Abstract
Background
Furan has been considered to be toxic and carcinogenic. In this research, a new method was applied for the pre-concentration of furan in beverage samples followed by high-performance liquid chromatography-ultraviolet detection (HPLC-UV) determination.
Results
Determination was carried out by a UV detection at 205 nm. Some of the most important parameters, including extraction solvent, extraction time, sample temperature and volume, helium flow rate and salt effect were investigated and optimized. Linear calibration curve was obtained over a concentration range of 0.005 to 0.3 μg/mL with coefficient of determination (R2 = 0.998). Limit of detection and relative standard deviation were obtained as 0.001 μg/mL and 2.21% (n = 5), respectively.
Conclusions
The proposed method was applied successfully for the determination of furan in beverage samples. Among them, cola had a detectable amount of furan.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Background
Chemical food safety is a topic with an increasing concern within the respective community, although due to the present risk management procedures, the acute threat posed by harmful chemicals is very uncommon. Instead, there has been much more interest focused on the minute concentrations and consequent potential chronic symptoms. It has been postulated that chemical contaminants found in foods (and feeds) might be etiological factors behind several diseases (e.g., cancer), the causes of which are still greatly unknown. In addition to the environmental and agrochemical residues as well as natural toxins, contaminants produced during the food processing or storage, such as acrylamide, polycyclic aromatic hydrocarbons (compounds), benzene and furan, are becoming an issue.
Furan (C4H4O), a lipophilic contaminant, is formed during heating process for the manufacture of foods. Furan is an aromatic heterocyclic compound, which is highly volatile with a boiling point of 31°C. Furan has been identified in hot-air dried, baked, fried and roasted food items, such as cereal products and coffee as well as can- or jar-prepared foods, at low concentration levels (<200 ng/g) [1]. Its presence in food is known since the 1960s as one of the Maillard reaction products formed in the thermal processes of foods such as birch syrup, caramel, coffee, bread and cooked meat [2–10]. Nowadays, the formation of furan remains unclear, although it is known that its generation is related to the thermal degradation of carbohydrates, thermal oxidation of polyunsaturated fatty acids and/or thermal decomposition of ascorbic acid or its derivatives [11–14]. In recent years, the study of furan in foods has received special attention by several international food organizations such as the US Food and Drugs Administration and the European Food Safety Authority [15–17]. Due to its toxicity and carcinogenicity, furan has been included by the International Agency for Research on Cancer in the Group 2B as possibly carcinogenic to humans [18–20] and has been classified by the US Department of Health and Human Services as a human pathogen [21]. Therefore, there is a need to propose reliable methods for its determination in foods in order to obtain information about its occurrence as well as the source of exposure, mechanisms of formation and toxicology. The analysis of furan in food is not easy due to its high volatility, low molecular weight and low concentration levels in food products.
The most common analytical technique used to determine the concentration of furan is headspace extraction combined with gas chromatography-mass spectrometric (GC-MS) detection [15, 22, 23]. Later on, solid-phase microextraction followed by GC-MS has been developed [22, 24]. Both approaches are convenient sample preparation techniques as they are very simple and demand no expensive equipment for sample extractions/concentration.
In the present study, purge-and-trap method combined with HPLC was applied for the determination of the amount of furan in food samples. The detection wavelength for the absorption of furan was set at 205 nm [25]. A systematic study of the effective parameters on the extraction and desorption processes was used to obtain maximum sensitivity and selectivity. In addition, we investigated the applicability of the proposed method for the quantification of furan in solid and liquid food samples. The proposed method provides relatively low limits of detection (LOD; 0.001 μg/mL).
Methods
Experimental
Materials and instruments
Furan, water, methanol and acetonitrile of HPLC grade were purchased from Merck and Co. Inc. (Darmstadt, Germany). All Other reagents were of analytical grade and obtained from Merck.
Chromatographic conditions for sample analysis
Chromatographic Separation Were Carried out on a HPLC Instrument (Waters, Model 600, USA) equipped with a Waters 2487 UV-Vis Detector. A Zorbax Extend C18 column (250 mm × 5 mm, with 5 μm particle size) was purchased from Agilent Technologies Inc. (Wilmington, DE, USA). A mixture of water/methanol/acetonitrile (70:20:10 vol.%) at the flow rate of 1 mL/min was used as the mobile phase in the isocratic elution mode. The injection volume was 20 μL, and the detection wavelength was set at 205 nm.
Extraction procedure
A balloon containing standard solutions was connected to the trapping unit which was placed in the ice bath (Figure1). Trapping unit was filled with glass beads to increase its trappig efficiency. From the other side of the balloon, helium gas was purged through the heated sample solution under stiring condition. The interface between the balloon and trapping unit was heated by industrial drier to prevent furan condensation and it's return to the balloon.
Results and discussion
Selection of an appropriate extraction solvent
The selection of an appropriate solvent is important to achieve the highest sensitivity. Different solvents such as dimethy formamide, acetonitrile, methanol were studied as the extraction solvents. Among them, methanol showed the largest peak area and was used as the extraction solvent (Figure2A).
