Introduction

Tracheal intubation is a primary lifesaving procedure for emergency department (ED) or intensive care unit (ICU) patients associated with respiratory dysfunction or decreased airway protection. Under the urgent situation, however, airway management can be challenging due to decompensated cardiopulmonary physiology, inadequate provision of tools and skilled staffs, unfasted state, simultaneous performance of cardiopulmonary resuscitation (CPR) or other medical procedures, difficult access to the patient head, and various anatomic features of difficult airways [1]. It has been reported that failure rate of urgent tracheal intubations in the ED and ICU is significantly higher compared to tracheal intubations performed in the operating room [2]. Furthermore, the results of the Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society shows that at least one in four major adverse airway events in a hospital are likely to occur in the ED or ICU and adverse airway events leading to death or brain damage are approximately 30-fold and 60-fold more frequent in the ICU and ED than in the operating room [1]. Thus, the strategies or alternative methods to reduce the occurrence of adverse events associated with urgent tracheal intubations in the ED and ICU patients are needed.

Traditionally, tracheal intubation is performed using direct laryngoscope (DL), which requires the alignment of the oral, pharyngeal and laryngeal axes to allow direct visualization of the glottis. Thus, tracheal intubation with DL is generally regarded as a difficult skill to acquire and maintain. Video laryngoscope (VL) is a new device that contains a miniaturized camera at the blade tip to indirectly visualize the glottis. Because of proven advantages of a fast learning curve, an improved laryngeal visualization and an increased success rate, VL has been widely used for tracheal intubation in emergency and critical situations [3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22]. There have been several meta-analyses comparing VL with DL for tracheal intubation in the ED and ICU patients, but they have mainly focused on the success rate of tracheal intubation and provided different outcomes [23,24,25,26,27,28,29]. Most important, it is unclear whether the use of VL can reduce the adverse events associated with tracheal intubations in the ED and ICU patients. Thus, this systematic review and meta-analysis was carried out to determine whether VL compared with DL could reduce the adverse events associated tracheal intubation in the ED and ICU patients.

Materials and methods

This systematic review and meta-analysis was conducted following the recommendations of The Cochrane Handbook for Systematic Reviews of Interventions and reported according to the PRISMA statement [30] (www.prisma-statement.org). The protocol was registered on the PROSPERO (http://www.crd. york.ac.uk/PROSPERO, ID: CRD42018100562).

Data sources and search strategy

The current issue of the Cochrane Central Register of Controlled Trials (CENTRAL; 2017, Issue 9), PubMed (1946 to October 30th, 2018), EMBASE (1974 to October 30th, 2018), and Web of Science (1900 to October 30th, 2018) were searched. Study authors were mailed for any useful information. The reference lists of all eligible trials and reviews were screened for additional citations. No language restriction was imposed. The search strategies of the four electronic databases were provided in Table 1.

Table 1 Search strategy for four databases

Eligibility criteria

RCTs, quasi-RCTs, and observational (prospective or retrospective) studies comparing VL and DL and reporting the adverse events of tracheal intubations in the ED and ICU patients were included. Conference abstracts with available data were also included. Studies in which VL or DL was used as a rescue device were excluded. Pre-hospital study, manikin study, cadaver study, simulated study, or case reports were also excluded. Participants were in-hospital non-surgical patients needing urgent tracheal intubations in the ED and ICU. Patients with a suspected laryngeal trauma or an extensive maxillofacial injury requiring an immediate surgical airway, supraglottic airway, or awake fiberoptic intubation were excluded. Patients in the intervention group used a VL. When the C-MAC or McGrath MAC laryngoscope was used for DL by inexperienced operator, the attempt was considered a VL, regardless of whether the operator looked at the monitor. In this case, the supervising attending physician was able to view the video monitor and assist with the tracheal intubation [5]. For patients in the control group, a DL was used. Optimizing maneuvers such as the external laryngeal pressure, the use of intubation stylet or introducer, could be initiated at the discretion of the operator.

