Introduction

Electroencephalography (EEG) is an almost century-old, non-invasive method deployed for electrophysiological brain investigations. In the past decade, its use has significantly increased within intensive care units (ICU) in view of improving prognostic performance after irreversible cerebral damage, secondary to a cardiac arrest [1, 2]. Recently, EEG has proven to be efficient for prognostication among septic ICU patients or those under VA-ECMO as a result of cardiogenic shock and/or refractory cardiac arrest [3, 4].

Patients under VA-ECMO experience a mortality rate close to 60% and are at risk of numerous neurological complications with dramatic consequences on the functional prognosis [5]. A recent study highlighted the role of standard 21-electrode EEG (stdEEG) background abnormalities (slowing, discontinuity, and absence of reactivity) in such a population for early prognostication of 28-day mortality and 90-day functional outcomes [4]. Moreover, the recent use of the Synek score in other critical conditions, such as cardiac arrest [6, 7], sepsis-associated encephalopathy [3] and traumatic brain injury [8] suggests that the use of such EEG scores could be of interest in adult VA-ECMO patients.

Nevertheless, in practice, installing 21 electrodes is time-consuming and requires constant adjustments to reduce artifacts. In addition, this method requires 24/7 availability of qualified medical personnel able to install the scalp electrodes and of trained neurologists for interpretation. In this context, simplifying the EEG montage while maintaining good clinical performance at establishing prognosis would represent a considerable advantage, as it may significantly reduce the time to interpretation in clinical practice. However, whether a simplified EEG montage is associated with a more straightforward interpretation at the bedside remains unknown, as EEG interpretation (by a neurologist or an intensivist) in itself requires specific training.

A recent study confirmed that a simplified 6-electrode montage (F3, T3, P3, F4, T4 and P4) can detect major background abnormalities after cardiac arrest [9]. A simplified montage is already used to guide the depth of sedation in an ICU [10] and some studies suggest that a frontal montage could be as effective as 21 electrodes for the detection of EEG patterns associated with a poor prognosis in post-anoxic comatose patients [11]. However, the use of a simplified montage for detection of background abnormalities in adults under VA-ECMO patients has never been investigated.

We aimed to evaluate whether a simplified 4-frontal electrode montage (4-frontEEG) could detect EEG patterns associated with poor outcomes in adult patients under VA-ECMO, as compared to interpretation made on stdEEG recordings. Moreover, we aimed to explore the performance of Synek scores calculated on this 4-frontEEG montage to predict outcome.

Methods

Patients

We conducted a reanalysis of EEG data from a prospective cohort of 118 adult patients treated with VA-ECMO, in whom EEG was performed on admission to intensive care. The study was conducted in two intensive care units of the Bichat-Claude Bernard University Hospital, Paris, France, between November 2013 and November 2017 [4].

The local ethics committee (IRB 00006477, study number 14-050) approved this research. Patients meeting the inclusion criteria were cases of refractory cardiogenic shock or refractory cardiac arrest requiring at least 24 h of VA-ECMO and having undergone a 21-electrode EEG within the first 3 days of treatment. We analyzed baseline demographic data, clinically relevant data before VA-ECMO cannulation and on the day of the EEG monitoring, including doses of sedatives. ICU severity scores including Simplified Acute Physiology Score 2 (SAPS2), Sepsis-Related Organ Failure Assessment (SOFA) at ECMO cannulation and survival after veno-arterial ECMO were also recorded.

Study protocol

EEG was recorded with system plus evolution, Micromed® (Modigliano Veneto, Italy). EEG recordings consisted of a 30-min standard EEG (stdEEG) using 21 electrodes placed according to the international 10–20 system and interpreted by two neurophysiologists (RW/ART), who were blinded to the patient’s outcome. They were classified as either “reactive” or “non-reactive” as well as “continuous” or “discontinuous” and a frequency range was determined for the background rhythm, as previously described [4]. The simplified montage (4-frontEEG) included only four electrodes from the stdEEG montage: Fp1-Fz and F7-Fz on the left and Fp2-Fz and F8-Fz on the right (Fig. 1). Interpretation of the 4-frontEEG was performed by an intensivist (CT) who was blinded to the existing stdEEG classification and had previously completed a 1-year training in critical care EEG interpretation.

