Abstract
In the context of the COVID-19 pandemic, wearable sensors are important for early detection of critical illness especially in COVID-19 outpatients. We sought to determine in this pilot study whether a wearable in-ear sensor for continuous body temperature and heart rate monitoring (Cosinuss company, Munich) is sufficiently accurate for body temperature and heart rate monitoring. Comparing with several anesthesiologic standard of care monitoring devices (urinary bladder and zero-heat flux thermometer and ECG), we evaluated the in-ear sensor during non-cardiac surgery (German Clinical Trials Register Reg.-No: DRKS00012848). Limits of Agreement (LoA) based on Bland–Altman analysis were used to study the agreement between the in-ear sensor and the reference methods. The estimated LoA of the Cosinuss One and bladder temperature monitoring were [-0.79, 0.49] °C (95% confidence intervals [-1.03, -0.65] (lower LoA) and [0.35, 0.73] (upper LoA)), and [-0.78, 0.34] °C (95% confidence intervals [-1.18, -0.59] (lower LoA) and [0.16, 0.74] (upper LoA)) of the Cosinuss One and zero-heat flux temperature monitoring. 89% and 79% of Cosinuss One temperature monitoring were within ± 0.5 °C limit of bladder and zero-heat flux monitoring, respectively. The estimated LoA of Cosinuss One and ECG heart rate monitoring were [-4.81, 4.27] BPM (95% confidence intervals [-5.09, -4.56] (lower LoA) and [4.01, 4.54] (upper LoA)). The proportion of detection differences within ± 2BPM was 84%. Body temperature and heart rate were reliably measured by the wearable in-ear sensor.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Facing the daunting coronavirus 2019 (COVID-19) pandemic in the last two years, with limited ICU beds, resources and health care personnel, it is essential to detect critical illness in COVID-19 patients before they begin to experience shortness of breath to prevent the pneumonia from progressing to a dangerous level [1]. Especially, there is a need of continuous monitoring of hospitalized or outpatient COVID-19 patients. Thus, defined attributes of an ideal monitoring should be non-invasiveness, accuracy during large changes in physiological conditions, independence of operator and technique, simplicity of use, and provide continuous monitoring [2].
Wearable sensors have been widely used for fitness tracking as well as daily life and may be used to continuously monitoring for the detection if worsening symptoms. Detecting patients with threatening respiratory deterioration, wearable sensors could bridge the gap between home isolation and normal ward, respectively, and the need for ICU. On the other hand, wearable sensors have the potential for monitoring healthy individuals who had exposure [3]. In this context, body temperature and heart rate play a crucial role for monitor early warnings signs. Body temperature rises during a viral infection due to immune activation. Therefore, body temperature monitoring has been widely accepted for detection of early signs of Covid-19 infection, and several wearable devices are available for monitoring of individuals who had the risk of exposure [4]. Due to a viral infection an increase of heart rate and change in pulse waveforms indicate physiological stress. Several studies showed the feasibility of wearable devices for continuous heart rate monitoring and detection of cardiac events [5, 6]. For instance, using a Fitbit or a Huami sensor, two studies showed the prediction of influenza-like illness by analyzing the resting heart rate [7, 8]. Studies have also indicated that COVID-19 infection are associated with cardiac events like arrhythmias, heart failure and myocarditis, for instance [9].
However, different devices are required for continuous monitoring vital signs and the challenge with these solutions is that the gadgets they use can only measure or track one or two symptoms of COVID-19. Additionally, there is a lack of data evaluating the available wearables devices with the standard of care monitoring devices. Passler et al. demanded studies with respect to measurement accuracy and precision [10]. Continuous ECG is the general accepted standard of care of intraoperative and ICU heart rate monitoring. Continuous body temperature measurement turns out to be more complicated, because there is an ongoing discussion of the most accurate body temperature monitoring site [11]. Although there are different monitoring sites the implantation of continuous non-invasive body temperature monitoring is difficult [12,13,14]. In general, there is a lack of data proving sufficient measurement accuracy of the wearable sensors compared to the standard of care monitoring devices.
Therefore, the purpose of this observational study is the evaluation of the new-developed in-ear sensor for continuous body temperature and heart rate monitoring and comparison with the standard of care body temperature and heart rate monitoring.
