1 Introduction

The quantitative analysis of uranium in minerals is the foundation for uranium exploration, environment evaluation, and radiation contamination.

The widely used quantitative analysis of uranium includes gamma spectroscopy, X-ray fluorescence, atomic absorption spectrometry (AAS), inductively coupled plasma mass spectrometry method (ICP-MS) and so on. Ammonium vanadate redox titration is often used to measure the uranium content in rock samples in China, where the uranium is reduced to a tetravalent state with divalent iron after decomposition of the sample with phosphoric acid, oxidation of excess divalent iron with nitric acid, destruction of excess nitric acid with urea, and finally titration of tetravalent uranium with a standard solution of ammonium vanadate using sodium diphenylamine sulfonate as an indicator. Different analysis methods are used with samples of different contents [1,2,3,4,5]. However, some difficulties like tedious pretreatment and complicated process, and high cost at faced. The gamma spectroscopy method is a semi-quantitative analysis method and is widely used to geological logging. With the AAS method, only monochromatic light in the atomic spectrum can be applied, so only one element at a time can be detected [6]. ICP-MS has high sensitivity and low detection limit, but it is only suitable for analyzing mineral samples with low-concentration-level uranium for the instruments ranges that cannot meet the requirements [3, 7,8,9]. ICP-MS is complicated to perform and requires precise adjustment, which is not conducive to improving the efficiency of sample testing.

The ICP-OES has a wide linear range [10]. Therefore, it is a fast analysis method with low-cost and high efficiency [11]. Thus it has been widely used in qualitative and quantitative analysis of minerals during geological exploration, such as rare poly-metallic ores, water sediments, soil, and feldspar [12].

When ICP-OES is applied for the quantitative analysis of uranium in mineral samples, the condition of RF power, atomizer flow rate, and sample pumping rate will affect the accuracy of results [13, 14]. In order to improve the accuracy of the ICP-OES for the uranium, this paper selects the optimal wavelength of the spectrum peaks after analyzing the spectral line of uranium; by adjusting the RF power, atomizer flow rate and sample pumping rate respectively, the standard series of uranium concentration gradients are measured. Then the results are compared and analyzed so that the three parameters are optimized. Apart from that, ICP-OES, ICP-MS, and volumetric methods are used to determine the references to verify the optimization effect.

2 Experimental

2.1 Reagents

Purified water was used for samples and standard solutions, which were ion-exchanged and re-distilled. Water (≥ 18.2 MΩ) was used in sample preparation.

The standard uranium solution was prepared from U3O8 a reference. The U3O8 reference with a purity higher than 99.99% was dried at 900 ℃ for 2 h. Then 1.1794 g of U3O8 was dissolved into 1000 mL water. The uranium concentration in the reserve liquid was 1000 µg/mL. The solution was diluted to 1 mg/L, 5 mg/L and 10 mg/L standard solution. The other reagents in the experiments were analytical pure reagent. The instrument used was an Optima 5300 DV inductively coupled plasma emission spectrometer made by PerkinElmer, U.S.A.

2.2 Sample preparation

  1. (1)

    A sample with 0.1000 g U3O8 was placed in a Poly tetra fluoroethylene (PTFE) crucible and wet with water. Then 1.0 mL of nitric acid, 2.0 mL of perchloric acid, and 10.0 mL of hydrofluoric acid were added into the PTFE crucible, after which was heated on a hot plate at 160–170 ℃.

  2. (2)

    When the sample produced a large amount of white perchlorate smoke, the temperature of the hot plate was raised so that the perchloric acid fully digest the organics in the samples. This step was not finished until no more perchlorate smoke was produced.

  3. (3)

    One mL of nitric acid and water was need to clean the sidewalls of crucible, after which the crucible was heated on a hot plate at low temperature to dissolve the sample. Two or four drops of hydrochloric acid were added and heating was continued. When the crucible was removed from the hot plate after wet salt had been formed.

  4. (4)

    Four mL of nitric acid and 10–15 mL of water were added to the sample, and a transparent solution was formed. If no insoluble matter was presented, 2–4 drops of hydrochloric acid were added and heating was continued until a transparent solution has been formed. Then solution was diluted to 100 mL.

2.3 Instrument and conditions

The instrument used a 5300DV inductively coupled plasma emission spectrometer (Perkin Elmer, U.S.A), the operation conditions were a radiofrequency (RF) power of 1200–1400 W, a working frequency of 40.68 MHz, a wavelength range of 165–782 nm, a wavelength resolution accuracy of ± 0.1 nm; the optical resolution was 0.006 nm (at 200 nm); the atomizer flow was 0.70–0.85 L/min, the auxiliary gas flow of 0.2 L/min, the plasma gas flow rate was 15 L/min, and the sample pumping rate was 1.2–1.8 mL/min. The reading delay time was 30 s, the cleaning time was 30 s, measurements were repeated for 3 times, and the plasma observation method was an axial observation.

3 Results and discussion

Three peak positions are needed in the ICP-OES analysis of uranium. The most intense peak is at 409.014 nm. The second peak is at 424.167 nm, and the last peak is at 367.007 nm, which is close to 367.003 nm (Fe), 367.006 nm (Th), and 367.021 nm (Ca). Fe is the major element in rock, and uranium is often accompanied by thorium. In this experiment, the wavelengths of 409.014 nm and 424.167 nm were chosen for the analysis.