Study selection and data extraction

The titles and abstracts were independently screened by two study authors (J.J.; N.K.). After retrieving the full-texts or conference abstract of any potentially relevant studies, their eligibilities were determined. Any disagreements between the two review authors were resolved by discussion with other authors until a consensus was obtained. A PRISMA flow diagram was completed to record the selection process in detail [31].

Data was independently extracted by two review authors (J.J.; N.K.). All the outcomes were dichotomous variables, the number of events occurred and the sample sizes were extracted. Any disagreement on data extraction was resolved by discussion with a third author (F.S.X.) until a consensus was reached.

Primary and secondary outcomes

The esophageal intubation (EI) is rare, but is one of mostly severe adverse events associated with tracheal intubation in the ICU and ED patients [3]. Thus, the rate of EI was used as primary outcome. If the rates of EI at the first-attempt and at any attempt were given, the rate of EI at the first-attempt was only used. The secondary outcomes were incidences of hypoxemia (SpO2 < 90%) and severe hypoxemia (SpO2 < 80%) during the intubation procedure, incidence of aspiration (any witnessed aspiration of gastric contents during the intubation attempt or defined by the original author), incidence of new-onset cardiac arrest (during or immediately after intubation), short-term (within 24 h) and long-term (28 d or in-hospital) all-causes mortality.

Risk of bias assessment

The risk of bias for each eligible study was independently assessed by two review authors (J.J. and N.K.). For RCTs (including quasi-RCT), the Cochrane Collaboration’s tool was used to assess the risk of bias [32]. Each of the seven domains, such as random sequence generation, allocation concealment, blinding of participants and personnels, blinding of outcome assessment, incomplete outcome data, selective reporting, and other biases, was judged as either low, high, or unclear. If all domains were assigned to the “low risk” of bias category, the study was classified as “low risk”; if one or more domains were assigned to the “unclear risk” of bias category, the study was classified as “unclear risk”; if one or more domains were assigned to the high risk of bias category, the study was classified as “high risk” [32]. For observational studies, the ACROBAT-NRSi Cochrane Risk of Bias Tool was used [33]. Each of the seven domains, such as confounding, selection of participants, classification of interventions, deviations from intended interventions, missing data, measurement outcomes, and selection of the reported outcomes, was judged as either low, moderate, severe, or unclear. The possible confounding domains are “experience of operators, difficult airways, number of patients with CPR, and the use of neuromuscular blockades”. No co-interventions were considered. An overall judgement of the risk of bias for each study was reached as low, moderate, serious, critical or no information on the risk of bias. Reporting bias was also assessed by using funnel plot if the result of primary outcome was from at least 10 trials [34].

The criteria of the Grading of Recommendations Assessment, Development and Evaluation (GRADE) system (study limitations, consistency of effect, imprecision, indirectness, and publication bias) were applied to assess the quality of evidence associated with all outcomes [35, 36]. Then a “Grade evidence profile” table was developed by using the GRADE software (www.guidelinedevelopment.org) to rate these outcomes as high, moderate, low, or very low quality. Since different types of studies were involved in this review, the quality of the evidence was assessed for the RCTs (including quasi-RCT) and non-RCTs, respectively. The quality of evidence was downgraded by one or two levels when serious or very serious deficiencies were considered in these criteria.

Trial sequential analysis (TSA) was performed for all adverse events from RCTs [37, 38]. The information size required was calculated to provide an estimate of how many more patients would be required to make a reliable conclusion. A conventional calculation for sample size estimation, with conventional values for α and β error (0.05 and 0.20), low bias-based relative risk reduction, the incidence in control arm, and a model variance-based heterogeneity correction was used.

