Abstract
Cardiac radioablation is a promising treatment for cardiac arrhythmias, but accurate dose delivery can be affected by heart motion. For this reason, real-time cardiac motion monitoring during radioablation is of paramount importance. Real-time ultrasound (US) guidance can be a solution. The US-guided cardiac radioablation workflow can be simplified by the simultaneous US and planning computed tomography (CT) acquisition, which can result in US transducer-induced metal artifacts on the planning CT scans. To reduce the impact of these artifacts, a new metal artifact reduction (MAR) algorithm (named: Combined Clustered Scan-based MAR [CCS-MAR]) has been developed and compared with iMAR (Siemens), O-MAR (Philips) and MDT (ReVision Radiology) algorithms. CCS-MAR is a fully automated sinogram inpainting-based MAR algorithm, which uses a two-stage correction process based on a normalized MAR method. The second stage aims to correct errors remaining from the first stage to create an artifact-free combined clustered scan for the process of metal artifact reduction. To evaluate the robustness of CCS-MAR, conventional CT scans and/or dual-energy CT scans from three anthropomorphic phantoms and transducers with different sizes were used. The performance of CCS-MAR for metal artifact reduction was compared with other algorithms through visual comparison, image quality metrics analysis, and HU value restoration evaluation. The results of this study show that CCS-MAR effectively reduced the US transducer-induced metal artifacts and that it improved HU value accuracy more or comparably to other MAR algorithms. These promising results justify future research into US transducer-induced metal artifact reduction for the US-guided cardiac radioablation purposes.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Cardiac arrhythmias are a major global health concern and it has been predicted that atrial fibrillation, which is the most common type of cardiac arrhythmia, will affect 6–12 million people in the USA by 2050 and 17.9 million people in Europe by 2060 [1]. Among currently available curative treatment methods for cardiac arrhythmias, catheter ablation is the most widely used [2]. However, catheter ablation is an invasive procedure which can have a long treatment time, and its success rate can be low due to the inaccessibility of some arrhythmogenic tissues [3].
Cardiac radioablation is a promising treatment method that could potentially overcome the limitations faced by catheter ablation. The primary aim of cardiac radioablation is to treat arrhythmias by noninvasively delivering a radiation dose to the arrhythmogenic tissues using external beam radiation therapy. Among the external-beam radiation therapy, photon and proton beams are investigated for cardiac radioablation in pre-clinical [4,5,6,7,8,9] and clinical [10,11,12,13,14,15,16] research studies.
A cardiac radioablation workflow typically consists of a simulation stage, where the treatment is planned, and of a treatment delivery stage. At the simulation stage, planning computed tomography (CT) scans are acquired and subsequently used to delineate the treatment target (arrhythmogenic tissue), and organs-at-risk (OARs). Then, the electron density information derived from the CT Hounsfield units (HU) is used for the calculation of the radiation dose distributions inside the delineated structures [17]. At the treatment delivery stage, the external beam delivers the treatment dose to the arrhythmogenic tissue, while sparing the OARs as much as possible. However, intra-fractional motion which occurs during the treatment delivery, due to heartbeat and respiration may negatively impact the accuracy of the dose delivery [18, 19]. This makes real-time monitoring of both cardiac and respiratory motion during the treatment of paramount importance.
Ultrasound (US) imaging is a non-ionizing radiation based real-time imaging modality which is clinically used for treatment guidance in radiation therapy of oncological targets [20,21,22]. US-based treatment guidance is currently being investigated for cardiac radioablation, where it has potential for dose delivery accuracy improvement [23,24,25]. This type of image guidance relies on US imaging of the cardiac tissue position during the radiation dose delivery in order to compare this position with the corresponding position at the simulation stage based on which the treatment was planned. To reduce the number of workflow steps and the time needed during the simulation stage, the US scan could be acquired simultaneously with the planning CT scans. The presence of the US transducer when the CT scan is acquired, though, can result in transducer-induced metal artifacts on the planning CT scans, which are caused by metal components located inside the US transducer [26, 27].
To reduce the negative impacts of this type of artifacts induced by metal, Metal Artifact Reduction (MAR) algorithms have been investigated and/or proposed in the literature, mainly focusing on the artifacts generated by implanted metal structures. Some of these algorithms are commercially available, and some are only research-based [28]. Many of these algorithms use the principle of sinogram inpainting technique [29]. Sinograms, or in other words projection data, are acquired during the CT scan acquisition and consist of the attenuation profiles of X-ray respective to an angle of the X-ray beam. During the sinogram inpainting technique, the projection data which are affected by metal components are treated as missing values. This technique uses interpolation methods, typically linear interpolation, to “paint” the missing values from surrogate data. When surrogate data are near to metal components or consist of high contrast structures, especially bone, the direct interpolation of a sinogram tends to induce new artifacts [30]. To minimize or avoid the creation of new artifacts, a normalization step is introduced during the sinogram interpolation in the normalized MAR (NMAR) method [31]. NMAR requires information on the regions corrupted by metal artifacts for the artifact reduction. For this, tissue types in the CT scan which are corrupted by metal artifacts (CTart) are classified into different clusters to create a clustered scan. The clustered scan has similar content as the CTart scan, but it has less or even no metal artifacts. After generation of the clustered scan, it is forward projected to the sinogram space and used to normalize the sinogram of the CTart scan for the interpolation of missing values. The performance of the NMAR method depends on the quality of clustered scan, which is often affected by the number of metal artifacts.
