Abstract
Due to the risk of mobilizing plaque fragments, transfemoral TEVAR is a potentially dangerous procedure in patients with a coral reef aorta. We describe a practical method for transfemoral TEVAR in a patient with a degenerative thoracic aneurysm and a coral reef aorta. After placing a filter catheter in the abdominal aorta via a contralateral percutaneous femoral access, a working channel through the distal thoracic aorta was created with a balloon-expandable stent graft in the coral reef segment. Thereafter, transfemoral TEVAR could be performed successfully, without any complications. The additional use of a percutaneously placed filter catheter potentially allows reduction of peripheral embolism and hence may prevent patients from more invasive treatment.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
A so-called coral reef aorta typically contains extensive calcified plaques that protrude far into the vessel lumen [1,2,3]. Hence, endovascular manipulations in such a vessel bear a certain risk of releasing emboli. We describe the use of an intravascular filter catheter for transfemoral thoracic endovascular aortic repair (TEVAR) of a degenerative aneurysm in a patient with a coral reef aorta.
Case Report
A 60-year-old woman presented to a tertiary medical center with dry cough and upper back pain for one week. Long-term arterial hypertension was known, but recently, difficult to control with her actual twofold antihypertensive medication. At admission, arterial blood pressure was 185/95 mmHg. She had no past medical history of nicotine consumption, diabetes or other additional vascular risk factors.
CT angiography (Fig. 1) revealed a 5-cm-wide, saccular aneurysm of the proximal descending thoracic aorta. The distal thoracic and the suprarenal abdominal aorta were affected by extensive wall calcifications that extended far into the lumen, typical for a coral reef aorta. At the distal thoracic segment, the protruding plaques formed a near-occlusive plug. The celiac trunk and superior mesenteric artery revealed no significant luminal stenoses. The renal ostia were not affected by the calcifications and, also free of lumen alterations.
Due to the possible causal connection with the patient´s upper back pain, early treatment of the aneurysm was scheduled. To act as minimally invasive as possible, transfemoral TEVAR was attempted. For the procedure, a filter catheter should be placed in the abdominal aorta to protect from eventual embolism. Then, a balloon-expandable stent graft should be deployed at the near-occlusive coral reef aortic segment to create a working channel for the introduction of an aortic endoprosthesis.
Perioperatively, the patient received intravenous antibiotic therapy with 3 × 1 g meropenem. Intravenous heparin was also given by the anesthesiologist. In general anesthesia, via an open surgical right femoral access a hydrophilic guidewire and a pigtail angiography catheter were cautiously advanced into the ascending aorta. After digital subtraction angiography (DSA, Fig. 2A), an 11-French sheath was placed just below the coral reef segment. Then a 10-French sheath of the filter catheter system was introduced via a percutaneous left groin access. The sheath was carefully advanced over a stiff guide wire until its tip was just below the obstructive calcifications. Then, the filter catheter (Capturex®, Straub Medical AG, Wangs, Switzerland) was introduced and deployed immediately distal to the obstructive coral reef section. In this way, the filter covered all visceral branches of the abdominal aorta. Further, a balloon-expandable 16/59 mm BeGraft® aortic stent graft (Bentley InnoMed GmbH, Hechingen, Germany) was introduced via the ipsilateral 11-French sheath and deployed at the obstructive coral reef aortic segment (Fig. 2B) [4]. Thus, having created a safe working channel, a 32/164 mm Relay® PRO thoracic stent graft (Bolton Medical, Inc., Sunrise, Fl, U.S.A.) was introduced via the right common femoral artery and deployed over the aneurysm. After a control angiogram (Fig. 2C), the filter catheter could be collapsed and removed percutaneously without significant resistance. Upon inspection, some solid fragments were found inside the filter that were histologically classified as up to 12 × 5 mm calcified plaques (Fig. 3). The right femoral arterial access was closed surgically. At the left puncture site, hemostasis was achieved by a percutaneous suture device (ProstarTMXL Percutaneous Surgical Vascular System, Abbott Vascular Inc. Santa Clara, CA, USA). After completion of the procedure, the patient was transferred to the intensive care unit for further surveillance. She did not develop any symptoms of central, visceral or peripheral embolism. Heparin was continued for 48 h; daily oral clopidogrel, 75 mg, for 3 months and aspirin 100 mg, indefinitely, were also prescribed. Due to ongoing hypertensive episodes, the patient’s medication had to be adjusted. A control CT angiography six days after intervention showed complete thrombosis of the aneurysm sac, an open stent graft in the coral reef segment as well as free perfusion of the visceral arteries (Fig. 4). Finally, she was discharged without complaints.
