Abstract
Purpose
Failure to restore the femoral offset of the native hip is a potential cause of dysfunctional hip arthroplasty. The aim of this study was to report our experience of using a modular head–neck adapter in revision THA, specifically analyzing its usefulness as a tool to correct a slightly diminished femoral offset.
Materials and methods
This was a retrospective single-center study including all hip revisions performed at our institution from January 2017 to March 2022 where the BioBallTM head–neck metal adapter was used. The preoperative and one year follow-up modified Merle d’Aubigné hip score was used to evaluate functional outcomes.
Results
Of a total of 34 cases included for revision, the head–neck adapter system was used specifically in six patients (17.6%) to increase femoral offset, retaining both the acetabular and femoral components. In this subgroup of patients, mean offset decrease after primary THA was 6.6 mm (4.0–9.1), equivalent to a mean 16.3% femoral offset reduction. The median modified Merle d’Aubigné score went from 13.3 preoperatively to 16.2 at one year follow-up.
Conclusion
The use of a head–neck adapter is a safe and reliable procedure that may allow the surgeon to easily correct a slightly diminished femoral offset in a dysfunctional THA without the need to revise well-fixed prosthetic components.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Total hip arthroplasty (THA) is one of the most successful surgeries, with well-reported survival rates. According to national registry reports, a hip replacement is expected to last 25 years in around 58% of patients [1, 2]. However, there are several factors leading to a painful prosthesis that may eventually require revision. The most prevalent causes of revision include infection and inflammatory reactions (20.1%), followed by instability (18.3%), aseptic loosening (15.9%), and mechanical complications (14.9%) [3].
One of the reasons for a dysfunctional hip arthroplasty is a failed restoration of the native femoral offset, which is essential to restore patients’ predisease hip biomechanics and abductor function. In fact, suboptimal restoration of femoral offset results in increased joint reaction forces caused by a reduction in abductor moment arm, which may eventually lead to limping and potential instability [4,5,6,7]. Furthermore, decreased femoral offset has been related to poorer functional outcomes [8, 9].
Retaining a well-fixed femoral stem significantly decreases not only the morbidity of the procedure but also blood loss, infection rates, and operating time [9,10,11]. A head–neck adapter that permits an increase in femoral offset is a useful tool that typically allows the surgeon to easily correct a slightly diminished femoral offset in a dysfunctional THA without the need to revise well-integrated prosthetic components [12].
According to a recent systematic review by Novoa et al [13], the main indication for the use of a head–neck adapter system is isolated acetabular revision (71.6%) followed by prosthetic dislocation (22.2%). The specific use of these adapters with the goal of restoring suboptimal hip offset has not been previously reported. Our aim was to report our experience using a modular head–neck adapter in revision THA and to analyze its usefulness as a tool to specifically increase femoral offset while retaining both the cup and the stem.
Materials and methods
Study design
This was an observational retrospective single-center study including all hip revisions performed at our institution from January 2017 to March 2022 that required the use of a modular head–neck adapter system. All the procedures were performed by at least one of the hospital’s joint reconstruction surgeons. Digital templating was carried out in all cases. All preoperative digital radiographs of the pelvis were taken in an anterior–posterior orientation. Since external rotation of the femur usually leads to underestimation of femoral offset [14], patients were positioned supine with both hip joints rotated inward by approximately 10–15°. A dual calibration marker ball system (KingMarkTM) was routinely used as a reference for determining the individual magnification factor [15]. Digital templating was performed as described by Bono et al. [16, 17], using the TraumaCadTM software (BrainLab, Chicago, IL, USA). A femoral offset was defined as the distance from the center of rotation of the femoral head to a line bisecting the long axis of the femur [18]. All revisions were performed through the same approach used in the primary surgery. Data regarding patient demographics, body mass index (BMI), comorbidities, and indication for revision were recorded. The preoperative and 1-year follow-up modified Merle d’Aubigné [19] hip score was used to evaluate the functional outcome in those cases where the revision was performed solely using the head–neck adapter to specifically increase femoral offset. The modified Merle d’Aubigné score evaluates pain, mobility, and the ability to walk, thus providing a reliable overall assessment of the hip function [20].
The head–neck adapter
The BioBallTM (Merete Medical, Berlin, Germany) head–neck metal adapter was used in all cases. The BioBall™ is a titanium implant (TiAl6VA4) that comes in different lengths, is versatile enough to adapt to diverse Morse tapers, and is able to add 7.5° offset in any direction (Fig. 1). The adapter comes with modular heads, which are available in different materials and sizes.