In general, the application of a MAR algorithm on a CTart scan results in an artifact-corrected CT (CTcor) scan. The commercially available MAR algorithms of Orthopaedics Metal Artifact Reduction (O-MAR, Philips Health System, Cleveland, USA) and iterative Metal Artifact Reduction (iMAR, Siemens Healthcare, Forchheim, Germany) were evaluated for photon and proton therapy applications [32,33,34,35,36,37,38,39,40]. These phantom and clinical studies revealed that the O-MAR and iMAR algorithms could be beneficial for the improvement of therapy treatment planning either in terms of image quality or dosimetric outcome. Research-based MAR algorithms have also been evaluated for radiotherapy applications. In particular the metal deletion technique (MDT) [41], MAR with hardware adaptation [42], and ker MAR [43] have been investigated. It was shown that these algorithms can be used to reduce the errors on photon and proton range calculation during treatment planning [42, 44, 45].
Another possible approach for metal artifact reduction is through the acquisition of a dual-energy CT (DECT) scan instead of a single energy scan. In this method, two discrete energy beams (typically 90 kVp and 140 kVp) are used rather than a polychromatic X-ray beam (average energy of 120 kVp) of a conventional single energy CT (SECT) [46, 47]. The utilization of a discrete high energy beam reduces the beam-hardening effect [48], which is one of the factors that contributes to the creation of metal artifacts on CT scans [49]. In literature, the application of a commercial MAR algorithm, especially iMAR, on both SECT scans and DECT scans was evaluated for metal artifact reduction. Generally, the application of a MAR algorithm on DECT scans better reduced the artifacts than the application of these algorithms on SECT scans [50, 51].
To the best of our knowledge, none of the studies in literature proposed or investigated MAR algorithms for the reduction of US transducer-induced metal artifacts on planning CT scans. Therefore, the aim of this work was to develop a specialized automatic MAR algorithm which can be used to correct these US transducer-induced metal artifacts on SECT and DECT scans. As O-MAR, iMAR and MDT have been widely investigated for photon and proton therapy applications in the literature, the performance of the proposed MAR algorithm has been compared with the performance of these commercial and research-based MAR algorithms.
Materials and methods
CCS-MAR algorithm design
CCS-MAR is a sinogram inpainting-based MAR algorithm which combines image processing strategies with the NMAR method [31] to replace artifact corrupted areas in CT scans. The NMAR method requires a CT scan which is clustered according to tissue types as input. The clustering approach utilized in this work was inspired by the studies conducted by Wu et al. [52] and Luzhbin and Wu [53]. These studies utilized the k-means clustering algorithm [54], which is an unsupervised iterative method being used for classification tasks, for the creation of clustered CT scans. As the performance of the NMAR method depends on the quality of the clustered CT scan, the creation of the clustered scan without metal artifacts is crucial. Therefore, the key step of the CCS-MAR algorithm is the creation of an artifact-free clustered CT scan from the CTart scan.
The CCS-MAR algorithm was implemented using MatLab (version 9.7, The MathWorks Inc, Natick, MA, USA), and its workflow is shown in Fig. 1. It is a fully automatic algorithm which does not require any manual contouring of artifact-corrupted areas. CCS-MAR utilizes a two-stage correction process for metal artifact reduction. The first stage of correction process and the second stage correction process is indicated using black, and dotted black arrows in Fig. 1, respectively. The input of the algorithm is the planning CTart scan, which will be referred to as ‘original’ scan in the rest of this section. During the first correction process of the algorithm’s workflow, HU value thresholds of 2000 HU and -950 HU are used to identify and segment the metal component (see Fig. 1) and air region, respectively. A fixed threshold value of 2000 HU has been chosen for the metal segmentation based on previous studies published in the literature [55, 56]. However, a slight change in the HU value threshold from 2000 to 2500 HU for the metal segmentation also has been checked. In order to create an initial clustered scan, the k-means clustering algorithm is applied to the original scan after the metal and air segmentation. As the original scan is acquired of the thoracic region, this scan primarily consists of soft tissue, bone, and lung tissue. By considering these three tissue types, a total of three clusters are chosen during the k-means clustering. After the clustering, the segmented air region from the first step is added to the clustered scan. Subsequently, the original scan, the metal component, and the clustered scan are forward projected to generate their respective sinograms.
The metal sinogram is then modified through the application of both dilation with a disk structuring element [57] and smoothing through Gaussian filtering. Since the exact HU value of the metal component in the transducer is unknown, dilation is applied to inflate the metal data to minimize the dependency on the 2000 HU value threshold. The smoothing is applied to remove statistical photon noise which may cause streak artifacts during the reconstruction of CT scans.
In the next step of the algorithm, the NMAR method [31] is applied to the created sinograms (see the blue box in Fig. 1). The main steps of NMAR are normalization, interpolation, and denormalization of the sinograms. In the normalization step, the original sinogram is divided by the clustered sinogram and a normalized sinogram is created. Then, the dilated and smoothed metal sinogram is used as a mask to identify the metal region in the normalized sinogram for the linear interpolation. Subsequently, the interpolated normalized sinogram is multiplied by the clustered sinogram during the denormalization step resulting in a corrected sinogram. Filtered back projection is finally used to reconstruct the first corrected CT scan from the corrected sinogram.