Discussion
Transfemoral TEVAR is usually the treatment of choice for patients with aneurysms in the descending thoracic aorta. However, this current case shows that the coral reef calcification posed a certain obstacle for an aortic endoprosthesis. Since any unprotected manipulation in a coral reef aorta may be vicious by mobilizing plaque fragments, an endovascular safety filter was placed first. This Capturex® filter catheter consists of a self-expanding nitinol filter basket on a wire-reinforced catheter shaft. It is inserted via a 10-French sheath and can be deployed or re-captured by moving the sheath. As a temporary filter, it was primarily designated to catch thrombi in the vena cava, but it can also be used in the aorta [5]. Secondly, a safe access route was generated for the aortic endoprosthesis, by placing of a balloon-expandable stent graft over the protruding calcified plaques. When we passed the introducer system of the aortic endoprosthesis along the deployed filter, the filter basket was likely to be pushed aside from the aortic wall. However, the patient’s lack of embolic symptoms indicates that this loss of filter alignment was not of relevance.
Two similar cases with filter protection during TEVAR have been published. In the first study, a large Wallstent was introduced via a surgical femoral access, kept partially opened during the procedure and re-captured and withdrawn thereafter [6]. In the other case, a homemade intra-aortic filter device was created with a braided nitinol basket where a polyester fabric was embedded. Through a 12-F sheath, it was also inserted and removed via an open surgical femoral access [7]. Both procedures were performed successfully. The technique with the partially opened Wallstent seems to be quite practical. However, when deployed too extensively, a minimum risk of losing the Wallstent may exist. The second technique requires some more handwork and thus appears more inconvenient. In contrast to these techniques, the filter catheter that we used is commercially available and can be applied percutaneously.
Conclusion
In the presented case, the use of a filter catheter allowed endovascular manipulations for TEVAR in a thoracoabdominal coral reef aorta without complications. Hence, it features a valuable tool to prevent patients from more invasive alternative access methods.
References
Qvarfordt PG, Reilly LM, Sedwitz MM, Ehrenfeld WK, Stoney RJ. “Coral reef” atherosclerosis of the suprarenal aorta: unique cinical entity. J Vasc Surg. 1984;1:903–9.
Gatta E, Berretta P, Pagliariccio SS, Carbonari L. Emergency endovascular treatment of coral reef aorta with an unconventional technique. Ann Vasc Surg. 2020;63:456.e5-456.e9.
Schulte K-M, Reiher L, Grabitz K, Sandmann W. Coral reef aorta: a long-term study of 21 patients. Ann Vasc Surg. 2000;14:626–33.
Isoda R, Morita I, Mano M, Ishida A. Balloon-expandable stent graft for coral reef aorta. J Vasc Surg Cases Innov Tech. 2020;6(3):422–4.
Haine A, Rosenov A, Mertineit N, Baumgartner I, Schindewolf M. Catheter directed thrombectomy under temporary catheter based filter protection in renal vein thrombosis and contraindication to anticoagulation. EJVES Vasc Forum. 2020;49:16–9.
George RK, Singh Y, Ram BL. An endovascular filter to prevent embolization during aortic stenting. J Vasc Surg. 2016;63:548–51.
Sanada J, Matsui O, Ohtake H, Kimura K, Kusanagi M, Terayama N. Distal embolus protection with an intra-aortic filter during stent-graft repair of a severely atherosclerotic thoracic aortic aneurysm. J Endovasc Ther. 2005;12:642–6.
Funding
Open access funding provided by Medical University of Graz. This study was not supported by any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Consent for Publication
Consent for publication was obtained for every individual person’s data included in the study.
Ethical Approval
For this type of study, formal consent is not required.
Informed Consent
For the described medical procedures, informed consent was obtained from the patient.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Portugaller, R.H., Steiner, J., Schmid, F. et al. The Use of a Commercially Available Endovascular Filter Catheter (Capturex®) for Thoracic Endovascular Aortic Repair (TEVAR) in a Patient with a Coral Reef Aorta. Cardiovasc Intervent Radiol 45, 879–883 (2022). https://doi.org/10.1007/s00270-022-03147-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-022-03147-6