Abstract
Background
The DiamondTemp ablation (DTA) system is a novel temperature-controlled irrigated radiofrequency (RF) ablation system that accurately measures tip-tissue temperatures for real-time power modulation. Lesion morphologies from longer RF durations with the DTA system have not been previously described. We sought to evaluate lesion characteristics of the DTA system when varying the application durations.
Methods
A bench model using porcine myocardium was used to deliver discrete lesions in a simulated clinical environment. The DTA system was power-limited at 50 W with temperature set-points of 50 °C and 60 °C (denoted Group_50 and Group_60). Application durations were randomized with a range of 5–120 s.
Results
In total, 280 applications were performed. Steam pops were observed in five applications: two applications at 90 s and three applications at 120 s. Lesion size (depth and maximum width) increased significantly with longer applications, until 60 s for both Group_50 and Group_60 (depth: 4.5 ± 1.2 mm and 5.6 ± 1.3 mm; maximum width: 9.3 ± 2.7mm and 11.2 ± 1.7mm, respectively). As lesions transition from resistive to conductive heating (longer than 10 s), the maximum width progressed in a sub-surface propagation. Using a “Time after Temperature 60 °C” (TaT60) analysis, depths of 2–3 mm occur in 0–5 s and depths plateau at 4.6 ± 0.8 mm between 20 and 30 s.
Conclusions
The DTA system rapidly creates wide lesions with lesion depth increasing over time with application durations up to 60 s. Using a TaT60 approach is a promising ablation guidance that would benefit from further investigation.
Similar content being viewed by others
1 Introduction
Over the past three decades, radiofrequency (RF) catheter ablation has become a prominent method of treating cardiac arrhythmias [1, 2]. Early research focused on the RF biophysics of lesion formation and firmly established the therapeutic temperature range between irreversible tissue damage (> 50 °C) and the prevention of interstitial steam formation (< 100 °C) [3,4,5,6,7,8,9]. Since this time, catheter designs have evolved. Particularly, open irrigation and the inclusion of force sensors have facilitated more efficient lesion formation and improved safety. While improvements have been made, lesion durability and arrhythmia recurrence remain challenging [10,11,12]. The safety improvements accompanying irrigated catheter designs resulted in a shift away from tissue temperature as the primary metric of effective lesion delivery. This catheter paradigm triggered iterative surrogates to facilitate feedback on energy transfer (i.e., contact force, stability), lesion durability (i.e., ablation index, lesion-size index), and procedural efficiency (i.e., high-power short-duration (HPSD), inter-lesion distance) [13,14,15,16,17,18,19].
From the continued need for procedural improvement, a novel catheter (ground-up design) focused on accurate temperature feedback was developed [20, 21]. Through externally located thermocouples, incorporation of industrial diamond, and a low irrigation rate, the DiamondTemp ablation (DTA) system (Medtronic Inc, Minnesota, MN) has established a tip-tissue interface temperature accuracy of 2–4 °C [22]. Recent clinical successes have been demonstrated for atrial arrhythmias with ablation parameters akin to HPSD [23,24,25,26,27]. However, the data for ventricular or thicker tissue applications using a similar approach are less understood [28, 29]. This study sought to evaluate lesion morphologies across a broad range of ablation parameters (i.e., temperature set-points and application duration) to better understand system performance for a range of tissue thicknesses.
2 Methods
2.1 Ablation system
The study was conducted using the DTA system FASTR RF generator (model CEDTG200), which has been previously described [21, 25]. In brief, the temperature-controlled system takes readings every 20ms from 6 external thermocouples; 3 located on the distal tip and 3 proximal to the RF electrode. Upon reaching the programmed temperature set-point (default 60 °C), the generator automatically modulates the power to maintain the tip-tissue temperature. Additionally, the catheter 4.1 mm split-tip design contains a chemical vapor deposition diamond network which allows for more efficient thermal transfer properties; ultimately allowing for a constant 8ml/min irrigation rate and accurate temperature measurements at the tip-tissue interface. The FASTR RF generator is the second-generation temperature-controlled system and incorporates an increased power modulation responsiveness, ramping to 50 W (default upper limit) within 1 s. In this experimental setup, the temperature set-point and the application duration were varied to create a range of lesions, and all other system properties were held constant, based on the device labeling.