Even though most of the artifacts are reduced in the first corrected scan, typically the intense dark streaks which are in the region close to the transducer remain unchanged. Also, the HU value of this region becomes lower than the HU value of the corresponding region on the original scan, because the intense dark streaks are modelled as lung tissue in the clustered scan. Therefore, further correction is required, for this, the second stage which is the final correction process is incorporated to reduce the errors in the first corrected scan. In the beginning of the second correction process, the pixel-wise absolute difference between the original scan and the first corrected scan is calculated. An empirically chosen threshold value of 200 HU is used to identify the pixels which are in the region with the dark streaks. Subsequently, those pixels are identified on the first corrected scan, and they are replaced with a 0 HU value. A value of the threshold in the range of 150 HU to 200 HU is suitable for the HU value replacement. Indeed, it was found that the threshold value below 150 HU will replace the HU values of pixels which do not have metal artifacts on the first corrected scan. This contributes to the generation of an inaccurate second clustered scan. On the other hand, a value above 200 HU will not contribute to the artifact-free combined clustered scan. The region with intense dark streaks may cause a wrong classification during k-means clustering. Thus, spatial relationships [52] are incorporated with k-means clustering, calculated as:
where \(P\) denotes a set of pixels covered by a 3 × 3 mask centred around pixel \(i\), and \(|.|\) is the number of members in the set after the k-means clustering. \({C}_{j}\) and \(k\) are the \(j\)th cluster and the total number of clusters, respectively. During this procedure, pixel \(i\) is reassigned to the cluster \({C}_{i}\) that has the maximum number of members within the mask. With this method, the clustered scan with spatial relationships is created, and then the air region which was segmented in the first step is added. For the mask size, the minimum size of 3 × 3 produced the best result for incorporating spatial relationships. On the other hand, the mask with a larger size smeared the clustered scan, which would then produce the second corrected scan with induced secondary artifacts. Even though incorporating spatial relationships allows to reduce the inaccuracies in clustering, especially in the artifact regions, it may also induce errors in pixel reassignment in the artifact-free regions. For this reason, during the initial k-means clustering, the spatial relationships are not incorporated.
In the next step, the clustered scans which are created with and without incorporating the spatial relationship are combined to generate the artifact-free combined clustered scan. For the combination, the absolute difference between those scans is calculated and the difference is added to the initial clustered scan to reduce the intense dark streaks in the region close to the US transducer. Afterwards, the combined clustered scan is forward projected to generate the combined clustered sinogram. Utilizing the combined clustered sinogram, the original sinogram, and the smoothed metal sinogram, the second corrected scan is reconstructed using filtered back projection after the NMAR method. The dilation of the metal sinogram may cause over interpolation or blurring especially in the region which is adjacent to the US transducer in the second corrected scan. In order to balance this effect, during the creation of second corrected scan, only smoothing and no dilation is applied to the metal sinogram. The corrected scan is an average of the first and second corrected scans. To preserve the noise texture, the result from the application of a high-pass filter on the original scan is added to the corrected scan for the creation of the final corrected scan (CTcor scan).
CT scan data collection for performance evaluation
CT scanning and acquisition parameters are summarized in Table 1. In total, three types of adult anthropomorphic phantoms were used: a Model ART-211 male phantom (ART, Radiology Support Devices, Long Beach, CA, USA), an ATOM® male phantom (CIRS, Model-701, Norfolk, VA, USA), and a CT torso phantom (CT Torso, Model CTU-41, Kyoto Kagaku Ltd, Japan). The phantoms were positioned on the CT table in a head-first position (Fig. 2) and scanned first without, and then with US transducers in place. As can be seen in Table 1, three different CT scanners from Philips and Siemens were used for CT scanning.
The size of an US transducer influences the amount and the appearance of metal artifacts on CT scans [26]. To investigate this effect, a total of three US transducers with various sizes were used: (1) single-plane phased array: Telemed P5-1S15-A6 from Telemed (UAB, Vilnius, Lithuania); (2) bi-plane phased array: Terason XY mini from Teratech Corporation (Burlington, MA, USA); and (3) linear volumetric array: Philips VL13-5 from Philips Healthcare (Andover, MA, USA). The maximum width of the metal part inside the transducers as measured on CT scans of above transducers were 4 cm, 2 cm, and 6 cm, respectively. To fix the US transducer on the phantoms, custom-built transducer holders (consisting of 3D printed plastic parts and elastic straps) which were developed in collaboration with Usono (Eindhoven, the Netherlands) were used.
For each phantom, first a CT scan was acquired without the transducer on the phantom, which resulted in a reference artifact-free CT (CTref) scan. Then, without changing the position of the phantom, an US transducer was placed into the transducer holder and a CT scan with the US transducer-induced metal artifacts (CTart) was acquired. The transducers were placed in a position considered to be suitable for imaging the heart. For all CT scans, a CT thorax protocol was utilized, and the scanning parameters are given in Table 1.
The commercial MAR algorithms iMAR and O-MAR were applied during the CT scan reconstruction on the scanner, as these algorithms are available on Siemens and Philips scanners, respectively. iMAR was not applied to the Siemens PET-CT scans, because it was not available on this particular scanner. In addition to the application of these commercial MAR algorithms, MDT and CCS-MAR algorithms were also applied to all CTart scans.
Image quality metrics analysis
In order to evaluate the effectiveness of a MAR algorithm for metal artifact reduction, image quality metrics including, structural similarity (SSIM) index, root mean square error (RMSE) of the HU values, and peak signal-to-noise ratio (PSNR) [58, 59] were calculated between the CTref scan and corresponding CTart and CTcor scans. To not bias these image quality metric results, pixels containing the details of the US transducer on CTart scans were copied and transferred to the corresponding CTref scans. Higher SSIM which ranges from 0 to 1, and higher values of PSNR indicate better image quality. On the other hand, the calculated RMSE values should be low for better artifact reduction. In the end, mean values of these image quality metrics calculated for all the CTart and CTcor scans were used for analysis.
HU value restoration evaluation
Quantitative comparisons were performed to evaluate HU values on the CTart and CTcor scans in comparison with the CTref scan. Initially, through visual inspection, axial CT slices, which were affected by metal artifacts were chosen from each CTart scan. Then, the corresponding axial CT slices were selected from the CTref and CTcor scans. Regions of interest (ROI) were defined as elliptical areas of ≈ 0.5 cm2 and the ROI-based HU values of pixels were extracted from the regions of the heart, lung, and bone from those selected CT slices. Figures 3, 4 and 5 shows the ROI placements. A paired sample t-test was used to compare the ROI-based HU value measurements of pixels on both CTart and CTcor with CTref slice at a significance level of 0.05.