Statistical analysis

The study authors of the original report were contacted for important missing statistics. For the participants missing due to dropout, if “missing at random”, analysis was performed based on the available data; if not, an available case analysis was performed and the potential bias was explained in discussion section. If a study did not mention withdrawals, no drop-out was assumed.

Both relative risk (RR) and 95% confidence interval (CI) were used for dichotomous data. P < 0.05 was considered statistically significant. Review Manager was applied to perform the pooled analyses for the outcomes from more than one study. A Chi-squared test with the I2 statistic (with statistical significance set at the level of two-tailed 0.10) was used to describe the percentage of the total variance across studies from heterogeneity rather than from chance. If I2 is < 40%, namely there was no statistical heterogeneity among studies, and a fixed-effect model was used; otherwise, a random-effect model was used. For the results that could not be analyzed via meta-analysis, only a qualitative systematic review was planned instead of excluding that study.

Clinical and methodological heterogeneities were considered before performing pooled analysis. In the presence of a statistical heterogeneity or an indication of clinical heterogeneity, subgroup analysis was carried out for all outcomes according to the following possible heterogeneous factors: the type of studies, RCTs or non-RCTs; whether a specialized CPR study; and operators’ expertise: experienced (certificated anesthesiologist, physicians of emergency medical services with no less than three years of clinical experience, physicians performed 50 successful tracheal intubations, or according to the judgment of authors) or inexperienced. If both experienced and inexperienced operators were involved in one study, the sub-grouping of the expertise was determined by the majority part.

Results

Characteristic of included studies

Using search strategy, a total of 1729 papers were identified. Of them, 1661 were excluded during title and abstract screening due to duplicates and being irrelevant to our research question. Sixty-eight studies were selected for full-text assessment using inclusion and exclusion criteria. Forty-five studies were further removed because of no available data (n = 27), overlapping data sets (n = 9), or irrelevant (n = 9). Finally, 23 studies (n = 13,117) including one conference abstract were eventually included in the review for data extraction. Authors from three studies [3, 6, 14] were contacted for detailed information on important data, only one of them replied [3]. The flow chart of included and excluded studies is shown in Fig. 1.

Fig. 1
figure 1

Flow chart of included and excluded studies

The characteristics of included studies are listed in Table 2. Of the 23 included studies, 9 were carried out in the ICU [4, 5, 7, 9, 14, 16, 17, 19, 20], 11 in the ED, one in both the ICU and wards [6], one in general wards [8], and the other in out-of-operating room location [39]; 4 studies enrolled only patients with cardiac arrest [8, 10, 11, 18] and 4 excluded patients with cardiac arrest [16, 19, 21, 22]; 6 studies used a VL with a standard blade (C-MAC) [9, 12, 13, 15, 21, 40], 12 used a VL with an angled blade (GlideScope, McGrath, UEscope) [4, 6, 7, 11, 14, 16, 18,19,20, 22, 39, 41], and 5 combined several types of VLs (Angled: GlideScope and McGrath; Standard: C-MAC; Channeled: AirwayScope, or KingVision) [3, 5, 8, 10, 17]; tracheal intubation was performed mostly by experienced operators in 7 studies [9, 13, 14, 18, 21, 22, 39], the meeting abstract did not mention the expertise of the operators; Rapid sequence induction (RSI) with sedatives and neuromuscular blockades was chosen as needed in most included studies except for the CPR patients or the studies performed in the CPR patients, 2 studies did not mention the anesthesia induction method [39, 41].

Table 2 Characteristics of 23 included studies

Risk of bias assessment for included studies

Detailed description regarding the risk of bias of the included studies is shown in Table 3. Of the 23 included studies, 9 were RCTs [13,14,15,16,17,18,19, 21, 22], one was quasi-RCT [20], and the others were observational studies. The overall risk of bias for most of the included RCTs was rated as low, but the overall risk of bias for the observational studies was rated as moderate or serious mainly due to confounding domains. The funnel plot obtained from primary outcome with its visually symmetrical distribution qualitatively indicated a low risk of publication bias (Additional file 1: Fig. S1). The GRADE system showed that the quality of evidence from most RCTs was high or moderate, whereas the quality of evidence from most non-RCTs was very low. The quality of evidence was downgraded manly due to inconsistency from moderate or high level of heterogeneity and imprecision from few participants and few events. The results regarding the quality of evidence for different adverse events were listed in Additional file 2: Table S1.

