Clinicians working on the frontlines have continued to adapt to the ever-changing environment that is living and working during the COVID-19 pandemic. Despite the many precautions that clinicians have taken against COVID-19, there are almost 10,000 healthcare workers that have been diagnosed with coronavirus as of April 17th, 2020, and at least half of these were contracted through patient contact [1].

In light of these historic events, many surgical organizations including the Society of American Gastrointestinal and Endoscopic Surgeons (SAGES), the European Association of Endoscopic Surgery (EAES), the American College of Surgeons (ACS), and many others have issued recommendations to postpone elective surgeries [2,3,4,5,6,7,8]. In addition to these recommendations, SAGES encourages the use of minimally invasive surgery and robotic surgery techniques whenever possible. Implementing minimally invasive surgical techniques help reduce hospital stay time for patients, allowing a greater capacity of empty beds for the treatment of COVID-19 while simultaneously minimizing clinician exposure to potentially contagious patients [2]. Extreme intervention efforts are being made in the medical community to mitigate healthcare worker exposure to the contagious virus, with notable focus on exposure in the Operating Room (OR) [9].

Aerosolized viral particles in surgery

While the use of minimally invasive surgery is thought to mitigate the exposure of clinicians to contagious patients compared to other clinical settings, it is not without risk! Greater concentrations of coronavirus particles have been found in peritoneal fluid than in the respiratory tract, which demonstrates a previously unknown risk to surgeons and their staff during surgery [10]. The laparoscopic environment also lends itself to the aerosolization of tissue particles. Specifically, the use of certain laparoscopic instrumentation such as the ultrasonic scalpel, electrocautery, and laser ablation have been shown to aerosolize tissue particles into smoke during surgery [11]. Furthermore, active viral particles have been shown to aerosolize via the same mechanism. Even more concerning is the fact that this space is not yet well-understood or -explored, as few studies have been performed to understand this issue at levels sufficient enough to garner strong understanding and scope of the potential problem [11,12,13,14,15,16,17]. However, those few studies have shown that viruses such as Hepatitis B and HPV have been shown to aerosolize in the smoke evacuated during laparoscopic surgery as well as during open surgery [11, 18,19,20,21,22,23,24,25]. The known aerosolization of these other viruses combined with several new articles documenting COVID-19 in the peritoneal fluid have led numerous international surgical societies to encourage exercising caution when performing minimally invasive surgery [2,3,4,5,6,7,8].

In addition to evacuated smoke, there is evidence to suggest that leaked CO2 gas from the insufflation of the pneumoperitoneum could contain aerosolized virus particles [11]. This leaked CO2 may contain viral particles as shown in the previous studies, suggesting a serious concern to the health and safety of clinical professionals in the OR, especially given the growing concern regarding the potential aerosolization of coronavirus particles [2, 3, 11, 18, 19, 21, 22, 26]. Due to the highly contagious nature of COVID-19, and our relative lack of understanding of its mode of transmission via aerosolization, Kwak et al. [18] further recommend that any measures taken to minimize gas leakage during laparoscopic surgery should be considered for the safety of operating room healthcare providers. Relevant studies of coronavirus concerning the aerosolization of COVID-19 and infection of surgeons or other medical staff through these aerosols remain relatively unexplored, but the issue continues to be a growing concern [15, 27,28,29,30]. Further support of this notion is the fact that both SAGES and the EAES strongly recommend the considered use of devices capable of ultra-filtration regarding released CO2 and smoke due to the concern for aerosolized COVID-19 particles [2].

As shown, even with limited data surrounding this problem, there is significant concern among surgical societies and operating rooms across the globe. Although data is limited, numerous studies suggest reducing exposure to laparoscopic gases in order to mitigate exposure of COVID-19, and other viruses, to operating room personnel [31,32,33,34]. As such, it is clear that more research is needed in these areas to help understand and quantify potential exposure concerns in the OR.

