Abstract
Introduction and objectives
This Swiss LithoClast® Trilogy lithotrite is a new lithotrite for percutaneous nephrolithotomy (PCNL). It has four modifiable settings; impact, frequency, ultrasound and suction. We aim to determine the optimal device settings for the fastest stone clearance.
Materials and methods
Kidney stone phantoms were made with Begostone in a powder to water ratio (15:3–15:6). Complete stone clearance (seconds) was calculated and impact and frequency were adjusted and repeated N = 3. Intra renal pressure (IRP) was then measured in a porcine kidney model.
Results
Stone phantoms with physical properties similar to struvite were cleared best with 100% impact and frequency of 12 Hz. Both uric acid stone phantoms and calcium phosphate stone phantoms were cleared most efficiently with an impact of 30% and a frequency of 4 Hz. The mean time to clear uric acid stone phantoms was 83 s versus 217 s for calcium phosphate stone phantoms. Similarly, for calcium oxalate stone phantoms, an impact of 30% and a frequency of 4 Hz was associated with the fastest clearance time, mean 204 s. However, the differences between 4, 8 and 12 Hz were not statistically significant. At a suction level of 60% or higher, IRP became negative.
Conclusion
These results indicate that stone phantoms of hard kidney stones are cleared more efficiently at lower impact and frequency settings. With regard to suction, a setting of ≤ 50% appears to be the optimal setting.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Percutaneous nephrolithotomy (PCNL) is the preferred treatment for renal stones larger than 20 mm in diameter [1, 2]. Many devices have been used for stone removal during PCNL and were termed lithotrites. Early lithotrites in PCNL used compressed air to accelerate a projectile inside a hand piece, which hit a probe to elicit a shock wave that moved through the probe to the stone to cause fragmentation (ballistic lithotripsy). Other single probe lithotrites used piezoceramic crystals to creating ultrasonic waves that caused the probe’s tip to vibrate and achieve fragmentation (ultrasonic lithotripsy) [3]. These fragments were then removed by a grasper.
Dual-energy single probe lithotrites were developed, combining ballistic lithotripsy and ultrasound lithotripsy. While this was a modernisation, a downside was that these devices were mounted within one another. This resulted in reduced stone clearance speeds. Another modification was the addition of suction to the probe, increasing stone clearance, but devices could only use this with a single fragmentation modality, i.e. ballistic and suction or ultrasonic and suction only [4,5,6]. The latest generation of lithotrite is called the LithoClast® Trilogy (Electro Medical Systems S.A., Nyon, Switzerland). This device has combined ultrasonic lithotripsy, ballistic lithotripsy and suction capability in single probe. Several different probe sizes are also available. Studies have suggested that the Trilogy offers faster stone clearance than other ultrasonic and combination ultrasonic devices [3, 7,8,9].
The Trilogy lithotrite has four modifiable settings; impact (0–100%), frequency (2–12 Hz), ultrasonic lithotripsy (0–100%) and suction (0–100%). The exact combination of these variables for the most efficient stone clearance of different stone types is unknown. The objective of this study was to determine the optimal settings (impact and frequency) for four of the most common kidney stone compositions. The secondary aim was to determine the effect of different suction settings on intrarenal pressure (IRP) in porcine kidney model to determine the optimal suction setting.
Methods
Artificial kidney stone production
Begostone (BEGO USA, Lincoln, USA) was mixed in ratios of 15:3, 15:4, 15:5 and 15:6 of powder to sterile water. In these ratios, 15:3 was comparable to calcium oxalate monohydrate (COM), 15:4 to calcium phosphate (CaP), 15:5 to uric acid and 15:6 to struvite [10]. Begostone and water were mixed for at least 60 s until the resulting mixture was homogeneous. The mixtures were then placed on a contact vibrator to remove air bubbles. The resulting mixture was poured into 2 cm3 silicone moulds and allowed to sit for at least 8 h. An ultrasonic flaw detector (KRAUTKRÄMER USM GO + , Waygate Technologies, Ahrensburg, Germany) was used to measure the longitudinal wave speeds for each stone type. Results were then compared to the known physical properties of kidney stones.