Results
Visual comparison of the CT scans
Single-plane phased array transducer
Figure 3 shows an example of 2D slices from the SECT (a) and DECT (b) scans of the ART phantom, and SECT scans of the ATOM® phantom (c) with and without the single-plane phased array transducer in place. Overall, the application of MAR algorithms reduced the metal artifacts on the corresponding CTcor scans. However, residual artifacts were observed on CTcor scans after the application of iMAR and O-MAR (see red arrows in Fig. 3). In addition, a modification in the edges of soft tissues and bone can be seen on the corresponding absolute differences of the CTcor scan after the application of MDT and CCS-MAR algorithms (see blue arrows in Fig. 3).
Bi-plane phased array transducer
For the bi-plane phased array transducer, the application of MAR algorithms on SECT and DECT scans of the ART phantom, and SECT scans of the ATOM® phantom are shown in Fig. 4. Due to its smaller size, the bi-plane phased array transducer produced fewer metal artifacts than the single-plane phased array transducer (see previous section). In addition, CCS-MAR better preserved the bone details than the iMAR and MDT algorithms (see red arrows in ‘b’ Fig. 4). However, the application of CCS-MAR and MDT and algorithms slightly modified the edges of soft tissues and bones (see blue arrows in ‘c’ Fig. 4).
Linear volume array transducer
Figure 5 shows the CT scans of the torso phantom with and without the volume array transducer. A commercial MAR algorithm was not applied to reduce the metal artifacts on this torso phantom scans since it was not available on the particular CT scanner used for CT scanning.
Image quality metrics
Single-plane phased array transducer
For the single-plane phased array transducer, the image quality metrics SSIM, RMSE and PSNR were calculated for the CT scans of the ART and the ATOM® phantom (see Table 2). Overall, the application of MAR algorithms on the CTart scans improved SSIM and PSNR, while reducing the RMSE values. This shows that the metal artifact reduction had the desired effect on the improvement of image quality and HU values,
For the CT scans of ART phantom and ATOM® phantom, the calculated mean SSIM values were well improved after the application of CCS-MAR algorithm. The mean RMSE values for the CTart scans of the SECT and DECT scans of ART phantom, and the ATOM® phantom scans were 87.94, 85.27, and 112.90, respectively, and these values were reduced to the lowest after the application of CCS-MAR (31.21) and iMAR (32.06) and the MDT (57.62) algorithms, respectively. The highest mean PSNR (dB) values were observed for the CTcor scans after the application of CCS-MAR for both SECT and DECT scans of the Atom phantoms.
Bi-plane phased array transducer
Table 3. shows the image quality metrics for the ART and the ATOM® phantom scans with the bi-plane phased array transducer. On the ATOM® phantom scans, the CTcor after the application of CCS-MAR (29.96) had lower mean RMSE values than the CTcor scans after the application of O-MAR (42.52) and MDT (32.33) algorithms. For the same phantom, the mean PSNR (dB) value for the CTart scan was 19.02 and it improved to 29.02, 31.03 and 31.38 after the application of O-MAR, MDT, and CCS-MAR algorithms, respectively.
Linear volume array transducer
The calculated image quality metrics for the torso phantom scans with the volume array transducer in place are shown in Table 4. The application of the CCS-MAR on the CTart scans improved the mean SSIM and PSNR values better than the MDT algorithm while reducing the mean RMSE values. The mean RMSE value for the CTart scan was 114.12 and it was reduced to 63.72 and 55.67 after the application of MDT and CCS-MAR algorithms, respectively.
HU value restoration evaluation
Single-plane phased array transducer
The ability of a MAR algorithm to restore the HU value on the ART and ATOM® phantom scans is given in Table 5. Even though the MAR algorithms improved the ROI-based mean HU value measurements on the phantom scans, significant differences (p < 0.05) were still identified, especially in ROI 1 and ROI 3 between the CTcor scan and the CTref scan.
In the case of the ATOM® phantom scans, for the ROI 1, the application of CCS-MAR algorithm better reduced the difference between the HU value measurement on CTref and CTcor scans. For these scans, the measured HU value (mean ± STD) on CTart scan was -179.40 ± 77 HU and it improved to -63.20 ± 56 HU, -54.04 ± 35 HU, and -49.37 ± 34 HU on the CTcor scan after the application of O-MAR, MDT and CCS-MAR, respectively, while it was 25.08 ± 16 HU on the corresponding CTref scan. The analysis of HU in the ROI 3 depicted that the application of MDT on both the SECT and DECT scans of the ART phantom, and the application of CCS-MAR on the ATOM® phantom scans performed well in terms of HU value restoration.
Bi-plane phased array transducer
Table 6 summarises the comparison of the ROI-based mean HU values of the ART and the ATOM® phantom scans with and without the bi-plane phased array transducer. In this study, ROI 1 was placed in the heart region and therefore it was assumed to be the target. On the other hand, ROI 2 and ROI 3 were placed in lung and bone regions, respectively. For all ROI measurements on the SECT and DECT scans of the ART phantom, the MDT and CCS-MAR algorithms brought the mean HU value closer (p > 0.05) to the mean HU values on the CTref scan. In ROI 1 on ATOM® phantom scans, significant differences (p < 0.05) were identified on the HU value measurements on CTcor after the application of the O-MAR algorithm (5.04 ± 28 HU, [mean ± STD]), compared to the CTref scan (25.94 ± 16 HU, [mean ± STD]).