Statistical analysis
Continuous variables were expressed as means or medians and standard deviation (SD) or interquartile range (IQR), whereas qualitative variables were described by absolute frequencies and percentages. Analyses were made and plots were drawn using the SPSS® v. 20.0 statistical package (SPSS, Inc. Chicago, IL, USA).
Declarations
The study was reviewed by a Research Ethics Committee, which approved it under code HCB/2023/0067, and was conducted in accordance with the 1964 Helsinki Declaration. No funding was sought or secured in relation to this manuscript. At the time of publication, none of the authors disclosed any potential conflicts of interest. None of the authors received any financial support for the research, authorship, and/or publication of this article. The retrospective nature of this study made it unnecessary to obtain the participants’ informed consent. Data was codified, and patients’ anonymity was preserved at all times.
Results
A total of 34 cases were included for revision. Median patient age was 71.2 years; 61.8% (21 out of 34) of the sample were male. The most frequent indication for surgery was revision of the acetabular cup (n = 26, 76.5%) followed by head replacement (while retaining both cup and stem) in order to achieve a higher offset (n = 6, 17.6%). There were two cases (5.9%) where the head–neck adapter was used during primary THA implantation. The adapter was not used in any case where all the prosthetic components had to be revised. An anterolateral Hardinge approach was used in 76.6% of cases (n = 26). As regards the implants’ characteristics, the most used neck length was 2XL (+ 10.5 mm) (30.3% of cases (n = 10)), followed by L (+ 3 mm) and 3XL (+ 14 mm) (27.3% (n = 9) and 18.2% (n = 6)), respectively. Regarding the taper, a 12/14 taper was used in 88.2% of cases (n = 30). The most frequently used head sizes were 36 mm (47.1% of patients (n = 16)) and 28 mm (44.1% of patients (n = 15)). The 7.5° offset adapter was used in 76.5% (n = 26) of patients. The head material was ceramics in 50% (n = 17) of cases, and dual-mobility heads were used in 41.2% of patients (n = 14). The characteristics of the patient sample are shown in Table 1.
As mentioned earlier, the head–neck adapter was used in six patients (17.6%) specifically to increase femoral offset. Of these six patients, five presented with hip pain that improved after prosthetic revision. A total of five patients presented with walking difficulties, which resolved completely in three cases. There was one patient that reported no improvement following surgery and one patient with multiple sclerosis that progressively worsened in terms of function even though a certain degree of pain relief was achieved. All six patients were operated through an anterolateral Hardinge approach. A partial gluteus medius tear was found in all cases, which was repaired with direct transosseous suture. The median offset decrease after primary THA had been 6.6 mm (4.0–9.1), causing a median femoral offset reduction of 16.3% (Figs. 2 and 3). The median femoral Merle d’Aubigné score rose from 13.3 preoperatively to 16.2 at 1-year follow-up. The detailed characteristics of these six patients are shown in Table 2.
Discussion
An accurate restoration of native femoral offset is often regarded as an essential requirement to obtain an optimal prosthetic joint. However, a certain amount of controversy still persists among surgeons. It is well known that patients with decreased offset present with poorer joint function at one-year follow-up [9]. Even though a 15% or > 5-mm decrease in offset has been related to hip abductor weakness [21, 22], its association with lower functional scores or the use of a walking aid [9, 23] remains unclear. In our series, the six cases specifically reoperated due to decreased offset experienced a mean femoral offset reduction of 6.6 mm (mean decrease of 16.3%) and they all presented with abductor weakness. Bullen et al. [24] found that an offset decrease greater than 20 mm typically resulted in worse pain and lower motion scores. Five of the six cases in the subgroup specifically operated due to decreased offset obtained significantly improved Merlé d’Aubigné scores following revision with a head–neck adapter.
The head–neck adapter has become a widely used tool, mainly in partial hip revisions where the cup is replaced and the adapter is used to restore the taper surface on a previously damaged cone. During the procedure, the existing taper is replaced by the adapter’s own taper, hence allowing retention of the primary stem. Moreover, the stability resulting from addition of a 7.5° offset may be used to address recurrent dislocation [25,26,27]. For this reason, as reported by Kock et al. [28], the adapter may be also used in primary THA in cases with a high risk of dislocation. In our series, the head–neck adapter was used in primary THA in two cases: in one case (case 10), the adapter was used to solely increase femoral neck length, whereas in the other case (case 12), the adapter added a 7.5° offset, which provided a more accurate centre of rotation.
