Introduction

Virtual Histology intravascular ultrasound (VH-IVUS) allows measurements of both atherosclerotic plaque and vessel size and plaque composition in vivo [110]. Plaque composition of target lesions prior to percutaneous coronary intervention (PCI) may impact clinical outcome following PCI [1119]. A relatively large necrotic core volume in target lesions as assessed with VH-IVUS is an independent predictor of creatine kinase and troponin elevation after PCI [11, 15]. Microembolization of such plaque components may lead to myocardial injury during stent implantation [11, 14, 16, 18, 20]. Certain plaque components or phenotypes as assessed with VH-IVUS are related to no-reflow phenomenon post-PCI which may be triggered by such microembolization and adversely affects prognosis [14, 18]. Therefore, volumetric VH-IVUS assessment of target lesions prior to PCI may have the potential to identify lesions at particularly high risk for such complications and may help to tailor such procedures. However, an important pre-requisite for clinical VH-IVUS studies assessing such relations in advanced coronary lesions prior to PCI is an adequate measurement reproducibility [2].

We recently reported significant but small differences for VH-IVUS measurements of mild-to-moderately diseased coronary segments between different European IVUS centers [21]. But so far, no data is available on the between-center reproducibility of VH-IVUS measurements of advanced lesions prior to PCI. Therefore, in the present study we assessed the between-center reproducibility of volumetric VH-IVUS measurements of target lesions, that were consecutively stented, by comparing data from repeated analyses as performed by independent analysts at 4 independent IVUS centers in Europe.

Methods

Study population

In order to assess the between-center reproducibility of segmental volumetric VH-IVUS data analysis of diseased atherosclerotic human coronary arteries, we performed IVUS analyses in 20 mm-long segments of coronary lesions prior to PCI in patients with unstable angina pectoris or stable angina. Coronary segments had to meet the following inclusion criteria: no severe plaque calcification, no major vessel tortuosity, and/or major side-branches. In this multicenter study design, 4 experienced analysts independently performed VH-IVUS analyses of all segments included in the study. Each IVUS center was invited to submit IVUS pullbacks of 10 coronary segments obtained from patients in sinus rhythm in whom a clinically driven cardiac catheterization required IVUS assessment prior to PCI. Participating centers were Odense University Hospital, Odense, Denmark; Essen University Hospital, Essen, Germany; Central University Hospital, Warsaw, Poland; and Thoraxcentrum Twente, Enschede, The Netherlands.

IVUS data acquisition

Acquisition of IVUS data was performed in an ECG-gated way with commercially available phased-array IVUS catheters (Eagle Eye Gold 2.9F 20 MHz, Volcano Corporation, Rancho Cordova, CA, USA) with a dedicated console [22]. The IVUS transducer was advanced ≥10 mm distal to the most distal side-branch. Angiographic cine runs were performed to define the position of the IVUS transducer. After intracoronary injection of 200 μg nitroglycerin, a continuous pullback of the IVUS catheter was performed using a motorized pullback device at 0.5 mm/s (TrackBack-II, Volcano Corporation). IVUS image data were stored on digital video disk (DVD) for offline analysis.

VH-IVUS data analysis

Description of the technique and the validation of VH-IVUS have previously been published [1, 3, 4, 6]. In brief, tissue maps are reconstructed from spectral IVUS radiofrequency data and classify the plaque in four major components: calcium, fibro-lipidic, fibrous, and necrotic core. Tissue components were displayed according to a color code (white, light-green, green, and red, respectively).

Offline VH-IVUS analysis of all cases was performed by one experienced IVUS analyst per center (Fig. 1). The IVUS sequences were provided with a predefined region of interest (20 mm length) that contained a (part of a) target lesion for PCI. Segments were located between two adjacent side-branches and contained no major calcification that could have limited reliable detection of the external vascular boundary.

Fig. 1
figure 1

Image acquisition and data analysis. Motorized pullbacks at 0.5 mm/s (including image frame acquisition at the time of the R-wave peak) were performed by each IVUS center (AD). These pullback sequences (obtained from 10 patients) were exchanged in order to obtain a ‘pullback pool’ of 40 pullbacks, which each analyst independently analyzed (I). Automated contour detection and, if required, manual correction of the lumen and vessel border were performed on all frames using the VH-images and software (II). Volumetric VH-IVUS data of vessel and plaque geometry and plaque composition were automatically generated for the analyzed segment based on the application of Trapezium method (III)

