Introduction

While there is a well established relationship between degree of stenosis and plaque burden and ischemic symptoms in patients with stable angina, plaque composition, especially necrotic core, is more relevant in terms of occurrence of acute coronary syndromes [14]. Virtual Histology-intravascular ultrasound (VH-IVUS), which permits the analysis of coronary plaque composition in vivo, is increasingly used in clinical research [512]. Despite its demonstrated utility in quantifying coronary plaque components with high predictive accuracy both in vitro and in vivo [13, 14], VH-IVUS is infrequently used routinely in the catheterization laboratory for clinical decision making partly because it is not easy to immediately draw the contours. In particular, lumen contour detection takes a long time and it is especially challenging in patients with high risk lesions (i.e., acute coronary syndrome) because there is no clear separation between luminal thrombi and intima. The aim of this study is to assess a new measurement method (Shin’s method) that readily provides on-line (i.e., immediately after acquisition) necrotic core and calcium measurements to help interventionalists to better assess coronary lesions irrespective of clinical presentation. Therefore, in the present study we compared the new measurement method (Shin’s method) and the conventional measurement method in quantifying necrotic core and calcium in atherosclerotic lesions using VH-IVUS.

Methods

Study patients

In this study, 1,401 frames and 59 segments of culprit lesions were imaged in a total of 57 consecutive patients with unstable angina who underwent coronary artery stenting. Specifically, measurements of necrotic core and dense calcium (i.e., area and volume) by VH-IVUS were considered. Unstable angina was defined according to the Braunwald classification [15], and patients with class IIB and IIIB were included in the study. Patients were excluded if they had an IVUS-detected thrombotic lesion, left main disease and chronic total occlusion. Culprit lesion is defined as the lesion related to the patient’s symptom that was treated with stenting. We determined the analysis times needed for both approaches and report the mean times of analysis. This study was performed with the patient’s written informed consent and approval of the institutional review board.

VH-IVUS

Details of the radiofrequency (RF) IVUS technique have been extensively validated [9, 10, 13, 14]. In brief, spectral analysis of VH-IVUS data is used to create tissue maps to classify atherosclerotic plaques into four major components (fibrous: green; fibrofatty: light-green; dense calcium: white; and necrotic core: red). In this study, only dense calcium and necrotic core were considered because they are commonly found in high-risk plaques; both were assessed using the conventional method and Shin’s method.

IVUS data acquisition

IVUS data were acquired with commercially available phased-array IVUS catheters (Eagle Eye Gold 2.9-F 20 MHz, Volcano Corporation, Rancho Cordova, USA) by a dedicated VH-IVUS console (Volcano Corporation, Rancho Cordova, USA). Image acquisition was ECG-gated. After intracoronary injection of 200 μg nitroglycerine, continuous pullback of the IVUS catheter was performed using a motorized pullback device at 0.5 mm/s. Data of both pullbacks were stored on hard disk for off-line analysis.

VH-IVUS data analysis

RF-IVUS analysis was performed off-line by an independent core laboratory (Cardialysis BV, Rotterdam, The Netherlands). The region of interest comprises the most diseased 10-mm segment within the culprit lesion that was contained the largest plaque burden. This approach mimics a real clinical situation in which the interventionalists must make a clinical decision on the most stenotic segment using not only geometrical but also compositional analysis.

Conventional method of analysis

The contours were traced using a semiautomatic contour detection program (PcVH 2.2, Volcano Corporation, Rancho Cordova, USA). The contours of each cross sectional image per frame were manually corrected if required to meet a high standard of accuracy. The lumen contour was detected by tracing the leading edge of the intima, and the vessel contour by tracing the leading edge of the adventitia. For each frame geometrical and compositional analyses were performed; the mean area of the vessel and lumen were obtained and for each segment, the vessel and lumen volumes were generated (Figs. 1, 2).

