifact correction in retinal spectral-domain optical coherence tomography

  • Research Article
  • Published:
Frontiers of Optoelectronics Aims and scope Submit manuscript

Abstract

A numerical method that compensates image distortions caused by random fluctuations of the distance to an object in spectral-domain optical coherence tomography (SD OCT) has been proposed and verified experimentally. The proposed method is based on the analysis of the phase shifts between adjacent scans that are caused by micrometer-scale displacements and the subsequent compensation for the displacements through phase-frequency correction in the spectral space. The efficiency of the method is demonstrated in model experiments with harmonic and random movements of a scattering object as well as during in vivo imaging of the retina of the human eye.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Fercher A F, Hitzenberger C K, Kamp G, El-Zaiat S Y. Measurement of intraocular distances by backscattering spectral interferometry. Optics Communications, 1995, 117(1–2): 43–48

    Article  Google Scholar 

  2. Gelikonov V M, Gelikonov G V, Terpelov D A, Shilyagin P A. Electronic interface systems for goals of spectral domain optical coherence tomography. Instruments and Experimental Techniques, 2012, 55(3): 392–398

    Article  Google Scholar 

  3. Rajabi H, Zirak A. Speckle noise reduction and motion artifact correction based on modified statistical parameters estimation in OCT images. Biomedical Physics & Engineering Express, 2016, 2 (3): e035012

    Article  Google Scholar 

  4. Kang W, Wang H, Wang Z, Jenkins M W, Isenberg G A, Chak A, Rollins A M. Motion artifacts associated with in vivo endoscopic OCT images of the esophagus. Optics Express, 2011, 19(21): 20722–20735

    Article  Google Scholar 

  5. de Kinkelder R, Kalkman J, Faber D J, Schraa O, Kok P H B, Verbraak F D, van Leeuwen T G. Heartbeat-induced axial motion artifacts in optical coherence tomography measurements of the retina. Investigative Ophthalmology & Visual Science, 2011, 52(6): 3908–3913

    Article  Google Scholar 

  6. Zawadzki R J, Miller D T. Retinal AO OCT. In: Drexler W, Fujimoto J G, eds. Optical Coherence Tomography: Technology and Applications. 2nd ed. Switzerland: Springer International Publishing, 2015, 1849–1920

    Chapter  Google Scholar 

  7. Gelikonov V M, Gelikonov G V, Shilyagin P A. Optimization of Fizeau-based optical coherence tomography with a reference michelson interferometer. Bulletin of the Russian Academy of Sciences. Physics, 2008, 72(1): 93–97

    Google Scholar 

  8. Kraus M F, Potsaid B, Mayer M A, Bock R, Baumann B, Liu J J, Hornegger J, Fujimoto J G. Motion correction in optical coherence tomography volumes on a per A-scan basis using orthogonal scan patterns. Biomedical Optics Express, 2012, 3(6): 1182–1199

    Article  Google Scholar 

  9. Kraus M F, Liu J J, Schottenhamml J, Chen C L, Budai A, Branchini L, Ko T, Ishikawa H, Wollstein G, Schuman J, Duker J S, Fujimoto J G, Hornegger J. Quantitative 3D-OCT motion correction with tilt and illumination correction, robust similarity measure and regularization. Biomedical Optics Express, 2014, 5(8): 2591–2613

    Article  Google Scholar 

  10. Chen Z, Shen Y, Bao W, Li P, Wang X, Ding Z. Motion correction using overlapped data correlation based on a spatial-spectral encoded parallel optical coherence tomography. Optics Express, 2017, 25(6): 7069–7083

    Article  Google Scholar 

  11. Chen Y, Hong Y J, Makita S, Yasuno Y. Eye-motion-corrected optical coherence tomography angiography using Lissajous scanning. Biomedical Optics Express, 2018, 9(3): 1111–1129

    Article  Google Scholar 

  12. Potsaid B, Gorczynska I, Srinivasan V J, Chen Y, Jiang J, Cable A, Fujimoto J G. Ultrahigh speed spectral/Fourier domain OCT ophthalmic imaging at 70000 to 312500 axial scans per second. Optics Express, 2008, 16(19): 15149–15169

