Introduction

The combination of functional cardiovascular magnetic resonance (CMR) and late gadolinium enhancement (LGE) is evolving as an important diagnostic [1, 2] and prognostic [35] modality in patients with ischemic heart disease. Because the availability of high field MR systems is increasing, the need arises to evaluate the performance and clinical value of these systems in cardiovascular disease. Earlier reports already suggested that 3.0 T MR systems offer higher temporal and spatial resolution, due to an increased signal-to-noise ratio, which would be especially advantageous in CMR scanning [68]. However, these studies were all assessed in healthy volunteers and in patients with non-ischemic heart disease or suspected coronary artery disease, and without the presence of coronary stents [610]. There is limited to no data about safety and image quality at 3.0 T CMR scanning in the acute or chronic phase after myocardial infarction (MI), in patients treated with percutaneous coronary intervention and primary stenting. In these patients B0 inhomogeneities (induced by the heart-lung interface or from coronary stents), flow artifacts, inhomogeneity of normal myocardial suppression in LGE (induced by B1 inhomogeneity) and poor cardiac triggering may interfere with the gain offered by the higher magnetic field.

The aim of this study was to test whether CMR scanning at 3.0 T is safe and feasible in patients with coronary stents in the acute and chronic phase after MI, and to prospectively compare image quality at 3.0 T with the current clinical standard at 1.5 T. Furthermore, the presence and significance of different image artifacts are considered.

Methods

Patient population

Patients were eligible for the study if they had been admitted with a first ST-elevation acute MI, according to standard electrocardiographic and enzymatic criteria [11], and had undergone successful primary PCI with stent implantation. Exclusion criteria were electrocardiographic evidence of reinfarction, haemodynamic or other clinical instability or (relative) contraindications for CMR such as claustrophobia, pacemakers, intracerebral aneurysm clips or very irregular heart rhythm. Patients were treated with aspirin, heparin, abciximab, clopidogrel, statins, beta-blockade and ACE-inhibitors, according to ACC/AHA practice guidelines [12]. The study was approved by the local ethics committee and all patients gave written informed consent.

CMR parameters

Eighteen CMR scans were performed with a 3.0 T MR system (Intera, Philips, Best, The Netherlands), with a gradient performance of 30 mT/m and slew rate of 150 T/m/s, using a six element cardiac phased array surface coil. Another 18 CMR scans were acquired with a 1.5 T MR system (Magnetom Sonata, Siemens, Erlangen, Germany), with a gradient performance of 40 mT/m and slew rate of 200 T/m/s, using an eight element cardiac phased array surface coil.

Cine imaging was performed at both field strengths using a SSFP pulse sequence, without parallel imaging. Long axis views, as well as short axis views covering the entire left ventricle were acquired during repeated breath-holds in expiration. At 3.0 T, cine SSFP sequence parameters were a temporal resolution between 25 ms and 50 ms, excitation angle of 45°, receiver bandwidth 868 Hz/pixel, TR/TE of 3.8/1.9 ms, matrix 192 × 155 and voxel size of 1.5 × 1.8 × 6.0 mm3. At 1.5 T, the temporal resolution was between 35 ms and 50 ms with an excitation angle of 60°, receiver bandwidth 930 Hz/pixel, TR/TE of 3.2/1.6 ms, matrix 256 × 156 and voxel size of 1.4 × 1.9 × 6.0 mm3.

LGE images were acquired in mid-diastole using a 2D segmented inversion-recovery gradient-echo pulse sequence, 10–15 min after intravenous injection of 0.2 mmol/kg of a gadolinium chelate (Dotarem, Guerbet, Roissy, France). The LGE images were obtained in exactly the same orientation as the cine images. Sequence parameters at 3.0 T were an excitation angle of 25°, receiver bandwidth 434 Hz/pixel, TR/TE of 3.5/1.3 ms, matrix 192 × 119, voxel size of 1.7 × 2.1 × 6.0 mm3, triggering every other heart beat, and an inversion time between 250 ms and 350 ms to null remote myocardium. At 1.5 T, sequence parameters were an excitation angle of 25°, receiver bandwidth 130 Hz/pixel, TR/TE of 9.6/4.4 ms, matrix 256 × 166, voxel size of 1.4 × 1.7 × 6.0 mm3, triggering every other heart beat, and an inversion time between 220 ms and 300 ms to null remote myocardium.

