Introduction

Energy drinks are one of the fastest growing soft drink markets around the world (Euromonitor 2013). They have been recently consumed in different societies, especially by younger consumers, because they give power throughout the day due to caffeine, vitamins, well-absorbed monosaccharides, and other ingredients such as taurine (Euromonitor 2013; Sather and Vernig 2011). The results of a European survey of 52,000 respondents from 16 countries showed that approximately 30 % of adults are regularly drinking caffeine-enriched beverages (Higdon and Frei 2006). Vitamins and dietary supplements are also a relevant category in the field of consumer’s health. In 2013, their worldwide value was estimated at US$84.4 billion (Euromonitor 2014).

Caffeine is a xanthine alkaloid acting as a central nervous system stimulant. It temporarily increases blood pressure and eliminates drowsiness. Caffeine is classified by the Food and Drug Administration as generally recognized as safe (GRAS), and its common consumption, below 400 mg per day (Zucconi et al. 2013), has low health risk or even its protective effects against some diseases, including Parkinson disease (Prediger 2010), and certain types of cancer have been reported (Uccella et al. 2013; Miura et al. 2014). However, the overconsumption of this drug could potentially be harmful (Jiang et al. 2013).

Water-soluble vitamins include nine water-soluble groups of compounds (eight B vitamins and vitamin C) showing diverse biochemical functions. Each “vitamin” refers to a number of vitamin derivatives that all show the biological activity associated with a particular vitamin. Although the Recommended Dietary Allowances (RDAs) for vitamins are relatively small, even the mild deficiency of one of them can lead to serious health problems. Nutritional deficit of vitamin B1 (thiamine) can lead to beriberi disease manifested by cardiovascular disorders and impairment of digestive and nervous systems (Sriram et al. 2012). Deficiencies of vitamin B2 (riboflavin and its co-enzymes: flavin mononucleotide (FMN) and flavin adenine dinucleotide (FAD)), vitamin B3 (niacin), vitamin B6 (pyridoxine, pyridoxamine, pyridoxal), and vitamin B7 (biotin) mainly occur as skin inflammation and discomfort from the gastrointestinal tract (Powers 2003; Ball 2006). Since vitamin B5 (pantothenic acid) is found in most food products, its deficiency is relatively rare (Ball 2006).

Several methods exist for the determination of caffeine as a single component or in combination with other drugs in pharmaceutical formulations or biological matrix of food (Patil 2012). The most popular include chromatographic techniques such as high-performance liquid chromatography (HPLC), high-performance thin layer chromatography (HPTLC), or capillary electrophoresis with ultraviolet (UV) or mass spectrometry (MS) detection. They allow the determination of caffeine as a single compound in herbal products and energy drinks or its separation from other alkaloids in chocolate, tea, coffee, urine, or serum (Abourashed and Mossa 2004; Tokusoglu and Kemal 2002; Jafari et al. 2011; Perrone et al. 2008; Peri-Okonny et al. 2005; Zhao and Lunte 1997). Caffeine can be separated from polyphenols in tea and coffee (Hadad et al. 2012; Samanidou et al. 2012; Poerner and Bragagnolo 2013) or from other drugs in pharmaceutical preparations (Gámiz-Gracia and Luque de Castro 1997; Zen and Ting 1997; Sultan et al. 2013).

For vitamin-enriched energy drinks, a micellar electrokinetic chromatography method was proposed to determine 2-aminoethanesulfonic acid, nicotinamide, pyridoxine, caffeine, riboflavin, and thiamine (Okamoto et al. 2002). Sather and Vernig (2011) and Leacock et al. (2011) described the HPLC method for the determination of caffeine and vitamin B6 (pyridoxine). The HPTLC for determination of riboflavin, pyridoxine, nicotinamide, caffeine, and taurine in energy drinks was developed by Aranda and Morlock (2006), but HPTLC is a less commonly used technique than HPLC.

In the present study, a new HPLC method with photodiode array (PDA) detection was proposed for determination of caffeine and water-soluble vitamins. The method with fluorescence (FL) detection for determination of vitamins B2 (in the form of FMN and riboflavin) and B6 (as pyridoxal, pyridoxamine, and pyridoxine) was also described. According to the authors’ knowledge, there is no chromatographic method that allows separation and determination of caffeine in combination with most of the water-soluble vitamins. Moreover, simultaneous PDA and FL detection of these compounds cannot be found in the literature. The proposed HPLC-PDA-FL method is dedicated for the determination of caffeine and water-soluble vitamins in energy drinks. The method can be also an alternative for existing methods of water-soluble vitamin determinations in pharmaceutical preparations.