Determination of intrarenal pressure in conventional PCNL with Trilogy
Whole intact urinary tracts were harvested from Landrace pigs slaughtered for the food chain by a licenced veterinarian. Three well dissected porcine kidneys were used for testing to ensure comparable results and validity. A 5Fr cystometry abdominal pressure line connected to an external strain gauge was placed into the renal pelvis and sutured in place with a purse-string suture. The intrarenal pressure was then calibrated to zero representing atmospheric pressure. Pressure readings were recorded using calibrated cystometry software. The pig kidney was punctured and dilated to 26Fr, and a 26Fr Amplatz sheath was placed. A 26Fr nephroscope was then placed into the kidney. A 3L bag of saline was set at 100 cm and 60 cm above the level of the kidney with the irrigation fluid channel fully open on the scope. The Trilogy lithotripter was then placed into the kidney via the nephroscope and the effect of the suction from the device on IRP was determined.
Determination of optimal settings for stone clearance
A water bath was filled with normal saline 0.9%. A fixed inner container was filled with sponge. This sponge had a cylindrical shape carved from the centre. This held a small soft plastic container that contained 2 cm3 artificial kidney stones. This set up is shown in Fig. 1B. All of these stones were soaked in water for 2 mins prior to testing. The outflow from the filtered suction fluid was secured in the water bath, so the saline level in the system remained constant. Trilogy probes of 3.4 mm × 340 mm were used. As per manufacturers’ guidelines, suction and ultrasound were kept constant at 40% and 100%, respectively. Impact and frequency were adjusted for a combination of 9 different settings. The 9 different settings used were 30% impact and 4 Hz frequency, 30% and 8 Hz, 30% and 12 Hz, 60% and 4 Hz, 60% and 8HZ, 60% and 12 Hz, 100% and 4 Hz, 100% and 8 Hz and 100% and 12 Hz. Each setting was repeated N = 3 for the four different artificial kidney stone mixes. The primary outcome was time to total stone clearance. This outcome was measured by a single operator using a consistent technique. The suction tubing, filter, probe and handpiece were cleared of stone fragments between runs to ensure testing conditions remained constant. Time was also taken to make sure the suction tubing was clear of blockages during runs.
Statistical analysis
Statistics were calculated using SPSS. All data were normally distributed and therefore given as mean (± standard deviation). An Independent t test was used to compare mean values for normally distributed data. One-way ANOVA (Turkeys multiple comparison) was used when three or more independent variables were compared. A p value of < 0.05 was considered statistically significant.
Results
Comparison of artificial kidney stone parameters to known parameters of kidney stones
Begostone and water when mixed in a ratio of 15:3 and set into a stone phantom had comparable physical properties to that of COM kidney stones. The longitudinal wave speeds by ultrasonic flaw detector were similar 4201 ± 27 m/s for the stone phantom vs 4535 ± 58 m/s for COM stones measured by Zhong et al. [10]. Density was also similar 2056 ± 6 kg/m3 vs 2038 ± 34 kg/m3. All percentage differences between stone phantoms and actual kidney stones in terms of longitudinal wave speed and density were less than 10%, except for the density of CaP stones and longitudinal wave speed of struvite stones. CaP stone phantoms were 15.4% more dense than actual kidney stones, while struvite phantoms had a longitudinal wave speed 15.8% higher than actual struvite stones.
Intrarenal pressure in conventional PCNL with a suction lithotripter
The Trilogy Lithotripter with a 3.9 mm × 340 mm probe was inserted into a pig kidney via a 26Fr sheath and 26 nephroscope. The change in IRP was assessed at different suction settings (0–100%) at both 60 cm and 100 cm irrigation heights. Suction strengths ≥ 60% resulted in a negative IRP. IRP was significantly lower using a 60 cm irrigation height compared to 100 cm at all suction strengths. At an irrigation height of 100 cm, IRP was 22 cmH20 at 20% suction, 15.25 at 30%, 8.5 at 40% and 3.25 at 50%. At a height of 60 cm, IRP was 11.8 cmH20 at 20% suction, 8.6 at 30%, 3.6 at 40% and 0.5 at 50%.
