Abstract
Purpose
Overhanging facial nerve (FN) may be challenging in imaging diagnosis. The purpose of the study is to investigate the imaging clues for overhanging FN near the oval window on ultra-high-resolution computed tomography (U-HRCT) images.
Methods
Between October 2020 and August 2021, images of 325 ears (276 patients) were included in the analysis obtained by an experimental U-HRCT scanner. On standard reformatted images, the morphology of FN was evaluated and its position was quantitatively measured using the following indices: protrusion ratio (PR), protruding angle (A), position of FN (P-FN), distance between FN and stapes (D-S), and distance between FN and anterior and posterior crura of stapes (D-AC and D-PC). According to the FN morphology in imaging, images were divided into overhanging FN group and non-overhanging FN group. Binary univariate logistic regression analysis was used to identify the imaging indices independently associated with overhanging FN.
Results
Overhanging FN was found in 66 ears (20.3%), which manifested as downwards protrusion of either local segment (61 ears, 61/66) or the entire course near the oval window (5 ears, 5/66). D-AC [odds ratio: 0.063, 95% CI 0.012–0.334, P = 0.001) and D-PC (odds ratio: 0.008, 95% CI 0.001–0.050, P = 0.000) were identified as independent predictors of FN overhang (area under the curve: 0.828 and 0.865, respectively).
Conclusion
Abnormal morphology of the lower margin of FN, D-AC and D-PC on U-HRCT images provide valuable diagnostic clues for FN overhang.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The tympanic segment of the facial nerve (FN) is one of the important structures in the medial wall of the tympanic cavity in the temporal bone. It is located at the upper edge of the oval window niche, above the stapes. However, the position of the FN may vary, which is one of the key challenges hindering adequate exposure in oval window-related surgery [1]. The low position of the FN, also referred to as overhanging FN, may cause narrowing of the oval window niche, making surgical treatment challenging [2]. Iatrogenic facial paralysis is a very rare but always threatening complication of oval window-related surgery [1, 3,4,5]. Preoperative imaging can be used to provide information regarding FN anatomy. However, the diagnosis of overhanging FN is challenging for non-otological radiologists, and heavily depends on the expertise of the viewer. A good anatomical knowledge of FN and objective measurement methods can provide in-depth characterization of overhanging FN and improve the diagnostic accuracy.
Computed tomography (CT) is widely used in temporal bone imaging. However, to the best of our knowledge, there is a paucity of studies that have focused on overhanging FN near the oval window [6, 7]. We did not find any detailed imaging studies of overhanging FN. This may be attributable to the limited spatial resolution of CT for delineating the fine anatomical changes of the FN. Ultra-high-resolution computed tomography (U-HRCT), a newly designed imaging device based on cone-beam computed tomography (CBCT), features a significantly higher spatial resolution [8]. It has been shown to provide better delineation of stapes and the FN than conventional 128-section multislice CT (MSCT) [9].
The purpose of this study was to identify the imaging diagnostic clues for overhanging FN using this new U-HRCT.
Methods
Study population
This retrospective, single-center study was approved by the medical ethics committee of the hospital (IRB: 2020-P2-061-02) and informed consent was obtained from all the participants. Between October 2020 and August 2021, U-HRCT images of ears of patients were collected and those who qualified the following inclusion and exclusion criteria were enrolled in the study. The inclusion criteria were: (1) age ≥ 18 years; (2) absence of any symptoms of facial nerve lesion; (3) without lesions in the tympanic cavity and inner ear in images. The exclusion criterion was poor image quality (presence of motion artifacts or incomplete images).
Image acquisition
The ears were examined by an experimental U-HRCT scanner (Ultra3D-SE, LargeV, Beijing, China). The device was integrated with a small-focus X-ray generator, a high-stability motion control unit, and a high-resolution flat-panel detector (unit size: 0.0748 mm × 0.0748 mm). Unilateral temporal bone images were obtained on each scan. The scan protocols were as follows: 100 kV; 3.5 mA; voxel size: 0.1 mm × 0.1 mm × 0.1 mm; field of view: 65 mm × 65 mm. exposure time: 40 s; the number of images: 370.