Methods
Study design
This observational pilot study was designed to evaluate the accuracy of the wearable in-ear sensor Cosinuss One (Cosinuss Company, Munich). To determine whether this wearable sensor is sufficiently accurate, we performed a comparison with anesthesiologic standard of care monitoring zero-heat flux temperature and urinal bladder monitoring, and heart rate measured by ECG, respectively. Minimalizing the strain on the study participants, especially the continuous body temperature measurement, we chose surgical procedures of non-critical ill patient for this study. Avoiding an influence on bladder temperature measurement by visceral surgery, non-cardiac thoracic procedures were chosen. In order to eliminate potential differences in temperature found between sexes due to female hormone cycle, we enrolled ten male patients in this study. All patients underwent thoracic (non-cardiac) operations including thoracotomy.
Ethical approval for this study was provided by the University of Luebeck Ethics committee (Reg.-Nr.: 16–324). The study was registered in the German Clinical Trials Register (Reg.-No: DRKS00012848). All patients gave their written consent to be included in this observational study. Patients’ demographic and surgical details were recorded.
After introduction of general anesthesia patients received a bladder catheter with temperature monitoring unit (Rüsch Sensor 400, Teleflex Medical Ltd., Ireland). Patients were placed in a lateral position with slight flexion of the torso. All patients received a perioperative warming system under the patient’s body and an additional warming unit on the lower extremities (MoeckWarmingSystem. Moeck & Moeck GmbH, Germany). After patient positioning, the Cosinuss One sensor was placed in the external ear canal. The zero-heat flux sensor (TCoreTM, Dräger, Germany) was positioned on the skin of the forehead. We confirmed adequate positioning at intercurrent controls throughout the study period. Initial five minutes of measurement were discarded to allow sensors to equilibrate, and measurements were restricted to the operation period.
Vesical bladder temperature, zero-heat flux, and ECG (InfinityTM Delta Monitor, Dräger, Germany) were recorded electronically in one-minute intervals by Dräger Infinity monitor. The in-ear temperature and heart rate were recorded electronically in one second intervals wirelessly using the Cosinuss Lab app. Operation theatre temperature was measured at the beginning and ending of each surgical procedure.
After surgery, we removed the zero-heat flux and in-ear sensors and examined the patients’ skin and external ear canal for injuries.
Wearable Cosinuss One in-ear temperature and heart rate sensor
The Cosinuss One in-ear sensor is developed and commercially distributed by Cosinuss Company, Munich, Germany. It is composed of a sensor element connected to an evaluation unit behind the ear via a connecting cable that is customized to the anatomy of the external ear (Fig. 1). The size of the sensor is 45 × 38 × 18mm and it weights 6.5 g (Table 1). A resistance thermometer is used for body temperature monitoring with an accuracy of ± 0.2 °C reported by the manufacturer. Using photoplethysmogram (PPG) with green light, the sensor measures the pulse rate in the external auditory canal by means of reflection measurement. Minimizing movements artefacts, the wearable sensor is available in different sizes. The accuracy is ± 1BPM for heart rate monitoring, reported by the manufacturer. The sensor probe is placed in the external auditory canal. The tip of the sensor probe does not touch the tympanic membrane. Recorded data are sent in real-time from the sensor unit to the Cosinuss Lab app wirelessly. After cleaning the Cosinuss One sensor with alcohol-free disinfection wipes, the device can be reused.
Data management and statistical analysis
All analyses were performed using the statistical software environment R 3.6.1 [15]. Since the Cosinuss One in-ear sensor measured several times per minute, we averaged the measurements in windows of one minute, centered around the time points of the Dräger® Monitor (zero-heat flux and bladder temperature and ECG heart rate monitoring) measurements. We set a priori acceptable agreement between Cosinuss One and bladder temperature and zero-heat-flux monitoring, respectively, to be 0.5 °C. We chose this limit due to normal human circadian temperature variation and occurrence of clinical complications starting from a temperature difference > 0.5 °C [2, 16, 17]. We set a priori acceptable agreement between Cosinuss One and ECG heart rate monitoring to be 2 BPM.