Fig. 1
figure 1

A The conventional method entails drawing the lumen contour (white arrows). B Instead of measuring the lumen, Shin’s method entails drawing behind the catheter. While there is a “flare” (ring-down artifact) around the catheter, the artificial contour must be drawn away from this flair (white arrows)

Fig. 2
figure 2

A and B Corresponding IVUS frames. Shin’s method (B) shows similar necrotic core and dense calcium area compared with the conventional method (A). To compare only the influence in terms of dense calcium and necrotic core of the two different approaches the same contour of the vessel was used (FI fibrous, FF fibrofatty, DC dense calcium, NC necrotic core)

Shin’s method of analysis

Shin’s method also used the off-line software PcVH 2.2. Instead of measuring the lumen as described above for the conventional method, in Shin’s method the lumen contour is drawn around the IVUS catheter (hereafter termed catheter contour), that is, without following the leading edge of the interface lumen intima. When there is a “flare” (ring-down artifact) around the IVUS catheter, this catheter contour must be drawn away from this flare (Fig. 1). The catheter contour was manually detected. To compare only the influence of the two different approaches (Conventional vs. Shin’s), in terms of dense calcium and necrotic core, the same contour of the vessel was used (Figs. 1, 2).

Using Shin’s method the following information can be obtained:

  1. 1.

    Necrotic core and dense calcium areas and volumes.

  2. 2.

    Vessel area and volume.

However, using Shin’s method the following information can NOT be obtained:

  1. 1.

    Fibrous and fibrofatty areas and volumes.

  2. 2.

    Lumen and plaque areas and volumes.

Data analysis

For all comparisons, two levels were considered, the frame level and the segment level.

First, we compared the two methods (necrotic core and calcium) by one expert analyst. To assess the reproducibility of Shin’s method, eight segments (189 frames) were randomly selected and the same segments were analyzed twice by the same analyst with a 3-week time difference. To this end, the vessel contour was kept from the conventional method, so that only the catheter contour was drawn twice.

Statistical analysis

Analyses were performed with SPSS 16.0 (SPSS Inc., Chicago, IL). Dichotomous data are presented as frequencies. Quantitative data are presented as mean ± SD and compared using Student’s t-test, and linear regression analysis for repeated measures. The Spearman’s correlation coefficient was used to assess associations between measured parameters determined by the two methods. According to Bland and Altman [16], the agreement between two measurements were assessed by determining the mean ± 2 SD of the between measurement differences. A two-sided P value < 0.05 was considered significant.

Results

Study patients

Table 1 shows the patients’ characteristics. The mean age was 60.1 ± 8.6 years, and 57.6% of participants were men. The study vessels were the left anterior descending artery in 55.9%, the left circumflex artery in 10.2% and the right coronary artery in 33.9% of cases. The length of the segments was 10.2 ± 0.3 mm with a volumetric plaque burden of 60.7 ± 0.8% (43.1–75.5%). There was no complication related to IVUS imaging.

Table 1 Study population

Comparison between the conventional and Shin’s methods

By design, the mean length of analysis and the mean vessel area are the same (Table 2). There were no significant differences in the mean area and volume of necrotic core and dense calcium between the two methods. The correlation coefficients (R) were 0.99 (P < 0.001) for all compared parameters (Fig. 3). Between methods, the absolute differences in mean area and volume of necrotic core were 0.02 ± 0.02 mm² and 0.34 ± 0.29 mm³, respectively, while for mean area and volume of dense calcium, the absolute differences were 0.04 ± 0.07 mm² and 0.36 ± 0.52 mm³, respectively. Although it did not achieve statistical significance, mean dense calcium was consistently smaller and mean necrotic area was larger in Shin’s method as compared to the conventional method (Fig. 4). A total of three segments in necrotic core and five segments in dense calcium had an absolute difference in volume of more than 1 mm³. The correlation of the mean areas and volumes of necrotic core and dense calcium between Shin’s and the conventional method was expressed by a linear least-squares fit in Fig. 3.

