Abstract
Introduction
We developed Eforto®, an innovative system for (self-)monitoring of grip strength (GS) and muscle fatigability (Fatigue Resistance (FR = time until GS decreased to 50% of maximum during sustained contraction) and grip work (GW = area under the strength-time curve)). The Eforto® system consists of a rubber bulb that is wirelessly connected to a smartphone-based application, and a telemonitoring platform. The aim was to evaluate the validity and reliability of Eforto® to measure muscle fatigability.
Methods
Community-dwelling older persons (n = 61), geriatric inpatients (n = 26) and hip fracture patients (n = 25) were evaluated for GS and muscle fatigability. In community dwellers fatigability was tested twice in the clinic (once with Eforto®, once with Martin Vigorimeter (MV), standard analog handgrip system) and for six consecutive days as a self-assessment at home with Eforto®. In hospitalized participants, fatigability was tested twice using Eforto®, once by a researcher and once by a health professional.
Results
Criterion validity was supported by good to excellent correlations between Eforto® and MV for GS (r = 0.95) and muscle fatigability (FR r = 0.81 and GW r = 0.73), and no significant differences in measurements between both systems. Inter-rater and intra-rater reliability for GW were moderate to excellent (intra-class correlation: 0.59–0.94). The standard error of measurement for GW was small for geriatric inpatients and hip fracture patients (224.5 and 386.5 kPa*s) and higher for community-dwellers (661.5 kPa*s).
Discussion/conclusion
We established the criterion validity and reliability of Eforto® in older community-dwelling persons and hospitalized patients, supporting the implementation of Eforto® for (self-)monitoring of muscle fatigability.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Loss of intrinsic capacity, the composite of all the physical and mental capacities that an individual can draw on [1], is a common health condition associated with ageing [1, 2], which leads to a loss of independence. Intrinsic capacity comprises all the mental and physical capacities that an individual can rely on. The intrinsic capacity level is affected by several factors such as the presence of diseases, injuries and age-related changes [3]. Various hypotheses on the aging process have been proposed, including genomic instability, epigenetic and metabolic disarray, oxidative stress, DNA damage, telomere shortening, inflammation, apoptosis, lipotoxicity, and mitochondrial modifications. Mitochondrial modifications have been extensively studied in aging conditions since they play an important role in different cellular pathways such as oxidative stress regulation, adenosine triphosphate biogenesis, mitophagy, and apoptosis [4,5,6]. These ageing processes affect muscle composition and function. Vice versa, muscle function may be a sensitive marker of progression of the ageing process and thus intrinsic capacity. The rapid increase in prevalence of older adults with compromised intrinsic capacity puts significant pressure on social and health care systems [7]. Solutions for early-onset detection and prevention of decreasing intrinsic capacity are currently unavailable. Existing technologies focus primarily on the detection of late-stage symptoms like frailty and dependency. Their assessment requires professional involvement, which makes them less suitable for long-term monitoring. Early identification and monitoring of changes in intrinsic capacity are key to better prevention and health management. Based on this need, we constructed a prototype smartphone-based system to assess muscle reserve recruitment capacity in persons with or without frailty, acting as an early warning system for the decline in intrinsic capacity. The Eforto® system (shown in Fig. 1) is designed to monitor muscle fatigability (fatigue resistance (FR) and grip work (GW)) as an ecological and dynamic marker of a person’s physical reserve capacity and resilience. The system facilitates (self-)assessment of muscle fatigability by measuring the maximal grip strength (GSmax) an individual can apply and sustain until their GS drops to 50% of its maximum (i.e., FR-test [8,9,10]). Test results can be consulted remotely through a secured web-based platform.
Protocols for the measurement of FR and GW were originally developed and validated using the Martin Vigorimeter (MV, KLS Martin Group, Germany), which is a classic analog pneumatic handgrip system [8, 9]. Pneumatic systems, such as rubber bulbs, are better tools to measure muscle fatigability and more valid to discriminate between subjects of varying degrees of frailty compared with clamp-based instruments such as the Jamar dynamometer [11, 12]. Measurements with MV require involvement of trained professionals and is therefore less suitable for remote monitoring than Eforto®, which allows self-assessments with application support.