Safety and quality analysis

To evaluate patient safety, a physician was present at the CMR scanner throughout the scan. Heart rhythm was monitored continuously. The patient was asked to report any discomfort and symptoms during the scan procedure. Additionally, a repeat catheterization was performed in 16 patients who underwent CMR examination in the acute phase after MI (9 at 3.0 T, 7 at 1.5 T). Stent position and patency after CMR scanning were visually assessed on the repeat catheterization and compared to the primary PCI. Occurrence of repeat intervention or hospitalization within 30 days of the CMR examination was recorded.

Different types of artifacts were reviewed and marked as being clinically relevant if the artifact interfered with visualisation of the myocardium. Flow related artifacts in SSFP imaging [13], and artifacts due to static field inhomogeneities such as the heart-lung interface were scored on cine images. Separately, we looked at artifacts caused by the coronary stent on the cine images, and measured the maximum artifact diameter, perpendicular to the length of the stent, using an appropriate cardiac view. In addition, the potential effect of B1 inhomogeneity on the homogeneous suppression of viable myocardium was visually assessed on LGE images.

Furthermore, image quality of the cine and LGE images were scored on a separate workstation (Centricity Radiology v6.1, GE Medical Systems, Zeist, the Netherlands) by four independent observers, who were blinded for MR system and clinical history. To distinguish between image quality that is satisfactory for clinical use or high quality images for research purposes, images were scored on a scale from 1 to 4: 1 not clinically useful; 2 clinically useful, but of insufficient quality for quantitative analysis; 3 clinically useful and of sufficient quality for quantitative analysis; 4 excellent quality. The following definitions were used for evaluation of the cine images: 1 poor quality, extensive artifacts, completely obscuring endocardial borders; 2 moderate quality, assessment of global function is possible, partly using assumptions, but regional wall thickening is not possible in all segments, due to interfering artifacts; 3 good quality, assessment of global and regional function is possible, despite some small artifacts; 4 excellent quality, functional analysis is possible and there is no interference of artifacts. And for the evaluation of LGE images: 1 poor quality, extensive artifacts, infarcted myocardium is not visible; 2 moderate quality, infarcted myocardium is visible, but delineation is not possible in all segments; 3 good quality, infarcted myocardium is easy to distinguish from viable myocardium, despite some small artifacts; 4 excellent quality, no artifacts. Two scores were given for each scan: 1 for the set of cine images and 1 for the set of LGE images. After completing the independent image quality assessment, all 4 observers exchanged their scores for each case and agreed on a consensus score.

Statistical analysis

Continuous variables with normal distribution are expressed as mean ± SD, or as median (25th–75th percentile). Comparison of the baseline characteristics and stent artifact diameter was done by using an unpaired Student’s t-test. The consensus score of image quality of both field strengths was assessed for statistical difference by a Mann-Whitney U test. All statistical tests were two-sided with a significance level of P < 0.05. SPSS 12.0.1 for Windows (SPSS Inc., Chicago, USA) was used for analysis.

Results

Twenty-five consecutive patients with 36 CMR studies were included in the study. One patient was studied on both MR systems in the acute phase, 9 patients (5 on 3.0 T, 4 on 1.5 T) were studied both in the acute phase and in the chronic phase, and 1 patient (3.0 T) was studied twice in the chronic phase after MI. The baseline characteristics of all 25 patients are listed in Table 1. On each system, 12 CMR scans were performed in the acute phase after MI, at 4 ± 2 days after primary PCI, and 6 CMR studies in the chronic phase, at 125 ± 23 days after primary PCI. One patient refused contrast injection during scanning in the chronic phase.

Table 1 Patient characteristics

ECG electrodes were positioned and displaced until an optimal ECG signal was acquired. Although it took more time to obtain a stable ECG waveform with a clearly delineated R wave at 3.0 T than at 1.5 T (10–15 min vs. <2 min respectively), it did not affect image quality, since scans were repeated when trigger problems occurred.

Stent safety and artifacts at 3.0 T

No patient reported any discomfort or symptoms during the CMR scan procedure. In addition, there were no clinical events during, or shortly after scanning at 3.0 T, and none of the patients underwent a repeat intervention or hospitalization for any reason within 30 days.

The mean number of stents implanted per patient per scan at 3.0 T was 1.3 ± 0.7, with a mean stent length of 19 ± 5 mm and diameter of 3.2 ± 0.4 mm. Further specifications of the implanted coronary stents are listed in Table 2. Repeat catheterization was performed 2 (1–4) days after scanning in 9 of the 12 patients who underwent 3.0 T CMR scanning in the acute phase after MI. Angiographic evaluation revealed no differences in stent position and patency compared to the initial result after primary PCI.