Fig. 1
figure 1

Montage description. Difference between frontal montage (4 electrodes: Fp1, Fp2, F7 and F8, blue electrodes) and standard montage (21 electrodes/system 10–20)

Briefly, training consisted of regular meetings with the neurophysiologists of the Bichat Hospital (one afternoon a week), including interpretation and discussion of ICU EEG tracings. After a year of training, the analysis of the relevant 118 4-frontEEG recordings of the study was carried out. None of the EEGs included in this study had been analyzed and/or discussed during the year of training. The intensivist evaluated on the 4-frontal electrode EEG reactivity, continuity, background rhythm and the Synek score. 4-frontEEG was compared to stdEEG which was considered as the gold standard. In addition, we used the Synek classification to score 4-frontEEG and stdEEG recordings between 1 (normal) and 5 (fatal) (Additional file 1) [12].

EEG terminology

The American Clinical Neurophysiology Society’s (ACNS) standardized critical care EEG terminology was used for the study [13]. A discontinuous background activity was defined when > 10% of the record consisted of periods of lower voltage, i.e., attenuation (> 10 μV) or suppression (< 10 μV). For each patient, EEG reactivity was tested at bedside according to a standardized protocol, in line with a recent international consensus [14].

The Synek score was determined on stdEEG and 4-frontEEG by expert neurophysiologists and by an intensivist, respectively, as described in the Additional file 1 [12].

Outcome variables

stdEEG, 4-frontEEG and clinical variables associated with unfavorable outcome were investigated. 28-day mortality following the start of VA-ECMO and 90-day poor functional outcome (i.e., severe disability or death at 90 days, defined by a score of 4–6 on the modified Rankin scale) were analyzed. Outcomes at 90 days were assessed during a telephone interview by a physician blinded to the EEG results [4].

Statistical analysis

The main outcome measure was the degree of correlation between 4-frontEEG and stdEEG montages to identify EEG patterns of interest. Values were expressed as percentages for qualitative variables and medians [inter quartile range: (IQR)] for quantitative variables. We used non-parametric Kruskal–Wallis and Mann–Whitney tests for quantitative data, and the Fisher’s test for qualitative variables. To compare interpretation of the two EEG montages, we constructed a 2 × 2 confusion matrix for reactivity and continuity, and computed Cohen’s kappa coefficients. We used dummy variables with 1 for reactivity and continuity and 0 otherwise. We then reported Fisher’s score, sensitivity and specificity. We used Spearman rank-based correlation to assess correlation between the Synek score computed from 4-front and stdEEG. The significance level used in this study was α = 0.05.

The performance of the Synek score, calculated on 4-frontEEG and stdEEG montages, to predict outcomes (i.e., 28-day mortality and poor 90-day functional outcome) was investigated in a secondary exploratory analysis. To evaluate propensity to predict the 28-day mortality outcome, we drew receiver operating characteristic (ROC) curves obtained from univariate logistic regressions. For each univariate regression, we provided the odds ratios (OR) with its 95% confidence interval (CI) and the areas under curves (AUC) for the ROC curves. The association between each EEG variable (background rhythm, and Synek scores on 4-frontEEG and stdEEG) and outcomes (day-28 mortality and day-90 mRS score ≥ 4) was evaluated using logistic regression analysis after adjustment for age, non-neurological SOFA and pre-ECMO cardiac arrest, as previously described [4].

Results

Patients

One hundred eighteen patients [58 (48; 66) years old, 32% female] were included (Table 1). The median SOFA on admission was 12 [10; 14.75]. Thirty-five (30%) had suffered cardiac arrest before ECMO cannulation. At the time of the EEG, 106 patients were sedated (90%) and none were in therapeutic hypothermia. Sixty-two (53%) had died at 28 days (D28). The median Rankin score of the 56 patients alive at 90 days was 3 [2; 5]. Of the 21-electrode EEGs, 8% (9 patients) were unreactive, 25% were discontinuous (30 patients), the median background rhythm was 6 Hz [6, 7] and the median Synek score was 2 [2, 5]. It should be noted that no electrical seizures were detected on the 21-electrode EEGs (stdEEG) by the experts.