To compare the paired measurements of the devices, we estimated limits of agreement according to the Bland and Altman analysis method [18]. We used the true value varies extension of the Bland and Altman analysis for repeated measurements [19]. We estimated the bias (\(B\)) and the standard deviation of the differences of the paired measurements (\({\upsigma }_{d}\)) to obtain the limits of agreement \(\mathrm{LoA}=\mathrm{B}\pm 1.96\cdot {\upsigma }_{\mathrm{d}}\). The confidence interval for \(B\) was constructed by assuming a Student’s \(t\)-distribution with \(n-1\) degrees of freedom, where \(n\) is the total number of paired measurements. The confidence intervals for the upper and lower \(LoA\) were based on a modified mean sum of squares to estimate the between-subject variance proposed by Thomas & Hultquist [20], if the ratio of between-subject variance and total variance is less than 1/3 as suggested by Olofsen et al. [19]. Construction of confidence intervals was performed by the method of variance estimates recovery (MOVER) [21, 22]. Further details on the Bland–Altman analysis can be found in the Supplement.
In addition, we followed an alternative approach based on linear mixed models to estimate limits of agreement, concordance correlation coefficient and the total deviation index while allowing for linear trends of the differences of the paired measurements [23]. Details on this method and the respective results can be found in the Supplement.
Results
Patients’ characteristics
In order to eliminate bias of body temperature between sexes, the Cosinuss One was evaluated in ten male patients undergoing non-cardiac thoracic surgical procedures (8 Lobectomy, 1 Pneumonectomy, 1 Pleurectomy + HITHOC). Patients and surgical characteristics are shown in Table 2. Overall surgical time was 190 ± 102 min (range: 102 – 455 min). The wearable sensor was well tolerated in all patients. After surgery and in-ear sensor removal examination of the outer ear canal and auricle showed no tissue damage. None the patients complained any pain of the ear after the surgery.
Temperature monitoring
Temperature data were obtained from all patients. In one case zero-heat-flux temperature monitoring was lost due to storage failure of the Dräger monitor. Zero-heat-flux (TB) and Cosinuss One (TC) temperature sensor needed a short time of equilibration. Bladder temperature monitoring (TU) showed a period of decrease of temperature during the first minutes of surgery and more slowly temperature increase than zero-heat-flux and Cosinuss One monitoring (Fig. 2). A total of 1423 pairs of TC and TU, and 1288 pairs of TC and TB were collected, ranging from 34.8 °C to 38.2 °C.
Cosinuss One (TC) vs. bladder temperature (TU) monitoring
The Bland–Altman analysis (Table 3) estimated an average bias of -0.15 °C (values of TU are lower on average than TC temperature readings, 95% confidence interval [-0.3, -0.002]) between TU and TC, suggesting a significant difference on average. 95% of the readings are in the range of [-0.79, 0.49] (LoA) with 95% confidence intervals [-1.03, -0.65] (lower LoA) and [0.35, 0.73] (upper LoA). The range between the lower confidence bound of the lower LoA and the upper confidence bound of the upper LoA is 1.76 °C, greater than the pre-defined acceptable limit of agreement of ± 0.5 °C (Fig. 4A and Table 3). 89% of measurement pairs have an absolute difference ≤ 0.5 °C. The within-subject variance is estimated as 0.08 ± 0.003, higher than 0.03 ± 0.01, the between-subject variance. These results were supported by the alternative analysis (results shown in Supplement).
The diagnostic plots (Fig. 5 in Supplement) show no major deviations from the underlying assumptions (normality of differences, constant variability). However, the higher absolute differences for higher temperatures indicate a linear trend of the bias. Thus, the LoA based on the Bland–Altman analysis may not be adequate.
Cosinuss One (TC) vs. Dräger® TCore (TB) temperature monitoring
The Bland–Altman (Table 3) analysis estimates an average bias of -0.22 °C (values of TB are lower on average than TCtemperature readings, 95% confidence interval [-0.40, -0.04]) between TB and TC, suggesting a significant difference on average. 95% of the readings are in the range of [-0.78, 0.34] (LoA) with 95% confidence intervals [-1.18, -0.59] (lower LoA) and [0.16, 0.74] (upper LoA). The range between the lower confidence bound of the lower LoA and the upper confidence bound of the upper LoA is 1.92, greater than the pre-defined acceptable limit of agreement of ± 0.5 °C (see Fig. 4B and Table 3). 79% of measurement pairs have an absolute difference ≤ 0.5 °C. The within-subject variance is estimated as 0.03 ± 0.001, lower than 0.05 ± 0.03, the between-subject variance. These results were supported by the alternative analysis (results shown in Supplement).
