1 Introduction

Metal plates have been used as implants to fix or replace lost organs as well as organs with unsatisfied functions [1]. Lane first introduced this method in 1890, where metal plates were used for bone fracture fixation [2]. There were issues of low yield strength and poor corrosion resistance facing metal implants in the past until several research works [3,4,5,6] that were carried out in the 1920s that led to the disco very of 18-8 stainless steel for potential applications [7,8,9]. The most widely used alloy for biomedical implant in recent times is 316L stainless steel (316L SS) due to its low cost as well as its ability to provide reasonable corrosion resistant [2]. However, when there is a need for high wear resistance material for applications such as artificial joints, the CoCrMo alloys, and titanium alloys are usually considered [2]. Titanium and Ti–6Al–4V are much used in orthopedics because of their high shear strength and wear resistance [9].

For the past decades, the world has witnessed a surge in the number of road accidents [10], therefore, there is a high demand for metallic bone plates for bone fractures [10]. However, with more youths coming down with need for bone implant due to accidents and the average service life of an implant is between 10 and 15 years, so there is high probability that orthopedic surgical patients will need revision surgeries in their lifetime. Hence, Ti-alloys with porous structure could be studied to provide solutions to fixation of bone fractures associated with the high rates of accidents.

The short service life of implants is associated with micro-shifts at the interface between the implant and the bone due to the discrepancy between the two mechanical moduli [11]. There is a dissimilarity between Young’s modulus of titanium implant (103–120 GPa) and bone (10–30 GPa) which can result into stress shielding that can affect the healing and remodeling of the bone [12]. There is a direct relationship between Young’s modulus and other mechanical properties of Ti–6Al–4V, therefore, a reduction in the elastic modulus of the alloy will also lead to decrease in its strength; once the strength is enhanced, the elastic modulus also increases [12]. So, there is a need to strike a balance between the elastic modulus and the strength of the Ti–6Al–4V orthopedic implant.

In this article, the mechanical properties of porous Ti–6Al–4V was evaluated by varying the sintering time and particle sizes of porous Ti-alloy powder while the hydrophilicity of the samples were examined by contact angle measurement.

2 Materials and methods

2.1 Materials

Spherical Ti–6Al–4V alloy powder was procured from Phelly Material Inc. (New Jersey, USA), alumina ceramic crucibles were purchased from Sigma Aldrich Co. (St. Louis, MO, USA), whiles British Standard (BS) sieve, muffle furnace were both obtained from ELE International (Model no: 88–0110/02, England, UK). A setup for contact angle/wettability studies comprised of a black box, 60 Watts incandescent bulb, a converging lens, a digital camera, phosphate-buffered saline solution (disodium hydrogen phosphate (1.44 g/L), potassium chloride (0.2 g/L), potassium dihydrogen phosphate (0.24 g/L) and sodium chloride (8.0 g/L)) with pH of 7.00 as bio-fluid and micropipettes were locally purchased in Abuja, Nigeria.

3 Experimental procedures

3.1 Sintering process

The as-received Ti–6Al–4V powder was sieved into three different particle sizes ranging from 0–150 µm and 151–300 µm using the BS sieves. The different particle sizes were then placed in alumina ceramic crucibles and heated in a muffle furnace to 980 °C at a ramp rate of 20 °C/min and were held at this temperature for 3 h, 4 h and 5 h. The furnace was then allowed to cool down with the samples in it at the ramp rate of 20 °C/min.

3.2 Sample preparation for optical and SEM characterizations

The sintered samples were first sprayed with ethanol, then compressed air was used on each sample to remove unattached Ti–6Al–4V particles from their surfaces. No other metallographic procedures were carried out on the sintered samples before the optical and SEM examinations.

3.3 Contact angle measurement

The contact angles, θe were used to determine the wettability of the sintered Ti–6Al–4V samples. This measurement was developed by Young in 1805 [13] by establishing the angle that a liquid drop will make with a solid surface. The contact angle is defined by the mechanical equilibrium of the drop under the influence of three interfacial tensions which includes the solid–vapor (\({\gamma }_{sv})\), solid–liquid (\({\gamma }_{sl})\) and liquid–vapour \(({\gamma }_{lv})\) surface tensions. The equilibrium equation is known as Young’s equation [14]:

$${\gamma }_{lv}\mathrm{cos}{\theta }_{e}= {\gamma }_{sv}-{\gamma }_{sl}$$
(1)

where \({\theta }_{e}\) is known as Young contact angle.

