Abstract
To treat bone fractures, implant manufacturers produce 2D anatomically contoured plates. Unfortunately, existing plates only fit a limited segment of the population and/or require manual bending during surgery. Patient-specific implants would provide major benefits such as reducing surgery time and improving treatment outcomes but they are still rare in clinical practice. In this work, we propose a patient-specific design for the long helical 2D PHILOS (Proximal Humeral Internal Locking System) plate, used to treat humerus shaft fractures. Our method automatically creates a custom plate from a CT scan of a patient’s bone. We start by designing an optimal plate on a template bone and, with an anatomy-aware registration method, we transfer this optimal design to any bone. In addition, for an arbitrary bone, our method assesses if a given plate is fit for surgery by automatically positioning it on the bone. We use this process to generate a compact set of plate shapes capable of fitting the bones within a given population. This plate set can be pre-printed in advance and readily available, removing the fabrication time between the fracture occurrence and the surgery. Extensive experiments on ex-vivo arms and 3D-printed bones show that the generated plate shapes (personalized and plate-set) faithfully match the individual bone anatomy and are suitable for clinical practice.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
References
White, T.D., Folkens, P.A.: The human bone manual. Elsevier (2005)
Goyal, K.S., Skalak, A.S., Marcus, R.E., Vallier, H.A., Cooperman, D.R.: Analysis of anatomic periarticular tibial plate fit on normal adults. Clin. Orthop. Relat. Res. 1976–2007(461), 245–257 (2007)
Schmutz, B., Rathnayaka, K., Albrecht, T.: Anatomical fitting of a plate shape directly derived from a 3D statistical bone model of the tibia. J. Clin. Orthopaed. Trauma 10, S236–S241 (2019)
Kozic, N., et al.: Optimisation of orthopaedic implant design using statistical shape space analysis based on level sets. Med. Image Anal. 14(3), 265–275 (2010)
Hwang, J.H., Oh, J.K., Oh, C.W., Yoon, Y.C., Choi, H.W.: Mismatch of anatomically pre-shaped locking plate on Asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study. Arch. Orthop. Trauma Surg. 132(1), 51–56 (2012)
Bou-Sleiman, H., Ritacco, L.E., Nolte, L.-P., Reyes, M.: Minimization of intra-operative shaping of Orthopaedic fixation plates: a population-based design. In: Fichtinger, G., Martel, A., Peters, T. (eds.) MICCAI 2011. LNCS, vol. 6892, pp. 409–416. Springer, Heidelberg (2011). https://doi.org/10.1007/978-3-642-23629-7_50
Schmutz, B., Wullschleger, M.E., Kim, H., Noser, H., Schütz, M.A.: Fit assessment of anatomic plates for the distal medial tibia. J. Orthop. Trauma 22(4), 258–263 (2008)
Schmutz, B., Wullschleger, M.E., Noser, H., Barry, M., Meek, J., Schütz, M.A.: Fit optimisation of a distal medial tibia plate. Comput. Methods Biomech. Biomed. Engin. 14(04), 359–364 (2011)
Harith, H., Schmutz, B., Malekani, J., Schuetz, M.A., Yarlagadda, P.K.: Can we safely deform a plate to fit every bone? Population-based fit assessment and finite element deformation of a distal tibial plate. Med. Eng. Phys. 38(3), 280–285 (2016)
Carrillo, F., Vlachopoulos, L., Schweizer, A., Nagy, L., Snedeker, J., Fürnstahl, P.: A time saver: optimization approach for the fully automatic 3D planning of forearm osteotomies. In: Descoteaux, M., Maier-Hein, L., Franz, A., Jannin, P., Collins, D.L., Duchesne, S. (eds.) MICCAI 2017. LNCS, vol. 10434, pp. 488–496. Springer, Cham (2017). https://doi.org/10.1007/978-3-319-66185-8_55
Wu, X., et al.: Preliminary exploration of a quantitative assessment index for the matching performance of anatomical bone plates using computer. J. Orthop. Surg. Res. 14(1), 1–8 (2019)
Tkany, L., Hofstätter, B., Petersik, A., Miehling, J., Wartzack, S., Sesselmann, S.: New design process for anatomically enhanced osteosynthesis plates. J. Orthop. Res. 37(7), 1508–1517 (2019)
Zenker, M., et al.: Quantifying osteosynthesis plate prominence-mathematical definitions and case study on a clavicle plate. Comput. Methods Biomech. Biomed. Eng. 25, 1–10 (2022)
Schulz, A.P., et al.: Evidence based development of a novel lateral fibula plate (variAx fibula) using a real CT bone data based optimization process during device development. Open Orthop. J. 6, 1–7 (2012)
Gill, D.R., Torchia, M.E.: The spiral compression plate for proximal humeral shaft nonunion: a case report and description of a new technique. J. Orthop. Trauma 13(2), 141–144 (1999)
Da Silva, T., Rummel, F., Knop, C., Merkle, T.: Comparing iatrogenic radial nerve lesions in humeral shaft fractures treated with helical or straight philos plates: a 10-year retrospective cohort study of 62 cases. Arch. Orthop. Trauma Surg. 140(12), 1931–1937 (2020)
Petersik, A., et al.: A numeric approach for anatomic plate design. Injury 49, S96–S101 (2018)
Meng, D., Keller, M., Boyer, E., Black, M., Pujades, S.: Learning a statistical full spine model from partial observations. In: Reuter, M., Wachinger, C., Lombaert, H., Paniagua, B., Goksel, O., Rekik, I. (eds.) ShapeMI 2020. LNCS, vol. 12474, pp. 122–133. Springer, Cham (2020). https://doi.org/10.1007/978-3-030-61056-2_10
Eisenberger, M., Lahner, Z., Cremers, D.: Smooth shells: multi-scale shape registration with functional maps. In: Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition, pp. 12265–12274 (2020)
Ahmad, M., Nanda, R., Bajwa, A., Candal-Couto, J., Green, S., Hui, A.: Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? Injury 38(3), 358–364 (2007)
Acknowledgements
This work was supported by the project CAMed (COMET K- Project 871132) which is funded by the Austrian Federal Ministry of Transport, Innovation and Technology (BMVIT) and the Austrian Federal Ministry for Digital and Economic Affairs (BMDW) and the Styrian Business Promotion Agency (SFG). Michael J. Black (MJB) has received research gift funds from Adobe, Intel, Nvidia, Meta/Facebook, and Amazon. MJB has financial interests in Amazon, Datagen Technologies, and Meshcapade GmbH. MJB’s research was performed solely at MPI. We thank Karoline Seibert at Hofer GmbH & Co KG, Fürstenfeld, Austria for the 3D printing of bones and plates.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
1 Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Copyright information
© 2023 The Author(s), under exclusive license to Springer Nature Switzerland AG
About this paper
Cite this paper
Keller, M. et al. (2023). Optimizing the 3D Plate Shape for Proximal Humerus Fractures. In: Greenspan, H., et al. Medical Image Computing and Computer Assisted Intervention – MICCAI 2023. MICCAI 2023. Lecture Notes in Computer Science, vol 14226. Springer, Cham. https://doi.org/10.1007/978-3-031-43990-2_46
Download citation
DOI: https://doi.org/10.1007/978-3-031-43990-2_46
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-031-43989-6
Online ISBN: 978-3-031-43990-2
eBook Packages: Computer ScienceComputer Science (R0)