Introduction

Positron emission tomography/computed tomography (PET/CT) has been established for tumor imaging and alters staging and treatment management in nearly 40% of the patients [1, 2]. Increased glycolysis and metabolism rate has been described in different tumor types allowing to use [18F]Fluorodeoxyglucose ([18F]FDG) as the main oncologic PET tracer [3]. [18F]FDG is transported by glucose transport proteins (GLUT) together with glucose in cells and after phosphorylation by the hexokinase trapped as [18F]FDG-6-phosphate [4]. Different GLUT subtypes with high glucose affinity (class I including GLUTs 1–4) have been identified. Their differing Km (Michaelis constant, substrate concentration at half the maximum velocity) partly predicts their different functions [5]. GLUT 1 and 3, which are overexpressed in tumor and brain tissue, have low Km (1–2 mmol/l) resulting in low Vmax, whereas GLUT 2 and 4 which are expressed in the liver and muscles have high Km (GLUT 2, ~ 20 mmol/l; GLUT 4, 5 mmol/l) resulting in high Vmax [6].

In conditions with elevated fasting blood glucose (BG) (> 8.0 mmol/l), it is assumed that GLUTs, especially with a low Km (e.g., GLUT 1 in tumor cells and GLUT 3 in brain tissue), are saturated. Consecutively, hyperglycaemia can result in reduced [18F]FDG uptake in the tumor and brain as it competes with the endogenous glucose excess. Furthermore, sink effect can occur if the bio-distribution of [18F]FDG is altered [7].

Eskian et al. reported that the liver as the key organ for glucose regulation showed increased [18F]FDG uptake in hyperglycaemia [3]. Furthermore, brain uptake was significantly reduced in patients with elevated blood glucose. Moreover, GLUTs can be insulin-dependent (e.g., GLUT 4 in skeletal muscle), which may result in reduced [18F]FDG uptake in patients with diabetes due to an insulin resistance of the cells [3].

Many patients scheduled for an [18F]FDG PET/CT examination present with elevated BG levels due to an increasing incidence of pre-diabetic and diabetic disorders [8]. Therefore, the Societies of Nuclear Medicine in the USA (SNMMI) and in Europe (EANM) suggest to either delay the PET/CT examination or inject short acting insulin or hydrate the patient in cases of elevated BG (8.3–11.1 mmol/l and > 11.1 mmol/l) or reschedule the scan [9, 10]. An upper blood glucose limit of 8.9 mmol/l for [18F]FDG brain imaging was defined by Varrone et al. [11]. To date, there is uncertainty about the effect of the elevated BG on tumor uptake and background activity [12]. While some authors showed that tissue uptake (e.g., lung, bone marrow, spleen, fat, bowel and stomach) was not affected by high plasma glucose levels prior to [18F]FDG injection, whereas muscle uptake was slightly increased in hyperglycaemia [13]. Other works reported that tumor uptake is not significantly impaired by elevated blood glucose levels [3, 13,14,15,16].

In the last decade, digital detector technology and enhanced time-of-flight capabilities have resulted in significant improvements in PET sensitivity and resolution [17]. Further advancements were achieved by the recent introduction of long-axial field-of-view (LAFOV) PET-scanners with axial lengths over 100 cm [17,18,19,20]. With increasing length of the scanner, more lines of response (LOR) emitted from the patient hit the detectors [21], resulting in noise reduction and an improvement in sensitivity which may improve lesion conspicuity [19, 22,23,24]. These advantages might be clinically useful when image quality is expected to be impaired, as in hyperglycaemic patients receiving [18F]FDG.

In this study, we aim to evaluate the improvement of image quality when using a LAFOV PET system in patients with elevated blood glucose and aim to investigate whether this allows to compensate to some degree the altered distribution of [18F]FDG in elevated blood glucose situations.

Materials and methods

Patient population, radiopharmaceutical, and imaging

Nine hundred thirty-three oncologic patients undergoing clinical routine [18F]FDG-PET/CT on a LAFOV PET scanner were retrospectively evaluated. As per clinical routine, patients fasted for at least 6 h prior to the [18F]FDG-PET/CT. Blood glucose levels were confirmed by blood glucose sampling. In total, 3.0 ± 0.19 MBq/kg [18F]FDG was injected intravenously. All patients received skull base to mid thighs PET scans on the Biograph Vision Quadra (Siemens Healthineers) LAFOV PET/CT system. A 360 s list-mode (LM) acquisition was utilized for all scans.