2.2 Ex vivo model
The ex vivo setup consisted of a ~20 L tank filled with 10 L of normal saline, and salt or water was added to adjust to a floating catheter impedance of 100–120 Ω. A continuous-flow pump and heating circuit were placed in the tank to heat the solution to 37 °C. An indifferent electrode was placed on the opposite end of the tank from the tissue sample, at a distance between 8 and 12 inches.
The DTA system was set up in accordance with the device labeling. The power limit was set to 50 W with an irrigation rate of 8 ml/min during RF delivery. Power modulation is predicated upon the tip-tissue temperature reaching the programmed temperature set-point. Applications were grouped by setting the temperature set-point to 50 °C and 60 °C (Group_50 and Group_60, respectively). With a lower temperature set-point, power modulation is expected to trigger earlier, potentially impacting lesion morphologies. The application duration varied between pre-determined values ranging from 5 to 120 s. All applications were delivered in a perpendicular catheter orientation.
Freshly excised porcine cardiac tissue was obtained, and the right ventricles were sectioned for use. Prior to ablation, each section was prepped, placed in the tank, and allowed to acclimate for approximately 5 min. Ablations were delivered in a randomized order, leaving adequate space between applications for measurements. If a steam pop occurred, the catheter was repositioned to unablated tissue and the application was repeated, and two sequential steam pops resulted in a failed attempt.
2.3 Visualization and comparison of lesion data
The methods for visualization and measurement have been previously described [30,31,32]. In brief, upon completion of a given tissue sample, the tissue was placed in a warmed 2,3,5-triphenyltetrazolium chloride stain for lesion evaluation. The surface width and orthogonal length were measured for each lesion. All lesions were then dissected through their width and re-stained for cross-sectional measurements including overall lesion depth, maximum width, and depth at maximum width.
2.4 Statistics
Temperature data from the DTA system was evaluated for each individual application. The lesion dimensions were also categorized by their application settings. All continuous variables are represented as a mean ± standard deviation. Comparisons between groups were done using either an unpaired t-test or a one-way analysis of variance (ANOVA) for normally distributed variables. Post hoc pairwise comparison via Tukey’s test was performed in response to ANOVA significance, where a p-value of <0.05 was considered significant. Continuous variables were fit to regression models to describe lesion growth.
3 Results
DTA applications were delivered in a randomized order across a series of bench experiments. In total, 280 applications were attempted, 141 in Group_50 and 139 in Group_60. Five (1.8%) applications resulted in steam pop and were removed from this analysis due to application disruption and tissue sample distortion. Two 90-s and three 120-s applications in Group_60 resulted in steam pops at an average of 61.1 s. The applications in Group_50 presented average temperatures, powers, and impedance drops of 49.2 ± 1.0 °C, 30.3 ± 7.6 W, and 15 + 8 Ω, respectively. In the Group_60 cohort, the same measures were 57.9 ± 2.6 °C, 39.9 ± 7.5 W, and 18 ± 7 Ω, respectively.
3.1 50 °C temperature set-point applications
In Group_50 (Table 1, Fig. 1), lesion depths increased from 0.9 ± 1.0 mm at 5 s to 4.5 ± 1.2 mm at 60 s, with no additional significant increase for 90- or 120-s applications. Similarly, maximum width increased with ablation duration until 60 s, with no significant depth increases beyond a 60-s application. Surface width plateaued after 10 s (5.3 ± 1.6 mm), at which time, the depth (at maximum width) continued to increase from 0.3 ± 0.7 mm to 2.3 ± 1.0 mm at 120 s. As the application durations extended, the temperature set-point was reached allowing for more power titration. In 5-s applications, average power was 34.9 ± 7.9 W, whereas in 30- and 60-s applications, the power averaged 29.7 ± 4.4 W and 29.4 ± 7.4 W respectively.