By use of the VH-images and software, IVUS measurements of geometry and plaque composition were made. The lumen and vessel borders were traced using a semi-automated, computer-assisted contour detection program (Volcano pcVH software program version 2.2; Volcano); to exclude potential between version differences of the analysis software, all four centers used the latest version of the analysis software. The leading edge of the blood-intima acoustic interface and the leading edge of the media-adventitia interface were automatically detected to define the lumen and vessel borders, respectively, and manually corrected as required. Plaque & media was used as a measure of atherosclerotic plaque and was calculated as the difference between vessel and lumen cross-sectional area. For each coronary segment, volumetric data of the vessel, lumen, plaque and plaque burden, as well as fibrous, fibro-lipidic, necrotic core, and calcified tissue were obtained from the application of the trapezium method to the cross-sectional area measurements [21, 23]. The pcVH software program uses this method that considers the exact spacing between image frames to calculate volumes from the cross-sectional area data in the most accurate way. In addition, the relative amount of the four plaque components was calculated. To account for any differences in length, all volumetric data were normalized for 10 mm-long segments [21, 23].

In order to assess the between-center reproducibility of volumetric VH-IVUS analyses, comparisons were performed on a two-observer basis (i.e., center A vs. B, C and D respectively; center B vs. C and D respectively; and center C vs. D).

Statistical analysis

Statistical analysis was performed with SPSS 15.0 (SPSS Inc., Chicago, IL, USA). Dichotomous variables are presented as frequencies, while quantitative data are presented as mean ± 1SD (or ±1SEM). Quantitative data were compared by use of a paired t test after Normality of the data was confirmed. A two-sided P-value < 0.05 was considered as significant. According to Bland and Altman, the agreement between two measurements was assessed by determining the mean ± 2SD of the between measurement differences [24]. Intra-class correlations were calculated for comparisons at an individual level. An intra-class correlation >0.90 was considered good.

Results

Study population

A total of 40 coronary segments (6 right; 27 left anterior descending; and 7 left circumflex coronary arteries) were analyzed from 40 patients (age 64 ± 10 years, 75% were male) who were in sinus rhythm during the IVUS pullback. IVUS examination was performed in both patients with stable syndromes (65%) and unstable syndromes (35%). Eighty-eight percent of the patients was on statin therapy prior to the cardiac catheterization, and 35% was diabetics. There was no complication related to IVUS imaging.

VH-IVUS analysis and data

Independent analysis of each of the 40 coronary segments was performed by each of the four analysts and volumetric data was obtained to evaluate the reproducibility of volumetric VH-IVUS measurements (50 ± 14 frames/segment). This approach required the analysis of a total of 8,052 cross-sectional frames (2,013 frames per analyst). Manual correction of the automatic border detection was required in almost all frames for a total analysis time of 60 ± 6 min/segment. The length of the analyzed segments was 20.0 ± 0.3 mm (range: 19.1–20.6 mm). On average, the atherosclerotic segments contained predominantly fibrous (57.4%), fibro-lipidic (13.1%), and necrotic core (19.1%) tissue. Geometric and compositional IVUS data obtained by the analysts of the 4 different centers (A–D) are presented in Table 1.

Table 1 VH-IVUS data from four different centers (A–D)

VH-IVUS measurement variability of plaque and vessel geometry

VH-IVUS measurements of vessel geometry were highly correlated for the different comparisons. Overall intra-class correlation and range of the individual between-center comparisons (within brackets) was for vessel, lumen, plaque volume and plaque burden 0.99 (0.98–1.00), 0.92 (0.84–0.97), 0.96 (0.93–0.98), and 0.83 (0.75–0.90), respectively. Although there were significant differences for vessel, lumen, and plaque volumes, these differences were only moderate in size (Tables 1, 2; Fig. 2). Plaque burden also showed significant differences for the comparisons between two centers. However, the differences were only moderate in size, ranging from 0.8 to 4.6%. Statistical analysis using Anova with Holm-Bonferroni correction did not show significant differences, except for plaque burden measurements of center C versus D (adjusted P-value 0.048), no significant differences were seen (P > 0.1, Table 1). The limits of agreement (i.e., 2 SD of mean difference) were relatively small and similar for the different comparisons; measurement reproducibility of vessel volume was particularly high (Fig. 3a, b). Measurement reproducibility was not influenced by the size of the measured volumes.