The contact angle measurement was obtained from a direct photograph, through the deflection (or reflection) of light rays by the liquid prism. The surface of each sample was cleaned with ethanol and compressed air to get rid of any loose Ti–6Al–4V particles. The sample was then carefully placed on the sample holder in a carton box, while the bulb was switched on. The micropipette was used to draw 50 µL of the prepared solution, and few drops of the buffer solution were gently dropped on the substrates through the provided opening on the box. The biofluid stayed for 2 s on each sample before the digital images were recorded. The recorded images were subsequently transferred to Gwyddion 2.45 software package to determine the contact angles between the liquid drop and the sample.

3.4 Determination of Young’s modulus

The Young’s moduli of the samples were characterized using nanoindentation techniques [7]. This method was chosen because of its non-destructive nature compared to other mechanical testing methods. Elastic modulus, E, and hardness, H, were the mechanical properties frequently measured using load and depth sensing indentation techniques. The data were extracted from one complete cycle of loading and unloading [15]. The deformation during loading is assumed to be both elastic and plastic as the indenter formed permanent impressions on the samples. During unloading, elastic displacement was expected to be recovered and that aid in analysis. This method is therefore not applicable to materials in which plasticity reverses during unloading. The Young’s modulus of the specimen was extracted using Eqs. (2) and (3), respectively.

$$S= \frac{2}{\sqrt{\pi }}{E}_{r }\sqrt{A}$$
(2)
$$\frac{1}{{E}_{r}}=\frac{(1-{v}^{2 })}{E}+ \frac{( 1- {v}_{i}^{2})}{{E}_{i}}$$
(3)

where S is the measured unloading stiffness and A is the contact area, \({E}_{r}\) is known as the reduced modulus, which takes into account the effect of non-rigidity of the indenters. \(E\) and \(v\) are Young’s modulus and Poisson’s ratio for the specimen, respectively. \({E}_{i}\) and \({v}_{i}\) are the same parameters for the indenter.

3.5 Compressive strength measurement

Since the application of the implant in the body will be subjected to compressive stresses, the compressive strength of the samples was obtained. The compressive strength of each sample was determined using a universal testing machine. Each sample has a hemisphere shape-like, and the load rate was 3000 mm/min (Fig. 1).

Fig. 1
figure 1

Contact angle parameters (Modified after [13])

4 Results and discussion

4.1 Optical and SEM characterizations

The optical results from ProScope® optical instrument are presented in Fig. 2; it was observed that as the sintering time increases, the pores between the Ti–6Al–4V powder particles disappeared gradually, this is more evident if you compare Fig. 2a and b. More detailed images are shown in Fig. 3 below using SEM micrographs. The micrographs in Fig. 3 are for same particle size range (300–212 µm) sintered at different temperatures; the contrast between the void’s sizes in Fig. 3b and f illustrates the elaborate effect of sintering time on pores within the samples.

Fig. 2
figure 2

a Optical image of As-received Ti–6Al–4V powder. b Optical image of Ti–6Al–4V with particle size 212–150 μm, sintered for 4 h

Fig. 3
figure 3

SEM micrographs of Ti–6Al–4V with particle size: a 300–212 µm sintered for 3 h (× 500), b 300–212 µm sintered for 3 h (× 1000) c 300–212 μm sintered for 4 h (× 500), d 300–212 μm sintered for 4 h (× 1000). e 300–212 μm sintered for 5 h (× 500). f 300–212 μm sintered for 5 h (× 1056)

4.2 Contact angles and wettability

The effect of particles size and sintering time on the contact angle, which signifies the wettability of the substrate are presented in Fig. 4. The wettability for particles sizes 212–300 µm was much improved with their contact angles less 25\(^\circ .\) This angle was also less than the contact angles for < 150 µm, and 212–300 µm. For particles sizes 212–300 µm, the contact angles, \(\theta\) at 3 h, 4 h and 5 h are respectively \(\sim 7^\circ , \sim 20^\circ ,\) and 25 \(^\circ\). For particles sizes 150–212 µm, the contact angles, \(\theta\) at 3 h, 4 h and 5 h are \(\sim 15^\circ , \sim 25^\circ\) and 52\(^\circ\), respectively whiles \(\theta\) for < 150 µm at 3 h, 4 h, and 5 h were \(\sim 18^\circ , \sim 45^\circ ,\) and 57 \(^\circ\), respectively.