3.2 60 °C temperature set-point applications
In Group_60 (Table 2, Fig. 1), lesion depths ranged from 1.9 ± 0.9 mm at 5 s to 5.6 ± 1.3 mm at 60 s, with no significant increase beyond a 60-s application. Additionally, it should be noted there was no significant change in depth between the 30-, 45-, and 60-s applications. However, significant increases were observed when comparing 30- and 45-s applications to 90- and 120-s applications. Surface width plateaued after 10 s (6.1 ± 1.3 mm), and at this time, the depth (at maximum width) continued to increase from 0.6 ± 0.7 mm to 2.9 ± 1.2 mm at 120 s. As the application durations extended, the temperature set-point was reached allowing more power titration, though to a lesser degree than in Group_50. In 5-s applications, the average power was 46.9 ± 3.9 W, whereas in 30- and 60-s applications, the power averaged 38.8 ± 6.8 W and 37.1 ± 5.1 W respectively.
3.3 Comparison of lesion morphologies for temperature set-points of 50 and 60 °C
Applications longer than 60 s in both Group_50 and Group_60 showed no marked increase in lesion morphology. Comparative analyses excluded the 90- and 120-s applications to focus on clinically relevant application durations. Applications in Group_60 generally produced deeper and wider lesions than those in Group_50. However, when comparing the lesion morphologies between Group_50 and Group_60, a pattern emerges. When application durations for Group_50 were twice the duration of Group_60, there were no significant differences in lesion depth and maximum width (Fig. 2a–c). This relationship remains consistent across the range of application durations: 10 s vs. 5 s, 30 s vs. 15 s, and 60 s vs. 30 s for Group_50 and Group_60. When adjusting the temperature set-point to 50 °C, the ability to deliver a comparable lesion took twice as long. In making this adjustment to the system, the average power and average temperature are also markedly different between Group_50 and Group_60.
3.4 Time after temperature 60 °C analysis
Additional analysis of the Group_60 data indicated a strong relationship with time after reaching a tip-tissue temperature of 60 °C (time after temperature 60 °C—denoted TaT60) and total lesion depth, independent of the total application duration (Fig. 3A). The concept of “time after temperature 60 °C” is illustrated in Fig. 3A. During this single 30-s application, a tip-tissue temperature of 60 °C was reached after 8 s from RF onset; thus, the measured TaT60 in this application was 22 s. When TaT60 measured under 5.0 s, the average depth was 2.4 ± 1.1 mm. Lesion depth exhibited a non-linear relationship and the rate of lesion growth slowed beyond 20–30 s TaT60 where depth averaged 4.6 ± 0.8 mm (n=19) (Fig. 3B). Depth measured 5.1 ± 1.2 (n=31) when TaT60 was between 30 and 60 s, which was no different to the 20–30 s TaT60 lesions (p=0.11).
4 Discussion
Temperature-controlled, power-modulated irrigated RF ablation with the DiamondTemp ablation system has been reported for both atrial and ventricular ablation [24, 26, 27, 29]. The focus of this study utilizing an ex vivo porcine cardiac model was to explore lesion morphologies at temperature set-points of 50 °C or 60 °C and at a range of ablation durations that may aid in developing dosing strategies. The key findings from this study were as follows:
-
1.
In both study groups, lesion depth plateaus after 60 s and location of maximum lesion width continues to deepen as application durations extend.
-
2.
Lesion morphologies in Group_60 were generally larger than those of the same duration in Group_50. Group_60 applications created similar-sized lesions in half the time compared to Group_50.
-
3.
“Time after temperature 60 °C”—TaT60 is a promising parameter to determine lesion depth. Lesion depth plateaus after reaching a temperature of 60 °C for 20–30 s.
