Introduction

Implantable cardioverter-defibrillators (ICDs) have become standard therapy to prevent sudden cardiac death, but have recently been associated with serious short- and long-term morbidity, such as lead failure, premature depletion, endocarditis and inappropriate shocks. A totally subcutaneous ICD (S-ICD) system has been developed to address some of these concerns [1]. We report the initial clinical experience of the first 31 patients who received this device in our department.

Methods

Patient population

Files of all patients receiving a totally subcutaneous cardioverter-defibrillator in our institution were reviewed for this report. Patients were selected for an S-ICD if they had an ICD indication according to the ACC/AHA/ESC 2006 guidelines for primary or secondary prevention. A total of 31 patients (24 males, 7 females) received the S-ICD system. The patient’s mean age was 53 ± 4 years. Left ventricular ejection fraction was 38 ± 15%, which was relatively high due to the number of patients with prophylactic implantations for channelopathy or genetic disease (32% had LVEF >50%). Coronary artery disease was present in 18 patients and the indication was primary prophylaxis in 21 patients. Coronary artery disease was present in 18 patients (58%), dilated cardiomyopathy in 4 (13%), Brugada disease in 2, idiopathic ventricular fibrillation in 3, idiopathic VT in 2 and non-compaction cardiomyopathy and valvular disease were diagnosed each in 1 patient. Demographic data are given in Tables 1, 2.

Table 1 Clinical characteristics of patients implanted with an S-ICD
Table 2 Patient characteristics

ECG showed an infarction in 13 patients. The mean PR interval was 165.2 ± 31.6 ms with one patient in atrial fibrillation at the time of implant. The mean QRS width was 105 ± 16.2 ms. No QT abnormalities were observed; three patients had an incomplete left bundle branch block (LBBB), three had a complete LBBB and one an intermittent RBBB.

In addition, we checked a 24 h recording to exclude evident bradycardia or high degree block at the time of implant. The mean heart rate during the 24 h Holter was 70 ± 8 beats/min. The minimal and maximal heart rates were 48 ± 5 and 111 ± 17 beats/min, respectively; the mean longest RR interval was 1,281 ± 149 ms. Two patients previously had a transvenous ICD explantation, one for hematoma and painful deep venous thrombosis, and the other for a pace-sense lead break.

Most patients were admitted on the day of the procedure and discharged on the following day. All gave informed consent and were aware of the innovative aspects, limitations and potential advantages and disadvantages of the device.

Beta-blocking agents and antiarrhythmic drugs were continued. Prior to implantation, oral anticoagulation was discontinued until the INR was normalized. Patients with frequent ventricular arrhythmias necessitating ICD therapy and those with a pacing indication were not considered for this device. The first 11 patients were included in the study by Bardy et al. [1] which only reported a follow-up of 3 months.

The device and its programming

The subcutaneous ICD system (model SQ-RX 1010; Cameron Health Inc., San Clemente, CA, USA), comprises a pulse generator and a subcutaneous lead (model Q-Trak 3010; Cameron Health Inc.) with two sensing electrodes to record the electrical activity of the heart. These electrodes represent three vector projections of electrical conduction occurring through the heart (A-to-CAN, B-to-CAN, and A-to-B) (Fig. 1). The S-ICD system operates by analyzing both the rate and morphological characteristics of the detected rhythm. The device can be programmed as single zone or dual zone. The device calculates heart rate based on the average of the last four intervals. As soon as the heart rate crosses the lowest programmed detection zone, further analysis is performed to determine whether therapy is required. Therapy consists of an 80 J shock, with potential temporary transthoracic back-up pacing for 30 s.

Fig. 1
figure 1

S-ICD vector configuration shown on the X-ray of a patient after implantation. The drawing on the right shows how the QRS and the T-wave are assessed by the device and the physician before implantation, to ensure that a correct vector will be selected

Implantation procedure

General anesthesia was used for the first two patients, and for the rest local anesthesia was given in combination with sedation, as done routinely for transvenous implantations. Antibiotic prophylaxis was administered 1 h before the procedure. The subcutaneous defibrillator was implanted without fluoroscopy using anatomical landmarks only. The pocket was created in the left axillary region at the level of the sixth rib (Fig. 2). The parasternal defibrillation lead was placed left of the sternal midline using small incisions at the xiphoid and at the sternal–manubrium junction. The lead was pulled into position by tunneling from xiphoid to the pocket, and subsequently from the xiphoid to the sternal–manubrium junction. Fixation of the lead was performed by a suture at the tip in all patients. After the first 15 patients, an additional suture sleeve was used at the xiphoid position. For the defibrillation efficacy testing (“DFT”), etomidate was given before VF was induced. Two consecutive shocks of 65 J were required as per protocol in the initial 13 patients. Polarity was reversed in case of failure. After the initial series, only one effective shock was performed. One day after the procedure, a chest X-ray was taken to control ICD position.