Again, the diagnostic plots (Fig. 6 in Supplement) show no major deviations from the underlying assumptions (normality of differences, constant variability). However, the higher absolute differences for higher temperatures indicate a linear trend of the bias. Thus, the LoA based on the Bland–Altman analysis may not be adequate.
Heart rate
Cosinuss One (HC) vs. ECG (HD) heart rate monitoring
There were between 48 and 456 paired measurements over time by the Cosinuss One (HC) and the ECG heart rate monitoring (HD) on each sample (Fig. 3), ranging from 43 to 96 BPM. The Bland–Altman analysis (Table 3) estimates an average bias of -0.27 BPM (values of Dräger Monitor are lower on average than Cosinuss Sensor readings, 95% confidence interval [-0.51, -0.03]) between Dräger Monitor and Cosinuss Sensor, suggesting a significant difference on average. 95% of the readings are in the range of [-4.81, 4.27] (LoA) with 95% confidence intervals [-5.09, -4.56] (lower LoA) and [4.01, 4.54] (upper LoA). The range between the lower confidence bound of the lower LoA and the upper confidence bound of the upper LoA is 9.63, greater than the pre-defined acceptable limit of agreement of 2 BPM (Fig. 4C and Table 3). 84% of measurement pairs have an absolute difference ≤ 2 BPM. The within-subject variance is estimated as 5.30 ± 0.18, higher than 0.06 ± 0.05, the between-subject variance. These results were supported by the alternative analysis (results shown in Supplement).
The diagnostic plots (Fig. 7 in Supplement) show some deviation from the underlying assumption of normality of differences. Thus, the LoA based on the Bland–Altman analysis may not be adequate.
Discussion
In this study, we were able to show that the wearable in-ear sensor offers reliable continuous body temperature and heart rate monitoring. Focusing on comparable monitoring sites and reducing patients’ stress, we will discuss the study results in the context of intraoperative body temperature and heart rate monitoring as well.
Observing body temperature by bladder, zero-heat flux and Cosinuss One sensor monitoring, three aspects were noticeable. First, the Cosinuss One enabled a reliable continuous measurement of body temperature. We could not detect outliers of temperature monitoring. Second, we found an inter-subject variation of temperature curves measured by all three devices. This inter-subject variation can be assumed by circadian rhythm, individual thermoregulation, and impairment of thermoregulatory control due to anesthetic drugs [24, 25]. Sessler reviewed that in the first period after induction of general anesthesia body temperature decreased significantly due to anesthetic-induced vasodilation [24]. Accordingly, in our observation, bladder temperature monitoring revealed a decrease at the beginning of surgery. We assume that all patients had a decrease of body temperature after induction of general anesthesia which stopped after warming patient’s body. However, bladder temperature showed a lag of responsiveness. A review about bladder temperature monitoring by Fallis supports our assumption [26]. However, this site of temperature monitoring is an useful indicator of total body rewarming [27]. Warming all patients during surgery, temperature monitoring of all sensors showed an increase of body temperature. Last, due to the kind of temperature measurement, zero-heat flux and in-ear temperature monitoring needed an equilibration.
The limits of agreement of Cosinuss One vs. bladder temperature monitoring and Cosinuss One vs. zero-heat flux temperature monitoring, respectively, were greater than the a priori acceptable limit of 0.5 °C. A closer look reveals that TB and TU also vary by more than 0.5 °C (see supplement). Comparing zero-heat flux temperature monitoring to urinary or pulmonary artery catheter temperature monitoring site, several studies also reported a temperature difference > 0.5 °C. In these studies, the authors argued that a high proportion of differences supported a sufficient agreement of the sensors [28,29,30]. In this respect, it must be considered that different temperature measuring methods and different measurement sites are also included in the comparison. As a result, the temperature site remains often part of the discussion in the literature. In order to establish a classification of the non-invasive sensors, we have used two non-invasive or minimally invasive measuring methods and locations.
