Abstract
Objective
The aim of this study was to propose and verify a universal method of left ventricular myocardium segmentation, able to operate on heart gated PET data with different sizes, shapes and uptake distributions. The proposed method can be classified as active model method and is based on the BEAS (B-spline Explicit Active Surface) algorithm published by Barbosa et al. The method was implemented within the Pmod PCARD software package. Method verification by comparison with reference software and phantom data is also presented in the paper.
Methods
The proposed method extends the BEAS model by defining mechanical features of the model: tensile strength and bending resistance. Formulas describing model internal energy increase during its stretching and bending are proposed. The segmentation model was applied to the data of 60 patients, who had undergone cardiac gated PET scanning. QGS by Cedars-Sinai and ECTb by Emory University Medical Centre served as reference software for comparing ventricular volumes. The method was also verified using data of left ventricular phantoms of known volume.
Results
The results of the proposed method are well correlated with the results of QGS (slope: 0.841, intercept: 0.944 ml, R2: 0.867) and ECTb (slope: 0.830, intercept: 2.109 ml, R2: 0.845). The volumes calculated by the proposed method were very close to the true cavity volumes of two different phantoms.
Conclusions
The analysis of gated PET data by the proposed method results in volume measurements comparable to established methods. Phantom experiments demonstrate that the volume values correspond to the physical ones.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Segmentation is one of the most important steps of image processing in cardiac diagnostics. It allows volume calculation [1, 2] and plays an important role when deriving regional time activity curves for signal modeling [3,4,5,6] or for heart muscle strain approximation [7, 8]. The aim of this study was to propose a flexible method for human gated PET left ventricular myocardium segmentation, which operates on heart data of different sizes, shapes and uptake distributions.
The proposed method can be classified as an active model segmentation algorithm. The idea of active model segmentation is to modify position and shape of the model, so that it best satisfies some energy functions [9]. The classic formulation of active model segmentation was described by Kass et al. [10]. They proposed a model restricted by its stiffness and elasticity, which is deformed in the energy field derived from image gradient. The idea represents a group of methods called edge-based active models. They were further developed in several works, e.g. [11,12,13,14]. Region-based active models represent another approach, which is less sensitive to image noise. It was first proposed by Chan and Vese [15], who defined the model as dependent on the energy derived from differences in pixel mean values inside and outside the model boundary. Based on their work, many features and energy functions were formulated [16,17,18]. To increase the method efficiency, Lankton and Tannenbaum [19] proposed a localized approach, in which just the neighborhood of the model is taken into account during image feature analysis. Such a solution connected with an explicit form of model description [20, 21] was used by Barbosa et al., developing the B-spline Explicit Active Segmentation (BEAS) algorithm [22, 23]. They described in a spherical coordinate system a B-spline surface model stretched on a set of nodes, whose angle coordinates are locked and their positions described by only one explicit coordinate—the radius (r) [24]. To make the method more effective in epicardium and endocardium segmentation, an extended definition of the BEAS model energy is proposed in this study.
The model was implemented and tested using the software package Pmod PCARDP, Pmod Technologies LLC [25]. This paper presents the model definition, describes its application to human gated PET data as well as physical phantom data and presents the results of left ventricular volume measurements compared to the ones obtained using the reference software packages: QGS—quantitative gated SPECT [26] and Emory cardiac toolbox ECTb [27].
Materials and methods
Model energy definition
Barbosa et al., proposed in [22] the following model description:
where \(\varphi\) and \(\theta\) are angle coordinates and r is a radius coordinate in the spherical coordinate system with origin inside the model, \(\beta^{d}\) uniform symmetric B-spline function in two-dimensional space, \(c[k]\) the B-spline coefficients, and \(p\) is a point of coordinates \(\left( {\varphi ,\theta } \right)\). Model nodes k are located on a rectangular grid with spacing h.