All images were exported in DICOM file format and archived in a three-dimensional workstation using special software (RadiAnt DICOM Viewer, RadiAnt, Poland) for image analysis.
Image analysis
(1) Quantitative measurement of FN position
Using the RadiAnt software, standard coronal images of tympanic segment of the FN were reformatted. The coronal plane was perpendicular to the lateral semicircular canal, and this is a commonly used reformatted section in clinical work [10,11,12].
To evaluate the position of the tympanic segment of the FN, the following indices were measured on coronal images (Fig. 1):
(a) Protrusion ratio (PR). The vertical diameter of FN (D-FN) and protrusion distance relative to upper border of oval window niche (D-OWN) were measured; PR was calculated as D-OWN/D-FN. The measurements were done on the image passing through the middle of the stapes footplate.
(b) Protrusion angle (A). The angle between the line connecting the lower margin of FN and the upper edge of oval window and the line of the upper border of oval window niche was measured on the same image as point (a).
(c) Position of FN (P-FN). The position of FN relative to the oval window was measured. The distance between the lower margin of FN and the plane of the upper point of oval window was measured on the same image as point (a).
(d) Distance between FN and stapes (D-S). The distance between the lower point of FN and the stapes was measured on the same image as point (a).
(e) Distance between FN and anterior and posterior crura of stapes (D-AC and D-PC), which was measured on the images demonstrating the anterior or posterior crus.
(2) Overhanging FN observation
Using the RadiAnt software, coronal and oblique sagittal images were reformatted and used to observe the tympanic segment of FN. The coronal plane for FN was perpendicular to the tympanic segment of FN. The oblique sagittal plane was parallel to the tympanic segment of the FN and was perpendicular to the posterior crus of the stapes (Fig. 2). The assessment range of the tympanic segment of FN near the oval window was defined as the region between the cochleariform process and the anterior edge of the pyramidal eminence.
Based on the morphological description of overhanging FN reported in published literature [5, 13], when the lower margin of the FN near the oval window was partially or completely moving downward on images, it was defined as overhanging FN. The observation was made on the oblique sagittal and coronal images reformatted for FN.
According to the observation results, the FN images were divided into two groups: overhanging FN group and non-overhanging FN group. Then, the two groups were statistically analyzed.
The U-HRCT images were independently evaluated by two radiologists with 5-year experience (R.W.T.) and 20-year experience (Z.Y.Z.), respectively, in head and neck imaging diagnosis. Continuous variables presented as the average of the results measured by the two radiologists. If there was a difference in the observation of prolapse, the conclusion was drawn by consensus.
Statistical analysis
Statistical analysis was conducted using IBM SPSS Statistics version 22.0. The Kolmogorov–Smirnov test was used to test for normality of distribution of variables. Continuous variables are expressed as mean ± standard deviation or median (interquartile range) depending on the normality of distribution. Interrater repeatability was assessed by calculating the intraclass correlation coefficient (ICC). The ICC values were interpreted using the benchmarks suggested by Landis and Koch [14]: 0.81–1 almost perfect, 0.61–0.8 substantial, 0.41–0.6 moderate, 0.21–0.4 fair, and 0–0.2 slight.
Between-group differences with respect to continuous variables were assessed using the Mann–Whitney U test, while those with respect to categorical variables were assessed using the Chi-squared test. Variables that showed significant between-group differences (P < 0.05) were selected for multivariate analysis. Binary univariate logistic regression analysis according to a forward stepwise method was used to identify the imaging indices independently associated with overhanging FN. MedCalc (version 19.0, Ostend, Belgium) was used to plot the receiver operator characteristic (ROC) curve and the area under the curve (AUC) was calculated to analyze the diagnostic performance of different imaging indices for overhanging FN. P values < 0.05 were considered indicative of statistical significance.
Results
U-HRCT images of 340 ears (285 participants) were initially evaluated. Motion artifacts were found in the images of 15 ears (9 participants) which qualified the exclusion criteria. Finally, 325 ears (left 135, right 190) in 276 participants (128 male, 148 female; mean age 40.4, age range 18–83) were enrolled in this study.
