Introduction

In cochlear implantation, best possible speech perception is desirable. Along with several other favourable prognostic factors in adults such as the motivation of the patient, medical history with occurrence of postlingual hearing loss and a short period of deafness, it is assumed that best speech perception can be reached by the correct electrode selection and an appropriate cochlear coverage [1,2,3,4]. Therefore, it is necessary to have sufficient data on the cochlear anatomy preoperatively, especially on the cochlear duct length (CDL). In recent years, new imaging technologies, models and formulas have been developed to measure the CDL [5,6,7,8,9,10,11,12,13,14]. Most of these studies were based on measurements of the CDL by Escudé in 2006 [5]. Most frequently, MSCT [5, 12] or cone-beam CT [8,9,10, 13, 14] have been used. In some studies, results were compared to more precise imaging, like micro-CT [9, 11, 13] or synchrotron radiation phase-contrast imaging [6, 7] as a reference method.

The flat-panel volume computed tomography (fpVCT) system has an innovative design which allows imaging of entire organs in one axial acquisition with an ultra-high spatial resolution [15]. Initial studies found superior image quality of the fine osseous temporal bone structures than in currently available MSCT scanners [16]. The first application of fpVCT in patients revealed a proper resolution regarding the critical structures for cochlear implantation [17]. Other studies first showed a significantly higher overall image quality compared to MSCT and a reduction of the effective dose of approximately 40% compared to 64-section- [18] and 128-section-MSCT [19]. Further advantages of fpVCT are portability-enabling the intraoperative use and reduced metallic artifacts, allowing postoperative position analyses, and more accurate frequency mapping of the electrode contacts [20,21,22]. As reported by Pearl et al. fpVCT offers the possibility to enhance image quality by secondary reconstructions (fpVCTSECO) [23], which can reach the same accuracy in measuring the CDL compared to experimental Micro-CT [24].

A specially designed software for otological surgical planning has been used, inter alia, to evaluate the possibility of facial nerve segmentation for otological training [25] and surgical planning of cochlear implantation in cases with post-meningitis ossification [26]. The clinical applicability of the software, in particular an angular insertion depth prediction for preoperative electrode selection has been demonstrated [27]. In various recently published studies, the CDL was also measured with this software, using MSCT and MRI [28,29,30]. They revealed low inter- as well as intraobserver variability and reliability [28, 29]. So far, evaluation of pre and postoperative clinical cochlear data created with fpVCT using the otological software has not been performed.

In the present study, different aspects should be investigated. Firstly, does the application of the fpVCT or fpVCTSECO facilitate the preoperative measurements of the CDL in comparison to MSCT in the otological planning software, secondly, is it possible herewith to have a reliable measure of the postoperative CDL with fpVCT or fpVCTSECO and thirdly, is there a difference in estimating the relative position of the electrode contacts in the comparison between fpVCT, fpVCTSECO and MSCT?

Methods

Subjects and groups

In this retrospective single-center study, 30 patients from two cohorts were included, who were divided into three groups as depicted in Table 1.

Table 1 Data of the cohorts of patients used in this study

In one cohort, 20 patients were included who were implanted with a MED-EL FLEX28 electrode (28 mm electrode with 12 single contacts each 2.1 mm apart). All cochlear implantations were performed by experienced otosurgeons (KR, RH). For all patients, this was the first implanted ear. Mean age of the patients was 64 years (standard deviation (SD): 14.9 years). Surgery was performed for 12 patients on the right and for eight on the left ear. Every patient had a preoperative MSCT of the temporal bone and a postoperative fpVCT for position control of the implanted electrode. In addition, fpVCTSECO was performed using the data of the fpVCT. Since radiological examinations of the patients include both temporal bones, different investigations with the three imaging modalities could be performed: in non-implanted ears (Group 1) and in preoperative non-implanted and postoperative implanted ears (Group 2).

Further images were evaluated in a second cohort of 10 patients, in which a MSCT and an fpVCT of an implanted ear was available. Consequently, the comparison of the three imaging modalities in implanted ears (Group 3) was possible. The mean age of this cohort was 57 years (SD: 21.2 years). Four subjects received their implant on the right side, six on the left side.

The retrospective anonymized study was conducted in concordance with local guidelines and principles of the Declaration of Helsinki and Good Clinical Practice and was approved by the local ethics committee at the University of Wuerzburg (2019020401).

Imaging

FpVCT scans were performed using an angiographic unit (Axiom Artis; Siemens Healthcare AG, Erlangen, Germany) with commercially available software (Syngo DynaCT, Siemens).