The energy function, which drives the model shape modification, was proposed as follows:
where \(H_{\phi } (x)\) is the Heaviside function [28], \(\varOmega\) an image space and \(f_{\text{in}}\) and \(f_{\text{out}}\) are formulae defining energy inside and outside the model:
where \(I\left( x \right)\) is the pixel intensity in point \(x\), and \(u\) and \(v\) are mean pixel intensities inside and outside the model. The authors proved that the direction of the function (2) minimization can be determined from its gradient with respect to B-spline coefficients [22]:
where \(\varGamma\) is the model surface and
Finally, when running the segmentation algorithm, model shape is modified iteratively by B-spline coefficient recalculation for minimization of EI:
where \(t\) is an iteration number and \(\lambda\) is a step scaling factor defined by the user as one of the method parameters. The above formulas express the global idea of the BEAS algorithm. Authors have also presented a localized version of the formulas, which take into account only pixel values in the neighborhood of the model nodes. This makes an algorithm more precise and faster.
The model description is extended in this work by the features analogous to mechanical tensile strength and bending resistance. These additional elements are called internal model energy in what follows.
According to the model definition (1), surface nodes can move only along lines defined by their angle coordinates, so model stretching corresponds to increasing the difference in \(r\) values for the neighboring nodes. We propose the following formulation of the energy Es resulting from model stretching:
where \(\alpha \left( x \right)\) is a constant corresponding to model stiffness in the point \(x\). Energy for a node defined by the angle coordinates can be expressed as the superposition of the finite differences of \(\psi \left( x \right)\) in both directions \(\left( {\left( {r\left( {\varphi ,\theta } \right) - r\left( {\varphi - h,\theta } \right)} \right)/h} \right)^{2}\) and \(\left( {\left( {r\left( {\varphi ,\theta } \right) - r\left( {\varphi ,\theta - h} \right)} \right)/h} \right)^{2}\) [29]. Minimization direction and node speed are then determined from the energy gradient:
Stretching energy defined in this way will be minimal, when the model takes the shape of a sphere.
The definition of the bending energy EB is expressed using a second derivative of the model function:
where \(\beta \left( x \right)\) is a constant corresponding to model bending resistance in point \(x\). Energy for the node defined by the angle coordinates can be expressed as the superposition of the second-order finite differences of \(\psi \left( x \right)\).
Analogous to (8), direction and speed of model adjustments leading to minimal energy \(E_{\text{E}}\) are determined from fourth derivative of the model with respect to its parameters (9):
Overall energy of the proposed model contains the following components: image energy defined in BEAS method \(E_{\text{I}}\) (2), stretching energy \(E_{\text{S}}\) (7) and bending energy \(E_{\text{B}}\) (9):
Direction of the energy minimization and speed of the model nodes is the resultant of all component energy gradients:
where \(\alpha \left( x \right)\), \(\beta \left( x \right)\) and \(\gamma \left( x \right)\) weigh the influence of the different components on the overall energy.
Segmentation
The segmentation algorithm works similarly as the one presented in [22]. In each iteration, it calculates the overall model energy as well as direction and speed of modification of the model. Modification vectors for all model nodes are multiplied by a step factor \(\lambda\) before being applied. If a step leads to energy decrease, \(\lambda\) is multiplied by a speed increase factor \(\lambda_{\text{d}}\) and the algorithm moves to the next iteration step. If the step leads to energy increase, the model modification is not applied and \(\lambda\) is divided by a speed decrease factor \(\lambda_{\text{z}}\). The algorithm is stopped either after a maximal number \(\left( {\hbox{max} I} \right)\) of iterations has been reached, or if the energy could not be reduced by defined number \(\left( {\hbox{max} I_{\text{z}} } \right) {\text{of trials}}\). To balance the role of the different energy components, each of them is normalized to the range [0,1] using the following formula:
where \(E_{k{\min} }\) and \(E_{k{\max} }\) are minimum and maximum values of the energy across the whole model. Taking (13) into consideration, Eq. (11) gets the following form:
The model and algorithm have been implemented in the Java programming language within Pmod PCARD tool [25]. As the left ventricular wall needs to be segmented, the method is applied to epicardium and endocardium segmentation separately. The space between these segments is considered as the myocardium segment. The implementation puts both epicardial and endocardial models in the same coordinates system, i.e. point \(r = 0\) is common for both of them. It also supports mutual restrictions of the models, preventing one model from moving too far from another one. In practice, endocardium segmentation is performed first. In the subsequent epicardial segmentation, the r coordinates of the model are restricted to a physiologically justified wall range. This restriction is required because the epicardial wall is often not well delineated due to signal from neighboring organs (e.g. liver), or due to lack of signal in regions with abnormal perfusion (e.g. because of infarct scar).