4.1 Lesion biophysics with temperature-controlled RF
With a temperature-controlled system, operators can maximize resistive heating to create wide lesions similar to a HPSD approach, while allowing power modulation to maintain tissue temperature within a therapeutic range. The findings of this study show that lesion morphologies (depth and maximum width) in the Group_50 and Group_60 groups incrementally expanded between 5- and 60-s applications and showed no substantial increase with 90- or 120-s applications. Group_50 applications demonstrated incremental lesion depths from 0.9 ± 1.0 mm at 5 s to 4.5 ± 1.2 mm at 60 s, with no significant increases beyond a 60-s application. Similarly, the maximum width also increased with application duration until 60 s. In the Group_60 applications, lesion depths ranged from 1.9 ± 0.9 mm at 5 s to 5.6 ± 1.3 mm at 60 s, with no significant increases beyond a 60-s application. Similarly, maximum width also increased with ablation duration until 60 s. Dr. Winkle’s assertion that resistive heating transitions to conductive heating within the first 10 s [33] is further supported by the shift in lesion shape as the maximum width surpasses the surface width and continued deepening of the depth of maximum width between 5 and 10-s applications. In a study by Verma and colleagues [21], they assessed lesion morphology in multiple test models with a temperature set-point of 60 °C. In their work, a porcine thigh model demonstrated average lesion depths of 3.2, 4.4, and 5.2 mm for 5-, 10-, and 15-s applications, respectively. While reaching a similar maximum temperature, these dimensions were larger than the findings in the present study, suggesting the model may play a role in overall performance. The DTA system was used in an in vivo canine model to deliver discrete lesions in all chambers of the heart. Using similar application parameters, ventricular lesions with durations of 17.9 ± 3.4 s achieved more comparable lesion depths of 6.0 ± 1.6 mm [28]. The observed differences in the lesion biophysics may be related to the TaT60 of the applications. Analysis of 15-s applications in Group_60 resulted in lesion depths of 3.1 ± 0.8 mm. However, applications with TaT60 of 20 s resulted in lesion depths of 4.6 ± 0.8 mm.
Recently, reports of a novel ablation system utilizing an ultra-HPSD (90W × 4s) approach have surfaced, demonstrating predominantly transmural atrial lesions with widths ranging from 5 to 10 mm [34, 35]. The ventricular lesions presented with this ultra-HPSD approach were reported to be shallow (2.7–3.8 mm) and wide (6.9–7.8 mm) in nature. More recently, a modified ultra-HPSD 10-s step-down approach (90W × 4s titrated to 50W × 6s) demonstrated lesion depths of 3.9 mm and widths of 7.7 mm [36]. The lesion characteristics from the present study stand up considerably against the ultra-HPSD approach with the addition of a maximum power delivery of 50 W. Furthermore, as longer application times are required to generate deeper lesions, the DTA system delivers high power with power titration reducing the risk of tissue superheating. This system accurately measures the temperature at the tip-tissue interface and modulates power accordingly. This may slow the rate of temperature change within the tissue but does not inhibit the rise in inter-tissue temperature. In this study of 280 temperature-controlled RF applications, only five steam pops (1.8%) were observed with an average duration to pop of greater than 60 s.
Contact force has proven to be influential for lesion formation in power-controlled radiofrequency ablation [13, 14]. The role contact force has on lesion formation with DTA may differ based on application duration. Lesion dimensions from 5 to 15-s applications with the DTA system have been shown to be independent of the level of applied force [21]. Counter to this, Sasaki et al. [37] demonstrated increased applied force positively correlated with lesion dimensions with 6-s applications; however, these measurements appear to be inclusive of steam pop events. Interestingly, this study also presented a non-linear relationship between lesion dimensions as application duration lengthens. Our results support these findings that the ability to titrate power based on an accurate tip-tissue temperature will ultimately result in consistent lesion formation across a range of clinically relevant applied forces. Temperature-controlled ablation produces similar-sized lesions to power-controlled ablations and increasing contact force in the latter is associated only with an increased risk of steam pop [21, 34, 37].