Imaging morphology of overhanging FN
The morphology of the tympanic segment of FN near the oval window was well demonstrated on oblique sagittal images. In 66 ears (66/325, 20.3%), the lower margin of the FN protruded downwards partially (61 ears, 61/66) or totally (5 ears, 5/66) (Fig. 3), which is the direct sign of overhanging FN. On the coronal images reformatted for FN, all the overhanging FN exhibited a sudden change in morphology of FN, beyond the original position (Fig. 3). These 66 ears were classified in overhanging FN group. The other 259 ears were included in non-overhanging FN group. The specific site of protrusion of FN and the demographic characteristics of the two groups are summarized in Table 1.
Quantitative indices of overhanging FN
The values of the indices are listed in Table 2. The protrusion ratio (PR) and the angle (A) of FN protrusion relative to upper edge of oval window niche were significantly larger in the overhanging FN group (P < 0.05). The distances between the lower margin of the FN and the stapes, anterior and posterior crura were significantly smaller in the overhanging FN group (P = 0.000) (Table 2) (Fig. 4).
All the measured indices were significantly different in the overhanging FN group and were included in the binary univariate logistic regression model. Finally, D-AC [odds ratio (OR): 0.063, 95% CI 0.012–0.334, P = 0.001) and D-PC (OR: 0.008, 95% CI 0.001–0.050, P = 0.000) were identified as independent predictors of FN overhang. On ROC curve analysis, the AUCs of D-AC and D-PC were 0.828, and 0.865, respectively (P < 0.001) (Fig. 5). The sensitivity and specificity of D-AC and D-PC for predicting FN overhang, and the optimal cutoff values are shown in Table 3.
Discussion
In the present study, we used a new device U-HRCT to demonstrate the imaging characteristics of overhanging FN in the oval window niche region. Using the direct imaging sign of overhanging FN as the standard, we identified the quantitative indices with high predictive value for FN overhang. Our findings may facilitate preoperative diagnosis of overhanging FN, which is important for oval window-related surgery.
FN overhang refers to a kind of variation in the course of the FN, which appears as variable degrees of inferior displacement, in close proximity to the stapes. In cases with FN overhang, the FN may partially or completely cover the stapes [5, 13, 15]. The overhanging FN may narrow the oval window increasing the difficulty in performing surgery. Therefore, FN overhang is of much concern for otology surgeons and the associated challenges are described in surgical studies [1, 4, 5, 13, 16]. Preoperative detection of the change in the anatomical course of the FN may help the surgeon in predicting an intraoperative challenge. Temporal bone imaging is the best method to evaluate the FN canal [7, 10, 17, 18]. However, the imaging characteristics of overhanging FN are not well characterized in the contemporary literature. In the current study, we found that the U-HRCT images could demonstrate the downward protrusion of the lower margin of the tympanic segment of FN, which was consistent with the appearance described in previous surgical studies [5, 13]. The degree of protrusion of FN was variable, from slight to severe prolapse enough to cover the stapes, which was in accordance with the previous reports [5, 13]. The reported prevalence of overhanging FN in previous studies ranged from 2.5 to 32%, which were all based on intraoperative observation [2, 13, 19]. In the present study, overhanging FN was identified in 66 ears (66/325, 20.3%) using direct imaging signs. The differences in the prevalence rates may be attributable to differences in research methods and study populations. In our experience, the reconstructed oblique sagittal plane which is parallel to the tympanic segment of the FN and is perpendicular to the posterior crus of the stapes is recommended for better delineation of the overhanging FN. This plane can help detect the abnormal morphology of the lower margin of the FN and can simultaneously display the relationship between the overhanging FN and the stapes. On the coronal images reformatted for FN, overhanging FN should be continuously observed from the front section to the back, and presents as a sudden downward protrusion of FN from its previous course. But these two special reformation methods may be difficult and complicated. While on the commonly used coronal images which are perpendicular to the lateral semicircular canal, the FN is in an oblique section, and some small morphological changes may be difficult to observe. Therefore, we tried to develop quantitative indices to facilitate diagnosis. Our findings suggest that the distance between the FN and the anterior or the posterior crus of the stapes are both good predictors of overhanging FN. If the distance is smaller than 0.975 mm or 0.69 mm, respectively, an overhanging FN should be taken into consideration.