Optimal frequency and impact for the clearance of artificial kidney stones
The fastest total stone clearance time for COM stone phantoms was with an impact of 30% and frequency of 4 Hz (mean 204 s). A clearance rate of 571.42 mm3/min was noted. The slowest stone clearance time for COM phantoms was 60% and 8 Hz (mean 269 s or 446.10 mm3/min). Reducing the impact from 100 to 60% and 30% resulted in significantly faster stone clearance rimes (p < 0.001). Reducing the Hz appeared to have a lesser effect. Data are shown in Fig. 2A. CaP phantoms were cleared fastest at the 30% impact and a frequency of 4 Hz (mean 217 s), while the slowest stone clearance was at the 30% and 12 Hz setting (mean 261 s). These findings were similar to COM stone phantoms where reducing the Hz appeared to have a lesser effect. Data are shown in Fig. 2B. Uric acid phantoms were also fragmented fastest at the 30% impact and 4 Hz frequency (mean 83 s or 1445.79mm3/min). This was similar to COM and CaP phantoms but fragmented times were faster. 60% and 12 Hz was the slowest setting for breaking up this stone type (mean 104 s). Data are shown in Fig. 2C. In contrast to the other stone phantoms, Struvite phantoms fragmented most efficiently at the 100% impact and 12 Hz. Mean clearance time of 83 s. The slowest setting for struvite stone clearance was at 30% and 4 Hz (mean 108 s or 1111.11 mm3/min). Interestingly, this setting was the fastest for the other stone types. Data are shown in Fig. 2D. The average time to total stone clearance across the nine settings and four stone types was (889.8125 mm3/min ± 227.40 SD:454.93).
Discussion
Experientially and clinically the Trilogy has performed excellently at kidney stone fragmentation and removal [11]. However, to date there has been no study to determine the optimal settings for different kidney stones types. This is the first study to our knowledge that aims to determine the best settings for the most efficient stone clearance for four of the most common stone types.
For hard kidney stones phantoms (calcium oxalate monohydrate and calcium phosphate), the fastest total stone clearance times were associated with low impact and low frequency. Uric acid is considered an intermediate stone with regard to hardness. Similar to COM and CaP, the fastest stone clearance times were seen with low impact and low frequency. In contrast to the other kidney stones, struvite, a soft stone, was cleared most efficiently with the highest impact and frequency settings.
These results are somewhat unexpected and are the inverse to the predicted results. The lowest impact and frequency settings tested for hard and intermediate stones were the fastest stone clearance times. For soft stones, the highest impact setting and highest frequency resulted in a fast clearance time. We theorised that higher impact and frequency settings were potentially associated with the reduced transfer of ultrasonic energy due to a reduced probe to stone contact time associated with these. Lower energy settings also result in reduced dispersion of stone fragments, allowing the Trilogy’s suction to function more effectively. Struvite stones broke in to larger fragments during testing due to the stone’s lower tensile failure strength [12]. This minimised the dispersion effect of high energy settings. This may explain the efficiency of high impact and frequency settings in this group.
Overall fragmentation times of Begostone in this ex-vivo study were significantly longer than those described in other benchtop studies, such as those by Carlos et al. and Bader et al. These two studies placed 1 cm3 Begostone stone phantoms of the 15:3 consistency (calcium oxalate monohydrate) in hemispherical silicone supports in a water bath for testing. Carlos et al. calculated an average total time to stone clearance of (23.79 s or 2522.07 mm3/min) while Bader et al. calculated an average of (26 s or 2307.87 mm3/min) [3, 8]. Across our nine different settings for calcium oxalate, the average time to total stone clearance was (240.77 s or 498.4 mm3/min ± 17.67). The likely explanation for this was using a different, soft plastic container, allowing for greater dispersion of stone fragments. We felt this better represented a stone sitting in a renal pelvis. We do not believe this is a difference in the efficacy of the Trilogy between studies but a difference in experimentation setup.