The aims of the present study were: (1) to assess the criterion validity of Eforto® against the analog MV measurement as a reference system in measuring muscle fatigability, and (2) to investigate the inter-rater and intra-rater reliability of muscle fatigability measured with Eforto® in community-dwelling older persons, hospitalized geriatric patients and older patients with hip fracture.
Materials and methods
Study sample
One hundred and twelve older persons from three ongoing studies were enrolled (Table 1): 61 community-dwellers aged 80+ years participating to the BrUssels sTudy on The Early pRedictors of FraiLtY (BUTTERFLY [13, 14], Vrije Universiteit Brussel & Universitair Ziekenhuis Brussel, Belgium), 26 hospitalized geriatric patients participating in the Bedside Resilience Registry (BRR, Radboud University Medical Center, The Netherlands) and 25 patients with hip fracture participating in the RESilience Hip fracture PatiEnts study (RESHAPE, Ziekenhuis Groep Twente, The Netherlands). Subjects consecutively enrolled in the study between August 2020 and November 2020 (inclusive). The study was conducted ethically in accordance with the World Medical Association Declaration of Helsinki. Following the COSMIN consensus guidelines for psychometric studies [15] we included a minimum of 50 participants per context for the psychometric evaluation for sufficient statistical power (patients are considered as 1 group here).
In- and exclusion criteria
a.Community-dwelling older persons
Older volunteers were recruited among participants of the BrUssels sTudy on The Early pRedictors of FraiLtY (BUTTERFLY). Subjects were eligible if they were aged 80 and over, could walk independently, lived independently, if they were mentally fit (MMSE > 23/30), and not frail according to the Groningen Frailty Indicator (GFI < 4/15) [16], Rockwood Frailty Index (RFI < 0.25/10) [17], and/or the adapted Fried Frailty Index (FFI < 3/4) [18]; (exhaustion, weight loss, gait speed, and grip strength). Participants were excluded if they underwent surgery or any radiotherapy or chemotherapy during the past six months. Participants with CRP > 10 mg/l were excluded, as this refers to an acute inflammatory state and not to a chronic low-grade inflammatory profile [19]. The Eforto® assessments were added to the study protocol by means of an amendment, which was approved by the Medical Ethics Committee of the Universitair Ziekenhuis Brussel (B.U.N. 143,201,424,976). All participants provided written informed consent. All assessments took place at the Vrije Universiteit Brussel and the Universitair Ziekenhuis Brussel (Brussels, Belgium).
b.Hospitalized geriatric patients
Geriatric patients were consecutively recruited among participants of the Bedside Resilience Registry (BRR). All new admissions to the geriatrics ward meeting the in- and exclusion criteria (65+ years, expected hospital stay of more than two days, life expectancy of more than 14 days, speaking and understanding Dutch, no contact isolation, no severe cognitive impairment, no physical impairments for GS measurements) were eligible for the BRR study and the current sub study. The study was reviewed by the research ethics committee of the Radboud university medical center (file 2018-4973). It did not fall within the remit of the Medical Research Involving Human Subjects Act (WMO). The ethics committee approved the study based on the Dutch Code of conduct for health research, the Dutch Code of conduct for responsible use, the Dutch Personal Data Protection Act, and the Medical Treatment Agreement Act. All participants provided written informed consent. All assessments took place at Radboud University Medical Center (Nijmegen, The Netherlands).
c. Patients with hip fracture
Old patients with hip fracture were recruited for the RESilience Hip fracture PatiEnts study (RESHAPE) in the multidisciplinary Centre for Geriatric Traumatology at the Department of Trauma Surgery at Ziekenhuisgroep Twente Almelo-Hengelo. Participants meeting the in- and exclusion criteria (age > 70 years, surgically treated for a hip fracture, no total hip replacement, no pathological or periprosthetic fracture, no severe cognitive impairment, no terminal illness with life expectancy < 14 days and no physical impairments for grip strength measurements) were eligible for the RESHAPE study. The study was reviewed by the Local ethics committee Almelo (file ZGT20-42) and Medical Research Ethics Committee United, Nieuwegein (file W20.195). It did not fall within the remit of the Medical Research Involving Human Subjects Act (WMO). All participants provided written informed consent. All assessments took place at the Department of Trauma Surgery at Ziekenhuisgroep Twente (Almelo, The Netherlands).