Table 3 Risk of bias assessment of 23 included RCTs and non-RCTs

Rate of EI

Eighteen studies reported the rate of EI [3,4,5,6,7, 9,10,11,12, 14, 15, 17,18,19,20,21, 39, 41]; among them, 2 studies reported the rate of EI based on the number of intubations rather than the number of patients [4, 12], one study reported based on the attempts instead of the number of patients [3]. We emailed the original authors for the data based on patients and the initial intubation devices, but the data were not available. After discussion with other authors in our study, we decided to use the current data and do a sensitive analysis by excluding the data from these 3 studies [3, 4, 12].

Pooled analysis showed a significant difference in the rate of EI between VL and DL (RR, 0.24; 95% CI, 0.18–0.32; n = 11,187; P < 0.01; high-quality evidence for RCTs and very low-quality evidence for observational studies). There was no significant heterogeneity among studies (P = 0.05; I2 = 39%) (Additional file 1: Fig. S2). Sensitive analysis by excluding the data from 3 studies mentioned above did not change the pooled result. Subgroup analyses based on the type of studies, whether a CPR study, or operators’ expertise showed a lower rate of EI by using VL compared with DL in all subgroups (P < 0.01) except for experienced operators (5 studies; RR, 0.44; 95% CI, 0.17–1.15; n = 1100; P = 0.09; Fig. 2, Additional file 1: Fig. S3-S4).

Fig. 2
figure 2

Forest plot for comparison of rate of esophageal intubation based on the type of studies between video laryngoscope (VL) and direct laryngoscope (DL). M-H, Mantel-Haenszel

Other adverse events

Results of adverse events including hypoxemia, severe hypoxemia, aspiration, new-onset cardiac arrest, 24 h-mortality, and 28 d-mortality are summarized in Table 4 and Additional file 1: Fig. S5-S19.

Table 4 Results of meta-analysis for all adverse events between direct laryngoscope and video laryngoscope

Three studies reported the incidence of hypoxemia [9, 14, 19]. One study, which reported the rate of decline in oxygen saturation greater than 10% from baseline [5], was also included in the pooled analysis. All of them are non-CPR studies. Eight studies reported the incidence of severe hypoxemia [4, 6, 7, 9, 14, 17, 19, 20], all of them are non-CPR studies. Thirteen studies reported the incidence of aspiration [4, 5, 7, 9, 13, 14, 17, 19,20,21,22, 39, 40]. Seven studies reported the incidence of new-onset cardiac arrest [4, 5, 12, 14, 17, 19, 20]. Six studies reported short-term mortality within 24 h [4, 7, 8, 14, 19, 20], and 7 studies reported long-term mortality (28 d or in-hospital) [4, 8, 13, 16, 17, 19, 22]; among them, one reported the data based on the number of intubations rather than the number of participants [4]. The available data was used because both the number of intubations and that of participants were quite similar. Pooled analyses for all these outcomes showed no significant differences between VL and DL (P > 0.05). For these adverse events, however, there was no significant heterogeneity among studies (I2 < 40%). Subgroup analyses for all these adverse events based on the type of studies, whether a CPR study, or operators’ expertise showed no significant difference between VL and DL in all subgroups (P > 0.05) except for the incidence of hypoxemia when intubated by inexperienced operators (P = 0.03).