Need for more research to accurately quantify the problem

While these suggestions attempt to address concerns (primarily) regarding “controlled” gas leakage, there are a number of uncontrolled gas leak mechanisms that have been postulated by the medical community, yet also remain relatively unexplored, around the trocar port and through instrument changes. Early stage studies have shown prominent attention given to gas leakage through the trocar (port) [2, 35]. And no data currently exists in literature that attempts to quantify gas leakage during instrument use and exchange, despite the fact the literature shows that such events are overly abundant in every laparoscopic procedure [36,37,38]. More work is needed to better understand the level of risk of clinician safety in today’s OR to guide us to a safer OR of the future around this rising issue, brought to the forefront of people’s minds during this global COVID-19 pandemic. The proposed study focuses on better quantifying gas leakage regarding instrument–port related events in order to better understand potential risks to today’s OR clinicians.

Materials and methods

IRB approval was not required for this study. The report details a test setup and results evaluating the gas leaked while simulating realistic scenarios happening during a laparoscopic surgery using two size trocar configurations, 5 mm and 12 mm trocar ports (Applied Medical Kii Balloon Blunt Tip System 12 × 100 mm & Applied Medical Kii Balloon Blunt Tip System 5 × 100 mm; Ranch Santa Margarita, CA), and two laparoscopic instruments, laparoscopes measuring 5 mm (Stryker 502-555-030; Kalamazoo, MI) and 10 mm (Richard Wolf 8934-442 Panoview; Vernon Hills, IL) each.

Test setup

A photo of the benchtop test setup can be seen in Fig. 1 [39]. The setup uses a compressor (California Air Tools CAT-1P1060S; San Diego, CA) combined with a regulator (30222 Fairchild Model 30 Compact Precision Regulator; Winston-Salem, NC) and pressure gauge (General Electric Druck DPI 104-IS; Leicester, UK) to simulate a laparoscopic insufflator to maintain a consistent and reliable pressure. The regulator has an on/off valve that controls the flow of compressed air to a mass flow meter (Model FMA1820A; Norwalk, CT) that is used to measure the leak flow rate of the system and records data via a data logger.

Fig. 1
figure 1

Physical benchtop and diagram outlining experimental setup and correlating setup in an OR environment

Next, gas travels to an insufflation chamber, comprised of a plastic tube with an elastic seal at the proximal end to mimic compliant tissue through which a trocar port is placed for testing. Data were logged using a National Instrument NI4350 data logger. The insufflation chamber was made using a 1 ft PVC tubing, on which the distal end was connected to a compliant rubber tubing from which to maintain insufflation pressure and monitor any insufflation gas leakage.

The simulated benchtop pressure-management system essentially performs the same function as a laparoscopic insufflation system, proving quite adequate as simulated OR equipment. The plastic tube of the insufflation cavity is rigid (unlike a patient’s abdomen) and the thin elastic seal is notably compliant (though not quite as compliant as human tissue). While this is a shortcoming compared to testing in a real patient or cadaver model, the benchtop setup is still sufficient enough to yield functional results surrounding data trends. Limitations of the benchtop experiment could be improved in future iterations but suffice for this exploratory experiment that aims to determine potential health hazards for clinicians in today’s OR.

Exploratory test configurations are outlined in Table 1 for instrument–port sizes and combinations. Shorthand for laparoscopic instrument levels are designated by “L” followed by dimension (i.e. 5 mm instrument level henceforth known as L5 and 10 mm as L10). Shorthand for trocar port levels are designated by “T” followed by dimension (i.e. 5 mm trocar port level henceforth known at T5, and 12 mm as T12).

Table 1 Test configurations [39]

The desired trocar configuration was inserted into the elastic seal end of the PVC pipe. The full length of the trocar was inserted creating a seal around the inner dimeter port the outer diameter of the trocar stem, simulating a best-case scenario seen in laparoscopic surgery between the port and tissue. Once the desired configuration was setup up the following steps were followed for each clinical situation. A “null” (i.e. trocar-only) configuration was tested to understand potential trocar port or elastic seal-trocar port interaction leakage. Such data was used to confirm that the test setup was capable of sustaining pressure to ensure the test setup had no leaks before instrument–port interaction was initiated.