Study design and primary outcomes
Eforto® and muscle fatigability assessment
The Eforto® system comprises of the original large rubber bulb of the MV, connected to an analog Honeywell TruStability HSCMANN100PGAA3 gauge type pressure sensor, with signal axial barbed port, 0–100 PSI measurement range, and accuracy of ± 0.25% FSS BFSL (Full Scale Span Best Fit Straight Line). The sensor was connected to an analog-to-digital conversion (resolution = 10-bit and sample frequency = 100 Hz) and data transmission system based on the BITalino (r)evolution, which has been previously validated[20]. The system was connected via Bluetooth to the Eforto® application on a smartphone. Eforto® can be set to the professional or self-test modes. In the professional mode, a trained assessor provides the standardized instructions (same as MV) and encouragements during the test. The assessor can monitor the GS readings on the smartphone screen during the measurement. In the self-test mode, the participant is guided through the steps by the application providing standardized auditory and visual instructions and motivational cues. In both settings, GS readings are not visible to the participant during the measurement. Standardized instructions for the positioning of the bulb in the hand and fingers was provided for all test procedures, by the assessors (when using the MV and Eforto®) as well as by the smartphone app (both in the professional mode as in the self-test mode): bulb positioned in the middle of the hand with the tube of the MV or the Eforto® case pointing upwards, the four fingers along one side of the bulb and the thumb along the other side of the bulb.
Eforto® was calibrated prior to each assessment. The test protocol was based on previous studies conducted at VUB/UZB [8,9,10]. Briefly, participants were asked to maintain the following position: shoulder adducted and neutrally rotated, elbow flexed at 90°, forearm in neutral position, and wrist in slight extension (0°–30°). The GSmax-test was repeated three times. Each time the participants squeezed the bulb as hard as possible with their dominant hand for five seconds. The highest score (in kPa) of the three attempts was noted as GSmax. Next, for the FR-test, participants were instructed to squeeze the bulb as hard as possible and for as long as possible, until the GS dropped below 50% of its maximum. Standardized verbal encouragement (‘harder, harder’) was provided by the assessor (professional mode) or by the Eforto® application (self-test mode) each time the GS dropped by ≥ 2 kPa relative to the previous value. In both modes, the application verified the GS peak during the first five seconds of the FR-test and if the GS peak was less than 80% of the predetermined GSmax, the test was aborted and restarted after at least 30 s of rest. In case the GS peak obtained during the FR-test exceeded the predetermined GSmax, this score was set as the new GSmax. The initial three GSmax measurements are required to obtain the maximum GS, which serves as the reference point for the FR test, which is a maximal endurance test. All data were automatically transferred to, and securely stored on, the online Eforto® platform.
Criterion validity
Sixty-one community-dwelling older persons performed the fatigability tests at the clinical study center once using the analog (MV) and once with Eforto® (professional mode). Tests were performed on the same day, in a random order with at least a 1-h interval. Each participant was tested by the same well-trained assessor (n = 4) for both measurements, except for four cases in which the assessor was changed between measurements due to unavailability. These four cases (2 women and 2 men) were excluded for the statistical analysis to prevent inter-assessor bias. Another case (1 man) was excluded from the statistical analysis due to technical problems leading to lost data on the Eforto® platform (Table 1).
Intra-rater reliability
Thirty community-dwelling older persons of the previous cohort performed the tests using Eforto® at home (self-test mode) without professional supervision. After an individual information session explaining how to use Eforto®, participants were asked to perform the test daily for one week starting from the first day after the measurements at the study center. For 28 participants, complete sets of measurements of six consecutive days were available and used for statistical analysis (Table 1).
Inter-rater reliability
Data from 29 participants of the same community-dwellers cohort were used to analyze inter-rater reliability by comparing the first home self-assessment with the measurement that took place at the clinical study center. One man had no study center data. In addition, 26 hospitalized geriatric patients and 25 patients with hip fracture completed two assessments (in professional mode) on the same day with at least a 1-h interval. Three cases were excluded (one geriatric patient and two patients with hip fracture, all women) due to data recording issues. One test was supervised by a researcher (n = 3) and one by a health care professional (n = 4). The test sequence depended on the availability of the health care professional (Table 1). In total 77 measurements were used to assess the inter-rater reliability.
