Ohmic Pulse-Heating Experiments
High-purity aluminum wires (99.999 at. pct) with a diameter of 0.5 mm (Catalogue No. AL501115, Advent) were investigated using an ohmic pulse-heating technique. The details of the experimental pulse-heating setup have already been described extensively in References 1 through 3.
The samples with about 60 mm length were treated with abrasive paper (grade 1200), cleaned with acetone, and subsequently resistively heated under N2 atmosphere at a pressure of 2.3 bar, starting at room temperature. A current pulse peaking at about 10 kA was discharged through the specimens and measured using an inductive coil (Model Number 3025, Pearson Electronics). At the same time, the voltage drop against common ground was measured using two Mo-foil voltage knives attached horizontally to the wire with subsequent voltage division. Due to the high heating rates of about 2 × 108 K s−1 needed to avoid a loss of contact, the experiments had a relatively short duration of about 35 μs.
To relate the time-resolved voltage drop and current behavior to a temperature, the surface radiance was simultaneously monitored by a Peltier-cooled IR pyrometer operating at a mean effective wavelength of λ
eff = 2315 nm with approximately 390 nm effective bandwidth. In order to in-situ calibrate the pyrometer, the surface radiance at the melting plateau, observed in the radiance-over-time development, was assigned to the melting temperature of Al, i.e., T
m = 933.47 K (660.32 °C).[4] Expressing the measured pyrometer signals by Planck’s law of radiation, the surface radiance J(T) measured at any time t can then be related to the observed radiance J(T
m) at the melting point by simply taking the ratio and solving for T:
$$ T = \frac{{c_{2} }}{{\lambda_{\text{eff}} \cdot \ln \left( {\frac{{J\left( {T_{\text{m}} } \right)}}{J(T)} \cdot \frac{{\varepsilon \left( {\lambda_{\text{eff}} ,T} \right)}}{{\varepsilon \left( {\lambda_{\text{eff}} ,T_{\text{m}} } \right)}} \cdot \left\{ {e^{{\frac{{c_{2} }}{{\lambda_{\text{eff}} \cdot T_{\text{m}} }}}} - 1} \right\} + 1} \right)}}, $$
(1)
where c
2 = 0.014388 m K is the second radiation constant. Due to the lack of spectral emissivity data at the extreme measuring wavelength, it was further assumed that the emissivity ε takes the value that is true at the melting point, i.e., \( \frac{{\varepsilon \left( {\lambda_{\text{eff}} ,T} \right)}}{{\varepsilon \left( {\lambda_{\text{eff}} ,T_{\text{m}} } \right)}} = 1 \), over the entire temperature range. Note that the assumption of constant emissivity is mostly feasible in the liquid phase, while this is usually not the case for the solid phase. The data presented in this publication thus focus on the liquid phase, while the solid phase data are also given as an overall estimate.
To obtain data for the density D(T), an adapted fast CCD system taking shadowgraph images of the backlit expanding wire was used (details in Reference 5). The shadow images with an exposure time of 600 ns were captured about 5 μs apart. The diameter d(T) as a function of temperature can then be obtained by evaluating the full-width-at-half-maximum in the calculated intensity profile of each image. By relating this temperature-dependent diameter to the diameter d
0 imaged before the experiment, the density can be derived using a room-temperature value of D
0 = 2.70 × 103 kg m−3.[6] The constraint \( \frac{d_{0}}{d[298\;{\text{K}}(25\;^\circ {\text{C}})]} = 1 \) and, therefore, D[298 K (25 °C)] = D
0 was applied for the solid phase data fitting.
$$ D\left( T \right) = D_{0} \cdot \left( {\frac{{{{d}}_{0} }}{{{{d}}\left( T \right)}}} \right)^{2} . $$
(2)
Note that using this approach, longitudinal expansion must be inhibited to a high degree. This is ensured by applying high heating rates, which have shown to result in an increased radial thermal expansion at the cost of absent longitudinal expansion.[7,8] Furthermore, it is crucial to achieve a nonmoving vertical liquid metal column in order to deduce precise diameters from the intensity profiles. Therefore, density measurements were conducted as separate experiments without voltage knives that can push the metal column. In addition, the wire length was shortened to about 40 mm to increase its stability.[8]
For nondensity experiments conducted with the pulse-heating apparatus, the electrical resistivity and the specific enthalpy can be derived. From the measured time-dependent voltage drop U(t) across the investigated length l, the time-dependent current I(t), and the diameter at room temperature d
0 (measured with a Keyence LS-7010 laser micrometer), the specific resistivity at initial geometry (IG) is obtained as
$$ \rho_{\text{IG}} \left( t \right) = \frac{U\left( t \right)}{I\left( t \right)} \cdot \frac{{d_{0}^{2} \pi }}{4 \cdot l}. $$
(3)
In order to correct for thermal volume expansion (VE), the radial expansion data obtained for density measurements are used. The correction is made under the assumption of absent longitudinal expansion, which is justified due to the experimental constraints, as described previously.