Test procedure for each use case scenario

Each configuration and use case included in this study was designed to simulate typical events observed in today’s laparoscopic OR. The Null configuration scenario was used to simulate a trocar port placed in a patient without any instrument being used with that port at a specific time. While this configuration is not common during surgery (no human instrument or scope driver is often “perfectly still”), it might be representative in nature (though not in number) to robotic surgeries, where a robot is able to hold a scope near-perfectly still. This configuration also doubled as a baseline test for the experimental setup. The Dynamic Instrument–Port Interaction scenario represented a common use case where a clinician might be moving a laparoscope, cutters, cautery, or other such elongated instrument within a trocar to perform surgery. Essentially, in laparoscopic and robotic surgeries, instruments are nearly constantly moving and interacting with trocar ports/valves, suggesting that these results indicate a real-time representation of OR instrument usage. The Instrument Full Insertion/Removal scenario was used to simulate an instrument exchange or other such removal and/or insertion event, which is commonly experienced during laparoscopy for each and every case. For example, scope removals are often performed to clean the lens, and instruments are commonly removed to achieve access through a different port or to exchange them out for a new instrument [36, 37, 40,41,42,43]. Each instrument–port configuration also helped to represent a few different levels of such combinations seen in today’s OR.

  1. (1)

    Null configuration scenario:

    1. 1.

      Open valve and stabilize insufflation chamber pressure.

    2. 2.

      Rotate valve counterclockwise to the closed position to depressurize chamber.

    3. 3.

      Confirm stable pressure remains and no leaks are present in the test setup.

  1. (2)

    Dynamic Instrument–Port Interaction scenario:

    1. 1.

      Insert laparoscopic instrument by ¾ of its full length through the port (~ 9 inches) (Fig. 2).

      Fig. 2
      figure 2

      Dynamic instrument–port interaction setup with 3 inch offset from full device insertion

    2. 2.

      Open valve and stabilize insufflation chamber pressure.

    3. 3.

      Insert remaining partial length, approximately 3 inches, of the laparoscopic instument until it bottoms out on the trocar.

    4. 4.

      Wait approximately 10 s for pressure to stabilize.

    5. 5.

      Remove the instrument partially 3 inches.

    6. 6.

      Wait approximately 10 s for pressure to stabilize.

    7. 7.

      Repeat steps 3–6 two more times.

    8. 8.

      Wait 5 s for pressure to stabilize.

  2. (3)

    Instrument Full Insertion/Removal scenario:

    1. 1.

      Open valve and stabilize insufflation chamber pressure.

    2. 2.

      Rotate valve counterclockwise to the closed position to depressurize chamber.

    3. 3.

      Fully insert the laparoscopic instrument through the port.

    4. 4.

      Wait approximately 10 s for pressure to stabilize.

    5. 5.

      Fully remove the laparoscopic instrument from the port.

    6. 6.

      Wait approximately 5 s for the pressure to stabilize.

Results

Null configuration scenario

5 mm and 12 mm trocar ports were used for each configuration (n = 1 device and n = 3 trials per configuration). Results confirm that both ports have negligible gas leaks when integrated with the test setup. The 5 mm port registered an average 0.02 L/min leak rate, while the 12 mm port registered no leak. While the reason for the small 5 mm leak could be attributed to either (1) a manufacturing/device design trait or (2) a test setup interaction with the port, the result of a negligible leak remains consistent and the setup was shown to be highly reliable.

Dynamic instrument–port interaction scenario

The dynamic instrument–port interaction scenario involved three advancement and retraction events per trial and was performed for each instrument–port configuration (n = 3 trials per configuration). Figure 3 shows graphs displaying results for both individual and averaged trials. T5–L5 averaged 43.67 mL of leaked gas per trial while T12–L5 and T12–L10 averaged 41.32 mL and 55.68 mL of leaked gas, respectively.

Fig. 3
figure 3

Gas leaked during dynamic interaction of instrument–port devices. Best viewed in color

Instrument full insertion/removal scenario

The laparoscopic instrument full insertion/removal scenario outlined above was performed for each instrument–port configuration, with additional trials performed in comparison with previous test configurations in an attempt to counteract additional noise variables experienced for this protocol (n = 10 trials per configuration). Graphs observed in Fig. 4 show results for individual and averaged trials. Test configuration combinations show different levels of gas leakage for simulated instrument insertion and removal events: T5–L5 = 25.97 mL, T12–L5 = 29.47 mL, and T12–L10 = 58.59 mL.

