Abstract
Radiochromic film is a self developing two-dimensional dosimeter system that is widely used in radiotherapy. Since its development some 40 years ago many improvements have been made; however the sensitivity can vary across the film and a non-linear dose response remains. The former can be addressed using a double exposure technique; however this technique assumes that the response is linear so that the dose distribution is incorrectly measured. We are proposing the use of two homogenous exposures; one performed prior, the other after the irradiation to be measured. In this ‘multiple exposure technique’ (MET), the first homogenous exposure can be used to correct for variations in response in different parts of the film, while the second homogenous exposure allows correction for non-linearity of response with dose. The MET was tested with a 60° wedged field and an Intensity Modulated Radiation Therapy fluence map produced by a computerised treatment planning system. In the wedge field measurements, the MET profile showed agreement within 0.6 cGy for 80% of the field compared to ionisation chamber dose values. A comparison of a TPS generated fluence map dose distribution with one measured using conventionally calibrated EBT, and another measured with MET calibrated EBT, showed similar agreement. The MET would be particularly useful for the assessment of highly inhomogenous dose distributions with high maximum dose such as encountered in hypofractionated radiation therapy.
References
ISP (2007) Gafchromic® EBT self-developing film for radiotherapy dosimetry corporate whitepaper. International speciality products, New Jersey. http://online1.ispcorp.com/_layouts/Gafchromic/content/products/ebt/pdfs/EBTwhitepaper.pdf. Accessed 21 March 2011
Khan F (2006) The physics of radiation therapy, 3rd edn. Lippincott Williams & Williams, Hagerstown
Fuss M, Sturtewagen E, Wagter CD, Georg D (2007) Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 52:4211
Fiandra C, Ragona R, Ricardi U, Anglesio S, Giglioli FR (2008) Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse. Med Phys 35:5463
Butson MJ, Cheung T, Yu PKN (2006) Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range. Appl Radiat Isot 64(1):60–62
ISP (2009) Gafchromic® EBT2 self-developing film for radiotherapy dosimetry corporate whitepaper. International speciality products, New Jersey. http://online1.ispcorp.com/_layouts/Gafchromic/content/products/ebt2/pdfs/GAFCHROMICEBT2TechnicalBrief-Rev1.pdf. Accessed 21 March 2011
Kairn T, Aland T, Kenny J (2010) Local heterogeneities in early batches of EBT2 film: a suggested solution. Phys Med Biol 55:L37
Hartmann B, Martišíková M, Jäkel O (2010) Technical note: homogeneity of Gafchromic EBT2 film. Med Phys 37:1753
Zhu Y, Kirov AS, Mishra V, Meigooni AS, Williamson JF (1997) Quantitative evaluation of radiochromic film response for two-dimensional dosimetry. Med Phys 24:223
Niroomand-Rad A, Blackwell C, Coursey B, Gall K, Galvin J, McLaughlin W, Meigooni A, Nath R, Rodgers J, Soares C (1998) Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55. American association of physicists in medicine. Med Phys 25(11):2093
Klassen NV, Van Der Zwan L, Cygler J (1997) GafChromic MD-55: investigated as a precision dosimeter. Med Phys 24:1924
Roozen K (1999) Radiochromic film: automated dosimetry and uniformity. University of Canterbury, Christchurch
Andreo P, Izewka J, Shortt K, Vatnitsky S, Van Dyk J, Cramb J, Fraass BA, Ionescu-Farca F, Levin V, Mijnheer B, Rosenwald J-C, Scalliet P (2004) Commissioning and quality assurance of computerized planning systems for radiation treatment of cancer. International Atomic Energy Agency technical report series no. 430, IAEA, Vienna
Ferreira B, Lopes M, Capela M (2009) Evaluation of an epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry. Phys Med Biol 54:1073
Saur S, Frengen J (2008) GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis. Med Phys 35:3094
Paelinck L, Neve WD, Wagter CD (2007) Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry. Phys Med Biol 52:231
Menegotti L, Delana A, Martignano A (2008) Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure. Med Phys 35:3078
Low DA, Harms WB, Mutic S, Purdy JA (1998) A technique for the quantitative evaluation of dose distributions. Med Phys 25:656
Low DA, Dempsey JF (2003) Evaluation of the gamma dose distribution comparison method. Med Phys 30:2455
Jiang SB, Sharp GC, Neicu T, Berbeco RI, Flampouri S, Bortfeld T (2006) On dose distribution comparison. Phys Med Biol 51:759
Millar M, Cramb J, Das R, Ackerly T, Brown A, Webb D (1997) Recommendations for the safe use of external beams and sealed brachytherapy sources in radiation oncology. Australas Phys Eng Sci Med Suppl 20(3)
Vial P, Greer PB, Oliver L, Baldock C (2008) Initial evaluation of a commercial EPID modified to a novel direct-detection configuration for radiotherapy dosimetry. Med Phys 35:4362
Pedraza Muriel V (2007) Hypofractionation in radiotherapy. Clin Transl Oncol 9(1):21–27
Macías V, Biete A (2009) Hypofractionated radiotherapy for localised prostate cancer. Review of clinical trials. Clin Transl Oncol 11(7):437–445
Khan A, Haffty BG (2010) Hypofractionation in adjuvant breast radiotherapy. The Breast 19(3):168–171
Brown S, Rodger A, Orton C (2009) Point/Counterpoint. Hypofractionation is a proven safe and effective modality for postoperative whole-breast radiotherapy for early breast cancer patients. Med Phys 36(6):1927
Jin JY, Kong FM, Chetty IJ, Ajlouni M, Ryu S, Ten Haken R, Movsas B (2010) Impact of fraction size on lung radiation toxicity: hypofractionation may be beneficial in dose escalation of radiotherapy for lung cancers. Int J Radiat Oncol Biol Phys 76(3):782–788
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Roozen, K., Kron, T., Haworth, A. et al. Evaluation of EBT radiochromic film using a multiple exposure technique. Australas Phys Eng Sci Med 34, 281–289 (2011). https://doi.org/10.1007/s13246-011-0067-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-011-0067-3