Parameters describing contribution of different types of energy to the overall value (α, β, γ) were calculated using the routine based on the Powell’s method of function minimization [30]. To find the best fitting parameters, a series of segmentations with α, β and γ values changing within the range [0,1] were performed and their results were compared to the reference segments outlined manually. Quality of the segmentation was quantified using Dice index [31] between the found myocardium segment and a reference one. The minimized cost function was defined as the one’s complement of the Dice index, i.e. 1-Dice index. The starting point of the minimization algorithm was set to α = 0.5, β = 0.5 and γ = 1 with the default step for all three values 0.1. The calculated energy contribution parameter values as well as the other method parameters are listed in Table 1.
The proposed segmentation method requires an initial model definition. In the PCARD tool a separate algorithm obtains it. It requires a priori setting of parameters describing the left ventricle approximately: long axis length, base radius and wall thickness. Furthermore, it requires that the data are resliced to short-axis orientation, with the heart long axis along the z-direction. The wall approximations are obtained by tracing rays from the long axis and analyzing the uptake profiles along them. An example of the resulting epi- and endocardial segments outlined on gated PET data averaged across all gates is shown in Fig. 1. They form the input for the energy-based segmentation described above. An example of the results applied to human gated PET data (eight bins) is shown in Fig. 2.
In PET data, the valve that closes the ventricular base is not visible. Therefore, additional post processing is required to approximate the valve plane. The following algorithm was implemented. First a mask of the myocardium is obtained using the Otsu threshold determined from the pixel histogram [32]. Having the data oriented in the short-axis projection, the center of mass in each slice between the apical and the basal plane is determined from the mask. From each center, rays are cast at equal angles. For each angle, the most basal slice is determined, for which the ray intersects with the mask. It is assumed that the intersection point in the middle of the mask wall belongs to the valve plane (valve point). The valve points in all angular directions provide an approximation of the valve plane. Hence, a plane is fitted to all valve points. An example of the initial segmentation after trimming by the fitted valve plane is shown in Fig. 3.
Validation
To validate the new segmentation method, clinical as well as phantom data were processed and the results compared to those obtined using two software packages widely used in the clinical practice, QGS—Quantitative Gated SPECT, Cedars-Sinai Medical Center [26], and Emory Cardiac Toolbox (ECTb), Emory University Medical Center [27].
The new method was incorporated in the Pmod PCARDP Tool, version 3.902. It represents a localized version [19] calculating image energy \(E_{\text{I}}\) (2) only in a cubic neighborhood of the current node.
The phantom data obtained during acquisition of two left ventricle phantoms of a known volume were analyzed. The phantoms were: an anthropomorphic phantom (Heart/Thorax Phantom, RSD Inc.) and a simplified isolated ventricular phantom Cardiac Insert Model ECT/CAR/I (BIODEX Medical Systems). The first phantom will be further referred to as the phantom A and the latter will be referred to as the phantom B.
The phantom A was filled with 18FDG, using a total activity of 37 MBq (1 mCi). To reproduce the proportions of activity usually observed in a human study, the heart wall (including the three defects provided by the producer to simulate perfusion defects) was filled with a concentration similar to the liver, whilst the concentration of the cardiac cavities (atria and ventricles) was one-fourth and that of the thorax one-tenth; finally the lungs were kept free of activity. After completion of the filling procedure, the phantom was laid on the PET/CT Gemini TOF scanner couch and the acquisition was performed. A CT study for attenuation correction was collected and then the data were acquired using the gated PET protocol, with the electrodes connected to a volunteer outside the scanner room to simulate the ECG signal for gating. The gated PET acquisition was reconstructed using the iterative method 3D-RAMLA (45 slices, matrix 144 × 144, voxel dimensions: 4 × 4 × 4 mm) with two iterations and three subsets, after decay, attenuation, random, scatter and time of flight correction. Before the final reconstruction of the images, the alignment of CT and PET images was controlled and corrected if necessary. Since the producer provides just the total volume of the cardiac cavities, the measurement of the left ventricular volume was performed by weighing the empty heart insert with a high precision balance, and then repeating the measurement after having slowly filled the sole left ventricular cavity with colored water through the proper injection hole, taking care of avoiding the overflow of the water in the nearby right ventricular cavity that communicates with the left one above the septum. The resulting volume was 75 ml.