Utilized conditions: furan 0.01 μg mL−1, temperature 40°C, time 45 min, 1,000 rpm, helium flow rate 100 mL min.−1. Effect of extraction solvent on the peak area ( A); effect of the sample volume ( B); effect of extraction solvent volume ( C); effect of sample solution temperature, extraction solvent 4 mL methanol ( D); effect of helium flow rate, extraction solvent 4 mL methanol ( E); effect of extraction time, extraction solvent 4 mL methanol ( F).
Optimization of the studied parameters
Sample volume
Sample volume was studied in the range of 0.5 to 2 L. Peak area increased with the increase of the sample volume. The volume of 1 L was chosen as the optimum sample volume. More than 1 L leads to poor purging capacity and precision (Figure2B).
Volume of the extraction solvent
The volume of the extraction solvent was studied in the range of 2 to 5 mL. Four milliliter was chosen as an optimum volume for the extraction solvent. More than 4 mL leads to a decrease in analytical sensitivity basically due to dilution (Figure2C).
Extraction temperature
Temperature affects the purging efficicency and transfer rate of analyte from sample solution to gas phase. Purging efficicency increases with increase in the temperature of the sample solution mainly because of increase in the vapor pressure. Extraction temperature was studied in the range of 40°C to 80°C. At temperatures above 80°C, helium temperature increases which subsequently decreases the trapping efficiency and precision. The optimum temperature was obtained as 70°C (Figure2D).
Flow rate
Helium flow rate was studied in the range of 50 to 100 mL/min. By increasing the helium flow rate, volatile species were transferred more rapidly into the trapping solvent, and as a result, the concentration was increased. At higher flow rate of helium, trapping solvent splashed out. The optimum value was obtained as 100 ml/min (Figure2E).
Extraction time
Furan cannot remain in the trapping solvent for a long time due to its volatility. Extraction time was studied in the range of 45 to 240 min. Forty-five minutes was chosen as the optimum extraction time due to the increase peak area. The peak area was decreased after 45 min (Figure2F).
Salt effect
The effect of salt was investigated by adding the sodium chloride to the aqueous samples. Peak areas were slightly increased because of salting-out effect. A concentration ration of 0.3% NaCl was selected throughout the experiments.
Analytical performance of the method
According to the above optimum conditions obtained for each factor, the furan chromatogram can be achieved as shown in Figure 3 and in order to evaluate the linearity of the method, a calibration curve over a concentration range of 0.001 to 0.300 μg/ml of furan was obtained (Figure 4). In order to evaluate the linearity of the method, a calibration curve over a concentration range of 0.001 to 0.300 μg/mL of furan was obtained. Linear dynamic range, coefficient of determination, LOD, limit of quantification (LOQ) and relative standard deviation (% RSD) are shown in Table1. For calculation of LOD, it is considered to be the concentration of analyte that gives a signal that is equal to three times the standard deviation of the blank. This is equivalent to defining the LOD as that concentration at which the signal-to-noise ratio equals 3 at 99% confidence limit:
Similarly, LOQ in the above equation is defined as 10σblank. As shown in Table1, only small deviations between sequential determinations were found.
Validation of the method and analysis of beverage samples
The proposed methodology was applied for the determination of furan in several beverage samples (cola, coffee, beer, grape, sour cherry) in which, among them, cola was found to have a significant amount of furan (Table2). The beverage samples were spiked to assess the matrix effect. The high recoveries of furan at a level of 0.03 μg/mL showed no matrix interferences. The chromatograms of cola without spiking (a) and with spiking (b) are illustrated in Figure5.
Conclusions
In this study, a new pre-concentration method combined with HPLC technique was applied for extraction and determination of furan in beverage samples. Some effective parameters such as the extraction solvent, extraction time, sample temperature and volume, helium flow rate and salt effect were investigated and optimized. LOD and RSD were obtained as 0.001 μg/mL and 2.21% (n = 5), respectively. The proposed methodology was applied successfully for the determination of furan in cola, coffee, beer, grape and sour cherry which, among them, cola was found to have a significant amount of furan.
References
Zoller O, Sager F, Reinhard H: Furan in food: headspace method and product survey. Food Addit Contam 2007,24(S1):91–107.
Merritt C, Bazinet ML, Sullivan JH, Robertson DH: Mass spectrometric determination of the volatile components from ground coffee. J Agric Food Chem 1963, 11: 152–155.