Table 2 Measurement differences of VH-IVUS data from four different centers (A–D)
Fig. 2
figure 2

Example of differences in contour detection in a single frame. The same coronary segment was analyzed by 2 different centers; please note that differences in contour detection resulted in differences in plaque composition. Center A (Greyscale (a) and VH-IVUS (b) analysis) versus Center B (c and d, resp)

Fig. 3
figure 3figure 3

a Agreement of repeated VH-IVUS measurements of geometric volumes. Agreement of repeated VH-IVUS measurements of geometry between center A versus B (left), center A versus C (mid) and center A versus D (right). Δ = difference. b Agreement of repeated VH-IVUS measurements of geometric volumes. Agreement of repeated VH-IVUS measurements of geometry between center B versus C (left), center B versus D (mid) and center C versus D (right). Δ = difference

VH-IVUS measurement variability of plaque composition

VH-IVUS measurements of plaque composition were highly correlated for the different comparisons. Overall intra-class correlation and range of the individual between-center comparisons (within brackets) for (absolute) fibrous, fibro-lipidic, necrotic core and calcified volume was 0.96 (0.94–0.98), 0.94 (0.91–0.97), 0.98 (0.96–0.99), and 0.99 (0.98–0.99), respectively. There were significant differences for all plaque components (Tables 1, 2). The limits of agreement (i.e., 2 SD of mean difference) for compositional data were higher compared to geometrical data, but similar for the different comparisons (Fig. 4a, b). Fibrous tissue and necrotic core showed the highest measurement reproducibility. Smaller fibrous and fibro-lipidic volumes tended to show somewhat higher between-center variability.

Fig. 4
figure 4figure 4

a Agreement of repeated VH-IVUS measurements of compositional volumes. Agreement of repeated VH-IVUS measurements of plaque composition between center A versus B (left), center A versus C (mid) and center A versus D (right). Δ = difference. b Agreement of repeated VH-IVUS measurements of compositional volumes. Agreement of repeated VH-IVUS measurements of plaque composition between center B versus C (left), center B versus D (mid) and center C versus D (right). Δ = difference

Discussion

In the present study, we evaluated the between-center reproducibility of VH-IVUS measurements of target lesions prior to PCI. VH-IVUS measurements of plaque and vessel dimensions and plaque composition were highly correlated for the different comparisons (overall ICC 0.83–0.99). However, we found significant between-center differences for vessel dimensions (relative differences of ≤4%, ≤12%, ≤9%, and ≤9%; vessel, lumen, and plaque volume and plaque burden, respectively) and plaque composition (relative differences of ≤15%, ≤25%, ≤11%, and ≤6%; fibrous, fibro-lipidic, necrotic core, and calcified plaque components, respectively). The highest measurement reproducibility for plaque components was found for fibrous tissue and necrotic core.

Rationale of VH-IVUS assessment of coronary lesions prior to PCI

Percutaneous coronary interventions may cause microembolization of unstable plaque material, which can lead to no-reflow phenomenon, myocardial micro-infarcts with cardiac marker release and impaired left ventricular function, and significantly impaired clinical outcome [11, 14, 16, 18, 20]. Several VH-IVUS studies demonstrated that differences in plaque composition of target lesions are related to such post-PCI events [1119]. Hong et al. showed that relative necrotic core volumes were significantly larger in patients with elevated troponin levels following PCI (19.8 ± 10.4% vs. 12.8 ± 8.4%, P = 0.015) [15]. Böse et al. reported that cardiac marker release following PCI was much greater in plaques with a large necrotic core volume (>10.81 mm³; P < 0.001) with a trend for a lower increase of cardiac markers in patients treated with statins before PCI [11]. Certain plaque components or phenotypes as assessed with VH-IVUS are related to no-reflow phenomenon post-PCI which may be triggered by such microembolization and adversely effects prognosis [14, 18]. Hong et al. reported that relative necrotic core volume serves as an independent predictor of no-reflow post-stenting in patients with acute coronary syndrome (odds ratio = 1.126; 95%CI 1.045–1.214, P = 0.002) [14].

Therefore, volumetric VH-IVUS assessment of target lesions prior to PCI may have the potential to identify lesions at particularly high risk for such complications and may help to tailor interventions (i.e., embolic protection device, direct stenting vs. pre-dilatation etc.) and may help to optimize systemic medical therapy (i.e., therapy with very high dose statins, or novel anti-atherosclerotic or platelet inhibiting drugs) [25, 26]. However, only large multicenter trials will be able to determine the clinical value of the assessment of plaque composition with VH-IVUS prior to PCI. An important pre-requisite for clinical VH-IVUS studies assessing such relations in advanced coronary lesions, is an adequate measurement reproducibility of VH-IVUS especially when pooled IVUS data is used in multicenter trials or registries [2].