Secondary outcomes
Participants’ characteristics
Age, sex and living situation were noted for all participants. For the community-dwelling cohort, home medication was noted. For all participants Charlson’s Comorbidity Index [21] was computed.
Statistical analysis
GW was calculated using two methods: the estimated GW (GWestimated) and actual measured GW (GWmeasured).
For the analog MV measure of GW, as the integral cannot be calculated we used the GWestimated to evaluate the criterion validity, using the previously published equation [9,10,11,12, 22,23,24,25,26]:
with GWestimated being estimated grip work (kPa*s), GSmax being the highest maximal grip strength reached during GSmax and FR-test (kPa), FR being fatigue resistance (time (s) during which GS dropped to 50% of its maximum).
GWmeasured was calculated by integrating the actual GS over time during the FR-test [10, 12, 26]:
With GWmeasured is the measured grip work (kPa*s), GS is the grip strength (kPa), t is the time-interval (at 100 Hz = 0.01 s).
The difference in GSmax, and muscle fatigability (FR and GWestimated) measured with the analog MV and Eforto® was analyzed using a paired t test. Bland–Altman plots with limits of agreements and a linear regression line were plotted to verify proportional differences between both systems. Additionally, Pearson correlation coefficients were computed.
Intra-rater reliability of the GSmax, and muscle fatigability (FR and GWmeasured) were assessed using intra class correlations (ICC) derived from a 2-way mixed effects, absolute agreement, single measurement (measurements in self-test mode of six consecutive days) as well as a multiple measurements model (mean of two consecutive measurements during six days, i.e. mean day 1 + 2, day 3 + 4 and day 5 + 6) [27]. In addition, inter-rater reliability of the GWmeasured was assessed using ICC, but derived from a 2-way random effects, absolute agreement, single rater/measurement model [27]. Finally, standard error of measurement (SEMagreement = √(variance in repetition + measurement error variance)), smallest detectable change (SDC = 1.96*\(\surd 2\)*SEMagreement), systematic error proportion (SEP = variance in repetition/(subject variance + variance in repetition + measurement error)), and residual error proportion (REP = measurement error/(subject variance + variance in repetition + measurement error)) were calculated [28] (Table 1).
For interpretation of the ICC-values, the following criteria were used: 0.60 ≤ ICC < 0.70 = moderate, 0.70 ≤ ICC < 0.80 = good, 0.80 ≤ ICC < 0.90 = very good, and ICC ≥ 0.90 = excellent [8]. Significance was set a priori at p < 0.05 (Fig. 2).
Results
Participants’ characteristics are presented in Table 1. In brief, 61 community-dwelling old persons (mean age = 85.6 ± 3.3 years), 26 hospitalized geriatric patients and 25 patients with hip fracture participated (mean age = 84.1 ± 6.4 years and 81.9 ± 7.0 years).
Criterion validity compared to MV system
As shown in Table 2, no significant differences were found between MV and Eforto®, neither in GSmax nor in muscle fatigability (FR and GWestimated). Moreover, MV and Eforto® showed good to excellent correlations for the three outcomes (ranging from 0.67 to 0.95, Table 3). Correlations were generally the highest for GS and lowest for GWestimated. Bland–Altman plots (shown in Fig. 3) showed no evidence for a proportional difference in GSmax or muscle fatigability between both systems. Similar results were obtained when analyzing male and female separately (shown in Table 2 and Supplementary Fig. 1).
Intra-rater reliability
Self-assessment of GWmeasured on six consecutive days showed good reliability in community-dwellers (ICC = 0.73 [0.61–0.85]); which improved to excellent (ICC = 0.94 [0.89–0.97]) when the average values of two consecutive days were considered. The relatively small SEMagreement indicates minimal random and systematic error (Table 4).