Patient characteristics and matched pair recruitment

Thirty-one of 933 consecutively scanned patients showed elevated blood glucose levels (> 8.0 mmol/l) prior to the [18F]FDG injection. Patients were ask to refrain antidiabetic medication at least 72 h prior to the PET/CT. A total of 5/31 patients (16%) had blood glucose levels higher than 11 mmol/l, while 26/31 (84%) patients had blood glucose levels of 8–11 mmol/l. Patients with normal blood glucose levels (3.8–8.0 mmol/l) were matched to the group of high blood glucose patients with regard to baseline tumor entity, age, gender, and body mass index (BMI). Figure 1 shows the study flow-chart. Patients’ characteristics showing the different tumor features, which were included in the analysis, are outlined in Table 1.

Fig. 1
figure 1

Study flowchart showing patient recruitment and distribution of blood glucose elevation

Table 1 High blood glucose (hBG) and normal blood glucose (nBG) patients were matched with regard to the tumor entity (n = 31 per group). Patient cohorts include melanoma, lung cancer, lymphoma, breast cancer, genitourinary cancer (GU), head and neck cancer (HNC), and gastrointestinal cancer (GI). Patients’ characteristics (mean ± SD) as age (years), body mass index (BMI), normalized activities, gender, and blood glucose levels (mmol/l) are given

Imaging protocol

The PET LM data were adapted using 360 s, 240 s, 120 s, and 60 s durations to simulate shorter acquisition times. The images were reconstructed using the ultra-high sensitivity mode (UHS, maximum ring difference MRD 322). To simulate the statistics and the image quality of a short-axial field-of-view PET-system (sSAFOV), the LM data were further reconstructed with 120 s and 60 s duration in the high sensitivity mode (HS, MRD 85) as previously described [24, 25]. The choice of sSAFOV was based on the coincident yield and sensitivity profile of the Biograph Vision Quadra to ensure reliable comparison between a SAFOV and LAFOV scanner [25, 26]. These reconstructions were used to compare tumor uptake, tumor-to-background ratio (TBR), and image quality between a LAFOV and a sSAFOV PET in hBG and nBG patients [22].

Image reconstruction was performed using the vendor’s standard reconstruction software (e7-tools, Siemens Healthineers), with OSEM (4 iterations, 5 subsets), point-spread-function (PSF), and TOF enabled. Images were reconstructed with a 440 × 440 × 644 image matrix with a voxel size of 1.65 × 1.65 × 1.65 mm3. A Gaussian post-reconstruction filter with 2-mm full width at half maximum (FWHM) was applied to the images. For attenuation correction, low-dose non-enhanced CT data were used. 3D scatter correction was performed using a 3D residual-based method which was especially developed for the MRD 322 [27] both for MRD 85 and 322 images [28].

CT scans were performed with equivalent parameters with slice thickness of 1.0 mm, pitch factor 1, bone and soft tissue reconstruction kernels, and maximum of 120 kV and 90 mAs by applying CARE kV and CARE Dose as previously published [22].

Image evaluation

Three nuclear medicine physicians (CM, LW, and TP) performed image evaluation. For image evaluation, appropriate workstation [29] and software for quantitative image analysis and identification of target lesion were used (Syngo.via, Siemens Healthineers, Erlangen, Germany). Tumor-lesion uptake and metabolic tumor volumes were calculated by placing a volume-of-interest (VOI) with a 40%-iso-contour approach around the lesion as previously described [22, 30].

Peak and maximum standardized uptake values (SUVpeak/max) were used to evaluate target lesion knowing that SUVpeak has been shown to be less susceptible for variation at different acquisition duration compared to SUVmax [31]. Lesions were evaluated at different scan times as LM data allowed different reconstructions (HS, 60 s, 120 s; UHS, 60 s, 120 s, 240 s, 360 s) [22, 32].