The MSCT datasets were acquired using a SOMATOM Definition AS + (Siemens) with commercially available software (Syngo CT, Siemens). The standard application (inner ear high-resolution program) was applied using the following parameters: tube current = 38 mA; tube voltage = 120 kV; collimation = 0.6 mm; pitch = 0.55; slice thickness = 600 µm.

The fpVCT datasets were acquired using the following parameters: 20 s DCT Head protocol; tube current = 21 mA; tube voltage = 109 kV; rotation angle = 200 degree; pulse length = 3.5 ms; frame angulation step = 0.5 degree/frame; slice thickness = 466 µm.

From these datasets, fpVCTSECO were designed according to the findings of Pearl et al. [23] using the following settings: 512 × 512 section matrix; HU kernel types; sharp image characteristics; slice thickness = 99 µm. Visual comparison of different imaging modalities and settings is presented in Fig. 1A–F.

Fig. 1
figure 1

“Cochlear view” of different imaging modalities and settings. Representative image of MSCT (600 µm) is shown in panel (A), display of fpVCT (466 µm) in panel (B). Image of fpVCTSECO (99 µm) is presented in panel (C). Respectively, panels (DF) demonstrate the “Cochlear View” [31] with an implanted electrode. MSCT: multislice computed tomography, fpVCT: flat-panel volume computer-tomography, fpVCTSECO: fpVCT secondary reconstruction

Measurement of cochlear parameters and calculating CDLOC

To determine pre- and postoperative cochlear parameters, OTOPLAN® software (CAScination AG (Bern, Switzerland) in cooperation with MED-EL (Innsbruck, Austria), version 2) was used. The data sets were converted to the DICOM standard by the PACS network of the hospital and transferred anonymously to the software using a memory stick. In all images initially, the coronal oblique view, typically referred to as “Cochlear View” [31, 32], was created by 3-dimensionally rotating against the axial, sagittal and coronal axis (Fig. 1). After attempting to receive the most optimized view, the cochlear diameter (A) and width (B) were measured. Both, the process of orientation and landmark selection, were carried out in two test series with an interval of two weeks by one medical examiner, who is trained as an ENT-specialist and is very experienced in the evaluation of cochlear imaging. Patient’s data were anonymized and the order of the evaluation of both the imaging modality as well as the temporal bones was randomized. The cochlear A-value is defined as a straight line from the round window, passing the modiolus, to the furthest point on the opposite wall of the cochlea. The cochlear B-value is the straight line connecting the two opposite lateral walls of the cochlea, perpendicular to the cochlear diameter passing through the modiolus. Visualization of the parameters in the different body axes is shown in Fig. 2A–C. From these data, the CDL of the organ of Corti (CDLOC) was calculated from the otological software using Eq. (1), which has been originally described as the “elliptic circular approximation” [9]. Since the OC does not start at the center of the round window but reaches into the cochlear base, the equation in the software version used herein has been supplemented by the length of this so-called hook region of 1.58 mm as follows:

$${\text{CDL}}_{{{\text{OC}}}} \; = \;\left( {{1}.{71}\; \times \;\left( {{1}.{18}\; \times \;\left( {A_{{{\text{OC}}}} } \right)\; + \;{2}.{69}\; \times \;B_{{{\text{OC}}}} } \right)\; - \;\sqrt {\left( {0.72\; \times \;A_{{{\text{OC}}}} \; \times \;B_{{{\text{OC}}}} } \right)} \; \times \;0.{18}} \right)\; + \;{1}.{58}$$
(1)
Fig. 2
figure 2

Evaluation of cochlear parameters. Visualization of the cochlear parameters and the electrode contacts in the different body axes using fpVCTSECO in the otological planning software. The “axial view” is shown in panel (A), the “coronal”, known as the “Cochlear View” [32] in panel (B) and the “sagittal view” in panel (C). Three-dimensional display of the inserted implant electrodes after identifying them in the three-body axes is shown in panel (D). Diameter A is shown as green fiducials and line, width B, respectively, as blue fiducials and line. From these data, CDLOC was calculated. C1-12: Electrode contacts. RW Round window

Determination of electrode contact position

The electrode contact position (ECP) of the electrodes was determined to perform postoperative position control. Since all patients received a device with 12 electrodes, 12 fiducials were, respectively, placed by rotating through the 3D-illustrated “Cochlear View” [31]. The fiducials were placed as precisely as possible at the center of each electrode contact of the three body axes (Fig. 2A–C). From these data the otological planning software calculated the insertion depth in mm for every single electrode. For statistical analysis, the distances between every single electrode (inter-electrode distance, IED) were extracted. The entire electrode array can be seen after these maneuvers and rotated in every direction in a three-dimensional display, as seen in Fig. 2D.

