Abstract
Multi-Dynamic Multi-Echo (MDME) Sequence is a new method which can acquire various contrast-weighted images using quantitative relaxometric parameters measured from multicontrast images. The purpose of our study was to investigate the effect of scan parameters of MDME Sequence on measured T1, T2 values of phantoms at 3.0 T MRI scanner. Gray matter, white matter and cerebrospinal fluid simulation phantoms with different relaxation times (named GM, WM, CSF, respectively) were used in our study. All the phantoms were scanned 9 times on different days using MDME sequence with variations of echo train length, matrix, and acceleration factor. The T1, T2 measurements were acquired after each acquisition. The repeatability was characterized as the intragroup coefficient of variation (CV) of measured values over 9 times, and the discrepancies of measurements across different groups were characterized as intergroup CVs. The highest intragroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 1.36%, 1.75%, 0.74%, 1.41%, 1.70%, 7.79%, respectively. The highest intergroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 0.54%, 1.86%, 1.70%, 0.94%, 1.00%, 2.17%, respectively. Quantitative T1, T2 measurements of gray matter, white matter and cerebrospinal fluid simulation phantoms derived from the MDME sequence were not obviously affected by variations of scanning parameters, such as echo train length, matrix, and acceleration factor on 3T scanner.
Avoid common mistakes on your manuscript.
Introduction
Magnetic resonance imaging (MRI) is widely used in clinical practice for evaluating pathologies because of its excellent soft tissue contrast. Quantitative MRI has been gaining more interest because of its ability to provide absolute values for physical properties of different tissues, such as longitudinal relaxation time (T1), transverse relaxation time (T2) [1]. Several methods have been introduced for quantification of T1, T2 [2,3,4,5], but due to the unacceptable scan time, these methods had not been widely used in clinical practice.
Recently, a new method, which is now referred to as multi-dynamic multi-echo (MDME) sequence, enables acquisition of quantitative T1, T2 of the whole brain within approximately 6 min [3]. According to the quantitative data, various contrast images can be created with certain software by manipulating scanning parameters in the single acquisition. It shows promising results in imaging of central nervous system diseases, such as multiple sclerosis (MS) [6], brain tumor [7], Sturge-Weber syndrome [8], bacterial meningitis [9], and stroke [10], as well as diseases in musculoskeletal system [11], spine [12], prostate [13], and breast [14]. In neuroimaging field, brain tissues segmentation can be automatically performed and the volumes of different brain tissues can be calculated [15, 16]. So, it can be potentially used in normal aging and neurodegenerative diseases [17].
Since T1, T2 values represent physical constants that are presumably intrinsic to a given tissue or other material, changing image acquisition parameters theoretically should not alter them; however, model-based derivations of these parameters from real data cannot be expected to be perfectly reproducible. Previous studies showed good accuracy, repeatability, and reproducibility of T1, T2 measurement using different head coils on 1.5T scanner [18] and showed robustness of this sequence across different vendors of 3.0T scanner [19]. Recently, Kang et al. observed that some brain regions of T1 values are slightly changed according to different slice thickness or interslice gap [20]. Some studies showed that brain tissue and myelin volumetry derived from synthetic MRI were robust with different in-plane resolutions in 1.5 T [1] but differences were found in some brain regions in 3.0 T MRI scanner [21]. In clinical multicenter study, we may use MDME sequences with different scan parameters, or change scan parameters to achieve shorter scan time for patients who have difficulty cooperating with the examination. For further expansion of clinical applications of the MDME sequence, the effect of scan parameters, such as echo train length (ETL), acceleration factor and matrix, on the measured quantitative values needs to be investigated.
Hence, the aim of this study was to investigate the effect of scan parameters on T1, T2 relaxation times measured with MDME sequence using gray matter, white matter and cerebrospinal fluid simulation phantoms on 3 T scanner.