Table 2 Mean area and volume of necrotic core and dense calcium with conventional and Shin’s method
Fig. 3
figure 3

Graphs show significant correlations for the absolute measurement values of each VH-IVUS plaque component between the conventional method and Shin’s method for atherosclerotic coronary lesions

Fig. 4
figure 4

Differences of necrotic core and dense calcium volumes (Shin’s method minus conventional method). Mean necrotic core volumes were larger in Shin’s method as compared to the conventional method (A), while the mean dense calcium volume was consistently smaller in Shin’s method as compared to the conventional method (B)

Although fibrous and fibrofatty components are not correct by Shin’s method because this method can not differentiate between lumen contents (usually characterized as fibrous and fibrofatty) and the fibrous and fibrofatty that are part of the plaque contents, here we provide a comparison between conventional vs. Shin’s method: 3.37 ± 2.03 mm2 vs. 4.04 ± 2.16 mm2, P < 0.001 for fibrous area and 0.87 ± 0.93 mm2 vs. 1.68 ± 1.49 mm2, P < 0.001 for fibrofatty area, respectively.

Reproducibility of Shin’s method

Table 3 shows the reproducibility of Shin’s method. The linear least-squares fit of the relationship between the two measurements for necrotic core area and dense calcium area were R = 0.999, P < 0.001 and R = 1.0, P < 0.001, respectively (Fig. 5). For areas of the necrotic core and dense calcium, the means of the differences between the two measurements were 0.004 ± 0.019 and 0.002 ± 0.013 mm². The limits of agreement for area of the necrotic core and dense calcium were −0.034 to 0.042 mm² and −0.024 to 0.051 mm². The reproducibility coefficients were within 1% of the means of the two measurements for area and volume of the necrotic core and dense calcium.

Table 3 Reproducibility of necrotic core and dense calcium by Shin’s method
Fig. 5
figure 5

Assessment of Shin’s method reproducibility. The differences of the mean necrotic core and the mean dense calcium areas between the two measurements were nearly zero, and the reproducibility coefficients were within 1% of the means of the two measurements

Analysis time

To assess the measurement analysis time difference between the two methods, 12 segments were randomly selected. Manual correction of the automatic contour detection was required in almost all frames for a total analysis time of 26.8 ± 6.7 min/segment in the conventional method. However, analysis time was 3.3 ± 0.6 min/segment for Shin’s method (Table 4).

Table 4 Comparison of analysis time between the conventional method and Shin’s method (12 segments)

Discussions

This report introduces a new measurement method (Shin’s method) for necrotic core and dense calcium using VH-IVUS. The main findings of this study are: (1) necrotic core and dense calcium areas and volumes determined by Shin’s method show a good correlation to those derived from the conventional method; (2) Shin’s method has a high reproducibility, and (3) analysis time is shorter for Shin’s method.

The introduction of RF-based IVUS analysis permits quantitative assessment of atherosclerotic plaque composition [17]. This is of crucial relevance because necrotic core is a key determinant of plaque vulnerability. This imaging modality has been extensively scrutinized and studied in different clinical contexts, and most studies using this technique have shown consistency as compared with previous pathological studies [13, 14]. However, this imaging technique is very much influenced by the precision of contour detection, which in turn is highly time-consuming thus rendering its clinical use for online clinical decisions today not possible and not widely accepted. There was a need therefore to explore a different manner of necrotic core quantification that would allow information to be quickly obtained as accurate as possible and with excellent reproducibility.

Use of Shin’s method does not require training, and more rapid contour detection around the catheter can be done. In the future, with the implementation of an automatic catheter contour detection program, no user intervention will be needed. This catheter contour must be drawn beyond the ring-down artifact (i.e., ‘flare’) so that this is not misclassified as necrotic core or dense calcium.