Statistics

Before parametrical analyses were conducted, the normal distribution of all data series was confirmed by Kolmogorov–Smirnov and Shapiro–Wilk test. As a requirement for paired t-test, normal distribution of the differences between both paired groups was also checked. The paired t-test was used for reference comparison between the two measurement series of the particular imaging modalities. Differences with a p-value of less than 0.05 were considered to be statistically significant. For evaluation of the differences of the absolute mean values of the CDLOC in the three modalities and settings, one-way repeated measures analysis of variance (ANOVA) was applied. When the data were not normally distributed, the Friedman-Test was used.

Intraclass correlation (ICC) was performed for evaluating intraobserver variability in each modality and setting within all test series. Based on the publication of Wirtz et al., a two-way mixed model was performed, and single measure values of ICC were used [33]. ICC was tested for absolute agreement. As described by Cicchetti et al., ICC values were interpreted by applying the following scale: unacceptable (ICC < 0.4), fair (0.4 ≤ ICC < 0.6), good (0.6 ≤ ICC < 0.75) and excellent (0.75 ≤ ICC) [34]. The associated Cronbach’s alpha (CA) was calculated according to the following scale: unacceptable (α < 0.7), fair (0.7 ≤ α < 0.8), good (0.8 ≤ α < 0.9) and excellent (0.9 ≤ α) [34]. For ICC and CA only significant results were considered.

According to Koch et al., a clinically acceptable margin of divergence of ± 1.5 mm for CDLOC was assumed [7]. With a recalculation from the same study, the clinically acceptable variance of ± 0.09 mm for IED was derived.

Statistical analyses and creating diagrams were performed by GraphPad Prism (Version 8.4.0, San Diego, California, USA), as well as IBM SPSS Statistics (Version 25.0.0.0; IBM Corporation, Armonk, New York USA). Data are presented in bar charts and Bland–Altman-Plots.

Results

Population parameters

All parameters of the measurements carried out, including mean values, ranges, SD, 95%-confidence intervals (CI), ICC, CA and t-Test for both cohorts, are depicted in Table 2.

Table 2 Analysis of non-implanted and implanted ears

Group 1: determination of CDLOC in non-implanted ears by MSCT, fpVCT and fpVCTSECO

CDLOC were measured in non-implanted ears. The mean value for the length of the CDLOC were 34.55 mm (range: 31.20–6.95 mm, SD: 1.60 mm) measured with MSCT. For fpVCT, it was 34.63 mm (range: 31.55–37.60 mm, SD: 1.47 mm) and for fpVCTSECO, 35.84 mm (range: 32.95–38.45, SD: 1.36 mm) were determined (Fig. 3A). All values measured by fpVCTSECO differed significantly from those evaluated using MSCT and fpVCT (p < 0.001). There were no significant differences in the measurements between MSCT and fpVCT.

Fig. 3
figure 3

Analysis of non-implanted ears with comparison of different radiological settings and modalities in a clinical setting (Group 1). Bar shows absolute values of measurements of the CDLOC based on MSCT, fpVCT and fpVCTSECO (A). Bland–Altman plots were constructed to visualize discrepancies within the two test series of the various imaging settings and modalities (BD). According to Koch et al. [7], a clinically acceptable deviation of ± 1.5 mm was assumed analyzing CDLOC. Differences between the cohorts are indicated as significant, ***p ≤ 0.001. MSCT: multislice computed tomography, fpVCT: flat-panel volume computed tomography, fpVCTSECO: fpVCT secondary reconstruction

When examining CDLOC, there was a significant difference within the two test series using MSCT (p = 0.0011) and fpVCT (p = 0.0297), but not applying fpVCTSECO. ICC were consistently categorized as excellent for all imaging modalities. The lower limits of the CI were unacceptable (MSCT), fair (fpVCT), and excellent (fpVCTSECO). CA was classified as excellent for MSCT and fpVCTSECO, and as good for fpVCT (Table 2).

Considering corresponding Bland–Altman-Plots for CDLOC, which demonstrate clinically acceptable values, MSCT had four errors (Fig. 3B). Measurements with fpVCT showed two deviations (Fig. 3C) and assessing fpVCTSECO, there were no divergencies (Fig. 3D).