Fig. 4
figure 4

Wettability via contact angle versus particle sizes

However, it is evident that the largest particle size group (212–300 µm) has the lowest contact angles which means that it has the highest wettability compared to particles sizes < 150 µm and 150–212 µm. This observation is because coarse particles sinter slower than finer particles thereby creating micro-roughness, and besides, the biofluid stayed on the samples for 2 s which allowed the biofluid to fill the pores of the samples through capillary action. There is an apparent effect of sintering time and particles size on the contact angles. From our view, sintering time of 3 h has better wettability for the formation of porous Ti–6Al–4V samples because as the sintering time increases, the titanium alloy particles are more likely to become denser. Lower contact angles suggest the force of adhesion between the solid–liquid interface was more significant than the cohesive force of the liquid–liquid interaction.

5 Mechanical properties

A diamond Berkovich indenter was used for the analysis; the elastic modulus and the Poisson ratio of the indenter were 1141 GPa and 0.07, respectively [16]. The Young's moduli for the samples with different particle size ranges are presented in Fig. 5.

Fig. 5
figure 5

Effect of sintering time on Young's modulus for different particle sizes

The Young’s moduli of the samples increase with the sintering time due to an increase in packing density of the particles, which gradually eliminate the pores within the samples. This led to stiffer and denser material with relatively high Young’s modulus. This result coincides with what was reported by some researchers. Lucía Reig et al. observed that the stiffness and bending strength of sintered porous Ti–6Al–4V samples increases with sintering time [17, 18], while Jose Luis et al. established that there is a polynomial relationship between Young’s modulus and relative density of the sintered porous Ti alloy [19]. The negative effect of porosity on Young’s modulus was also corroborated by Martínez et al. [20].

In contrast, the above assertion was only true for samples with particle sizes < 150 µm and 212–300 µm. The Young’s modulus result from samples with particle size 150–212 µm was not consistence with what was reported in literatures even after repeated trials.

However, samples sintered from finer particle sizes tend to have higher Young’s modulus when compared to ones formed from coarse particle sizes, one of the reasons for this observation is that finer particles are likely to develop better interparticle necking during sintering which in turn will reduce the porosity and give rise to higher Young’s modulus. This is also in line with was reported various literatures [18, 19].

The Young’s modulus of samples with particles less than 150 µm sintered for 5 h was found to be ~ 35 GPa, and this is much closer to that of the natural bone.

5.1 Compressive strength

The compressive strength was extracted from the stress versus deformation graph using the data from the universal testing machine. The compressive strength indicates a point on the graph where the applied stress is no longer proportional to deformation; this is also known as the yield point. Figure 6 shows the compressive strengths obtained for different particle sizes and at different durations.

Fig. 6
figure 6

Effect of sintering time on Compressive strength for different particle sizes

The sample with sizes less than 150 µm sintered for 5 h has the highest compressive strength of 180 MPa; one of the reasons for this observation is that the less than 150 µm sample has the highest particle size variation thereby making it easy for smaller particles to diffuse into the pores in the sample during sintering.

6 Conclusion

This work presents the effect of sintering time on the mechanical properties of Ti–6Al–4V powder. The wettability of the Ti alloy samples was also evaluated. It was observed that the wettability decreases with an increase in sintering time. The smallest particle size group (< 150 µm) gives the highest contact angle while the particle sizes 212–300 µm has the lowest contact angle indicating highly wettable surface due to micro-roughness which enhances the spread of the bio-fluid drop.

Furthermore, 3 h sintering time has the lowest contact angle compared to other sintering time, and this suggests that 3 h sintering time would be optimum for the formation of porous Ti–6Al–4V but unfortunately, the samples sintered at this temperature do not have the mechanical capability to function properly in a load bearing area of the body.

In addition, Young’s modulus and compressive strength of the samples through Nano-indentation and universal testing machine were analyzed, respectively. It was found that the particles size range of < 150 μm sintered for 5 h has Young’s modulus of \(\sim\) 35 GPa which is most closer to that of the bone (bone Young’s modulus is between 10–30 GPa); when compared with Ti–6Al–4V datasheet (103–120 GPa), we observed that there is great disparity between the two moduli. One of the reasons for the large difference in the values of Young’s moduli is the presence of voids due to the sintering process of the samples, the voids make dislocation of the Ti alloy atoms easier, thereby resulting in the smaller Young’s modulus for the porous Ti alloy. Metallic powder sintered at lower temperatures tends to have lot of voids or porosity which reduces the mechanical properties and increases the wettability of the samples.

More relevant analysis should be carried out to determine how the particle sizes and the different sintering time of the samples may affect cell proliferation and focal adhesion of the osteoblast cells; to measure the effect of porosity on the mechanical properties of the samples. These are challenges for future work.