The background activity was measured by placing a 10-cm3 spherical VOI in the right lobe of healthy liver tissue as previously described [33]. The VOIs were copied to same anatomical locations in all different images obtained from different (list-mode) frame durations within the same patient to ensure that mistakes can be minimized. The coefficient of variation (COV) for the liver background was defined by the following formula where σ was the standard deviation of the liver VOI and μ was the SUVmean of the liver VOI [22]. TBR was defined as tumor uptake (SUVmax) per liver uptake (SUVmean). Contrast-to-noise ratio (CNR) was calculated as previously published [34].

$$COV\;=\;(\frac{\sigma\;}{\mu\;});\;TBR=\frac{SUVmax\;(tumor)}{SUV\;mean\;(liver)};\;CNR\;=\;\frac{SUVmean\;\left(lesion\right)\;-\;SUVmean\;(background)}{SD\;(background)}$$

Statistical analysis

Statistical analyses were performed using Excel (Microsoft, Redmond, Washington) and GraphPad Prism Version 8 (San Diego, CA) [35]. The data are presented either as mean ± standard deviation or as median and range. Qualitative comparison between hBG and nBG patients was analyzed using Fisher’s exact test, and for quantitative measurements (COV, CNR, SUV), were analyzed using unpaired Student’s T-test after verification of normal distribution. Comparison between different reconstructions in the same patient were analyzed using paired Student’s T-test and Bonferroni-correction was applied. p-values < 0.05, in cased of Bonferroni-corrections p < 0.001, were considered statistically significant.

Results

Overall, 78 nBG and 82 hBG tumor lesions were identified. The metabolic tumor volume (MTV) and total lesion glycolysis (TLG) did not differ significantly between both patient groups (median, range; hBG MTV, 0.94 ml, 0.01–7.45 ml vs. nBG, 1.23 ml, 0.12–18.48 ml; p = 0.06 and hBG TLG, 7.02 ml, 0.6–749.75 ml vs. nBG, 4.43 ml, 0.53–181.61 ml; p = 0.24). In the hBG patients, 70 lesions (85%) were detected in the subgroup of patients with BG 8–11 mmol/l and 12 (15%) in the subgroup of patients with BG > 11 mmol/l.

Background activity (liver and muscle uptake)

Liver uptake in hBG patients was significantly higher compared to nBG patients (SUVmean, 3.07 ± 0.41 vs. 2.37 ± 0.33; p = 0.03) whereas the brain (gray matter) uptake was significantly lower in hBG patients compared to nBG patients (SUVmax, 7.58 ± 0.74 vs. 13.38 ± 3.94; p < 0.001 and SUVpeak, 5.70 ± 0.25 vs. 9.91 ± 3.46; p < 0.001) (Fig. 2A, B).

Fig. 2
figure 2

Background uptake visualized with box plots in patients with high (hBG) and normal blood glucose (nBG). Shown are standardized uptake value (SUVmean) of liver uptake (A), SUVmax of brain (gray matter, B), and muscle uptake (shoulder, C; gluteal, D). Statistically significant differences are highlighted with an asterisk “*”; ns, not statistically significant

Muscle uptake was not statistically different between the hBG and nBG cohort neither in shoulder nor in gluteal muscle (shoulder SUVmax, 1.63 ± 0.47 vs. 2.09 ± 0.26; p = 0.91 and gluteal SUVmax, 1.47 ± 0.18 vs. 1.45 ± 0.12; p = 0.33) (Fig. 2C, D).

Tumor uptake in hBG and nBG patients

In all reconstructions (HS 60 s, HS 120 s, UHS 60 s, UHS 120 s, UHS 240 s, and UHS 360 s), hBG patients had slightly, but still significantly lower tumor uptake compared to the matched-pair equivalent with nBG (SUVmax hBG, 11.05 ± 2.29 vs. nBG, 12.83 ± 5.35; p < 0.001; SUVpeak hBG, 6.36 ± 2.16 vs. nBG, 8.85 ± 5.03; p < 0.001). TBR between hBG and nBG reconstructions was consecutively significantly lower in hBG patients (TBR hBG, 3.48 ± 0.74 vs. nBG, 5.29 ± 1.48; p < 0.001) (Fig. 3).

Fig. 3
figure 3

Tumor uptake visualized with violin plots in patients with high (hBG) and normal blood glucose (nBG) (AC). Shown are standardized uptake values (SUVmax/peak) and tumor to background ratios (TBR). Statistically significant differences are highlighted with an asterisk “*”

The results were confirmed in the subgroup analysis. Tumor uptake was lower in both subgroups compared to the nBG patients. Consecutively, TBR was significantly lower in both subgroups compared to nBG patients (TBR BG, 8–11 mmol/l, 3.88 ± 3.24 vs. nBG, 4.73 ± 3.87; p < 0.0001 and BG > 11 mmol/l, 4.11 ± 0.89 vs. nBG, 5.25 ± 1.42; p < 0.0001). Detailed visualization of the subgroup results can be found in the Supplementary Material.