Overall, fpVCTSECO was categorized as excellent and had no clinically unacceptable deviations when measuring cochlear parameters in cochleae without an inserted electrode.

Group 2: measurements of CDLOC by using non-implanted preoperative MSCT and postoperative fpVCT and fpVCTSECO with an inserted electrode

To evaluate whether it is possible to have a reliable postoperative measurement with fpVCT or fpVCTSECO despite the inserted electrode, CDLOC were measured pre- and postoperatively. The mean values for the length of the CDLOC were 34.29 mm (range: 30.50–37.15 mm, SD: 1.81 mm) measured with preoperative MSCT. For postoperative fpVCT, it was 36.43 mm (range: 33.55–39.30 mm, SD: 1.67 mm) and for postoperative fpVCTSECO 36.18 mm (range: 32.90–39.60, SD: 1.75 mm) (Fig. 4A). The values measured in MSCT differed significantly from those evaluated with fpVCT and fpVCTSECO (p < 0.001 for CDLOC). Between fpVCT and fpVCTSECO measurements, no significant differences were observed.

Fig. 4
figure 4

Analysis of preoperative non-implanted and postoperative implanted ears (Group 2). Bar shows absolute mean values of measurements of CDLOC based on MSCT, fpVCT and fpVCTSECO (A). Bland–Altman plots were constructed to visualize discrepancies within the two test series of the various imaging settings and modalities (BD). According to Koch et al. [7], a clinically acceptable deviation of ± 1.5 mm was assumed analyzing CDLOC. Notably, MSCT-measurements were done preoperatively without inserted electrode and fpVCT/ fpVCTSECO measurements postoperatively. Differences between the different cohorts are indicated as significant, ***p ≤ 0.001. MSCT: multislice computed tomography, fpVCT: flat-panel volume computed tomography, fpVCTSECO: fpVCT secondary reconstruction

Regarding CDLOC, there was a significant difference within the test series for MSCT (p = 0.0077), but not for fpVCT and fpVCTSECO. ICC and CA were consistently rated to be excellent. The lower limit of the CI was only excellent for fpVCTSECO and fair for MSCT and fpVCT (Table 2). Summarising, using fpVCTSECO to evaluate CDLOC postoperatively was exclusively excellent.

In the corresponding Bland–Altman-Plots, using MSCT produced two clinically unacceptable deviations for CDLOC (Fig. 4B). Measurements with fpVCT showed four aberrations for and (Fig. 4C) assessing fpVCTSECO, there were two outlier (Fig. 4D).

Group 3: comparison of CDLOC in cochleae with an inserted electrode and evaluation of the IED by using MSCT, fpVCT and fpVCTSECO

This investigation was performed in cases with an MSCT and fpVCT scan with an inserted electrode. The mean value for CDLOC was 37.59 mm (range: 34.35–42.40 mm, SD: 2.12 mm) using MSCT. Assessing fpVCT, it was 37.01 mm (range: 34.70–42.00 mm, SD: 2.06 mm). Measurements with fpVCTSECO revealed a CDLOC of 36.55 mm (range: 34.55–41.40 mm, SD: 2.07 mm) (Fig. 5A). All values demonstrated above did not differ significantly between the different imaging modalities.

Fig. 5
figure 5

Analysis of implanted ears (Group 3). Bars shows absolute values of measurements of CDLOC based on MSCT, fpVCT and fpVCTSECO (A). Bland–Altman plots were constructed to visualize discrepancies within the two test series of the various imaging settings and modalities (BD). According to Koch et al. a clinically acceptable error of ± 1.5 mm was assumed analysing CDLOC. Moreover, Bland–Altman plots of MSCT (E), fpVCT (F) and fpVCTSECO (G) were constructed to demonstrate variations within the two test series localizing the IED in postoperative analysed ears. Derived from Koch et al. a clinically acceptable error was calculated with ± 0.09 mm for IED. MSCT: multi-slice computed tomography, fpVCT: flat-panel volume computer-tomography, fpVCTSECO: fpVCT secondary reconstruction

For CDLOC, no significant differences were found within the two test series and ICC was consistently ranked as excellent for all imaging modalities. CA was stated excellent for fpVCT and fpVCTSECO but just good for MSCT. The lower limit of the CI was unacceptable for MSCT, while it was excellent for fpVCT and fpVCTSECO (Table 2).