4.2 Clinical implications
The DTA technology is predicated on modulating power once the set-point temperature is achieved. This is enabled through the enhanced thermal diffusivity of the industrial diamond elements embedded in the distal portion of the catheter, which allows for highly accurate tip-tissue temperature readings [20, 22]. From these accurate real-time temperature readings, the power is then modulated, thus allowing the resistive heating to remain in the optimal therapeutic range. Our findings show that at temperature set-points of both 50 and 60 °C, increasing ablation duration will provide greater lesion width and depth with a plateauing effect for depth at 60 s. In previous DTA atrial fibrillation clinical trials, ablation recommendations were a temperature set-point of 60 °C, and ablation termination guidance was 3 to 5 s beyond electrogram amplitude attenuation of 75 to 80% [20, 23,24,25]. During the DIAMOND-AF trial, the average application duration, with the first-generation DTA system, was 14.7 ± 5.3 s resulting in a 79.1% freedom from recurrence of any atrial arrhythmia during the 12-month follow-up. Similarly, in the FASTR-AF study, which introduced the updated generator with increased power modulation responsiveness, the average application duration was 10.3 ± 3.3 s, resulting in freedom from arrhythmia recurrence of 73.4%. By using the same ablation parameters, the present study shows that 10-s applications created lesion depths of 2.9 ± 0.8 mm and maximum lesion widths of 6.7 ± 1.3 mm, which fall well within the requirements for atrial transmurality. While shorter applications resulted in lesion depths better suited for atrial tissue, further testing should be conducted to evaluate the effectiveness of TaT60 as a metric for atrial applications.
Al Sheikhli and colleagues [29] reported the feasibility of using the DTA catheter system for ventricular tachycardia ablation in ischemic cardiomyopathy patients. Ablation was delivered with a temperature set-point of 60 °C for a maximum of 60 s and limited to bipolar electrogram amplitude attenuation of 75 to 80% or impedance reduction of no more than 20 Ω. Applying these ablation parameters, the present study findings would suggest lesion depths of 5.6 ± 1.3 mm, albeit in healthy tissue. The authors reported a single steam pop with a 44-s ablation lesion where the maximum temperature reached was 60 °C with an impedance reduction of 20 Ω. Interestingly, the maximum temperature was reached within the initial 10 s, and thus 34 s of ablation was delivered beyond reaching the pre-set temperature. The present study demonstrated a plateau in depth when TaT60 reaches 20 s. It is unknown if the steam pop would have been avoided had the ablation duration been limited to TaT60 of 20 s and if the necessary lesion size had been achieved.
Establishing stopping criteria for energy delivery is paramount for lesion formation to not extend beyond the cardiac tissue and to minimize tissue overheating. The plateau in depth identifies a point of diminishing return for continued RF delivery in ventricular tissue. Utilizing TaT60 criteria may also serve as a function of titrating RF time and thus potentially increasing efficiency in ventricular ablation procedures. Further clinical studies are ongoing applying this tailored DiamondTemp ablation approach where individual ablation application is based on the TaT60 duration.
5 Limitations
The ex vivo nature of this study may limit the ability to translate our findings to clinical outcomes. Experiments conducted on excised myocardial tissue lack cardiac motion and blood flow patterns that would be present in a clinical scenario. In this study, it was observed that when the temperature set-point of 50 °C or 60 °C was reached, as happened in most instances, it was maintained for the remainder of the application. In addition, the absence of perfused viable myocardium has an impact on intramuscular cooling as well as pre-ablation impedance values. This can have an impact on heat transfer into the tissue and may have an impact on lesion formation along with temperature recordings from the tip-tissue interface. Finally, slight variations in relative catheter contact and orientation at the tip-tissue interface can limit the rate at which temperature increases throughout the ablation, and the temperature set-point may not always be achieved. These limitations do not change the relationship between the measured temperatures presented in this study and the correlated lesion morphologies.
6 Conclusion
The DTA system rapidly creates wide lesions with lesion depth increasing over time with application durations up to 60 s. The introduction of “time after temperature 60 °C” demonstrates a potential clinical target for a stopping criterion for ventricular applications. With a temperature-controlled system, operators can maximize the resistive heating to create wide lesions, while allowing power modulation to maintain tissue temperature within a therapeutic range.