Linear volume array transducer
The comparison of ROI-based HU values on the CT scans of the torso phantom with and without the linear volume array transducer is given in the Table 7. Even though MDT and CCS-MAR improved the ROI-based mean HU value on the CTcor scan compared to the CTart scan, significant differences (p < 0.05) in HU value measurements were still observed between the CTcor and CTref scans, especially in ROI 1 and ROI 3. However, in ROI 2, the mean HU value (mean ± STD) was improved after the application of MDT (-927.43 ± 120 HU) and CCS-MAR (-928.09 ± 116 HU), and significant differences were not identified with the measurement on the CTref scan (-926.02 ± 119 HU).
Overall, the application of the CCS-MAR algorithm on the CTart scans effectively reduced the metal artifacts and improved the SSIM and PSNR values for all phantoms scans. Also, CCS-MAR better improved the HU values compared to the other MAR algorithms in most of the scenarios. HU value threshold for the metal segmentation is crucial for the performance of a MAR algorithm. For CCS-MAR, it has been verified that slight changes (range 2000–2500 HU) in the HU threshold value did not affect its performance for the metal artifact reduction. Since CCS-MAR does not utilize many iterations during the artifact reduction, for the investigated CT scans which included severe metal artifacts, the algorithm written in MatLab took an average of 1.5 min per CT slice on a i7-8665U, 2.11 GHz CPU. Therefore, for an average of 40 CTart slices, CCS-MAR is estimated to take about an hour to perform the artifact reduction task on a standard personal computer.
Discussion
In this work, a fully automated algorithm (CCS-MAR) was proposed which can potentially be used to reduce the negative impact of metal artifacts appearing on planning CT scans used during US-guided cardiac radioablation. Furthermore, the performance of the CCS-MAR algorithm for metal artifact reduction and HU value restoration has been compared to other commonly used commercial and research-based MAR algorithms.
In order to check the robustness of the CCS-MAR algorithm, this study utilized CT scans of different anthropomorphic phantoms. These phantoms were scanned with multiple types of CT scanners while US transducers with various sizes were or were not in place during the acquisition. The presented results show that, the CCS-MAR algorithm effectively improved the image quality metrics SSIM and PSNR and that it improved the ROI-based HU values accuracy more or comparably to other MAR algorithms. Also, the application of CCS-MAR induced fewer secondary metal artifacts on CTcor scan than the application of O-MAR, iMAR and MDT (see Figs. 3, 4). However, slight modification of the edges of the bone and soft tissues were identified on CTcor scan after the application of the CCS-MAR algorithm (see Figs. 3, 4, and 5). CCS-MAR utilizes a CTart scan to generate the original sinogram for the metal artifact reduction, while commercial MAR algorithms typically use the original sinogram directly from the CT scanner. This is a possible reason for these edge modifications.
The mean absolute differences of torso phantom scans show notable differences (Fig. 5), especially in the edges of the bone and soft tissues. These differences might be due to the large number of metal artifacts produced from the linear volume array transducer, or phantom movement between the acquisition of CTref and CTart scans. A lot of effort was put into not moving the phantom during CT scan acquisition; however, the occurrence of a small motion cannot be excluded. It has to be noted that the same phenomena were not observed or to a lesser extent for the other phantoms scans (Figs. 3, 4). Therefore, we hypothesize that it is more likely that the differences originate from the large number of artifacts. In order to draw final conclusion, additional evaluation is required.
It has been shown in literature that the size of a US transducer influences the creation of the metal artifacts [26]. In our case, the CT scans with the single-plane phased array transducer had higher calculated RMSE than the bi-plane phased array transducer. This strengthens the recommendation to use the smallest US transducer possible and subsequently apply an MAR algorithm to reduce the artifacts.
Generally, the application of the MAR algorithms on the DECT scans of the ART phantom with the bi-plane phased array transducer better reduced the RMSE value and better restored the HU value than the application of MAR algorithms on SECT scans of the ART phantom. These results align with the work conducted by Kovacs et al. [51] in which they reported that the application of MAR algorithms on DECT scans is the preferred choice to reduce the metal artifacts, especially for the metal artifacts created by dental, and hip implants, for radiotherapy applications. Thus, to further reduce the metal artifacts which are created by the smallest US transducer, the acquisition of DECT scan is recommended over the acquisition of a SECT scan.
This work has a few limitations which are related to the use of CT phantoms for the MAR algorithm evaluation. These phantoms are just simulations of human anatomies and therefore do not contain structures which are specific for cardiology patients, such as pacemakers and cardioverter defibrillators (ICDs). These implantable cardiac electronic devices often consist of metal parts, and therefore they also can produce metal artifacts on CT scans. This study focused solely on the artifacts produced by the US transducer and did not take into account the metal artifacts produced by the implantable cardiac devices itself. In addition, the phantoms used in this work could not be imaged with US. For this reason, the positioning location and angle of the US transducers on the phantoms was chosen based on educated guess. In order to draw final conclusions regarding the performance of the proposed MAR algorithm, further evaluation of the algorithm using clinical data is required.
This study primarily investigated the impacts of MAR algorithms on HU value restoration on planning CT scans affected by metal artifacts. In addition to HU value restoration, the dosimetric impact of metal artifact reduction on the target (arrhythmogenic tissue) and OARs during the treatment planning is also crucial. Future work may therefore include an evaluation of the dosimetric impacts of the CCS-MAR algorithm.
Cone-Beam Computed Tomography (CBCT) is often used in radiation therapy, especially in the image-guided radiation therapy for patient positioning prior to the treatment [60]. This positioning is performed by matching the CBCT scans to the planning CT scan. Instead, or in addition to simultaneous acquisition of the planning CT scan and US images, simultaneous acquisition of CBCT scans and US images may be another way to implement US-based image guidance for cardiac radioablation. Therefore, the performance of the CCS-MAR algorithm should also be evaluated for transducer-induced metal artifact reduction on CBCT scans.