The overhanging FN moves downwards in close proximity to or covering the stapes [1, 4]. In the current study, we measured the FN protrusion ratio, angle, the position of FN relative to the oval window and the distance between FN and stapes to evaluate FN position. These quantitative indices were chosen because they are simple and convenient to measure using precise and reproducible anatomic landmarks, making the evaluation visual and repeatable. The results showed that in ears with overhanging FN, the FN protruded downwards to the oval window niche, and the distances between FN and crura of stapes were significantly smaller, which was consistent with the intraoperative findings reported in previous studies. Among the six indices used in the study, the other four items showed no statistical significance for diagnosis of FN prolapse, although their values differed between the prolapse and non-prolapse groups. To the best of our knowledge, overhanging FN has not been evaluated in images in previous studies. Some scholars have measured the height of oval windows and studied its correlation with the degree of surgical difficulty [4, 10, 20]. Overhanging FN was also described to be one of the main causes leading to oval window niche narrowing; however, there were no overhanging FN measurements. The indices evaluated in this study will facilitate understanding of the imaging manifestations of overhanging FN and provide a basis for diagnosis. In previous studies, overhanging FN was categorized as dehiscent overhanging FN and bony overhanging FN [2, 5, 19]. In the former, the FN canal was dehiscent leading to FN herniation, while in the latter, the FN canal was intact with a low location. Owing to the thin wall of the FN canal, it is typically poorly displayed in images. The results of previous imaging studies on the bone continuity of FN canal have not been consistent [7, 17, 21]. In our study, the FN canal wall was not evaluated to avoid observation errors. In cases with obvious overhanging FN in the present study, the lower margin of the FN was significantly protruded, deviating from the running area of the original FN canal, and the local wall was not demonstrated. We conjectured that this was a manifestation of dehiscent FN canal and overhanging FN. A comparative study with intraoperative findings will be further explored.
Although CT is the preferred method for temporal bone imaging, fine and complex structures are not displayed well because of its limited spatial resolution [6, 22]. U-HRCT is a newly developed high-resolution equipment for temporal bone imaging, which is based on CBCT. CBCT offers advantages over the current-generation 64- or 128-slice MSCT in temporal bone imaging in terms of higher spatial resolution and lower radiation dose [23, 24]. U-HRCT is currently being evaluated in clinical trials. Its spatial resolution is equal to or greater than 4 line pairs per millimeter (l p/mm), with the routine reconstructed voxel size of 0.1 mm × 0.1 mm × 0.1 mm [8]. This ultrahigh spatial resolution helps identify small imaging changes in the temporal bone. In this study, U-HRCT was used to gain clear images of the tympanic segment of FN that allowed us to observe the morphological changes of overhanging FN, which was an imaging supplement to understand overhanging FN.
Some limitations of our study should be acknowledged. First, the overhanging FN was observed in images without comparison with intraoperative findings. The U-HRCT is still an experimental equipment. There were not enough cases for which our findings could be compared with intraoperative findings. The morphology of the overhanging FN described in the study was consistent with our previous experience and literature reports. We will accumulate more imaging-surgical comparison cases for further study. Second, the radiation dose of U-HRCT was not provided in this study. The U-HRCT device used in the study does not have the capability for real-time detection of radiation doses. In a previous study, the radiation dose of this U-HRCT with the same scanning conditions as used in the study was found to be approximately 1/3 of that of conventional MSCT [25]. Lastly, owing to the very small size of the measured anatomical structures, the possibility of measurement error cannot be ruled out. However, to minimize the error, two radiologists separately measured the indices using standardized measurement methods. The ICC values showed moderate to almost perfect agreement.
Conclusion
We identified some imaging diagnostic clues for overhanging FN in U-HRCT images, which included the abnormal morphology of the lower margin of FN and the distance between FN and the anterior and posterior crura of stapes. All these indices can provide valuable information for imaging diagnosis of overhanging FN.
Availability of data and materials
No additional data are available.