The incidences of hypoxemia and severe hypoxemia in both groups were almost 2–3 times higher in the non-RCTs than in the RCTs (21–30% vs. 11–12% for hypoxemia; 10–12% vs. 6–6.7% for severe hypoxemia, P > 0.05). The incidence of aspiration in both groups was higher in the non-RCTs than in the RCTs (3–4.5% vs. 2.5–2.7%, P > 0.05).

Trial sequential analysis

The TSA of a diversity-adjusted required information size for the rate of EI was 668 patients. The cumulative z-curve crossed the boundary of required information size and TSA monitoring boundary for favoring VL. Thus, this pooled analysis from RCTs is conclusive, namely, the use of VL reduces the rate of EI. For incidences of aspiration and new-onset cardiac arrest, and short-term mortality, the cumulative z-curve did not cross the boundary of required information size and TSA monitoring boundary. Thus, the pooled analyses from RCTs for these adverse outcomes are inconclusive. For incidence of hypoxemia and long-term mortality, boundary TSA is ignored due to too little information size. Thus, the pooled analyses from RCTs for these adverse outcomes are also inconclusive (Additional file 1: Fig. S20A-G and Additional file 2: Table S2).

In summary, pooled analysis showed that the use of VL reduced the rate of EI, especially for inexperienced operators. As to the other adverse outcomes, however, no significant differences were identified, except for the incidence of hypoxemia when intubated by inexperienced operators.

Discussion

This systematic review and meta-analysis is focused on adverse events of tracheal intubation using VL compared with DL in the ED and ICU patients by including both RCTs and observational studies. Pooled analysis showed that the use of VL reduced the rate of EI; as to the other adverse outcomes, however, no significant differences were identified.

The rate of EI was chosen as the primary outcome in this analysis, as even a single episode of recognized EI is significantly associated with desaturation, increased risk of aspiration and cardiac arrest. It has been shown that patients with EI have higher incidences of aspiration (6.1 times), dysrhythmia (6.4 times), hypotension (3.1 times), and hypoxemia [3]. In our analysis, however, a significant lower rate of EI by using VL did not result in any significant difference for other adverse events. This is probably due to small sample size, leading these results inconclusive, as the TSA has proved. Even though, a lower trend for the incidences of hypoxemia and aspiration with VL can still be identified. Inexperienced operators benefited more from using VL, with significant lower incidences of EI and hypoxemia. For inexperienced operators who have not performed the tracheal intubation with VL and DL, visualization of the airway on VL screen can allow their supervisors to directly assist them in completing tracheal intubation themselves, thus improving the success rate [42]. In contrast, experienced operators with extensive training and experience on the tracheal intubation using DL might overshadow the benefits of VL, especially when patients have a normal airway or there are some VL-related difficult scenarios like secretions or blood in the airway or certain VL design-related deficiencies which may bate their benefits [19, 22, 43, 44].

Different from above non-fatal adverse outcomes, an increased trend was otherwise shown for incidence of new-onset cardiac arrest and short-term mortality by using VL. However, it should be noted that among the studies reporting the results of new-onset cardiac arrest and short-term mortality, only one was the study in which  emergency tracheal  intubation was performed by experienced operators [14]. The longer duration of intubation with VL by inexperienced operators may result in an increased risk of severe life-threatening adverse outcomes [19]. In addition, the pooled result of the short-term mortality was mainly from a CPR study [8]. Although the participants were randomized, the baseline clinical characteristics for both groups was not comparable, with more ischemic heart diseases in the VL group, leading the patients in this group to be at a higher risk of severe adverse outcomes. It has been reported that the use of a VL can reduce the chest compression interruptions for both experienced and inexperienced operators [11, 18]. However, visualization of VL can be unfavorably compromised by increased amount of secretions and emesis in the upper airway, which are common during the CPR, leading to a longer duration of intubation which may worsen the patient’s prognosis [16, 22]. Besides traditional airway suctioning, other techniques to decrease influences of massive secretions and emesis on airway visualization and intubation procedure, such as intentional esophageal intubation (IEI) [45] and suction-assisted laryngoscopy and airway decontamination (SALAD) [46], have been described. Furthermore, the use of airway decontamination technique provides improved intubation conditions with a VL. However, these methods appear mostly in case report or simulated mannequin study and have not been generalized in clinical practice. Nowadays, even for skilled operators like trained anesthesiologists, DL might still be the first choice for urgent tracheal intubation during the CPR, especially in patients without difficult airways.