    Article  Google Scholar 

  13. Lezama J, Mukherjee D, McNabb R P, Sapiro G, Kuo A N, Farsiu S. Segmentation guided registration of wide field-of-view retinal optical coherence tomography volumes. Biomedical Optics Express, 2016, 7(12): 4827–4846

    Article  Google Scholar 

  14. Camino A, Zhang M, Dongye C, Pechauer A D, Hwang T S, Bailey S T, Lujan B, Wilson D J, Huang D, Jia Y. Automated registration and enhanced processing of clinical optical coherence tomography angiography. Quantitative Imaging in Medicine and Surgery, 2016, 6(4): 391–401

    Article  Google Scholar 

  15. Baghaie A, Yu Z, D’Souza R M. Involuntary eye motion correction in retinal optical coherence tomography: hardware or software solution? Medical Image Analysis, 2017, 37: 129–145

    Article  Google Scholar 

  16. Camino A, Zhang M, Gao S S, Hwang T S, Sharma U, Wilson D J, Huang D, Jia Y. Evaluation of artifact reduction in optical coherence tomography angiography with real-time tracking and motion correction technology. Biomedical Optics Express, 2016, 7(10): 3905–3915

    Article  Google Scholar 

  17. Lang A, Carass A, Al-Louzi O, Bhargava P, Solomon S D, Calabresi P A, Prince J L. Combined registration and motion correction of longitudinal retinal OCT data. In: Proceedings of SPIE, Volume 9784, Medical Imaging 2016: Image Processing. San Diego: SPIE, 2016, 97840X

    Google Scholar 

  18. Watanabe Y, Takahashi Y, Numazawa H. Graphics processing unit accelerated intensity-based optical coherence tomography angiography using differential frames with real-time motion correction. Journal of Biomedical Optics, 2013, 19(2): 021105

    Article  Google Scholar 

  19. Shemonski N D, Ahn S S, Liu Y Z, South F A, Carney P S, Boppart S A. Three-dimensional motion correction using speckle and phase for in vivo computed optical interferometric tomography. Biomedical Optics Express, 2014, 5(12): 4131–1143

    Article  Google Scholar 

  20. Lee J, Srinivasan V, Radhakrishnan H, Boas D A. Motion correction for phase-resolved dynamic optical coherence tomography imaging of rodent cerebral cortex. Optics Express, 2011, 19(22): 21258–21270

    Article  Google Scholar 

  21. Carrasco-Zevallos O M, Nankivil D, Viehland C, Keller B, Izatt J A. Pupil tracking for real-time motion corrected anterior segment optical coherence tomography. PLoS One, 2016, 11(8): e0162015

    Article  Google Scholar 

  22. Braaf B, Vienola K V, Sheehy C K, Yang Q, Vermeer K A, Tiruveedhula P, Arathorn D W, Roorda A, de Boer J F. Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO. Biomedical Optics Express, 2013, 4(1): 51–65

    Article  Google Scholar 

  23. Montuoro A, Wu J, Waldstein S, Gerendas B, Langs G, Simader C, Schmidt-Erfurth U. Motion artefact correction in retinal optical coherence tomography using local symmetry. In: Proceedings of MICCAI International Conference on Medical Image Computing and Computer-Assisted Intervention. Boston: Springer, 2014, 17, 130–137

    Google Scholar 

  24. Hu Z, Rollins A M. Fourier domain optical coherence tomography with a linear-in-wavenumber spectrometer. Optics Letters, 2007, 32 (24): 3525–3527

    Article  Google Scholar 

  25. Gelikonov V M, Gelikonov G V, Shilyagin P A. Linearwavenumber spectrometer for high-speed spectral-domain optical coherence tomography. Optics and Spectroscopy, 2009, 106(3): 459–465

    Article  Google Scholar 

  26. Shilyagin P A, Ksenofontov S Y, Moiseev A A, Terpelov D A, Matkivsky V A, Kasatkina I V, Mamaev Y A, Gelikonov G V, Gelikonov V M. Equidistant recording of the spectral components in ultra-wideband spectral-domain optical coherence tomography. Radiophysics and Quantum Electronics, 2018, 60(10): 769–778