Materials and Methods

Materials

Thiamine hydrochloride (vitamin B1) and pyridoxine hydrochloride (vitamin B6) were purchased from Fluka (Buchs, Switzerland). Pyridoxamine dihydrochloride, pyridoxal hydrochloride (vitamin B6), cyanocobalamin (vitamin B12), riboflavin and riboflavin 5′-monophosphate sodium salt hydrate (FMN, vitamin B2), nicotinic acid and nicotinamide (vitamin B3), folic acid (vitamin B9), pantothenic acid sodium salt (vitamin B5), and caffeine were purchased from Sigma (St. Louis, MO, USA). Ascorbic acid (vitamin C) was from Merck (Darmstadt, Germany). Methanol of HPLC grade was from Chempur (Piekary Śląskie, Poland).

Certified reference materials (CRM; Cerilliant, TX, USA), caffeine solution (1 mg mL−1 in methanol), pyridoxine hydrochloride (vitamin B6) solution (1 mg mL−1 in methanol), and nicotinamide (vitamin B3) solution (1 mg mL−1 in methanol), were purchased from Sigma-Aldrich (Munich, Germany). CRMs of other vitamins were not commercially available or were withdrawn from the market.

Demineralized water was obtained from a Hydrolab System (Hydrolab, Wiślina, Poland) and filtered by Millipore system using 0.45 μm nylon filters (Waters, Milford, MA, USA). Energy drinks and vitamin supplements were purchased in local supermarkets and pharmacies.

Sample Preparation

Liquid energy drinks were analyzed directly after ultrasonic degassing for 10 min. Energy drinks and vitamin supplements in tablets were weighted, dissolved in 150 mL of demineralized water, and sonicated for 10 min. Subsequently, the sample was transferred to a 200-mL volumetric flask and filled up with water to the mark. Samples were protected from light by aluminum foil and analyzed using HPLC. Before the HPLC analysis, the samples were centrifuged for 5 min at 14,000 g (microcentrifuge Eppendorf MiniSpin plus, Warsaw, Poland). No supernatant for the tested products was observed; thus, the recovery studies were not necessary. For each type of product, three independent samples were prepared.

HPLC Determination of Caffeine and Water-Soluble Vitamins

A Waters 600 high-performance liquid chromatograph (Waters, Milford, MA, USA) equipped with Nova-Pak C18 column (150 mm × 3.9 mm, 5 μm) fitted with μBondapak C18 cartridge guard column (Waters, Milford, MA, USA) was used. A gradient of mobile phase consisting of methanol (solvent A) and 0.05 M NaH2PO4 containing 0.005 M hexanesulfonic acid, pH 3.0 (solvent B) was developed and used according to the following program: linear increment starting with 10 % A to 40 % A for 20 min and the return to the initial conditions within the next 10 min with the flow rate of 1 mL min−1. The injection volume was 20 μL. The eluate was detected using a Waters 996 PDA detector set at the wavelength characteristic for appropriate vitamin or caffeine. Moreover, a Waters 474 scanning fluorescence detector set at an excitation wavelength of 290 nm and emission at 390 nm for vitamin B6 or 450 nm excitation and 530 nm emission wavelengths for vitamin B2 was used. Emission slit width was 10 nm, fluorometer gain was 10, and attenuation was 1. Caffeine and vitamins were identified by comparison of their retention times to those of corresponding standards. Additionally, UV or ultraviolet/visible (UV/VIS) absorption spectra measured using the PDA detector were used to confirm the identity and purity of the compounds.

Method Validation

Validation of the HPLC method included linearity, sensitivity, accuracy, and instrument precision. Quantification of caffeine and vitamins was performed using an external standard method. At least seven-point calibration curves (n = 3) were prepared with the aqueous solutions of standards at the levels covering those in the assessed samples.

Accuracy of the method was verified using solutions of CRMs: nicotinamide (vitamin B3), pyridoxine (vitamin B6), and caffeine. From each CRM, five different solutions in demineralized water were prepared (1, 5, 10, 25, and 50 μg mL−1 for nicotinamide and pyridoxine and 50, 100, 200, 300, and 500 μg mL−1 for caffeine), and their concentrations were determined using an appropriate calibration curve. Each concentration was prepared in triplicate. The accuracy of the method was expressed as an average percentage of the vitamin/caffeine concentration in CRM.