Inter-rater reliability
Inter-rater reliability of GWmeasured showed good to very good ICC-values for both hospitalized geriatric patients and patients with hip fracture (ICC = 0.86[0.72–0.94] and 0.79[0.58–0.91]), whereas the ICC was moderate for the community-dwellers (ICC = 0.59[0.29–0.78]) (Table 4). SEMagreement values were small in geriatric inpatients and patients with hip fracture (respectively 224.5 and 386.5 kPa*s), indicating minimal random and systematic error, but greater in community-dwellers (661.5 kPa*s). FR-values followed the same trend as GWmeasured while GSmax was not that different between the three populations. See Supplementary Table 1 and 2 for intra- and inter-rater reliability analysis of GSmax and FR.
Discussion
The first objective of this study was to evaluate the criterion validity of Eforto® to measure muscle fatigability using the classic analog MV measure as the reference standard. The good to excellent correlation between the two instruments in GSmax and muscle fatigability confirms the validity of Eforto® in community-dwelling older persons. The correlation between the instruments for GSmax was higher than for FR. This difference in correlation is explained by the fact that the GSmax-test was performed four times (three consecutive GSmax attempts followed by a fourth effort during the FR test), while the FR-test was performed only once. Hence, the variation in repeated GSmax values is reduced by averaging over the four attempts. Previous evidence showed that the evaluation of sustained voluntary contractions is more reproducible after earlier familiarization with maximal voluntary contractions [29]. On the other hand, we cannot exclude the possibility that Eforto® may measure FR more accurately than the analog method, because the data registration starts and stops automatically based on the strength increase and subsequent strength decay as programmed in the application. Conversely, when assessing FR with the analog MV, the time from the start of the test until GS dropped to 50% of its maximum is measured by the assessor using a handheld stopwatch, which may be less accurate than the Eforto® application. Consequently, the correlation between Eforto® and MV for GWestimated is smaller than that for FR. Nevertheless, the small and non-significant differences between the two systems for GSmax and muscle fatigability indicate the absence of a systematic measurement error.
The second aim of this study was to examine the inter-rater and intra-rater reliability of the GWmeasured results obtained with Eforto®. Both inter-rater and intra-rater reliability of GWmeasured were moderate to excellent, with ICC-values ranging from 0.59 to 0.94. This is in line with a previous study that reported good to excellent inter- and intra-assessor reliability ICC-values of FR obtained with the analog MV in hospitalized geriatric patients ranging from 0.77 to 0.94 [8]. Moreover, we found that intra-rater reliability of GWmeasured improved when using the average of the values of two consecutive days compared to the values of six consecutive days, thus it seems that for long term monitoring, two measurements per week are sufficient. But it could also mean that there is substantial day-to-day variation. Further research should clarify if the day-to-day variation gives clinically relevant information or not. Additionally, inter-rater reliability ICC-values for muscle fatigability in old community-dwellers are lower compared to the other two population groups, whereas the ICC-values for GSmax are less different between the three populations. Moreover, the SEM-values for FR and GWmeasured were higher for community-dwellers than the hospitalized populations. This might be due to differences in test modes of Eforto® that were used (professional mode in the hospitals versus self-test mode in the community-dwellers). Also, the repeated tests were performed on the same day in the hospitals and on consecutive days in the community-dwellers cohort. Finally, our results show that the ICC-values for GSmax are high but those of FR remain lower, probably because of the bigger residual error. In future studies, the error variance could likely be reduced by shortening the FR-test protocol (i.e., until 25% drop instead of until 50% drop in GSmax) as suggested in a previous study[26].
According to the WHO, “fostering Healthy Ageing will require a much better understanding of common trajectories of intrinsic capacity and functional ability, their determinants and the effectiveness of interventions to modify them”. This is exactly what the Eforto® system aims to do: to offer a method for measuring and monitoring the intrinsic capacity [30, 31] of a person. Unique to Eforto® is that rather than focusing on a statistic measure (GS), it utilizes information from the dynamic performance of the individual (GW) to recruit or actualize that potential. This is more sensitive to identify early decline in intrinsic capacity[32] and enables to detect changes in intrinsic capacity long before frailty or dependency are detected by other methods. GW opens a window of opportunity for preventive measures, e.g., diet [33] and exercise [34, 35], to optimize the level of intrinsic capacity before the decline becomes problematic.