References
Fernandez IJ, Bonali M, Fermi M, Ghirelli M, Villari D, Presutti L (2019) The role of endoscopic stapes surgery in difficult oval window niche anatomy. Eur Arch Otorhinolaryngol 276:1897–1905. https://doi.org/10.1007/s00405-019-05401-z
Daniels R, Krieger L, Lippy W (2001) The Other Ear: findings and results in 1800 bilateral stapedectomies. Otol Neurotol 22:603–607
Zourntou S, Makridis KG, Tsougos C, Skoulakis C, Vlychou M (2021) Facial nerve: a review of the anatomical, surgical landmarks and its iatrogenic injuries. Injury 52:2038–2048. https://doi.org/10.1016/j.injury.2021.05.029
Parra C, Trunet S, Granger B, Nguyen Y, Lamas G, Bernardeschi D, Hervochon R, Tankere F (2017) Imaging criteria to predict surgical difficulties during stapes surgery. Otol Neurotol 38:815–821. https://doi.org/10.1097/MAO.0000000000001417
Senturk E, Eren SB, Aksoy F, Dagistanli N, Yenigun A, Degirmenci N, Dogan R, Ozturan O (2020) Stapedectomy in patients with dehiscent and prolapsed facial nerve. Am J Otolaryngol Head Neck Med Surg 41:102580. https://doi.org/10.1016/j.amjoto.2020.102580
Bremke M, Luers JC, Anagiotos A, Gostian AO, Dorn F, Kabbasch C, Unkel C, Höllering J, Beutner D (2015) Comparison of digital volume tomography and high-resolution computed tomography in detecting superior semicircular canal dehiscence-a temporal bone study. Acta Otolaryngol 135:901–906. https://doi.org/10.3109/00016489.2015.1042041
Zhang Z, Yin H, Wang Z, Li J, Lv H, Zhao P, Yang Z, Wang Z (2019) Imaging re-evaluation of the tympanic segment of the facial nerve canal using cone-beam computed tomography compared with multi-slice computed tomography. Eur Arch Otorhinolaryngol 276:1933–1941. https://doi.org/10.1007/s00405-019-05419-3
Yin HX, Zhao PF, Zhang KP, Zhang L, Wu HX, Wang YJ, Lv H, Wang Z, Zhang P, Wang ZC (2020) An experimental study on the ability of newly developed CT equipment for temporal bone to display fine bony anatomy. Chin J Radiol 54:763–768. https://doi.org/10.3760/cma.j.cn112149-20200320-00427
Tang R, Yin H, Wang Z, Zhang Z, Zhao L, Zhang P, Li J, Zhao P, Lv H, Zhang L, Yang Z, Wang Z (2021) Stapes visualization by ultra-high resolution CT in cadaveric heads: a preliminary study. Eur J Radiol 141:109786. https://doi.org/10.1016/j.ejrad.2021.109786
Ukkola-Pons E, Ayache D, Pons Y, Ratajczak M, Nioche C, Williams M (2013) Oval window niche height: quantitative evaluation with CT before stapes surgery for otosclerosis. Am J Neuroradiol 34:1082–1085. https://doi.org/10.3174/ajnr.A3354
Chen JY, Mafee MF (2014) Computed tomography imaging technique and normal computed tomography anatomy of the temporal bone. Oper Tech Otolaryngol Head Neck Surg 25:3–12. https://doi.org/10.1016/j.otot.2013.11.002
Razafindranaly V, Truy E, Pialat J, Martinon A, Bourhis M, Boublay N, Faure F, Ltaïef-Boudrigua A (2016) Cone beam CT versus multislice CT: radiologic diagnostic agreement in the postoperative assessment of cochlear implantation. Otol Neurotol 37:1246–1254. https://doi.org/10.1097/MAO.0000000000001165
Neff BA, Lippy WH, Schuring AG, Rizer FM (2004) Stapedectomy in patients with a prolapsed facial nerve. Otolaryngol Head Neck Surg 130:597–603. https://doi.org/10.1016/j.otohns.2004.01.011
Landis J, Koch G (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174
Tange RA, De Bruijn AJG (1997) Dehiscences of the horizontal segment of the facial canal in otosclerosis. Orl 59:277–279. https://doi.org/10.1159/000276953
Singal A, Sahni D, Gupta T, Aggarwal A, Gupta AK (2020) Anatomic variability of oval window as pertaining to stapes surgery. Surg Radiol Anat 42:329–335. https://doi.org/10.1007/s00276-019-02347-z
