Introduction

Exhaled breath is a carrier of comprehensive information about the metabolic state of humans [1, 2] and the collection of samples significantly affects the diagnostic value of a measurement. Usually sampling is adapted to the requirements of a measurement and to the relevant medical condition for reproducible and reliable sampling. Significantly different findings of chemical composition of breath can be expected between samples obtained from the upper airways and from those taken at the end-tidal volume. Only one of these sampling methods will be used in clinical studies, depending on the application, e.g., the upper airway volume for Halitosis studies or the end-tidal volume for human metabolite studies.

The value of ion mobility spectrometry (IMS) coupled with a rapid pre-separation (muti-capillary columns or MCC) for the analyses of human breath has been described in detail recently with examples where MCC/IMS has been applied with medical benefits [3]. The method is characterized by sensitivities down to the pptV range and selectivity based on ionization properties of the analyses and can be made using only 8 mL samples of breath. Characteristic pattern for diseases can be observed as statistically valid from authentic breath samples obtained in clinical trials at the Lung Hospital in Hemer, Germany [47]. In other studies, serum concentrations of anaesthetics were determined from breath concentrations at the Department of Anaesthesiology, University of Göttingen, Germany and suggested the possibility for on-line anaesthesia control using MCC/IMS monitoring of exhaled breath [8].

In hospital studies from early developments with MCC/IMS, an operator controlled the sampling step and adapted the method to a subject’s physical condition. In a next development, e.g. for use during anaesthesia, the patients were ventilated and the sampling was automated from concentrations of respired carbon dioxide (CO2) [811]. Since these sensors are comparable expensive and slow responding with times of about 300 ms, already within the range of a single expiration at high respiratory rates, a comprehensive investigation was merited on the suitability of parameters and sensors (including CO2, humidity, flow and volume of the exhaled breath) in breath sampling.

Since breath measurement with MCC/IMS has been successful developed, applied clinically, and adapted for sensor controlled sampling, studies and comparisons of the sampling variables was made using this analytical method; however, findings here should be transferable to other comparable breath measurement methods. These may include on-line methods of proton transfer reaction/mass spectrometry [12, 13] or IMR/MS [14] and off-line methods such as gas chromatography/mass spectrometry based on sample pre-concentration with solid phase micro-extraction or polymer trap adsorption, [15] and batch sampling with Tedlar-bags [16].

Experimental details

Instrumentation

An inlet for sampling breath using a gas switching valve and ion mobility spectrometer coupled with a multi-capillary column has already been described and was used here without only minor modifications [3]. All experiments were made using a sample loop of 8 mL and the inlet was fitted with sensors to detect humidity, flow, and carbon dioxide (CO2), as summarised in Table 1.

Table 1 Sensors used for the detection of the control parameters humidity, flow and carbon dioxide (CO2)

Sampling procedures

The study with ventilated patients was approved by the ethics committee of the University of Göttingen (protocol No. 01/07/06). All patients gave written informed consent to participate in the investigation.

Operator controlled sampling

The subject breathes directly through the sample loop (see Fig. 1a). The operator chooses the portion of breath which should be analysed based on observation of the breathing pattern of the subject—which is generally the end-tidal breath—and manually initiates the analysis. This is accomplished when the volume of the sample loop is passed into the pre-separation column (see Fig. 1c).

Fig. 1
figure 1

Scheme of the sampling system used for the present study. The 6-way valve may be switched manually for operated controlled sampling as well as automatically for sensor controlled sampling. a) The subject exhales directly through the sample loop. b) The subject exhales through a mouthpiece and the sensor signal is used to decide when a sample is drawn by help of a pump at (4) from the mainstream through the sample loop. c) After switching the 6-way valve from “sampling” to “sample introduction”, the volume of the sample loop is introduced into the MCC at (1) and after pre-separation into the IMS by help of the carrier gas at (1)

Sensor controlled bypass sampling

Parameters used to control sampling were monitored continuously for the flow of breath using sensors for inspired or expired air in the main stream (see Fig. 1b). The data acquisition software enabled the setting of a threshold value for these control parameters and a sample is drawn from the mainstream through the sample loop only when a threshold was exceeded. Sampling was made for a pre-determined time duration which the sample loop was flushed with sample. After the sampling period, the 6-way valve was automatically switched, thus starting the measurement (see Fig. 1c). When the volume is used as control parameter, the data from the flow sensor is integrated over time.

Comparisons of sampling by direct, operator control and by flow sensor control were made using a single subject during a few minutes interval using two MCC/IMS instruments of the same design and dedicated inlet designs. The measurements were repeated by changing the control between the 2 MCC/IMS and by varying the sequence to avoid sampling biases or artefacts. The threshold for the flow sensor was set to half of the maximum, thus avoiding significant influence of the breath from the upper airways. In operator control of sampling, the operator sought to introduce the sample only for end-tidal breath.

Laboratory studies

A humid gas volume of air was prepared using 2-nonanone as a synthetic sample to probe sampling efficiency with an uncomplicated measurement using a proved calibration gas generator (HovaCAL 3834SC VOC, Inspire Analytical Systems GmbH, Frankfurt am Main, Germany, see [17]).