A limitation of this study is that we did not collect data on body composition including presence of sarcopenia. The exclusion criteria may result in a somewhat selected sample, which could limit generalizability of the findings. However, the reliability was evaluated in three distinct patient groups, with ICCs being in the same range for hospitalized geriatric patients and patients with hip fracture. In this study we used pressure readouts to measure GS, FR and GW since the pressure measured in the bulb reflects the muscle strength produced by the participant. However, Pneumatic systems, such as rubber bulbs, are better tools to measure muscle fatigability and more valid to discriminate between subjects of varying degrees of frailty compared with clamp-based instruments such as the Jamar dynamometer [11, 12]. In addition, it has been previously shown that GS measured by the Jamar dynamometer is even more dependent on hand anthropometry than measurement with the Martin Vigorimeter which is a Pneumatic device with exactly the same rubber bulb as used in Eforto® [36]. Finally, standardized instructions for the positioning of the bulb in the hand and fingers was provided for all test procedures, by the assessors (when using the Martin Vigorimeter and Eforto®) as well as by the smartphone app (both in the professional mode as in the self-test mode).
A strength is that validity and reliability were evaluated in three distinct clinically relevant settings, which demonstrates the clinimetrics of Eforto in every day clinical practice.
Conclusion
Overall, we conclude that Eforto® is a valid and reliable tool to measure muscle fatigability, which can be used in older persons both in supervised clinical settings as well as in unsupervised community settings.
Data availability statement
All data generated or analyzed during this study are included in this article or its supplementary material files. Further enquiries can be directed to the corresponding authors.
References
Beard JR, Officer A, de Carvalho IA et al (2016) The World report on ageing and health: a policy framework for healthy ageing. Lancet 387:2145–2154. https://doi.org/10.1016/s0140-6736(15)00516-4
Morley JE (2016) Frailty and sarcopenia: the new geriatric giants. Rev Invest Clin 68:59–67
Noppen M, Vincken W, Meysman M et al (1993) Clinical usefulness and safety of bronchoalveolar lavage (BAL) in elderly patients. Arch Gerontol Geriatr 16:33–38. https://doi.org/10.1016/0167-4943(93)90024-c
Lippi L, de Sire A, Mezian K et al (2022) Impact of exercise training on muscle mitochondria modifications in older adults: a systematic review of randomized controlled trials. Aging Clin Exp Res 34:1495–1510. https://doi.org/10.1007/s40520-021-02073-w
Kawai H, Kera T, Hirayama R et al (2018) Morphological and qualitative characteristics of the quadriceps muscle of community-dwelling older adults based on ultrasound imaging: classification using latent class analysis. Aging Clin Exp Res 30:283–291. https://doi.org/10.1007/s40520-017-0781-0
Inoue T, Shimizu A, Ueshima J et al (2022) Diet-induced inflammation is associated with sarcopenia and muscle strength in older adults who visit a frailty clinic. Aging Clin Exp Res. https://doi.org/10.1007/s40520-022-02195-9
Vermeiren S, Vella-Azzopardi R, Beckwee D et al (2016) Frailty and the prediction of negative health outcomes: a meta-analysis. J Am Med Dir Assoc 17:e1-1163. https://doi.org/10.1016/j.jamda.2016.09.010
Bautmans I, Mets T (2005) A fatigue resistance test for elderly persons based on grip strength: reliability and comparison with healthy young subjects. Aging Clin Exp Res 17:217–222. https://doi.org/10.1007/BF03324600
Bautmans I, Onyema O, Van Puyvelde K et al (2011) Grip work estimation during sustained maximal contraction: validity and relationship with dependency and inflammation in elderly persons. J Nutr Health Aging 15:731–736. https://doi.org/10.1007/s12603-010-0317-1
De Dobbeleer L, Beyer I, Hansen AM et al (2019) Grip work measurement with the Jamar dynamometer: validation of a simple equation for clinical use. J Nutr Health Aging 23:221–224. https://doi.org/10.1007/s12603-019-1155-4