Selecting bladder and zero-heat flux thermometry for reference temperature monitoring sites, we focused on the avoidance of patients’ stress and comparable monitoring sites. Zero-heat flux temperature monitoring was well tolerated by the study participants [29]. Every patient per protocol got a bladder catheter with temperature monitoring undergoing extended thorax surgery (e.g., lobectomy). Thus, there was no additional stress for the patients. Several studies comparing zero-heat flux and bladder temperature monitoring to invasive methods (e.g., pulmonary artery catheter, esophageal probe, arterial catheter) showed comparable results of body temperature [2, 26, 29, 30]. The advantage of bladder temperature monitoring during thoracic surgery is an intact abdomen. There is no impairment of measurement. In contrast, esophageal body temperature monitoring might be impaired by open chest and lavage during surgery [31].
In the light of the above, the inaccuracy of the sensors has to be considered in addition to the a priori limit. Additionally, in 89% of measurements, in-ear temperature monitoring was within 0.5 °C of bladder temperature monitoring. 78.87% of measurements were within the priori limit of 0.5 °C of zero-heat flux monitoring. Considering that bladder temperature monitoring site is more valid than zero-heat flux as body core temperature monitoring site, our results appear to reasonably estimate body temperature.
Our results show a reliable agreement of in-ear heart rate monitoring and ECG. ECG is the gold standard of heart rate monitoring during anesthesia, surgery, and intensive care. Several studies showed a good agreement between PPG in-ear heart rate monitoring and ECG [32, 33]. Using green spectrum of light, the Cosinuss One sensor receives a stronger PGG signal. Passler et al. showed good levels of agreement between Cosinuss One and ECG heart rate monitoring [10]. Additionally, several studies showed that wearable technologies enabled detection of cardiac events [34, 35]. Breteler et al. concluded in their study that wearable continuous monitoring may have the potential to contribute to early recognition of physiological decline in high-risk patients [36].
Our study has a few limitations. First, we enrolled a small number of patients. The University of Luebeck Ethics committee only allowed to evaluate ten male patients. A strength, on the other hand, is that each patient was analyzed in a standardized set-up due to the surgical and anesthesia standards for minimum 45 min, yielding up to 1423 pairs of body temperature and 456 pairs of heart rate monitoring, respectively. Second, we did not evaluate the wearable sensor during rapid body temperature changes. Several studies showed that there is a greater bias of zero-heat flux and bladder temperature monitoring during rapid body temperatures changes (e.g., hypothermia during cardiopulmonary bypass, HIPEC) [37, 38]. Both reference temperature monitoring sites are not designed for rapid body temperature changes monitoring. Second, the in-ear sensor measured body temperature and heart rate once a second. Bladder and zero-heat flux temperature monitoring were recorded only once a minute due to manufacture determined monitoring setting. There might be a greater bias between the in-ear sensor and bladder and zero-heat flux temperature monitoring, respectively, because we estimated the minute means of body temperature and heart rate recorded by the in-ear sensor. Finally, we compared the in-ear sensor with the ECG. But each ECG signal has not be converted into a pulse wave during an extra-systole.
The advantage of this wearable sensor is the continuous body temperature and heart rate monitoring. It is well tolerated by patients and offers a non-invasive monitoring, especially for outpatient monitoring. The advantage of wearable sensors is monitoring clinical parameters without interfering with the current activity. Body temperature and heart rate monitoring are two cornerstones of detection of critical illness in COVID-19 patients. Oxygen saturation is the third cornerstone completing the wearable non-invasive monitoring. All three parameters are important to detect the silent hypoxia in Covid-19 pneumonia and the shift from non-critical illness to critical illness. Due to the COVID-19 pandemic several studies evaluate the application of wearable non-invasive devices [1, 39]. The next generation of the Cosinuss sensors will allow for monitor oxygen saturation (Cosinuss Two sensor). A clinical study initiated by Schmidt et al. at the Klinikum rechts der Isar, Munich, investigates the Cosinuss Two sensor monitoring in COVID-19 outpatients (Tele-COVID study – Remote Patient Monitoring for Covid-19) [40].