    Article  Google Scholar 

  27. Terpelov D A, Ksenofontov S Y, Gelikonov G V, Gelikonov V M, Shilyagin P A. A data-acquisition and control system for spectral-domain optical coherence tomography with a speed of 91 912 A-scans/s based on a USB 3.0 interface. Instruments and Experimental Techniques, 2017, 60(6): 868–874

    Article  Google Scholar 

  28. Leitgeb R A, Wojtkowski M. Complex and coherence-noise free Fourier domain optical coherence tomography. In: Drexler W, Fujimoto J G, eds. Optical Coherence Tomography: Technology and applications. 2nd ed. Switzerland: Springer International Publishing, 2015, 195–224

    Chapter  Google Scholar 

  29. Gelikonov V M, Gelikonov G V, Kasatkina I V, Terpelov D A, Shilyagin P A. Coherent noise compensation in spectral-domain optical coherence tomography. Optics and Spectroscopy, 2009, 106 (6): 895–900

    Article  Google Scholar 

  30. Ai J, Wang L V. Synchronous self-elimination of autocorrelation interference in Fourier-domain optical coherence tomography. Optics Letters, 2005, 30(21): 2939–2941

    Article  Google Scholar 

  31. Leitgeb R A, Hitzenberger C K, Fercher A F, Bajraszewski T. Phase-shifting algorithm to achieve high-speed long-depth-range probing by frequency-domain optical coherence tomography. Optics Letters, 2003, 28(22): 2201–2203

    Article  Google Scholar 

  32. Zhang J, Nelson J S, Chen Z. Removal of a mirror image and enhancement of the signal-to-noise ratio in Fourier-domain optical coherence tomography using an electro-optic phase modulator. Optics Letters, 2005, 30(2): 147–149

    Article  Google Scholar 

  33. Matkivsky V A, Moiseev A A, Ksenofontov S Y, Kasatkina I V, Gelikonov G V, Shabanov D V, Shilyagin P A, Gelikonov V M. Medium chromatic dispersion calculation and correction in spectral-domain optical coherence tomography. Frontiers of Optoelectronics, 2017, 10(3): 323–328

    Article  Google Scholar 

  34. Gelikonov G V, Gelikonov V M. Measurement and compensation for the amplitude and phase spectral distortions of an interference signal in optical coherence tomography for the relative optical-spectrum width exceeding 10%. Radiophysics and Quantum Electronics, 2018, 61(2): 135–145

    Article  Google Scholar 

  35. Matveev L A, Zaitsev V Y, Gelikonov G V, Matveyev A L, Moiseev A A, Ksenofontov S Y, Gelikonov V M, Sirotkina M A, Gladkova N D, Demidov V, Vitkin A. Hybrid M-mode-like OCT imaging of three-dimensional microvasculature in vivo using reference-free processing of complex valued B-scans. Optics Letters, 2015, 40(7): 1472–1475

    Article  Google Scholar 

  36. Moiseev A, Ksenofontov S, Sirotkina M, Kiseleva E, Gorozhantseva M, Shakhova N, Matveev L, Zaitsev V, Matveyev A, Zagaynova E, Gelikonov V, Gladkova N, Vitkin A, Gelikonov G. Optical coherence tomography-based angiography device with realtime angiography B-scans visualization and hand-held probe for everyday clinical use. Journal of Biophotonics, 2018, 11(10): e201700292

    Article  Google Scholar 

  37. Huo L, Xi J, Wu Y, Li X. Forward-viewing resonant fiber-optic scanning endoscope of appropriate scanning speed for 3D OCT imaging. Optics Express, 2010, 18(14): 14375–14384

    Article  Google Scholar 

  38. Moon S, Lee S W, Rubinstein M, Wong B J F, Chen Z. Semi-resonant operation of a fiber-cantilever piezotube scanner for stable optical coherence tomography endoscope imaging. Optics Express, 2010, 18(20): 21183–21197

    Article  Google Scholar 

  39. Park H C, Seo Y H, Jeong K H. Lissajous fiber scanning for forward viewing optical endomicroscopy using asymmetric stiffness modulation. Optics Express, 2014, 22(5): 5818–5825

    Article  Google Scholar 

  40. Chen Y, Hong Y J, Makita S, Yasuno Y. Three-dimensional eye motion correction by Lissajous scan optical coherence tomography. Biomedical Optics Express, 2017, 8(3): 1783–1802