Stoffelsma J, Simpa G, Kettenes DK, Pypker J: New volatile components of roasted coffee. J Agric Food Chem 1968, 16: 1000–1004. 10.1021/jf60160a010
Walter RH, Fagerson IS: Volatile compounds from heated glucose. J Food Sci 1968, 33: 294–297. 10.1111/j.1365-2621.1968.tb01371.x
Tatum JH, Shaw PE, Berry RE: Degradation products from ascorbic acid. J Agric Food Chem 1969, 17: 38–40. 10.1021/jf60161a008
Persson T, von Sydow E: Aroma of canned beef: gas chromatographic and mass spectrometric analysis of the volatiles. J Food Sci 1973, 38: 377–385. 10.1111/j.1365-2621.1973.tb01436.x
Maga JA: Furans in foods. Pre-Concentration of Ultra-Trace Furan in Beverage Samples and its determination by high performance liquid cromatography 1979, 11: 355–400.
Kallio H, Leino M, Salorinne L: Analysis of the headspace of foodstuffs near room temperature. J High Resolution Chromatogr 1989, 12: 174–177. 10.1002/jhrc.1240120311
Shimoda M, Shibamoto T: Isolation and identification of headspace volatiles from brewed coffee with an on-column GC/MS method. J Agric Food Chem 1990, 38: 802–804. 10.1021/jf00093a045
Holscher W, Steinhart H: Investigation of roasted coffee freshness with an improved headspace technique. Z Lebensm Unters Forsch 1992, 195: 33–38. 10.1007/BF01197836
Locas CP, Yaylayan VA: Origin and mechanistic pathways of formation of the parent furan: a food toxicant. J Agric Food Chem 2004, 52: 6830–6836. 10.1021/jf0490403
Becalski A, Seaman S: Furan precursors in food: a model study and development of simple headspace method for determination of furan. J Assoc Anal Communities Int 2005,88(1):102–106.
Fan XJ: Formation of furan from carbohydrates and ascorbic acid following exposure to ionizing radiation and thermal processing. J Agric Food Chem 2005, 53: 7826–7831. 10.1021/jf051135x
Mark J, Pollien P, Lindinger C, Blank I, Mark T: Quantitation of furan and methylfuran formed in different precursor systems by proton transfer reaction mass spectrometry. J Agric Food Chem 2006, 54: 2786–2793. 10.1021/jf052937v
US FDA: Determination of furan in foods. 2008.http://www.cfsan.fda.gov/dms/furan.html 7 May 2004(updated 27 October 2006) [cited: 2008 Jul 21], Available from
US Food and Drug Administration (FDA): Furan in foods - thermal treatment. 2004. www.cfsan.fda.gov/~lrd/fr040510.html
European Food Safety Authority (EFSA): Report of the scientific panel on contaminants in the food chain on provisional findings on furan in food. EFSA J 2004, 137: 1–20.
Glatt H, Schneider H, Liu Y: A cell line for the sensitive detection of genotoxic effects induced by carbohydrate pyrolysis products and other food-borne chemicals. Mutat Res 2005, 580: 41–52. 10.1016/j.mrgentox.2004.11.005
Byrns MC, Vu CC, Neidigh JW, Abad JL, Jones RA, Peterson LA: Detection of DNA adducts derived from the reactive metabolite of furan, cis-2-butene-1,4-dial. Chem Res Toxicol 2006, 19: 414–420. 10.1021/tx050302k
International Agency for Research on Cancer (IARC): Dry cleaning some chlorinated solvents and other industrial chemicals. Monogr Eval Carcinog Risks Hum 1995, 63: 393.
US Department of Health and Human Service: 11th Report on Carcinogens, Public Health Service, National Toxicology Program, Research Triangle Park. , NC; 2005. www.ntp.niehs.nih.gov/ntp/roc/eleventh/profiles/s090fura.pdf
Bianchi F, Careri M, Mangia A, Musci M: Development and validation of a solid phase micro-extraction-gas chromatography-mass spectrometry method for the determination of furan in baby-food. J Chromatogr A 2006, 1102: 268–272. 10.1016/j.chroma.2005.10.056
Hasnip S, Crews C, Castle L: Some factors affecting the formation of furan in heated foods. Food Addit Contam 2006,23(3):219–227. 10.1080/02652030500539766
Goldmann T, Périsset A, Scanlan F, Stadler RH: Rapid determination of furan in heated foodstuffs by isotope dilution solid phase micro-extraction-gas chromatography-mass spectrometry (SPME-GC-MS). Analyst 2005, 130: 878–883. 10.1039/b419270b
Hyman D, Sluiter A, Crocker D, Johnson D, Sluiter J, Black S, Scarlata C: Determination of acid soluble lignin concentration curve by UV-Vis spectroscopy. Natl Renewable Energy Laboratory (NREL) Analytical Proced 2008, 1–13. Technical Report NREL/TP-510–42617 Technical Report NREL/TP-510-42617
Acknowledgment
The authors would like to thank Dr. M.H. Mousazadeh for his helpful comments.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no competing interests.
Authors’ contributions
HK conceived of the study and participated in coordination. AK participated in experimental design and optimization. EK participated in preconcentration experiments. MT participated in graphical and tables preparation. SK participated in HPLC experiments. MB and HP participated in the statistical analysis of data. all authors read and approved the final manuscript.
Authors’ original submitted files for images
Below are the links to the authors’ original submitted files for images.