Hernandez-Trejo AF, Cuellar-Calderon KP, Treviño-Gonzalez JL, Yamamoto-Ramos M, Campos-Coy MA, Quiroga-Garza A, Guzman-Avilan K, Elizondo-Riojas G, Elizondo-Omaña RE, Guzman-Lopez S (2020) Prevalence of facial canal dehiscence and other bone defects by computed tomography. Eur Arch Otorhinolaryngol 277:2681–2686. https://doi.org/10.1007/s00405-020-06013-8
Shin KJ, Gil YC, Lee JY, Kim JN, Song WC, Koh KS (2014) Three-dimensional study of the facial canal using microcomputed tomography for improved anatomical comprehension. Anat Rec 297:1808–1816. https://doi.org/10.1002/ar.22977
Inserra MM, Mason TP, Yoon PJ, Roberson JB (2004) Partial promontory technique in stapedotomy cases with narrow niche. Otol Neurotol 25:443–446. https://doi.org/10.1097/00129492-200407000-00007
Zdilla MJ, Skrzat J, Kozerska M, Tarasiuk J, Liberty W, Virginia W, Virginia W, Science AC (2019) Oval window size and shape: a micro-CT anatomical study with considerations for stapes surgery. Otlo Neurotol 39:558–564. https://doi.org/10.1097/MAO.0000000000001787.Oval
Kozerska M, Skrzat J, Spulber A, Walocha J, Wroński S, Tarasiuk J (2017) Micro-CT study of the dehiscences of the tympanic segment of the facial canal. Surg Radiol Anat 39:375–382. https://doi.org/10.1007/s00276-016-1744-4
Majdani O, Thews K, Bartling S, Leinung M, Dalchow C, Labadie R, Lenarz T, Heidrich G (2009) Temporal bone imaging: comparison of flat panel volume CT and multisection CT. Am J Neuroradiol 30:1419–1424. https://doi.org/10.3174/ajnr.A1560
Conte G, Scola E, Calloni S, Brambilla R, Campoleoni M, Lombardi L, Di Berardino F, Zanetti D, Gaini LM, Triulzi F, Sina C (2017) Flat panel angiography in the cross-sectional imaging of the temporal bone: assessment of image quality and radiation dose compared with a 64-section multisection CT scanner. Am J Neuroradiol 38:1998–2002. https://doi.org/10.3174/ajnr.A5302
Piergallini L, Scola E, Tuscano B, Brambilla R, Campoleoni M, Raimondi G, Lombardi L, Di Berardino F, Zanetti D, Sina C, Triulzi F, Conte G (2018) Flat-panel CT versus 128-slice CT in temporal bone imaging: assessment of image quality and radiation dose. Eur J Radiol 106:106–113. https://doi.org/10.1016/j.ejrad.2018.07.013
Yin HX, Lu WK, Li WH, Xu N, Zhao L, Sun J, Zhao PF, Zhang L, Wang ZC (2021) Experimental comparison of temporal bone imaging protocol optimization between 10 μm otology CT and multi-slice CT. Natl Med China 101:3890–3896. https://doi.org/10.3760/cma.j.cn112137-20210816-01843
Acknowledgements
We thank Medjaden Inc. for scientific editing of this manuscript.
Funding
This study was funded by [Beijing Scholar 2015 (Zhenchang Wang)], and [Capital's Funds for Health Improvement and Research (No. 2022-1-1111)].
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by [ZZ], [RT] and [QW]. The first draft of the manuscript was written by [ZZ] and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Ethical approval
All procedures performed in the study were in accordance with the ethical standards of the institutional research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. The study was approved by the medical ethics committee of Beijing Friendship Hospital (No. 2020-P2-061–02).
Informed consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, Z., Tang, R., Wu, Q. et al. An exploratory study of imaging diagnostic clues for overhanging facial nerve in ultra-high-resolution CT. Eur Arch Otorhinolaryngol 280, 3643–3651 (2023). https://doi.org/10.1007/s00405-023-07879-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-023-07879-0