The average stone clearance rate across all of the experiments in this study was (889.81 mm3/min ± 227.46). This stone clearance rate is quite similar to the in-vivo mean stone clearance rate recorded in a European multicentre prospective study on behalf of the European section of UroTechnology. This study included data from 157 PCNL procedures and calculated the mean stone clearance rate as 65.55 mm2/min or 945 mm3/min based on calculated 3D volume [11]. Nottingham et al. recorded a more efficient stone clearance rate to Thakare et al. of 68.9 mm2/min, while another prospective study by Sabnis et al. recorded an average stone clearance rate of 590.7 ± 250 mm3/min [13, 14]. A multi-institutional prospective randomised controlled trial by Large et al. calculated stone clearance rates across 51 PCNL procedures using the Trilogy. Their mean clearance rate was (1220 mm3/min ± 1670). [9]
The suction effect on IRP during PCNL with The Swiss Lithoclast® Trilogy suction is also an unknown variable. Suction can vary from 0 to 100% on the consoles touch screen. Our data show that negative pressure is generated at suction strengths above 60%, and the renal pelvis collapses. This causes loss of vision intra-operatively and can also lead to bleeding within the renal pelvis. The author's opinion is that suction strengths above 60% are unlikely to be required for PCNLs while using the larger probe sizes (3.4 mm/3.9 mm). A suction strength of 30–40% is likely optimal to maintain an IRP so that vision is not impaired and fragments are removed. The effect of suction on IRP is intuitive and supported by a 2017 study by Abourbih et al., which found that increasing the suction on a nephroscope was more effective than making two tracts in a porcine model [15]. Another interesting finding from the data shown is that increasing suction allows an increase in the height level of irrigating fluids while maintaining a safe IRP; this thus allows the operator to speed up irrigation safely. A similar study on IRP in a porcine model found the optimal suction for PCNL with the Trilogy to be 10% to maintain a constant pressure within the kidney [16].
This study does have some limitations. These experiments were conducted in controlled conditions using a single-operator and consistent technique. However, human and device errors cannot be disregarded. Human error in this experiment consisted of subtle changes in the fragmentation of artificial kidney stones, with some stones fragmenting in a more physically favourable manner. Steps taken to reduce this error included ensuring the Trilogy’s device tip was in constant contact with the stones and ensuring the device and tubing were clear of fragments during and between runs. Another limitation was that a single probe size was used. However, we feel this size is representative of all large Swiss Lithoclast® Trilogy probes.
These findings have changed practice in our institution and intra-operative stone clearance times have improved subjectively. The authors are currently in the process of recruiting patients for a trial to assess the efficiency of these settings in clinical practice.
Conclusion
This experiment shows that with harder stones, lower frequency and impact settings may result in more efficient fragmentation. The 30% impact function was proven to fragment calcium oxalate monohydrate stones most efficiently. After investigating the effect of the Trilogy's suction on IRP, the investigators conclude that suction of ≤ 50% is optimal.
References
Türk C, Petřík A, Sarica K, Seitz C, Skolarikos A, Straub M, Knoll T (2016) EAU guidelines on interventional treatment for urolithiasis. Eur Urol 69(3):475–482. https://doi.org/10.1016/j.eururo.2015.07.041
Fernström I, Johansson B (1976) Percutaneous pyelolithotomy. A new extraction technique. Scand J Urol Nephrol 10(3):257–259. https://doi.org/10.1080/21681805.1976.11882084
Carlos EC, Wollin DA, Winship BB, Jiang R, Radvak D, Chew BH, Gustafson MR, Simmons WN, Zhong P, Preminger GM, Lipkin ME (2018) In vitro comparison of a novel single probe dual-energy lithotripter to current devices. J Endourol 32(6):534–540. https://doi.org/10.1089/end.2018.0143
Rai RS, Patrulu K, Rai R, Gupta E, Kayastha A, Sawhney S (2008) Lithoclast(®) master in intracorporeal lithotripsy during percutaneous nephrolithotomy : our experience. Med J Armed Forces India 64(3):232–233. https://doi.org/10.1016/s0377-1237(08)80100-1
Cho CO, Yu JH, Sung LH, Chung JY, Noh CH (2010) Comparison of percutaneous nephrolithotomy using pneumatic lithotripsy (lithoclast®) alone or in combination with ultrasonic lithotripsy. Korean J Urol 51(11):783–787. https://doi.org/10.4111/kju.2010.51.11.783