The wall cavity of the phantom B was filled with 185 MBq (5 mCi) of 18FDG, and the images were acquired with a standard gated PET protocol for 10 min on a Discovery 710 PET scanner (GE) and then iteratively reconstructed with VUE point HD (18 subsets, two iterations). The ventricular cavity volume, reported by the producer to be about 60 ml, was measured as above described and found to be exactly 63 ml.
Besides the phantom data, a patient population of 60 patients, who had undergone cardiac PET scanning in both stress and rest conditions, has been analyzed. In particular, there were 29 patients studied because of suspected microvascular dysfunction in the setting of hypertrophic or dilated cardiomyopathy, or of secondary left ventricular hypertrophy, and 31 patients studied because of known or suspected coronary artery disease. Within the population, there were no patients with wide and severe resting or inducible perfusion defects.
Patients were studied with a standard resting and stress protocol using a PET/CT Gemini TOF scanner (Philips) as previously described [33]. Patients were submitted to CT imaging for attenuation correction, followed by resting study with administration of 370 MBq of 13NH3 in slow bolus and dynamic list mode acquisition lasting 9 min. Immediately thereafter, an eight-frame gated PET acquisition was started for additional 5 min. After 60 min, the stress study was performed using similar modalities, with the administration of 0.56 mg/kg of dipyridamole over 4 min. After 3 min of dipyridamole completion, 370 MBq of 13NH3 was injected and a second dynamic study acquired, again followed by a gated PET acquisition. No electrocardiographic or respiratory gating was applied. The study was then reconstructed using the same modalities described above for the phantom studies. Rest and stress end-systolic and end-diastolic volumes were obtained using the software mentioned above.
Results
Physical models
The phantom data were analyzed using QGS and PCARD using the proposed algorithm. The results compared to the real phantom volume are presented in Table 2. The mean relative error between QGS measurements and the physical volume is 6.66% and the mean relative error for PCARDP is 3.59%.
Clinical data
For logistic reasons, it was not possible to analyze all 60 patients using all three programs. The following pair-wise comparisons were feasible: PCARD compared to QGS (31 patients), PCARD compared to ECTb (18 patients), ECTb compared to QGS (18 patients). For each patient, two volumes (EDV, ESV) were calculated under stress and rest conditions, resulting in four observations for each patient. The results are shown as scatter plots in Figs. 4, 5 and 6. Each scatter plot presents the comparison of the measured LV volume between two programs as well as the result of the linear regression analysis performed using the least squares method. All pair-wise comparisons result in R2 higher than 0.8 and slope direction slightly smaller than 1.
Some points on the scatter plots are visibly more distant from the regression line than the points. All of them are the examples of not healthy patients. In Fig. 7, the data of a patient with a severe form of hypertrophic cardiomyopathy (HCM), already in a dilative phase, are shown. In Figs. 8 and 9, examples of patient data with septal hypertrophy are displayed. Although the patient suffers from a septal hypertrophy, the segmentation results are correct in the septal sectors. In Fig. 9, the lateral basal sector inaccuracy shows underestimation of the left ventricular volume.
Discussion
The presented scatter plots and linear regression results demonstrate a good correlation (\(R^{2} > 0.8\)) between results obtained by the new segmentation method and the reference results obtained by clinically used software. The linear regression equations suggest that there is a scaling factor of approximately 0.84 between PCARD and the reference results. It is unclear where this factor comes from. When processing phantom data, PCARD only slightly overestimated the true volumes, whereas QGS results were not consistent: for the simplified phantom QGS estimated the exact volume, whereas it overestimated the volume by 13% for the anthropomorphic phantom.
