Many studies have investigated how to eliminate image blurring caused by cardiac and respiratory motions in cardiac single-photon emission computed tomography (SPECT) and positron emission tomography (PET) images. Cardiac electrocardiographic (ECG)-gated SPECT and PET are established methods with which to monitor the beating heart, which allows evaluation of cardiac functions such as left ventricular ejection fraction (LVEF), cardiac volume, and the contraction indices. Consequently, an algorithm was created to generate images that are free of cardiac motion, and it has been clinically applied.1 Image blurring caused by cardiac motion was eliminated while myocardial count density was maintained. By contrast, some medical devices can monitor and measure abdominal and thoracic movements during respiration,2 and a data-driven algorithm to detect respiratory motion is also available.3 Therefore, clinical dual-gated myocardial perfusion SPECT and PET image acquisitions could become routine considering these technological advancements.

However, since organ displacement cannot be accurately measured due to cardiac and respiratory motions in clinical patients and precise organ movements cannot be reproduced in an acrylic phantom, algorithms and software for detecting respiratory and cardiac motions are difficult to validate. A mathematical digital phantom is a powerful simulation tool that can evaluate the effects of cardiac and respiratory motions on the quality of SPECT and PET images. A 4D extended cardiac-torso (XCAT) anthropomorphic digital phantom based on a realistic model of patient anatomy is available from the Duke University.4 Cardiac and respiratory motions can be simulated in this phantom based on gated X-ray computed tomography data, and amplitudes and timing for each motion can be mathematically controlled. We acquired virtual myocardial perfusion SPECT images using the 4D XCAT digital phantom through simulating medical imaging nuclear detectors (SIMIND) Monte Carlo software5 to authenticate a novel program that can evaluate ECG-gated myocardial perfusion SPECT data.6 Figure 1 shows an example of normal myocardial perfusion images derived from a clinical patient and a digital phantom. The image quality of the digital phantom was essentially equivalent to that of the clinical patient. This is because not only cardiac 99mTc activity but also hepatic, pulmonary, and background activities in the digital phantom were considered in that comparison (Figure 1A, B). Moreover, cardiac and other organ movements were also represented. Polar maps created out of data generated from the digital phantom and the patients corresponded in terms of ECG-gated SPECT functions such as perfusion at end-diastolic and end-systolic timings, wall motion, and wall thickening (Figure 1C).

Figure 1
figure 1

Examples of 99mTc normal myocardial perfusion distribution in a digital phantom and a clinical patient. Apical, mid, and basal short-axis, and vertical, and horizontal long-axis displays of myocardial perfusion distribution derived from a male digital phantom (A) and a 72-year-old man with a low likelihood of coronary artery disease (B). Ejection fractions of patient and digital phantoms were 62% and 61%, respectively. Polar maps of perfusion in ECG-non-gated images, ED, ES, wall motion, and thickening (C). ECG, electrocardiographic; ED, end-diastolic; and ES, end-systolic

The study by Zhang and colleagues published in this issue of the Journal of Nuclear Cardiology evaluated respiratory-gated methods to reduce motion-induced blurring in myocardial perfusion SPECT images.7 Alternative time- and amplitude-based respiratory curves can be used to acquire respiratory-gated images from list-mode datasets. That cited study compared respiratory amplitude (AG)-, equal count (EG)-, and equal time- or phase (TG)-gated images acquired from 10 digital phantom models and clinical patients. Acquired counts are typically limited in myocardial perfusion SPECT, but appropriate count statistics can be obtained by image registration processing in which each respiratory bin is registered to a reference bin. Zhang et al. fit respiratory bins to the reference bin using an affine and b-spline nonrigid registration method. Respiratory motion was reduced more in AG and EG than in TG. Cardiac and respiratory motions will require compensation to precisely determine myocardial perfusion distribution. Compensating dual-gated PET myocardial perfusion images with cardiac and respiratory motions has improved image resolution, contrast, and contrast-to-noise ratios.8 However, further investigation of routine clinical dual-gated myocardial perfusion SPECT images are required, since very few studies have addressed this matter.9

In addition to image-registration processing, respiratory-gated imaging using a cadmium zinc telluride (CZT) gamma camera might be promising,10 because this type of camera has high-sensitivity detectors that can monitor respiratory motion even during short acquisition periods.3 However, myocardial perfusion distribution differs according to the structure of CZT and Anger gamma cameras, as well as the scan position and time.11 Therefore, the quality of respiratory-compensated myocardial perfusion images needs to be confirmed for each instrument. Moreover, the applicability of dual-gated myocardial perfusion imaging will require multicenter validation studies using digital phantoms and clinical patients. Normal databases for the calculation of summed stress/rest/difference scores might be updated for dual-gated imaging, and the effects of dual-gated imaging on normal values for LVEF, volume, and contraction indices should be also investigated.

Clinical applications of cardiac and respiratory-gated myocardial perfusion SPECT and PET imaging methods are just emerging. When ECG gating was introduced during the 1990s, it was rapidly accepted for clinical practice, which resulted in more quantitative functional evaluation. The need for ECG gating is not in any doubt. Respiratory gating is theoretically desirable to reduce artifacts caused by complicated cardiac movements, but the technology will need to improve before it can become clinically applicable. A digital phantom is suitable for dual-gated myocardial perfusion studies to confirm the performance of a novel or improved compensation algorithm for cardiac and respiratory motions. Since digital phantoms and Monte Carlo simulation codes are widely available, the accuracy of cardiac and respiratory-gated SPECT and PET imaging can continuously improve. Perhaps dual-gated imaging will become commonplace in clinical practice within 10 years, but quality, reliability, and practicality will require enhancement in the meantime.