Although in the conventional method we would have included thrombi that are undetected by IVUS, recent papers showed that thrombus was always classified as fibrous and fibrofatty tissues by VH-IVUS [18, 19], and therefore even when the luminal contour would have included some thrombi, the necrotic core and calcium measurement using Shin’s method is not affected by a variable contour detection.

Although this study showed that the sum of necrotic core and calcium was very similar using both methods, the mean dense calcium area was consistently smaller and the mean necrotic area was larger in Shin’s method as compared to the conventional method. In particular, a total of five segments in dense calcium and three segments in necrotic core had an absolute difference in volume of more than 1 mm³. Possible explanations for this difference between two methods are as follows: (1) Thin plaque is treated as a grey medial stripe by VH-IVUS in the conventional method. However, Shin’s method is not influenced by the artificial thickness of the medial stripe. Therefore, Shin’s method shows more necrotic core even though there is a thin plaque (Fig. 6). (2) In the case of mixed plaques at the lumen, in the conventional method, the proximity of the contour with the plaque, will characterize the adjacent tissue to the contour using one of the two classification trees of the virtual histology system, resulting in more calcium or more necrotic core at that area (Fig. 6). In Shin’s method, a longer distance will be present from the catheter contour to the plaque, and therefore the second classification tree would be applied resulting in a slightly different tissue characterization. In a previous study, in vivo VH-IVUS images suggested an increase in dense calcium in the acute coronary syndrome group when compared with the stable angina group. Although in general the in vivo VH-IVUS data correlated well with in vitro histology (predictive accuracy of 96.5%), the RF data analysis overestimated the frequency of calcifications. The authors mentioned that one possible explanation was that the artifact is colored with white because the software for VH analysis is obliged to assign one of the four colors for each pixel [13].

Fig. 6
figure 6

A Shin’s method shows some necrotic core in the grey medial stripe (white arrow) that was not shown with the conventional method (white empty arrow). B In case of superficial calcium, there is much more calcium apparent behind the luminal contour in the conventional method (white arrow) while in Shin’s method, superficial calcium shows very finely and some decrease of calcium area. Also, some necrotic core shows behind the luminal contour (white arrow heads) in the conventional method (DC dense calcium, NC necrotic core)

Previous data with the conventional method showed acceptable reproducibility for plaque composition; likewise, repeated measurements for necrotic core and calcium with Shin’s method were highly correlated. Awareness of inter-observer differences may be particularly important in serial pharmacological intervention trials where small changes in plaque composition (i.e., necrotic core) may be expected [20, 21]. Therefore, Shin’s method may be a more adequate system to assess the change of necrotic core in longitudinal studies.

Shin’s method for measurement of necrotic core and dense calcium using VH-IVUS demonstrated a good correlation with the conventional method and excellent reproducibility. In the future, since Shin’s method could be automatically applied using an automated catheter contour detection with a computer program, and analysis time will be further decreased. This would also mean that the technique can be used in the catheterization laboratory to help operators in decision making based on content of necrotic core or calcium.

Limitations

This study has several limitations. (1) Because patients in the present study all suffered unstable angina, we cannot know how well the two methods would correlate in other atherosclerotic lesions. (2) The study did not include a validation component as therefore the accuracy of Shin’s method remains to be determined (3) Shin’s method does not show the lumen area and plaque area. Therefore, greyscale should remain the gold standard to assess plaque size changes (geometrical analysis), while virtual histology should be used to obtain compositional analysis, so as complementary information to greyscale. (4) Shin’s method does not provide correct quantification of fibrous or fibrofatty tissues.

Conclusions

Shin’s method for measurement of necrotic core and dense calcium using VH-IVUS demonstrated a good correlation with the conventional method and a high reproducibility. Also, Shin’s method required a significantly shorter analysis time than the conventional method. Therefore, Shin’s method could replace the conventional method for necrotic core and calcium measurement in atherosclerotic lesions, and it might be useful in the catheterization laboratory for online clinical decision.