Methods
MRI acquisition
MRI examinations were performed on a 3 T scanner (SIGNA Pioneer; GE Healthcare, Milwaukee, USA) using a 32-channel head coil. Quantitative MRI was performed using MAGiC (MAGnetic resonance image Compilation) sequence [22]. This sequence is a multisection, multiecho, multisaturation delay method of saturation recovery acquisition that uses a fast spin-echo readout. A single basic block of this quantification sequence consists of 2 phases. In the first saturation phase, a slice-selective saturation pulse with flip angle θ is performed on slice n, followed by subsequent spoiling the signal (“saturation”). In the second acquisition phase, a slice-selective fast spin-echo acquisition is performed on another slice m (“acquisition”), consisting of multiple echoes which are acquired to measure transverse relaxation time (T2). By shifting between slice m and n, a desired delay time can be set between the saturation and acquisition of each specific slice. The longitudinal relaxation time (T1) after a saturation pulse can be retrieved from multiple scans, by using different delay times. Since the number of scans and delay times can be freely chosen, the dynamic range of T1 can also be set as desired [3]. In this way, two echo times and four delay times were used to quantify longitudinal T1 and transverse T2 relaxation times and eight complex images per slice were produced. To retrieve T1, T2 maps, while accounting for B1 inhomogeneity, a least square fit was performed on the signal intensity (I) of images by minimizing the following Eq. (1):
where α is the applied excitation flip angle (90°) and θ is the saturation flip angle (120°). A is an overall intensity scaling factor that takes into account several elements, including sensitivity of the coil, amplification of the radiofrequency chain, and voxel volume [19].
Phantom study
Gray matter, white matter and cerebrospinal fluid simulation phantoms (named GM, WM, CSF, respectively) produced by Wandong (Beijing Wandong Medical Technology Co., Ltd) were included in the study. GM (WD-TP001) consisted of NiCl2·6H2O 0.25 g, Agarose 9.6 g, Potassium Sorbate 0.1 g, H2O 1000 g; WM (WD-TP002) consisted of NiCl2·6H2O 0.5 g, Agarose 11 g, Potassium Sorbate 0.1 g, H2O 1000 g; CSF (WD-TP003) consisted of CuSO4·5H2O 0.02 g, H2O 1000 g.
Three phantoms were placed together and scanned using MDME sequences with different scan parameters for quantitative MRI (Fig. 1). In order to observe the effect of scan parameters on the measured T1, T2 values, we changed the ETL, matrix, and acceleration factor in a certain range, while trying to keep the other scan parameters unchanged. We selected 4 different ETLs (group ETL1–ETL4), 4 different matrixes (group matrix 1–matrix 4), and 4 different acceleration factors (group phase 1–phase 4). Totally, 12 sets of different scan parameters of MDME sequence were set. TRs (repetition time) and TEs (echo time) would be slightly automatically adjusted with the variations of matrix. The slice thickness was 4 mm, the interslice gap was 1 mm, and the field of view (FOV) was 240 mm × 240 mm.The other detailed sequence parameters were shown in Tables 1, 2, and 3.
The phantoms were scanned 9 times repeatedly on the same scanner, and each scan cycle lasted 3 days (3 acquisition groups) with 12 sets of different scan parameters of MDME sequence based on the modifications of ETL, matrix and phase, respectively. This arrangement should reduce the impact of long acquisition time on measured quantitative values. The entire experiment lasted nearly 1-month (day 1,4,7,10,13,16,19,21,24 for group ETL; day 2,5,8,11,14,17,20,23,26 for group matrix; day 3,6,9,12,15,18,21,24,27 for group phase). The phantoms were placed in scanner for 30 min before each scan. T1, T2 maps were acquired and processed using SyMRI software (SyntheticMR AB, version 8.0.4, Linköping, Sweden). A circle region of interest (ROI) (18.0 cm2) was placed in the center of each phantom on T1, T2 maps using RadiAnt DICOM Viewer software (Version 2021.2) to include as much of the circle as possible while avoiding partial volume effect. ROIs were copied and pasted on the images acquired at different times.
Statistical analysis
According to different sets of scan parameters, the measured values were divided into 3 groups: ETL, Matrix, and Phase. Mean values and standard deviations (SD) for T1, T2 in each group were determined. Coefficients of variation (CVs) were calculated within each group (intragroup CV) and across different groups (intergroup CV). The intergroup CV was calculated using the average values from each of the groups. Statistical analysis was performed using SPSS software (SPSS for Windows, 23.0.0.0, IBM).
Results
Table 4 showed the mean values and SD of all the T1, T2 measurements on three phantoms in each group. Tables 5, 6, and 7 showed the intragroup and intergroup CVs of T1, T2 measurements on GM, WM, and CSF acquired using MDME sequence with different ETLs (Table 5), matrixes (Table 6), and acceleration factors (Table 7). The intragroup CVs of all the T1, T2 measurements were less than 3%, except for T2 values of CSF. The intergroup CVs of all the quantitative T1, T2 values were less than 3%. For different ETLs, the highest intergroup CV was 1.00% for T1 and 0.80% for T2 (Table 5; Fig. 2). For different matrixes, the highest intergroup CV was 1.70% for T1 and 1.86% for T2 (Table 6; Fig. 3). For variations of acceleration factors, the highest intergroup CV was 0.83% for T1 and 2.17% for T2 (Table 7; Fig. 4).