As can be seen from the Fig. 1, in the experimental environment described above, when ICP-OES is used for analysis and measurement, the interference of adjacent spectral lines with the spectrum peak of the target element can be avoided, which makes it highly sensitive. According to the correlation coefficients of the recommended wavelength of the element to be measured, such as the SNR, spectral intensity, relative sensitivity, etc., provided by the line library of the inductively coupled plasma emission spectrometer, the optimal measuring wavelength was found to be 409.014 nm and 424.167 nm by linear regressions.

Fig. 1
figure 1

Three spectrum peaks at different concentrations

3.1 Optimization experiment of instrument parameters

Table 1 shows that the main parameters of general ICP-OES for uranium analysis are the RF power, atomizer flow rate, and sample pumping rate [14]. The proposed parameters for the experiments in this research is shown in Table 2.

Table 1 Main parameters of the ICP-OES for uranium analysis
Table 2 Proposed parameters of the ICP-OES for uranium analysis

In the experiment, standard solutions with uranium concentrations of 0 mg/L, 1 mg/L, 5 mg/L, and 10 mg/L were used to build a standard curve, and reference samples were used to verify the method. The experiment was performed using the nine groups of the parameters in Table 2. Each group of the experiments was repeated three times to determine the average of uranium concentration. The standard deviation analysis of the experimental results was carried out by measuring the relative deviation between the mean value and the standard value obtained using the volumetric method. The results were listed in Table 3.

Table 3 Results of references for different parameters

Table 3 shows that three parameters influenced the accuracy and those three parameters have impacts from high to low on the nebulizer flow, pump rate, and RF power. The RF power was fixed at 1400 W, and the flow rate of the atomizer and the sample pump rate were optimized.

The parameter that impacted the atomizer flow was the temperature of the central plasma channel, the residence time of the sample in the plasma channel, and the atomization efficiency of the solution. When the RF power was fixed at 1400 W a sample pumping rate of 1.5 mL/min and atomizer flow rates of 0.7 L/min, 0.75 L/min, 0.8 L/min, 0.85 L/min were selected for experiments. The results for the reference samples are shown in Table 4.

Table 4 Results of references at different atomizer flow rates

From Table 4, when the concentration in the references were 0.194% and 0.703%, the standard deviation at an atomizer flow rate of 0.75 L/min was the smallest. Therefore, the optimal atomizer flow rate in the experiments was 0.75 L/min.

The sample pumping rate mainly impacts the injection volume and the atomization efficiency of the solution. The magnitude of the injection volume further affects the intensity of the spectral line. For a fixed RF power of 1400 W and an atomization flow rate of 0.75 L/min, sample pumping rate of 1.4 mL/min, 1.5 mL/min and 1.6 mL/min were selected for optimization experiments. Table 5 showed the experimental results.

Table 5 Results of references at different sample pumping rates

From Table 5, that the concentrations in the reference were 0.194% and 0.703%. When analyzed, the GBW04113 the standard deviation at a sample pumping rate of 1.6 L/min was the smallest. When analyzed the GBW04114, standard deviation at three sample pumping rates was not very different. Therefore, the optimal sample pumping rate in the experiments was 1.6 L/min.

From the experiments, the optimal parameters for a uranium analysis using ICP-OES are the atomizer flow rate of 0.75 L/min, a sample pumping rate of 1.6 mL/min, and a RF power of 1400 W.

3.2 Verification of reference

The five references, from GBW04110 to GBW04114, were selected to verify the accuracy of the ICP-OES analyses using the optimal parameters as listed. Table 6 shows the results for the five references with standard values between 0.0285 and 0.703% were measured using ICP-OES. The standard deviation of the results is between 0.05 and 1.47%.

Table 6 ICP-OES experimental results

3.3 Verification of rock samples

Twenty-six rock samples with different uranium concentrations were selected for the experiments to verify the accuracy of the results obtained using ICP-OES at optimal parameters. The sample processing method described above was used to process the samples, then ICP-OES, ICP-MS, and the volumetric method were used to analyze sample solutions.

As shown in Fig. 2, 26 samples with uranium concentrations between 0.02 and 0.26% were analyzed by using ICP-OES and ICP-MS. The standard deviations of the two results were between 0.00 and 1.61%. The data shows that the results obtained using ICP-OES are similar to those obtained using ICP-MS when analyzing uranium samples.

Fig. 2
figure 2

ICP-OES and ICP-MS experimental results

Figure 3 shows the results for 26 samples with uranium concentrations between 0.02 and 0.271% are obtained by using ICP-OES and the volumetric method. The standard deviation of the two sets of results was between 0.00 and 1.27%. The results obtained using ICP-OES can be seen to be similar to those obtained using the volumetric method when analyzing uranium samples.

Fig. 3
figure 3

Experimental results from the ICP-OES and from the volumetric method

3.4 Spike recovery of adding standard

Six different samples with uranium concentrations between 350 and 1900 mg /kg were selected. 0. Samples of 0.1000 g were accurately weighed, pretreatment and standardization experiments were performed, and the volume was set to 100 mL. The spike recovery was calculated based on the measurement results and concentrations. The results are shown in Table 7.

Table 7 Sample added recovery

4 Conclusion

In this paper, we discussed the uranium concentrations of the samples analyzed by using the ICP-OES method and found that the spectrum peaks at 409.014 nm and 424.167 nm could be used for that purpose.

We tested the RF power, atomization flow rate, and sample pumping rate by using a control variate method and found the following optimal parameters an atomizer flow rate of 0.75 L/min, a sample pumping rate of 1.6 mL/min, and a high-frequency power of 1400 W. Based on the results from the experimental samples and the references, we compared the accuracy of results achieved by using a volumetric method, ICP-MS method, and ICP-OES method with optimized parameters. The results show that convenient, quick, and effective quantitative analysis of uranium concentrations in ores.