Correlation between QGS and ECTb is better than with PCARD. It is notable, however, that three different data sets had to be used for the comparisons, and manual interventions in PCARD were restricted to the short-axis reorientation. The scatter plots indicate some clear outliers. In practice, in the case of an erroneous segmentation result due to unusual heart geometry, the geometric parameters axis length and base radius of the method would be adjusted to improve the segmentation output. In Figs. 7, 8 and 9, the data correlating to the outlier point can be seen. The segmentation results in Figs. 7 and 8 are favorable for the proposed segmentation method.
Among the limitations of the proposed method, one should mention the deficiency in valve plane fitting algorithm if there is lack of perfusion in the heart basal layer either on lateral or septal side. Such regions with low signal level, resulting from e.g. infarction may result in detected valve plane shift towards heart mid layer and in LV volume underestimation.
Despite existence of the scaling factor for absolute volume calculations, the proposed method can be applied for left ventricular gated PET analysis. Phantom experiments show that the volume values obtained using PCARD are in agreement with the real values.
This study was not conceived for assessing the clinical implications of the proposed segmentation method nor to evaluate possible advantages over other established approaches. However, since each of so-far implemented methodologies can face difficulties in specific patients, the availability of another segmentation method could be useful. The phantom measurements seem favorable for PCARD; however, due to the small number of measurements, the results are not conclusive.
The segmentation method is derived from the BEAS algorithm, dedicated for ultrasound imaging. Its development proposed in this paper made it appropriate for gated PET data. Further stiffness and elasticity parameters modifications may allow adjusting the method to other imaging modalities, such as magnetic resonance or computed tomography.
Conclusions
The analysis of gated PET clinical data by the proposed method results in volume measurements comparable to established methods. As the phantom experiments also demonstrate that the volume values correspond to the physical ones, it can be concluded, that the proposed segmentation method can be recognized as a good alternative for the reference methods.
References
Miller TR, Wallis JW, Landy BR, Gropler RJ, Sabharwal CL. Measurement of global and regional left ventricular function by cardiac PET. J Nucl Med. 1994;35(6):999–1005.
Bonow RO. Identification of viable myocardium. Circulation. 1996;94(11):2674–80.
Degrado T, Hanson M, Turkington T, Delong D, Brezinski D, Vallee J, et al. Estimation of myocardial blood flow for longitudinal studies with 13N-labeled ammonia and positron emission tomography1. J Nucl Cardiol. 1996;3(6):494–507.
Choi Y, Sung-Cheng H, Hawkins RA, Kim JY, et al. Quantification of myocardial blood flow using (13N)-ammonia and PET: comparison of tracer models. J Nucl Med. 1999;40(6):1045.
Hutchins GD, Schwaiger M, Rosenspire KC, Krivokapich J, Schelbert H, Kuhl DE. Noninvasive quantification of regional blood flow in the human heart using N-13 ammonia and dynamic positron emission tomographic imaging. J Am Coll Cardiol. 1990;15(5):1032–42.
van der Hoff J, Burchert W, Borner AR, Fricke H, Kuhnel G, Meyer GJ, et al. [1-(11)C] acetate as a quantitative perfusion tracer in myocardial PET. J Nucl Med. 2001;42(8):1174–82.
Werys K, Blaszczyk L, Kubik A, Bogorodzki P. Gabor-filter based longitudinal strain estimation from tagged magnetic resonance imaging. In: Bilski P, Guerreiro F, editors. Computer Systems for Healthcare and Medicine. River Publisher Series in Information Science and Technology; 2015. p. 187–91.
Cygan S, Werys K, Błaszczyk Ł, Kubik T, Kałużyński K. Left ventricle phantom and experimental setup for MRI and echocardiography – Preliminary results of data acquisitions. Biocybern Biomed Eng. 2014;34(1):19–24.
Haidekker MA. Advanced biomedical image analysis. New Jersey: Wiley; 2011.
Kass M, Witkin A, Terzopoulos D. Snakes: active contour models. Int J Comput Vis. 1988;1(4):321–31.
Malladi R, Sethian JA, Vemuri BC. Topology-independent shape modeling scheme. In: Proceedings of SPIE 2031, geometric methods in computer vision II. 1993. https://doi.org/10.1117/12.146630.