Discussion
The MDME sequence enabled simultaneous acquisition of the physical tissue parameters T1, T2 in an acceptable scan time, and subsequent reconstruction of synthetic images. Promising application of the sequence has been previously reported in neuroimaging [8, 18, 23,24,25]. Clinical experience is still limited and most relies on the accurate and reproducible study [18, 19].
In this study, we changed the scan parameters of MDME sequence, including ETL, matrix, and acceleration factor within the range of routine clinical use, and evaluated the differences of measured T1, T2 values according to the variation of scan parameters in phantoms using 3T scanner. The observed intergroup CVs were lower than 3%, even lower than the intragroup CVs, indicating that MDME sequence is robust even across different scan parameters in a certain range.
Several studies have investigated the repeatability and reproducibility of MDME sequence. In Hagiwara’s study [19], they measured the quantitative values of the NIST/ISMRM (National Institute of Standards and Technology/International Society for Magnetic Resonance in Medicine) phantom across different scanners and showed that the highest intrascanner CVs was 2.07% for T1 values, 7.60% for T2 values. In our study, the intragroup CVs of T1 values on GM, WM, CSF and T2 values on GM, WM were very low (less than 3%), indicating that the MDME sequence is very reliable in measuring these values. The intragroup CVs of T2 values of CSF showed the highest intragroup CV (7.79%) which was consistent with previous study [19]. This variation is independent of scan parameters and may be related to the MDME sequence itself. The B1 inhomogeneity profiles differ per scan and imperfect gradient refocusing due to eddy currents may decrease signal readout, potentially resulting in an apparently altered T2 relaxation [19]. Another reason may be the limited number of relaxation data points in MDME sequence which makes it less reliable for examining tissues with very short or very long T2 relaxation times. So, it should be cautious to evaluate the T2 measurement of CSF or any other lesions containing fluid using this sequence.
In previous study, Hagiwara et al. [19] demonstrated that the inter-scanner CVs of T1, T2 were 10.86% and 15.27% in phantom data. In our study, the intergroup CVs were less than 3%. In Hagiwara’ study, the inter-scanner CVs of T1, T2 measurements were calculated using data from three different vendors, and in our study, the intergroup CVs of T1, T2 measurements were calculated from the same scanner with different scan parameters. This may be one reason for the bias. On the other hand, the phantoms we used were different from Hagiwara’s and the ROI sizes in our study were much larger, which might result in lower intergroup CVs. In our study, the intergroup CVs of T2 values for CSF were much lower than the intragroup CVs. This suggested that changing scan parameters of MDME sequence, such as ETL, matrix, and acceleration factors, has no obvious effect on the measured quantitative values. This may provide a basis for multicenter study using MDME sequence, and accelerate the acquisition when needed in emergent situations.
There are some limitations in our study. First, we only examine accuracy in T1, T2 values relative to MAGiC sequence variations and not across vendor platforms. The impact of scan parameters on measured quantitative values may be different across different vendors. Second, we only performed phantom measurements but not in vivo measurements. Human brain is much more complex, and the R1 and R2 relaxation might be multiexponential in one voxel under such various microscopic environments. Therefore, the effect of scan parameters on quantitative values of brain should be further investigated.
Conclusions
In conclusion, our single center study showed that changing the scan parameters of MDME sequence, such as ETL, matrix, and acceleration factor, has no obvious influence (within the difference of 3%) on the measured quantitative T1, T2 values of gray matter, white matter and cerebrospinal fluid simulation phantoms on 3.0T scanner.