Fig. 4
figure 4

Gas leaked during insertion and removal of a laparoscopic instrument through a trocar. Best viewed in color

Discussion

Insufflation leakage during laparoscopic surgery has become a prevalent topic of conversation in surgical societies around the globe in light of the COVID-19 outbreak. The data presented quantifies the amount of potential leakage that may be attributed to instrument interactions with the trocar port, including typical surgical dynamic use as well as insertion/removal events during laparoscopic surgery.

Specifically, for the full insertion/removal scenario, once the instrument was inserted into the trocar flow rate began to increase after initial leakage from opening of the trocar port valve occurred, and then stabilizes with a stationary instrument. This suggests that a stationary instrument inside the trocar port still produces a constant gas leak. Then, as the instrument was removed, the flow rate appears to reach a peak, demonstrating a higher amount of gas leakage during dynamic port valve interaction with instrumentation (further supported by data from the Dynamic Instrument–Port Interaction protocol). After the instrument was fully removed, the flow rate then approaches its baseline as the trocar port valve is allowed to fully seal again since no instrumentation is present to keep it open. Data from this study show that instrument interaction with a trocar port, especially considering instrument–port dynamic interaction/motion as well as full instrument insertion and removal events, stands to produce gas leakage across 5 mm and 10 mm instruments in 5 mm and 12 mm trocars. Further work should consider additional levels of instrumentation-port interactions and configurations regarding different laparoscopic instrument motions and sizes/dimensions. Additionally, similar testing of instrument–port interactions in a clinical setting would more directly show the amount of gas leaked during a surgery from this leakage source and should further look to understand volumes of aerosolized viral particles exposed to the OR environment during surgery to better understand potential instances of exposure to airborne particulates. In addition, only “insufflation pressures” of 15 mmHg were tested. This is a common pressure used during laparoscopic surgery [44]. Currently, recommendations from surgical societies such as SAGES that consider decreasing leakage include using lower intra-abdominal pressures when technically possible. It is possible that higher or lower insufflation pressures could affect results. Additionally, the test setup might benefit from additional alterations to more closely overcome the limitations discussed in ‘Materials and methods’ of this article.

Further extrapolation of the presented data can be considered for insufflation leakage estimated to be released during 1 h of laparoscopic surgery (Table 2). The dynamic port–instrumentation interaction over 1 h of operating time may be estimated at over 2000 mL for the proposed configurations. For a typical laparoscopic operation with four ports, this translates to the same volume as 4 two-liter bottles of soda, potentially containing aerosolized contaminants, being released into the OR environment from the patient’s surgical cavity. Instrument insertion and removal events (conservatively estimated at 6 events/h in the OR) appear to release additional potentially harmful gas volumes equal to 1–2 cans of twelve ounce sodas [43]. Using an average of 6 instrument removals per hour, results from this study might then suggest a range of 324–684 mL of potentially contaminated gases released during 1 h of surgery (i.e. roughly one or two 12 oz cans of soda).

Table 2 Clinically translatable estimates

There are opportunities to stop the leakage of gases during laparoscopic surgery in order to protect the OR team from potential viral exposures. This experiment demonstrates that both dynamic instrument interaction with the trocar and instrument insertion and removal will contribute leaked gas into the operating room. Concerns for exposing clinical personnel to COVID-19 significantly reduced elective surgery around the globe for months while COVID restrictions were put in place, which had huge economic impact to hospitals across the world; Ascension, which has 150 hospitals, posted a loss of $2.7 billion in the first quarter [45]. Products that allow for reduction of laparoscope and laparoscopic instrument removal would eliminate the need to remove the scope and intstruments during surgery and have significant impact on decreasing gas leakage. Additionally, modifications to the ports could result in decreasing leakage when the instruments are being used. In light of COVID-19 and the expressed danger that may be inherent in laparoscopic surgery due to aerosolized viral particles, serious consideration should be given to new technologies or strategies that minimize risk to the clinicians caring for these patients.