The application of deep learning for the metal artifact reduction on CT scans for the radiotherapy applications gained significant interest and has been actively studied in recent years [61,62,63,64]. These studies have shown promising results for the metal artifact reduction and claimed that the deep learning algorithms are powerful in handling the complex metal artifacts patterns, while preserving the anatomical features. Further, it was found that these algorithms often work on the image domain and thus create fewer number of secondary artifact and/or image blurring. Therefore, the potential application of deep learning for the reduction of the US transducer induced metal artifacts on CT scans for radiation therapy applications may be beneficial and worth exploring.
Conclusion
The purpose of this work was to propose and evaluate an MAR algorithm which can be used to reduce metal artifacts resulting from the presence of a US transducer during CT scan acquisition. Visual comparison of CT scans, analysis of image quality metrics, and evaluation of the HU value restoration have shown that the proposed MAR algorithm, CCS-MAR, effectively reduces the negative impacts of US transducer-induced metal artifacts on planning CT scans of CT phantoms. These promising results justify further research into US transducer-induced MAR on CBCT scans as well as extensive (dosimetric) evaluations using clinical CT scans.
References
Lloyd-Jones D, Adams RJ, Brown TM, Carnethon M, Dai S, De Simone G et al (2010) Executive summary: heart disease and stroke statistics–2010 update: a report from the American Heart Association. Circulation 121(7):948–954
Haegeli LM, Calkins H (2014) Catheter ablation of atrial fibrillation: an update. Eur Heart J 35(36):2454–2459. https://doi.org/10.1093/eurheartj/ehu291
Cronin EM, Bogun FM, Maury P, Peichl P, Chen M, Namboodiri N et al (2020) 2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias. Hear Rhythm 17(1):e2–e154. https://doi.org/10.1016/j.hrthm.2019.03.002
Suzuki A, Deisher AJ, Rettmann ME, Lehmann HI, Hohmann S, Wang S et al (2020) Catheter-free arrhythmia ablation using scanned proton beams: electrophysiologic outcomes, biophysics, and characterization of lesion formation in a porcine model. Circ Arrhythm Electrophysiol 13(10):e008838
Hohmann S, Deisher AJ, Suzuki A, Konishi H, Rettmann ME, Merrell KW et al (2019) Left ventricular function after noninvasive cardiac ablation using proton beam therapy in a porcine model. Hear Rhythm 16(11):1710–1719
Richter D, Lehmann HI, Eichhorn A, Constantinescu AM, Kaderka R, Prall M et al (2017) ECG-based 4D-dose reconstruction of cardiac arrhythmia ablation with carbon ion beams: application in a porcine model. Phys Med Biol 62(17):6869–6883. https://doi.org/10.1088/1361-6560/aa7b67
Lehmann HI, Deisher AJ, Takami M, Kruse JJ, Song L, Anderson SE et al (2017) External arrhythmia ablation using photon beams. Circ Arrhythmia Electrophysiol 10(4):1–8
Sharma A, Wong D, Weidlich G, Fogarty T, Jack A, Sumanaweera T et al (2010) Noninvasive stereotactic radiosurgery (CyberHeart) for creation of ablation lesions in the atrium. Hear Rhythm 7(6):802–810. https://doi.org/10.1016/j.hrthm.2010.02.010
Zei PC, Wong D, Gardner E, Fogarty T, Maguire P (2018) Safety and efficacy of stereotactic radioablation targeting pulmonary vein tissues in an experimental model. Hear Rhythm 15(9):1420–1427
Widesott L, Dionisi F, Fracchiolla F, Tommasino F, Centonze M, Amichetti M et al (2020) Proton or photon radiosurgery for cardiac ablation of ventricular tachycardia? Breath and ECG gated robust optimization. Phys Medica 78:15–31
Goddu SM, Hilliard J, Knutson N, Zhao T, Hugo GD, Mutic S et al (2018) Feasibility of noninvasive cardiac ablation utilizing intensity modulated proton therapy to treat ventricular tachycardia. Int J Radiat Oncol 102(3):S58. https://doi.org/10.1016/j.ijrobp.2018.06.166
Neuwirth R, Cvek J, Knybel L, Jiravsky O, Molenda L, Kodaj M et al (2019) Stereotactic radiosurgery for ablation of ventricular tachycardia. Europace 21(7):1088–1095
Gianni C, Rivera D, Burkhardt JD, Pollard B, Gardner E, Maguire P et al (2020) Stereotactic arrhythmia radioablation for refractory scar-related ventricular tachycardia. Hear Rhythm 1–17.