Table 1 Patient characteristics and EEG findings with respect to 28-day survival

Pattern concordance between stdEEG and 4-frontEEG

Among the 118 patients, we identified 9 (7.6%) unreactive EEG patients with stdEEG versus 12 (10.2%) with the 4-frontEEG. The accuracy of the frontal monitoring to detect reactivity compared to the standard montage was statistically similar [\(\kappa\) = 0.739 (0.522; 0.957)]. We then compared how the continuity of the EEG signal could be assessed based on the 4-frontEEG montage; both frontal and standard classifications were similar [\(\kappa\) = 0.955 (0.894; 1)]. Concerning the background rhythm, we reported statistically similar results, with 6 Hz [6, 7] and 7 Hz [5, 8] (r = 0.655, p < 0.001, Spearman rank test) for frontal and standard montages, respectively. Finally, we compared the Synek score estimated from 4-frontEEG and stdEEG. Synek scores between the two groups were not significantly different (p non-significant, Wilcoxon signed-rank paired test) and were strongly correlated (r = 0.795, p < 0.001, Spearman rank test) (Fig. 2).

Fig.2
figure 2

Similarity between EEG variables computed from the standard and frontal montages. a Spearman correlation matrix showing strong association between standard (std, red fonts) and frontal (4-front, blue fonts) derived Synek score (ρ = 0.79, p < 0.001). On the same matrix, we also show strong association between std and 4-front estimated background rhythm (ρ = 0.66, p < 0.001). Confusion tables between std and 4-front montages computed for the absence of reactivity (b) and the discontinuity (c)

stdEEG vs 4-frontEEG to predict mortality and Rankin-based neurological outcome

We found that discontinuity, background rhythm and Synek scores obtained from stdEEG and 4-frontEEG were significantly linked to 28-day mortality (Table 1, p < 0.005). In addition, we found that for both stdEEG and 4-frontEEG, the Synek scores were significantly correlated with continuity (r = − 0.783 and r = − 0.762) and background rhythm (r = − 0.776 and r = − 0.782, all p < 0.0001, Spearman rank correlation) (Fig. 2). Therefore, we decided to retain only the Synek score as the EEG marker, which was the feature with the best-balanced performance to analyze the prediction using the ROC curve.

We found that the stdEEG Synek score was associated with 28-day mortality (Fig. 3, AUC = 0.68, accuracy: 64%, p < 0.001, sensitivity = 0.45 and specificity = 0.84) and with a 90-day Rankin score ≥ 4 (AUC = 0.66, accuracy: 66%, p = 0.004, sensitivity = 0.30 and specificity = 0.85). With 4-frontEEG, the Synek score was also associated with 28-day mortality and a 90-day Rankin ≥ 4 pts (AUC = 0.71, balanced accuracy 66%, p < 0.001, sensitivity = 0.57, specificity = 0.77 and AUC = 0.71, 68%, p < 0.001, sensitivity = 0.40, specificity = 0.82, respectively). A comparison of ROC curves revealed no difference between this biomarker when 4 or 21 electrodes are used (Delong test, p = NS).

Fig. 3
figure 3

Performance of standard (stdEEG) and frontal (4-frontEEG) montages for prediction of 28-day mortality and 90-day poor functional outcome. The Synek score, determined by a trained intensivist on 4 frontal electrodes, was associated with 28-day mortality (AUC = 0.71, specificity 0.77 and sensitivity 0.57) and 90-day poor functional outcome (Rankin score ≥ 4) (AUC = 0.71, specificity 0.82 and sensitivity 0.40) with a precision comparable to an expert’s interpretation on 21 electrodes (AUC = 0.68 specificity 0.84 and sensitivity 0.45 and AUC = 0.66, specificity 0.85 and sensitivity 0.30, respectively)

Association between EEG parameters and outcomes (Table 2)

Table 2 Association between EEG parameters and outcomes

Synek scores determined by 4 or 21 electrodes were independently associated with 28-day mortality and poor 90-day functional outcome, with no significant differences in the association estimates (Table 2). Background rhythm expressed as a continuous variable (in Hertz), was significantly, independently associated with mortality at 28 days but not with poor functional outcome, as previously described [4].

Discussion

In this study, we evaluated how poor prognosis among ICU patients under VA-ECMO could be assessed using a 4-electrode frontal EEG. Working with EEG features such as reactivity, discontinuity, background rhythm and the Synek classification, we obtained similar results in prognostication using the simplified (4-frontEEG) as with the classical montage (stdEEG). In particular, we confirmed that the 4-frontEEG showed comparable precision with stdEEG for predicting 28-day mortality, but also 90-day functional outcome based on the Rankin score. In addition, we further simplify prognostication by suggesting that an intensivist with a training in neurophysiology could interpret the simplified 4-frontEEG. At the same time, our secondary exploratory analysis found that after adjustment for age, non-neurological SOFA and pre-ECMO cardiac arrest, the 4-electrode and 21-electrode Synek scores showed similar performances in predicting mortality and poor functional outcome. In summary, a trained intensivist’s interpretation of EEGs based solely on 4 frontal electrodes can, in the context of VA-ECMO patients, provide information about patient survival and neurological prognosis. This strategy is not intended to fully replace a thorough neurological examination, but may enhance patient management, and therefore prognosis, especially as the presence of neurophysiologists or personnel able to set up a classical EEG is often lacking. In our study, experienced EEG technicians placed the electrodes, but one may expect that the use of dedicated neuromonitoring devices with 4 frontal electrodes will improve patients’ management and prognostication.

