Suggested baseline ranges for performance testing of the Fujifilm FCR Capsula XL computed radiography system

  • Technical Paper
  • Published:
Australasian Physical & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

Due to the difficulty in obtaining expected baseline values from the supplier for the performance testing of the Fujifilm FCR Capsula XL computed radiography system, performance test results from the acceptance testing of 42 Capsula systems were reviewed. Guidance ranges are provided for nine system performance tests which can be applied to future testing of calibrated systems.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Report of the American Association of Physicists in Medicine (AAPM) (2006) Task Group 10, AAPM Report No. 93: Acceptance testing and quality control of photostimulable storage phosphor imaging systems. American Association of Physicists in Medicine, USA

  2. Institute of Physics and Engineering in Medicine (2005) Report 91: Recommended standards for the routine performance testing of diagnostic x-ray imaging systems, Institute of Physics and Engineering in Medicine, Fairmount

  3. King’s Centre for the Assessment of Radiological Equipment (KCARE) (2010) Protocol for the QA of computed radiography systems commissioning and annual QA tests. http://www.kcare.co.uk/userfiles/Education/CR(Physics)protocolDRAFT8.0.doc. Accessed 26 Oct 2010

  4. Fujifilm (2006) FCR quality assurance program plus guidebook for medical institutions—Image reader, 3rd edn

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to David L. Thiele.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Thiele, D.L., Schick, D.K. & Barr, S. Suggested baseline ranges for performance testing of the Fujifilm FCR Capsula XL computed radiography system. Australas Phys Eng Sci Med 33, 335–340 (2010). https://doi.org/10.1007/s13246-010-0052-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-010-0052-2

Keywords

Navigation