In conclusion, the Cosinuss One sensor enables a good validity of body temperature and heart rate monitoring used in the ear. Body temperature and heart rate were reliably measured by the wearable in-ear sensor compared to clinical standard of care monitoring devices. In this context, this observational study is one of the first with respect to measurement accuracy and precision of wearable sensors compared to standard and certificated clinical monitoring devices. Wearable sensors are promising technique that enables out-patient monitoring. The COVID-19 pandemic demonstrated the need of such monitoring devices for monitoring non-critical ill patients and detecting patients with threatening respiratory deterioration. In this regard, we think that guidelines and general standards for wearable sensors toned to be introduced.
Abbreviations
- BPM:
-
Beats per minute
- HITHOC:
-
Hyperthermic Intrathoracic Chemotherapy
- Hc :
-
Heart rate in-ear sensor
- HD :
-
Heart rate Dräger monitor (ECG)
- LoA:
-
Limits of Agreement
- PPG:
-
Photoplethysmogram
- TU :
-
Urinary bladder temperature
- TB :
-
Zero-heat flux brain temperature
- TC :
-
In-ear sensor temperature
References
Teo J (2020) Early Detection of Silent Hypoxia in Covid-19 Pneumonia Using Smartphone Pulse Oximetry. Journal of Medical Systems 44 (8):134. https://doi.org/10.1007/s10916-020-01587-6
Eshraghi Y, Nasr V, Parra-Sanchez I, Van Duren A, Botham M, Santoscoy T, Sessler DI (2014) An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients. Anesth Analg 119 (3):543-549. https://doi.org/10.1213/ANE.0000000000000319
Lukas H, Xu C, Yu Y, Gao W (2020) Emerging Telemedicine Tools for Remote COVID-19 Diagnosis, Monitoring, and Management. ACS Nano 14 (12):16180-16193. https://doi.org/10.1021/acsnano.0c08494
Channa A, Popescu N, Skibinska J, Burget R (2021) The Rise of Wearable Devices during the COVID-19 Pandemic: A Systematic Review. Sensors (Basel) 21 (17). https://doi.org/10.3390/s21175787
Jacobsen M, Dembek TA, Kobbe G, Gaidzik PW, Heinemann L (2020) Noninvasive Continuous Monitoring of Vital Signs With Wearables: Fit for Medical Use? J Diabetes Sci Technol:1932296820904947. https://doi.org/10.1177/1932296820904947
Tanwar G, Chauhan R, Singh M, Singh D (2020) Pre-Emption of Affliction Severity Using HRV Measurements from a Smart Wearable; Case-Study on SARS-Cov-2 Symptoms. Sensors (Basel) 20 (24). https://doi.org/10.3390/s20247068
Radin JM, Wineinger NE, Topol EJ, Steinhubl SR (2020) Harnessing wearable device data to improve state-level real-time surveillance of influenza-like illness in the USA: a population-based study. Lancet Digit Health 2 (2):e85-e93. https://doi.org/10.1016/S2589-7500(19)30222-5
Zhu G, Li J, Meng Z, Yu Y, Li Y, Tang X, Dong Y, Sun G, Zhou R, Wang H, Wang K, Huang W (2020) Learning from Large-Scale Wearable Device Data for Predicting the Epidemic Trend of COVID-19. Discrete Dynamics in Nature and Society 2020:6152041.
Malaty M, Kayes T, Amarasekera AT, Kodsi M, MacIntyre CR, Tan TC (2021) Incidence and treatment of arrhythmias secondary to coronavirus infection in humans: A systematic review. European journal of clinical investigation 51 (2):e13428. https://doi.org/10.1111/eci.13428
Passler S, Müller N, Senner V (2019) In-Ear Pulse Rate Measurement: A Valid Alternative to Heart Rate Derived from Electrocardiography? Sensors (Basel) 19 (17). https://doi.org/10.3390/s19173641
Sessler DI (2021) Perioperative Temperature Monitoring. Anesthesiology 134 (1):111-118.