De Dobbeleer L, Theou O, Beyer I et al (2018) Martin Vigorimeter assesses muscle fatigability in older adults better than the Jamar Dynamometer. Exp Gerontol 111:65–70. https://doi.org/10.1016/j.exger.2018.07.004
DeDobbeleer L, Beckwee D, Arnold P et al (2020) Muscle fatigability measured with Pneumatic and Hydraulic handgrip systems are not interchangeable. Exp Gerontol 136:110950. https://doi.org/10.1016/j.exger.2020.110950
CaoDinh H, Bautmans I, Beyer I et al (2019) Association between immunosenescence phenotypes and pre-frailty in older subjects: does cytomegalovirus play a role? J Gerontol A Biol Sci Med Sci 74:480–488. https://doi.org/10.1093/gerona/gly135
Vermeiren S, Beckwee D, Vella-Azzopardi R et al (2019) Evaluation of appendicular lean mass using bio impedance in persons aged 80+: a new equation based on the BUTTERFLY-study. Clin Nutr 38:1756–1764. https://doi.org/10.1016/j.clnu.2018.07.029
DeVet HC, Terwee CB, Mokkink LB et al (2011) Measurement in medicine: a practical guide. Cambridge University Press, Cambridge
Steverink NSJ, Schuurmans H, van Lis M (2001) Measuring frailty: developing and testing the GFI (Groningen frailty indicator). Gerontologist 41:236–237
Collerton J, Martin-Ruiz C, Davies K et al (2012) Frailty and the role of inflammation, immunosenescence and cellular ageing in the very old: cross-sectional findings from the Newcastle 85+ Study. Mech Ageing Dev 133:456–466. https://doi.org/10.1016/j.mad.2012.05.005
Sirola J, Pitkala KH, Tilvis RS et al (2011) Definition of frailty in older men according to questionnaire data (RAND-36/SF-36): the Helsinki Businessmen Study. J Nutr Health Aging 15:783–787. https://doi.org/10.1007/s12603-011-0131-4
Sproston NR, Ashworth JJ (2018) Role of C-reactive protein at sites of inflammation and infection. Front Immunol 9:754. https://doi.org/10.3389/fimmu.2018.00754
Batista D, Plácido da Silva H, Fred A et al (2019) Benchmarking of the BITalino biomedical toolkit against an established gold standard. Healthc Technol Lett 6:32–36. https://doi.org/10.1049/htl.2018.5037
Charlson ME, Pompei P, Ales KL et al (1987) A new method of classifying prognostic comorbidity in longitudinal studies: development and validation. J Chronic Dis 40:373–383. https://doi.org/10.1016/0021-9681(87)90171-8
Bautmans I, Gorus E, Njemini R et al (2007) Handgrip performance in relation to self-perceived fatigue, physical functioning and circulating IL-6 in elderly persons without inflammation. BMC Geriatr 7:5. https://doi.org/10.1186/1471-2318-7-5
Bautmans I, Njemini R, Predom H et al (2008) Muscle endurance in elderly nursing home residents is related to fatigue perception, mobility, and circulating tumor necrosis factor-alpha, interleukin-6, and heat shock protein 70. J Am Geriatr Soc 56:389–396. https://doi.org/10.1111/j.1532-5415.2007.01571.x
Beyer I, Bautmans I, Njemini R et al (2011) Effects on muscle performance of NSAID treatment with piroxicam versus placebo in geriatric patients with acute infection-induced inflammation. A double blind randomized controlled trial. BMC Musculoskelet Disord 12:292. https://doi.org/10.1186/1471-2474-12-292
Beyer I, Njemini R, Bautmans I et al (2012) Inflammation-related muscle weakness and fatigue in geriatric patients. Exp Gerontol 47:52–59. https://doi.org/10.1016/j.exger.2011.10.005
De Dobbeleer L, Beyer I, Njemini R et al (2017) Force-time characteristics during sustained maximal handgrip effort according to age and clinical condition. Exp Gerontol 98:192–198. https://doi.org/10.1016/j.exger.2017.08.033
Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15:155–163. https://doi.org/10.1016/j.jcm.2016.02.012
Kim HY (2013) Statistical notes for clinical researchers: evaluation of measurement error 1: using intraclass correlation coefficients. Restor Dent Endod 38:98–102. https://doi.org/10.5395/rde.2013.38.2.98
Hunter S, White M, Thompson M (1998) Techniques to evaluate elderly human muscle function: a physiological basis. J Gerontol A Biol Sci Med Sci 53:B204–B216