It has been reported that patients operated due to hip dysplasia are at a high risk of experiencing an excessively increased offset after THA. Indeed, dysplasia is typically accompanied by decreased femoral offset, which is usually excessively augmented when regular necks are used [9, 29]. These patients constitute a significant challenge in terms of offset restoration, particularly in the presence of a short varus and an anteverted neck. Two of the six patients analyzed presented with a paediatric hip abnormality and in five of them a standard offset neck was used in the primary THA, which was eventually replaced by an offset head-neck adapter. In Cassidy et al. [9], 22.6% of the patients in the decreased offset group still exhibited a decreased offset postoperatively despite the use of an extended offset stem. This possibility should be carefully considered, as one of the main objectives of any reconstruction procedure is to normalize femoral offset irrespective of the initial diagnosis.
To the best of our knowledge, this is the first study to report on the use of the BioballTM head–neck adapter to specifically correct a slightly diminished and hence suboptimal femoral offset. However, the present study is subject to some inherent limitations. Most importantly, because of its retrospective nature, certain biases may have influenced the results. The study’s retrospective design may have resulted in an inadequate control over confounding variables and limited the amount of information obtained. Secondly, it is impossible to ascertain whether the outcome of the procedure can be strictly attributed to the revision of the investigated component or whether it was also related to the way the periarticular soft tissues were managed. Given that all six cases presented with gluteus medius tears of various sizes, the authors acknowledge that the benefit might have been at least in part related to the direct tendon repair performed. Finally, the limited number of cases inherent in a single-center study on a low-prevalence condition requires the performance of further research to corroborate the findings obtained. All these limitations may affect the extent to which the results presented here can be generalized beyond the specific cases studied.
In conclusion, a head–neck adapter is a safe and reliable tool that may allow the surgeon to easily correct a slightly diminished offset in a dysfunctional THA without the need to revise well-integrated prosthetic components. The use of the component might be considered whenever a revision of a decreased-offset arthroplasty is performed.
References
Chang RW, Pellisier JM, Hazen GB (1996) A costeffectiveness analysis of total hip arthroplasty for osteoarthritis of the hip. JAMA 275:858–865
Evans JT, Evans JP, Walker RW, Blom AW, Whitehouse MR, Sayers A (2019) How long does a hip replacement last? A systematic review and meta-analysis of case series and national registry reports with more than 15 years of follow-up. Lancet 393(10172):647–654
Kelmer G, Stone A, Turcotte J, King P (2021) Reasons for revision: primary total hip arthroplasty mechanisms of failure. J Am Acad Orthop Surg 29(2):78–87
Asayama I, Naito M, Fujisawa M, Kambe T (2002) Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty 17:747e51
Zahar A, Rastogi A, Kendoff D (2013) Dislocation after total hip arthroplasty. Curr Rev Musculoskelet Med 6:350e6
Moharrami A, Mirghaderi SP, Hoseini-Zare N, Kaseb MH, Moazen-Jamshidi SMM, Mansour AK, Mortazavi SMJ (2022) Restoring femoral medial offset could reduce pelvic obliquity following primary total hip arthroplasty, an observational study. Int Orthop Dec 46(12):2765–2774
de Thomasson E, Conso C, Mazel C (2021) A well-fixed femoral stem facing a failed acetabular component: to exchange or not? A 5- to 15-year follow-up study. Orthop Traumatol Surg Res 98:24–29
Vaishy A, Arif M, Aanand R, Singh K, Rulaniya HK, Kumawat N, Choudhary R, Choudhary P (2022) Implant positioning in total hip arthroplasty: influence of horizontal and vertical offset on functionality of prosthetic hip joint. Int Orthop 46(7):1501–1506
Cassidy KA, Noticewala MS, Macaulay W, Lee JH, Geller JA (2012) Effect of femoral offset on pain and function after total hip arthroplasty. J Arthroplasty 27:1863e9
Kershaw CJ, Atkins RM, Dodd CA, Bulstrode CJ (1991) Revision total hip arthroplasty for aseptic failure. A review of 276 cases. J Bone Joint Surg Br 73:564–568 18
Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T (2011) Should the well-fixed, uncemented femoral components be revised during isolated acetabular revision? Arch Orthop Trauma Surg 131:481–485
Dabis J, Hutt JR, Ward D, Field R, Mitchell PA, Sandiford NA (2020) Clinical outcomes and dislocation rates after hip reconstruction using the Bioball system. Hip Int 30(5):609–616