Langham GE, Maheshwari A, Contrera K, You J, Mascha E, Sessler DI (2009) Noninvasive temperature monitoring in postanesthesia care units. Anesthesiology 111 (1):90-96. https://doi.org/10.1097/ALN.0b013e3181a864ca
National Collaborating Centre for Nursing and Supportive Care (publisher)(2008. 29; Addendum 65.1 Dec. 2016. https://www.nice.org.uk/) Hypothermia: prevention and management in adults having surgery, NICE Clinical Guideline 65.
Torossian A, Brauer A, Hocker J, Bein B, Wulf H, Horn EP (2015) Preventing inadvertent perioperative hypothermia. Dtsch Arztebl Int 112 (10):166-172. https://doi.org/10.3238/arztebl.2015.0166
R-Core-Team (2018) R: A language and environment for statistical computing. R Foundation for Statistical Computing (Publisher)
Sastre JA, Pascual MJ, Lopez T (2019) Evaluation of the novel non-invasive zero-heat-flux Tcore thermometer in cardiac surgical patients. J Clin Monit Comput 33 (1):165-172. https://doi.org/10.1007/s10877-018-0143-2
Winkler M, Akca O, Birkenberg B, Hetz H, Scheck T, Arkilic CF, Kabon B, Marker E, Grubl A, Czepan R, Greher M, Goll V, Gottsauner-Wolf F, Kurz A, Sessler DI (2000) Aggressive warming reduces blood loss during hip arthroplasty. Anesth Analg 91 (4):978-984
Bland JM, Altman DG (2007) Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 17:571-582
Olofsen E, Dahan A, Borsboom G, Drummond G (2015) Improvements in the application and reporting of advanced Bland–Altman methods of comparison. Journal of Clinical Monitoring and Computing 29 (1):127-139. https://doi.org/10.1007/s10877-014-9577-3
Thomas JD, Hultquist RA (1978) Interval Estimation for the Unbalanced Case of the One-Way Random Effects Model. The Annals of Statistics 6 (3):582-587.
Donner A, Zou G (2012) Closed-form confidence intervals for functions of the normal mean and standard deviation. Statistical Methods in Medical Research 21 (4):347-359. https://doi.org/10.1177/0962280210383082
Zou G (2013) Confidence interval estimation for the Bland–Altman limits of agreement with multiple observations per individual. Statistical Methods in Medical Research 22 (6):630-642. https://doi.org/10.1177/0962280211402548
Choudhary PK, Nagaraja HN (2017) Measurement Agreement. Wiley & Sons, Inc., Hoboken, USA.
Sessler DI (2016) Perioperative thermoregulation and heat balance. Lancet 387 (10038):2655-2664. https://doi.org/10.1016/S0140-6736(15)00981-2
Weinert D (2010) Circadian temperature variation and ageing. Ageing Res Rev 9 (1):51-60. https://doi.org/10.1016/j.arr.2009.07.003
Fallis WM (2002) Monitoring bladder temperatures in the OR. AORN J 76 (3):467-476, 481-466, 488-469
Ramsay JG, Ralley FE, Whalley DG, DelliColli P, Wynands JE (1985) Site of temperature monitoring and prediction of afterdrop after open heart surgery. Can Anaesth Soc J 32 (6):607-612
Schell-Chaple HM, Liu KD, Matthay MA, Puntillo KA (2018) Rectal and Bladder Temperatures vs Forehead Core Temperatures Measured With SpotOn Monitoring System. Am J Crit Care 27 (1):43-50. https://doi.org/10.4037/ajcc2018865
Boisson M, Alaux A, Kerforne T, Mimoz O, Debaene B, Dahyot-Fizelier C, Frasca D (2018) Intra-operative cutaneous temperature monitoring with zero-heat-flux technique (3M SpotOn) in comparison with oesophageal and arterial temperature: A prospective observational study. Eur J Anaesthesiol 35 (11):825-830. https://doi.org/10.1097/EJA.0000000000000822
Dahyot-Fizelier C, Lamarche S, Kerforne T, Benard T, Giraud B, Bellier R, Carise E, Frasca D, Mimoz O (2017) Accuracy of Zero-Heat-Flux Cutaneous Temperature in Intensive Care Adults. Crit Care Med 45 (7):e715-e717. https://doi.org/10.1097/CCM.0000000000002317
Russell SH, Freeman JW (1996) Comparison of bladder, oesophageal and pulmonary artery temperatures in major abdominal surgery. Anaesthesia 51 (4):338-340