Abbreviations
- DTA :
-
DiamondTemp ablation
- HPSD :
-
High-power short-duration
- RF :
-
Radiofrequency
- TaT60 :
-
Time after temperature 60 °C
References
Calkins H, Hindricks G, Cappato R, et al. HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14(10):e275–444.
Breithardt G, Borggrefe M. The dawn of radiofrequency catheter ablation for cardiac arrhythmias. Heart Rhythm. 2021;18(3):485–6.
Haines DE, Verow AF. Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium. Circulation. 1990;82:1034–8.
Haines DE. The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring. Pacing Clin Electrophysiol. 1993;16(3 Pt 2):586–91.
Nath S, DiMarco JP, Haines DE. Basic aspects of radiofrequency catheter ablation. J Cardiovasc Electrophysiol. 1994;5(10):863–76.
Haines DE, Watson DD. Tissue heating during radiofrequency catheter ablation: a thermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall. Pacing Clin Electrophysiol. 1989;12(6):962–76.
Calkins H, Prystowsky E, Carlson M, Klein LS, Saul JP, Gillette P. Temperature monitoring during radiofrequency catheter ablation procedures using closed loop control Atakr Multicenter Investigators Group. Circulation. 1994 Sep;90(3):1279–86.
Dinerman JL, Berger RD, Calkins H. Temperature monitoring during radiofrequency ablation. J Cardiovasc Electrophysiol. 1996;7(2):163–73.
Kongsgaard E, Steen T, Jensen O, Aass H, Amlie JP. Temperature guided radiofrequency catheter ablation of myocardium: comparison of catheter tip and tissue temperatures in vitro. Pacing Clin Electrophysiol. 1997;20(5 Pt 1):1252–60.
Virk SA, Ariyaratnam J, Bennett RG, Kumar S. Updated systematic review and meta-analysis of the impact of contact force sensing on the safety and efficacy of atrial fibrillation ablation: discrepancy between observational studies and randomized control trial data. Europace. 2019;21(2):239–49.
Hussein A, Das M, Riva S, Morgan M, Ronayne C, Sahni A, Shaw M, Todd D, Hall M, Modi S, Natale A, Dello Russo A, Snowdon R, Gupta D. Use of ablation index-guided ablation results in high rates of durable pulmonary vein isolation and freedom from arrhythmia in persistent atrial fibrillation patients: the PRAISE study results. Circ Arrhythm Electrophysiol. 2018;11(9):e006576.
Das M, Wynn GJ, Saeed Y, Gomes S, Morgan M, Ronayne C, Bonnett LJ, Waktare JEP, Todd DM, Hall MCS, Snowdon RL, Modi S, Gupta D. Pulmonary vein re-isolation as a routine strategy regardless of symptoms: the PRESSURE randomized controlled trial. JACC Clin Electrophysiol. 2017;3(6):602–11.
Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT, Kantipudi C, Mansour MC, Melby DP, Packer DL, Nakagawa H, Zhang B, Stagg RB, Boo LM, Marchlinski FE. Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol. 2014;64(7):647–56.
Reddy VY, Dukkipati SR, Neuzil P, Natale A, Albenque JP, Kautzner J, Shah D, Michaud G, Wharton M, Harari D, Mahapatra S, Lambert H, Mansour M. Randomized, controlled trial of the safety and effectiveness of a contact force-sensing irrigated catheter for ablation of paroxysmal atrial fibrillation: results of the TactiCath contact force ablation catheter study for atrial fibrillation (TOCCASTAR) study. Circulation. 2015;132(10):907–15.
Fujiwara R, Imamura K, Kijima Y, Masano T, Nagoshi R, Kohzuki A, Shibata H, Tsukiyama Y, Takeshige R, Yanaka K, Nakano S, Fukuyama Y, Shite J. The importance of catheter stability evaluated by Visitag during pulmonary vein isolation. J Interv Card Electrophysiol. 2016;46(2):161–6.