To demonstrate a clinically arguable zone, corresponding Bland–Altman-Plots for CDLOC were created. For MSCT, 2 clinically unacceptable deviations became apparent (Fig. 5B). For fpVCT, no deviances were observed (Fig. 5C) and using fpVCTSECO no aberrations were measured (Fig. 5D).

Furthermore, for the purpose of evaluating the accuracy of the determination of electrode contact positions in the different imaging modalities, distances between every single electrode (IED) of the implant were extracted from the otological planning software. A statistical workup of the mean was not performed since different electrodes in different patients have been used. Therefore, a comparison of the three imaging modalities by ICC was carried out. No significant difference was found within the two test series. ICC was good and CA fair for MSCT. In contrast, the same parameters were excellent when using fpVCT and fpVCTSECO (Table 2). The corresponding Bland–Altman-Plots demonstrate four clinically unacceptable deviations for MSCT, two clinically unacceptable aberrations for fpVCT and one clinically unacceptable value for fpVCTSECO (Fig. 5E–G).

Discussion

In the field of cochlear implantation there is a growing interest in determining the cochlear anatomy and the intracochlear position of an implanted electrode, as it is assumed that an anatomically optimal location of the electrode can lead to a better hearing outcome after surgery [35]. For the preoperative planning and the postoperative position control, new high-resolution imaging techniques as well as newly developed otological planning softwares can be used. In order to evaluate whether fpVCT and its secondary reconstruction (fpVCTSECO) contributes to anatomically based cochlear implantation compared to the conventionally used MSCT, data of the different imaging modalities were analyzed using a newly developed otological planning software in a clinical setting.

The three different imaging modalities were compared measuring the clinically most important parameter CDLOC. The comparison of MSCT and fpVCT using raw data stacks with similar slice thicknesses revealed only moderate clinical discrepancies. In contrast, clinical relevant differences were seen, when compared to fpVCTSECO with low slice thickness, which was the only one that was categorized as excellent in all clinical settings, including postoperative scans with an implanted electrode. Moreover, even the lower limits of the 95%-CI were consistently ranked as excellent. In comparison to MSCT, where unacceptable deviations in non-implanted ears occurred (20% in Group 1 (Fig. 3B) and 10% in Group 2 (Fig. 4B)), there were no clinically unacceptable deviations measuring CDLOC-values using fpVCTSECO (Fig. 3D). Moreover, in implanted ears (Group 3), there were no divergencies using fpVCTSECO (Fig. 5D) compared to deviations in 20% of the cases using MSCT (Fig. 5B). These results indicate low intra-variability for fpVCTSECO measuring CDLOC and thus good clinical applicability. This is in accordance with Rathgeb et al., who was among the first to evaluate a good clinical observer variability of the planning software [27]. FpVCT without secondary reconstruction is considered a precise method for assessing the parameters of the inner ear with the same or even less radiation dose than conventional methods [18, 19]. This might reduce the risk of radiogenic damage in the long term [36, 37] while increasing diagnostic image quality [18, 19, 22]. The option of generating secondary reconstructions limits the need for further radiation exposure, while the image quality is even more greatly improved. Indeed, the use of fpVCTSECO in this study, revealed the above mentioned advantages, in addition to the obvious reduced metal artifacts [20, 21]. The demonstrated results are in accordance with recently published data on measurements of CDLOC using fpVCTSECO in temporal bone specimen [24] using multi-planar reconstructions [38]. This might be well explained by the higher resolution of fpVCTSECO with a slice thickness of 99 µm in comparison to the other image qualities. Summarized, the findings of this study implicate that the application of fpVCTSECO facilitates the measurement of the CDLOC in clinical imaging and makes them more reproducible in comparison to MSCT using this otological planning software.