On a daily basis, the prognostication in ECMO patients remains a complex task requiring clinical examination, EEG, serum biomarker assay and neuroimaging. A multimodal neurological assessment upon ICU admission appears indispensable to assist clinicians with the level of care [15, 16]. Unfortunately, within hospital structures, we are often confronted with a decrease in the number of personnel specialized in EEG and more departments are turning towards private structures at increased costs. We have demonstrated here that prognostication based on EEG can be simplified using a 4-electrode montage that consists of electrodes applied to the forehead rather than the conventional electrodes positioned on the scalp. This technical adjustment led to a simplified interpretation of the EEG, that we have shown can be made by an intensivist who has been trained for only one year. Highly malignant patterns (suppression or burst suppression) have recently predicted poor outcome without false prognostications in comatose patients after cardiac arrest [7, 17]. According to our results, 4 frontal electrodes are sufficient to detect such cortical electrical alterations and could be used for prognostication after cardiac arrest.

Our study has certain limitations that we should point out. Firstly, an important issue concerns the detection of epileptic episodes using the simplified montage. Indeed, in our cohort no seizures were detected using the 21-electrode montage. This blind spot prevented us from evaluating the simplified montage upon such a critical feature. Besides, several studies showed limited usefulness of a reduced montage for the detection of seizures that were previously identified using the full montage [18,19,20]. A 21-electrode set-up should therefore be considered when there is a clinical or anamnestic suspicion of an epileptic seizure in an ICU, and expert interpretation is indispensable in such a case. Furthermore, it should be noted that our study only included patients without focal brain injury. Therefore, our results may not be extrapolatable to patients with stroke or traumatic brain injury. In our study, if an unreactive EEG was perceived by both 4 and 21 electrodes, it was not associated with mortality. The maintenance of this pattern in the ICU prognostication is therefore questionable since other EEG patterns may already contain sufficient information for reliable prognostication [15, 21]. A recent study shows that reactivity testing in itself is not sufficiently reliable for poor outcome prediction. However, for prediction of good outcome, it appears to have added value [14].

We show that 4 frontal electrodes can independently predict mortality, but the method is not optimal for this type of conclusion. Indeed, 90% of the patients were under sedation at the time of the EEG. By their inhibitory properties, sedative drugs disturb the EEG and can slow down the tracing and cause burst suppression pattern. The specificity of the prediction could therefore be increased if the method imposed the absence of sedation at the time of the EEG. In addition, it should be stressed that the sensitivity of these biomarkers on mortality remains poor [4] confirmed by our results. Recent studies underline the importance of the multimodal approach combining multiple prognostication tests, in particular to improve sensitivity [22]. Further studies are therefore necessary to evaluate the differences in prognosis among VA-ECMO patients, or more generally after cardiac arrest, between 21 and 4 electrodes in a multimodal approach (combining clinical examination, blood neuron-specific enolase, and imaging).

Although a prior study showed that ACNS terminology can be used by unexperienced reviewers [23] other studies point out the importance of experience in EEG interpretation in some situations [24]. While certain patterns can be detected by intensivists after assiduous training with experts on a limited number of electrodes, reading a higher resolution trace to detect injury or functional abnormalities undeniably requires a greater expertise. More generally, the exchange of knowledge between specialties is essential to grasp all the issues involved in the management of patients in ICU and for the prognostication challenges often associated. This study is therefore part of an effort to encourage communication and the sharing of tasks between neurophysiologists and intensivists.

Conclusion

In adult patients under VA-ECMO, a simplified 4-frontal electrode EEG montage interpreted by an intensivist, detected common EEG patterns associated with poor outcomes, with a performance similar to that of a standard EEG montage interpreted by expert neurophysiologists. This simplified montage could be implemented as part of a multimodal evaluation for bedside prognostication. These findings deserve validation in multicenter settings.