Results and discussion

Results using a humidified vapour sample with 2-nonanone to simulate breath in controlled concentrations provided a comparison of both sampling methods (see Fig. 2) with a chemically simple and stable sample. While response was observed with both methods of sampling and MCC-IMS, results were qualitatively identical, showing no influence of the sampling on retention time or drift time. However, strong variations in signal height or intensity of the response to 2-nonanone could be observed. A quantitative comparison of measurements from each sampling method is shown in Fig. 3 where the sensor controlled sampling resulted in higher response and better reproducibility than from manual sampling. The normalised concentrations showed a standard deviation of 66% for operator controlled sampling compared to 13% from flow controlled bypass sampling. Since both studies were operated under stable conditions by a single individual, the variations may be attributed to variations in manually sampling only, specifically with variation in the start of sampling. Though manual sampling is significantly worse than automated sampling, variations still greater may result in routine clinical applications where various analysts could introduce errors in judging a subjects physical condition and the moment to stop sample collection. As a consequence, a sensor controlled sampling should be used when quantitative and medically defensible results are desired.

Fig. 2
figure 2

MCC/IMS chromatogram of the synthetic breath sample, of 2-nonanone in humid air, (indicated with the cross line) derived from direct, operator controlled sampling (top) and from flow controlled bypass sampling (bottom)

Fig. 3
figure 3

Comparison of quantitative MCC/IMS response for 2-nonanone determination in humid air by direct sampling with an operator and with an automated sensor controlled bypass sampling using a flow sensor in this case and a constant concentration of 1 ppbV provided by HovaCAL

The response of sensors and parameters for monitoring respiration are shown in Fig. 4 where flow and humidity were sufficiently fast for real-time control of sampling, fitting close within the ranges of physiologic activity. The sensor for CO2 exhibited a response time of 300 ms and the signal shown in Fig. 4 was adjusted in time scale so comparisons can be made in slope and shapes of response to the other parameters. This type of delay is typical and sufficient for most of the available capnographic sensors since in general this concentration is not used to control any other analytical or sampling procedure during breathing; however, this sensor can be used only for sampling when the breathing frequency is <2 s. For higher frequency as quite usual for ventilated patients, other and faster sensors—at least few of such sensors are available—are necessary for an automated sampling inlet. The slope of all 4 parameters is presented in Fig. 4 and several conclusions can be made from comparisons of response patterns:

Fig. 4
figure 4

Comparison of the control parameters and of the influence of threshold values in the breath of a ventilated patient as an example. Inspiration, expiration and the end - tidal volume are indicated. The response delay of the CO2 sensor was corrected

Tidal flow

Using a unique threshold limit for the flow, expiration and inspiration can be differentiated, though as can be seen in Fig. 4, this will always lead to a predominant sampling of the air of the upper airways as the flow follows a steep increase followed by a slowly decrease. If the threshold is set to zero, the entire tidal volume will be sampled and will lead to a mixture of the end-tidal and upper airway breath. This could be solved by applying two different threshold limits for sampling “on” and “off”. Setting both of them in the decreasing slope of the curve would best enable end-tidal sampling; however, this requires elaborate programming of the data acquisition software and furthermore, a calibration to the particular subject is required in all cases.

Humidity

The humidity signal rises during expiration and thus enables a differentiation of the breath from the upper airways from the end-tidal volume. A tailing of response on the decrease during inspiration is significant and slow resulting in the collection of an increasing partition of inspiration air for lower threshold limits.

CO2

This parameter is an appropriate measure of the origin of the samples air as the concentration is almost independent on all other parameters such as tidal flow and from the background level as well. The slope is repeatable and the changes rapidly follow the tidal flow. Thus, a clear distinction of end-tidal breath and air from the upper airways is possible and enables the exclusive sampling of both partitions of breath.

Tidal volume

Tidal volume can be derived online from integration of flow rates and with suitable thresholds limits, the switching of sampling “on” and “off” enables an exclusive sampling of end-tidal breath as well. This is almost identical to the use of CO2. However, calibration for each particular subject is needed for tidal volume as significant physical differences may lead to different tidal flow and volume.

Conclusions

Approaches for human breath sampling through direct, operator controlled sampling and through sensor controlled bypass sampling showed that sensor controlled bypass sampling with a flow sensor exhibited higher reproducibility than the direct operator controlled sampling. Direct, operator controlled sampling showed detection of pattern of metabolites, as with the sensor control and only the quantitative performance was degraded.

A comparison of the simultaneously determined slope of control parameters humidity, tidal flow, tidal volume and CO2 concentration—all measured in human breath—showed that a flow sensor can be applied optimally for the sampling of breath from the upper airways only. During expiration, the slope follows a steep increase in the beginning followed by a more or less exponential decrease. Therefore, with a relatively high threshold value, the first part of the volume of the exhaled breath will be sampled.

In contrast, for analysis of end-tidal breath, the control using the volume or CO2 as parameters is required. The concentration in end-tidal breath ends up a plateau and can therefore easily be distinguished by the sensor from the first part of breath. However, using appropriate threshold values, the tidal volume—obtained from integration of the tidal flow—and CO2 obtain identical results.

The humidity follows the slope of inspiration and expiration with a delay and therefore is less suitable.

As a consequence, CO2 and the expired volume are the most suitable control parameters for end-tidal breath sampling with CO2 being the most appropriate parameter if the sensor signal is available after less then 50 ms as it is independent on the subjects physiognomy. Indeed, this is also the most expensive solution compared to expired volume.

Due to the response time of 300 ms the CO2 sensor used in the present study is not suitable for the breath analysis using MCC/IMS in all situations. Depending on the physical conditions or—in case of mechanical ventilation on the setup of the ventilator—the breathing frequency can rise easily to more than 1 Hz. Then the duration of expiration is less than 500 ms and cannot be resolved properly by a sensor with 300 ms. Such sensors with a response time in the range of 50 ms are not widely spread, however, some of such sensors are in fact available. Although all results were obtained using a MCC/IMS instrument, finding should be useful for other on-line analytical methods and for off-line sampling as well.