Blake A, Isard M. Active contours: the application of techniques from graphics, vision, control theory and statistics to visual tracking of shapes in motion. London: Springer; 1998.
Paragios N, Deriche R. Geodesic active regions and level set methods for motion estimation and tracking. Comput Vis Image Underst. 2005;97(3):259–82.
Caselles V, Kimmel R, Sapiro G. Geodesic active contours. Int J Comput Vis. 1997;22(1):61–79.
Chan TF, Vese LA. Active contours without edges. IEEE Trans Image Process. 2001;10(2):266–77.
Yezzi A, Tsai A, Willsky A. A fully global approach to image segmentation via coupled curve evolution equations. J Vis Commun Image Represent. 2002;13(1–2):195–216.
Michailovich O, Rathi Y, Tannenbaum A. Image segmentation using active contours driven by the Bhattacharyya gradient flow. IEEE Trans Image Process. 2007;16(11):2787–801.
Rousson M, Deriche R. A variational framework for active and adaptative segmentation of vector valued images. In: Workshop on Motion and Video Computing, 2002. Proceedings, Orlando, USA; 2002, pp. 56–61.
Lankton S, Tannenbaum A. Localizing region-based active contours. IEEE Trans Image Process. 2008;17(11):2029–39.
Duan Q, Angelini ED, Laine AF. Surface function actives. J Vis Commun Image Represent. 2009;20(7):478–90.
Duan Q, Angelini ED, Laine AF. Real-time segmentation by active geometric functions. Comput Methods Programs Biomed. 2010;98(3):223–30.
Barbosa D, Dietenbeck T, Schaerer J, D’hooge J, Friboulet D, Bernard O. B-Spline explicit active surfaces: an efficient framework for real-time 3-d region-based segmentation. IEEE Trans Image Process. 2012;21(1):241–51.
Barbosa D, Heyde B, Dietenbeck T, Houle H, Friboulet D, Bernard O, et al. Quantification of left ventricular volume and global function using a fast automated segmentation tool: validation in a clinical setting. Int J Cardiovasc Imaging. 2013;29(2):309–16.
Unser M. Splines: a perfect fit for signal and image processing. IEEE Signal Process Mag. 1999;16(6):22–38.
PMOD Cardiac PET Modeling Tool, Pmod Technologies LLC. 2019. http://doc.pmod.com/pcardp/pcardp.htm. Accessed 28 Feb 2019.
Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su HT, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med Off Publ Soc Nucl Med. 1995;36(11):2138–47.
Garcia E, Faber T, Cooke C, Folks R, Chen J, Santana C. The increasing role of quantification in clinical nuclear cardiology: the Emory approach. J Nucl Cardiol. 2007;14(4):420–32.
Zhao H-K, Chan T, Merriman B, Osher S. A variational level set approach to multiphase motion. J Comput Phys. 1996;127(1):179–95.
Williams DJ, Shah M. A Fast algorithm for active contours and curvature estimation. CVGIP Image Underst. 1992;55(1):14–26.
Press WH, editor. Numerical recipes in C: the art of scientific computing. 2nd ed. Cambridge ; New York: Cambridge University Press; 1992. p. 994.
Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26(3):297–302.
Otsu N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern. 1979;9(1):62–6.
Sciagrà R, Calabretta R, Cipollini F, Passeri A, Castello A, Cecchi F, et al. Myocardial blood flow and left ventricular functional reserve in hypertrophic cardiomyopathy: a 13NH3 gated PET study. Eur J Nucl Med Mol Imaging. 2017;44(5):866–75.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Ethical approval
The clinical data were all obtained and processed exclusively by the Nuclear Medicine, ECBS Department, University of Florence. The data were anonymized and after processing tabulated using the related code. Only the anonymized tabulated data were transferred. The protocol approved by the competent Ethics Committee has the number SPE 16.270, approval date 23 March 2017.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Kubik, T., Kałużyński, K., Burger, C. et al. Novel 3D heart left ventricle muscle segmentation method for PET-gated protocol and its verification. Ann Nucl Med 33, 629–638 (2019). https://doi.org/10.1007/s12149-019-01373-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12149-019-01373-6