Background noise (COV) and image quality (CNR)

In order to compare background noise, liver COV in the hBG and nBG group was determined. We found no statistical difference between the COV of each reconstruction comparing hBG vs. nBG patients (COV hBG, 0.09 ± 0.39 vs. nBG, 0.10 ± 1.23; p = 0.10) (Fig. 4). COV decreased with acquisition time as with lowest COV after 360 s acquisition in UHS mode (hBG, 0.07 ± 0.37 vs. nBG, 0.07 ± 0.86; p = 0.16). All obtained reconstructions showed comparable COV in situations of high and normal blood glucose.

Fig. 4
figure 4

Background noise of the liver visualized with violin plots in patients with high (hBG) and normal blood glucose (nBG). Shown is the coefficient of variation (COV) in liver tissue. Brackets summarize the comparison between each reconstruction in hBG and nBG and indicate that differences were not statistically significant (ns) between all corresponding reconstructions

With regard to CNR, we noted that CNR behaved similar to the tumor uptake and was significantly higher in nBG compared to hBG comparing corresponding reconstructions (CNR hBG, 12.17 ± 4.86 vs. nBG, 23.31 ± 12.22; p < 0.001) (Fig. 5A). However, CNR in nBG patients receiving sSAFOV (HS 120 s, Fig. 5A (green)) and hBG patients receiving LAFOV (UHS 360 s, Fig. 5B (red)) was similar (sSAFOV nBG, 17.79 ± 9.3 vs. LAFOV hBG, 19.84 ± 8.4; p = 0.08).

Fig. 5
figure 5

Image quality of all reconstructions between high (hBG) and normal blood glucose patients (nBG). Visualized are contrast to noise ratios (CNR) in all reconstructions (A). B Chosen reconstructions for comparison UHS 360 s for LAFOV PET and HS 120 s for sSAFOV PET. Statistically significant differences are highlighted with an asterisk “*”; ns, not statistically significant. Brackets summarize the comparison between each reconstruction in hBG and nBG and indicate that differences were significant between all corresponding reconstructions

Subgroup analysis with regard to tumor uptake and image quality

Subgroup analysis of patients with hBG 8–11 mmol/l revealed no statistical difference in comparison of hBG LAFOV (UHS 360 s) vs. nBG sSAFOV (HS 120 s) reconstructions neither in TBR nor in CNR (TBR hBG LAFOV, 3.70 ± 3.12 vs. nBG sSAFOV, 4.86 ± 3.95; p = 0.06 and CNR hBG LAFOV, 20.59 ± 31.92 vs. nBG sSAFOV, 13.10 ± 16.34; p = 0.09). This finding was consistent with the overall analysis, which did not reveal statistical differences between the CNR of hBG LAFOV and nBG sSAFOV scans (p = 0.08) (Fig. 6).

Fig. 6
figure 6

Tumor uptake und image quality in comparison of long-axial field-of-view (LAFOV) high blood glucose (hBG, A) and simulated short-axial field-of-view (sSAFOV) normal blood glucose (nBG, B) patients. Box plots (C, D) show the subgroup with BG 8–11 mmol/l and (E, F) BG > 11 mmol/l. Statistically significant differences are highlighted with an asterisk “*”; ns, not statistically significant

In patients with hBG > 11 mmol/l, we found different results. The TBR was significantly lower in hBG LAFOV compared to nBG sSAFOV (TBR hBG LAFOV, 3.8 ± 0.9 vs. nBG sSAFOV, 5.32 ± 1.41; p = 0.03), whereas the CNR was not different in the hBG 8–11 mmol/l group between hBG LAFOV compared to nBG sSAFOV (CNR hBG LAFOV, 20.6 ± 8.4 vs. nBG sSAFOV, 17.10 ± 9.15; p = 0.46) (Fig. 6). An example image of hBG patients shows that CNR is significantly higher in LAFOV compared to sSAFOV (Fig. 7).