York NE, Borofsky MS, Chew BH, Dauw CA, Paterson RF, Denstedt JD, Razvi H, Nadler RB, Humphreys MR, Preminger GM, Nakada SY, Krambeck AE, Miller NL, Terry C, Rawlings LD, Lingeman JE (2017) Randomized controlled trial comparing three different modalities of lithotrites for intracorporeal lithotripsy in percutaneous nephrolithotomy. J Endourol 31(11):1145–1151. https://doi.org/10.1089/end.2017.0436
Krambeck AE, Miller NL, Humphreys MR, Nakada SY, Denstedt JD, Razvi H, Preminger GM, Nadler RB, Matlaga BR, Paterson RF, Chew BH, Munch LC, Handa SE, Lingeman JE (2011) Randomized controlled, multicentre clinical trial comparing a dual-probe ultrasonic lithotrite with a single-probe lithotrite for percutaneous nephrolithotomy. BJU Int 107(5):824–828. https://doi.org/10.1111/j.1464-410X.2010.09567.x
Bader MJ, Eisel M, Strittmatter F, Nagele U, Stief CG, Pongratz T, Sroka R (2021) Comparison of stone elimination capacity and drilling speed of endoscopic clearance lithotripsy devices. World J Urol 39(2):563–569. https://doi.org/10.1007/s00345-020-03184-1
Large T, Nottingham CU, Brinkman JE, Agarwal DK, Ferrero A, Sourial MW, Stern KL, Rivera ME, Knudsen BE, Humphreys MR, Krambeck A (2021) Multi-institutional prospective randomized control trial of novel intracorporeal lithotripters: ShockPulse-SE vs TrilogyTM trial. J Endourol. https://doi.org/10.1089/end.2020.1097
Zhong P, Chuong C, Preminger G (1993) Characterization of fracture toughness of renal calculi using a microindentation technique. J Mater Sci Lett 12:1460–1462
Thakare N, Tanase F, Saeb-Parsy K, Atassi N, Endriss R, Kamphuis G, Pérez-Fentes D, Hasan M, Brehmer M, Osther P, Jung H, Turney B, Finch W, Burgess N, Irving S, Dragos L, Liatsikos E, Knoll T, Cauni V, Wiseman O (2021) Efficacy and safety of the EMS Swiss LithoClast® Trilogy for PCNL: results of the European multicentre prospective study on behalf of European Section of UroTechnology. World J Urol 39(11):4247–4253. https://doi.org/10.1007/s00345-021-03710-9
Esch E, Simmons WN, Sankin G, Cocks HF, Preminger GM, Zhong P (2010) A simple method for fabricating artificial kidney stones of different physical properties. Urol Res 38(4):315–319. https://doi.org/10.1007/s00240-010-0298-x
Nottingham CU, Large T, Cobb K, Sur RL, Canvasser NE, Stoughton CL, Krambeck AE (2020) Initial clinical experience with swiss LithoClast trilogy during percutaneous nephrolithotomy. J Endourol 34(2):151–155. https://doi.org/10.1089/end.2019.0561
Sabnis RB, Balaji SS, Sonawane PL, Sharma R, Vijayakumar M, Singh AG, Ganpule AP, Desai MR (2020) EMS Lithoclast Trilogy™: an effective single-probe dual-energy lithotripter for mini and standard PCNL. World J Urol 38(4):1043–1050. https://doi.org/10.1007/s00345-019-02843-2
Abourbih S, Alsyouf M, Yeo A, Martin J, Vassantachart JM, Lee M, Maldonado J, Lightfoot M, Li R, Olgin G, Baldwin DD (2017) Renal pelvic pressure in percutaneous nephrolithotomy: the effect of multiple tracts. J Endourol 31(10):1079–1083. https://doi.org/10.1089/end.2017.0298
Strittmatter F, Nagele U, Schachtner J, Bader MJ (2021) Evaluation of intrarenal pressure stability at different lithotripter suction settings in a porcine kidney model. World J Urol. https://doi.org/10.1007/s00345-021-03679-5
Acknowledgements
The authors would like thank Boston Scientific and Electro Medical Systemts (EMS) for the use of the SwissLithoclast Trilogy device as well as probes for testing.
Funding
Open Access funding provided by the IReL Consortium. No funding was received. However, EMS provided probes to be used in the experiments.
Author information
Authors and Affiliations
Contributions
CJOC: protocol/project development, data collection or management, data analysis, manuscript writing/editing; DH: data analysis, manuscript writing/editing; LCY: protocol/project development, data collection or management, manuscript writing/editing; LL: data collection or management, Manuscript writing/editing; DBH: protocol/project development, data collection or management, data analysis, manuscript writing/editing.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflicts of interests.
Ethical approval
This study was approved by the School of Medicine Sub-Committee of the Social Research Ethics Committee (SREC) on the 1st of April 2021.
Research involving human participants and/or animals
All procedures performed in research involving animals were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
There was no requirement for informed consent in this research.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
O’Connor, C.J., Hogan, D., Yap, L.C. et al. An ex-vivo assessment of a new single probe triple modality (Trilogy) lithotripter. World J Urol 40, 2561–2566 (2022). https://doi.org/10.1007/s00345-022-04127-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00345-022-04127-8