Measurement reproducibility and studies with multicenter design

Several prior single-center studies demonstrated sufficient reproducibility of VH-IVUS measurements between observers, catheters, and repeated pullbacks [2, 8, 9, 21, 23].

Rodriguez-Granillo et al. assessed the measurement variability of VH-IVUS cross-sectional data obtained form 16 mild coronary plaques in a single-center design. The relative intra-observer difference was <11% for plaque cross-sectional area and compositional measurements showed a variability up to 24% [9]. Prasad et al. assessed volumetric VH-IVUS measurement variability of 16 significant coronary plaques (mean plaque burden of 55%) prior to PCI and reported high agreement for geometrical and compositional measurements (Spearman’s correlation >0.8) in a single-center design [8]. A relatively high intra- and interobserver reproducibility of volumetric geometric and compositional VH-IVUS data was previously described by Hartmann et al. in a series of 33 mild-to-moderate diseased coronary segments [2].

However, the aforementioned studies assessed VH-IVUS measurement variability of analysts from a single center or core lab with the same training background. In a previous multicenter study by this group of investigators, we demonstrated in coronary segments with mild-to-moderate coronary atherosclerosis a higher measurement variability and small but significant measurement differences between analysts from 4 different IVUS centers with different training programs [21]. In the current study, in much more advanced atherosclerotic lesions, VH-IVUS measurements of vessel geometry and plaque composition showed, despite good correlation between the four IVUS centers, significant differences for both vessel geometry and plaque composition. These differences could affect the results of multicenter trials with VH-IVUS.

Implications for multicenter VH-IVUS studies of coronary lesions prior to PCI

VH-IVUS measurement differences may have resulted from systematic over or underestimation of the lumen and/or vessel borders, differences in the interpolation of the vessel contour at the site of side-branches or behind calcified areas, and misinterpretation of thrombus and soft plaque adjacent to the lumen, which could theoretically have introduced larger measurement variability. The nature of such advanced lesions prior to PCI with more calcification, complex lesions and the IVUS probe against vessel wall/plaque, may result in increased difficulties of border detection in contrast with mild-to-moderate diseased coronary segments. It is likely that small differences or failures in contour detection will translate into (larger) errors of measurements of plaque composition. Therefore, our present study underlines the necessity to analyze the VH-IVUS data of advanced coronary lesions prior to PCI at a single site or by an experienced core-lab to avoid measurement errors and misinterpretation of acquired data.

Recently, Shin et al. described a simplified contour detection method for the assessment of plaque composition with on VH-IVUS [27]. In Shin’s method the lumen contour is drawn around the IVUS catheter, that is, without following the leading edge of the interface lumen intima. With this method only calcium and necrotic core can be assessed and related to total vessel area [27]. Shin et al. were able to demonstrate that coronary lesions with a greater amount of dense calcium volume and percent necrotic core to external elastic membrane show more post-PCI myocardial injury [28]. Nevertheless, exclusive application of this method is limited by incompleteness of the geometrical and compositional data (e.g., no data on lumen and plaque size, and no information on fibrous or fibro-lipidic tissue).

Limitations

The present study was performed in an offline setting in coronary segments with advanced coronary atherosclerosis prior to PCI; therefore, our findings cannot be extrapolated to VH-IVUS reproducibility studies in an online setting in the catheterization laboratory or following stent implantation. Similar to other studies with IVUS and VH-IVUS, we excluded very tortuous and severely calcified vessels that could have led to non-uniform pullbacks and/or inability to detect the (external) vessel contour [9, 21, 23]. We used the pcVH software (available in all four centers) to perform IVUS image analyses in entire coronary segments. Meanwhile a newer computer software is available, however, this does not affect the importance of the present study. The purpose of out present study was to examine systematic between-center differences when analyzing lumen and vessel contours with a semi-automated IVUS analysis program in this specific set of coronary lesions; it was not our aim to test a specific software release. In addition, the between-center differences of the present study were mainly explained by differences in manual contour editing, which is still generally not abandoned by the most recent IVUS analysis software. The variability between different IVUS catheters or pullback devices, repeated pullbacks, and independent selections of the segment of interest were not addressed in our present multicenter study, as this study focused on the assessment of measurement reproducibility in predefined coronary segments that contained target lesions prior to PCI.

Conclusions

A central analysis for VH-IVUS multicenter studies of lesions prior to PCI should be pursued. Moreover, it may be problematical to pool VH-IVUS data of individual trials that are analyzed by independent IVUS centers.