Fig. 7
figure 7

Example SUV image of a 46 year/o female patient referred to our PET/CT center for restaging of endometrial cancer. Prior to the [18F]FDG scan, elevated blood glucose was elevated (BG, 9.4 mmol/l). In this study, a new liver metastasis was detected (marked with a red arrow). Shown are maximum intensity projections (A, B) and axial images (C, D) of long-axial field-of-view (LAFOV, A, C) and simulated short-axial field-of view (sSAFOV (B, D)) PET-reconstructions. Of note, contrast-to-noise ratio (CNR) of the liver lesion was significantly higher in the LAFOV compared to sSAFOV images. Scale bars indicate the SUV window used for this patient

Discussion

The rising prevalence of pre-diabetic and diabetic patients results in an increasing portion of patients with elevated blood glucose levels despite fasting 6 h prior to the [18F]FDG injection [8]. As described above, hBG leads to altered [18F]FDG uptake in both healthy tissue and neoplasms, resulting in a generally reduced image quality. The recently introduced LAFOV PET/CT scanners enable performing scans with higher sensitivity and lower noise as well as higher lesion conspicuity [22, 24, 36]. We present the first matched pair analysis including patients with hBG and nBG comparing the background, liver, tumor uptake, and image quality on LAFOV and sSAFOV PET/CT.

Our results are in keeping with previous reports on [18F]FDG bio-distribution. We found that liver uptake (SUVmean) was significantly higher in hBG patients compared to nBG patients. That was consistent with the findings of Lindholm et al. who report a significant correlation between increased BG levels and liver uptake using SUVmean (correlation coefficient, 0.12; p < 0.001) [13]. Furthermore, Büsing et al. report a negative correlation between patients with hBG and nBG in brain uptake (Pearson coefficient, − 0.44; p < 0.01); this is congruous with our findings as we found significantly decreased brain uptake in the cohort of hBG patients. This effect of elevated BG in [18F]FDG PET/CT resulted in the recommendation of an upper BG limit of 8.9 mmol/l for [18F]FDG brain imaging (e.g., dementia, epilepsy) [14, 15].

Our cohort showed no significant difference in muscle uptake (neither deltoid nor gluteal) in hBG and nBG patients. Thus, we found a tendency toward increased uptake in the gluteal muscle in hBG patients, but not statistically significant (SUVmax, 1.47 ± 0.18 vs. 1.45 ± 0.12; p = 0.33). Previous publications report discrepant results. Lindholm et al. and Büsing et al. showed a weak positive correlation between BG levels and muscular [18F]FDG uptake, whereas Eskin et al. report different findings. Muscular SUVmax was inversely correlated with increasing BG and SUVmean showed no significant correlation to serum BG [3]. Accordingly, Webb et al. found no significant difference in muscle uptake between patients with nBG and hBG [37]. As possible confounders, gender, age, BMI, diabetes, [18F]FDG dose, and scan time were identified. We tried to avoid those inconsistencies using a matched pair study design and matching with regard to tumor entity, gender, age, and BMI. Still, we noticed that patients with hBG had higher BMI in our cohort compared with nBG patients (28.22 ± 5.35 kg/m2 vs. 26.25 ± 4.53 kg/m2) [38].

Our cohort of hBG patients showed significantly lower tumor uptake compared to the nBG cohort over all reconstructions and consecutively significantly lower TBR. Subgroup findings were consistent with overall findings. These findings are perfectly in line with the literature. The majority of publications show that tumor [18F]FDG uptake decreases with increasing serum BG [3, 14, 37]. Moreover, significant TBR reduction is the consequence of decreased tumor and increased liver uptake. That is the reason why SNMMI and EANM guidelines suggest to reschedule [18F]FDG PET/CT in cases of elevated serum BG [9, 10].

Rescheduling the PET/CT examination leads to increased cost for the healthcare system, inconvenience for the patients, and can result in an unnecessary delay of the tumor staging which may affect the therapy of the patient [3]. Therefore, it is important, especially for oncologic patients to avoid unnecessary delay of the PET/CT. We tried to address this problem by analyzing the image quality of [18F]FDG on a LAFOV compared to a (simulated) SAFOV scanner in cases of high blood glucose. sSAFOV scan parameters were determined by the performance characteristics of the Biograph Vision Quadra PET-system. Prenosil et al. reported by evaluating the performance characteristics of the Biograph Vision Quadra according to the National Electrical Manufacturers Association NU 2–2018 (NEMA) standard increased sensitivity and noise equivalent count rates (NECR), which were by 5 and 10 times higher compared to a SAFOV system (Biograph Vision 600) in MRD 85 and MRD 322 mode, respectively [25]. Further publication with patient data showed that a scan with an acquisition time of 120 s in MRD 85 was equivalent comparing to a the mentioned SAFOV system [22]. Moreover, further increase in sensitivity and NECRs was reported with the use of the LAFOV PET/CT’s full acceptance angle (MRD 322) resulting in the possibility to reduce acquisition times and injected doses even further compared to the SAFOV system [24].