Qian PC, Azpiri JR, Assad J, Gonzales Aceves EN, Cardona Ibarra CE, de la Pena C et al (2020) Noninvasive stereotactic radioablation for the treatment of atrial fibrillation: first-in-man experience. J Arrhythmia 36(1):67–74
Cuculich PS, Schill MR, Kashani R, Mutic S, Lang A, Cooper D et al (2017) Noninvasive cardiac radiation for ablation of ventricular tachycardia. N Engl J Med 377(24):2325–2336
Bhaskaran A, Downar E, Chauhan VS, Lindsay P, Nair K, Ha A et al (2019) Electroanatomical mapping–guided stereotactic radiotherapy for right ventricular tachycardia storm. Hear Case Rep 5(12):590–592. https://doi.org/10.1016/j.hrcr.2019.09.007
Vyfhuis MAL, Onyeuku N, Diwanji T, Mossahebi S, Amin NP, Badiyan SN et al (2018) Advances in proton therapy in lung cancer. Ther Adv Respir Dis 12:1–16
Guzhva L, Flampouri S, Mendenhall NP, Morris CG, Hoppe BS (2019) Intrafractional displacement of cardiac substructures among patients with mediastinal lymphoma or lung cancer. Adv Radiat Oncol 4(3):500–506
Dusi V, Russo G, Forte GI, De Ferrari GM (2020) Non-invasive ablation of cardiac arrhythmia. Is proton radiation therapy a step forward? Int J Cardiol 313:64–66
Fontanarosa D, Van Der Meer S, Bamber J, Harris E, O’Shea T, Verhaegen F (2015) Review of ultrasound image guidance in external beam radiotherapy: I. Treatment planning and inter-fraction motion management. Phys Med Biol 60(3):R77-114
Camps SM, Fontanarosa D, de With PHN, Verhaegen F, Vanneste BGL (2018) The Use of ultrasound imaging in the external beam radiotherapy workflow of prostate cancer patients. Biomed Res Int 2018:7569590
O’Shea T, Bamber J, Fontanarosa D, Van Der Meer S, Verhaegen F, Harris E (2016) Review of ultrasound image guidance in external beam radiotherapy part II: Intra-fraction motion management and novel applications. Phys Med Biol 61(8):R90-137. https://doi.org/10.1088/0031-9155/61/8/R90
Blanck O, Ipsen S, Chan MK, Bauer R, Kerl M, Jacobi V, et al (2016) Treatment planning considerations for robotic guided cardiac radiosurgery for atrial fibrillation. Cureus 8(7).
Casula M, Dusi V, Camps S, Gringet J, Benoit T, Garonna A et al (2022) Feasibility of an automatic ultrasonographic image acquisition system associated with an artificial intelligence algorithm for real-time monitoring of cardiac motion during cardiac radio-ablation. Front Cardiovasc Med. https://doi.org/10.3389/fcvm.2022.849234
Perrin R, Maguire P, Garonna A, Weidlich G, Bulling S, Fargier-Voiron M et al (2022) Case report: treatment planning study to demonstrate feasibility of transthoracic ultrasound guidance to facilitate ventricular tachycardia ablation with protons. Front Cardiovasc Med. https://doi.org/10.3389/fcvm.2022.849247
Ipsen S, Bruder R, García-Vázquez V, Schweikard A, Ernst F (2019) Assessment of 4D ultrasound systems for image-guided radiation therapy - image quality, framerates and CT artifacts. Curr Dir Biomed Eng 5(1):245–248
Schlosser J, Hristov D (2016) Radiolucent 4D ultrasound imaging: system design and application to radiotherapy guidance. IEEE Trans Med Imaging 35(10):2292–2300
Puvanasunthararajah S, Fontanarosa D, Wille ML, Camps SM (2021) The application of metal artifact reduction methods on computed tomography scans for radiotherapy applications: a literature review. J Appl Clin Med Phys 22(6):198–223
Bal M, Spies L (2006) Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering. Med Phys 33(8):2852–2859
Müller J, Buzug TM (2009) Spurious structures created by interpolation-based CT metal artifact reduction. Med Imaging 7258:72581Y
Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieß M (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37(10):5482–5493. https://doi.org/10.1118/1.3484090
Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E (2015) Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors. Br J Radiol 88(1052).
Andersson KM, Dahlgren CV, Reizenstein J, Cao Y, Ahnesjö A, Thunberg P (2018) Evaluation of two commercial CT metal artifact reduction algorithms for use in proton radiotherapy treatment planning in the head and neck area. Med Phys 45(10):4329–4344
Axente M, Paidi A, Von Eyben R, Zeng C, Bani-Hashemi A, Krauss A et al (2015) Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy. Med Phys 42(3):1170–1183
Kwon H, Kim KS, Chun YM, Wu HG, Carlson JNK, Park JM et al (2015) Evaluation of a commercial orthopaedic metal artefact reduction tool in radiation therapy of patients with head and neck cancer. Br J Radiol 88(1052):13–22
Bär E, Schwahofer A, Kuchenbecker S, Häring P (2015) Improving radiotherapy planning in patients with metallic implants using the iterative metal artifact reduction (iMAR) algorithm. Biomed Phys Eng Express 1(2):025206
Maerz M, Mittermair P, Krauss A, Koelbl O, Dobler B (2016) Iterative Metallartefakt Reduktion verbessert die Genauigkeit der Dosisberechnung: Phantomstudie mit Zahnimplantaten. Strahlentherapie und Onkol 192(6):403–413
Huang JY, Followill DS, Howell RM, Liu X, Mirkovic D, Stingo FC et al (2016) Approaches to reducing photon dose calculation errors near metal implants. Med Phys 43(9):5117–5130
Righetto R, Clemens LP, Lorentini S, Fracchiolla F, Algranati C, Tommasino F et al (2019) Accurate proton treatment planning for pencil beam crossing titanium fixation implants. Phys Medica 2020(70):28–38. https://doi.org/10.1016/j.ejmp.2020.01.003
Akdeniz Y, Yegingil I, Yegingil Z (2019) Effects of metal implants and a metal artifact reduction tool on calculation accuracy of AAA and Acuros XB algorithms in small fields. Med Phys 46(11):5326–5335
Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259(3):894–902
Jin Y, Giantsoudi D, Fu L, Verburg J, Gjesteby L, Wang G et al (2018) Metal artifact reduction for radiation therapy: a simulation study. Med Imaging 2018:25