The limit of detection (LOD) and limit of quantification (LOQ) were calculated on the basis of signal-to-noise ratio (S/N) of three or ten, respectively. Instrument precision (intra- and interday) was checked from six consecutive injections of liquid and tablet energy drink samples.

Statistical Analysis

All determinations were carried out at least in triplicate and the results are presented as means ± standard deviation.

Results and Discussion

Analytical Characteristics of the HPLC Method

In the present study, the HPLC-PDA method for the simultaneous separation and determination of caffeine and 12 water-soluble vitamin compounds was proposed. Moreover, the HPLC-FL method for vitamins B2 (in the form of FMN and riboflavin) and B6 (as pyridoxal, pyridoxamine, and pyridoxine) was described.

Detection wavelengths and retention times for caffeine and water-soluble vitamins separated using the proposed method are shown in Table 1. Typical HPLC chromatograms for standards and energy drink are shown in Figs. 1 and 2, respectively. The UV or UV/VIS absorption spectra of all separated compounds are also included (Fig. 3). The elution order was as follows: vitamin C (ascorbic acid), B3 (nicotinic acid, nicotinamide), B5 (pantothenic acid), B6 (pyridoxal, pyridoxamine, pyridoxine), B1 (thiamine), B2 (FMN), caffeine, vitamin B9 (folic acid), B12 (cyanocobalamin), and B2 (riboflavin). Reproducibility of the retention times did not exceed 0.20 min (for thiamine). All tested compounds were separated using a gradient of mobile phase within 30 min including re-equilibration of the column before the next injection.

Table 1 HPLC detection wavelengths and retention times of caffeine and water-soluble vitamins
Fig. 1
figure 1

The HPLC separation of caffeine and water-soluble vitamin standards monitored at 254 and 202 nm and using a fluorescence detector (290/390 nm for vitamin B6 and 450/530 nm for vitamin B2). Asterisk, the change of the excitation and emission wavelengths. 1, ascorbic acid (vitamin C); 2, nicotinic acid (vitamin B3); 3, nicotinamide (vitamin B3); 4, pantothenic acid (vitamin B5); 5, pyridoxal (vitamin B6); 6, pyridoxamine (vitamin B6); 7, pyridoxine (vitamin B6); 8, thiamine (vitamin B1); 9, riboflavin 5-phosphate (FMN, flavin mononucleotide; vitamin B2); 10, caffeine; 11, folic acid (vitamin B9); 12, cyanocobalamin (vitamin B12); and 13, riboflavin (vitamin B2)

Fig. 2
figure 2

HPLC chromatograms of energy drink monitored at 254 nm and using a fluorescence detector. Asterisk, the change of the excitation and emission wavelengths. Legend as in Fig. 1

Fig. 3
figure 3

Absorption spectra of water-soluble vitamins and caffeine

Nicotinamide and pyridoxine hydrochloride are the forms of vitamin B3 and vitamin B6, respectively, used in the production of energy drinks and vitamin supplements. Riboflavin and riboflavin 5′-phosphate (flavin mononucleotide, FMN) are the forms of vitamin B2 which are also used as a dye in selected beverages. These two forms of vitamin B2 can be well separated and quantified using both the HPLC-PDA and HPLC-FL. The HPLC-PDA method also allows the quantification of both forms of vitamin B3 (nicotinic acid and nicotinamide). Moreover, all vitamin B6 forms—pyridoxal, pyridoxamine, and pyridoxine—can be determined. It is not excluded that the proposed methods could be useful for water-soluble vitamin determination in a matrix containing all forms of vitamins B2, B3, and B6 after appropriate sample preparation and purification. The developed HPLC-PDA and HPLC-FL methods were validated for linearity and sensitivity (Table 2), accuracy (Table 3) as well as for intra- and interday instrument precision (Table 4).

Table 2 Linearity and sensitivity of HPLC-PDA and HPLC-FL methods for determination of caffeine and water-soluble vitamins
Table 3 The accuracy of the HPLC method expressed as an average percentage of the vitamin/caffeine concentration in CRM
Table 4 Intra- and interday instrument precision for determination of caffeine and water-soluble vitamins in energy drinks

Quantification of the tested compounds was performed using the external standard method. The relative standard deviation (RSD) values of the standard curve slopes were below 5.3 % (for pantothenic acid at the wavelength of 202 nm). Linear correlation coefficients (r) for standard curves for all compounds were not lower than 0.997 (Table 2). The order of sensitivity of the HPLC-PDA, expressed as a LOD, was as follows: ascorbic acid < pyridoxine < nicotinic acid = folic acid < caffeine = cyanocobalamin (at 207 nm) < thiamine < pyridoxamine < riboflavin < pyridoxal < nicotinamide = cyanocobalamin (at 360 nm) < FMN < pantothenic acid. The LOD determined using the FL detector was as follows: riboflavin < FMN < pyrydoxine < pyridoxamine < pyridoxal (Table 2).