Beijersbergen CM, Granacher U, Vandervoort AA et al (2013) The biomechanical mechanism of how strength and power training improves walking speed in old adults remains unknown. Ageing Res Rev 12:618–627. https://doi.org/10.1016/j.arr.2013.03.001
Bautmans I, Van Puyvelde K, Mets T (2009) Sarcopenia and functional decline: pathophysiology, prevention and therapy. Acta Clin Belg 64:303–316. https://doi.org/10.1179/acb.2009.048
Gijzel SMW, van de Leemput IA, Scheffer M et al (2019) Dynamical indicators of resilience in postural balance time series are related to successful aging in high-functioning older adults. J Gerontol A Biol Sci Med Sci 74:1119–1126. https://doi.org/10.1093/gerona/gly170
de Sire A, Ferrillo M, Lippi L et al (2022) Sarcopenic dysphagia, malnutrition, and oral frailty in elderly: a comprehensive review. Nutrients. https://doi.org/10.3390/nu14050982
Coelho-Júnior HJ, Uchida MC, Picca A et al (2021) Evidence-based recommendations for resistance and power training to prevent frailty in community-dwellers. Aging Clin Exp Res 33:2069–2086. https://doi.org/10.1007/s40520-021-01802-5
Carvalho LP, Pion CH, El Hajj BG et al (2019) Effect of a 12-week mixed power training on physical function in dynapenic-obese older men: does severity of dynapenia matter? Aging Clin Exp Res 31:977–984. https://doi.org/10.1007/s40520-018-1048-0
Desrosiers J, Hebert R, Bravo G et al (1995) Comparison of the Jamar dynamometer and the Martin vigorimeter for grip strength measurements in a healthy elderly population. Scand J Rehabil Med 27:137–143
Acknowledgements
This study was jointly supported by EU, INNOVIRIS, VLAIO and ZonMW by means of an Active Assisted Living Programme Small Collaborative Project grant (grant n° AAL-2019-6-165-SCP). This study was partly funded by an “Interdisciplinary Research Program” grant from the research council of the Vrije Universiteit Brussel (VUB), Belgium.
Funding
This study was jointly supported by EU, INNOVIRIS, VLAIO and ZonMW by means of an Active Assisted Living Programme Small Collaborative Project grant (grant n° AAL-2019–6-165-SCP). This study was partly funded by an “Interdisciplinary Research Program” grant (grant n° IRP12) from the research council of the Vrije Universiteit Brussel (VUB), Belgium.
Author information
Authors and Affiliations
Contributions
LDD contributed to the data collection, data analysis and writing of the manuscript; MMS contributed to the data collection, data analysis and writing of the manuscript; MAJG contributed to the data collection and reviewed the manuscript; RJB contributed to the data collection and reviewed the manuscript; AJSJ contributed to the data collection and reviewed the manuscript; RT contributed to the development of the Eforto® hardware and software and reviewed the manuscript; HS contributed to the development of the Eforto® hardware and software and reviewed the manuscript; SL reviewed the manuscript; KB reviewed the manuscript; GP contributed to the data collection and reviewed the manuscript; MMRVH contributed to the study design and reviewed the manuscript; RJFM designed the study, contributed to the data analysis and writing of the manuscript; IB designed the study, contributed to the data analysis and writing of the manuscript. All authors approved the final draft of the manuscript.
Corresponding authors
Ethics declarations
Conflict of interest
The authors have no conflicts of interest to declare.
Statement of human and animal rights
The study was conducted ethically in accordance with the World Medical Association Declaration of Helsinki. No statement of animal rights are mandatory for this study since the participants are humans and no animals.
Informed consent
All participants provided written informed consent.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
De Dobbeleer, L., Swart, M.M., Geerds, M.A.J. et al. Validity and reliability of Eforto®, a system to (self-)monitor grip strength and muscle fatigability in older persons. Aging Clin Exp Res 35, 835–845 (2023). https://doi.org/10.1007/s40520-023-02365-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s40520-023-02365-3