Our study included RCTs and observational studies. The overall risk assessment of bias for the included RCTs was classified as low risk. Although blinding was not adopted in the most RCTs, we judged “no blinding” as low risk, as it seems impossible to blind personnel in the urgent situations at times. The overall risk assessment of bias for the included observational studies was classified as moderate or severe risk, mainly due to the confounding factors. For most outcomes, the level of heterogeneity was low, but subgroup and sensitivity analyses based on some potential clinical heterogeneous factors had been performed in our analysis. Furthermore, the TSA ensured the credibility of the result of primary endpoint from RCTs.

There are some limitations in our analysis that deserve special attentions. First and foremost, the inclusion of observational studies inevitably introduces selection bias, leading the possibility that patients in the VL and DL groups differ significantly in terms of operator’s expertise, anesthesia methods, certain difficult airway characteristics, clinical scenarios, and even basic characteristics of patients [4,5,6, 8,9,10]. Besides the disparities between groups, a reporting bias might be also present due to self-reported or recall property of the data in the observational studies, leading to the inaccuracy of data collection. Second, some of the included studies were quality-improvement process within a before–after study [4, 9]. The better result may probably not only from the switch from DL to VL, but also from the improved patient management in quality-improvement studies. Third, during the searching process, 12 observational studies were found to be carried out in a same ED or ICU from a same hospital with an overlapped study period [3, 5, 10, 47,48,49,50,51,52,53,54,55]. Only the latest 3 large studies [3, 5, 10] covering the majority of other 8 studies were decided to be included. This might lose some validated participants while avoiding repeated enrolment. Fourth, definition of the expertise used in our analysis was somewhat arbitrary. A Cochrane review defines an experienced operator as a clinician with more than 20 tracheal intubations with each device and thence obtains a fewer failed intubations when using VL [56]. However, it is suggested that the number of procedures required to achieve proficiency with tracheal intubation using DL in an controlled environment like OR is approximately 50 [57, 58]. Given the challenges of limited cardiopulmonary reserve, difficulty on airway assessment, few chances of tracheal intubation practice in the ED or ICU, it is even hard to know how many procedures are required to achieve a definitive competence with tracheal intubation in this patient population. Anyhow, having one VL available for all urgent tracheal intubation could potentially offer a major safety advantage. Fifth, there might be still some other heterogeneous factors in our analysis, such as different types of VLs used and different in-hospital setting, i.e. ED, ICU, or general ward. About one third of our included studies used more than one type of VL [3,4,5, 8, 10, 17, 19], making subgroup analysis on this factor difficult. Although ED and ICU are two main in-hospital settings for urgent tracheal intubation, indications, intubation conditions, and severity of patients’ illness would not differ significantly, at least not as much as the difference between pre-hospital and in-hospital setting. Lastly, some studies used the episode of intubations or attempts instead of participants [3, 4, 12]. When calculating incidence of adverse outcomes, this would be inappropriate. However, sensitivity analysis by excluding these studies did not change the result of primary endpoint.

Conclusions

This systematic review and meta-analysis reveals that VL can reduce the risk of EI during urgent tracheal intubations in the ED and ICU patients, but does not provide significant benefits on other adverse events associated with tracheal intubation. Further studies are needed to demonstrate whether severe adverse events like cardiac arrest or mortality are significantly different between two devices. Furthermore, well-designed RCTs are further needed to focus on a specific scenario and should stratify some other prognostic indicators such as length of hospital stay and cost.