Nielsen JS (2019) MR-based CT metal artifact reduction using Bayesian modelling Magnetic resonance-based computed tomography metal artifact reduction using Bayesian modelling. Phys Med Biol 64:245012
Li X, Kantor M, Zhu X, Frank S, Sahoo N, Li H (2014) SU-E-T-396: dosimetric accuracy of proton therapy for patients with metal implants in CT scans using metal deletion technique (MDT) artifacts reduction. Med Phys 41(6 Part 17):316
Nielsen JS, Van Leemput K, Edmund JM (2019) MR-based CT metal artifact reduction for head-and-neck photon, electron, and proton radiotherapy. Med Phys 46(10):4314–4323
Johnson TRC (2012) Dual-energy CT: general principles. AJR Am J Roentgenol 199(5 Suppl):3–8
Omoumi P, Verdun FR, Guggenberger R, Andreisek G, Becce F (2015) Dual-energy CT: basic principles, technical approaches, and applications in musculoskeletal imaging (Part 2). Semin Musculoskelet Radiol 19(5):438–445
Yadava GK, Pal D, Hsieh J (2014) Reduction of metal artifacts: beam hardening and photon starvation effects. Med Imaging 9033:90332V
Wellenberg RHH, Donders JCE, Kloen P, Beenen LFM, Kleipool RP, Maas M et al (2018) Exploring metal artifact reduction using dual-energy CT with pre-metal and post-metal implant cadaver comparison: are implant specific protocols needed? Skeletal Radiol 47(6):839–845
Long Z, DeLone DR, Kotsenas AL, Lehman VT, Nagelschneider AA, Michalak GJ et al (2019) Clinical assessment of metal artifact reduction methods in dual-energy CT examinations of instrumented spines. Am J Roentgenol 212(2):395–401
Kovacs DG, Rechner LA, Appelt AL, Berthelsen AK, Costa JC, Friborg J et al (2018) Metal artefact reduction for accurate tumour delineation in radiotherapy. Radiother Oncol 126(3):479–486. https://doi.org/10.1016/j.radonc.2017.09.029
Wu J, Shih CT, Chang SJ, Huang TC, Sun JY, Wu TH (2011) Metal artifact reduction algorithm based on model images and spatial information. Nucl Instrum Methods Phys Res Sect A 652(1):602–605. https://doi.org/10.1016/j.nima.2011.01.041
Luzhbin D, Wu J (2020) Model image-based metal artifact reduction for computed tomography. J Digit Imaging 33:71–82
Hartigan JA, Wong MA (1979) Algorithm AS 136: A K-means clustering algorithm. J R Stat Soc Ser C 28(1):100–108
Hahn A, Knaup M, Brehm M, Sauppe S, Kachelrieß M (2018) Two methods for reducing moving metal artifacts in cone-beam CT. Med Phys 45(8):3671–3680
Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieß M (2012) Frequency split metal artifact reduction (FSMAR) in computed tomography. Med Phys 39(4):1904–1916. https://doi.org/10.1118/1.3691902
Rodríguez-Gallo Y, Orozco-Morales R, Pérez-Díaz M (2018) Metal artifact reduction by morphological image filtering for computed tomography BT—world congress on medical physics and biomedical engineering. Springer, Singapore, pp 219–222
Wang Z, Bovik AC, Sheikh HR, Simoncelli EP (2004) Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process 13(4):600–612
Zhang Y, Yu H (2018) Convolutional neural network based metal artifact reduction in X-ray computed tomography. IEEE Trans Med Imaging 37(6):1370–1381
De Los SJ, Popple R, Agazaryan N, Bayouth JE, Bissonnette JP, Bucci MK et al (2013) Image guided radiation therapy (IGRT) technologies for radiation therapy localization and delivery. Int J Radiat Oncol Biol Phys 87(1):33–45. https://doi.org/10.1016/j.ijrobp.2013.02.021
Koike Y, Anetai Y, Takegawa H, Ohira S, Nakamura S, Tanigawa N (2020) Deep learning-based metal artifact reduction using cycle-consistent adversarial network for intensity-modulated head and neck radiation therapy treatment planning. Phys Medica 78:8–14. https://doi.org/10.1016/j.ejmp.2020.08.018
Nakamura M, Nakao M, Imanishi K, Hirashima H, Tsuruta Y (2021) Geometric and dosimetric impact of 3D generative adversarial network-based metal artifact reduction algorithm on VMAT and IMPT for the head and neck region. Radiat Oncol 16(1):96. https://doi.org/10.1186/s13014-021-01827-0
Nakao M, Imanishi K, Ueda N, Imai Y, Kirita T, Matsuda T (2020) Regularized three-dimensional generative adversarial nets for unsupervised metal artifact reduction in head and neck CT images. IEEE Access 8:109453–109465
Gjesteby L, Shan H, Yang Q, Xi Y, Jin Y, Giantsoudi D et al (2019) A dual-stream deep convolutional network for reducing metal streak artifacts in CT images. Phys Med Biol 64(23):235003
Acknowledgements
We thank F. Edward Boas, MD, Ph.D. for providing the MDT algorithm. We also thank Giovanna Dipasquale and Nikolaos Koutsouvelis from the Division of Radiation Oncology at the Geneva University Hospital as well as Nicoletta Lomax and Roger Hälg from Aarau Kantonspital for their support during the data acquisition of the ART and ATOM® phantoms for this study. Further, we thank Peta Gray from the Herston Medical Research Institute (HIRF) for providing support to collect data from the CT torso phantom.
Funding
Open Access funding enabled and organized by CAUL and its Member Institutions. This publication is part of a project that has received funding from the European Union’s Horizon 2020 research and innovation program under grant agreement No. 954783. The data acquisition at the Geneva University Hospital, Geneva, Switzerland, was supported by Innosuisse Project No. 40854.1 INNO-LS. Data collection at HIRF, Queensland, Australia was funded by a grant under the HIRF Project Support Scheme 2020.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Puvanasunthararajah, S., Camps, S.M., Wille, ML. et al. Combined clustered scan-based metal artifact reduction algorithm (CCS-MAR) for ultrasound-guided cardiac radioablation. Phys Eng Sci Med 45, 1273–1287 (2022). https://doi.org/10.1007/s13246-022-01192-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-022-01192-6