    Article  Google Scholar 

  41. Chauhan B C, Stevens K T, Levesque J M, Nuschke A C, Sharpe G P, O’Leary N, Archibald M L, Wang X. Longitudinal in vivo imaging of retinal ganglion cells and retinal thickness changes following optic nerve injury in mice. PLoS One, 2012, 7(6): e40352

    Article  Google Scholar 

  42. Taibbi G, Peterson G C, Syed M F, Vizzeri G. Effect of motion artifacts and scan circle displacements on Cirrus HD-OCT retinal nerve fiber layer thickness measurements. Investigative Ophthalmology & Visual Science, 2014, 55(4): 2251–2258

    Article  Google Scholar 

  43. Bezerra H G, Costa M A, Guagliumi G, Rollins A M, Simon D I. Intracoronary optical coherence tomography: a comprehensive review clinical and research applications. JACC: Cardiovascular Interventions, 2009, 2(11): 1035–1046

    Google Scholar 

  44. Ksenofontov S, Vasilenkova T. Method of optimizing maximum intensity projection technique for rendering scalar three-dimensional data in static mode, in interactive mode and in real time. Patent of Russian Federation RU 2533055, 2014

  45. Ksenofontov S Y. Application of the method of multiple mutual synchronization of parallel computational threads in spectral-domain optical coherent tomography systems. Instruments and Experimental Techniques, 2019, 62(3): 317–323

    Article  Google Scholar 

Download references

Acknowledgements

This work was supported by the State task for IAP RAS (project No. 0035-2019-0013) in part of method development by the Russian scientific foundation (project No. 17-15-01507) in part of model experiments and setup creating, the Russian Federal target program (project 14.610.21.0014 unique No. RFMEFI61017X0014) in part of retinal imaging experiments.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Pavel A. Shilyagin.

Ethics declarations

The authors report no conflict of interest.

Additional information

Sergey Yu. Ksenofontov, Ph.D., Graduated Nizhny Novgorod University in 1993. Ph.D. in Engineering since 2017. Research interests: optical measurement techniques, OCT, computationally efficient algorithms. Current work: computer solutions for OCT.

Pavel A. Shilyagin, Ph.D., Graduated Nizhny Novgorod University in 2005, master of Radiophysics. Ph.D. in Radiophysics since 2009. Research interests: optical measurement techniques, OCT, optical instrumentation, medical physics. Current work: optical solutions for OCT.

Dmitry A. Terpelov, Ph.D., Graduated Nizhny Novgorod Technical University in 2004. Ph.D. in Experimental physics since 2018. Research interests: optical measurement techniques, OCT, computationally efficient algorithms. Current work: electronics for OCT.

Valentin M. Gelikonov, Ph.D., D.Sc. Graduated Nizhny Novgorod University in 1966, engineer with “Radiophysics and electronics” specialization. Scientific degree: 1985-Ph.D. 2007-D.Sc. State Prize of Russian Federation in Science and Technology, 1999. Head of department “Nano-optics and high precision optical measurements” Institute of Applied Physics at the Russian Academy of Sciences (IAP RAS). Research interests: natural fluctuations of gas laser, nonlinear laser spectroscopy, fiber-optical interferometry, high precision optical measurements, bioimaging. Current work: polarization effects in OCT, multimodal OCT.

Grigory V. Gelikonov, Ph.D., D.Sc. Graduated Nizhny Novgorod University in 1992. Ph.D. in Radiophysics since 2007, D.Sc. since 2018. Research interests: development and creation of single-mode fiber elements, laboratory and clinical imaging of animal and human tissues, experimental study of middle-infrared laser ablation of biotissue with OCT monitoring, development of improved OCT technique, including cross-polarization OCT and two-wavelength OCT. Current work: multimodal OCT, clinical OCT approaches.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ksenofontov, S.Y., Shilyagin, P.A., Terpelov, D.A. et al. Numerical method for axial motion artifact correction in retinal spectral-domain optical coherence tomography. Front. Optoelectron. 13, 393–401 (2020). https://doi.org/10.1007/s12200-019-0951-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12200-019-0951-0

Keywords

Navigation