Martin A, Voix J (2018) In-Ear Audio Wearable: Measurement of Heart and Breathing Rates for Health and Safety Monitoring. IEEE Trans Biomed Eng 65 (6):1256-1263. https://doi.org/10.1109/TBME.2017.2720463
He DD, Winokur ES, Sodini CG (2012) An ear-worn continuous ballistocardiogram (BCG) sensor for cardiovascular monitoring. Conference proceedings : Annual International Conference of the IEEE Engineering in Medicine and Biology Society IEEE Engineering in Medicine and Biology Society Annual Conference 2012:5030-5033. https://doi.org/10.1109/EMBC.2012.6347123
Inui T, Kohno H, Kawasaki Y, Matsuura K, Ueda H, Tamura Y, Watanabe M, Inage Y, Yakita Y, Wakabayashi Y, Matsumiya G (2020) Use of a Smart Watch for Early Detection of Paroxysmal Atrial Fibrillation: Validation Study. JMIR Cardio 4 (1):e14857. https://doi.org/10.2196/14857
Karmen CL, Reisfeld MA, McIntyre MK, Timmermans R, Frishman W (2019) The Clinical Value of Heart Rate Monitoring Using an Apple Watch. Cardiol Rev 27 (2):60-62. https://doi.org/10.1097/CRD.0000000000000243
Breteler MJMM, Huizinga E, van Loon K, Leenen LPH, Dohmen DAJ, Kalkman CJ, Blokhuis TJ (2018) Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. BMJ Open 8 (2):e020162. https://doi.org/10.1136/bmjopen-2017-020162
Kiya T, Yamakage M, Hayase T, Satoh J, Namiki A (2007) The usefulness of an earphone-type infrared tympanic thermometer for intraoperative core temperature monitoring. Anesth Analg 105 (6):1688-1692. https://doi.org/10.1213/01.ane.0000289639.87836.79
Rotello LC, Crawford L, Terndrup TE (1996) Comparison of infrared ear thermometer derived and equilibrated rectal temperatures in estimating pulmonary artery temperatures. Crit Care Med 24 (9):1501-1506
Klum M, Urban M, Tigges T, Pielmus AG, Feldheiser A, Schmitt T, Orglmeister R (2020) Wearable Cardiorespiratory Monitoring Employing a Multimodal Digital Patch Stethoscope: Estimation of ECG, PEP, LVETand Respiration Using a 55 mm Single-Lead ECG and Phonocardiogram. Sensors (Basel) 20 (7). https://doi.org/10.3390/s20072033
Cosinuss GmbH MG (2022) Telecovid – Remote Patient Monitoring for COVID-19. https://www.cosinuss.com/en/portfolio-items/study-on-remote-patient-monitoring-for-covid-19. Accessed 09/20 2022
Funding
Open Access funding enabled and organized by Projekt DEAL. This research received no specific Grant from any funding agency in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
DE contributed to conception and design, data acquisition, interpretation of data, and drafting and revision of article. DG and MK contributed to interpretation of data, drafting and revision of the article. All authors have provided final approval of the version submitted.
Corresponding author
Ethics declarations
Ethics approval
All procedures conducted in our study involving human participants were in accordance with the ethical standards of the institutional review board and with the Helsinki declaration and its amendments. Written informed consent was obtained from all patients. Ethical approval for this study was provided by the University of Luebeck Ethics committee (Reg.-Nr.: 16–324).
Informed consent
Informed consent was obtained from all individual participants included in the study.
Conflicts of interest
The authors have no known conflicts of interest associated with this study and there has been no significant financial support for this work that could have influenced its outcome.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Mobile & Wireless Health
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ellebrecht, D.B., Gola, D. & Kaschwich, M. Evaluation of a Wearable in-Ear Sensor for Temperature and Heart Rate Monitoring: A Pilot Study. J Med Syst 46, 91 (2022). https://doi.org/10.1007/s10916-022-01872-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10916-022-01872-6