Fig. 2
figure 2

Lateral (at the left) and frontal (at the right) view of a patient who received an S-ICD many years after coronary artery bypass grafting (with a midsternal scar). The device and the lead are almost invisible. The lateral incision was closed with seven discontinuous stitches. The manubrial and xyphoid wounds are barely visible

Follow-up

After discharge, patients were seen in the device outpatient clinic of our institute. After implantation, patients were seen pre-discharge, 10 days and 2 months after implantation. Subsequently, patients underwent a regular ICD interrogation at 6-monthly intervals or shorter if clinically required. At each follow-up visit, arrhythmic events with stored subcutaneous electrocardiograms (S-ECGs) were retrieved from the device’s memory.

Results

Patients

A total number of 31 patients received an S-ICD between December 2008, and October 2010.

Implantation

The implantations were usually done within 100 min (average of 101 ± 33 min) including the DFT (with 3 min between each attempt and up to three attempts before external defibrillation was used). They were performed by a group of three electrophysiologists, including one in the first year of training, without using fluoroscopy. No particular surgical problems were encountered. The procedures under local anesthesia were relatively well tolerated; only one patient was put under deep sedation for excessive pain. The most time-consuming step was the making of a lateral pocket. The DFT was performed after positioning of the generator in the pocket without closing it completely. All patients could be converted from induced VF to sinus rhythm, with the initial polarity (Fig. 3a), except for two patients in whom reversed polarity was programmed. Non-sustained AF was induced by the shock in three cases, but reverted spontaneously before the end of the procedure and never triggered further activation of the device. Post-shock bradycardia occurred in one patient, which was resolved with pacing by the device (Fig. 3b). All devices were programmed with maximal shocks (80 J).

Fig. 3
figure 3

a Detection of induced VF and subsequent shock during defibrillation threshold testing. b Transthoracic post-shock pacing after conversion of induced ventricular fibrillation

Follow-up

No record of unexplained syncope or death in our group exists till now, with a median follow-up of 286 days (range 30–638) and a total number of 350 patient-months under observation (till 1 October 2010).

Surgical problems

A late procedure-related complication was observed in two of the first 15 patients implanted, with dislocation of the lead due to migration of the electrode. One of these patients suffered from two inappropriate shocks due to myopotentials. The other was detected with an X-ray during routine control. Initial implanting recommendations did not utilize an electrode suture sleeve at the xyphoid level, but this was integrated into the implantation protocol thereafter. In both cases, repositioning of the wires was done under local anesthesia, with the addition of a dedicated suture sleeve (Fig. 4) No further inappropriate therapy was noted in the patient (239 + days since implant). There was one case of pocket infection with tissue necrosis, which led to explantation of the entire system (without difficulty). This occurred in a 14-year-old boy in a cachectic state after treatment with extracorporeal membrane oxygenation for severe heart failure and an ongoing skin infection. The implantation was carried out despite the infectious risk to allow revalidation and because his life vest had become intolerable to him.

Fig. 4
figure 4

Initial position of lead (at the left), obvious lead displacement (in the middle) and after correction (at the right)

Inappropriate shocks

Five patients had inappropriate shocks. Two were clearly due to myopotentials, one during vigorous sawing and the above-mentioned lead displacement. These events occurred prior to the software update, which specifically addressed sensing during myopotentials. Since update, no inappropriate therapy has occurred (506+ days). There was a case of oversensing during heavy coughing and one case of double counting after the appearance of a complete right bundle branch block. The fifth patient had an inappropriate shock on a very fast sinus tachycardia, the sense vectors were evaluated and a new vector was selected with an updated template. No further inappropriate therapy has been noted in this patient. These events are summarized in Table 3.

Table 3 Inappropriate shocks

Arrhythmias

All ventricular arrhythmias, detected in four patients, were treated successfully (Fig. 5). This included 11 appropriate shocks for ventricular fibrillation in the patient with the infection and correct detection of four episodes of non-sustained arrhythmias in the VF zone. In one patient, six shocks were delivered for ventricular tachycardia (VT) [cycle lengths (CL) of 240, 250 and 260 ms], with correct detection of seven episodes of non-sustained VT (CL 220, 240 and 260 ms). In this patient, the S-ICD was explanted and replaced by a transvenous system to allow ATP and also pacing for a later acquired symptomatic bradycardia. There was detection of a non-sustained VT in two other patients.

Fig. 5
figure 5

Detection and termination of spontaneous fast ventricular tachycardia (coded as T) with a shock, 18 s after initiation of arrhythmia

Discussion

Implantable cardiac defibrillators are indicated in patients with a high risk for sudden death. This is confirmed after the initial scientific studies by extensive meta-analysis and reconfirmed in most international and national guidelines [25]. Nevertheless, it became recently clear that some disadvantages of ICDs can outweigh the advantages of a prophylactic implantation [6]. Therefore it seems reasonable to consider less invasive systems to offer the advantage of internal defibrillation to patients with a moderate risk.