In the present work, we showed that the tumor uptake and TBR decrease with increasing BG on a LAFOV scanner, which is consistent to the published data. Accordingly, overall CNR was significantly lower in hBG compared to nBG patients. However, background noise was not significantly different between hBG and nBG patients. Moreover, and most importantly, hBG patients receiving LAFOV PET/CT showed comparable image quality (p = 0.08) as measured by CNR as that of matched nBG patients receiving sSAFOV reconstructions (Fig. 6).

Subgroup analysis revealed that tumor uptake differed only significantly in LAFOV compared to sSAFOV in patients with BG levels > 11 mmol/l. In patients with BG 8.0–11.0 mmol/l, no statistical significant difference was found between LAFOV and sSAFOV PET. Our data regarding the tumor uptake are in agreement with so far published findings and guideline recommendation. Eskian et al. found that only the group with BG > 11.1 mmol/l (Eq. 200 mg/dl) had significantly lower SUV compared to euglycemic patients and EANM guideline states that [18F]FDG PET/CT can be performed for clinical studies if plasma glucose is lower than 11 mmol/l [3, 9]. Furthermore, Boellaard et al. acknowledge that elevated BG can occur in situations of poorly controlled diabetes or can be associated with infection. Therefore, they argue that hyperglycaemia should not represent an absolute contraindication [9].

In our cohort, LAFOV PET/CT was able to obtain PET-image quality in situations with elevated (BG, 8–11 mmol/l) or highly elevated (> 11 mmol/l) BG compared to nBG patients scanned with SAFOV PET/CT (Fig. 6). An example of a subtle liver metastasis in a female patient with hBG shows how the image quality (CNR) improves significantly in LAFOV PET/CT compared to sSAFOV PET/CT (Fig. 7). We therefore argue that higher sensitivity and lower background noise in LAFOV PET/CT can compensate to some degree the altered bio-distribution of [18F]FDG in oncologic patients with elevated blood glucose. Nevertheless, SUV measurements in cases of elevated BG should be used carefully when comparison with follow-up PET-imaging is required. Patient reports should particularly mention this limitation to draw clinicians’ attention to the hBG PET-examination.

Furthermore, we note several limitations of our study. First, the matched-pair study design introduced some selection bias. We tried to address all necessary oncological features of each patient to adjust the cohorts as best as possible. Moreover, we matched each patient according to age, gender, and BMI. A prospective, preferably head-to-head study design would have avoided this bias. However, due to the retrospective nature of the study, a head-to-head evaluation was not possible, which would have required two independent scans on two different days leading to further delay of the oncological treatment. Secondly, we could not compare LAFOV with SAFOV PET/CT directly. Instead, we used reconstructions simulating SAFOV PET/CT [22, 24]. However, using sSAFOV enabled to perform an intra-individual comparison of all reconstructions and avoided further bias. Lastly, the matched-pair cohort was limited to a relatively small number of patients, which resulted from only 3.32% of investigated patients meeting the inclusion criteria. Many of our PET/CT examinations were delayed due to high BG levels. Nevertheless, possible non-[18F]FDG-avid lesions might have been missed, especially in situations of hBG patients.

With the limitations in mind, our data suggest that the higher sensitivity of LAFOV may compensate for the reduced CNR in cases of elevated blood glucose, and that in borderline clinical circumstances, rescheduling the study or maneuvres to reduced blood glucose levels may not be necessary.

Conclusion

In a matched-pair analysis of patients with elevated blood glucose levels (> 8 mmol/l) and patients with normal blood glucose prior to examination with [18F]FDG PET/CT on a long axial field-of-view scanner, we found decreasing brain uptake, increasing background activity (e.g., liver uptake), and decreasing tumor uptake. However, we showed that LAFOV PET/CT could compensate to some degree the altered bio-distribution in hBG patients using a higher sensitivity profile and a lower noise rate compared to SAFOV PET/CT.

We, therefore, argue that using a LAFOV PET/CT might be advantageous in patients with high BG levels when decisions to delay imaging would lead to increased costs, inconvenience for the patient, and delay in oncological treatment. In such cases, especially if BG is ≥ 11 mmol/l, we recommend performing [18F]FDG LAFOV PET/CT imaging with longer acquisition duration whenever possible.