References
Andica C, Hagiwara A, Hori M et al (2018) Automated brain tissue and myelin volumetry based on quantitative MR imaging with various in-plane resolutions. J Neuroradiol 45:164–168
Newbould RD, Skare ST, Alley MT et al (2010) Three-dimensional T(1), T(2) and proton density mapping with inversion recovery balanced SSFP. Magn Reson Imaging 28:1374–1382
Warntjes JB, Leinhard OD, West J et al (2008) Rapid magnetic resonance quantification on the brain: optimization for clinical usage. Magn Reson Med 60:320–329
Ehses P, Seiberlich N, Ma D et al (2013) IR TrueFISP with a golden-ratio-based radial readout: fast quantification of T1, T2, and proton density. Magn Reson Med 69:71–81
Deoni SC, Peters TM, Rutt BK (2005) High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2. Magn Reson Med 53:237–241
Hagiwara A, Hori M, Yokoyama K et al (2017) Utility of a multiparametric quantitative mri model that assesses myelin and edema for evaluating plaques, periplaque white matter, and normal-appearing white matter in patients with multiple sclerosis: a feasibility study. AJNR Am J Neuroradiol 38:237–242
Kang KM, Choi SH, Hwang M et al (2018) Application of synthetic MRI for direct measurement of magnetic resonance relaxation time and tumor volume at multiple time points after contrast administration: preliminary results in patients with brain metastasis. Korean J Radiol 19:783–791
Hagiwara A, Nakazawa M, Andica C et al (2016) Dural enhancement in a patient with sturge-weber syndrome revealed by double inversion recovery contrast using synthetic MRI. Magn Reson Med Sci 15:151–152
Andica C, Hagiwara A, Nakazawa M et al (2017) Synthetic MR imaging in the diagnosis of bacterial meningitis. Magn Reson Med Sci 16:91–92
Duchaussoy T, Budzik JF, Norberciak L et al (2019) Synthetic T2 mapping is correlated with time from stroke onset: a future tool in wake-up stroke management? Eur Radiol 29:7019–7026
Yi J, Lee YH, Song HT et al (2018) Clinical feasibility of synthetic magnetic resonance imaging in the diagnosis of internal derangements of the knee. Korean J Radiol 19:311–319
Jiang Y, Yu L, Luo X et al (2020) Quantitative synthetic MRI for evaluation of the lumbar intervertebral disk degeneration in patients with chronic low back pain. Eur J Radiol 124:108858
Cui Y, Han S, Liu M et al (2020) Diagnosis and grading of prostate cancer by relaxation maps from synthetic MRI. J Magn Reson Imaging 52:552–564
Jung Y, Gho SM, Back SN, et al (2018) The feasibility of synthetic MRI in breast cancer patients: comparison of T2 relaxation time with multiecho spin echo T2 mapping method. Br J Radiol 20180479
West J, Warntjes JB, Lundberg P (2012) Novel whole brain segmentation and volume estimation using quantitative MRI. Eur Radiol 22:998–1007
Warntjes M, Engstrom M, Tisell A et al (2016) Modeling the presence of myelin and edema in the brain based on multi-parametric quantitative MRI. Front Neurol 7:16
Lou B, Jiang Y, Li C, et al (2021) Quantitative analysis of synthetic magnetic resonance imaging in Alzheimer’s disease. Front Aging Neurosci 13
Krauss W, Gunnarsson M, Andersson T et al (2015) Accuracy and reproducibility of a quantitative magnetic resonance imaging method for concurrent measurements of tissue relaxation times and proton density. Magn Reson Imaging 33:584–591
Hagiwara A, Hori M, Cohen-Adad J et al (2019) Linearity, bias, intrascanner repeatability, and interscanner reproducibility of quantitative multidynamic multiecho sequence for rapid simultaneous relaxometry at 3 T: a validation study with a standardized phantom and healthy controls. Invest Radiol 54:39–47
Kang KM, Choi SH, Kim H et al (2019) The effect of varying slice thickness and interslice gap on T1 and T2 measured with the multidynamic multiecho sequence. Magn Reson Med Sci 18:126–133
Saccenti L, Andica C, Hagiwara A et al (2019) Brain tissue and myelin volumetric analysis in multiple sclerosis at 3T MRI with various in-plane resolutions using synthetic MRI. Neuroradiology 61:1219–1227
Tanenbaum LN, Tsiouris AJ, Johnson AN et al (2017) Synthetic MRI for clinical neuroimaging: results of the magnetic resonance image compilation (MAGiC) prospective, multicenter, multireader trial. AJNR Am J Neuroradiol 38:1103–1110
West J, Blystad I, Engstrom M et al (2013) Application of quantitative MRI for brain tissue segmentation at 1.5 T and 3.0 T field strengths. PLoS ONE 8:e74795
Granberg T, Uppman M, Hashim F et al (2016) Clinical feasibility of synthetic MRI in multiple sclerosis: a diagnostic and volumetric validation study. AJNR Am J Neuroradiol 37:1023–1029
Hagiwara A, Hori M, Yokoyama K et al (2017) Synthetic MRI in the detection of multiple sclerosis plaques. AJNR Am J Neuroradiol 38:257–263
Funding
This work was supported by Beijing Scholar Program [No. (2015) 160].
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by ZZ, YH and LY. The first draft of the manuscript was written by ZZ and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Ethical approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zheng, Z., Yang, J., Zhang, D. et al. The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study. Phys Eng Sci Med 45, 657–664 (2022). https://doi.org/10.1007/s13246-022-01128-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-022-01128-0