The mean values of all measurements, depicted in Fig. 3A and Fig. 4A (regardless with or without implanted electrode) for the CDLOC using fpVCTSECO, were 36.14 mm and differed significantly between measurements with MSCT (34.55 mm) and fpVCT (35.66 mm). That is in contrast to studies on measurements of CDLOC using the corresponding software. The authors described a shorter mean length for CDLOC with 32.91 mm [39] and 32.84 mm [29]. This may be explained by the fact that only conventional MSCT was used. In addition, as the exact version of the software was not mentioned in these publications, it might be possible that older versions have been used that did not include the hook region yet in the equation as it was in Version 2 of the software in this study. Moreover, it was suggested that any effect of slice thickness on the CDL measurement is likely rather small. However, it could not be ruled out the possibility that variation in other parameters affecting CT quality could result in greater variability in CDL estimates as this present study has now confirmed [29]. A systematic error was described in recent studies when the points at the lateral wall are placed more medially. This occurs mainly in scans with lower resolution than in high-resolution scans of micro-CT or pictures of histological sections [9, 10]. This induces shorter values of cochlear parameters. This phenomenon might interestingly also explain why the values measured by MSCT with implanted electrodes, as depicted in Fig. 5A–C, are not even more significantly different and even slightly higher than those measured with fpVCTSECO. Due to the larger artifacts of the electrodes in MSCT, the points marking the lateral wall might be set more laterally than in the true anatomical set, responsible for the discrepancies in different parameters. Therefore, clinical analyses using fpVCTSECO probably represents the true length, since assessments with micro-fpVCT and micro-CT in temporal bone specimens have shown no significant different CDL at the lateral wall values up to over 40 mm [13, 24], which corresponds to a CDLOC of about 35 mm. In line with this, in other studies, shorter CDL values have been reported in clinical modalities [8, 13]. Based on these and the presented findings, it should be carefully considered whether it might be better to use longer electrodes than currently assumed for covering the correct CDLOC, when measured by MSCT or other low-resolution images. This might also result in even more accurate predictions for preoperative angular insertion depths, some of which have already been calculated with the planning software [27].

In order to verify the postoperative location of the implant, IED was computed from the data generated in the otological planning software. It was analyzed which imaging modality had the highest accuracy to detect the IED of 2.1 mm of the FLEX28-electrode. Major differences between MSCT and fpVCTSECO were observed. Regarding the reproducibility, MSCT was categorized as good, whereas fpVCTSECO as excellent. The differences become even clearer, when focusing on the lower limit of the 95%-CI interval, where MSCT was unacceptable, compared to excellent for fpVCTSECO. Moreover, the presented study revealed that MSCT showed clinically unacceptable deviations in 40% of the cases (4/10) (Fig. 5E), while using fpVCTSECO, there was only one clinically unacceptable value (10%) (Fig. 5G). To the best of our knowledge, up to date, this is the first study presenting the IED measured in the used otological planning software. However, in several studies, it was shown that fpVCT is very appropriate for postoperative examination following cochlear implantation, in particular, to determine the final position of the electrode arrays [22, 23, 40], but in none of these publications the intra-electrode variables were measured. The increased clinically unacceptable errors using MSCT in this study are presumably due to radiological artifacts. Indeed, it has already been shown that the real diameter of the electrode is half in comparison to the radiological one in CBCT [41]. Even rather, a study reported that single electrode contacts are only visible in fpVCT and not in MSCT [22]. This may be of clinical relevance, as it is assumed that a regular tonotopic stimulation of the cochlea by the electrode could lead to improved hearing performance and a better speech perception [42,43,44]. Therefore, the wide variation of the measurements using MSCT could lead to an incorrect calculation of frequency mapping. Jiam et al. demonstrated that based on fpVCTSECO imaging findings, 83% of the electrode contacts, in which standardized frequency maps were used, might be improved by reprogramming and concluded that individual pitch mapping should be performed [35]. These pitch maps would be even more accurate if the particular electrode position was located as precisely as possible.

It needs to be mentioned that only one electrode type from one implant-firm was used for measuring IED. This is due to the use of the otological planning software, which only allows the use of the manufacture’s electrodes. However, it is known that certain electrodes are more prone to generate artefacts, which are believed to depend on the thickness and material of the metal [45]. Therefore, we cannot exclude that the problems of MSCT measuring the IED may be more a problem of the electrode rather than of the modality.

One potential limitation of this study is the sample size concerning pre- and postoperative measurements. In addition, the study has the limitation to observe only intraobserver variability, but low interobserver variability in the use of this otologic planning software has been shown before [27,28,29]. We also cannot rule out a possible influence of the investigator’s training curve. However, since there was a blinded and randomized sequence not only of the patient data but also of the imaging modalities, such an effect would affect all modalities.

Conclusion

The results of the presented study suggest that the combination of fpVCTSECO and otological planning software will enable further progress in the development of an anatomically based cochlear implantation. It might be beneficial to perform the preoperative planning on cochlear implantation regarding the correct electrode selection with fpVCTSECO utilizing the otological planning software. Furthermore, postoperative control of the IED with fpVCTSECO will enhance accuracy of creating individual pitch maps through the software. This might further improve the exact determination of cochlear anatomy and refine postoperative frequency mapping for a better hearing outcome.