Accuracy of the HPLC-PDA and HPLC-FL methods (Table 3) was confirmed using CRMs of nicotinamide (vitamin B3), pyridoxine (vitamin B6), and caffeine. It was found that the accuracy for caffeine and nicotinamide determination, expressed as an average percentage of the caffeine/vitamin concentration in CRM, was 99.1 and 98.2 %, respectively. The concentration of pyridoxine could be determined with the accuracy of 97.2 % when the PDA detector was used. The accuracy of the HPLC-FL was higher (99.4 %).

Instrument precision (Table 4) was checked from six consecutive injections of the tested sample. The intraday (the daily) and interday (day-to-day) RSD values of the system were not higher than 1.87 and 1.73 %, respectively, for liquid energy drink. For tablet energy drink, intra- and interday instrument precision was 1.44 and 1.83 %, respectively.

Quantification of Caffeine and Water-Soluble Vitamins in Energy Drinks and Vitamin Supplements

The validated HPLC-PDA and HPLC-FL methods were applied to 12 different energy drinks: seven liquid drinks and five in the form of tablet. Moreover, three vitamin supplements were included in the study for extending the method for vitamins which cannot generally be found in energy drinks (vitamins B1 and B9). Table 5 presents the declared amount of caffeine and water-soluble vitamins and the values obtained using the proposed HPLC-PDA and HPLC-FL methods. Caffeine and vitamins declared by the producers were detected in all the tested products with the exception of vitamin B12. The declared level of maximal 5 ng of vitamin B12 per mL is too low for its detection (LOD for vitamin B12 in applied conditions was 23 and 63 ng mL−1 at the wavelengths of 207 and 360 nm, respectively). Ascorbic acid can be also separated from other water-soluble vitamins, but it cannot be properly quantified in some products because of the influence of interferences (quantitative results not shown). Ascorbic acid was not present in liquid energy drinks.

Table 5 The content of caffeine and water-soluble vitamins in energy drinks and vitamin supplements [mg 100 mL−1 or mg in tablet]: producer declaration and the content determined using PDA and FL detectors (in parenthesis)

The results obtained by the proposed HPLC method and the declared values are generally comparable with the exception of caffeine in liquid energy drinks. The results obtained in this study are slightly higher than the amount declared by the producers. The purity of the caffeine peak was confirmed, which may indicate that some energy drink producers use “technological allowance” in order to prevent the possible decrease of this compound during prolonged shelf-life.

Conclusions

The aim of the present study was to develop a chromatographic method for the separation and quantification of caffeine and water-soluble vitamins. The new HPLC-PDA and HPLC-FL methods were validated and applied to 15 products: energy drinks and vitamin supplements. The proposed HPLC method allows separation of 13 compounds: ascorbic acid, nicotinic acid, nicotinamide, pantothenic acid, pyridoxal, pyridoxamine, pyridoxine, thiamine, riboflavin 5′-phosphate (flavin mononucleotide), caffeine, folic acid, cyanocobalamin, and riboflavin in a relatively short time (30 min including re-equilibration time). The HPLC-FL method is selective for the determination of vitamins B2 (as FMN and riboflavin) and B6 (as pyridoxal, pyridoxamine, and pyridoxine). The proposed HPLC-PDA method is suitable for quantification of caffeine and water-soluble vitamins (nicotinamide, pantothenic acid, pyridoxine, thiamine, FMN, folic acid, and riboflavin) in energy drinks and vitamin supplements. The declared level of vitamin B12 in these products is too low for its quantification, whereas ascorbic acid, if present in tablet energy drinks, may co-elute with interfering compounds overestimating its quantity.

The proposed separation method is simple and sensitive; thus, it can be carried out in any laboratory equipped with HPLC-PDA or HPLC-FL systems for routine investigations whether energy drinks meet the quantity requirements. They can also be an alternative for existing methods of water-soluble vitamin determinations in pharmaceutical preparations. It cannot be also excluded that these methods may be useful for water-soluble vitamin determinations in other matrices containing all forms of B2, B3, and B6 vitamins after appropriate sample preparation.