Das M, Loveday JJ, Wynn GJ, Gomes S, Saeed Y, Bonnett LJ, Waktare JEP, Todd DM, Hall MCS, Snowdon RL, Modi S, Gupta D. Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values. Europace. 2017;19(5):775–83.
Mattia L, Crosato M, Indiani S, Causin E, Licciardello C, Maria Squasi PA, De Leo A, Calzolari V. Prospective evaluation of lesion index-guided pulmonary vein isolation technique in patients with paroxysmal atrial fibrillation: 1-year follow-up. J Atr Fibrillation. 2018;10(6):1858.
Santoro F, Metzner A, Brunetti ND, Heeger CH, Mathew S, Reissmann B, Lemeš C, Maurer T, Fink T, Rottner L, Inaba O, Kuck KH, Ouyang F, Rillig A. Left atrial anterior line ablation using ablation index and inter-lesion distance measurement. Clin Res Cardiol. 2019;108(9):1009–16.
Winkle RA, Mead RH, Engel G, Patrawala RA. Atrial fibrillation ablati“n: “perpetual mo”ion” of open irrigated tip catheters at 50 W is safe and improves outcomes. Pacing Clin Electrophysiol. 2011;34(5):531–9.
Iwasawa J, Koruth JS, Petru J, et al. Temperature-controlled radiofrequency ablation for pulmonary vein isolation in patients with atrial fibrillation. J Am Coll Cardiol. 2017;70(5):542–53.
Verma A, Schmidt MM, Lalonde JP, Ramirez DA, Getman MK. Assessing the relationship of applied force and ablation duration on lesion size using a diamond tip catheter ablation system. Circ Arrhythm Electrophysiol. 2021;14(7):e009541.
Stoffregen WC, Maddox W, Cuoco F, McElderry HT. Abstract B-PO02-094, Temperature sensing accuracy and lesion characteristics of a novel, diamond tip, temperature controlled open irrigated radiofrequency catheter. Heart Rhythm. 2018;15(5):S218.
Starek Z, Lehar F, Jez J, Pesl M, Neuzil P, Sediva L, Petru J, Dujka L, Funasako M, Kautzner J, Peichl P, Aldhoon B, Albenque JP, Combes S, Boveda S, Dukkipati SR, Reddy VY. Efficacy and safety of novel temperature-controlled radiofrequency ablation system during pulmonary vein isolation in patients with paroxysmal atrial fibrillation: TRAC-AF study. J Interv Card Electrophysiol. 2022;64(2):375–81.
Kautzner J, Albenque JP, Natale A, Maddox W, Cuoco F, Neuzil P, Poty H, Getman MK, Liu S, Starek Z, Dukkipati SR, Colley BJ 3rd, Al-Ahmad A, Sidney DS, McElderry HT. A novel temperature-controlled radiofrequency catheter ablation system used to treat patients with paroxysmal atrial fibrillation. JACC Clin Electrophysiol. 2021;7(3):352–63.
Neuzil P, Poty H, de Chillou C, Petru J, Getman MK, Liu S, Funasako M, Durand-Dubief A, Combes S, Albenque JP. Radiofrequency ablation using the second-generation temperature-controlled diamond tip system in paroxysmal and persistent atrial fibrillation: results from FASTR-AF. J Interv Card Electrophysiol. 2023;66(2):343–51.
Ramak R, Lipartiti F, Mojica J, Monaco C, Bisignani A, Eltsov I, Sorgente A, Capulzini L, Paparella G, Deruyter B, Iacopino S, Motoc AI, Luchian ML, Osorio TG, Overeinder I, Bala G, Almorad A, Ströker E, Sieira J, et al. Comparison between the novel diamond temp and the classical 8-mm tip ablation catheters in the setting of typical atrial flutter. J Interv Card Electrophysiol. 2022;64(3):751–7.
Rottner L, My I, Schleberger R, Moser F, Moser J, Kirchhof P, Ouyang F, Rillig A, Metzner A, Reissmann B. Temperature-controlled ablation of the mitral isthmus line using the novel DiamondTemp ablation system. Front Cardiovasc Med. 2022;24(9):1046956.