Table 2 Type and number of coronary stents used per CMR examination

During cine imaging, the coronary stent was visible in 14 of the 18 scans (78%), with a susceptibility related signal loss of 12.8 ± 4.3 mm in diameter. Due to severe interference of the coronary stent in 6 scans (33%), analysis of surrounding myocardium was impossible (Fig. 1A). Through-plane flow artifacts were present in 14 scans (from the aorta in 9 scans, from the pulmonary trunk in 7 scans, from the left ventricle in 3 scans), and hindered visualization of the myocardium in 7 scans (39%). In-plane flow artifacts were present in 4 scans, but were not clinically relevant. In the majority of the scans (94%) dark band artifacts appeared at the transition between myocardium and lung, with a predilection for the anterolateral wall, which resulted in severe image distortion in 8 scans (44%). Overall, artifacts were responsible for clinically relevant image deformation in 12 scans (67%), of which half was caused by the coronary stent (in 6 scans).

Fig. 1
figure 1

Short axis cine SSFP images in different patients, demonstrating a signal void from a coronary stent with myocardial interference at 3.0 T (A, white arrow head), and a smaller signal void without interference at 1.5 T (B, white arrow head)

No effect could be observed of B1 inhomogeneity on the suppression of viable myocardium on LGE images at 3.0 T.

Stent safety and artifacts at 1.5 T

Also during scanning at 1.5 T, no patient reported any discomfort or symptoms, and there were no clinical events during, or shortly after scanning. None of the patients underwent a repeat intervention or hospitalization for any reason within 30 days after CMR scanning.

The mean number of implanted stents per patient per scan at 1.5 T was 1.2 ± 0.5 (see also Table 2). The mean stent length was 19 ± 5 mm with a mean diameter of 3.2 ± 0.4 mm, which was comparable with the stent length and diameter scanned at 3.0 T (p = ns). In 7 of the 12 patients who underwent CMR scanning at 1.5 T in the acute phase, repeat catheterization at 2 (2–4) days after scanning revealed no differences in stent position and patency compared to post-PCI.

During cine imaging, the coronary stent was visible in 6 of the 18 scans (33%), with a susceptibility related signal loss of 5.5 ± 0.3 mm in diameter, which was significantly smaller than at 3.0 T (P < 0.01). Furthermore, visible stents did not cause clinically important image deformation (Fig. 1B). Besides the small signal void from coronary stents, no other artifacts were observed at 1.5 T.

Also at 1.5 T there was no effect of B1 inhomogeneity on the suppression of viable myocardium on LGE images.

Evaluation of image quality

With scanning at 3.0 T, localized shimming was compulsory in every scan to obtain better image quality (Fig. 2). At 1.5 T, there was no need for adjustments in sequence parameters or the use of local shimming to optimize image quality. At 3.0 T, in 44% of the cases, image quality was not sufficient for quantitative analysis and assessment of regional function of the left ventricle (Table 3). For clinical purposes and global assessment of left ventricular function, image quality at 3.0 T and 1.5 T were comparable. Image quality of LGE images was not sufficient for quantitative analysis in 53% of the cases at 3.0 T, and 12% of the images were not useful for clinical purposes. Both image quality of cine and LGE imaging at 3.0 T were significantly lower compared to 1.5 T (P < 0.001).

Fig. 2
figure 2

Four chamber cine SSFP images at 3.0 T in end-diastole (A & C) and end-systole (B & D), at the same slice position in one patient. Upper panels are without and lower panels with localized shimming. Dark band artifacts (apex) and flow artifacts (around the atrioventricular valves) are reduced with localized shimming

Table 3 Consensus score of image quality of cine and late gadolinium-enhanced images for 1.5 T and 3.0 T

Discussion

In this paper we report on our initial experience with CMR scanning at 3.0 T in a clinical situation. We found that it is safe and feasible to perform CMR scanning at 3.0 T in the acute and chronic phase after MI in patients treated with primary stenting. Image quality of cine imaging at 3.0 T is of sufficient quality for global assessment of left ventricular function, however, quantitative analysis is not possible in almost half of the patients, due to dark band, flow and stent artifacts. Image quality of LGE studies were significantly better at 1.5 T. Therefore further optimization of pulse sequences at 3.0 T is essential.