Novoa CD, Citak M, Zahar A, López RE, Gehrke T, Rodrigo JL (2018) The Merete BioBall system in hip revision surgery: a systematic review. Orthop Traumatol Surg Res 104(8):1171–1178
Blümel S, Stadelmann VA, Brioschi M, Küffer A, Leunig M, Rüdiger HA (2021) The trochanteric double contour is a valuable landmark for assessing femoral offset underestimation on standard radiographs: a retrospective study. BMC Musculoskelet Disord 22(1):310
King RJ et al (2009) A novel method of accurately calculating the radiological magnification of the hip. J Bone Joint Surg Br 91(9):1217–1222
Bono JV (2004) Digital templating in total hip arthroplasty. J Bone Joint Surg Am 86-A(Suppl 2):118–122
Gallart X, Daccach JJ, Fernández-Valencia JÁ, García S, Bori G, Rios J, Riba J (2012) Study of the consistency of a system for preoperative planning digital in total arthroplasty of the hip. Rev Esp Cir Ortop Traumatol 56(6):471–477
Lecerf G, Fessy MH, Philippot R, Massin P, Giraud F, Flecher X, Girard J, Mertl P, Marchetti E, Stindel E (2009) Femoral offset: anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty. Orthop Traumatol Surg Res 95(3):210–219
Charnley J (1972) The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg Br 54(1):61–76
Ugino FK, Righetti CM, Alves DP, Guimarães RP, Honda EK, Ono NK (2012) Evaluation of the reliability of the modified Merle d’Aubigné and Postel method. Acta Ortop Bras 20(4):213–217
Asayama I, Chamnongkich S, Simpson KJ, Kinsey TL, Mahoney OM (2005) Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 20:414–420
Sato H, Maezawa K, Gomi M, Kajihara H, Hayashi A, Maruyama Y et al (2020) Effect of femoral offset and limb length discrepancy on hip joint muscle strength and gait trajectory after total hip arthroplasty. Gait Posture 77:276–282
Mahmood SS, Mukka SS, Crnalic S, Wretenberg P, Sayed-Noor AS (2016) Association between changes in global femoral offset after total hip arthroplasty and function, quality of life, and abductor muscle strength. A prospective cohort study of 222 patients. Acta Orthop 87:36–41
Bullen ME, Babazadeh S, van Bavel D, McKenzie DP, Dowsey MM, Choong PF (2022) Reduction in offset is associated with worse functional outcomes following total hip arthroplasty. J Arthroplasty 38(2):329–334
Hoberg M, Konrads C, Huber S, Reppenhagen S, Walcher M, Steinert A, Barthel T, Rudert M (2015) Outcome of a modular head-neck adapter system in revision hip arthroplasty. Arch Orthop Trauma Surg 135(10):1469–1474
Woelfle JV, Fraitzl CR, Reichel H, Wernerus D (2014) Significantly reduced leg length discrepancy and increased femoral offset by application of a head-neck adapter in revision total hip arthroplasty. J Arthroplasty 29:1301–1307
Vaishya R, Sharma M, Chaudhary RR (2013) Bioball universal modular neck adapter as a salvage for failed revision total hip arthroplasty. Indian J Orthop 47(5):519–522
Kock HJ, Cho C, Buhl K, Hillmeier J, Huber FX (2019) Long-term outcome after revision of hip arthroplasty with the BioBall® adapter system in multimorbid patients. J Orthop Translat 20(22):43–49
Tokuhara Y, Kadoya Y, Kim M, Shoundou M, Kanno T, Masuda T (2011) Anterior knee pain after total hip arthroplasty in developmental dysplasia. J Arthroplasty 26(6):955–960
Funding
Open Access funding provided thanks to the CRUE-CSIC agreement with Springer Nature.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation and data collection were performed by Ernesto Muñoz-Mahamud, Clara Chimeno, Alfonso Alías, and Adrià Serra. Statistical analysis and figures were performed by Yury Postnikov. The first draft of the manuscript was written by Ernesto Muñoz-Mahamud, Jenaro A. Fernández-Valencia, and Andrés Combalia, and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Ethics Committee of Hospital Clínic of Barcelona and University of Barcelona (register number: HCB/2023/0067).
Consent to participate
According to Institutional Review Board, the informed consent is not required since all information is anonymized and the submission does not include images that may identify the person.
Competing interests
The authors declare no competing interests.
Additional information
The work submitted has not been published before. It is not under consideration for publication anywhere else. Its publication has been approved by all co-authors, if any, as well as by the responsible authorities—tacitly or explicitly—at the institute where the work has been carried out. The publisher will not be held legally responsible should there be any claims for compensation.
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chimeno, C., Fernández-Valencia, J.Á., Alías, A. et al. Contribution of the BioballTM head–neck adapter to the restoration of femoral offset in hip revision arthroplasty with retention of a well-fixed cup and stem. International Orthopaedics (SICOT) 47, 2245–2251 (2023). https://doi.org/10.1007/s00264-023-05833-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-023-05833-7