Maddox W, Dhanjal T, Getman MK, McElderry HT, Schmidt MM, Verma A. AFS2022-33, Characterizing lesion formation of temperature-titrated radiofrequency dosage with a novel diamond tip ablation system. J Cardiovasc Electrophysiol. 2022;33(4):801–2.
Al-Sheikhli J, Patchett I, Lim VG, Marshall L, Foster W, Kuehl M, Yusuf S, Panikker S, Patel K, Osman F, Banerjee P, Lellouche N, Dhanjal T. Initial experience of temperature-controlled irrigated radiofrequency ablation for ischaemic cardiomyopathy ventricular tachycardia ablation. J Interv Card Electrophysiol. 2023;66(3):551–9. https://doi.org/10.1007/s10840-022-01158-4.
Nakagawa H, Yamanashi WS, Pitha JV, Arruda M, Wang X, Ohtomo K, Beckman KJ, McClelland JH, Lazzara R, Jackman WM. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995;91(8):2264–73.
Yokoyama K, Nakagawa H, Wittkampf FH, Pitha JV, Lazzara R, Jackman WM. Comparison of electrode cooling between internal and open irrigation in radiofrequency ablation lesion depth and incidence of thrombus and steam pop. Circulation. 2006;113(1):11–9.
Hwang YM, Lee WS, Choi KJ, Kim YR. Radiofrequency induced lesion characteristics according to force-time integral in experimental model. Medicine (Baltimore). 2021;100(10):e25126.
Winkle RA. HPSD ablation for AF high-power short-duration RF ablation for atrial fibrillation: a review. J Cardiovasc Electrophysiol. 2021;32(10):2813–23.
Leshem E, Zilberman I, Tschabrunn CM, Barkagan M, Contreras-Valdes FM, Govari A, Anter E. High-power and short-duration ablation for pulmonary vein isolation: biophysical characterization. JACC Clin Electrophysiol. 2018;4(4):467–79.
Takigawa M, Kitamura T, Martin CA, Fuimaono K, Datta K, Joshi H, Constantin M, Bourier F, Cheniti G, Duchateau J, Pambrun T, Denis A, Derval N, Sacher F, Cochet H, Hocini M, Haïssaguerre M, Jaïs P. Temperature- and flow-controlled ablation/very-high-power short-duration ablation vs conventional power-controlled ablation: comparison of focal and linear lesion characteristics. Heart Rhythm. 2021;18(4):553–61.
Rozen G, Ptaszek LM, Zilberman I, Douglas V, Heist EK, Beeckler C, Altmann A, Ruskin JN, Govari A, Mansour M. Safety and efficacy of delivering high-power short-duration radiofrequency ablation lesions utilizing a novel temperature sensing technology. Europace. 2018;20(FI_3):f444–50.
Sasaki W, Matsumoto K, Higuchi S, Mori H, Fukaya H, Kawano D, Tanaka N, Narita M, Tsutsui K, Ikeda Y, Arai T, Nakano S, Kato R. Detailed analysis of the lesion formation using a diamond tip catheter in an ex vivo experimental model. J Cardiol. 2023;82(1):76–83. https://doi.org/10.1016/j.jjcc.2023.03.006.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Ethics approval
Not required because of ex vivo nature of the study
Conflict of interest
Prof. Tarvinder Dhanjal reports consultancy services, educational grant awards, and speaker honoraria fees for Abbott, Biosense Webster, and Medtronic. Dr. Melissa Robinson reports consultancy services and speaker honoraria for Abbott, Biosense Webster, Boston Scientific, and Medtronic. Megan M. Schmidt and Michael K. Getman are employed by and stockholders of Medtronic.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Dhanjal, T.S., Schmidt, M.M., Getman, M.K. et al. Characterizing lesion morphology of a novel diamond-tip temperature-controlled irrigated radiofrequency ablation catheter. J Interv Card Electrophysiol 67, 293–301 (2024). https://doi.org/10.1007/s10840-023-01595-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-023-01595-9