MR imaging is considered to be contraindicated in patients with ferromagnetic implants, primarily because of the potential risks associated with migration, the induction of an electrical current and heating of the implant [14, 15]. According to earlier reports and the American Society for Testing and Materials International, the overall magnetic field interaction and heating for coronary stents is limited or absent at 1.5 T and 3.0 T, as tested in vitro [1618]. In vivo studies concerning stent safety at 1.5 T demonstrated that CMR scanning is safe in patients early after coronary artery stent placement [1921]. In this in vivo study using both 1.5 T and 3.0 T, no clinical signs were observed by the attending physician, and no substantial side effects or clinical events occurred during or within 30 days of CMR scanning. In addition, in the patients who underwent a repeat catheterization there was no angiographic evidence of stent migration, confirming the in vitro data.

As discussed by Schär et al. [7], dark band and flow artifacts in SSFP cine imaging can be solved using optimized sequence parameters, localized shimming and correct water resonance frequency adjustment. In the clinical setting of the present study, it was not always possible to solve these artifact problems despite sequence optimization and localized shimming, especially in the presence of stent artifacts. The larger and diagnostically interfering artifacts of coronary stents at 3.0 T as opposed to the smaller stent artifacts at 1.5 T are an important issue in patients after MI, since primary PCI with stent implantation is the method of choice to re-establish coronary flow [22]. For clinical application, it may be more advantageous to return to spoiled gradient-echo cine imaging at 3.0 T, which includes longer acquisition times and lower contrast between blood and myocardium, but is less sensitive for off-resonance artifacts and equally accurate as SSFP cine imaging (Fig. 3) [9, 23].

Fig. 3
figure 3

Short axis cine SSFP image (A) and cine spoiled gradient-echo image (B), at the same slice position in one patient at 3.0 T. The coronary stent artifact on a SSFP image is larger than on a spoiled gradient-echo image (white arrows heads). Flow artifacts (A, asterisk) are less visible with spoiled gradient-echo imaging at 3.0 T

We initially intended to evaluate differences in signal- and contrast-to-noise ratios (SNR and CNR) of both pulse sequences as well. At 1.5 T good noise estimates could be made using the technique described by Constantinides et al. [24]. However, this method did not work at 3.0 T due to a different reconstruction algorithm, even without parallel imaging, and by switching off clear and image enhancement filters. Subtraction methods with two consecutive scans to estimate noise are inaccurate in a mobile tissue as the heart [25]. Other methods were beyond the scope of this initial study. Despite the fact that SNR’s and CNR’s were therefore omitted from this study, it is interesting to mention that an important reason for inferior quality of LGE images at 3.0 T was that infarcted myocardium was sometimes difficult to delineate from the left ventricular cavity (Fig. 4).

Fig. 4
figure 4

Short axis LGE images 15 min after injection of contrast at 3.0 T (A) and 1.5 T (B), in different patients. Visual assessment of infarct extent and location was more difficult at 3.0 T than at 1.5 T, since it was sometimes difficult to delineate infarcted myocardium from the left ventricular cavity at 3.0 T

We used the same contrast agent and dose at both field strengths. The contrast dose has been optimized for 1.5 T in the past [26, 27], but might be different at 3.0 T as T1 relaxation rates are in general field strength dependent. However, a recent study of Sharma and colleagues showed that there were only minor differences in post contrast myocardial T1 relaxation times between 1.5 T and 3.0 T, using a contrast dose of 0.2 mmol/kg [28]. As long as inversion times are set appropriately at each field strength as done in this study, it is not to be expected that significant differences in LGE image quality are caused by contrast dose effects.

In conclusion, our study demonstrates that it is safe and feasible to perform CMR scanning at 3.0 T in the acute and chronic phase after MI in patients treated with primary stenting. Although cine imaging at 3.0 T is of sufficient quality for clinical use, quantitative assessment is less reliable compared to 1.5 T, mainly due to dark band, flow and stent artifacts. Further optimization of pulse sequences at 3.0 T is essential to make 3.0 T CMR scanning suitable for clinical cardiology.

Limitations

The use of different MR systems, from different vendors, with different coils and sequence parameters, of course introduces confounding factors for a comparison. However, the sequences were optimized for their field strength to evaluate safety, feasibility and image quality rather than technical differences. A second limitation is that patients did not undergo a CMR examination at both 1.5 T and 3.0 T. Because the MR systems were on two different locations, it was not feasible to study the same patient twice in the acute phase after MI.