General advantages of SQ systems

The procedural advantages of subcutaneous systems are evident [8]. The risk of vascular damage during implantation is minimized. This implies that implantation is possible in patients with congenital heart disease or after extensive thrombosis or infection of the venous access. The subcutaneous system was effective in all cases in our experience. This was true for induced and spontaneous arrhythmias. In our patient group, for two patients this ICD was selected due to a complication from an endovascular ICD (lead break and deep vein thrombosis). After a complication implicating a transvenous device, patients are reluctant to undergo a second implantation, and the stress of possible complications related to explantation makes the subcutaneous system even more attractive. Indeed, there is no risk of endocarditis with infection and cardiac or venous damage with an explantation. After implantation, patients do not have to limit arm movements for 6 weeks, but only for about a week to 10 days until the stitches are removed. After implantation, there are no particular limitations in movement of the arms, and this is indeed appreciated by younger and more active patients. In our experience, this has been of particular interest to younger patients with a primary prevention indication in hereditary diseases such as Brugada syndrome or genetic idiopathic ventricular fibrillation. A recent study involving 61 patients diagnosed with Brugada syndrome suggests that programming a single high-rate VF zone may be associated with reduced inappropriate defibrillator discharges [9].

Transvenous implantation can otherwise be associated with right-sided venous thrombosis, endocarditis, surgical complications, pocket bleeding and pneumothorax when subclavian puncture is necessary [10]. Further, it became clear in the recent years that lead related problems were very often present and it is even said that after 10 years only 80% of leads still function as anticipated [7, 11]. We did not observe any of these disadvantages in our patients. However, it must be noted that the longevity of subcutaneous leads is not yet well established. The only surgical problems we encountered were two lead migrations. This occurred in the era before a suture sleeve was applied at the xyphoid, which prevented migration in all the later implanted patients. Further, we had one hematoma in a very thin and generally ill patient in whom a life vest was given because a transvenous device was judged to be an impossible approach.

Specific advantages in comparison with conventional devices

The subcutaneous device has incorporated some new technology specifically for interpretation of the electrocardiogram and claims good arrhythmia discrimination in the absence of transvenous leads. We observed inappropriate shocks in the beginning of this series. All events were easily explained and by upgrading the template or the software, spurious shocks could be further prevented. Whether this incidence is higher or lower, in comparison with conventional devices, has to be analyzed over a longer period or in a randomized study. At this moment, we have the impression that the incidents are reasonably further avoided with optimal programming, better vector selection and the introduction of the extra suture sleeve [12].

Specific disadvantages of subcutaneous systems

A potential disadvantage is that no bradycardia pacing and anti-tachycardia pacing is possible. The absence of pacing could more or less be considered an advantage for this therapy, as it was shown that a certain amount of pacing is associated with a decrease in left ventricular function. This was initially shown for pacemaker patients and also for patients requiring ICD therapy [13, 14]. No pacing also implies no anti-tachycardia pacing. This could potentially be of lesser concern when prophylactic implantations are considered in particular where the target is indeed ventricular fibrillation. We only had one (secondary prophylaxis) patient in whom it was regretted that no ATP facilities were present. He had also developed symptomatic bradycardia.

The implantation procedure can in itself be quite painful, there are three scars instead of one, and the box is heavier and larger and in a more sensitive part of the thorax. Patients reported pain and the need for pain medication, but it was mostly limited to the first week after implantation.

Patient selection

We did not change our selection process greatly for determination of suitable candidates for this device. An important concern is the need for pacing or ATP. However, by using Holter recordings and analyzing them in a very conventional way, we are able to exclude patients with evident or symptomatic bradycardia, high degree AV block or having pauses so that pacing function was necessary in addition to the ICD. Further, patients with monomorphic tachycardia who could have been candidates for ATP were not selected. Another consideration could be that the presence of transvenous leads might be a disadvantage when an ablation may be indicated. Recently, it was proposed that a pacemaker may be associated with the S-ICD if necessary [15].

Limitations

This single center experience was not randomized and shows the short term, initial experience of our group. This implies that no real data on long-term performance of the device and its longevity were available.

Conclusions

We had the opportunity to use a new device, which was associated with mostly reversible complications and an efficacy that seemed reasonable for the reported follow-up. Inappropriate shocks occurred during this short observation time, all cases were elucidated and an efficient solution (so far) was found. It is worthwhile to perform larger trials to gain experience with this new prophylactic device. There was only one death in our group due to an aggressive lung carcinoma, and there were no episodes of unexplained syncope or hospitalizations for heart failure. This device is also a potential alternative for patients who have had an ICD-related complication, have a limited vascular access or are reluctant to undergo a new implantation.