$$ \rho_{\text{VE}} \left( T \right) = \rho_{\text{IG}} \left( T \right) \cdot \left( {\frac{d\left( T \right)}{{d_{0} }}} \right)^{2} $$
(4)
Furthermore, the time-dependent specific enthalpy H(t) starting from room temperature can be derived by integrating the power over time and relating it to the sample mass m that is calculated from room-temperature density D
0 and diameter d
0, as well as the distance between the voltage knives l,
$$ H\left( t \right) = \frac{1}{m} \cdot \mathop \smallint \limits_{0}^{t} U\left( {t^{'} } \right) \cdot I\left( {t^{'} } \right){\text{d}}t'. $$
(5)
Due to the isobaric characteristics of the experiment, the specific heat capacity at constant pressure c
p can be evaluated from the slope of the H(T) curve:
$$ c_{\text{p}} = \left( {\frac{\partial H}{\partial T}} \right)_{\text{p}} .$$
(6)
In order to estimate thermal conductivity λ(T), the Wiedemann–Franz law was used. The Lorenz number was assumed constant at the theoretical value of L = 2.45 × 10−8 V2 K−2. This assumption is justified, as the phonon conductivity of Al is reported to be very small (for more information, see Klemens and Williams.[9])
$$ \lambda (T) = \frac{L \cdot T}{{\rho_{\text{VE}} (T)}}. $$
(7)
Thermal diffusivity a(T) can be estimated using thermal conductivity λ(T), specific heat capacity at constant pressure c
p, and density D(T). Note that applying the Wiedemann–Franz law to the respective equation and inserting Eqs. [4] and [2] yields an expression independent of thermal expansion.
$$ a(T) = \frac{\lambda (T)}{{c_{\text{p}} \cdot D(T)}} \equiv \frac{L \cdot T}{{c_{\text{p}} \cdot \rho_{\text{IG}} (T) \cdot D_{0} }}. $$
(8)
Thermal diffusivity, therefore, can be estimated with a relatively low uncertainty.[10]
Electromagnetic Levitation Experiments
An electromagnetic levitation (EML) setup was used to investigate the surface tension of liquid aluminum as well as to obtain additional reference data for the density of aluminum in the liquid phase. A detailed description of the EML setup was already part of precedent publications.[11,12,13,14]
In the EML experiments, small aluminum samples of high purity (99.999 at. pct) with a mass in the range of 100 to 140 mg were investigated. The samples were cut off from a high-purity aluminum rod with a diameter of 5.0 mm (Catalogue No. AL501907, Advent). Each sample was cleaned with acetone in an ultrasonic bath, followed by determining the mass on a precision balance.
A clean environment in the probe chamber was ensured by first evacuating the chamber to a pressure lower than 5 × 10−6 mbar and then flooding it close to atmospheric pressure (850 mbar) with high-purity gas mixtures of argon with 2.4 vol. pct hydrogen (AirLiquide Arcal10) and helium with 4 vol. pct hydrogen (AirLiquide custom gas mixture). The ratio of the gas mixtures was adapted throughout the experiment in order to control the heat dissipation from the sample and, thus, realize different sample temperatures. The idea of using hydrogen-enriched gas mixtures was to hinder aluminum oxide (alumina) formation and reduce already present aluminum oxide on the sample surface back to aluminum and vapor.[15]
The contactless temperature measurement was performed via a commercial NIR pyrometer (IMPAC IGA 6 Advanced, LumaSense), operating in the bandwidth of 1.45 to 1.80 μm. The temperature values recorded by the software (InfraWin 5.0.1.52) relate to the blackbody temperature (T
bb) of the detected radiance. In order to obtain the true temperature, the emissivity at a reference temperature was determined by assigning the recorded blackbody temperature at the solidification plateau of the sample under investigation to the real melting temperature of Al, i.e., T
m = 933.47 K (660.32 °C),[4] using formula [9], as given in Reference 16
$$ \varepsilon = { \exp }\left( {\frac{{c_{2} }}{\lambda } \cdot \left( {\frac{1}{{T_{\text{m}} }} - \frac{1}{{T_{\text{bb}} }}} \right)} \right) $$
(9)
with c
2 = 0.014388 m K, the second radiation constant. Assuming that emissivity ε does not change its value in the liquid phase over the entire temperature range, the true temperature (T
t) of each recorded blackbody temperature value can be recovered:[16]
$$ T_{\text{t}} = \left( {\frac{1}{{T_{\text{bb}} }} + \frac{\lambda }{{c_{2} }} \cdot \ln \varepsilon } \right)^{ - 1} . $$
(10)
In the EML experiment, a high-frequency current (≈380 kHz) is applied to the levitation coil that generates an inhomogeneous radio frequency electromagnetic field, inducing eddy currents in the sample material. These eddy currents, according to Lenz’s rule, generate an opposing electromagnetic field leading to a repulsive force that pushes the sample towards areas of lower field strength.[17]
Simultaneously, the ohmic losses of the eddy currents heat the sample to the liquid phase. Stable temperatures are obtained when heat dissipation by the atmosphere in the chamber and induced heating power are balanced.
