Abstract
Objective
Although simultaneous measurements of PET and magnetic resonance imaging (MRI) can provide interesting results in molecular imaging research, most of the combined systems are huge and animal handling in the system is not easy. To minimize these problems, we developed a compact integrated PET/MRI (iPET/MRI) system for small animals.
Methods
For the iPET/MRI system, a new MR-compatible PET and a permanent magnet open MRI were designed. In the MRI, a tunnel is opened at the yoke of the magnet. The position-sensitive photo-multiplier tubes (PSPMTs) of the MR-compatible PET are positioned at the back of the yoke where the magnetic field is sufficiently low. The scintillators for the PET system are positioned at the center of the MRI magnets, and the direction of the scintillation photons is changed by slanted light guides, and they are fed to the PSPMTs by 75 cm long optical fiber bundles. The PET detectors employed two types of LGSO crystals (1.9 mm × 2.2 mm × 6 mm and 7 mm) with different decay times (33 and 43 ns) for depth of interaction detection. Sixteen optical fiber-based block detectors are arranged in a 112 mm diameter ring.
Results
The transaxial field-of-view (FOV) of the PET system is ~80 mm, and the axial FOV is 21 mm which can be enlarged by the axial motion of the PET detector ring during MRI acquisition. The transaxial and axial resolutions at the center of the PET system was 2.9 and 2.4 mm FWHM, respectively. The absolute sensitivity was 1.5% at the center of the axial FOV. Phantom images revealed no artifact in either the PET or MRI images. We successfully obtained simultaneously measured small animal images using the iPET/MRI system.
Conclusion
The open geometry of the developed iPET/MRI facilitates easy accessibility to the subject. The iPET/MRI system appears to be a promising tool for molecular imaging research.
Similar content being viewed by others
References
Shao Y, Cherry SR, Farahani K, Slates R, Silverman RW, Meadors K, et al. Development of a PET detector system compatible with MRI/NMR system. IEEE Trans Nucl Sci. 1997;44:1167–71.
Shao Y, Cherry SR, Farahani K, Meadors K, Siegel S, Silverman RW, et al. Simultaneous PET and MR imaging. Phys Med Biol. 1997;42:1965–70.
Garlick PB, Marsden PK, Cave AC, Parkes HG, Slates R, Shao Y, et al. PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat heart. NMR Biomed. 1997;10:138–42.
Slate R, Cherry SR, Boutefnouchet A, Shao Y, Dahlbom M, Farahani K. Design of a small animal MR compatible PET scanner. IEEE Trans Nucl Sci. 1999;46:565–70.
Yamamoto S, Murayama H. A block detector for a multi-slice depth of interaction MR-compatible PET. IEEE Trans Nucl Sci. 2005;52:33–7.
Imaizumi M, Yamamoto S, Kawakami M, Aoki M, Sugiyama E, Kanai Y, et al. Simultaneous imaging of magnetic resonance imaging and positron emission tomography by means of MRI-compatible optic fiber based PET: a validation study in ex vivo rat brain. Jpn J Radiol. 2009;27:252–6.
Raylman RR, Majewski S, Lemieux SK, Velan SS, Kross B, Popov V, et al. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol. 2006;51:6371–9.
Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med. 2006;47:639–47.
Judenhofer MS, Wehrl HF, Newport DF, Catana C, Siegel SB, Becker M, et al. Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med. 2008;14:459–65.
Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multi-slice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med. 2006;47:1968–76.
Catana C, Procissi D, Wu Y, Judenhofer MS, Qi J, Pichler BJ, et al. Simultaneous in vivo positron emission tomography and magnetic resonance imaging. Proc Natl Acad Sci USA. 2008;105:3705–10.
Gilbert KM, Handler WB, Scholl TJ, Odegaard JW, Chronik BA. Design of field cycled magnetic resonance systems for small animal imaging. Phys Med Biol. 2006;51:2825–41.
Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR compatible gamma detectors. IEEE Trans Nucl Sci. 2003;50:1683–5.
Yamamoto S. Optimization of the integration time of pulse shape analysis for dual layer detector with different amount of Ce. Nucl Instrum Methods A. 2008;587:319–23.
Schlyer D, Vaska P, Tomasi D, Woody C, Solis-Najera S, Southekal S, et al. Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph. Nucl Sci Symp Conf Rec IEEE. 2006;4:2340–4.
Acknowledgments
This work was supported by the National Institute of Biomedical Innovation, Osaka, Japan.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yamamoto, S., Imaizumi, M., Kanai, Y. et al. Design and performance from an integrated PET/MRI system for small animals. Ann Nucl Med 24, 89–98 (2010). https://doi.org/10.1007/s12149-009-0333-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12149-009-0333-6