When liquefied, oscillations of the sample around its equilibrium shape can be observed. Those variations in the radius of the sample can be described mathematically by spherical harmonics Y
m
l
. Lord Rayleigh deduced a relation between the frequency of the oscillation of a spherical droplet and the surface tension (γ) that acts as the restoring force, but it is only valid under nonterrestrial conditions where the sample is force free and not rotating.[18] The fundamental frequency, called Rayleigh frequency (v
R), is obtained for l = 2 and is fivefold degenerated (with M being the sample’s mass):
$$ \nu_{\text{R}}^{2} = \frac{8 \cdot \gamma }{3 \cdot \pi \cdot M}. $$
(11)
Under terrestrial conditions, those requirements are violated and the degeneracy is removed. Instead of one single oscillation frequency, up to five different oscillation modes can be observed for aspherical, rotating droplets. Cummings and Blackburn[19] derived a sum rule that recovers the original Rayleigh frequency from those five oscillation frequencies so that the surface tension can be calculated from the frequencies observed in the experiment by
$$ \gamma = \frac{3}{8} \cdot \pi \cdot M \cdot \left[ {\frac{1}{5} \cdot \left( {\nu_{2,0}^{2} + 2 \cdot \nu_{2,1}^{2} + 2 \cdot \nu_{2,2}^{2} } \right) - \nu_{\uptau }^{2} \cdot \left( {1.9 + 1.2 \cdot \left( {\frac{{z_{0} }}{a}} \right)^{2} } \right)} \right], $$
(12)
$$ z_{0} = \frac{g}{{2 \cdot \left( {2 \cdot \pi \cdot \nu_{\tau } } \right)^{2} }}, $$
(13)
$$ a = \sqrt[3]{{\frac{3 \cdot M}{4 \cdot \pi \cdot D}}}, $$
(14)
where z
0 is proportional to the relative position of the droplet in the field; g is the gravitational acceleration; and a is the radius of the sample, which can be calculated from the sample’s mass M and density D. The term \( \nu_{\uptau }\) in Eqs. [12] and [13] is the mean value of the squared translation frequencies in all three spatial dimensions denoted by v
i
with i = 1, 2, 3.
$$ \nu_{\uptau }^{2} = \frac{1}{3} \cdot \mathop \sum \limits_{i = 1}^{3} \nu_{i}^{2} . $$
(15)
The value v
2,0 in Eq. [12] is the observed frequency for the oscillation mode l = 2, m = 0. The terms v
2,1 and v
2,2 in Eq. [12] can be calculated by
$$ \nu_{2,m} = \frac{1}{2} \cdot \left( {\nu_{2, + m} ({\varOmega }) + \nu_{2, - m} ({\varOmega })} \right) $$
(16)
since the split of the observed frequencies for the oscillation modes l = 2, m = 1, 2 due to the rotation of the sample with frequency Ω denoted by ν
2,±m
(Ω) is symmetrical.
The sample movement and oscillations were recorded from the top taking advantage of the vertical axis symmetry. A high-speed camera recorded 4096 images per temperature point. The framerate of the camera, typically 600 frames per second (fps) at a resolution of 1024 px · 1024 px, had to be decreased to a value as low as 200 fps for low sample temperatures to allow shutter times as high as 5 ms. This adjustment was necessary in order to ensure a sufficient brightness on the recorded images, since the spectral radiance in the visible range declined to a very low level at temperatures close to the melting point.
The image series were analyzed using an edge-detection software that detects position and size of the sample on the images. The software generates a table that holds for each image the coordinates of the center of mass as well as the radii as a function of the azimuthal angle in steps of 5 deg. Using this table, a spectrum of the coordinates as well as of the radii can be obtained by applying a fast Fourier transform to the according time series.
In order to identify the five oscillation frequencies, additional spectra of the sum and difference of two arbitrary perpendicular radii were used, as presented in Reference 13.
For the density measurement, a series of 4100 shadowgraph images acquired from the side were recorded at a framerate of 120 fps for each temperature point. From these shadowgraph images, the mean shape of the sample was determined by a software using edge detection algorithms to determine the position and size of the sample in the images.
High-precision ball bearing spheres of known diameter levitated by an argon airflow through the cooling nozzle were used to relate the area of the sample on the shadowgraph images to a real (metric) quantity. Assuming vertical axis symmetry of the droplet, the volume is calculated and, with the known sample mass, the density is determined.